Sensor SAmpling Frequency
Sensor SAmpling Frequency
Sensor SAmpling Frequency
High levels.
Wide frequency content of pulses.
Transient characteristics of instrumentation.
As a result, each part of the instrumentation system should be evaluated and selected for:
Kulite pressure transducers provide a unique combination of high performance in all of these
critical characteristics.
In some applications, where true steady-state measurements are not required, and where low
frequency drift may be a problem, ac coupling can be used.
When ac amplifiers or other equipment with limited low frequency response are connected in a
system, the pulse wave shape will not be maintained. The nature of this inaccuracy can be seen
by examining the effect on a rectangular pulse of duration T and amplitude A applied to the input
of a signal conditioner which does not respond to dc (steady state) signals.
If this transient is passed through an ac system with first-order low frequency response, the
resultant output will be as shown in Figure 7-1.
A first-order system has the same low frequency response as a single resistor-capacitor high-
pass filter whose time constant in seconds is equal to RC (ohms x farads). Such a system exhibits
a low frequency cutoff equal to 1/ 2! RC and is 5% down at a frequency of 3/ 2! RC. A critically
damped mechanical system has a first-order response.
The output does not remain at the peak value for the full pulse duration, but decays
exponentially, "droop." The output amplitude at any time, t, (during the pulse) can be expressed
as:
where RC is the system time constant. At the termination of the pulse, the output does not return
to zero, but overshoots in a negative direction. Recovery from this "undershoot" occurs at the
same exponential rate as droop.
Figure 7-1: Response of System With First Order LF Response to a Rectangular Pulse
The ratio of total pulse height to droop is a function of the ratio RC/T. The larger this ratio, the
less error (and the less undershoot). For example, if this ratio is 20, there will be approximately
5% error in the rectangular pulse amplitude; if the ratio is 50, there will be only a 2% error.
Although slightly more complex to analyse, it can be shown that similar low frequency effects
occur for other pulse shapes. If the requirement for adequate RC/T is not satisfied, it is possible
to predict the degree of error for these pulses and apply appropriate correction factors to the data
obtained.
Figure 7-2: Response of System with First Order HF Response to a Rectangular Pulse
It is also of interest to note the effect of passing the rectangular pulse through a system
possessing second-order high frequency response (Such a system corresponds to the electrical
frequency response of a single LC low-pass filter combination, or a damped mechanical system).
Figure 7-3 shows the resulting output, A high-frequency ringing at approximately the resonance
frequency is superimposed on the pulse. The amplitude and duration of the ringing depends on
the damping factor.
Figure 7-3: Response of System with Second Order HF Response to a Rectangular Pulse
Fourier analysis show that short transients ,contain significant high-frequency components.
Faster rise-time transients contain higher frequency components. Both the transducer and
associated systems must have adequate high frequency response to avoid undesirable
measurement distortion.
Although the diaphragm is actually a higher-order structure with multiple high-frequency modes,
the high frequency response of a piezoresistive transducer with very little damping is
approximately a second-order function and is determined by the transducer fundamental (first)
modal resonance frequency. The use of such devices provides desirable high frequency response
along with minimum phase shift in the frequency range of interest.
Transients, however, may excite such a transducer's first mode to resonance; natural frequency
"ringing" will then be superimposed on the basic input transient.
In the case of short, rectangular or other transients with essentially zero rise time (very short rise
time in proportion to the natural period of the transducer), almost 100% overshoot on the
transient may occur along with subsequent excitation of transducer natural frequency. To
minimise or prevent these distortions, the transducer should have a natural period (the reciprocal
of the natural frequency) one-third the expected rise time or less.
Resonance frequencies should be as shown in the following table in order that the natural period
be one-fifth the pulse duration for undamped transducers measuring half-sine or sawtooth
transients:
The most common types of readout devices are: (1) Oscilloscope, (2) Computer based dynamic
data capture devices. The storage oscilloscope is probably the most versatile and easy to use. A
good quality scope will have a response from dc to above a gigahertz so that it will not introduce
significant errors.
In some cases, when rise times are extremely short, it may become necessary to resort to
electrical filtering. The data in the pass band of a low-pass filter will have quantitative value,
even if the transducer is resonating, as long as the filter has a linear phase-shift characteristic
(constant delay). When filtering is used in the measuring system, the actual and recorded
transients may differ widely. For this reason, it is strongly recommended that the unfiltered
response from the transducer be recorded, as well as the filtered signal. The introduction of
filtering into the measurement chain can cause phase shifting of the data.
In many transient measurement applications, only certain frequencies (within the Fourier
spectrum of a pulse) are of interest. Unwanted noises or system transients of one kind or another
must be eliminated. In these cases filtering is introduced in the system to allow only the desired
frequencies to pass. Even though the filter passes the desired components without attenuation,
the phase relation of a signal at one frequency with respect to a signal at another may be
changed; this causes an error in the composite wave shape. To minimise this, the filter should be
designed to have a constant time delay within its passband. In this way the phase relation
between the frequency components of a transient is maintained and only a time shift of the entire
composite wave shape will occur.
In essence, air blasts involve shock waves. A shock wave is defined as a pressure wave
characterised by a very steep, almost discontinuous, rise in pressure which occurs when a region
of high pressure overtakes a region of low pressure, with a consequent rapid compression of the
medium. The duration of a shock wave is distinguished by two phases. First, there is the positive
(or compression) phase during which the pressure rises very sharply to a value that is higher than
ambient and then decreases rapidly to the ambient pressure. The positive phase for the dynamic
pressure is somewhat longer than for overpressure, due to the momentum of the moving air
behind the shock front. The duration of the positive phase increases and the maximum (peak)
pressure decreases with increasing distance from an explosion of given energy yield. In the
second phase, the negative (or suction) phase, the pressure falls below ambient and then returns
to the ambient value. The duration of the negative phase is approximately constant throughout
the blast wave history and may be several times the duration of the positive phase. Deviations
from ambient pressure during the negative phase are never large, and they decrease with
increasing distance from the explosion.
A transducer that is exposed to a step pressure change will be excited at its resonance frequency
if the rate of pressure rise is sufficiently fast. The rise time for a shock wave will vary, depending
upon the blast source and intensity, distance from source and gas dynamic properties of the
shocked medium. Nonetheless, rise times can sometimes be less than a microsecond. Damped
oscillations will show up on the output of virtually any air blast pressure measurement that is not
overdamped since very few (if any) sensing elements have natural frequencies that would not be
excited
by high frequency energy in the shock wave. Many Kulite pressure transducers have natural
frequencies of 500 kHz or higher.
Referring to the example above, this ringing response is clearly evident in the initial response of
the transducer. Sources of this oscillation may be mechanical resonance of the transducer sensing
element, acoustical resonance of ports and cavities, or aerodynamic oscillations in the flow set
up by diffraction of the initial and succeeding shocks and rarefaction waves over the probe.
Three major problems are common to all pressure measurements in the extreme close-in region
to air blasts: (1) survival of the gage and cabling, (2) thermal isolation of the sensor, and (3)
minimising effects of mechanical shock motion on the transducer Survival needs include
pressure transducer. To provide the high frequency response, the transducer must be coupled to
the pressure in close-to-flush mounted configuration, but the diaphragm :must also be protected
from high speed particles. This is sometimes accomplished by using a mechanical screen or
baffle (a plate with offset holes in front of the diaphragm ). Thermal protection of the diaphragm
is also required for a short time period. Carbon-filled vacuum grease is often placed in front of
pressure transducer diaphragms to accomplish this. Mechanical shock motion can be at
thousands of g'S. To minimise errors, acceleration sensitivity of the transducer must be low
along with a high resonance frequency.
In reality, diaphragms of the type being discussed typically deflect in accordance with theory
applicable to rectangular plates. This theory does not predict only one natural frequency, but an
infinite number of natural frequencies for the various plate vibration modes.
Practically, only a finite number of plate natural frequencies are of importance due to damping in
the plate structural material.
Mounting threads should be fully engaged, but contact with the relieved area at the front of the
transducer should be avoided. Installation torque recommendations should be observed as well as
recommended specifications for mounting hole dimensions and perpendicularity. Only the 0-
rings or crushwashers provided or recommended by Kulite should be used for sealing.
The cable should be restrained to prevent whipping and flexing. Cable tension should be avoided
by providing strain relief cable routing and/or clamping. The cable should also be protected from
crushing. Where the cable might be subjected to high pressure pulses or shock waves, it should
be protected in a rigid conduit.