Laser Light Instructions
Laser Light Instructions
Laser Light Instructions
LightLas 532
Green Laser Photocoagulator With LCD
Control Panel
LightLas532 Operators Manual Rev. No 01 Page 1of 115
Operators Manual
for the
LightLas 532
Green Laser Photocoagulator With LCD
Control Panel
Directive 93/42/EEC
as amended by 2007/47/EC
Doc. No. : DC1001
Rev. No. : 01
CAUTIONS - Any modification to the Ophthalmic Laser will result in the necessity for
it to be reclassified
CAUTIONS - U.S. law restricts this device to sell by or on the order of a physician
Document Title: Operators Manual for the LightLas 532 Ophthalmic Laser
Document Number: DC1001
Document Revision History:
DRAFT March 2014 Draft prepared
01 August Release 01 version
Record Sheet
Appendix I
The Manufacturer and Distribution organization assume no liability through the use
of this Laser system.
All care has been taken in the preparation and checking of this manual however there
is no guarantee provided that all information is correct. The information provided in
this manual is subject to change without notice.
Only approved or authorized accessories may be used in the LightLas 532. The
Manufacturer and Distribution organization shall not be held liable or responsible for
damages or injury caused as a result of using non-approved accessories. This includes
all Optical Fiber systems, Laser Delivery Units, Safety Filters, Safety Glasses and Table
units.
All maintenance and service work must be carried out by authorized and trained
service agents and only those procedures outlined in the operator and service
manual are allowed. Any service work carried out by unauthorized persons will void
all warranties.
No circuit diagrams or component part lists are to be supplied for the LightLas 532. If
you require technical documentation that is not provided in this manual then please
contact the manufacturer or your local distributor in writing with your reasons for
wanting them and then a copy of the service manual may be provided.
Before using the LightLas 532 Photocoagulator Laser system the operator should
read this manual carefully and pay particular attention to the sections of Safety,
Operation and Maintenance.
The LightLas 532 has been designed to comply with the following standards:
Laser standards
IEC 60825-1 (2007)
USA 21 CFR 1040.10, 1040.11 (1997)
IEC60601-2-22 (1995)
Electrical standards
IEC 60601-1:2005
EN 60601-1:2006
EN 60601-1-2:2007
IEC60601-1-2 :2007
USA UL 2601
JIS T1001 (1992) and T1002 (1992)
Others
MDD 93/42/EEC (as amended by 2007/47/EC)
EN 60601-1-6:2010
IEC 60601-1-6 :2010
DO NOT use the Laser Console if the ambient temperature is outside the
range of 20 to 35C. This temperature range is the rated operating
temperature limits where the Laser system can be guaranteed to operate
without any interruptions to normal use. Outside this range of temperature
it is possible that the Laser will generate an error condition where the word
hold is displayed and the system goes to Standby until the internal
temperature returns to within normal limits then the Laser can be used again
but the error condition may reoccur unless the rated temperature comes
within limits.
LIGHTMED
LIO
LIO Delivery
Key
STOP
STOP
FOOTSWITCH DOOR
INTERLOCK
WARNING
Risk of fire
Replace only with
fuse as marked above
LASER RADIATION
AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
LASER RADIATION CLASS 4 LASER PRODUCT
AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
CLASS 4 LASER PRODUCT Green Laser 532nm 2.0W CW Max
Green Laser 532nm 2.0W CW Max Diode Aiming Laser 635-650nm 1mW CW Max
Diode Aiming Laser 635-650nm 1mW CW Max
Figure 2.1
Laser Console Safety and Control Labels
Lightmed Corporation
Lightmed Corporation
MODEL NO:
MODEL NO:
SERIAL NO: SERIAL NO:
Coiling
Lightmed Corporation
MODEL NO:
SERIAL NO:
FOOTSWITCH DOOR
INTERLOCK
WARNING
Risk of fire
Replace only with
fuse as marked above
Figure 2.2
Laser Console Safety Labels
CAUTION/DANGER
VISIBLE LASER RADIATION
AVOID EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED
RADIATION
CLASS 4 LASER PRODUCT IEC/EN 60825-1:2007
Figure 2.3(a)
Integrated Slitlamp LDU with labels (RH side)
DANGER / CAUTION
Visible laser radiation when open
AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
0434
Lightmed Corporation
Safety Filter Model No:
OD4@532nm Serial No:
DANGER / CAUTION
Visible laser radiation when open
AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
Figure 2.3(b)
Integrated Slitlamp LDU with labels (LH side)
Serial No:
Manufactured:
Lightmed Corporation
Safety Filter OD4@532nm
Model No:
Serial No:
Manufactured:
CAUTION/DANGER
VISIBLE LASER RADIATION DANGER / CAUTION
AVOID EYE OR SKIN EXPOSURE Visible laser radiation when open
AVOID EYE OR SKIN EXPOSURE TO
TO DIRECT OR SCATTERED DIRECT OR SCATTERED RADIATION
RADIATION
CLASS 4 LASER PRODUCT IEC/EN 60825-1:2007
Figure 2.5
Attachment LDU with labels
Lightmed Corporation
Safety Filter OD4@532nm 500 300 400 500
Model No:
Serial No:
Manufactured:
Lightmed Corporation
Safety Filter OD4@577nm
Model No:
Serial No:
Manufactured:
CAUTION/DANGER
VISIBLE LASER RADIATION
AVOID EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED
RADIATION
CLASS 4 LASER PRODUCT IEC825-1(1993)
CAUTION/DANGER
VISIBLE LASER RADIATION
AVOID EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED
RADIATION
CLASS 4 LASER PRODUCT IEC60825-1 2007
Figure 2.6
TruSpot Attachment LDU Labeling
Model No:
FIL
Lightmed Corporation FILTER
Manufactured: Safety Filter Model No:
OD4@532nm Serial No:
Figure 2.7
Microscope Safety Filter with labels
General Specification
Electrical Input : 100 to 230 Vac. 50/60 Hz Single phase
Power : 400W
Fuse rating : T3.15AH250V @ 100-230Vac (Time Lag)
Temperature Range : Transport: -10 to 70C
Operating: 15 to 30C
Storage: -10 to 55C
Relative Humidity Range : Operating: 30% - 85% non-condensing
Storage and Transport: up to 95% non-condensing
Atmospheric pressure : Operating: 800-1060 mbar
Storage and Transport: 500-1060 mbar
Cooling System : Fan cooled and TECs for Laser Diode and Crystal
Dimensions (Total) : 130mm(H) x 370mm(W) x 330mm(D)
Weight : 13 Kg (System) 20 Kg (Packed)
Treatment Laser
Laser Type : Diode Pumped Frequency Doubled YAG
Wavelength : 532 nm
Mode of Operation : CW
Power Output : 2 W Maximum
Power Adjustment : Variable from 0.05 to 2.0 W
Exposure Duration : 0.01, 0.02, 0.03, 0.04, 0.05, 0.06, 0.08, 0.1, 0.15 and
0.2 to 3.0s is 0.1s increment
Repeat Interval : Selectable from 0.01 to 3.0 secs and equal to or greater
than exposure duration in same discrete steps as
duration
SP Mode : Duty cycle selections - 7.5%, 150s On time
- 10%, 200s On time
- 12.5%, 250s On time
- 15%, 300s On time
- 17.5% ,350s On time
- 20%, 400s On time
Endo-ocular Probes
Probe Types :
Straight, Curved and Aspirating
Fiber Length 3m :
Safety Filters :
OD 4 @ 532nm in housing for installation to Operating
Microscope before Laser can be fired
Sterile : Sterilized by Ethylene Oxide and a single use
device only
Beam Divergence : 0.2 NA
___________________________________________________________________
Accessories
Safety Filters
Types : To suit Zeiss, Moeller, Leica and Topcon Operating
Microscopes
Safety Filter : OD4@532nm
The following are the Slitlamp Specifications for the CSO SL980 and SL990 Slitlamps
that are used in the LightLas 532 Photocoagulator Laser System. The SL980 is a Zeiss
clone and the SL990 is a Haag Streit clone. Both Slitlamps have very similar
specifications however the SL980 uses an illumination source below the viewing path
and the SL990 uses illumination from above the viewing path.
Microscope : Galilean
Magnification Set : 5 Step Drum Rotation
Eyepiece : 12.5X
Magnification Ratio : 6X, 10X, 16X, 25X, 40X
PD Range : 48.5-80mm
Diopter Adjustment : +/-8
Slit Illumination : 6V 20W Halogen Lamp
Slit Width : 0-14mm (SL980) and 0-12mm (SL990)
Slit Length : 1.8 12mm
Slit Apertures : 0.3, 5.5, 9, 14mm(SL980) and 0.2, 1, 3, 5, 9, 12mm(SL990)
Slit Angles : 0- 180
Filters : Red Free, Heat Absorbing, Cobalt Blue
Movement Ranges
Longitudinal (In/Out) : 113mm
Lateral (Left/Right) : 108mm
Vertical (Up/Down) : 35mm
Fine movement range : 10mm
Chin Rest Range : 70mm
The LightLas 532 Laser system has a wavelength of 532nm, which is in the visible
spectrum and is a green light. A red aiming beam is used to position the treatment
green light beam prior to delivery.
The word LASER is an acronym for Light Amplification by Stimulated Emission of
Radiation. The light from a laser has particular characteristics, which makes it a
valuable tool for medical applications.
The beam from a laser is collimated which means that the beam does not
diverge and can maintain a constant diameter over a long distance. This
means that the Laser beam can be focused to a very small spot with high
energy and power densities.
The beam is Monochromatic, which means that it is a single wavelength
beam and therefore the effects of the beam on tissue are very predictable
and reproducible.
The light waves are coherent which means they are in phase with each
other and do not interfere and generate losses in energy.
The LightLas 532 system consists of a laser console where the green laser is housed
along with the electronic control system and power supplies and accompanies along
with various Laser Delivery Units (LDUs). These LDUs include:
Slitlamp Integrated into CSO model SL980
Slitlamp Attachment for CSO model SL990 and other Haag Streit clones.
Slitlamp Attachment for Zeiss model SL30 Slitlamp
Laser Indirect Ophthalmoscope (LIO) using a Keeler II
Endo photocoagulation hand pieces (Endoprobes)
When using these LDUs a microscope Doctor Safety Filter (DSF) is required to
protect the doctor from unexpected reflections causing eye injury during the
treatments. The DSF is mounted in the beam path of the microscope.
The following paragraphs give a general description of the operation of the Laser
System
The laser system console generates a controlled beam of the 532nm wavelength light
that is focused to a small spot so that it can be delivered into an optical fiber that
then connects to one of the Delivery units. The Slitlamp LDUs optical fiber has a
diameter of 200m and is 2 meters long and 3 meters long for the LIO. Special care
must be taken with the fibers not to damage the jacket as this may create extra
losses and may allow the laser beam to be transmitted at the damaged place along
the fiber. Therefore the fiber should be kept off the floor and away from sharp edges.
The 532nm wavelength green laser light is primarily used as a source of energy to
heat the tissue and thereby cause photocoagulation. The laser beam is directly
applied to the treated tissue and absorbed by the melanin pigment within the retinal
pigment epithelium and the choroid. This absorption converts the light energy into
The system consists of two major parts: the console and the LCD touch control panel
integrated with computer platform
Inside the Laser Console there are several operating components that put together
to provide the output Laser beam such as:
Laser Diode (808nm)
Laser Cavity and optics system
Thermal Electric Coolers and Driver units
Electronic Microprocessor control system
Power supply
First of all, main source power must be connected to the laser console system before
the system is enabled to function accordingly. Secondly, a blue LED backlight power
switch, located at the bottom of the screen (refer to fig. 4.3), on the LCD control
panel display needed to be enabled and wait for system software to boot up before
proceed to the next procedure (refer to fig. 4.1 - 4.2). Once the software is properly
boot up, then the key-switch is inserted and turned to the ON position (with the
Emergency switch in the out or OFF position). And then the console system
microprocessor controller will perform some internal checks to verify that the
machine is functioning as it should be. Few warning messages such as, "BBF
Temperature Not Ready", "LBO Temperature Not Ready" ...etc. will display on the
LCD when the console power is ON. This process usually takes less than few
seconds. If the temperature setting up process time is out of specification (> 5mins),
there will be an error code shown on the LCD display (for more detailed refer to
troubleshooting section of this manual).
Laser Power
Laser Power Pulse Duration
Pulse Interval
Accumulated number of pulses
The type of LDU connected
Aiming intensity
Mode of operation
SP Mode Selection
Customized treatment configuration
Query or Help functionality
The mode of operation is an important function display because when the laser is
turned ON STANDBY mode is automatically selected, which prevents any accidental
firing of the LightLas 532 Laser System with LCD panel. In STANDBY mode, the
footswitch is disabled and the shutter module blocker will obstruct the beam path is
closed.
Only Standby button switch is toggled, the system will be in treat mode. Then the
footswitch and shutter are now enabled and the aiming beam is turned on. If the
footswitch is pressed, the Green laser beam will be delivered into the fiber. The
system will turn back to STANDBY mode in the following situations:
No controls are operated for 10 minutes
Any warning or error condition occurs
Prior to activate the Treat mode, it is recommended that all operating conditions are
to be set correctly such as patient positioning, power selection, pulses duration,
interval duration, aiming beam intensity, spot size and illumination intensity. This will
prevent the likelihood of accidental firing of the Green laser or unintentional delivery
during the set up stages.
Output power distribution can fine tune through an up/down arrow switch button on
the LCD touch control panel. The power setting will remain the same whenever the
power is on which means the default setting screen will be primary unless it is
replaced by another setting. The power can be adjusted from 50 to 2000 mWatts.
The pulse duration can be adjusted from 0.01 secs to 3.0 secs. by pressing the action
key switch buttons on the LCD touch control panel. Similarly, the Repeat Interval
LCD Power
On / Off
Button
LIGHTMED
LIO
LIO Delivery
Key
STOP
FOOTSWITCH DOOR
INTERLOCK
WARNING
Risk of fire
Replace only with
fuse as marked above
Figure 4.4
Laser Console Controls and LCD Displays
Legend
1. Emergency Stop Switch 2. Key Switch
3. Laser Aperture 4. Delivery Key Connector
5. LIO Illumination Control 6. LCD Control Panel
7. LIO Power 8. Main Power Inlet
9. Footswitch Connector 10. Remote Interlock Connector
11. LCD Control Panel Power Switch 12. Power Control Adjuster or Dial
Figure 4.5
LCD Panel Display and Control
Legend
13. Mode Indication (TREAT / STANDBY) 14. Type of Delivering System Indicator
15. User Configuration Indication 16. Date and Time Indicator
17. Total Laser Shot Counter Display 18. Counter Reset Switch
19. Aiming Beam Intensity Indicator Display 20. Aiming Beam Intensity Setting Switches
21. Manual Inquiry 22. Loading Profile Inquiry
23. Saving Inquiry 24. Repeat Interval Setting Switches
25. Repeat Interval Display 26. Exposure Duration Setting Switches
27. Exposure Duration Display 28. SP Mode & Selection Display Icon
29. Preset Power Display 30. SP Mode Selection Icon
31. Buzzer Control Icon
Note: There are cooling fans inside of the Laser Console. This forced air is
used to cool down the electronics component and specially the laser diode and
cavity in the system. Therefore it is very important that the intakes cooling fans
are not blocked otherwise the laser will overheat and an error condition will
generate.
CAUTION/DANGER
VISIBLE LASER RADIATION
AVOID EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED
RADIATION
CLASS 4 LASER PRODUCT EN60825-1(1994)
Figure 4.6
Integrated Slitlamp LDU Controls
Fig. 4.7b
Truspot Attachment Slitlamp LDU Controls (New version)
Slitlamp LDU system can be integrated with two types of the slitlamp that carry by
Lightmed are as:
10
3
2
This control is used to filter out application required by physician or user. There
are three types of filters such as, Red Free, Heat Absorbing, and Cobalt Blue.
The CSO Slitlamps can be used as a standard diagnostic tool and all the standard
Slitlamp functions are available.
The following figures 4.6 and 4.7 show all of the controls that can be found on the
Slitlamps and the description of the components.
Figure 4.10
SL990 Slitlamp Parts List and Controls
Figure 5.1
Packing Carton for Integrated LDU and Slitlamp (Optional)
LIGHTMED
Upon receiving the system inspect the packing carton for any signs of mishandling,
which must be reported to the freight handler before the instrument is unpacked. If
there is damage, then the manufacturer reserves the right to decline any warranty
claims that may be forthcoming so it is essential that the freight company will take
full responsibility for any damages.
If the outer cardboard packaging looks OK then you can proceed to remove the
internal packed assemblies (refer to Fig. 5.1). The contents of each layer are:
1. The top layer is just a protection layer.
2. The middle layer contains the Chinrest and Integrated LDU including the
Zoom assembly, binoculars, eyepieces, target rod, target plate, cross-slide
shaft and gears, gear covers, spare lamp, manual, mains power cable.
3. The bottom layer part contains the table top.
The packing checklist is used to confirm the individual layer contents and notify the
manufacturer if any discrepancy. The motorized table is an optional and may be
purchased from the manufacturer.
1. Laser Console
2. Integrated Laser Delivery Unit Microscope assembly with Zoom unit
3. LIO Assembly
4. Fiber Delivery cable
5. Endoprobe Plug
6. Endoprobe
7. 4 Different Delivery Key Connectors
8. Attachment Laser Delivery Unit (New One without Micromanipulator-Truspot)
9. Microscope Laser Safety Filter units
10. Footswitch (including a Signal Receiver and a Footswitch Cable)
LIGHTMED
1 LIO
2
LIO Delivery
Key DANGER / CAUTION
Visible laser radiation when open
STOP AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
10
4
WARNING
Only use with 532nm Laser
9 8 7 6
Figure 5.3
Laser System parts
LDU Systems
5.4.1 Slitlamp Integrated LDU
Prior to install the Slitlamp Integrated system, ensure all the corrected setup and
equipment are in placed to proceed the installation steps.
System Part(s):
Upper Arm Housing
Lower Base Housing
Table top
Accessories (binocular, cover, chinrest, and target rod)
Procedures:
1. Unpack all the packaging items from the cartoon box
2. Lay the ready to go table on the final destination location (well level) and ensure all
the accessories are installed including rail and joystick slide pad (refer to fig. 5.4)
3. Ensure the voltage and fuses rate setting are set up correctly, as default setting is
230VAC (refer to fig. 5.5a-d)
4. Take the Slitlamp upper arm housing and lower base housing with cross-slide shaft
assembled (refer to fig. 5.6)
5. Assembled the upper and lower housing together by securing the screw at rear
end of lower housing (refer to fig. 5.7)
6. Position the Slitlamp assy back to the gear rail and gently wheel it back and forth
to get it smooth out
7. Ensure the Slitlamp assy is horizontally parallel in x-axis
8. Place the gear cover
9. Install the Chinrest assy and connect the fixation lamp to the source box and plug
in the source power connection (refer to fig. 5.5a)
Figure 5.5 (c) Removal of fuses and check the fuses rate
Magnifier Assy or SD pc
Binocular
Figure 5.8
Slitlamp mounted onto Table Top ready for Integrated LDU
Figure 5.12 Align the keyways then fit the Zoom unit to the Delivery housing
Prior to install the Attachment Slitlamp delivery unit, ensure all the corrected setup
and equipment are in placed to proceed the installation steps.
System Part(s):
Tonometer Mounting block (Haag Streit style / Zeiss 30SL / SYL9000)
Attachment arm (Haag Streit style / Zeiss 30SL / SYL9000)
Mounting screws (Haag Streit style / Zeiss 30SL / SYL9000)
Safety Filters (only for combination system)
Micromanipulator joystick
New Modified Illumination Tower (Zeiss 30 SL type only)
Procedures:
1. Place the appropriate mounting post into the magnifier assy and secure with the
correct mounting screw (refer to fig. 5.15a-c)
2. Swing the slitlamp tower aside prior to fit the appropriate whole attachment body
into the mounting block (refer to fig. 5.16a-c)
If this is the Zeiss 30SL attachment LDU type (refer to fig. 5.14a-b), a minor
modification required prior move on to the next step:
2a. Place a target rod
2b. Turn on the slitlamp and set aperture setting to the smallest
2c. Mark on the aperture spot size with pencil
2d. Remove the old illumination tower and replace with newly modified one
2e. Verify the new tower against the mark on target rod aperture spot and align
and lock if it is confirmed
2f. Align the spot position by adjusting the mirror if required
Prior to install the LIO delivery unit, ensure all the corrected setup and equipment
are in placed to proceed the installation steps.
Accessories:
LIO fiber
LIO Delivery System set
Procedures:
1. Take a LIO delivery fiber and remove the fibers protector one end at a time
and clean it with alcohol prior to install to the console aperture and LDU (refer
to fig. 5.17)
2. Plug in its delivery key and light source power connector (refer to fig. 5.18)
3. Make sure the fiber and the cable are fitted to the strain relief fittings on the
LIO headpiece (refer to fig. 5.19)
Figure 5.17 Fiber tips cleaning Figure 5.18 LIO connections in console
The endoprobe is sterilized single use device therefore there will be no installation
procedures involved however, do want to ensure that the endoprobe packaging is
not being opened prior to the treatment.
Ensure Endoprobe plug is fitted prior to load to console laser aperture (refer to fig.
5.20)
Fiber
2
Fiber probe
Fiber Nut
Eyes Lockin
g
Accessories:
Power cord (110 or 220 VAC type)
Delivery key (Slitlamp / LIO / Endoprobes)
Footswitch (wire or wireless)
Optical fiber (link from console aperture to the LDU could be Endoprobe
or standard delivery one)
Interconnector (Y-modified combination connector / treatment door
switch / default dummy connector)
On/Off Switch Key
Procedures:
1. Position the laser console onto the mounting plate and to the right side of the
Slitlamp on the table top
2. Connect or fit all appropriate connectors (power cord, delivery key, interlock,
footswitch, key switch, and Endoprobe or standard delivery fiber)
2a. If this interlock is Y-modified connector, ensure the safety filter and YAG
interlock connections are well fitted (refer to fig. 5.21b)
3. Ensure all the connections are fitted no leftover since the improper connection or
no connection will result the system not functioning properly (refer to fig. 5.21a-c)
Note: Be extreme careful about the delivery fiber connection. The delivery fiber with
the large fiber holder end with delivery key attached is connected to the
console aperture and the another end is connected to appropriate delivery
system (refer to fig. 5.21c)
Take special care when preparing the fiber to not stress the cable. Hold the
connector when removing the protector caps from each end. Never pull on the
cable or the cable sleeve as the fiber may be damaged
Figure 5.21 (c) Sample of Delivery Fiber and Delivery Key Connection
Procedures:
Figure 5.22(a) System boot up sequential display #1 Figure 5.22(b) System boot up sequential display #2
Figure 5.22(c) Power on console & system is initializing Figure 5.22(d) System is ready to go
temperature process
Figure 5.24
Attach the Delivery fiber to the Zoom unit
Prior to pre-check and realign the integrated Slitlamp LDU, ensure all the corrected
installation or setup are in placed (including eye piece dioptre setting, fiber
connection, and console setup and on) prior to proceed the following steps.
Slit Integrity Checkout
1. Insert the Slitlamp target rod into the Slitlamp or set up a target to view the
Illumination spot
2. Set a narrow slit width with low intensity illumination and verify that the slit is
focused and that the two slit lines are aligned together for the Integrated
Delivery Unit Slitlamp. If not align, proceed to following steps:
2a. Unscrew the upper slit alignment screw slightly or lower slit alignment
depending on your preference (refer to fig. 5.25)
2b. Choose one end as your starting reference and realign the another
Note: Do not over loosen the alignment screws completely specially the upper slit
due to the slit prism module may fall off and break prism mirror
2c. Confirm it by illuminating the aperture to see if you have aligned them ok
2d. Process to the next step if it is aligned
Lower Alignment
X-axis Alignment
Y-axis Alignment
Prior to pre-check and realign the attachment Slitlamp LDU, ensure all the corrected
installation or setup are in placed to proceed the following steps.
Slit Integrity Checkout
1. Set a narrow slit width with low intensity illumination and verify that the slit is
focused. If the Slitlamp focus is OK then set it to a smaller spot size aperture and
observe the red aiming beam.
Aiming Integrity and Focus Checkout
2. Set spot size to the smallest by adjusting the zoom pc
3. Power on the console and toggle the mode to Treat mode
4. Verify if aiming round spot is reflected on the target rod
5. Adjust the spot size from 50 to 500 by rotating the zoom pc and verify if the spot
size is consistently enlarged from smallest to largest
6. Adjust Mounting arms black knob (old type version) or the Delivery Main Body
locking screw (new type version) to align the focus correctly and then secure the
screw tightly (refer to fig. 5. 28a-b)
7. Remove the target rod and place a piece of black tape or thermal paper on the
chinrest
8. Verify the slit and aiming beam focus are in focus together
Micromanipulator Joystick Movement Checkout (applicable only to old version type)
9. Place back the target rod and verify the micromanipulator joystick is moving within
the full range of the field view and in the center portion when illumination spot is
set at largest (Applicable only to old version type refer to fig. 5.29)
10. Ensure the illumination spot is reasonable in the center portion of field of view. If
not refer to below steps for fine tune-up:
10a. Loosen the mirror nut that clamp to Main body (refer to fig. 5.30) and low
or raise will center the beam correctly and retighten when achieved or
10b. Loosen the set screw on the side of Tonometer using 1.5 mm allen key and
adjust the rod on the Tonometer mount post (refer to fig. 5.30) in aside
way to achieve the centering beam and retighten when achieved the
position
10c. Ensure the set screw is tighten and the rod is removed (keep it in a safe
location) upon completion
Set screw
Y-axis Adjustment #1
LIO Intensity
Control Knob
Aiming spot
light up/down
Figure 5.31 (b) LIO Intensity Control Knob
5.5.2.1 Endoprobes
The Endoprobes are sterilized single use devices so it is not possible to do any testing
of them at installation time. It is recommended that if the doctor is to use
Endoprobes that one unit be used as a reference for checking the Console operation
and for Calibration testing.
2. Slitlamp Checks
2.1 Illumination Lamp Controls OK.
2.2 4 filter wheel settings check OK.
2.3 With Slit fully open 5 Aperture settings OK.
2.4 With Large Aperture Set, Slit fully open get full circular spot.
2.5 Slit is fully adjustable to closed position.
2.6 When slit is closed there is no Illumination showing.
o
2.7 Set narrow slit and check slit rotation + 90 around vertical.
2.8 Load the target rod and set up the correct individual's ocular.
2.9 Check that eyepieces fit firmly into binoculars and can be fully adjusted.
2.10 Binoculars can be adjusted for interpupillary distance (PD).
2.11 Check Slitlamp movements with joystick:
Vertical adjustment range OK.
Sideways adjustment range OK.
Forward adjustment range OK.
Movements should be smooth.
2.12 Illumination Tower and Binocular arm can be swung
from side to side OK.
2.13 The replacement Illumination Tower is fitted to the Zeiss SL130
(If applicable).
For the LightLas 532 Green Laser several key safety issues to consider are:
All laser treatment indications and contraindications are well documented and
readily available for the doctors to reference.
There is a long history already for the various treatments when using a Green
Laser, so all possible side effects and adverse reactions are well understood.
The doctor should always ensure they are adequately trained to perform the
Laser procedures in LightLas 532 Green Laser system.
100 mw 80 / 120 mw
1000 mw 800/1200 mw
1500 mw 1200/1800mw
8.1 Symptom
A completed listing of Lightlas system 532 symptoms or warnings and along with the
suggested corrective action is explained as following:
No lamp illumination -No power to the lamp -Check the lamp voltage
(SD only) distribution (+12)
-Lamp is burned out
-Check the lamp integrity
-Check that there is no dust and/or
fingerprint on the bulb
-Call the authorized service or
agent for assistance
Warning Message
Correction Action
Call authorized service if an error message is regularly shown on the system display.
Note: Under no circumstances should unauthorized or untrained personnel attempt
repairs. Refer to the warranty conditions for further details.
The error codes of the Lightlas system are usually shown in 3-digits or 4 digits codes
(refer to fig. 8.1).
Failure Types
Failure Description
Remedy Action(s)
Failure Codes
Err 5042 - Laser -Laser sensor failure -Power off ; call the authorized
Sensor Failure (opened or shorted service or manufacturer for further
sensor) assistance
Err 5043 - Cavity -Cavity Temperature -Reboot the system
Temperature Out of sensor can not reach to -If the problem persisted, call the
Tolerance +/- 2.5C the preset +/- 2.5C authorized service or manufacturer
for further assistance
Err 5044 - Cavity -Cavity temp. sensor -Call the authorized service or
Temperature failure or the sensor is manufacturer for further assistance
Sensor Failure open or shorted circuit
Err 5051 - Remote -No communication -Reboot the system
Communication between remote and the -If the problem persisted, call the
Error laser console within the authorized service or manufacturer
time of frame of 300ms for further assistance
2. Warranty
The LightLas 532 Frequency Doubled YAG Laser conforms to the requirements
of the MDD (93/42/EEC as amended by 2007/47/EC) and thereby in the EU
Countries the Warranty as specified in the applicable regulations will apply.
Your local Distributor will be able to confirm the details.
ber das Ergebnis der Prfung ist eine Bescheinigung auszustellen, die bei den
Gerteunterlagen (Medizinproduktebuch) aufbewahrt werden muss.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which LightLas 532 is used exceeds the applicable RF
compliance level above, LightLas 532 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating LightLas 532.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The following accessories can be purchased from the Distributor to use with the
LightLas 532 Photocoagulator Laser Product. The accessories are only available for
customers outside the EU Countries due to the requirements for CE Marking and the
manufacturer not having control over the use of the attachments that are available
and their indications for use.
Attention: The accessories are passive (do not emit laser radiation) and depend on
the main device (Lightmed LightLas 532).