Multiskan User Manual
Multiskan User Manual
Multiskan User Manual
Multiskan FC
User Manual
Rev. 1.4
Thermo Scientific
Multiskan FC
User Manual
Patents
This product is protected by the following patent: US 6111636, Device for Measuring Optical Density.
Trademarks
Immulon, Microtiter, Multiskan and SkanIt are registered trademarks of Thermo Fisher Scientific.
All other trademarks and registered trademarks are the property of their respective holders.
Disclaimer
Thermo Fisher Scientific reserves the right to change its products and services at any time to incorporate
technological developments. This manual is subject to change without prior notice as part of a continuous
product development. Although this manual has been prepared with every precaution to ensure accuracy,
Thermo Fisher Scientific assumes no liability for any errors or omissions, nor for any damages resulting from
the application or use of this information. This manual supersedes all previous editions.
Power failure
The system requires uninterrupted power supply in order to operate correctly. Thermo Fisher Scientific has no
responsibility whatsoever for system malfunctions arising from power failures.
About This User Manual
Intended users The Multiskan FC, as standalone or with SkanIt Software, can be used
in research and routine-test laboratories by professional personnel.
How to use this This user manual is for the following instruments, Multiskan
FC/Multiskan FC with incubator: Cat. no. 51119000, 51119020,
user manual 51119050, 51119100, 51119120 and 51119150. It has been designed
to give you the information you need to:
Review safety precautions
Install the Multiskan FC
Navigate and edit in the Multiskan FC user interface
Make a protocol and measure
Perform basic cleaning and maintenance procedures
Troubleshoot the instrument performance
This user manual also describes all the features and specifications of the
Multiskan FC instrument as well as ordering information.
Read the manual in its entirety before operating the instrument.
Keep the user manual for future reference. The user manual is an
important part of the instrument and should be readily available during
use of the instrument. Keep the user manual together with the
instrument in case you distribute it onwards.
For more For PC software-related issues, refer to the Thermo Scientific SkanIt
Software for Multiskan FC User Manual (Cat. no. N07713).
information
For the latest information on products and services, visit our websites
at:
http://www.thermo.com
http://www.thermo.com/readingroom
Safety symbols These symbols are intended to draw your attention to particularly
important information and alert you to the presence of hazards as
and markings indicated.
Safety symbols and The following symbols and markings appear on the type label and the
markings used on instrument itself.
the Multiskan FC
Power ON
Power OFF
Serial number
Catalog number
Date of manufacture
A black label with the following text (Figure 310 on page 21):
Warning and other The following symbols and markings appear in this user manual.
markings used in
the documentation
Chapter 10 Shutdown....................................................................................................... 85
Shutting down ................................................................................ 85
Chapter 13 Maintenance................................................................................................. 93
Regular and preventative maintenance ............................................ 93
Maintenance checklist .................................................................. 93
General ........................................................................................ 94
Immediate.................................................................................... 94
Cleaning a filter............................................................................... 95
Cleaning of the instrument ............................................................. 95
Disposal of materials ....................................................................... 96
Decontamination procedure............................................................ 96
Adding individual filters to the filter wheel...................................... 97
Changing the lamp........................................................................ 101
Refitting the transport lock ........................................................... 103
Maintaining a system log............................................................... 105
How to pack for service................................................................. 105
Disposal of the instrument ............................................................ 105
Warning The Multiskan FC weighs 8.5 kg [18.7 lbs.] and care must be
taken when lifting it.
What to do upon This section covers the relevant procedures to be carried out upon
arrival of the instrument.
delivery
How to unpack Move the packed instrument to its site of operation. To prevent
condensation, the instrument should be left in its protective plastic
wrapping until the ambient temperature has been reached. Unpack the
Multiskan FC instrument and accessories carefully with the arrows on
the transport package pointing upwards. Place the instrument onto a
laboratory bench.
Caution Do not touch or loosen any screws or parts other than those
specifically designated in the instructions. Doing so may cause
misalignment and will void the instrument warranty.
Checking delivery Check the enclosed packing list against order. Visually inspect the
for completeness or transport package, the instrument and the accessories for any possible
transport damage. If any parts are missing or damaged, contact your
damage local Thermo Fisher Scientific representative or Thermo Fisher
Scientific Oy.
Environmental When you set up your Multiskan FC, avoid sites of operation with
requirements excess dust, vibrations, strong magnetic fields, direct sunlight, draft,
excessive moisture or large temperature fluctuations. Make sure:
the working area is flat, dry, clean and vibration-proof, and leave
additional room for cables, covers, and so on
the ambient air is clean and free of corrosive vapors, smoke and dust
the ambient temperature range is between +10C (50F) and +40C
(104F)
the humidity is low so that condensing does not occur (relative
humidity is between 10% and 80%, non-condensing).
Installation setups This section describes the installation setups that have to be carried out
before instrument operation.
1. Open the lamp and filter wheel chamber cover (Figure 22).
Figure 22. Transport lock (inside the instrument) and transport lock tag
4. Keep the transport lock bar and tag (Figure 25 on page 16) for
future relocation or transportation of the instrument.
Installing the filter To install the filter wheel, follow these steps:
wheel
1. Unpack the filter wheel (Figure 26). It is delivered in a filter wheel
box. Check that all filters are clean and undamaged.
Position slot
Caution When installing the filter wheel, do not touch any other
mechanical or electronic part.
2. Place the filter wheel into the filter slot so that the filter wheel
numbers (1-8) face outwards (Figure 27). A magnet locking
mechanism will automatically lock the wheel in the correct position
and the optical filter position sensor will make sure the correct filter
is used during measurement.
Figure 27. Placing the Multiskan FC filter wheel into the filter wheel slot
If the filter wheel seems to be "jumping" in the filter wheel slot, you
have inserted the wheel incorrectly and the magnet is rejecting it.
Turn the wheel over.
Connecting the To connect the power supply cable, follow these steps:
power supply cable
Warning Never operate your instrument from a power outlet that has
no ground connection. Never use a power supply cable other than the
Thermo Scientific power supply cable designed for your region.
Connecting to a If you are using a computer with the Multiskan FC, connect the
communication cable to the USB port for the computer (Figure 28).
computer
Connecting to a If you are using a printer, connect it to the USB port for the printer
(Figure 28). The external USB printer protocol that can be used is
printer HP PCL5.
Switching on Check that all the cables are properly fitted according to the installation
instructions. Switch the instrument ON.
Performing the Before you take the instrument into use, make time to perform the
following operational check:
operational
check 1. Switch the instrument ON.
2. Check that the plate carrier comes out from the instrument and the
Self diagnostics passed text is shown for a while on the display.
4. Check that the plate carrier moves in and the measurement chamber
door closes properly.
The instrument starts the measurement. After the measurement the
plate carrier comes out. No error messages should appear on the display.
6. Press the Left arrow key twice to return to the Main menu.
Settings after If you want to change the language or add additional filters to the filter
wheel, modify the required settings in the Settings menu. Refer to the
installation instructions in Settings menu on page 58.
Installation of For installing SkanIt Software for Multiskan FC, refer to the SkanIt
Software for Multiskan FC User Manual (Cat. no. N07713).
SkanIt Software
The instrument has one computer interface: USB.
for Multiskan FC
When the instrument is connected to SkanIt Software, the Remote
mode active text is shown on the display.
Front view The front view of the Multiskan FC instrument is shown in Figure 39.
Display
Lamp and filter wheel Keypad
chamber cover
Power supply
connector
Warning marking
Type label
Side view The side view of the Multiskan FC instrument is shown in Figure 311.
USB memory The instrument is equipped with a USB port for an external memory
stick (Figure 28 on page 18). You can transfer measurement data to the
stick port PC and assay protocols from an instrument to another with the USB
memory stick.
Optical system The optical system of the Multiskan FC instrument comprises the
following components (Figure 312):
Light source
The light source is a quartz-halogen lamp (Osram 64222, 6V/10W).
The lamp is normally off between measurements to prolong the service
life the lamp switches on for the measurement.
Interference filters
The filter is a special interference filter.
The wavelength is selected from one to eight filters held in the filter
wheel.
Fiber
The fiber is flexible, quartz fiber.
Fiber end optics
Fiber end optics produces a highly focused light beam, which passes
through the well to the detector.
Detector housing
The detector housing consists of a silicon photodetector, an amplifier
and lens optics.
Signal processing
The measurement electronics employed has very low noise and a wide
dynamic range.
Autoblanking
The dynamic autoblanking procedure ensures that the lamp can be off
between measurements. This will lengthen the lifetime of the lamp.
Incubator The optional incubator is only available with the Multiskan FC with
incubator model. The incubator chamber consists of thermal elements
on the top and bottom of the incubator chamber. The upper electrical
thermal element prevents condensation on the microplate that is being
incubated.
Background incubation is also possible during kinetic measurements.
The incubation temperature ranges from 25 to 50C (max.).
Shaker The linear shaker operates at three different speeds (Table 31):
Table 31. Shaking speeds
Speed name Speed
Slow 5 Hz, amplitude 15 mm
Medium 11 Hz, amplitude 3 mm
Fast 20 Hz, amplitude 1 mm
Internal memory There is memory for 99 assay protocols and at least assay results for
100 microplates.
Clock Menu
Active row
Display
Function keys
Use the Left, Right, Up and Down arrow keys to navigate. You can
speed up the selection by holding down the arrow key continuously.
Use the OK button to select and edit the highlighted item. It is mostly
used for setting and accepting the active parameter.
Use the F1-F3 keys to select the corresponding action from the info
text bar (Figure 414). The information on the info text bar is updated
according to the active menu.
Function
button
Press the FILE key, for example, for saving the active protocol in the
Main menu. Depending on the level of the internal software, the FILE
key opens a list of actions possible for the current protocol: New, Save,
Save as, Export, Export all protocols, Import, Print and Delete.
Press the PLATE in/out button to move the plate carrier in or out.
Press the START button to start microplate reading using the currently
selected protocol.
Press the STOP button to terminate the reading process. Data from the
plate in progress is lost. It can also be used to terminate the possible
computer remote control.
Main menu You can specify the instrument-related parameters in the Main menu.
The Main menu contains the Protocol, Plate format, Measurement,
Shaking and Incubation parameters.
Note The parameters of the Main menu are locked when the run
(measured data) for the protocol exists. After the measurement, the plate
format, measurement, shaking and incubation parameters cannot be
changed. For more information, refer to Chapter 6: Starting Ready-
made Protocols.
Protocol
Protocol: Created: Modified: Runs:
Plate format
96 Wells
384 Wells
Measurement
Filter 1 (nm)
No filter
405 (#1)
450 (#2)
620 (#3)
Filter 2 (nm)
No filter
405 (#1)
450 (#2)
620 (#3)
Mode
Fast
Normal
No of readings (1-100)
1
Shaking
Mode
No
Before measurement
Background
Incubation
Temperature (C)
No
25-50
Wait for warm up
Yes
No
Time (hh:mm:ss)
00.00.00
Depending on the level of the internal software, the FILE key opens a list of actions possible for the
current protocol.
Protocol The Protocol row in the Main menu shows the name of the active
protocol. You can open another protocol by pressing the OK button on
the Protocol row or by pressing the FILE key. The list of protocols saved
in the internal software will appear.
You can also open a particular run of the protocol in question (protocol
with measured data) with the Menu function. Also the runtime is
shown on the Protocol row. For more details on opening a protocol,
refer to Opening a new protocol on page 65.
Protocol 2
Run 1
Run 2
Run 3
Plate format You can select the plate format 96 or 384 in the Plate format window.
parameters
Note The 384 format is only available if the instrument version
supports 384-well plate measurements.
Measurement You can define the measurement filter(s), number of readings and
parameters measurement mode in the Measurement window.
Filters Filter 1 defines the first (main) filter. Filter 2 defines the second
(reference) filter.
You can select the suitable filter from the list of available filters. You
must physically install the filter into the filter wheel, as well as introduce
the additional filter to the internal software of the instrument in the
Settings window. Refer to Introducing a filter to the internal software
on page 89.
Measurement modes There are two measurement modes: Normal and Fast.
The measurement head stops on each well in the Normal measurement
mode. The Normal mode produces the most accurate results with
measurement times that are fast enough for most purposes.
The head moves continuously in the Fast measurement mode. The Fast
mode produces fast results with a slightly narrower linear absorbance
range than with the Normal mode.
Number of readings sets the number of measurements per well. One
means that each well is measured only once, and two or more means
that the measurement is repeated (kinetic). The number of readings can
be anything between 1 and 100.
If the Number of readings is higher than 1, the measurement is
considered kinetic, and the interval time between the readings must be
set. Kinetic calculations can be added to the protocol.
For instructions on how to program a wavelength, for example, refer to
Programming a wavelength on page 65.
Shaking parameters You can define the shaking parameters in the Shaking window.
Incubation You can define the incubation parameters, incubation time and
parameters temperature in the Incubation window.
Processing You can specify the calculation parameters in the Processing menu.
These parameters supplement the parameters programmed in the Main
menu menu.
The Processing menu contains the following parameters:
Plate layout (with well information)
Preprocessing (for example, kinetic calculations)
Calculation (for example, curve fit)
Interpretation (for example, limit calculation)
Quality control (rules for quality control)
Plate layout You can set the plate map of the protocol in the Plate layout window.
parameters The calculations of the results are made based on the plate layout.
You can start filling the plate layout on any of the wells of the plate. You
start the fill procedure from the well marked with a blue square. You
can start filling a series of samples by pressing the OK button in the
highlighted well. You can edit and delete a single well with the Edit
function.
You can determine the sample type in Well type. The sample types are
Blank, Calibrator, Control, Unknown or Empty.
Note The results will not be calculated for the Empty sample type.
Fill direction can be either to the right or downwards from the active
well in the Plate layout window.
Fill replicate direction can be either the same or different from the fill
samples direction in the Plate layout window.
You can define the number of samples belonging to the group, for
example, the number of different calibrators or the number of different
unknown samples in Number of. You define the number of replicates
for each sample type in Replicate.
Note The Replicate number is the same for samples, for example, for
different controls when using Add series starting from the well
programming. Use the F3 (Edit) key to program each sample
individually, if a different replicate number for the samples are
needed.
When filling the plate with the Add series function, note that adding a
group of samples of the same type to another location will delete the
first group. If you want to add the same well type into several separate
locations on the plate layout, use the Edit function.
For instructions on how to program the plate layout, refer to
Programming a plate layout on page 68.
Preprocessing
Pre-Calculation
No
Mea1 Mea2
Mea1 / Mea2
Mea1 + Mea2
Mea1 * Mea2
Mea2 Mea1
Mea2 / Mea1
Kinetic
Type
No
Average rate
Kinetic rate
/s
/min
Ignore from beginning
0
Ignore from end
0
Maximum rate
Kinetic rate
/s
/min
Ignore from beginning
0
Ignore from end
0
Window
2
Reaction
Undefined
Increasing
Decreasing
Time to maximum rate
Kinetic rate
/s
/min
Ignore from beginning
0
Ignore from end
Continued
Kinetic calculations Kinetic calculations can be used when the number of readings is more
than 1 in the measurement parameters (see Measurement parameters on
page 33). Also the results of a kinetic calculation can be used as source
data for further calculations.
The available kinetic calculation types are:
Average rate
Maximum rate
Time to maximum rate
Time to change
Maximum of well
Time to maximum
Average rate
Average rate is also known as normal rate. The average kinetic rate
(slope of the absorbance (Abs) vs. time curve) is calculated by linear
regression (linear least squares method or LLS) using all the
measurement readings within the selected raw data and time range.
You can define the following settings:
Maximum rate
If the maximum rate is selected, the software searches the data for the
maximum rate found in each well. To obtain the maximum rate, a series
of linear curve fits will be performed for different segments of the
measurement value, absorbance vs. time curve (Figure 515). The first
segment starts at the first data point within the selected time and
measurement range, the second segment starts at the second data point,
and so on, until all the data points have been analyzed. All the rate
calculations are evaluated to determine the maximum rate. In other
words, the LLS fit of m span points are sequentially fitted through each
of the n data points. There will be n - m + 1 curves produced from this.
You can specify the number of data points in a segment with the
Window setting.
Time to change
The time to change is used for calculating the time required to reach a
defined change in the signal (in each well). The time is given in seconds
or minutes from the first reading.
First define the base value by entering the number of readings from the
start in baseline. The base value is the average of results of the given
baseline count from the beginning or end of results. Then select whether
the change from the base value is evaluated as a relative (%) or an
absolute value and define that percentage or value. This change is then
added to the base value to create a required change value. The average of
the sliding window is compared to the limit. The result is the exact
interpolated time where the given change occurs.
Maximum of well
The maximum of well is used to search for the maximum measurement
value in each well.
You can define the following settings:
Time to maximum
The time to maximum is used to calculate the time elapsed before the
maximum measurement signal in each well is reached.
You can define the following settings:
Calculation You can define the calculation parameters in the Calculation window.
parameters The available calculation types are: curve fits and factor.
You can select different calibrator curve fit types for quantitative
calculation or a factor. You can also use a stored curve for quantitative
calculation.
Note The plate layout must include calibrators to enable them for
further calculations, for example, curve fits. Refer to Curve fits on
page 48.
Linear regression
Linear regression is a term usually reserved for the simple linear model
involving a response, y, that is, a continuous variable and a single
explanatory variable, x, related by the equation
y = a + bx
where y denotes the expected value.
The minimum number of calibrators for this fit type is two.
Extrapolation is only available for the linear regression fit type.
ab
y =b+
1 + ( xc) d
In the equation above, y is the signal, x the concentration, a the response
at high asymptote, b the response at low asymptote, c the
1/concentration corresponding to 50% specific binding, and d the slope
factor.
The minimum number of calibrators for this fit type is five.
The Concentration logarithmic transformation is not available for this
fit type.
Extrapolation is not available for the four-parameter logistic fit type.
Cubic spline
This method is a smoothed point-to-point method where the adjacent
calibration points are connected together using cubic polynomials (see
the equation below) and optimizing the connecting points as smoothly
as possible to avoid sharp angles. The results are calculated by first
searching for the correct interval and then using a bisectional method to
find the answer from the corresponding equation. The minimum
number of calibrators for this fit type is five.
yi = ai + bixi + cixi2 + dixi3
Extrapolation is not available for the cubic spline fit type. Only the
means of the measured calibrator replicates can be used, not the
individual replicate values.
Point to point
The point-to-point method connects the adjacent response-
concentrations coordinates, that is, the calibration points, together using
a straight line (see the equation below), which is different for each of the
intervals. The results are calculated by first searching for the correct
interval and then using the corresponding equation. The minimum
number of calibrators for this fit type is two.
yi = ai + bixi
Extrapolation is not available for the point-to-point fit type. Only the
means of the measured calibrator replicates are used in the calculations.
The replicates are visible in the graph but they are not used in
calculations.
Factor Measured absorbances are multiplied with a factor given by the user to
obtain the concentrations. The concentration is calculated according to
the following equation:
Concentration = Factor x absorbance
Stored curve You can use a stored curve for an assay (a run). You can only use a
stored curve with the protocol it was created with. After the curve has
been stored, the plate layout may be modified to contain, for example,
only unknowns. You can save the curve from the curve window using
FILE/Save calibrators and load it using FILE/Load curve.
Note The stored curve must be older than the measured run data. This
means that if a new curve is stored, the earlier saved assay run data
cannot be recalculated against the new curve.
Interpretation
Source data
Interpretation 1
Limit 1
Sample
BLA
Raw Blanked Dual Kinetic Concentration*
CAL1-CAL12
Raw Blanked Dual Kinetic Concentration*
CTRL1-CTRL5
Raw Blanked Dual Kinetic Concentration*
Operator
+ - * / ( )
Constant
1.00
Function
SD CV%
Delete
Interpretation 2
Limit 2
Sample
BLA
Raw Blanked Dual Kinetic Concentration*
CAL1-CAL12
Raw Blanked Dual Kinetic Concentration*
CTRL1-CTRL5
Raw Blanked Dual Kinetic Concentration*
Operator
+ - * / ( )
Constant
1.00
Function
SD CV%
Delete
Interpretation 3
* The pop-up menu appears when you double-click Sample.
Depending on the level of the internal software, the FILE key opens a list of actions possible for the
current protocol.
Quality control Quality control (QC) is used to program rules for quality control and
parameters thereby ensure that the assay works as expected. You can set a maximum
of four QC rules per protocol.
Quality control
Enable quality control
Yes
Rule 1
Rule 2
Rule 3
Rule 4
No
Depending on the level of the internal software, the FILE key opens a list of actions possible for the
current protocol.
The rules (formulas) can be entered with a QC rule editor. The editor
contains the following items:
Sample Select the sample to be checked. It can be any of the
calibrators or controls present in the layout or the blank. If the
sample has replicates, the average value of these is automatically
selected for the QC check. Both raw and calculated data of a sample
can be used as a criteria.
Operator One of the following can be chosen to be used in the
formula: +, -, *, /, (, ), > or <.
Constant A constant can be entered with the number keys on the
keypad.
Function SD or CV% of a sample can be added. SD calculates the
standard deviation of the sample. CV calculates the coefficient of
variation of the sample.
Delete deletes the created formula digit by digit.
For instructions on how to add a QC rule to a protocol, refer to
Adding a QC rule to the protocol on page 75.
Results menu You can view the results of a measured run (assay) in several formats in
the Results menu, depending on the protocol settings in the Processing
menu: raw data in list and table format, calculated results in list and
table format, and graphs (calibration and kinetic curves).
The assay data that is linked to the protocol consists of raw data and all
the information needed for calculations programmed in the Main and
Processing menus.
The result of the previously run assay protocol (measured protocol) is
shown as default (if exists). When another protocol is selected from the
Main menu, the result view is cleared to avoid any conflicts.
Note Results over 4 Abs are calculated but they are struckthrough, and
interpretations for these results are not given.
Raw data You can see the unprocessed data (raw absorbances) of the assay in the
Raw data window.
You can disable/enable samples and sample replicates from the raw data
by pressing the OK button. Select Disable and press the OK button.
You can print or export the selected data. For further information, refer
to Chapter 8: Viewing Results.
Calculated results The calculated results are shown in the Calculated window.
Note The results of precalculations are not available in the data view.
They are only available in exported or printed reports.
The kinetic timing is relative. This means that each time point is
calculated from the first measurement point with the filter.
You can print or export the selected data. For further information, refer
to Chapter 8: Viewing Results.
Calibration curve The calibration curve is shown in the Calibration curve window.
The report also includes the calculated fit parameters of the curve.
You can use both the curve on the plate and a previously created
calibration curve of the same assay protocol. You can store a curve for
future use with the FILE function. You can load a saved curve for the
active protocol with the FILE function.
If a stored curve is in use for calculation, the calibrators measured with
the current plate can be taken into use with the Menu function.
Note The plate layout must include calibrators to enable them for
further calculations, for example, curve fits. Refer to Curve fits on
page 48.
You can disable and enable calibrator replicates from the curve by
pressing the OK button in the Calibration curve window.
You can print or export the selected data. For further information, refer
to Chapter 8: Viewing Results.
Quality control (QC) You can view the quality control data in the Quality control window.
If the QC rules are not fulfilled and the assay is therefore not valid, the
QC view is not opened until after the measurement.
You can print or export the selected data. For further information, refer
to Chapter 8: Viewing Results.
Results in list format The assay results can be viewed in list format in the Results in list
format window.
You can print or export the selected data. For further information, refer
to Chapter 8: Viewing Results.
Settings menu You can set the instrument parameters in the Settings menu.
The values shown in the Settings menu remain in the instrument
memory and are instrument-specific, not protocol- specific.
You can define System, Filters, Printer and Startup settings.
For instructions on how to modify different settings, refer to Chapter
11: Modifying Settings.
System
Date
22.09.2008 14:19:01
Date format
dd-mm-yyyy
dd/mm/yyyy
dd.mm.yyyy
yyyy-mm-dd
mm/dd/yyyy
Time format
12 hour
24 hour
Language
English
Russian
French
Spanish
Chinese
Japanese
Portugese
German
Filters
Wavelength
405 (1)
450 (2)
620 (3)
Comment
Printer
Printer parameters
Enable header
Yes
No
Header text
For example, Laboratory 123
Startup
Plate position
In
Out
Temperature (C)
No
25-50
To start a ready-made protocol shown on the info text bar with the
quick keys (F1-F3), follow these steps:
2. If the plate carrier is inside the instrument, press the PLATE in/out
button. Insert the assay microplate so that the A1 corner is
positioned in the top left corner of the plate carrier (Figure 618).
Note During the run you can press the STOP button to abort the
run.
7. Press the F2 key to close the result table, and then press the Left
arrow key twice to return to the Main menu.
Note The main parameters are locked when a run (measured data) for
this protocol exists.
Starting a ready-made To start a ready-made protocol from the protocol list, follow these steps:
protocol from the list
Caution 1) Do not switch the power off during Performing self
diagnostics. 2) Do not plug/unplug the printer cable during
Importing/Exporting data from/to the USB memory stick. 3) Do not
plug/unplug the USB memory stick during Printing or Performing
self diagnostics.
2. Select the ready-made assay protocol you want to run from the
protocol list using the Down arrow key, and then press the OK
button.
Note The protocol name selected is shown on the Protocol row in the
Main menu.
3. If the plate carrier is inside the instrument, press the PLATE in/out
button. Insert the assay microplate so that the A1 corner is
positioned in the top left corner of the plate carrier.
Note During the run, you can press the STOP button to abort the
run.
8. Press the F2 key to close the result table, and then press the Left
arrow key twice to return to the Main menu.
Note The main parameters are locked when a run (measured data) for
this protocol exists.
Note Ensure that you save the changes you make in each step. Refer to
Saving a new (active) protocol on page 76.
2. Press the OK button at the Filter 1 (nm) item and select, for
example, 450 as the filter 1 value using the Down arrow key, and
then press the OK button.
3. Press the F1 key to accept the selection and return to the Main
menu.
2. Press the OK button at the Mode item and select, for example,
Before measurement as the shaking mode using the Down arrow
key, and then press the OK button.
3. Select the Speed item using the Down arrow key and press the OK
button.
4. Select, for example, Fast as the shaking speed using the Down
arrow key, and then press the OK button.
5. Select the Time (hh:mm:ss) item using the Down arrow key and
press the OK button.
7. Press the F1 key to accept the selection and return to the Main
menu.
Programming You can set the incubation time and temperature in the Incubation
window.
incubation
Note The Incubation row is only active with the Multiskan FC with
incubator model.
1. Select the Incubation row in the Main menu using the Down
arrow key and press the OK button.
2. Press the OK button on the Temperature row and use the Down
arrow key to select the preferred temperature from the list. The
temperature range is from 25 to 50C. Then press the OK button.
3. If Wait for warm up is set to Yes, the assay will not start until the
temperature is reached. If Wait for warm up is set to No, the assay
will start.
4. Select the Time (hh:mm:ss) item using the Down arrow key and
press the OK button.
Programming a You can start filling the plate layout on any of the wells of the plate.
You start the fill procedure from the well marked with a blue square.
plate layout You can start filling a series of samples by pressing the OK button in the
highlighted well. You can edit and delete a single well with the Edit
function.
In this example
a blank sample
Calibrator 1 (Conc. 1)
Calibrator 2 (Conc. 2)
Calibrator 3 (Conc. 3)
a control sample, and
91 unknown samples
are programmed.
To create the plate layout of this example protocol, follow these steps:
1. Select the Plate layout row in the Processing menu using the
Right arrow key. Then press the OK button to open the Layout
window.
2. Press the OK button to start filling the plate from well A1.
3. Select the Well type item in the Add series starting from well: A1
window and press the OK button.
4. Select, for example, a Blank sample using the Up arrow key, and
then press the OK button.
5. Press the F1 key to accept the selection and return to the Layout
window (well A2).
6. Select well B1 using the Down and Left arrow keys and press the
OK button to start filling the plate from well B1.
7. Select the Well type item in the Add series starting from well: B1
window and press the OK button.
9. Select the Number of item using the Down arrow key and press
the OK button.
10. Select, for example, 3 by using the number keys, and then press the
to , OK button.
11. Press the F3 (Conc.) key to program the concentrations for the
calibrators.
12. Select the Cal 1 item using the Down arrow key and press the OK
button.
13. Enter, for example, 1 (1.00) as the Cal 1 concentration using the
to ,
number keys, and then press the OK button.
14. Select the Cal 2 item using the Down arrow key and press the OK
button.
15. Enter, for example, 2 (2.00) as the Cal 2 concentration using the
to , number keys, and then press the OK button.
16. Select the Cal 3 item using the Down arrow key and press the OK
button.
17. Enter, for example, 3 (3.00) as the Cal 3 concentration using the
to , number keys, and then press the OK button.
18. Press the F1 key to accept the concentrations and return to the
previous view, i.e., the Add series starting from well: B1 window.
19. Press the F1 key to accept the concentrations and to return to the
Layout window.
20. Press the OK button to start filling the plate from well E1.
21. Select the Well type item in the Add series starting from well: E1
window and press the OK button.
22. Select, for example, a Control sample using the Up arrow key, and
then press the OK button.
23. Press the F1 key to accept the selection and return to the Layout
window.
24. Press the OK button to start filling the plate from well F1.
25. Select the Well type item in the Add series starting from well: F1
window and press the OK button.
26. Select, for example, an Unknown sample using the Up arrow key,
and then press the OK button.
27. Select the Number of item using the Down arrow key and press
the OK button.
28. Select, for example, 91 using the number keys, and then press the
to , OK button.
29. Press the F1 key to accept the selection and return to the Layout
window. The filled wells will show the following colors: white for
blank wells, green for calibrators, bright green for controls, and
blue for unknowns.
30. Press the F1 key again to accept the plate layout and to return to
the Processing menu.
2. Select the Precalculation type and press the OK button. Select, for
example, Mea1-Mea2 using the Down arrow key and press the OK
button.
3. Select, for example, Blanked using the Down arrow key, and then
press the OK button.
5. Enter the category name, for example, Neg, by using the letter keys
and press the OK button.
to ,
6. Select Limit (negative) using the Up arrow key and press the OK
button.
The limit editor is opened.
8. Select CTRL 1 by using the Down arrow key and press the OK
button.
10. Select Operator using the Right arrow key and press the OK
button.
12. Select Constant using the Right arrow key and press the OK
button.
13. Enter the constant, for example, 0.050, using the number keys and
to , press the OK button.
15. Select Interpretation 2 using the Up arrow key and press the OK
button.
16. Enter the category name, for example, Pos (positive), using the
to , letter keys, and then press the OK button.
Adding a QC rule The rules are normally given in the kit insert and can read, for example,
the absorbance of Blank has to be smaller than 0.100 Abs. In this case the
to the protocol rules to be entered would be: Sample: Raw(BLA), Operator: <, Constant:
0.100.
The following example shows how to add a Quality Control (QC) rule
to a protocol.
1. Select the Quality Control row in the Processing menu using the
Down arrow key and press the OK button.
2. Press the OK button on the Enable quality control and select Yes
using the Up arrow key, and then press the OK button.
3. Select the Rule 1 row using the Down arrow key and press the OK
button.
7. Select Operator using the Right arrow key and press the OK
button.
8. Select < using the Down arrow key and press the OK button.
9. Select Constant using the Right arrow key and press the OK
button.
10. Enter the constant, for example, 0.100 using the number keys and
to , press the OK button.
12. Press the F1 key to accept the Quality control parameters and to
return to the Processing menu.
Saving a new To save a new (active) protocol that is open in the menu, follow these
steps:
(active) protocol
1. Press the FILE key in the Main menu.
2. Select Save As using the Down arrow key and press the OK button.
The Save Protocol As dialog opens.
3. Enter the protocol name, for example, Test1, by using the number
to , and letter keys, and then press the OK button.
Note The results of an Undefined protocol are not saved unless you
save the active protocol, including the measured run data with a new
name.
After the run has finished, the data view of a run is automatically
shown.
2. Select another data view using the Up and Down arrow keys and
press the OK button.
3. Press the FILE key in the data view to print or export a data view.
Select Export As Text or Print using the Down arrow key if
necessary, and then press the OK button.
5. Select another row from the Results menu using the Up and
Down arrow keys and press the OK button. To view different data
views or to print, refer to Steps 2-4.
Printing or This section provides information on how the active run data (measured
results) can be printed or exported to the USB memory stick. Follow
exporting data these steps:
1. If you are exporting data, insert the USB memory stick into the
USB memory stick port of the instrument
OR
If you are printing data, ensure that the printer is connected and
switched on.
2. In the Main menu, press the Right arrow key continuously till you
get to the Results menu.
Note The data format should be UTF 8 when the *.txt file is opened
in, for example, Microsoft Excel.
Exporting a protocol The calculated results or stored protocols can be exported to the USB
memory stick via the USB port (Figure 28 on page 18) by using the
FILE key. Runs, including protocol information and data, can be
exported. Protocols can be exported and imported to another Multiskan
FC instrument with the same configuration. Import of runs is only
possible to the same instrument as backup.
In the main level of Results, the results or protocols to be exported can
be defined by FILE/Export.
A protocol can be transferred from an instrument and imported to
another instrument. To export a protocol, follow these steps:
1. Insert the USB memory stick into the USB memory stick port of
the instrument.
3. Select the protocol you want to export from the protocol list using
the Up or Down arrow key.
4. Press the FILE key and select Export using the Down arrow key.
Then press the OK button.
5. If you also want to export the runs that are created (measured) with
the protocol, press the OK button. If not, press the Right arrow
key, and then the OK button.
6. Press the F2 key to close the protocol list and return to the Main
menu.
1. Insert the USB memory stick into the USB memory stick port of
the instrument.
3. Press the FILE key and select Import using the Down arrow key,
and then press the OK button.
4. Select the protocol you want to import from the USB protocol list
using the Up or Down arrow key and press the OK button.
Note You can only import the runs to the same instrument they were
created with. You can also import the protocol to an instrument with
the same configuration.
5. Press the F2 key to close the protocol list and return to the Main
menu.
2. Press the PLATE in/out button. Make sure the plate carriage goes
inside the instrument and check that the measurement chamber
door closes properly.
System settings The Multiskan FC instruments internal software version, serial number
and model are shown in the System settings window. Model 1 refers to
Multiskan FC and Model 2 to Multiskan FC with incubator.
You can set the date, date and time format, and language in the
System settings parameter window.
Select the item to be changed with the Up and Down arrow keys and
press the OK button.
Setting the date You can change the date in the System settings.
The Left and Right arrow keys are used to move from one position to
another in the Date setting. Enter the values with the number keys.
, to
Changing the You can set the internal software language by moving to Language and
language selecting the preferred language from the drop-down list.
To change the internal software language, follow these steps:
1. In the Main menu, select the Settings menu using the Left arrow
key.
3. Select the Language item using the Down arrow key and press the
OK button.
5. Press the F2 key to confirm the selection and to close the system
parameters.
Filters Multiskan FC comes with installed filters of 405 mm, 450 mm and
620 mm.
You can introduce a filter to the internal software by moving to the
Filters settings. The filter wavelength must be between 340 and
850 nm.
The Filters settings window shows the current filters of the filter wheel.
Introducing a filter This section provides an example on how to enter the information of an
to the internal additional filter, which has been added to the filter wheel. To introduce
a filter to the internal software, follow the given steps.
software
1. In the Main menu, select the Settings menu using the Left arrow
key.
2. Select the Filters row using the Down arrow key and press the OK
button.
3. Select the empty filter position by pressing the Right arrow key
until the filter wheel on the display rotates to the empty position
and the text Empty is shown.
Removing a filter You can remove a filter from the instrument settings by selecting the
position and pressing the F3 (Remove) key.
Note The filter must physically be removed from the filter wheel.
Printer You can add a header in the Printer parameter window, for example,
the laboratory information, to all of the reports to be printed. You can
do this by selecting Yes for the Enable header.
The maximum length of the header is about 50 characters.
Startup In the Startup parameters, the Plate position setting defines the plate
carrier position after the instrument is switched on.
You can set the startup temperature. The instrument starts to warm up
to the set temperature immediately after it has been switched on.
Note The set startup temperature is only activated when the instrument
is powered (switched on). The temperature set in the protocol overrules
the startup temperature. If the protocol does not include incubation, the
startup temperature remains.
Monthly
Weekly
Yearly
Daily
Keep the instrument free of dust. See General on page 94. 3
Avoid disturbing any of the optical system components, including 3
optical covers. See General on page 94.
Clean and check the condition of the filters and the filter wheel position 2x
slot.
Wipe away spilled saline solutions, solvents, acids or alkaline solutions 3
from outer surfaces immediately to prevent damage, and wipe with
deionized distilled water. See Immediate on page 94.
If any surfaces have been contaminated with biohazardous material, 3
disinfect with a mild sterilizing solution. See General on page 94.
Clean the case of the instrument periodically. See Cleaning of the 3
instrument on page 95.
Clean the plate carrier when necessary. See Cleaning of the 3
instrument" on page 95.
Change the lamp when blown. See Changing the lamp on page 101. 3
Ensure proper shutdown. See Chapter 10: Shutdown. 3 3
Decontaminate the instrument when relocating the instrument or 3
sending it for service. See Decontamination procedure on page 96.
Perform verification with the Multiskan Verification Plate, Cat. no. 2x
24072800.
Service the instrument regularly. See Cleaning of the instrument on 3
page 95 and Maintaining a system log on page 105.
3= depending on the laboratory conditions and the use and configuration of the instrument
Caution Do not use any other liquids to clean the optics. Avoid any
harsh treatment.
3. Clean the instrument outside and the plate carrier with a soft cloth
dampened with water or mild detergent.
Decontamination If you have spilt infectious agents, carry out the decontamination
procedure.
procedure
Warning The decontamination procedure should be performed by
authorized trained personnel in a well-ventilated room wearing
disposable gloves, protective glasses and clothing.
2. Empty the plate carrier. Ensure that you are wearing disposable
gloves.
3. Switch OFF the power and disconnect the power supply cable.
5. Place the instrument in a large plastic bag. Ensure that the lamp and
filter wheel chamber cover and measurement chamber door are
open and the plate carrier is out.
7. Close the bag firmly and leave the instrument in the bag for at least
24 hours.
Adding individual filters This section explains how to replace a filter in the filter wheel or add an
additional filter to the filter wheel.
to the filter wheel
Caution Only use filters approved by the instrument manufacturer.
Caution When handling the filter wheel, do not touch any other
mechanical or electronic part.
3. Lift the filter wheel from the filter wheel slot (Figure 27 on page
17 and Figure 1319 on page 98). Do not touch the filter surfaces.
4. Remove the filter spring by unscrewing the four (4) spring position
holding screws (Figure 1320 and Figure 1321 on page 99).
Caution Do not touch the filter glass surfaces with your bare hands.
Caution The magnet in the middle of the filter wheel attracts both the
screwdriver and the screws. Make sure the screwdriver or the screws do
not scratch the filters.
Filter spring
Filter
Filter wheel
Figure 1323. Inserting a filter into the next free position in the filter wheel
5. Insert a new filter or an additional optional filter into the next free
position in the filter wheel (Figure 1323 on page 99). When
inserting the filter, the arrow must be pointing upwards as in Figure
1323 on page 99. The wavelength of the filter is marked on the
filter package as well as on the filter side, that is, the three digits
before the last digit of the number sequence: xxxxx NNNx, where
NNN is the wavelength. The filter positions are marked on the
other side of the filter wheel. The filter wheel has at present three
filters factory installed (Table 1314).
Table 1314. Example of the filters in the filter wheel positions 1-8
Filter wheel position Wavelength (nm)
1 405 (example of factory installed filter)
2 450 (example of factory installed filter)
3 620 (example of factory installed filter)
4
5
6
7
8
6. Write down in Table 1314 the additional filters you have installed
into the filter wheel. It is recommended to add the filters in
ascending wavelength order.
7. Place the filter spring into its original position and fasten the four
screws symmetrically.
8. Slide the filter wheel back into the filter wheel slot with the filter
numbers (1-8) facing outwards (Figure 27 on page 17). The
magnet locking mechanism will automatically lock the wheel in the
correct position. Refer to Installing the filter wheel on page 16.
Changing the lamp To replace the lamp that has burnt out, follow these steps:
Caution Only use the lamp approved by the supplier: Cat. no.
1410101, Lamp, quartz-halogen lamp (Osram 64222, 6V/10W).
Warning If the instrument has been in use and you need to replace a
burned lamp, the lamp and its surroundings may be very hot. Wait for
the lamp to cool down before replacing it.
Caution When handling the lamp, do not touch any other mechanical
or electronic part.
3. Turn aside the two clamps, which keep the lamp in place
(Figure 1324 on page 102 and Figure 1325 on page 102).
5. Loosen the two screws of the terminal socket and pull the lamp
from the terminal socket.
Caution Do not touch the reflective surface of the new lamp or the
bulb itself.
6. Refit the terminal socket to the contacts of the new lamp approved
by the manufacturer (OSRAM 64222, 6V/10W).
7. Tighten the screws of the terminal socket and put the new lamp in
place.
Note Place the lamp and terminal socket into the correct position.
Refer to Figure 1325 on page 102.
8. Turn the two clamps in place to lock the lamp into position.
Refitting the When you relocate the instrument or ship it for service, make sure you
refit the transport lock.
transport lock
To refit the transport lock, follow these steps:
2. Press the PLATE in/out button so that the plate carrier goes in.
5. Push the transport lock bar through the hole in Figure 1326 into
the hole in the measurement head.
Figure 1326. Refitting the transport lock bar into the measurement head
7. Fasten the transport lock bar and tag clockwise using the Allen key
supplied (Figure 1327). Ensure that the transport lock bar is
completely tightened.
Maintaining a A system log, which includes a short summary of the use, maintenance
procedures, error messages and other information about the use of the
system log system can be very useful in properly maintaining the system. Refer to
Appendix A: System Log. Copy the table as many times as necessary,
but leave the blank original inside the user manual.
How to pack for To pack the Multiskan FC for service, follow the instructions presented
below:
service
Inform about the use of hazardous materials.
Remove any microplate before decontamination. Also remove the
filter wheel and put it into its transportation box. Decontaminate
the instrument.
Refit the transportation lock. Refer to Refitting the transport lock
on page 103.
Close the lamp and filter wheel chamber cover and the
measurement chamber door.
Pack the instrument according to the packing instructions.
Use the original packaging for shipping.
Enclose a dated and signed Certificate of Decontamination (see
Appendix B) both inside and attached to the outside of the package,
in which you return your instrument (or other items).
Enclose the return authorization number (RGA) given by your local
Thermo Fisher Scientific representative.
Indicate the fault after you have been in touch with your local
Thermo Fisher Scientific representative or the Thermo Fisher
Scientific technical service department.
Warning The used lithium (Li) battery is regulated waste and must be
disposed of according to strict EPA (Environmental Protection Agency)
requirements. The Li battery has to be changed by an authorized service
technician only. Instructions for changing the Li battery are described
in the service manual.
User interface The instrument can be under PC control and run on SkanIt Software for
Multiskan FC, which controls all the instrument functions and provides data
reduction as well as reporting functions.
Computer interface USB 1.1 (2.0 compatible)
Plate types 96 and 384-well plates
Shaker Linear shaking, 3 speeds: Slow (5 Hz, 15 mm), Medium (11 Hz, 3 mm) and
Fast (20 Hz, 1 mm)
Performance This section provides the performance specifications for the relevant
measurement technique and other instrument capabilities.
specifications
Table 1416. Photometry
Performance specifications / Photometry
Optical system Quartz-halogen lamp (Osram 64222, 6V/10W), interference filter (in the
filter wheel), fiber, fiber end optics, photodetector, signal processing
Wavelength range 340 850 nm
Filters 8-position filter wheel
The filters are optional. The standard filters are: 405 nm; 450 nm, and
620 nm. Additional filters can be ordered separately.
Half-bandwidth of filters 3 9 nm
Wavelength accuracy 2 nm
Detector One silicon photodetector
Linearity 0 4 Abs, 2%, normal mode
(96-well plate) 0 3 Abs, 2%, fast mode
(405 nm)
Linearity 0 3 Abs, 2% with normal mode
(384-well plate) 0 2.5 Abs, 2% with fast mode
(405 nm)
Measurement range 0 6 Abs
Absorbance resolution 0.001 Abs
Accuracy 1% (0.3 3 Abs)
(405 nm, normal mode) 2% (3 4 Abs)
Precision CV 0.2% (0.3 3 Abs)
(405 nm, normal mode) CV 1.0% (3 4 Abs)
Startup time Approx. 30 s
Measurement time 6 s (96-well plate), fast mode
12 s (96-well plate), normal mode
15 s (96-well plate), fast mode, dual wavelength
11 s (384-well plate), fast mode
33 s (384-well plate), normal mode
25 s (384-well plate), fast mode, dual wavelength
Long-term stability Instrument is stable due to autoblanking procedure.
Safety This section describes the safety specifications for the Multiskan FC
instrument.
specifications
In conformity with Multiskan FC bears the following markings:
the requirements Type 357
100 240 Vac, 50/60 Hz, 100 VA max.
CE mark
CSA monogram
Safety performance:
EN 61010-1:2001 (Ed. 2),
including US and CA National differences
EN 61010-2-010:2003 (Ed. 2)
EN 61010-2-101:2002 Particular Requirements for In Vitro Diagnostic (IVD) Medical
Equipment
The safety specifications are also met under the following environmental
conditions in addition to or in excess of those stated in the operating conditions:
Altitude Up to 2000 m
Temperature +5C to +40C
Humidity Maximum relative humidity 80% for temperatures up to
31C decreasing linearly to 50% relative humidity at 40C
Mains supply fluctuations 10% from nominal
Installation category II according to IEC 60664-1 (see Note 1)
(overvoltage category)
Pollution degree 2 according to IEC 60664-1 (see Note 2)
EMC performance:
EN 61000-6-3:2001 Generic standards
Emission standard for residential, commercial and light-
industrial environments
EN 61000-6-1:2007 Generic standards
Immunity standard for residential, commercial and light-
industrial environments
EN 61326-1:2006 Product family standard
Electrical equipment for measurement, control and
laboratory use
EN 61326-2-6:2006 Product family standard
Electrical equipment for measurement, control and
laboratory use; In vitro diagnostic (IVD) medical equipment
Note The instrument does not verify the logic flow of the received
commands.
Error and When an error is detected, the current operation is terminated. After an
error, it is best to abort the current run and restart from the beginning
warning codes after the problem is fixed. The error (Table 1518) and warning codes
(Table 1519) that may appear in SkanIt Software for Multiskan FC
are presented below.
Table 1518. Error codes reported
Code Explanation Suggested action
0 The command was executed successfully.
2 The instrument did not recognize the command it received. Contact the PC software vendor.
3 The arguments of the received command are not valid. Contact the PC software vendor.
4 Plate positioning error. Check that there is nothing preventing free plate
movement.
Contact service.
5 Measure head positioning error. Check that there is nothing preventing free measure
head movement.
Contact service.
6 Filter wheel positioning error. Check that the filter wheel is in place.
Contact service.
7 AD converter offset voltage is too high. Contact service.
8 AD converter noise level is too high. Contact service.
9 Lamp failure. Replace the lamp.
It is also possible to get this error if either one of the
measure head or plate motors has lost steps so that
the position is wrong.
10 Non-volatile parameters lost. Contact service.
11 Attempt to set the instrument serial number when it already Do not try to set the instrument serial number.
has been set.
12 Not enough memory for a new user-defined parameter. Using the PUT command, delete some unused user
Continued
Cont.
Code Explanation Suggested action
This error may be reported in response to a PUT command. parameters to make room for new parameters. The
internal memory is full. Take a backup by exporting
the protocols/runs. Refer to Chapter 9: Printing,
Exporting and Importing. Release space from the
internal memory by deleting protocols/runs from the
internal memory.
13 Such an error during startup, that the execution of the MEA When you connect to the instrument with the VER
and SCA commands is precluded. command, the instrument should not report any error
in the response. If it does, try to eliminate the error.
If you are unable to eliminate the error, contact
service.
14 The distance between plate wells is too short for scan Check the parameters of the PLA command used.
measurement.
15 The sampling time for a single result is too long for scan Use the MEA command instead of the SCA
measurement. The plate cannot be moved as slowly as the command. Use a shorter sampling time in the MPA
sampling time requires. command.
16 Failed to adjust a suitable lamp intensity for a filter. Check that there actually is a filter in all used filter
positions. If the error persists, contact service.
17 Too high dark signal level. The dark level is checked once This error may also be caused by a measure head or
for each measured plate row/column. plate position error see errors 4 and 5. Contact
service.
18 Measure signal saturation. Contact service.
19 No filters defined. Define at least one filter you are using.
20 The USB memory stick/printer interface is not working. Contact service.
21 The USB memory stick/printer interface firmware is Contact service.
corrupted.
22 XY table position calibration failed. This may be reported Contact service.
after startup.
23 Command parser optimization failure. You should never see this error. This is not fatal
the instrument still works but command parsing is a
bit slower.
Contact service.
24 Temperature reference is out of limits. Incubation is not Contact service.
available if this error is reported.
25 Parameters memory not found. Contact service.
26 Parameters memory erase failure. Contact service.
27 Parameters memory write failure. Contact service.
28 Temperature is not changing. Use a at least a 4C higher incubation temperature
than the environmental temperature. This applies to
both the protocol and startup temperature.
Contact service.
29 One or more heaters not ok. Contact service.
50 File open error. File not found. Contact service.
51 Filter 1 not selected in measure parameters. Select filter 1.
Continued
Cont.
Code Explanation Suggested action
52 Filter information in the protocol does not match the filter Edit the protocol or add the correct filter to the filter
wheel. wheel.
53 The file already exists. Save with another file name.
54 The file does not exist. Contact service.
55 The media is full. You have to delete some protocols or runs. If
Defragmenting file system messages start to
appear on the display, you should already at this
stage make more space by deleting some protocols
or runs.
56 End of file. Contact service.
57 Other file error (none of the previous). Contact service.
58 Firmware update aborted by user.
59 Printer not attached. Connect the external printer.
60 The printer is out of paper. Add paper to the printer.
61 Printing error. Retry and if the error persists, contact service.
Warnings and This instrument is designed to provide full user protection. When
correctly installed, operated and maintained, it will present no hazard to
cautions the user.
The following recommendations are given for added user safety.
Electrical Ensure that the power supply cable supplied with the unit is always
used. If a correct type of mains cable is not provided, use only cables
certified by the local authorities.
The power plug should only be inserted into a socket outlet provided
with a protective ground contact. Never use an extension cable without
a protective ground wire.
Defects and This section describes about defects and abnormal stresses.
abnormal stresses
Warning If the instrument is not functioning properly, it may create
electromagnetic perturbation, which could impair the operation of
other devices or equipment in the usual environment.
List of spare Table 1621. Codes for spare parts and accessories
parts and Code Item Quantity
PHOTOCOPIABLE
Name:
Address:
Tel./Fax:
Name: Serial no.:
A) I confirm that the returned items have not been contaminated by body fluids, toxic,
carcinogenic or radioactive materials or any other hazardous materials.
B) I confirm that the returned items have been decontaminated and can be handled without
exposing the personnel to health hazards.
Materials used in the unit: Chemicals + Biological Radioactive *)
Decontamination procedure1:
PHOTOCOPIABLE
1
Please include decontaminating solution used.
A E
absorbance, 11, 12, 33, 40, 41, 50, 55, 73, 75, 108, 123 enabling QC, 54, 75
accessories, 13, 116 endpoint measurement, 26, 123
accuracy, 2, 94, 108 environmental requirements, 14
adding a QC rule, 54, 75 error, 19, 105, 111, 112, 113, 123
adding individual filters to the filter wheel, 97 codes, 111
application, 2, 5, 12 messages, 19, 105, 123
autoblanking, 23, 108 exporting a protocol, 82
autocalibration, 23, 123 exporting results or protocols, 55, 56, 57, 80, 81, 82, 83
average rate, 33, 38, 39, 40
F
B factor, 47, 48, 49, 50
Baseline readings, 38, 45 filter wheel, 14, 16, 17, 19, 23, 33, 88, 89, 90, 93, 97, 98,
99, 100, 101, 103, 104, 105, 108, 111, 112
C filters, 16, 19, 23, 29, 33, 58, 59, 87, 88, 89, 93, 94, 95, 97,
98, 100, 101, 108, 112, 115
calculated results, 29, 54, 56, 79, 81, 82
four parameter logistic, 48, 49
calculation parameters, 29, 31, 35, 36, 38, 47, 48, 51, 54, 55,
68, 71, 72, 73, 75
calibration, 29, 49, 50, 54, 56, 57, 79, 112 G
curve, 29, 56, 57 glossary, 123
Change, 38, 45, 93
Change type, 38, 45 H
changing the lamp, 93, 101, 102 half-bandwidth, 108
changing the language, 87 header, 59, 90
cleaning the filters, 95 how to decontaminate the instrument, 85, 93, 95, 96, 97,
connecting to a computer, 18 105, 121
connecting to a printer, 18 how to install the Multiskan FC, 13, 14, 18, 19, 94, 109
cubic spline, 48, 49 how to maintain the instrument, 62, 64, 93
curve fit, 35, 39, 41, 47, 48, 57, 72 how to pack for service, 13, 105
cutoff, 51, 52 how to shut down, 85, 93
how to switch on, 18
D how to unpack, 13
decontamination, 85, 93, 95, 96, 97, 105, 121, 123
Certificate of, 97, 105, 121 I
procedure, 85, 93, 95, 96, 97, 105, 121 Ignore from beginning, 38, 41, 43, 44, 45, 46
detector, 23, 108, 123 Ignore from end, 38, 41, 43, 44, 45, 46
disposal of instrument, 105 importing a protocol, 83
disposal of materials, 96 importing results or protocols, 82, 83
www.thermo.com
N07710