Basic Principles of Magnetic
Basic Principles of Magnetic
Resonance Imaging
Robert-Jan M. van Geuns, Piotr A. Wielopolski, Hein G. de Bruin,
Benno J. Rensing, Peter M.A. van Ooijen, Marc Hulshoff,
Matthijs Oudkerk, and Pim J. de Feyter
The Source of the MR Signal From the Department of Cardiology, Thoraxcenter, and
the Department of Radiology, Dr Daniel den Hoedkliniek,
A correct description of what happens when University Hospital Rotterdam, Rotterdam, The Nether-
tissue is subjected to a magnetic field relies on lands.
quantum mechanics. Fortunately, all the theory Address reprint requests to Robert-Jan M. van Geuns, MD,
Department of Cardiology, Thoraxcenter, BD 377, University
necessary for MRI can be based on a simple
Hospital Rotterdam, PO Box 2040, 3000 CA Rotterdam, The
classical model in which certain nuclei that spin Netherlands; e-mail: [email protected].
around their own axes behave like small magnets. Copyright 1999 by W.B. Saunders Company
For clinical imaging, hydrogen is the most fre- 0033-0620/99/4202-0003$10.00/0
Progress in Cardiovascular Diseases, Vol. 42, No. 2 (September/October), 1999: pp 149-156 149
150 VAN GEUNS ET AL
tems of 1.5 Tesla (T), the Larmor frequency will patient, two phenomena occur: first, enough
be 63.75 MHz for hydrogen. It is of note that the protons absorb energy to jump from the parallel
phase of precession around the axis of the magnetic state to the higher level of the antiparallel state,
field is different for each individual nucleus (Fig 2B). and second, the spins are whipped to precess in
phase. The effect of all this is that the net
magnetization (Mz) flips 90 from the positive
Excitation
z-axis to transverse plane (Fig 3). The net magne-
The net magnetization vector from the nuclei tization in the transverse plane rotates around B0
inside the magnet in its equilibrium state is static at the Larmor frequency. This rotating transverse
and does not produce a measurable signal. To magnetization can be measured, because it will
obtain information from the spins, the direction induce an alternating current (AC) in the receiver
of the net magnetization vector has to be altered. coil placed around the patient.
For this the precessing spins are excited by
applying energy, in the form of radiofrequency
Return to Equilibrium
(RF) energy pulses of exactly the Larmor fre-
quency (resonance frequency). When an RF sig- After the RF frequency transmitter is switched off,
nal is given at the resonance frequency into the the equilibrium state will be sought (high energy
back to low energy). This means that the magneti- ation and longitudinal relaxation. Both processes
zation decays over time, which is represented by a are independent. The process of realignment to
decreasing magnitude of Mz in the transverse the external magnetic field is called the longitudi-
plane. Consequently, the induced signal in the nal relaxation process (Fig 5). It is characterized
receiver coil will decrease in time. This decreasing by the T1, relaxation time. The T1 relaxation time
signal is called the free induction decay (FID) is defined as the time required for the system to
(Fig 4). The time required for the signal to return recover to 63% of its equilibrium value after it has
to equilibrium is the relaxation time. been exposed to a 90 RF pulse (Fig 6). Various
Two relaxation processes exist: transverse relax- human tissues have different T1 values (Table 1).
The second process of relaxation, the trans-
verse relaxation, depends on the spins precessing
around the magnetization vector. Initially, after
the excitation by the RF pulse, the spins precess
completely in phase. However, as time passes, the
observed signal starts to decrease because the
spins begin to dephase due to small differences in
the Larmor frequency induced by random local
magnetic inhomogeneities, due to spin-spin inter-
action and inhomogeneity of the main static
magnetic field B0. This process is called the transverse
relaxation or spin-spin relaxation and is character-
ized by T2 relaxation time (Fig 5). The T2 relaxation
time is the time it takes for dephasing to decay the
signal to 37% of its original value (Fig 6). The T2
time from various tissues is different, but the T2
time is always shorter than the T1 time.
Spatial Encoding
Fig 4. The received signal detected by the receiver
coil, the FID, decreases over time when the net magne- To create an image, the MR signal from the
tization vector returns to its original orientation. H-protons has to contain information about where
152 VAN GEUNS ET AL
these H-protons are positioned in the patient. combinations of gradients in all three directions,
This is done in three steps: slice selection, fre- it is possible to acquire a slice in any arbitrary
quency encoding, and phase encoding. direction through the body.
To select an imaging slice through the body, a The frequency and phase encoding are used to
magnetic gradient (for example 25 milliT/m) is obtain information for the individual points within
added along the main magnetic field in the caudal a slice, the picture elements or pixels. For the
to cranial direction. Because the frequency of phase encoding, a short temporary change in the
precession, and thus the frequency at which the magnetic field is applied between the RF excita-
spins can be excited, is dependent on the local tion pulse and the readout of the signal. This
strength of the magnetic field, a narrow band of change in the magnetic field will influence the
frequencies will only excite a thin slice (3 to 8 frequency of precessing, resulting in a shift in the
mm) of spins through the body (Fig 7). With a phase of precessing of the spins dependent on the
change in the excitation frequency another paral- duration of this gradient switch. By repeating this
lel slice can be acquired later. To obtain slices in process with different duration of the temporary
other directions, for example vertical slices, the gradients, signals with a different phase encoding
direction of gradients for the slice encoding are are acquired (Fig 8).
altered to an anterior-posterior gradient. By using The frequency encoding is used to differentiate
pixels with the same phase encoding. A magnetic
gradient during readout of the signal results in a
specific shift of the resonance frequency, likewise
the effect of the slice-encoding gradient, for pixels
with the same phase shift (Fig 9).
Fig 12. Contrast between different tissues. Two tissues each with a different T1 and T2 value have different signal
intensity dependent on the time. The contribution of T1 to the signal intensity of a pixel is dependent on the TR used,
indicated in the figure for a short and a long TR. From these points, the signal intensity decays along the T2 curve.
Using a short TR and TE, the signal intensity and contrast will be mainly influenced by the differences in T1 value of
the tissues (T1 weighted imaging). A combination of a long TR and a short TE will result in contrast dependent on the
proton density, whereas a long TR and a long TE will pronounce difference in T2 value (T2 weighted imaging).
BASIC PRINCIPLES OF MRI 155
Resolution
In digitized imaging, such as MRI, pictures are
composed of a matrix of elements, called picture
elements or pixels. The image represents the field
of view (FoV). The image matrix defines the
Fig 13. In MR imaging, resolution is determined by the number of pixels used to construct an image that
pixel size, which is dependent on the number of lines is determined by the number of frequency encod-
in the x and y direction used in the matrix and the FoV
covered with this matrix. Pixel size FoV (x)/No. lines ings (128 or 256 on the x-axis) and the number of
(x) X FoV (y)/No. lines (y). phase-encoding steps used (128 or 256 on the
y-axis) for a certain FoV (Fig 13). Therefore, the
MR techniques using the combination of a 90 FoV, the matrix size used, and the slice thickness
and a 180 RF pulse to generate an echo signal are determine the volume of each pixel.
called spin-echo sequences. In practice the resolution is determined by the
pixel size (the smaller, the higher the resolution),
but the signal-to-noise ratio is the limiting factor
Contrast if the pixels become to small and do not contain
MRI has the potential to visualize the difference enough spinning protons to produce a measur-
in T1 and T2 of different tissues. Using these able signal.
differences, contrast between different soft tissues
in MRI is superb compared with x-ray computer
Summary
tomography. If the time for the next repetition of
RF pulses (time of repetition TR) is shorter To obtain a single MR image, a complex combina-
than the time necessary for total longitudinal tion of RF pulses and magnetic gradient switches
relaxation, the contrast in the image will be have to be applied. In spin-echo imaging, two RF
Fig 14. Example of a spin-echo sequence. An echo signal is evoked after a 90 and a 180 RF pulse. Spatial
encoding of the echo signal is performed with gradient switches in three directions. The slice selection gradient is
present during RF excitation, which will now only excite a thin slice through the body. The phase-encoding gradient
is shortly present between the excitation and the sampling of the echo signal. Finally, the frequency-encoding
gradient is activated during the echo acquisition. This combination of RF pulses and gradient switches is repeated
128 or 256 heartbeats with different phase-encoding steps. Imaging time will be approximately 3 to 4 minutes for a
single cardiac triggered image.
156 VAN GEUNS ET AL