ASTM D4739 - 2011 - Base Number Determination by Potentiometric Titration
ASTM D4739 - 2011 - Base Number Determination by Potentiometric Titration
ASTM D4739 - 2011 - Base Number Determination by Potentiometric Titration
INTRODUCTION
Currently, there are two ASTM test methods under the jurisdiction of Committee D02 used for
determining base number by potentiometric titration: Test Method D2896 and Test Method D4739.
They are both used throughout the petroleum industry. Test Method D2896 is for new oils and is used
in setting specifications since it is more accurate than Test Method D4739. Test Method D4739 shall
be used exclusively for the purpose of tracking base number loss as an oil proceeds in service. In many
cases, the test methods will provide different results. Some comparisons are given in the Scope and
the Significance and Use sections of this test method to distinguish between the two standards and to
provide guidance to users.
1. Scope* be used to indicate relative changes that occur in oil during use
1.1 This test method covers a procedure for the determina- under oxidizing or other service conditions regardless of the
tion of basic constituents in petroleum products and new and color or other properties of the resulting oil. The values
used lubricants. This test method resolves these constituents obtained, however, are intended to be compared with the other
into groups having weak-base and strong-base ionization values obtained by this test method only; base numbers
properties, provided the dissociation constants of the more obtained by this test method are not intended to be equal to
strongly basic compounds are at least 1000 times than that of values by other test methods. Although the analysis is made
the next weaker groups. This test method covers base numbers under closely specified conditions, this test method is not
up to 250. intended to, and does not, result in reported basic properties
that can be used under all service conditions to predict
1.2 In new and used lubricants, the constituents that can be performance of an oil; for example, no overall relationship is
considered to have basic properties are primarily organic and known between bearing corrosion or the control of corrosive
inorganic bases, including amino compounds. This test method wear in the engine and base number.
uses hydrochloric acid as the titrant, whereas Test Method
D2896 uses perchloric acid as the titrant. This test method may 1.4 This test method was developed as an alternative for the
or may not titrate these weak bases and, if so, it will titrate former base number portion of Test Method D664 (last
them to a lesser degree of completion; some additives such as published in Test Method D66481).
inhibitors or detergents may show basic characteristics. 1.4.1 Colorimetric test methods for base number are Test
Method D974, IP 139, and 5102.1 on acid and base number by
1.3 When testing used engine lubricants, it should be
extraction (color-indicator titration) of Federal Test Method
recognized that certain weak bases are the result of the service
Standard No. 791b. Test results by these methods may or may
rather than having been built into the oil. This test method can
not be numerically equivalent to this test method.
1
1.5 The values stated in SI units are to be regarded as
This test method is under the jurisdiction of ASTM Committee D02 on
Petroleum Products and Lubricants and is the direct responsibility of Subcommittee standard. No other units of measurement are included in this
D02.06 on Analysis of Lubricants. standard.
Current edition approved May 15, 2011. Published July 2011. Originally
approved in 1987. Last previous edition approved in 2008 as D473908 1. DOI: 1.6 This standard does not purport to address all of the
10.1520/D4739-11. safety concerns, if any, associated with its use. It is the
Copyright by ASTM Int'l (all rights reserved); Sun Sep 23 18:56:57 EDT 2012
2
Downloaded/printed by
D4739 11
period of 90 s is allowed to pass before another 0.100-mL 7.4 Buffer, Aqueous BasicCommercial pH 10 buffer solu-
increment of titrant is delivered. This procedure is repeated tion with a tolerance of 60.02 pH units at 25C. This solution
until the titration is completed. shall be replaced at regular intervals consistent with its stability
6.1.2 The precision of addition of the 0.100-mL increments or when contamination is suspected. Information related to the
of titrant must be 60.001 mL for automatic titrators. For stability should be obtained from the manufacturer.
manual buret, it should be 60.005 mL. A higher incremental 7.5 ChloroformReagent grade. (WarningToxic and
precision is required for an automatic buret, because the total suspected carcinogen.)
volume to the end point is summed from the individual
increments; it is read from a scale with a manual buret. 7.6 Hydrochloric Acid Solution, Standard Alcoholic
(0.1 M)Mix 9 mL of reagent grade hydrochloric acid (HCl,
6.2 Sensing Electrode, standard pH with glass membrane,
sp gr 1.19) (WarningToxic and corrosive), with 1 L of
suitable for non-aqueous titrations.
anhydrous isopropyl alcohol. Standardize frequently enough to
6.3 Reference Electrode, Silver/Silver Chloride (Ag/AgCl) detect normality changes of 0.0005 by potentiometric titration
reference electrode with sleeve junction, filled with 1 to 3 M of approximately 8 mL (accurately measured) of the 0.1 M
LiCl in ethanol. alcoholic KOH solution diluted with 125 mL CO2-free water.
6.3.1 Combination ElectrodeSensing electrodes may have
7.7 Ethanolreagent grade. (WarningFlammable and
the Ag/AgCl reference electrode built into the same electrode
toxic, especially when denatured.)
body, which offers the convenience of working with and
maintaining only one electrode. The combination electrode 7.8 Lithium Chloride Electrolyte Prepare a solution of 1 to
shall have a sleeve junction type of reference and shall use an 3 M LiCl in ethanol.
inert ethanol electrolyte, for example, 1 to 3 M LiCl in ethanol.
7.9 Potassium Hydroxide(WarningCauses severe
In the reference compartment, the sensing electrode part shall
burns.)
use a glass membrane designed for non-aqueous titrations.
These combination electrodes shall have the same response or 7.10 Potassium Hydroxide Solution, Standard Alcoholic
better response than a dual electrode system. They shall have (0.1 M)Add 6 g of reagent grade potassium hydroxide
removable sleeves for easy rinsing and addition of electrolyte. (KOH) (WarningToxic and corrosive), to approximately 1 L
(WarningWhen a movable sleeve is part of the electrode of anhydrous isopropyl alcohol. Boil gently for 10 min to effect
system, ensure that the sleeve is unimpaired before every solution. Allow the solution to stand for 2 days, and then filter
titration.) the supernatant liquid through a fine sintered-glass funnel.
6.4 Stirrer, Buret, Stand, Titration Vessel, as specified in Store the solution in a chemically resistant bottle. Dispense in
Annex A1, are required. a manner such that the solution is protected from atmospheric
carbon dioxide (CO2) by means of a guard tube containing
soda lime or soda non-fibrous silicate absorbent, and such that
7. Reagents
it does not come into contact with cork, rubber, or saponifiable
7.1 Purity of ReagentsReagent-grade chemicals shall be stopcock grease. Standardize frequently enough to detect
used in all tests. Unless otherwise indicated, it is intended that normality changes of 0.0005 by potentiometric titration of
all reagents shall conform to the specifications of the Commit- weighed quantities of potassium acid phthalate dissolved in
tee on Analytical Reagents of the American Chemical Society, CO2-free water.
where such specifications are available.5 Other grades may be
used, provided it is first ascertained that the reagent is of 7.11 Propan-2-ol (Isopropyl Alcohol)Anhydrous, (less
sufficiently high purity to permit its use without lessening the than 0.1 % H2O). (WarningFlammable.) If dry reagent
accuracy of the determination. cannot be procured, dry it by distillation through a multiple
plate column, discarding the first 5 % of material distilling over
7.2 Purity of WaterUnless otherwise indicated, references and using the 95 % remaining. Also, drying can be accom-
to water shall be understood to mean reagent water that meets plished using molecular sieves by passing the solvent upward
the requirement of either Type I, II, or III of Specification through a molecular sieve column using one part of molecular
D1193. sieve per ten parts of solvent. (WarningIt has been reported
7.3 Buffer, Aqueous AcidCommercial pH 3 buffer solution that, if not inhibited against it, propan-2-ol can contain
with a tolerance of 60.02 pH units at 25C. This solution shall peroxides. When this occurs, an explosive mixture is possible
be replaced at regular intervals consistent with its stability or when the storage vessel or other equipment such as a dispens-
when contamination is suspected. Information related to the ing bottle, are near empty and approaching dryness.
stability should be obtained from the manufacturer. 7.12 Commercially available solutions may be used in place
of laboratory preparations provided the solutions have been
certified as equivalent.
5
Reagent Chemicals, American Chemical Society Specifications, American
Chemical Society, Washington, DC. For Suggestions on the testing of reagents not
7.13 Alternate volumes of solutions may be prepared pro-
listed by the American Chemical Society, see Annual Standards for Laboratory vided the final solution concentration is equivalent.
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, 7.14 TolueneReagent grade. (WarningExtremely flam-
MD. mable.)
Copyright by ASTM Int'l (all rights reserved); Sun Sep 23 18:56:57 EDT 2012
3
Downloaded/printed by
D4739 11
7.15 Titration SolventIn a brown reagent bottle, add 9. Standardization of Apparatus
30 mL of water to 1 L of isopropyl alcohol, and mix thor- 9.1 Determination of Meter Readings for the Aqueous
oughly. Add 1 L each of toluene and chloroform, and mix Buffer SolutionEnsure comparable selection of end points
thoroughly. when definite inflection points are not obtained in the titration
curve. Determine daily, for each electrode pair, the meter
8. Preparation of Electrode System readings obtained with the aqueous acidic buffer solution to be
8.1 Maintenance and Storage of ElectrodesCleaning the used for the determination of base numbers and with the
electrode thoroughly, keeping the ground-glass joint free of aqueous basic buffer solution to be used for the determination
foreign materials, and regular testing of the electrodes are very of strong base numbers.
important in obtaining repeatable potentials, since the contami- 9.2 Prepare the electrodes as described in 8.2, immerse them
nation may introduce uncertain erratic and unnoticeable liquid in the appropriate aqueous buffer solution, and stir for at least
contact potentials. While this is of secondary importance when 2 min, maintaining the temperature of the buffer solution at a
end points are chosen from inflection points in the titration temperature within 2C of that at which the titrations are to be
curve, it may be very serious when end points are chosen at made. Read the cell voltage. The reading so obtained in the
experimentally determined cell potentials as outlined in the acidic buffer solution is taken as the end point for the base
procedure. number if an inflection is not observed as specified in 12.1, and
8.1.1 Clean the pH indicating electrode or the pH indicating the reading obtained in the basic buffer solution is taken as the
part of the combination electrode at frequent intervals based on end point for the strong base number.
use and type of samples being analyzed by immersing in
10. Preparation of Sample of Used Oil
non-chromium containing, strongly oxidizing cleaning solu-
tion. The electrode shall be cleaned periodically when in use or 10.1 Strict observation of the sampling procedure is neces-
when a new electrode is installed. Drain the LiCl electrolyte sary, since the sediment itself is acidic or basic or has absorbed
from the electrode at least once each week, and refill with fresh acidic or basic material from the sample. Failure to obtain a
LiCl electrolyte as far as the filling hole. Ensure that there are representative sample causes serious errors.
no air bubbles in the electrode liquid. If air bubbles are 10.1.1 When applicable, refer to Practice D4057 (Manual
observed, hold the electrode in a vertical position and gently Sampling) or Practice D4177 (Automatic Sampling) for proper
tap it to release the bubbles. Maintain the electrolyte level in sampling techniques.
the electrode above that of the liquid in the titration beaker at 10.1.2 When sampling used lubricants, the specimen shall
all times. be representative of the system samples and shall be free of
contamination from external sources. As used oil can change
8.1.2 When not in use, immerse the lower halves of the
appreciably in storage, test samples as soon as possible after
electrodes in either water (sensing) or the LiCl in isopropyl
removal from the lubricating system, and note the dates of
alcohol electrolyte (reference). Do not allow them to remain
sampling and testing.
immersed in titration solvent for any appreciable period of time
between titrations. While the electrodes are not extremely 10.2 Heat the sample (Note 1) of used oil to 60 6 5C in the
fragile, handle them carefully at all times. original container, and agitate the sample until all of the
8.1.3 Electrode LifeTypically, electrode usage is limited sediment is homogeneously suspended in the oil. If the original
to 3 to 6 months, depending upon usage. Electrodes have a container is a can, or if it is glass and more than three-fourths
limited shelf life and shall be tested before use (8.3). full, transfer the entire sample to a clear glass bottle having a
capacity at least one-third greater than the volume of the
8.2 Preparation of Electrodes: sample. Transfer all traces of sediment from the original
8.2.1 When Ag/AgCl reference electrode is used for the container to the bottle by vigorous agitation of portions of the
titration and it contains an electrolyte that is not 1 to 3 M LiCl sample in the original container.
in ethanol, replace the electrolyte. Drain the electrolyte from 10.3 After complete suspension of all sediment, strain the
the electrode, wash away all the salt (if present) with water and sample or a convenient aliquot through a 100-mesh screen for
then rinse with ethanol. Rinse several times with the LiCl the removal of large particles.
electrolyte solution. Finally, replace the sleeve and fill the
electrode with the LiCl electrolyte to the filling hole. When NOTE 1When samples are visibly free of sediment, the heating
procedures described may be omitted.
refitting the sleeve, ensure that there will be a free flow of
electrolyte into the system. A combination electrode shall be 11. Procedure for Base Number and Strong Base
prepared in the same manner. The electrolyte in a combination Number
electrode can be removed with the aid of a vacuum suction. 11.1 Calculate the quantity of sample required for its
8.2.2 Prior to each titration, soak the prepared electrodes in expected base number as follows:
water, pH 4.5 5.5 acidified with HCL, for at least 5 min.
A 5 7/E (1)
Rinse the electrode with propan-2-ol immediately before use,
and then with the titration solvent. where:
8.3 Testing of ElectrodesSee Appendix X2 for the proce- A = approximate mass of sample, g and
dure to check electrode performance. E = expected base number.
Copyright by ASTM Int'l (all rights reserved); Sun Sep 23 18:56:57 EDT 2012
4
Downloaded/printed by
D4739 11
11.1.1 Take a maximum of 5 g and a minimum of 0.1 g for
analysis. The precision of weighing is as follows:
Size of Sample, g Precision of Weighing, g
15 0.005
0.11 0.002
Copyright by ASTM Int'l (all rights reserved); Sun Sep 23 18:56:57 EDT 2012
5
Downloaded/printed by
D4739 11
ANNEX
(Mandatory Information)
A1. APPARATUS
A1.1 Apparatus for Manual Titration touching with a grounded lead (Note A1.1), any part of the
A1.1.1 MeterA voltmeter or potentiometer that will oper- exposed surface of the glass electrode, the glass electrode lead,
ate with an accuracy of 60.005 V and a sensitivity of the titration stand, or the meter. A desirable apparatus may
60.002 V, over a range of at least 60.5 V, when the meter is consist of a continuous-reading electronic voltmeter with
used with the electrodes specified in A1.1.2 and A1.1.3, and specified range, accuracy, and sensitivity, that is designed to
when the resistance between the electrodes falls within the operate on an input of less than 5 1012 A when an electrode
range from 0.2 to 20 MV. The meter shall be protected from system having 1000 MV resistance is connected across the
stray electrostatic fields so that no permanent change in the meter terminals, and that is provided with a satisfactory
meter readings over the entire operating range is produced by terminal to connect the shielded connection wire from the glass
Copyright by ASTM Int'l (all rights reserved); Sun Sep 23 18:56:57 EDT 2012
6
Downloaded/printed by
D4739 11
electrode to the meter without interferences from the presence A1.1.6 Titration StandA suitable stand to support the
of external electrostatic field. electrodes, stirrer, and buret. An arrangement that allows the
NOTE A1.1Grounded or connected to the ground means connected removal of the beaker without disturbing the electrodes, buret,
through a resistance of not more than 100 V to a standard ground potential and stirrer is desirable.
such as that of a water-service pipe.
A1.1.2 ElectrodesSee 6.2 and 6.3. A1.2 Apparatus for Automatic Titration
A1.1.3 StirrerA variable-speed mechanical stirrer of any A1.2.1 Automatic titration system shall be generally in
suitable type, equipped with a glass, propeller-type-stirring accordance with A1.1 and provide the following technical
paddle.A propeller with blades 6 mm in radius and set at a pitch performance characteristics of features:
of 30 to 45 is satisfactory. A magnetic stirrer is also satisfac-
tory. If electrical stirring apparatus is used, it must be grounded A1.2.1.1 The addition of titrant must be automatically
so that connecting or disconnecting the power to the motor will controlled to dispense discontinuously 0.100 6 0.001-mL
not produce a permanent change in meter reading during the increments of titrant, with a waiting period of 90 s between
courses of titration. increments.
A1.1.4 BuretA 5-mL buret graduated in 0.01-mL divi- A1.2.1.2 Interchangeable precision motor-driven burets
sions and calibrated with an accuracy of 60.005 mL. The buret with volume dispensing accuracy of 60.001 mL.
shall have a chemically-resistant stopcock and shall have a tip A1.2.1.3 A record of the complete course of a titration by
that extends 100 to 130 mm beyond the stopcock. continuously printing out the potential or change in potential,
A1.1.5 Titration BeakerA 250-mL beaker made of boro- with the addition of each increment of titrant versus volume of
silicate glass, or other suitable titration beaker. titrant added.
APPENDIXES
(Nonmandatory Information)
X1.1 A long, equilibration period of 90 s/increment was tions has not been determined.
selected for the base number titration because the titration
reaction and electrode equilibration are generally slow. This, of X1.2 There are many cases where the optimum in precision
course, can lead to long titration time/sample, with a maximum in the base number is not required, and in these cases the
time of 1 h based on a maximum volume of titrant of 4 mL and titration time can be shortened by taking a smaller sample. For
a rate of titrant addition of 0.1 mL/90 s. It is possible to example, for a base number of 2 and using a sample size
substantially reduce the titration time by predosing with rapid calculated from Eq 1, a total titration time of 24 min would be
addition of titrant until a potential within 25 mV of the buffer required. By taking only one half of the prescribed sample size
potential is reached, then allowing 90 s for equilibration and the titration time would be reduced to 12 min. The affect of
completing the titration under normal conditions. This proce- halving the sample size on the precision of this test method has
dure is not expected to have an adverse affect on the precision not been determined, but it would be expected to be small.
of this test method; however, the precision under these condi-
X2.1 The kinetic electrode test measures the kinetic re- reaches a stable point in less than 30 s and changes little from
sponse of the electrode. Electrodes can calibrate with accept- 30 to 60 s. Use buffers pH 4, pH 7, and pH 11 for this check,
able slope and intercept values yet still not have a response as needed.
good enough for titration. The speed of response and subse-
quent stability is important for a titration electrode. A manual X2.3 Procedure
method is described in X2.3 that can be carried out with a pH X2.3.1 Set the titrator or pH meter to read millivolts
meter or titrator set to read millivolts continuously. continuously. Have provision for stirring the buffer solution at
X2.2 The essence of the test method is to challenge the the same speed used for the titrations.
electrode coming from rest in a water solution with buffers and X2.3.2 Allow the electrode to stabilize for 1 min in distilled
measure the potential after 30 s and 60 s. A fast electrode or equivalent deionized water.
Copyright by ASTM Int'l (all rights reserved); Sun Sep 23 18:56:57 EDT 2012
7
Downloaded/printed by
D4739 11
X2.3.3 Remove the electrodes from the water, and place drift < 1 excellent
them in the pH 4 buffer. Start a stopwatch at about the moment 1 < drift < 2 good
2 < drift < 3 acceptable
when the buffer touches the electrode. After 30 s, note the 3 < drift < 4 questionable
potential. 4 < drift unacceptable
X2.3.4 After another 30 s, note the potential again. The
X2.5 The difference between the 60 s potentials for pH 4
difference between the two potentials is termed the drift.
buffer and pH 7 buffer should be greater than 162 mV, or
X2.3.5 Repeat the procedure for pH 7 buffer and pH 11 54 mV/pH number. Electrodes with a slope less than 54
buffer. mV/pH number are not reliable for titration.
X2.4 Calculate the drift for each of the three buffers. The
electrode response may be judged as follows:
SUMMARY OF CHANGES
Subcommittee
1
D02.06 has identified the location of selected changes to this standard since the last issue
(D473908 ) that may impact the use of this standard.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT/).
Copyright by ASTM Int'l (all rights reserved); Sun Sep 23 18:56:57 EDT 2012
8
Downloaded/printed by