Photon Beam Characteristics & Basic Concepts of Treatment Planning - Dr. K J Maria Das PDF
Photon Beam Characteristics & Basic Concepts of Treatment Planning - Dr. K J Maria Das PDF
Photon Beam Characteristics & Basic Concepts of Treatment Planning - Dr. K J Maria Das PDF
Dose distribution
Absorbed X
completely removed from beam
ceases to exist
Scattered
change in direction
no useful information carried
source of noise
Nothing
Photon passes unmolested
Interaction depends on
Basic Interactions:
Coherent Scattering
Compton Scattering
Photoelectric Effect
Pair Production
Photodisintegration
Beam Characteristics
Quantity
number of photons in beam
1, 2, 3, ...
~
~
~ ~
~
Beam Characteristics
Quality
energy distribution of photons in beam
~ ~ 10 20 30 40 50
Energy
60 70 80
~ ~
~
~ ~
~
Beam Characteristics
Intensity
weighted product of number and energy of
photons
depends on 324 mR
quantity
~
quality
~ ~
~ ~ ~
~ ~
Beam Intensity
-
~
+
Attenuation Coefficient
kVp
(as set on
generator)
X-Ray Beam Attenuation
Fraction .1
Transmitted Polychromatic
.01
Monochromatic
.001
Attenuator Thickness
Photoelectric vs. Compton
Small
Photoelectric vs. Compton
= coherent + PE + Compton
As photon energy
As atomic # increases
increases
Fraction of that is PE increases
Both PE & Fraction of that is Compton
Compton decrease decreases
Interaction
PE decreases faster Probability
Fraction of that
is Compton
increases
Fraction of that Compton
is PE decreases
Photoelectric
Photon Energy
Sources of radiation that determine dosimetric
characteristics of clinical photon beams
Inverse Square Law & Field Divergence
Ds = Surface dose
BUR
Dose buildup
Buildup of dose increases with
increase in energy of the beam.
The region between the surface
and the point of maximum dose
is called the dose buildup-
buildup-
region.
Kerma-- (1) kinetic energy released in the medium; (2) the energy
transferred from photons to directly ionizing electron; (3) maximum at the
surface and decreases with depth due to decreased in the photon energy
fluence; (4) the production of electrons also decreases with depth
Dd
Dmax = ------ x100
P
PDD depends on
beam energy
depth
field size d0
x Dd0
distance from source d
beam collimator system x Dd
PDD dependence on SSD Mayneord F Factor
2 2
Photon fluence from a point source
f 2 + d m f1 + d
varies inversely as a square of the
F =
f1 + d m f 2 + d
distance from the source. (SSD> 80 80cm)
cm)
PDD increase with SSD
Dd
P = 100
D max
F1+dm
F2+dm
F1+d F2+d
Normalized Depth Dose Data
Energy Dependence
Percentage Depth Dose
Characteristics
Percentage Depth Dose
Field Size Dependence
Percentage Depth Dose
Wedge/Open Comparison
Normalized Depth Dose Data
Wedge/Open Comparison
Effect of field size and shape
Field size increases the scatter increases. Scattered dose is greater at larger
depth than at the depth of Dmax. PDD increases with increasing field size.
Field size dependence of PDD is less for higher energy than for lower energy
beams
d t0
rd Dd rd Dt0
Properties of TMR
TMR is independent of SSD, increases with energy and field size.
TAR (d , rd )
TMR(d , rd ) =
BSF (rd )
TMR
3030
TMR(d ,0) = e ( d t0 )
1010
00
Is caused entirely by
the primary beam
Depth in water
TMR data for 10 MV x-ray beams
Collimator Scatter Factor (Sc)
The beam output measured in air depends on the
field size
Field size ; output ; collimator scatter
Output factor
Definition
Mini phantom
The ratio of the output in air for a given field to that
for a reference field (10 x 10 cm)
Direct measurement
Sc = D(r)/ D(10)
Phantom Scatter Factor (Sp)
The change in scatter radiation originating in the phantom
reference depth as the field side is changed
Definition
The ratio of the dose rate for a given field at a reference depth (e.g.
depth of Dmax) to the dose rate at the same depth of the reference
field size (10 x 10 cm), with the same collimator opening
Related to the change in the volume of the phantom irradiated
Photon Beam Penumbra
The penumbra region
The dose rate decreases rapidly as a function of lateral
distance from the beam axis.
The width of geometric penumbra depends on source size,
distance from the source, and source-to-diaphragm
distance.
10
Symmetry
within 2% over
80% of the field
Profile characteristics
15 MV Photon Beam, Field size of 15x15cm2, Depth 2.5, 5.0, 10, 15, 20 cm
The field flatness changes with depth. This is attributed to an increase in scatter to primary
dose ratio with increasing depth and decreasing incident photon energy off axis
Cross Beam Profile
6 MV Photon Beam, Depth of 5.0 cm, Field size of 4x4, 10.4x10.4, and 21x21cm2
The flatness of photon beams is extremely sensitive to change in energy of the incident beam.
A small change in the penetrative quality of a photon beam results in very large change in
beam flatness.
Effect of Electron Steering
Beam Flatness
Beam Quality
The depth of a given isodose curve increases with beam quality.
Greater lateral scatter associated with lower-energy beams
For megavoltage beams, the scatter outside the field is
minimized as a result of forward scattering and becomes more a
function of collimation than energy.
200 kVp,
SSD=50 cm
60Co, SSD=80 cm 4 MV, SSD=100 cm 10 MV, SSD=100 cm
Isodose distribution
Co-60, 6 & 15 MV Photon Beam, Field size of 10x10cm2
The field flatness changes with depth. This is attributed to an increase in scatter to primary
dose ratio with increasing depth and decreasing incident photon energy off axis
Isodose distribution
Co-60, 6 & 15 MV Photon Beam, Field size of 10x10cm2
The field flatness changes with depth. This is attributed to an increase in scatter to primary
dose ratio with increasing depth and decreasing incident photon energy off axis
Isodose distribution
Field size of 20x20cm2
Note contaminant electrons contribute to dose outside the field at shallow depths.
The magnitude and extent of dose outside the geometric edge of a field at shallow
depths increases with beam energy.
Cross section isodose distribution
Co-60, 6 & 15 MV Photon Beam, Field size of 10x10cm2
The field flatness changes with depth. This is attributed to an increase in scatter to primary
dose ratio with increasing depth and decreasing incident photon energy off axis
Problem in beam modification
Radiation reaching any point, is made
up of primary and scattered photons.
Used in Linac
Wedge profile at 5cm depth (45 degree)
Wedge angle
The wedge isodose angle () is the
complement of the angle through
which the isodose curve is tilted
with respect to the central ray of
the beam at any specified depth.
This depth is important because
the angle will decrease with
increasing depth.
The choice of the reference depth
varies:
10 centimeters.
1/2 - 2/3rd of the beam width.
At the 50% isodose curve
(kV).
It is angle is defined as the complement of the angle through which the isodose
curve with respect to the beam central axis at reference depth of 10cm.
(ICRU Report N0.24)
Beam modifiers
Field blocking and shaping devices:
Shielding blocks.
Custom blocks. The higher scatter
contribution to the overall
dose results in lower
Asymmetrical jaws. dosage adjacent to the
shielded area in kilovoltage
Multileaf collimators. radiation.
100%
Lesser amount of
scattered radiation with
250 KV megavoltage radiation
means that the
50%
attenuation produced by
shielding is also more.
4 MV
Shielding blocks
To spare the critical organ & Normal tissue Beam Quality 5 HVL Lead (cm)
Should be at least 5 HVL (3.125%) ; 1/2n = % Cs 137 3.0
transmission Co 60 5.0
Made of
4 MV 6.0
Lead 11.3gm/cm3
Cerro-bend alloy 9.4gm/cm3 6 MV 6.5
(Bi 50%, Pb 26.7%, tin 13.3% & Cd -10% ) 10 MV 7.0
15 MV 7.0