E 801 - 06 (2011)
E 801 - 06 (2011)
E 801 - 06 (2011)
Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
E801 06 (2011)
NOTE 1Use additional layers of shim material as required. The layers shall be 1 by 1.625 in. (25.4 by 41.275 mm). The additional shim material shall
be identified by the placement of lead numbers which denote the thickness immediately adjacent to the IQI during exposure, or as an alternative,
documented on the radiographic technique.
NOTE 2Tolerance is 60.001 in. (0.025 mm) where dimensions are 0.000 and 60.003 in. (0.076 mm) where dimensions are 0.00.
NOTE 3Bond materials together with cyanoacrylic or equivalent fast-drying epoxy.
NOTE 4Particle holes are 0.031 in. (0.787 mm) nominal diameter.
NOTE 5Tolerance on particle diameter is + 0.0003 in. (0.0076 mm).
NOTE 6Wire grooves are 0.007 in. (0.178 mm) depth with 90 inclusive angle.
NOTE 7The number hole is 0.25 in. (0.635 mm) nominal diameter and 0.125 in. (0.318 mm) deep.
FIG. 1 Image Quality Indicator for Electron Devices
higher power devices that have thicker coverings normally use thin coverings. Particle size is normally independent of device
larger interconnecting wires than small signal devices that use type, so these remain constant.
2
E801 06 (2011)
7. Positioning of IQIs
7.1 The IQIs shall be placed so that the plane of the IQI is
normal to the radiation beam.
7.1.1 The IQIs used for each exposure or field of view shall
be positioned at the outer edges of the film or field of view but
not impeding the area of interest being examined.
8. Number of IQIs
8.1 Two IQIs shall be used for each radiograph as shown in
Fig. 3. The IQIs shall be located at diagonally opposite corners
of the film as shown in Fig. 3. In radioscopic examinations, IQI
qualification is required at least once during examination of
similar parts. The procedure for radioscopic IQI qualification
shall be established to subject the IQIs to the same image
NOTE 1Permanently etched or engraved serial number.
NOTE 218 in. lead markers corresponding to serial number.
magnification and distortion as the parts will be subjected to
FIG. 2 Arrangement of IQI and Lead Serial Number for Calibration during inspection and the procedure shall be documented on
Image the radioscopic scan plan/technique.
8.1.1 The radiographic density or grey level of the IQIs
shall bracket the radiographic density or grey level of the
6.2 The IQI used shall be the one with a radiographic devices being examined. Additional shim(s) may be used as
density or grey level nearest to that of the electronic device described in 6.2 to achieve this requirement.
being examined. The radiographic density or grey level is
measured on an area of the IQI image that contains no wire or 9. Part Identification
particle images. It shall be measured using a calibrated 9.1 In radiography, the image of identification markers for
densitometer in the case of film, or by pixel grey level value in the coordination of the part with the film shall appear on the
the case of radioscopic examination. When necessary to meet film, so as not to interfere with the interpretation of the
the requirements of 8.1.1, additional shim stock shall be used. radiograph and so that it is evident that the complete part order
The shim(s) shall be of the same type of steel as in the IQI and was obtained. These marker positions shall coincide with the
shall be placed under the IQI. The additional shim(s) shall part serial number.
exceed the IQI outside dimensions by at least 0.125 in. (3.18
9.2 In radioscopy, the image header for each part image
mm) on at least one side. The thickness of the shims shall be
shall include a part serial number for traceability and to ensure
noted with lead numerals placed adjacent to the shimmed IQI
that the complete part order was obtained.
or documented on the radiographic technique.
NOTE 2Radiographic film densities shall be measured with a cali-
10. Identification of the Image
brated densitometer. When films are exposed simultaneously in one film 10.1 In both radiography and radioscopy, a system of
holder, density variations should be determined on the single or combined positive identification of the image shall be provided in
films, depending on the manner in which they are read and interpreted.
Radiographic film density shall be in the range of 1.0 to 3.5, unless
addition to Section 9. Any or all of the following may appear:
otherwise specified. the name of the examining laboratory, the date, the part
number, the serial number, the date code, the view, and whether
6.3 The image distortion shall not exceed 10 % unless a
original or subsequent exposure.
higher or lower percentage is agreed upon between the
purchaser and the supplier. Measurements for computation of 11. Multiple-Film Techniques
image distortion shall be made in the grid formed by the 11.1 Film techniques with two or more films of equal or
crossing wires. different speeds in the same holder will be permitted, provided
6.3.1 Determine percent image distortion (Note 3) as fol- that the appropriate wires and particles in the IQI for a specific
lows: area are demonstrated.
D 5 @ ~ S M 2 S A ! /S A # 3 100
12. Non-Film Techniques
where:
12.1 The use of non-film imaging techniques will be
D = percent distortion, permitted, provided that the applicable IQI wires and particles
SM = wire spacing as measured on the radiograph, and are demonstrated in the resultant image.
SA = actual wire spacing.
NOTE 3Distortion may be positive or negative. 13. Image Quality
6.4 Particles G through L shall be visible in the image when 13.1 The radiographic image shall be free of blemishes that
IQIs 1 through 7 are used and particles G through K shall be interfere with its interpretation.
visible in the image when IQI number 8 is used.
6.4.1 If shims are required, the number and size of particles 14. Source-Film Distance
that can be resolved shall be agreed on between the purchaser 14.1 Any source-film distance will be satisfactory provided
and the supplier. that the required IQI quality level is attained.
3
E801 06 (2011)
15. Records the grids formed by the cross wires and the parallel variances
15.1 Complete records of the radiological technique or scan within the grid. It shall also state the actual size of each
plan details, or both, shall be maintained by the examining particle.
laboratory.
16. Keywords
15.2 A calibration image shall be made by exposing the
individual semiconductor IQI at the proper distance from the 16.1 electronic devices; IQIs; radiography; radioscopy;
focal spot. The IQI shall be centered in the beam during this shim; wire
exposure. The certificate shall show the actual dimensions of
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/