LOGIQ™ 100 PRO Service Manual - 5305880 - 100 - r3 PDF
LOGIQ™ 100 PRO Service Manual - 5305880 - 100 - r3 PDF
LOGIQ™ 100 PRO Service Manual - 5305880 - 100 - r3 PDF
Technical
Publication
Direction 5305880-100
Revision 3
GE Healthcare
LOGIQ 100 PRO Service Manual
ApprovedDocument-53081TPH_r3.pdfPage2of197
GE HEALTHCARE
DIRECTION 5305880-100 , REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Your Notes:
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
Page i-ii -
ApprovedDocument-53081TPH_r3.pdfPage3of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
IMPORTANT PRECAUTIONS
LANGUAGE
- Page i-i
ApprovedDocument-53081TPH_r3.pdfPage4of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Page i-ii -
ApprovedDocument-53081TPH_r3.pdfPage5of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
- Page i-iii
ApprovedDocument-53081TPH_r3.pdfPage6of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write Damage
In Shipment on ALL copies of the freight or express bill BEFORE delivery is accepted or signed
for by a GE representative or hospital receiving agent. Whether noted or concealed, damage
MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation
company will not pay a claim for damage if an inspection is not requested within this 14 day period.
Page i-iv -
ApprovedDocument-53081TPH_r3.pdfPage7of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part,
without prior written permission of GE Medical Systems.
GE Medical Systems may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
COPYRIGHTS
All Material Copyright 1997-2007 by General Electric Inc. All Rights Reserved.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
- Page i-v
ApprovedDocument-53081TPH_r3.pdfPage8of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
Page i-vi -
ApprovedDocument-53081TPH_r3.pdfPage9of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Revision History
2 APRIL 24 2008 Changes incorporated for correcting the SPRs raised against DOC0421436.
Changes incorporated on contact details and added additional information based on service
3 August 4 2008
teams requirements.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
- Page i-v
ApprovedDocument-53081TPH_r3.pdfPage10of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
Page i-vi -
ApprovedDocument-53081TPH_r3.pdfPage1of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Table Of Contents
Chapter 1 Introduction
Section 1-1Overview ....................................................................................... 1-1
1-1-1 Overview ........................................................................................................................ 1-1
1-1-2 Purpose of Chapter 1..................................................................................................... 1-1
1-1-3 Purpose of Service Manual............................................................................................ 1-1
1-1-4 Typical Users of the Basic Service Manual ............................................................... 1-2
1-1-5 LOGIQ 100 PRO Models Covered by this Manual .................................................... 1-2
1-1-6 Purpose of Operator Manual(s) ..................................................................................... 1-2
Section 1-2 Safety .......................................................................................... 1-3
1-2-1 Warnings........................................................................................................................ 1-3
Section 1-3 Important Conventions.............................................................. 1-4
1-3-1 Conventions Used in Book ............................................................................................ 1-4
1-3-2 Standard Hazard Icons .................................................................................................. 1-5
1-3-3 Product Icons ................................................................................................................. 1-6
Section 1-4Safety Considerations .............................................................. 1-10
1-4-1 Introduction .................................................................................................................. 1-10
1-4-2 Human Safety .............................................................................................................. 1-10
1-4-3 Mechanical Safety ...................................................................................................... 1-11
1-4-4 Electrical Safety ........................................................................................................... 1-11
1-4-5 Labels Locations.......................................................................................................... 1-12
1-4-6 Dangerous Procedure Warnings ................................................................................. 1-18
1-4-7 Lockout/Tagout Requirements (For USA Only) ........................................................... 1-18
1-4-8 Classification................................................................................................................ 1-18
Section 1-5 EMC, EMI, and ESD ............................................................... 1-19
1-5-1 Electromagnetic Compatibility (EMC) .......................................................................... 1-19
1-5-2 CE Compliance ............................................................................................................ 1-19
1-5-3 EMC Performance ....................................................................................................... 1-19
1-5-4 Notice Upon Installation of Product ............................................................................. 1-19
1-5-5 General Notice ............................................................................................................. 1-20
1-5-6 Countermeasures against EMC related Issues ........................................................... 1-20
1-5-7 Notice on Service......................................................................................................... 1-20
1-5-8 Electrostatic Discharge (ESD) Prevention ................................................................... 1-21
Section 1-6Customer Assistance ................................................................ 1-21
1-6-1 System Manufacturer................................................................................................... 1-21
1-6-2 Contact Information ..................................................................................................... 1-22
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
ApprovedDocument-53081TPH_r3.pdfPage12of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Lighting 2
2-2-2 Electrical Requirements ................................................................................................ 2-2
2-2-2-1 LOGIQ 100 PRO Power Requirements ......................................................... 2-2
2-2-2-2 Site Circuit Breaker ........................................................................................... 2-2
2-2-2-3 Site Power Outlets............................................................................................. 2-2
2-2-2-4 Unit Power Plug................................................................................................. 2-2
2-2-2-5 Power Stability Requirements ........................................................................... 2-2
2-2-3 EMI Limitations 2-3
2-2-4 Probes Environmental Requirements ............................................................................ 2-4
2-2-5 Time and Manpower Requirements ............................................................................... 2-5
2-2-6 Checklist for Installation ................................................................................................. 2-5
2-2-6-1 Checklist for Installation .................................................................................... 2-5
Chapter 3 Installation
Section 3-1Overview........................................................................................ 3-1
3-1-1 Purpose of Chapter 3 ..................................................................................................... 3-1
3-1-2 Average Installation Time............................................................................................... 3-1
3-1-3 Safety Reminders 3-2
Section 3-2 Receiving and Unpacking the Equipment................................ 3-3
Section 3-3 Preparing for Installation ........................................................... 3-5
3-3-1 Physical Inspection ........................................................................................................ 3-5
3-3-1-1 Systems Voltage Settings ................................................................................. 3-5
3-3-1-2 Video Formats ................................................................................................... 3-5
3-3-2 EMI Protection................................................................................................................ 3-5
Section 3-4 Assembling LOGIQ 100 PRO .................................................. 3-5
3-4-1 Installation of the System............................................................................................ 3-5
Section 3-5 Completing the Installation ....................................................... 3-6
3-5-1 Transducer Connection .................................................................................................. 3-6
3-5-2 Installation of Optional Accessories ............................................................................... 3-7
3-5-3 Installation of Two Probe Adapter & LOGIQ a200 Probe Adapter (Optional) 3-9
3-5-4 Installation of LOGIQ 100 PRO with Global or Modular Trolley 3-11
3-5-5 Adjustment of Monitor Brightness & Contrast 3-17
3-5-6 Adjustment of System Clock/Date, Hospital Name & OB Version ............................... 3-17
Section 3-6 System Configuration ............................................................... 3-17
3-6-1 System Specifications .................................................................................................. 3-17
3-6-1-1 Physical Dimensions ....................................................................................... 3-17
Front View 18
3-6-1-3 Side View (Keyboard Closed) ......................................................................... 3-19
3-6-1-4 Side View (Keyboard Open)............................................................................ 3-19
3-6-1-5 Weight without Peripherals.............................................................................. 3-20
3-6-2 Electrical Specifications ............................................................................................... 3-20
3-6-2-1 Power Supply .................................................................................................. 3-21
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
Page TOC-2
ApprovedDocument-53081TPH_r3.pdfPage13of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage14of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting
Section 7-1Overview........................................................................................ 7-1
7-1-1 Purpose of Chapter 7 ..................................................................................................... 7-1
Section 7-2 Power On Diagnostics ................................................................ 7-3
7-2-1 DAVINCI Board Tests 7-4
7-2-2 FEB Test for LOGIQ 100 PRO ........... ....................................................................... 7-7
7-2-3 Keyboard Tests 7-9
7-2-4 Monitor Test 7-10
7-2-5 Key Operation History .................................................................................................. 7-13
Section 7-3 Error Reporting......................................................................... 7-14
7-3-1 Error Reporting............................................................................................................. 7-14
Section 7-4 Troubleshooting ........................................................................ 7-15
Page TOC-4
ApprovedDocument-53081TPH_r3.pdfPage15of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage16of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
10-6-6 Probe Leakage Current Test 10-19
10-6-6-1Definition....................................................................................................... 10-19
10-6-6-2Generic Procedure for Leakage current ....................................................... 10-19
10-6-6-3Meter Procedure Using Dale Meter to measure leakage current ................. 10-19
Meter Procedure using Probe Adapter to measure Probe isolation(Sink) Current 21
Section 10-7 When There's Too Much Leakage Current......................... 10-23
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
Page TOC-6
ApprovedDocument-53081TPH_r3.pdfPage17of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Chapter 1 Introduction
Section 1-1Overview
1-1-1 Overview
The LOGIQ 100 PRO is a compact portable ultrasound scanner which is designed for OB/GYN,
Abdomen, Urology and small part Scans using the convex, linear and micro-convex (Transvaginal)
probes. High quality images can be obtained by the proper selection of scan control parameters.
The diagnostic clarity is further enhanced by the different measurement and calculation packages
available in the system.
This chapter describes important issues related to safely servicing this ultrasound machine. The
service provider must read and understand all the information presented here before installing or
servicing a unit.
CONTENTS IN CHAPTER 1
This Service Manual provides installation and service information for the LOGIQ 100 PRO
Ultrasound Scanning System and contains the following chapters:
1.) Chapter 1, Introduction: Contains a content summary and warnings.
2.) Chapter 2, Pre-Installation: Contains any pre-installation requirements for the
LOGIQ 100 PRO.
3.) Chapter 3, Installation: Contains installation procedure with installation checklist.
4.) Chapter 4, Functional Checks: Contains functional checks that must be performed as
part of the installation, or as required during servicing and periodic maintenance.
5.) Chapter 5, Theory: Contains block diagrams and functional explanations of the
electronics.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
6.) Chapter 6, Service Adjustments: Contains instructions on how to make any available
adjustments to the LOGIQ 100 PRO.
7.) Chapter 7, Diagnostics/Troubleshooting: Provides procedures for running and
diagnostic or related routines for the LOGIQ 100 PRO.
8.) Chapter 8, Replacement Procedures: Provides disassembly procedures and
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage18of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
9.) Chapter 9, Replacement Parts: Contains a complete list of replacement parts for the
LOGIQ 100 PRO.
10.)Chapter 10, Quality Assurance: Provides periodic maintenance procedures for the
LOGIQ 100 PRO.
The Operator Manual(s) should be fully read and understood before operating the LOGIQ 100
PRO and also kept near the unit for quick reference.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage19of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1.) The operator manual should be read and understood before operating the LOGIQ 100
PRO and kept nearby for quick reference.
2.) Although the ultrasound energy transmitted from the LOGIQ 100 PRO transducer is
within AIUM/NEMA standards, unnecessary exposure should be avoided. Only trained
personnel should operate the LOGIQ 100 PRO.
3.) To prevent electric shock, the LOGIQ 100 PRO should be connected to a properly
grounded power receptacle.Do not use a three prong to two prong adapter. This defeats
safety grounding/
4.) Probes are fragile, Please handle with care.
5.) Concerning outside markings, refer to Figure 1-2 through 1-3.
6.) For the cleaning, disinfection, and sterilization, refer to Probe section in LOGIQ 100
PRO User Manual and Caution Sheet supplied with each probe.
NOTICE This medical equipment is approved, in terms of the prevention of radio wave interference,
to be used in hospitals, clinics and other institutions which are environmentally qualified. The use
of this equipment is an inappropriate environment may cause some electronic interference to radios
and televisions around the eqiupment. Proper handling of this equipment is required in order to
avoid such trouble according to the operator and service manuals. This eqiupments can be used in
residential areas only under the supervision of physicians and qualified technicians.
CAUTION Improper performance possibility. Do not use the following devices near this
equipment. Cellular phone, radio transciever, mobile radio transmitter, radio-controlled toy,
etc. Use of these devices near this equipment could cause this equipment to perform outside
the published specifications. Keep power to these devices turned off when near this
equipment.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage20of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Model Designations.
This manual covers the LOGIQ 100 PRO scanners.
(See LOGIQ 100 PRO Models Covered by this Manual on page 1-2.)
Icons.
Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
Safety Precaution Messages.
Various levels of safety precaution messages may be found on the equipment and in the service
information. The different levels of concern are identified by a flag word that precedes the
precautionary message. Known or potential hazards are labeled in one of three ways:
ApprovedDocument-53081TPH_r3.pdfPage21of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Important information will always be preceded by the exclamation point contained within a triangle,
as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may
be used to make you aware of specific types of hazards that could cause harm.
LASER
LIGHT
Other hazard icons make you aware of specific procedures that should be followed.
AVOID STATIC ELECTRICITY TAG AND LOCK OUT WEAR EYE PROTECTION
ApprovedDocument-53081TPH_r3.pdfPage2of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The following table describes the purpose and location of safety labels and other important
information provided on the equipment.
Identification and Rating Plate Manufacturer's name and Rear of console near power
address inlet
Date of manufacture
Model and serial numbers
Electrical ratings
CAUTION - This unit This precaution is intended to On the console where easily
weighs...Special care must be prevent injury that may result seen during transport?
used to avoid..." if one person attempt to move
the unit considerable
distances or on an incline due
to the weight of the unit.
anesthetic gases.
ApprovedDocument-53081TPH_r3.pdfPage23of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
CAUTION
This Power Switch DOES
NOT ISOLATE Mains
Supply
ApprovedDocument-53081TPH_r3.pdfPage24of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage25of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage26of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The following safety precautions must be observed during all phases of operation, service and
repair of this equipment. Failure to comply with these precautions or with specific warnings
elsewhere in this manual, violates safety standards of design, manufacture and intended use of the
equipment.
Note: This medical equipment is approved, in terms of the prevention of radio wave interference, to be
used in hospitals, clinics and other institutions which are environmentally qualified. The use of this
equip ment in an inappropriate environment may cause some electronic interference to radios and
televi sions around the equipment. Proper handling of this equipment is required in order to avoid
such trouble according to the operator and service manuals.
ApprovedDocument-53081TPH_r3.pdfPage27of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Always lock the Control Console in its parking (locked) position before moving the scanner
CAUTION around.
The LOGIQ 100 PRO weights 10kg or more, depending on installed peripherals, (309 lbs, or
more) when ready for use. Care must be used when moving it or replacing its parts. Failure to
CAUTION follow the precautions listed below could result in injury, uncontrolled motion and costly
damage.
NOTE: Special care should be taken when transporting the unit in a vehicle:
To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The
system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. If an extension cord is used with the system, make sure that the
total current rating of the system does not exceed the extension cord rating.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector meet international electrical standards.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage28of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
GHOST Label
ApprovedDocument-53081TPH_r3.pdfPage29of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage30of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage31of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage32of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
AV
ISO
VO
RS
ICH
T
ApprovedDocument-53081TPH_r3.pdfPage3of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage34of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Warnings, such as the example below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.
Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the plug.
1-4-8 Classification
ApprovedDocument-53081TPH_r3.pdfPage35of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1-5-2 CE Compliance
The LOGIQ 100 PRO unit conforms to all applicable conducted and radiated emission limits
and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and
power line transient requirements.
Applicable standards are: 47CFR Part 18, IEC 60112, and 80613.
For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps,
are in good condition, installed tightly without skew or stress. Proper installation following
all comments noted in this service manual is required in order to achieve full EMC
performance.
All types of electronic equipment may characteristically cause electromagnetic interference with
other equipment, ei ther through air or connecting cables. The term EMC (Electromagnetic
Compatibility) indicates capability of the equip ment, which curbs electromagnetic influence from
other equipment and at the same time does not affect other equip ment with similar electromagnetic
radiation from itself.
This product is designed to fully comply with the EN60601-1-2 (IEC601-1-2), in Medical electrical
equipment EMC regulations.
Proper installation following this service manual is required in order to achieve the full EMC
performance of the product.
The product must be installed as stipulated in 1-4-4, Notice upon Installation of Product.
In case of issues related to EMC, please follow procedures stated in 1-4-6, Countermeasures
against EMC-related Issues.
1.) Use either power supply cords provided by GEMS or ones approved by GEMS. Products
equipped with power source plug should be plugged into the fixed power socket which has the
protective grounding conductor
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
2.) Connect a three-pin plug to a three-pin socket without using a three-pin-to-two-pin converter.
3.) Keep the equipment as far as possible from other electronic equipment.
4.) Be sure to use the only cables provided by GEMS or ones approved by Wipro GE. Install the
unit/peripherals as per the installation procedure provided in Chapter 3 - INSTALLATION. The
SeethGEHCMyworkshpSystemodrinhaufc.
layout of the LOGIQ 100 PRO & other peripherals should be as per installation procedures
described in Chapter 3 - INSTALLATION.
Chapter 1 - Introduction Page 1-19
ApprovedDocument-53081TPH_r3.pdfPage36of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1.) Designation of Peripheral Equipment Connectable to this Product. The peripheral which
conforms to EN60601-1-2 (IEC601-1-2), can be connected up to the LOGIQ 100 PRO
without compromising its EMC performance. Failure to comply with the above instruction may
result in poor EMC performance of the product.
2.) Notice against User Modification: Do not modify this product. Unilateral user modification
may cause degradation in EMC performance. Modification of the product includes:
a.) Changes in cables (length, material, wiring etc.)
b.) Changes in system installation/layout
c.) Changes in system configuration/components
d.) Changes in means of fixing system/parts (cover open/close, cover screwing)
3.) Operate the system with all covers closed. If you open any cover for some reason, be sure to
shut it before starting/ resuming operation. Operating the system with any cover open may
affect EMC performance.
Generally it is very difficult to handle with issues related to EMC. It is time consuming and
expensive. General counter measures for Electromagnetic interference with other equipment
1.) Electromagnetic interference may be reduced by positioning other equipment far away from
the system.
2.) Electromagnetic interference may be reduced by changing the relative location (installation
angle) of the system and other equipment.
3.) Electromagnetic interference may be reduced by changing wiring locations of power/signal
cables of other equip ment.
4.) Electromagnetic influence may be reduced by altering the path of power supply for other
equipment.
Ensure all screws are tight after servicing. Loose screws may cause degradation in EMC
performance.
CAUTION Do not use the following devices near the LOGIQ 100 PRO system. Devices which
intrinsically transmit radio waves such as cellular phone, radio transceiver, mobile radio
transmitter radio-controlled toy, etc. Use of these devices could cause the LOGIQ 100 PRO
system to perform outside the published specifications. Keep power to these devices turned
OFF when near the system. Medical staff in charge of the LOGIQ 100 PRO system is
required to instruct technicians, patients and other people who may be around the system
to fully comply with the above regulation
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage37of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage38of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ASIA(China/India/Korea/SEA):
EUROPE:
GE Ultraschall Deutschland TEL:+49 212 2802 208
GmbH & Co. KG +49 212 2802 207
BeethovenstraBe 239 FAX: (49) 212.28.02.431
Postfach 11 05 60, D-42665 Solingen
Germany
ApprovedDocument-53081TPH_r3.pdfPage39of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Section 2-1Overview
2-1-1 Purpose of Chapter 2
This chapter provides the information required to plan and prepare for the installation of a
LOGIQ 100 PRO. Included are descriptions of the facility and electrical needs to be met by the
purchaser of the unit. A checklist is also provided at the end of this section to help determine
whether the proper planning and preparation is accomplished before the actual equipment
installation is scheduled.
CONTENTS IN CHAPTER 2
Operating temperature: 10 to 40 oC
ApprovedDocument-53081TPH_r3.pdfPage40of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
2-2-1-1 Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a
source of EMI which could degrade image quality. These controls should be selected to minimize
possible interference.
CAUTION POWER OUTAGE MAY OCCUR. The LOGIQ 100 PRO requires a dedicated single branch
circuit. To avoid circuit overload and possible loss of critical care equipment, make sure you
DO NOT have any other equipment operating on the same circuit.
ApprovedDocument-53081TPH_r3.pdfPage41of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The LOGIQ 100 PRO
complies with limits as stated on the EMC label. However there is no guarantee that interference
will not occur in a particular installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect.
These sources include:
medical lasers,
scanners,
cauterizing guns,
computers,
monitors,
fans,
gel warmers,
microwave ovens,
light dimmers
portable phones.
The presence of a broadcast station or broadcast van may also cause interference.
See Table 2-4 for EMI Prevention tips.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage42of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Be aware of RF Keep the unit at least 5 meters or 15 feet away from other EMI sources.
sources Special shielding may be required to eliminate interference problems
caused by high frequency, high powered radio or video broadcast
signals.
Ground the unit Poor grounding is the most likely reason a unit will have noisy images.
Check grounding of the power cord and power outlet.
Replace all screws, After you finish repairing or updating the system, replace all covers and
RF gaskets, covers, tighten all screws. Any cable with an external connection requires a
cores magnet wrap at each end. Install the shield over the front of card cage.
Loose or missing covers or RF gaskets allow radio frequencies to
interfere with the ultrasound signals.
Replace broken RF If more than 20% or a pair of the fingers on an RF gasket are broken,
gaskets replace the gasket. Do not turn on the unit until any loose metallic part is
removed.
Do not place labels Never place a label where RF gaskets meet the unit. Otherwise, the gap
where RF gaskets created will permit RF leakage. Or, if a label has been found in such a
touch metal position, move the label.
Use GE specified The interconnect cables are grounded and require ferrite beads and other
harnesses and shielding. Also, cable length, material, and routing are all important; do
peripherals not change from what is specified.
Take care with Cellular phones may transmit a 5 V/m signal; that could cause image
cellular phones artifacts.
Properly dress Do not allow cables to lie across the top of the card cage or hang out of
peripheral cables the peripheral bays. Loop the excess length for peripheral cables inside
the peripheral bays. Attach the monitor cables to the frame.
ELECTRONIC PAMPTE
Operation: 10 to 40 oC 5 to 42.7 oC
NOTICE: SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES
OF -20 TO + 50 deg C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage43of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Site preparation takes time. Begin Pre installation checks as soon as possible, if possible, six weeks
before delivery, to allow enough time to make any changes.
ApprovedDocument-53081TPH_r3.pdfPage4of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Your Notes:
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage45of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Chapter 3 Installation
Section 3-1Overview
3-1-1 Purpose of Chapter 3
This chapter contains information needed to install the unit. Included are references to a procedure
that describes how to receive and unpack the equipment and how to file a damage or loss claim.
How to prepare the facility and unit of the actual installation, and how to check and test the unit,
probes, and external peripherals for electrical safety are included in this procedure. Also included
in this section are guidelines for transporting the unit to a new site.
CONTENTS IN CHAPTER 3
Table 3-1 Contents in Chapter 3
Section Description Page Number
3-1 Overview 3-1
3-2 Receiving and Unpacking the Equipment 3-3
3-3 Preparing for Installation 3-5
3-5 Completing the Installation 3-6
3-6 System Configuration 3-17
3-7 Paperwork 3-28
ApprovedDocument-53081TPH_r3.pdfPage46of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use
CAUTION a three to two prong adapter. This defeats safety grounding.
Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
CAUTION present.
Do not use a 20 Amp to 15 Amp adapter on the 120 Vac units power cord. This unit requires
CAUTION a dedicated 20 A circuit and can have a 15A plug if the on board peripherals do not cause the
unit to draw more than 14.0 amps.
Do not operate this unit unless all board covers and frame panels are securely in place.
CAUTION System performance and cooling require this.
OPERATOR MANUAL(S)
CAUTION The User Manual(s) should be fully read and understood before operating the LOGIQ 100
PRO and kept near the unit for quick reference.
ApprovedDocument-53081TPH_r3.pdfPage47of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1.) Cut the two straps wrapped around the box. Refer Figure 3-1.
2.) Remove the adhesive tape. Refer Figure 3-1
3.) Open the carton flaps. Refer Figure 3-2.
4.) Remove the Operator Manual, Service Manual, Gel Bottle, Power Cord and other
accessories kept on either side of the equipment. Refer Figure 3-3.
5.) Lift the system (wrapped with stretch film) out of the box along with the packaging buffers.
Refer Figure 3-2
6.) Slide the packaging buffers out to free the system.
7.) Remove the stretch film which is wrapped around the system. Refer Figure 3-2
Straps
Adhesive Tap
ApprovedDocument-53081TPH_r3.pdfPage48of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Packaging Buffe
Carton Flaps
Note: The Service Manual, Operator Manual,Advanced Refernce Manual, Gel Bottle and
Power Cord are wrapped in bubble sheet and are kept along the sides of the
corrugated carton box. Refer Illustration below for details
ApprovedDocument-53081TPH_r3.pdfPage49of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
WARNING CONNECTING A LOGIQ 100 PRO SCANNER TO THE WRONG VOLTAGE LEVEL
WILL MOST LIKELY DESTROY THE SCANNER.
This unit has been designed to minimize the effects of Electro Magnetic Interference (EMI). Many
of the covers, shields, and screws are provided primarily to protect the system from image artifacts
caused by this interference. For this reason, it is imperative that all covers and hardware are
installed and secured before the unit is put into operation.
CAUTION It is strongly recommended that the equipment should be installed on an even and hard
surface to provide proper cooling, as the vents for air circulation are on the bottom cover.
1.) Locate, remove and unpack the probes located in the separate container.
2.) Make sure that system power is OFF before connecting or disconnecting the probe.
3.) Secure the probe to the receptacle as follows. Refer to Fig 3-4
a.) Ensure probe `twist lock' lever points towards the top of the probe connector.
a.) Install probe connector on the receptacle guide pin until it touches the receptacle mating
surface.
a.) Twist the probe connector `twist lock' lever in the clockwise direction to lock it in place
(Twist the lever anticlockwise direction to disconnect the probe).
4.) Connect the LOGIQ 100 PRO Power cable's female part to the power connector located on
the rear panel and the male part to a hospital grade power socket of a proper voltage. Never
use a three-to-two pin adapter; this defeats the purpose of safety ground.
5.) Unlock the keyboard by pressing the lock release on top of the system.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
CAUTION Do not move or lift the unit, holding the keyboard alone.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage50of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
2.) Connect the main power cable to a hospital grade power receptacle with the proper rated
voltage. Never use a three-to-two prong adapter; this defeats the safety ground.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage51of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1.) Unpack the foot switch and connect it to the connector on the rear panel. (Refer Figure 3-5)
2.) Connect the VCR to a suitable power outlet. Using the BNC to BNC cable connect the Video
Out on the rear panel of LOGIQ 100 PRO to the Video IN of the VCR.
3.) Connect the Video Graphic Printer (VGP) to a suitable power outlet. Using the BNC to BNC
cable connect the Video Out on the rear panel of LOGIQ 100 PRO to the Video IN of the
VGP. Connect the shutter cable from the VGP printer to the rear panel of the Logiq 100 PRO.
4.) The images and report pages of the Logiq 100 PRO can be printed on HP Laser jet
printer(Model HP1020) through the USB port provided on the rear panel.Images can be
transfered to a PC using USB Thumb drive.
Note: The LOGIQ 100 PRO has only one Video Out connection Therefore either a VGP
or a VCR only can be connected at a time
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage52of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage53of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
3-5-3 Installation of Two Probe Adapter & LOGIQ 200 Probe Adapter (Optional)
1.) Remove the adhesive tape to open the flaps of two probe port or LOGIQ 200 Probe Adapter
box
2.) Remove the PU Foams inside the two probe port or LOGIQ 200 Probe Adapter box (Refer
Figure 3-6)
3.) Take out the two probe port or LOGIQ 200 Probe Adapter (Refer Figure 3-6)
PU Foam
Box
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
Figure 3-6 Unpacking Two Probe Adapter or LOGIQ 200 Probe Adapter
Note: The two probe port or LOGIQ 200 Probe Adapter can be connected to or disconnected
from the system at any time regardless of whether the system is powered ON or OFF.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage54of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Installation of Two Probe Adapter & LOGIQ 200 Probe Adapter (Optional) (contd.,)
4.) Ensure that no probe is connected to the system. If connected remove it first.
5.) Remove the probe holder by lifting it vertically up
6.) Make sure that the two probe port or LOGIQ 200 Probe Adapter lock points to the 12 o'clock
position.
7.) Align the Two Probe Port or LOGIQ 200 Probe Adapter Connector with the System Probe
Port and carefully push it in.
8.) Align the slots on the upper flange to its counterparts on the top cover of the system.
9.) Turn the Two Probe Port or LOGIQ a 200 Lock to the 3 o'clock position to secure the Two
Probe Port or LOGIQ a 200 Probe Adapter to the system.
10.) Carefully replace the probe holder to its place and press firmly till the projections at
the bottom fit into the slots in the upper flange and the system.
Probe Holder
Figure 3-7 Installing Two Probe Port or LOGIQ 200 Probe Adapter
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage5of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Note: LOGIQ 100 PRO has two types of Trolleys, the Global Trolley and the Modular Trolley.
The Global Trolley comes in the assembled form and the Modular Trolley comes in a modular
form which needs assembly at field.
1.) Unpack the global trolley packed in a wooden box by opening the top cover, refer Figure 3-8
(trolley which is inside the box is seen in the diagram).
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
ApprovedDocument-53081TPH_r3.pdfPage56of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
2.) Unpack the Modular Trolley packed in a box by opening the top cover, refer Figure 3-8
ApprovedDocument-53081TPH_r3.pdfPage57of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage58of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
3.) Remove two M4X10 screws from the base ofLOGIQ 100 PRO and fix lock plate. Refer
Figure 3-11
4.) Place the unit with lock plate fixed on Trolley, ensure that lock plate gets into the projection on
the trolley. Refer Figure 3-11
Lock Plate
Detail X
Detail X
Console
Trolley
Lock Plate
Additional Details
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
ApprovedDocument-53081TPH_r3.pdfPage59of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
5.) Take the belt (provided in the kit) over the unit and lock it. Refer Figure 3-12
6.) Release the Keyboard by pressing the lock release and make it rest on the handle. Refer
Figure 3-12
7.) Insert the VGP inside the Printer Tray and fix from the bottom using M3X12 screw. Refer Figure
3-12
Handle
VGP
ApprovedDocument-53081TPH_r3.pdfPage60of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Figure 3-13 Installation of LOGIQ 100 PRO with Global or Modular Trolley
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage61of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Turn ON the Machine and then adjust the monitor contrast and brightness. Monitor contrast and
brightness should be adjusted according to the lighting in the room. For Monitor brightness and
contrast adjustment refer to the appropriate section in the User Manual.
Press CTRL-S-1 ENTER, an `Installation Setup window appears'. Set the system Clock/Date,
Hospital Name and OB Version. These settings can be modified by using the CONTROL-S-1 key if
required. Refer Section Control Keys of the User Manual for more details
CAUTION Functional checks in Chapter 4, to be carried out before handing over the system to the
customer
ApprovedDocument-53081TPH_r3.pdfPage62of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
3-6-1-2 Front View
ApprovedDocument-53081TPH_r3.pdfPage63of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
3-6-1-3 Side View (Keyboard Closed)
420.6 (16.5)
405 (15 9)
Figure 3-15 Side View (Keyboard Closed)
289(11.4)
596 (23.5)
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
ApprovedDocument-53081TPH_r3.pdfPage64of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
SIDE VIEW
ApprovedDocument-53081TPH_r3.pdfPage65of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
3-6-2-1 Power Supply
The Following Power Line requirement should be monitored one week before Installation. We
recommend that a Drantez model 605-3 Power line Monitor with options 101 to be used
Table 3-6
Parameter Limits
Decaying Oscillation Less than 15% of Peak Voltage for less than 1 millisecond
LOGIQ 100 PRO peripherals and accessories can be properly connected using the connectors
on the rear panel of the LOGIQ 100 PRO system.
External Optional Peripherals
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
ApprovedDocument-53081TPH_r3.pdfPage6of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Printing should not be done while live scanning patients. Also, the Printer cable should not be
attached to the LOGIQ 100 PRO during the scan.
For Europe Only: All devices connected to the LOGIQ 100 PRO must be CE Marked.
Located on the panel are video input & output connectors, footswitch connector, shutter connector,
power connector & control connections for Video Graphic Printer and USB Port for the
LOGIQ 100 PRO.
The Thumb Drive/Memory Stick (Model TS512MJF150), HP Laser Jet printer(HP1020), Sony VGP
Printer( Model UP-D897MD/ UP-897MD) can be connected through the USB slot provided in the
rear panel.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage67of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage68of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Installation of VGP:
1.) Place the SONY Digital Printer on the fixture.
2.) Fix one end of power cable and USB cable to the printer.
3.) Fix the other end of USB cable to the USB slot provided in the rear panel and power cable
to the Power socket.
4.) Switch ON the LOGIQ 100 PRO and VGP printer.
5.) Press the Recall button on the keyboard to view the stored images.
6.) Press Record button on the keyboard and check if the selected image matches with the
VGP print out.
ApprovedDocument-53081TPH_r3.pdfPage69of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Installation of VCR:
1.) Connect the SONY VCR
2.) Switch on the VCR and the LOGIQ 100 PRO unit.
3.) Press the Record key to record the image.
4.) Press the External video key on the keyboard to enable external video( i.e VCR Playback)to
be viewed on the LOGIQ 100 PRO system monitor.
ApprovedDocument-53081TPH_r3.pdfPage70of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Note : USB hard disk drive may/may not be recognized by the system.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage71of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Probe Freq in
Part No. Catalog No.
Type MHz
C36 2268634 H41252CF 3.5
* This Probes are available only for LOGIQ 100 PRO with LOGIQ 200 Probe Adapter.
Use only approved probes, peripherals or accessories.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage72of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Video specifications may be needed to be able to connect laser cameras to the LOGIQ 100 PRO.
General
SL Parameter NTSC PAL
7 Display width and Height [pixel * line] 608 X 464 608 X 472
Vertical Timing
SL Parameter NTSC PAL
Section 3-7Paperwork
Note: During and after installation, the documentation (i.e. User Manuals, Installation Manuals...) for the
peripheral units must be kept as part of the original system documentation. This will ensure that all
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
relevant safety and user informations are available during the operation and service of the complete
system.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage73of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Check that the correct User Manual(s), per software (sw) revision, for the system is included.
Fill out proper customer information on the Product Locator Installation Card. Refer to Illustration
below. Mail this Installation Card Product Locator to the address corresponding to your pole.
Note: The Product Locator Installation Card shown in Figure 3-25 may not be same as
the Product Locator card actually provided.
ApprovedDocument-53081TPH_r3.pdfPage74of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage75of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Section 4-1Overview
4-1-1 Purpose of Chapter 4
This chapter provides procedures for quickly checking major functions of the LOGIQ 100 PRO
scanner, diagnostics by using the built-in service software, and power supply adjustments. For
remaining functional checks please refer to the LOGIQ 100 PRO Operator's Manual.
CONTENTS IN CHAPTER 4
ApprovedDocument-53081TPH_r3.pdfPage76of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Note: After turning off the system, wait at least ten seconds before turning it on again. The system may
not be able to boot if power is recycled too quickly.
4-2-1-3 Power Up
1.) Connect the Power Cable at the rear of the System.
2.) Connect the Power cable to an appropriate mains power outlet .
Mains Power
Circuit Breaker
Cable Connector
ApprovedDocument-53081TPH_r3.pdfPage7of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
4-2-1-4 Front Panel ON/OFF
Press the Front Panel ON/OFF key once. to Switch ON/OFF
When power is applied to the Scanner, Power is distributed to the Fan, Control panel, Monitor,
Internal and External I/Os, Boards & Peripherals. Back End Processor starts and its software code
is distributed to initiate the scanner.
ApprovedDocument-53081TPH_r3.pdfPage78of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage79of1
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage80of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Table 4-2 describes basic controls which helps in checking the LOGIQ 100 PRO during
installation.
Connect the sector or convex transducer to connector on the right side of the base.
Press New Patient Key A pop-up window appears. Enter Patient Name/ID. It
presets the system
Press ID/Name A pop-up window appears. Enter Patient Name/ID. It
does not preset the system.
Press Dyn Range Up/Down Image grows softer and harder depending on position.
Rotate Gain Knob Image grows lighter and darker with rotation.
Press M key B/M Mode with M-line cursor appears. (Use trackball to
move the M-line cursor). Refer Figure 4-4
Roll Trackball The M-Line cursor should follow the trackball movement
and the real-time image varies on the M-Mode display.
Press M key a second time Only M-Mode image appears on the screen.
Press B Mode Key to exit M-Mode.
Press Freeze and Zoom Zoom is used to magnify an image. The magnification
factor for zoom is fixed at 2.0. The system adjusts all
Roll Trackball imaging parameters accordingly.
To select the ROI for Zoom
Zoom feature is avialable in the freezed mode only.
ApprovedDocument-53081TPH_r3.pdfPage81of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
CINE
(Press Freeze and Rotate B/M Gain/ The Cine Gauge, menu and the Cine frame number
Cine Scroll ) appears at the bottom of the display.
Rotate the Cine Scroll dial to move through the images
in Cine Memory.
Cine Menu: Cine frame number is displayed on the left side of the
screen.
Press 1 (Start Frame) Move the Cursor by rotating the Cine Scroll dial to the
frame you want and press 1 again to select the Start
frame.
Move the Cursor by rotating the Cine Scroll dial to the
Press 2 (End Frame) frame you want and press 2 again to select the End
frame.
Press 3 (Cine Loop) Enters into Cine loop within the selected Start & End
frames.
Press 4 (Cine Gauge) To toggle between the Cine Gauge display
ApprovedDocument-53081TPH_r3.pdfPage82of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage83of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The power on diagnostics or self tests are run every time the system is booted. This self tests
include testing the validity of the system software through EPROM checksum test and testing the
system RAM. It checks whether the keyboard and the trackball are properly interfaced to the
system. It also initiates the self test of analog subsystem. The two LED's and the beeper are used
to indicate error conditions.
Note: Power on Diagnostics works only when the system is in application (imaging) mode
and not while on service diagnostics.
LED - Location on FEB This LED indicates the scan lines,the DSP is scanning.The
FEB D65 LED glows when the cursor is moved in the B/M mode
LED - Location on FEB This LED indicates the probe connect signal is good.This
FEB D66 indicates the FEB is Good
LED - Location on FEB This LED indicates the B mode signal is given to the
FEB D67 display.The intensity of this LED is more during B mode.
LED - Location on Davinci This LED indicates the KEY press on the keyboard.It glows
DAVINCI CPU D10 intermittantly when there is a key pressed.
ApprovedDocument-53081TPH_r3.pdfPage84of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1. Check Date and Time setting Date and Time are correct Adjust the Date and Time setting
2. Check that Location (Hospital Name) is Location Name is correct Re-enter the correct Location Name
correct
2. Press RECORD Key on the Control The image displayed on the screen is
panel printed on B&W Video Graphic Printer.
4. Press SHIFT + RECORD Key on the To print the displayed image on the
Control Panel. screen to the USB B&W Video Graphic
Printer.
ApprovedDocument-53081TPH_r3.pdfPage85of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1. To test each delivered Probe, Connect it It will display the probe name on the
to the System screen
CAUTION:
Do not allow the probe head to hang
freely. Impact to the probe head may
result in irreparable damage.
6. Two Probes can be Switched by The LED toggles to indicate the probe
pressing the Key on the Two Probe Port change & a click is also heard to
indicate change over
The system initalizes the new probe &
the image from the newly selected
probe is displayed in the B-Mode
desired parameters
Press CTRL + W + Enter
8. To test the LOGIQ PROBE ADAPTER, Connect the Adapter to the Port
connect it to the system
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage86of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
9. Hold the probe connector horizontally It will display the Probe name on the
with the cable pointing Sideways. Screen
Turn the connector locking handle to the
Vertical position.
Align the connector with the probe port
and carefully push into place.
Rotate the locking handle to the full
Horizontal position to lock in place.
Position the probe cable so that it is not
resting on the floor
CAUTION:
Do not allow the probe head to hang
freely. Impact to the probe head may
result in irreparable damage.
WARNING Probes can be connected at any time, whether the unit is on or off
WARNING Do NOT touch the patient and any of the connectors on the ultrasound unit simultaneously,
including ultrasound probe connectors.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage87of19
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Section 5-1Overview
5-1-1 Purpose of Chapter 5
This chapter explains LOGIQ 100 PROs system concepts, component arrangement, and
subsystem function. It also describes the Power Distribution System (PDS) and probes.
CONTENTS IN CHAPTER 5
- Page 5-1
ApprovedDocument-53081TPH_r3.pdfPage8of197
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
6
7
LOGIQ 100 PRO is a portable ultrasound imaging scanner. It also has provisions for analog
input sources like The system can be used for:
- B-Mode Black and White imaging
- M-Mode Black and White imaging
- Combinations of the above
LOGIQ 100 PRO is a digital beam forming system and can handle up to 16 channel,64
element by use of multiplexing.
Signal flow from the Probe Connector to the Front End, then to the Mid Processors and Da
Vinci CPU and finally to the monitor and peripherals.
System configuration is stored on EPROM and all necessary software is loaded when powered
up.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage89of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
BUS
Control
Davinciboard
TRANSMIT Signal
TX ECO Signal
Probe Display
RX Front End Board
RECIEVE Signal
Patient
Patient
Isolated
I/O
Sensors
External I/O
1 0 Normal mode
0 1 Service mode
Table 4-7
DIP 3 MODE
0 NTSC
1 PAL
- Page 5-3
ApprovedDocument-53081TPH_r3.pdfPage90of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
FEB Block Diagram.
ApprovedDocument-53081TPH_r3.pdfPage91of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
- Page 5-5
ApprovedDocument-53081TPH_r3.pdfPage92of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage93of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The following table lists the Circuit boards on LOGIQ 100 PRO system.
Davinci Davinci Board For LOGIQ 100 PRO (All S/W Versions)
FEB Front End Board For LOGIQ 100 PRO (All S/W Versions)
PDB Power Distributor PCB For LOGIQ 100 PRO (All S/W Versions)
HVPS High Voltage Power Supply For LOGIQ 100 PRO (All S/W Versions)
Table 4-9
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
- Page 5-7
ApprovedDocument-53081TPH_r3.pdfPage94of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Figure 4-9 Power Distributor/Power Supply Wiring Diagram for LOGIQ 100 PRO.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage95of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Table 4-10 Power Distributor Connector Voltages for LOGIQ 100 PRO .
Pin 2 - +3.3V in Pin 2 - Gnd Pin 2 - Gnd Pin 2 - +12VD Pin 2 - +3.3V
Pin 7 - +12VD
Pin 8 - Gnd
Pin 9 - -12VD
Pin 10 - Gnd
Table 4-11 Power Distributor Input and Output Voltages of LOGIQ 100 PRO .
NOTE : The Test point information is provided on the Base of the console, which is visible after
opening the top cover.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
- Page 5-9
ApprovedDocument-53081TPH_r3.pdfPage96of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
5-4-0-2 Cable Assembly Diagrams
ApprovedDocument-53081TPH_r3.pdfPage97of1
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Cable Assembly Diagrams (Continued)
Figure 4-10
- Page 5-11
ApprovedDocument-53081TPH_r3.pdfPage98of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The Front End Board generates the strong transmit bursts, transmitted by the probes as
ultrasound into the body. It also receives weak ultrasound echoes from blood cells and body
structure, amplifies the signals and converts them to an 8 bit digital signal. FEB supports 16
channels Beamformer. The beamformed data is passed through CHACOM ASIC and
subsequently passed through Frame Averaging filter and Mean filter. The Output of FEB is fed to
the Davinci CPU PWA through a board-to-board connector. This connector also provides power
to the Davinci CPU PWA Board.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage9of17
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The Davinci CPU board uses TMS320DM6446 chip that contains an ARM Processor, DSP and
Video processors into a single chip.
2) Program Storage 16 Megabytes of NOR Flash is used to store the Boot loader code, Kernel,
and Application software.
3) Image Storage 64 Megabytes of NAND Flash is used to store the User Images. It is also
used to store a part of Application software and operating system code.
Optical Trackball TB_CLK and TB_DATA pins are connected to Davinci CPU PWA. Trackball
supports Ps2 standard.
GAIN Encoder - GAIN_CCW and GAIN_CW pins of the gain encoder are connected to the GPIO
pins of Davinci CPU.
5) On Screen Display (OSD), Video Out and Monitor Display - The DAVINCI CPU generates the
OSD along with the image in NTSC or PAL video output format (standard) to the Video Out and
Monitor connector.
6) USB port interface - The Davinci CPU PWA provides one USB 2.0 Host connector. This port is
used to interface Thermal Printer, DeskJet printer as well as to transfer the image to Memory
stick in JPEG format.
7) NTSC and PAL setting - A Dipswitch is provided to select NTSC or PAL standard. This switch
is available on Davinci CPU PWA.
8) Real Time Clock (RTC) - RTC chip is provided on Davinci CPU PWA that acts as a timekeeper
i.e. it maintains the system clock and date.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
- Page 5-13
ApprovedDocument-53081TPH_r3.pdfPage10of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
5-5-3 Peripherals
VCR, Black & White Video Graphic Printer and HP Laser Jet Printer 1020 can be connected to the
scanner. For more details on recommended peripherals refer Chapter 3.
ApprovedDocument-53081TPH_r3.pdfPage10of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
- Page 5-15
ApprovedDocument-53081TPH_r3.pdfPage102of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage103of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The LOGIQ 100 PRO does not contain service adjustable parts. In case of failure of any module
like the Power Supply, HV pcb it is recommended to replace the power supply unit instead of trying
to adjust the potentiometer or replacing the fuse.
No Service Adjustments.
WARNING Whenever Upgrade of LOGIQ 100 PRO is done at field, Please ensure patient data back up is
done before upgrading the machine.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage104of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Your Notes:
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage105of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting
Section 7-1Overview
7-1-1 Purpose of Chapter 7
This section describes how to setup and run the tools and software that help maintain image quality
and system operation. Very basic host, system and board level diagnostics are run whenever power
is applied. Some Service Tools may be run at the application level.
CONTENTS IN CHAPTER 7
ApprovedDocument-53081TPH_r3.pdfPage106of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
LED - Location on FEB This LED indicates the scan lines,the DSP is scanning.The
FEB D65 LED glows when the cursor is moved in the B/M mode
LED - Location on FEB This LED indicates the probe connect signal is good.This
FEB D66 indicates the FEB is Good
LED - Location on FEB This LED indicates the B mode signal is given to the
FEB D67 display.The intensity of this LED is more during B mode.
LED - Location on Davinci This LED indicates the KEY press on the keyboard.It glows
DAVINCI CPU D10 intermittantly when there is a key pressed.
ApprovedDocument-53081TPH_r3.pdfPage107of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Note: Power on Diagnostics works only when the system is in application (imaging) mode
and not while on service diagnostics.
Service Diagnostics
Check the S/W Version of your System prior to start of the service diagnostics. The S/W version
can be known during the power up Sequence.
MAIN MENU
1. DAVINCI BOARD TESTS
2.KEYBOARD TESTS
3. MONITOR TEST
4. KEYBOARD OPERATION HISTORY
5. FRONT END BOARD TESTS
Q. QUIT
Note: Option Q (Quit) can be used to come out of the diagnostic and enter the application
mode.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage108of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
All Tests
This option is used to run all tests in the menu. Once all the tests are completed, press any key to
exit.
ApprovedDocument-53081TPH_r3.pdfPage109of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Graphics Test
The pattern shown below is displayed on the monitor when this test is enabled.
FRAM Test
When this test is selected the following will appear
NAND FLASH
MANUFACTURER ID 0x20
MANUFACTURER NAME ST Micro
NUMBER OF BLOCKS 4096
NUMBER OF BAD BLOCKS 0
USABLE MEMORY (IN BYTES) 67108864
NOR FLASH
MANUFACTURER ID 0x1
MANUFACTURER NAME AMD AM29LV256M
ApprovedDocument-53081TPH_r3.pdfPage10of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
All Tests
This option is used to run all tests in the menu.
Note: This option must be run during the following instance without fail
A.) Take a back up of the archive before the system is upgarded to a higher version of software.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage1of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Main Menu
This option is used to return to the main menu.
Quit
This option is used to come out of the diagnostic and enter the application mode.
Main menu
This option is used to return to the main menu
Quit
This option is used to come out of the diagnostic mode and enter the application mode.
When self test for the FEB is selected, the DPRAM on the FEB Board will be written on and read
back. Read data is checked for validity. This will prove the communication between FEB & Davinci
boards.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
All Tests
This option is used to run all tests in the menu. Once all the tests are completed, press any key to
SeethGEHCMyworkshpSystemodrinhaufc.
exit.
Chapter 7 - Diagnostics/Troubleshooting Page 7-7
ApprovedDocument-53081TPH_r3.pdfPage12of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Change the modes to B/B, B/M and M modes and check for the proper display of the FEB Test
pattern.
ApprovedDocument-53081TPH_r3.pdfPage13of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
KEYBOARD TESTS
1.) ALL TESTS
2.) KEYBOARD ALIVE TEST
3.) KEYS TEST
4.) LEDS TEST
5.) TRACKBALL TEST
6.) GAINKNOB TEST
P.) MAIN MENU
Q.) QUIT
Keys Test
In this test, all the keys in the keyboard can be tested. The user is prompted to press the keys on
the keyboard and the pressed keys are highlighted on the screen. (This is a toggle operation).
Note: During Keys Test Press Ext. Video at the last. Pressing Ext. Video key
interrupts the test and return to main keyboard Tests Menu.
LEDS Test
During this test FREEZE field and EXTERNAL VIDEO field will be displayed on the monitor. The
FREEZE and EXTERNAL VIDEO LEDs blink for a while sequentially and the corresponding field
gets highlighted.
Trackball Test
A cursor is displayed on the monitor and the user can move the trackball and move the cursor over
the screen.
Main Menu
This option is used to return to the main menu.
Quit
SeethGEHCMyworkshpSystemodrinhaufc.
This option is used to come out of the diagnostic and enter the application mode.
Chapter 7 - Diagnostics/Troubleshooting Page 7-9
ApprovedDocument-53081TPH_r3.pdfPage14of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The following test patterns are displayed on the monitor when this test is enabled. Please match
the patterns displayed on your monitor with the given patterns figures 7-7 to 7-13. The Gain Knob
is used for selection of patterns 1-10. To exit out of any pattern and return to the main menu, press
EXTERNAL VIDEO.
ApprovedDocument-53081TPH_r3.pdfPage15of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage16of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage17of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Note: The patterns in 7-7 to 7-10 are written on the overlay where as the pattern in 7-11 to 7-13 is written
on the image memory.
7-7 to 7-9 are also available in reverse video
When this option is selected from the Main Menu the following appears indicating a maximum of 32
keys last activated. The history of keys used is stored even after powered off.
15. 31.
16. 32.
ApprovedDocument-53081TPH_r3.pdfPage18of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Errors Remarks
THE KEYBOARD SUBSYSTEM FAILED This Message is displayed during the diagnostics of the Keyboard. When the keyboard
TO RESPOND firmware is not alive, this message is displayed.
KEYBOARD NOT CONNECTED If the keyboard is not connected then this message is displayed during system boot up.
TRACKBALL INTERFACE ERROR This message is displayed during system boot up and the trackball interface to keyboard is not
proper.
NO TRACKBALL ACKNOWLEDGEMENT This message is displayed during system boot up and keyboard connection is not there.
PLEASE POWER OFF AND RESTART This message is displayed when the system boots up in the normal mode and any error which
THE SYSTEM! needs the system to be restarted comes up.
PRINTER IS BUSY This message comes up when shift + record is pressed without connecting the printer. Only
for LOGIQ 100 PRO
PRINTER ERROR This message comes up when an error occurs in the printing process.
PRINTING ABORTED This message is displayed during Printing & due to printer error. Only for LOGIQ 100 PRO
WARNING: THIS WILL ERASE ALL This message is displayed when the 3 (Clear ALL) is pressed after going to the recall screen.
IMAGES! 1:YES 2:NO Only for LOGIQ 100 PRO.
ARCHIVE FULL This message is displayed when the store key is pressed while the archive(USB) is full. Only
for LOGIQ 100 PRO.
USB IS NOT READY This message is displayed when shift + store is pressed without connecting the USB. Only for
LOGIQ 100 PRO.
USB TRANSFER ERROR This message is displayed when some error occurs in the USB transfer process. Only for
LOGIQ 100 PRO.
ApprovedDocument-53081TPH_r3.pdfPage19of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
TROUBLESHOOTING
No Display, total black out No AC Power Ensure power cable is ok and there is power
at the wall outlet
Circuit Breaker tripping Power Supply or PCB may be loading.
Replace the defective power supply or the
PCB
Brightness/Contrast knobs may be at end Turn the knobs to the suitable position
limit
No video signal to the monitor Ensure video cable connection from the
Davinci to the monitor pcb is intact
Monitor may be defective Feed an external video signal to confirm
Set right the same
Power Supply to the Monitor may be bad (12
Volts)
No trackball Acknowledgement or Trackball or the cable maybe defective Replace appropriately. Contact your local
Trackball Interface error - Error message service representative for further information
on this.
Keyboard or Davinci or interface cable may Replace appropriately
be defective
Keyboard not connected - Error message Keyboard to Davinci interface cable may be Replace the same
defective.
If the EXT Video LED and the FREZZE LED Replace the same
are glowing simultaneously during Boot Up
then the keyboard may be defective.
No System Mode Acknowledgement - Error Davinci interface may not be proper Replace FEB or Davinci pcb
message
No Echoes otherwise system is ok HV Power Supply maybe defective, Check for the cable assembly 30N.
Connection of the CA 30N may be loose. Replace the HV Power Supply
During some boot ups the DSP application If this problem appears reboot the system.
may not be fully loaded.
System does not respond to Gain variation Cable interfacing Davinci and KBD maybe Replace the same
defective.
Replace appropriately
Davinci or FEB maybe defective
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
ApprovedDocument-53081TPH_r3.pdfPage120of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
System Resets intermittently 3.3 Volt power supply may be less. a. First check for the 3.3V DC supply to the
CPU board & ensure Proper Voltage.
b. If the voltage is not at required level,
Proceed to check if the power supply is a
problem or the Davinci board.
c. If the voltage is ok then check for the
connection between the Davinci board and
FEB.
d. If still not working relplace the boards
approppriately.
Display shrinks and/or shakes Problem maybe with monitor or power supply Replace appropriately
to the monitor
Image blooming If only image area (Problem maybe with HV Replace appropriately
Power supply or FEB.
If problem is with full monitor display Replace the monitor
Ring artefacts on the monitor in the near field Due to pickup signal when the probe is Connect the probe and scan to check the ring
in Air scan. connected in air scan. artifacts on an organ.
Low Frequency moving artefacts AC line interference Change the AC input , Check for proper
grounding.
Video-In/Video Out not working Connection on the rear panel not ok Set right the same
Foot switch not working Connection on the rear panel not ok Set right the same
Remote Control not working Rear panel or Davinci maybe defective Replace appropriately
"Image Archive Full" message not The memory size allocated for Image storage Transfer all the archived images to memory
appearing. is full due to uneven size of the stored stick from the recall menu.
But the archive is full and image cannot be images. Go to service diagnostics and Erase Flash
stored. Memory in the Davinci Test menu.
ApprovedDocument-53081TPH_r3.pdfPage12of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Machine Not Booting
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage12of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage123of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Machine Resets Intermittently
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage124of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Junk Image in the Scan Area
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage125of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage126of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
No Scan Image, Probe is firing.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage127of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Your Notes:
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage128of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage129of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
CONTENTS IN CHAPTER 8
Page
Section Description Number
8-5-2 Disassembly/Assembly of FEB for LOGIQ 100 PRO (FRU 703) 8-18
ApprovedDocument-53081TPH_r3.pdfPage130of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Follow the steps below and refer Figure 8-1, For FRU Details refer 9-1-2
1.) Always Shut Down the System when disassembling the TOP COVER
2.) Disconnect the power cord from the mains and remove from the system.
3.) Remove Probe, which is at the right side of the base.
4.) Remove the probe holder.
5.) Remove two screws (a,b) at the side of the probe holder.
6.) Move the unit to the edge of the table and remove two screws (e,f) from the bottom of the unit.
7.) Remove the handle cap by pulling upward and remove the handle screws (c,d).
8.) Slide the top cover backward by holding the handle.
9.) For assembly follow the reverse order.
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage13of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Disassembly/Assembly (Contd.)
Handle Cap
d
Probe Holder
101 c
102
ApprovedDocument-53081TPH_r3.pdfPage132of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Section 8-3KEYBOARD
8-3-1 Detaching the Keyboard
Follow the steps below and refer Figure 8-2. For FRU Details refer 9-1-3
1.) Remove Front Panel. Refer 8-4-1.
2.) Remove 14-pin connector from the Davinci.
3.) Remove the black ground cable coming from the keyboard along with the 14 pin flat cable
connected to the chassis.
4.) Remove the TGC connector from the Davinci.
5.) Place a rubber mat below the unit and tilt the unit upward. Remove two screws (a,b) from
the bottom hinge of the keyboard and remove the bottom hinge cover refer Figure 8-2.
6.) Press the lock release and turn the keyboard 90. Move the keyboard downward and
carefully pull out the cable with the connector.
7.) Remove the 2 ring bushes refer Figure 8-2.
8.) For assembly follow the reverse order.
Service Manual
Section
Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage13of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage134of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Follow steps below and refer Figure 8-3 . For FRU Details refer 9-1-4
1.) Remove Top Cover. Refer 8-2.
2.) Detach the keyboard. Refer 8-3-1.
3.) Remove the spacer pad (refer FRU NO. 308 in the 8-7) which is near the lock release.
4.) Remove four screws a,b,c,d (b, is not visible in diagram).
5.) Remove keyboard top with PCB assembly.
6.) Remove the eight fixing screws (e to l) on the base plate.
7.) Remove the Gain knob (use allen key for M2.6 grub screw)
8.) Remove trackball mounting by removing two screws (man) and take out the trackball.
9.) Remove the TGC connector from the Davinci Board (for LOGIQ 100 PRO).
10.) Remove Gain Encoder connector.
11.) Remove the two screws (o,p) holding the Gain Encoder Unit.
12.) Remove eleven screws (q x 11 Nos.) holding the keyboard PCB to the base plate to separate
the keyboard PCB.
13.) For assembly follow the reverse order.
Note: When assembling the PCB to the base plate, M3x6 screws only should be used because a wrong screw
used will pierce the key sheet.
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage135of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
c 307
308 305
302
304
303
o 306
p l
e
n f
k
g
m
r
s
q (11 Nos) u t
i
309
h
301
ApprovedDocument-53081TPH_r3.pdfPage136of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
8-3-2-1 Optical Trackball Assembly
Note: When assembling the PCB to the base plate, M3x6 screws only should be used
because a wrong screw used will pierce the key sheet.
8.) For initializing the Optical Trackball Select Alps Trackball from the Trackball Selection
Menu in Service Diagnostics
C
A D
x y z
Figure 8-4 Optical Trackball Assembly
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage137of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
8-3-2-2 Optical Trackball Assembly
Note: When assembling the PCB to the base plate, M3x6 screws only should be used
because a wrong screw used will pierce the key sheet.
15.)For initializing the Optical Trackball Select Alps Trackball from the Trackball Selection
Menu in Service Diagnostics
C
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
ApprovedDocument-53081TPH_r3.pdfPage138of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Follow steps below and refer Figure 8-6 For FRU Details refer 9-1-5
1.) Remove Top Cover. Refer 8-2
2.) Lock the Keyboard before removing the front panel.
3.) Remove the keyboard connector from the Davinci Board.
4.) Remove the black ground cable coming from the keyboard .
5.) Remove Brightness and Contrast connector from the Brightness and Contrast PCB. (Refer
Detail 2 in the 8-11)
6.) Remove two screws (e,f) and detach the bottom hinge of the keyboard.
7.) Remove two screws (a,b) from the base along the rib.
8.) Remove two screws (c,d) from the support plate on the chassis.
9.) Disconnect the connector coming from the POWER ON Switch.
10.) Slide the front panel and keyboard sub-assembly together.
11.) For assembly follow the reverse order.
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage139of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
c d
402
403
401
e
404
f 405
Lock Hold
a Power Switch
b
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
ApprovedDocument-53081TPH_r3.pdfPage140of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Follow steps below and refer Figure 8-7. For FRU Details refer 9-1-6
1.) Remove Top Cover. Refer 8-2
2.) Remove Front Panel. Refer 8-4-1
3.) Disconnect the communication cables between the Analog and Digital circuits.
4.) Remove Shield Plate.
5.) Remove the communication cable - VIDEO IN (b), Brightness/Contrast (i,j) and the power
cable (a) on the CRT PCB.
6.) Remove the 2 screws (c,d) holding the CRT drive PCB to the chassis.
7.) Remove 4 screws (e,f,g,h) holding the CRT Assy connected to the chassis.
8.) Move the CRT forward along with CRT driver PCB assembly.
9.) For assembly follow the reverse order.
Note: Ensure that the monitor switch (all the four) near Fly Back Transformer is set to PAL side in case of
PAL systems and to NTSC side in case of NTSC systems.
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage14of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage142of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage143of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage14of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage145of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Dip Switch
Figure 8-10 Davinci board Dip Switch Setting for NTSC & PAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage146of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage147of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Mating connector
Davinci & FEB
ApprovedDocument-53081TPH_r3.pdfPage148of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Follow steps below and refer Figure 8-12. For FRU Details refer
1.) Remove Top Cover. Refer 8-2.
2.) Remove 2 screws (a,d) and loosen screws b,c from the chassis.
3.) Remove the connector (e) on the power distributor board which is connected to the fan.
4.) Remove Fan assembly from the unit.
5.) For assembly follow the reverse order.
.
ApprovedDocument-53081TPH_r3.pdfPage149of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage150of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
1.) Remove Top Cover. Refer 8-2. For FRU Details refer
2.) Remove Noise filter by removing screws (c,d) then disconnect the power cables from the noise
filter.
3.) Remove all five screws indicated in the fig below from the rear panel.
SeethGEHCMyworkshpSystemodrinhaufc.
4.) Remove the connectors from the Davinci board, FEB and CRT PCB.
ApprovedDocument-53081TPH_r3.pdfPage15of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
5.) Remove the connections to the circuit breaker.
6.) Remove the rear panel from the chassis.
7.) For assembly follow the reverse order.
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage152of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage153of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1. Remove Top Cover. Refer 8-2. For FRU Details refer 9-1-12
Service Manual
Section Functional Test/ Diagnostic Test Debrief Script
ApprovedDocument-53081TPH_r3.pdfPage154of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1201
f h
d
g
b
a e
c
1202
ApprovedDocument-53081TPH_r3.pdfPage15of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Section 9-1Overview
9-1-1 Purpose of Chapter 9
This chapter gives you an overview of Spare Parts for LOGIQ 100 PRO.
CONTENTS IN CHAPTER 9
ApprovedDocument-53081TPH_r3.pdfPage156of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
H41072LA INDIA
H41072LB CHINA, ANZ, SE-ASIA, MEXICO
H41072LC TAIWAN
H41072LD CHILE, KOREA
H41072LE EUROPE
H41072LF LA
H41072LS
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage157of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
5304117,5304414,5304240,5304248,5304441,5304596,5306834.
101
ApprovedDocument-53081TPH_r3.pdfPage158of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
5304117,5304414,5304240,5304248,5304441,5304596,5306834.
201
204
ApprovedDocument-53081TPH_r3.pdfPage159of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
303
305
304
309
311
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
310
ApprovedDocument-53081TPH_r3.pdfPage160of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage16of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
5304117,5304414,5304240,5304248,5304441,5304596,5306834.
Front Panel
401
ApprovedDocument-53081TPH_r3.pdfPage162of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
501
ApprovedDocument-53081TPH_r3.pdfPage163of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
5304117,5304414,5304240,5304248,5304441,5304596,5306834.
601
1.) Davinci is common for PAL & NTSC, Only dip switch setting has to be changed for the required
type.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage164of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
703
Table 9-7
704
Davinci board
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
FEB
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage165of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
801
802
Note: Replace Cable Assy 28N (5267859) & Cable Assy 29N (5265017) when replacing Power
Distributor PCB Assembly for the LOGIQ 100 PRO.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage16of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
5304117,5304414,5304240,5304248,5304441,5304596,5306834.
901
902
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage167of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1001
ApprovedDocument-53081TPH_r3.pdfPage168of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1301 LOGIQ 100 PRO Cable Set 5308178 1 1 For LOGIQ 100 PRO
Cable Assy 3N 5268930 1 Cable used for interfacing mains power, low voltage power supply, noise filter and
power switch Contains, apart from cable, Core Support Clamp and associated Screws
and Nuts
USB Cable 5269617 1 Davinci - Rear Panel USB Port cable for LOGIQ 100 PRO.
Assy
List of PROBES
Probe
Part No. Catalog No.
Type
C36 2268634 H41252CF
ApprovedDocument-53081TPH_r3.pdfPage169of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Notes:
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage170of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage17of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
This chapter describes Quality Assurance & Periodic Maintenance (PM) on the scanner and its
peripherals. These PM procedures are intended to maintain the quality of the ultrasound systems
performance. Read this chapter completely and familiarize yourself with the procedures before
starting a PM.
CONTENTS IN CHAPTER 10
Practice good ESD prevention. Wear an antistatic strap when handling electronic parts and
even when disconnecting/connecting cables.
ApprovedDocument-53081TPH_r3.pdfPage172of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
It is good business practice that ultrasound facilities maintain records of periodic and corrective
maintenance. The Ultrasound Periodic Maintenance Inspection Certificate provides the customer
with documentation that the ultrasound scanner is maintained on a periodic basis.
A copy of the Ultrasound Periodic Maintenance Inspection Certificate should be kept in the same
room or near the scanner.
In order to gain accreditation from organizations such as the American College of Radiology (USA),
it is the customers responsibility to have a quality assurance program in place for each scanner.
The program must be directed by a medical physicists, the supervising radiologist/physician or
appropriate designee.
Routine quality control testing must occur regularly. The same tests are performed during each
period so that changes can be monitored over time and effective corrective action can be taken.
Testing results, corrective action and the effects of corrective action must be documented and
maintained on the site.
Your GE service representative can help you with establishing, performing and maintaining records
for a quality assurance program.
The Periodic Maintenance Schedule specifies how often your LOGIQ 100 PRO should be
serviced and what items need attention. It is the customers responsibility to ensure the LOGIQ
100 PRO periodic maintenance is performed as scheduled in order to retain its high level of safety,
dependability and performance.
Your GE Service Representative knows your LOGIQ 100 PRO best and can provide competent,
efficient service. Please contact us for further information and to schedule GE Medical Systems
Ultrasound to perform this service for you.
The services and intervals shown in the maintenance schedule assumes that you use your
LOGIQ 100 PRO for an average patient load (10-12 per day).
If conditions exist which exceed typical usage and patient load, then it is recommended to increase
the periodic maintenance frequencies.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage173of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage174of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The following is a description of the Standard GE tool kit in the USA. Not all tools are required for PMs.
9-44579 Wrench Open End 1/2 - 9/16 9-XL9974MM Xcelite-hex Blade 2.5mm
9-44579 Wrench Open End 1/2 - 9/16 9-XL9975MM Xcelite-hex Blade 3mm
9-4518 Pliers, Long Nose, Miniature C2356E Screw starter - Kedman Quick Wedge
9-41581 Screwdriver, Blade 3/16 in. x 4 in. 9-45341 Pliers Vice Grip 10 inch
9-39451 20' Steel Tape, locking Spring load 9-3001 Xacto Pen Knife
9-GH807 Ratchet, Offset, Slotted 9-HT62002 Solder Aid, Fork and Hook
9-41584 Screwdriver, slotted 1/4 in.X 6 in. 9-GH300ME Steel Rule Metric 6 inch
9-46677 Hex Keys, 20 pc., Metric 9-XL9924 Xcelite-hex Blade 1/8 inch
9-34701 1/4 in. Standard.Socket set (19 pc) 9-XL9925 Xcelite-hex Blade 5/32 inch
9-43499 1/2 inch Socket 1/4 inch drive 9-XL9926 Xcelite-hex Blade 3/16 inch
ApprovedDocument-53081TPH_r3.pdfPage175of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal
9-39829 Torpedo Level, Magnetic 9-XL87CG Pliers - very fine needle nose-87CG
9-GH408 Tweezers
9-45381 Pliers, Arc Joint 9 1/2 inch 9-44067 Socket 1 1/16 in. for 1/2 in. drive
9-42884 Punch, Pin 5/32 inch 9-4258 3/8 inch to 1/2 inch Adapter
9-42886 Punch, Pin 1/4 inch 9-34374 3/8 inch Metric Socket Set - 12 PT
9-31277 File, Half Round, Bastard 8 inch 9-52068 Tap and Drill Set
ApprovedDocument-53081TPH_r3.pdfPage176of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Leakage Current Ultrasound Kit 2113015 For 120V and 220V Units
46194427P231 Kit includes antistatic mat, wrist strap and cables for 200 to
46194427P279 240 V system
Anti Static Kit 46194427P369 3M #2204 Large adjustable wrist strap
46194427P373 3M #2214 Small adjustable wrist strap
46194427P370 3M #3051 conductive ground cord
46194427P278 120V
Anti Static Vacuum Cleaner
46194427P279 230V
Safety Analyzer 46285652G1 DALE 600 KIT (or equivalent) for electrical tests
E7010GG 60 minute
SVHS VCR Cassette
E7010GF 120 minute
SVHS VCR Head Cleaner See VCR user manual for requirements
B/W Printer Cleaning Sheet See printer user manual for requirements
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
Disposable Gloves
ApprovedDocument-53081TPH_r3.pdfPage17of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The preliminary checks take about 15 minutes to perform. Refer to the system user documentation
whenever necessary.
1 Ask & Listen Ask the customer if they have any problems or questions about the equipment.
Fill in the top of the Periodic Maintenance (PM) Inspection Certificate. Note all probes and system
2 Paperwork
options.
Turn the system power on and verify that all fans and peripherals turn on. Watch the displays during
3 Power up
power up to verify that no warning or error messages are displayed.
ApprovedDocument-53081TPH_r3.pdfPage178of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The functional checks take about 30 minutes to perform. Refer to the system user documentation
whenever necessary.
Step Description
Verify basic B-Mode (2D) operation. Check the basic system controls that affect this mode of
B-Mode
operation.
M-Mode Verify basic M-Mode operation. Check the basic system controls that affect this mode of operation.
System Perform the Automatic Tests, see section 7-2 on page 7-3, to verify that all boards function
Diagnostic according to specifications.
Control Panel Perform the Control Panel Test Procedure, see section 7-2-3 on page 7-9, to verify that all keyboard
controls are OK. This is performed by the internal PC (backend processor) which does a normal
Test
keyboard run through.
Monitor Verify basic Monitor display functions. Refer to Chapter 3 of the User Manual.
Measurements Use the measurement controls to verify distance and area calculation accuracy. Refer to the User
Manual, Chapter 18, for measurement accuracy specifications.
ApprovedDocument-53081TPH_r3.pdfPage179of
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1 VCR Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
VGP
2 Verify hardcopy output of the video graphic printer. Clean heads and covers if necessary.
Printer
B/W
3 Verify hardcopy output of the B/W printer. Clean heads and covers if necessary.
Printer
Unplug
1 Disconnect the mains cable from the wall and system.
Cord
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no
3 Verify
strands may cause a short circuit.
10-5-4 Cleaning
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful
1 Console
not to get the cloth too wet so that moisture does not enter the console.
Probe
2 Clean probe holders. (they may need to be soaked to remove excess gel).
Holder
ApprovedDocument-53081TPH_r3.pdfPage180of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
1 Labeling Verify that all system labeling is present and in readable condition.
Scratches &
2 Inspect the console for dents, scratches or cracks.
Dents
3 Control Panel Inspect keyboard and control panel. Note any damaged or missing items.
Control Panel Verify ease of control panel (Operator I/O Panel) movement in all acceptable directions. Ensure that
4
Movement it latches in position as required.
Cables & Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
5
Connectors special attention to footswitch assembly and probe strain or bend reliefs.
Shielding & Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
6
Covers Missing covers and hardware could cause EMI/RFI problems while scanning.
7 External I/O Check all connectors for damage and verify that the labeling is good.
1 Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).
Thoroughly check the system probe connectors and remove dust from inside the connector sockets
2 Probes
if necessary. Visually check for bent, damaged or missing pins
ApprovedDocument-53081TPH_r3.pdfPage18of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
10-5-6-3 Basic Probe Cleaning
Refer to the Users Manual for details on probe cleaning.
To help protect yourself from blood borne diseases, wear approved disposable gloves. These are
made of nitrile derived from vegetable starch to prevent allergic latex reactions.
Failure to follow the prescribed cleaning or disinfection procedures will void the probes warranty.
DO NOT soak or wipe the lens with any product not listed in the User Manual. Doing so could result
in irreparable damage to the probe. Follow care instructions that came with the probe.Disinfect a
defective probe before you return it. Be sure to tag the probe as being disinfected.
The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC
60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety
evaluation of cord-connected, electrically operated, patient care equipment. If additional information
is needed, refer to the NFPA 99 (For USA) and IEC 60601-1 documents.
THE USER MUST ENSURE THAT THE SAFETY INSPECTIONS ARE PERFORMED AT
LEAST EVERY 12 MONTHS ACCORDING TO THE REQUIREMENTS OF THE PATIENT
SAFETY STANDARD IEC-EN 60601-1. ONLY TRAINED PERSONS ARE ALLOWED TO
PERFORM THE SAFETY INSPECTIONS MENTIONED ABOVE.
To avoid electrical shock, the unit under test must not be connected to other electrical
equipment. Remove all interconnecting cables and wires. The unit under test must not be
contacted by users or patients while performing these tests.
Possible risk of infection. Do not handle soiled or contaminated probes and other components
that have been in patient contact. Follow appropriate cleaning and disinfecting procedures
before handling the equipment.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage182of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Test the system, peripherals and probes for leakage current. Excessive leakage current can cause
injury or death in sensitive patients. High leakage current can also indicate degradation of insulation
and a potential for electrical failure. Do not use probes or equipment having excessive leakage
current.
To minimize the risk that a probe may shock someone the customer should:
Not use a probe that is cracked or damaged in any way
Check probe leakage current:
* once a year on surface probes
* twice a year on endocavitary probes
* whenever probe damage is suspected
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage183of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment
Safety Standards. These limits are GEMS standards and in some cases are lower than the above
standards listed.
Table 10-14 Type BF Applied Part Leakage Current Limits - Non-Conductive (Floating)
Surface and Cavity Probes
Table 10-15 Type CF Applied Part Leakage Current Limits - Surgical Probes and ECG
Connections
*Mains Applied refers to the sink leakage test where mains (supply) voltage is
applied to the part to determine the amount of current that will pass (or sink) to
ground if a patient contacted mains voltage.
The following tests are performed at the factory and should be performed at the site. These tests
are: grounding continuity, chassis leakage current, probe leakage current, and ECG leakage
current. All measurements are made with an electrical safety analyzer Model 600/600E built by Dale
Technology Corporation or equivalent device.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage184of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon
outlet tester and noting the combination of lights that are illuminated. Any problems found should
be reported to the hospital immediately and the receptacle should not be used.
Combination
of
Lights
Note: No outlet tester can detect the condition where the Neutral (grounded supply) conductor and
the Grounding (protective earth) conductor are reversed. If later tests indicate high leakage
currents, this should be suspected as a possible cause and the outlet wiring should be
visually inspected.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage185of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage186of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
10-6-5-1 Definition
This test measures the current that would flow in a grounded person who touched accessible metal
parts of the bedside station if the ground wire should break. The test verifies the isolation of the
power line from the chassis. The meter is connected from accessible metal parts of the case to
ground. Measurements should be made with the unit On and Off, with the power line polarity Normal
and Reversed. Record the highest reading.
Electric Shock Hazard. When the meter's ground switch is OPEN, don't touch the unit!
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged.
Figure 10-4 Set Up for Chassis Source Leakage Current, IEC 601-
1 Clause 19 - Continuos Leakage Currents and Patient,
Auxiliary Currents
When using the Microguard or a similar test instrument, its power plug may be inserted into the wall
outlet and the equipment under test is plugged into the receptacle on the panel of the meter. This
places the meter in the grounding conductor and the current flowing from the case to ground will be
indicated in any of the current ranges. The maximum allowable limit for chassis source leakage is
shown in Table 10-13.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage187of9
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
10-6-5-3 Dale 600 Meter Procedure
When measuring system chassis currents with the Dale 600, always use the CHASSIS selection of
the external/chassis function switch. This requires the ground clip lead and changing the meters
switches in accordance with the IEC 601-1.1. Refer to the Dale 600 Instruction Manual for meter
self-test and operation. Record the highest leakage current measured.
Follow these steps to test the unit for leakage current.
1.) Turn the LOGIQ 100 PRO unit OFF.
2.) Plug the unit into the meter, and the meter into the tested AC wall outlet
3.) Plug the black chassis cable into the meter's CHASSIS connector and attach the black
chassis cable clamp to an exposed metal part of the LOGIQ 100 PRO.
4.) Set the tester's FUNCTION switch to CHASSIS position.
Outlet Test
Lamps
Functions Normal
Selector Reverse
Polarity
Neutral
Open/Closed
Chassis
Cable
Clamp
TEST CONDITION
2 Wheel support
6.) Keep a record of the results with other hard copies of PM data kept on site.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage18of97
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
10-6-5-4 Data Sheet for Chassis Source Leakage Current
The test passes when all readings measure less than the value shown in Table 10-13. Record all
data on the PM Inspection Certificate.
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
Table 10-17 Typical Data Sheet for Chassis Source Leakage Current
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage189of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
10-6-6-1 Definition
This test measures the current that would flow to ground from any of the probes through a patient
who is being scanned and becomes grounded by touching some other grounded surface.
Note: Each probe will have some amount of leakage current, dependent on its design. Small
variations in probe leakage currents are normal from probe to probe. Other variations will
result from differences in line voltage and test lead placement.
6.) Put the saline probe and the probes face(imaging area of the probe) into the saline bath.
CAUTION To avoid probe damage and possible electric shock,do not immerse probes into any liquid
beyond the level indicated in the probe users manual.Do not touch the probe, conductive
liquid or any part of the unit under test while the LIFT GROUND switch is depressed.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage190of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ApprovedDocument-53081TPH_r3.pdfPage19of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
10-6-6-4 Meter Procedure using Probe Adapter to measure Probe isolation(Sink) Current
The Dale 600/600E provides a method for testing probes independently from the system.The meter
utilizes a probe adapter to apply a test potential commonly to all connector pins.
The probe's imaging area is immersed in a saline solution along with a grounding probe from
the meter to complete the current path.Saline solution is a mixture of water and salt. The salt adds
a free ion to the water, making it conductive. Normal saline solution is 0.9% salt or 1/2 gram salt per
1 litre of water. If saline is not available, a mixture of 1 quart of water with one or more grams of
table salt, mixed thoroughly, will substitute.
CAUTION To avoid probe damage and possible electric shock, do not immerse probes into any liquid
beyond the level indicated in the probe users manual.Do not touch the probe,conductive
liquid or any part of the unit under test while the LIFT GROUNDswitch is depressed.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
8.) Have unit power ON for the first part; turn it OFF for the second half.
9.) Depress the ISO TEST (Dale 600) or Mains on Applied Parts(MAP) (DALE 601) ROCKER
SWITCH and record the highest current reading.
10.) Follow the test conditions described in Table 10-18 for every probe.
11.) Keep a record of the results with other hand copies of Planned Maintenance data.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage192of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged
.
Probe Tested:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
Table 10-18 Typical Data Sheet For Probe Source Leakage Current
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage193of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
PROBE FAILS
Test the probe in another connector to isolate if the fault lies with the probe or the scanner.
Each probe will have some amount of leakage, dependent on its design. Small
variations in probe leakage currents are normal from probe to probe. Other
variations will result from differences in line voltage and test lead placement. The
maximum allowable leakage current for body surface contact probe differs from
inter-cavity probe. Be sure to enter the correct probe type in the appropriate space
on the check list.
If excessive leakage current is slot dependent, inspect the system connector for bent pins, poor
connections, and ground continuity.
If the problem remains with the probe, replace the probe.
PERIPHERAL FAILS
Tighten all grounds. Ensure star washers are under all ground studs.
Inspect wiring for bad crimps, poor connections, or damage.
STILL FAILS
If all else fails, begin isolation by removing the probes, external peripherals, then the on board ones,
one at a time while monitoring the leakage current measurement.
NEW UNIT
If the leakage current measurement tests fail on a new unit and if situation can not be corrected,
submit a Safety Failure Report to document the system problem. Remove unit from operation.
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage194of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
PM INSPECTION CERTIFICATE
Customer Name: System ID: Dispatch Number / Date Performed: Warranty/Contract/HBS
* Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other
Measurement Accuracy
GE Approved Peripherals
COMMENTS:
________________________________________________________________
________________________________________________________________
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
________________________________________________________________
________________________________________________________________
________________________________________________________
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage195of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
ELECTRICAL SAFETY
PROBES
Probe 1:
Probe 2:
Probe 3:
Probe 4:
Probe 5:
Probe 6:
Probe 7:
Probe 8:
Probe 9:
Final Check. All system covers are in place. System scans with all probes as expected.
Accepted by: ______________________________________________________
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage196of7
GE HEALTHCARE
DIRECTION 5305880-100, REVISION 3 LOGIQ 100 PRO SERVICE MANUAL
Your Notes:
State:RELEASE-DocumentisrladfoChangeControl.ChangesrubjctohECR/ECOProces.
SeethGEHCMyworkshpSystemodrinhaufc.
ApprovedDocument-53081TPH_r3.pdfPage197of