This document describes a new hybrid technique for real-time ECG data compression. It combines two existing techniques called SAPA-2 and Adaptive compression. The new approach is tested on single-lead ECG signals and compared to other well-known compression techniques. Testing shows the new hybrid technique achieves higher compression ratios while maintaining clinical information in the reconstructed ECG signals.

Uploaded by

Sneha Singh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
61 views3 pages

Korurek

This document describes a new hybrid technique for real-time ECG data compression. It combines two existing techniques called SAPA-2 and Adaptive compression. The new approach is tested on single-lead ECG signals and compared to other well-known compression techniques. Testing shows the new hybrid technique achieves higher compression ratios while maintaining clinical information in the reconstructed ECG signals.

Uploaded by

Sneha Singh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

A NEW DATA COMPRESSION TECHNIQUE

FOR ELECTROCARDIOGRAMS

Murat ZVIN and Mehmet KORUREK


Technical U n i v e r s i t y of I s t a n b u l
E l e c t r i c a l and E l e c t r o n i c s Engineering Faculty
80626 Ayazaga, I s t a n b u l , TURKEY

A b s t r a c t : During t h e l a s t 20 years, a broad One of the scan-along-polygonal approximation


spectrum of d a t a reduction algorithms have been proposed techniques (SAPA-2) has got the t h e o r e t i c a l bases t h a t
f o r Electrocardiaogram (ECG) d a t a compression. Such a l - the d e v i a t i o n between t h e s t r a i g h t l i n e s and t h e o r i g i n a l
gorithms, t h a t o p e r a t e i n r e a l time have been v i t a l i n s i g n a l i s never more than t h e p r e s e t e r r o r t o l e r a n c e 131.
reducing t h e d i g i t a l ECG d a t a volume f o r s t o r a g e , t r a n s -
mission and a n a y s i s without l o s i n g c l i n i c a l information I n t h e p r e s e n t paper, an hybrid appritaeh based on
c n n t e n t . I n t h i s paper,a new approach f o r real-time ECG two o t h e r d i r e c t ECG d a t a compression algorithms, ADAP-
d a t a compression i s presented. This technique i s a h y b r i d TIVE and SAPA-2, i s a p p l i e d t o d i f f e r e n t single-lead ECG
of w e l l known techniques c a l l e d SAPA-2 and Adaptive signals.
compression algorithm. A s p e c i a l decompression algorithm
and smooth-curve approximation a r e used t o r e c o n s t r u c t
t h e compressed d a t a i n t o o r i g i n a l form. I n o r d e r t o
compare t h i s new approach on t h e same b a s e s with o t h e r Data A c q u i s i t i o n
well known techniques which a r e widely used i n prac-
t i c e , a l l t h e techniques a r e a p p l i e d t o t h e same s i n g l e The d i f f e r e n t s i n g l e l e a d analog ECG s i g n a l s ,
l e a d ECG s i g n a l . The comparison i s given i n a t a b l e . which a r e obtained from an ECG s i m u l a t o r , a r e converted
by a PCL-812 A/D c o n v e r t o r , i n o r d e r t o process with a
Introduction PC computer.

Compression of d i g i t a l E C G ' s in a more r e l i a b l e , Preprocessing


a c c u r a t e and e f f i c i e n t way has received g r e a t a t t e n t i o n
because of continuing p r o l i f e r a t i o n of computerized ECG The o r i g i n a l sampled d a t a t o b e approximated a r e
processing systems along w i t h t h e increased f e a t u r e per- r e p r e s e n t e d by the d i s c r e t e sequence X(i) where t h e in-
formance requirements and demand f o r lower c o s t medical t e g e r i is t h e sample number. Denote hx and X+ as
care. maximum and minimum value of samples, of which i n i t i a l
Data compression i s t h e process of d e t e c t i n g and v a l u e s a r e set t o be e q u a l t o t h e f i r s t sample X(1).
e l i m i n a t i n g redundancies i n a given d a t a s e t and has the L e t +E(%) r e p r e s e n t allowable, adaptive e r r o r i n t h e
main goal t o achieve maximum d a t a volume r e d u c t i o n while approximation and T( i ) a d a p t i v e t h r e s h o l d . When begining
preserving t h e s i g n i f i c a n t s i g n a l morpology f e a t u r e s t h e approximation algorithm, t h e f i r s t d a t a p o i n t i s
upon r e c o n s t r u c t i o n . The p r a c t i c a l importance of d a t a s t o r e d . Upon r e c e i p t of l a t e r salqple a t i + n , x(i+n) w i l l
compression i n computerized e l e c t r o c a r d i o g r a p h y can be be compared w i t h hX and &inand the normalized s l o p e s
shown i n following a s p e c t s : 1) increased s t o r a g e c a p a c i t y of t h e t h r e e l i n e s j o i n i n g X(i) t o X(i+n)+E and t o
of ECG's a s databases f o r subsequent comparison o r X(i+n)-E and f i n a l l y t o X(i+n) a r e c a l c u l a t e d a s ;
e v a l u a t i o n 2) f e a s i b i l i t y of t r a n s m i t t i n g r e a l time
E C G ' s over t h e p u b l i c phone network 3) implementation of
c o s t e f f e c t i v e real-time rhythm algorithms 4 ) economical
r a p i d transmission of o f f - l i n e ECG's over p u b l i c phone
l i n e s t o a remote i n t r e p r e t a t i o n c e n t e r and 5) improved and
f u n c t i o n a l i t y of ambulatory ECG-monitors and r e c o r d e r s
Ill.
M3 = X(i.n)-x(i)
n
where (n = 1,2, ...)
The e x i s t i n g ECG d a t a compression methods can be I f t h e new sample X(i+n) i s g r e a t e r then bx, then
c l a s s i f i e d i n t o t h r e e c a t a g o r i e s : 1 ) d i r e c t d a t a handling hX
i s made equal t o t h e value of t h e sample. I f t h e
methods, 2) transformation methods and 3) parameter e x t - sample i s less then b i n , t h e n b i n i s made equal t o t h e
r a c t i o n methods 111, 12 I. I n t h i s paper, some of d i r e c t v a l u e of t h e sample. These p r o c e s s e s a r e repeated while
d a t a handling methods such a s AZTEC, TP, SAPA-2,ADAPTfVE -
t h e c o n d i t i o n s Ml>M3>M2 and Lx G i n < T ( i ) a r e v a l i d .
a r e compared with t h e proposed new approach. Whenever one of t h e s e c o n d i t i o n s a r e v i o l a t e d , t h e
inmediately previous d a t a p o i n t is s t o r e d a s ( b a n , t )
The amplitude zone time epoch coding (AZTEC) a l - where ,X, i s t h e amplitude of t h e l i n e c a l c u l a t e d a s
gorithm, developed f o r preprocessing real-time E C G ' s f o r t h e average value of samples X(i) i n t h e c o r r e n t region,
rhythm a n a l y s i s , became a popular d a t a r e d u c t i o n algo- and t is t h e l e n g t h of t h e l i n e r e p r e s e n t i n g t h e number
rithm f o r ECG monitors and d a t a b e s e s with an achieved of samples i n t h e region.
compression r a t i o of 1 O : l (500 Hz samped ECG with 12 b i t
1,
r e s o l u t i o n ) 11 121. However, t h e r e c o n s t r u c t e d s i g n a l The adaptive approximation e r r o r E ( i ) and t h r e s -
demonstrates d i s c o n t i n u i t i e s and d i s t o r s i o n , which occurs hold T ( i ) depend on t h e n a t u r e of t h e s i g n a l . The v a l u e s
i n t h e r e c o n s t r u c t i o n of t h e P and T waves due t h e i r of E ( i ) and T ( i ) f o r t h e low information regions ( t h e
slow varying s l o p e s . base l i n e ) w i l l be g r e a t e r then t h e v a l u e s f o r high in-
formation r e g i o n s (Pand T waves, ST segments, QRS complex
Adaptive compression a l g o r i t h m i s a modified o r tachycardia s i g n a l s ) .
AZTEC algorithm, whereby the e r r o r t h r e s h o l d i s made
a d a p t i v e t o the ECG s i g n a l v a r i a t i o n s 12 I.
This technique I n o r d e r t o r e c a l c u l a t e the a d a p t i v e E ( i ) and
has r e s u l t e d i n a s l i g h t improvement ( b e t t e r compromise T ( i ) values, t h e algorithm c a l c u l a t e s permeanently t h e
between compression r a t i o of 1 O : l and r e c o n s t r u c t e d s i g - following s t a t i s t i c a l parameters:
nal fidelity.)

InternationalBiomedical Engineering Days, 1992 / 92 $3.00 0 1992 IEEE


0-7803-0743-7 61
I n o r d e r t o obcain a good compressibn w i t h o u t
l o o s i n g t h e o r i g i n a l form of i n p u t d a t a h i g h CR and
low PRD a r e d e s i r e d . In o r d e r t o compare t h i s new hyb-
r i d Technique w i t h o t h e r compression a l g o r i t h m s such
a z Turning P o i n t (TP), AZTEC, SAPA and ADAPTIVE t e c h -
n i q u e s , we a l s o a p p l i e d t h e a l g o r i t h m s t o the same ECG
s i g n a l a n d gave t h e CR and PRD r e s u l t s of t h e a l g o r i t h m s
f o r d i f f e r e n t h e a r t r a t e s , i n T a b l e s 1 and 2 -

T a b l e 1- CR and PRD r e s u l t s of t h e compression


t e c h n i q u e s f o r 60 bpm ECG d a t a ( p , i T l i n g
frequency =500 Hz, r e s o l u t i o n 12 b i t s )

Technique ! CR ' PRD

TP 2.0 / 4.8

AZTEC 10.0 ! 8.7

SAPA-1 9.8 ' 8.2

SAPA-2 9.3 7.6


1 :,,L;)LivL, ap;>rottinition e r r o r and t h r e s h o l d ' d e i i n e d 2s
ADAPTIVE 10 .o 8.8
:( i) :( i-l)-a- (Cf( i ) -Cf
T( i ) --.C(i-l)-a:
( i-lk ( i-1)

(C, ( i ) -Cf( i-IYT( i-1)


(5)
(6)
HYBRID i 10.5 1 7.0

wllerc t h e c o n s t a n t s a , and a, can be determined empri-


C d l ly. Table 2 - CR and PRD r e s u l t s of the compression
techniques f o r 120 bpm ECG d a t a .
-
Decompress ion Algorithm

Compressed d a t a i s a sequence o i p a i r s o i d a t a
Technique 1 CR PRD
(.b,, t) . k c ~ u s c of s t e p l i n e q r r a n t i z a t i o n , t h e I
2 . 1 ~t o~ t e d i s p l a y e d won't be a c c e p t a b l e t a r p h y s i c i a n s . TP I 2.0 4.8
i11~~r,~to3 r c ,curve-smootiting a l s o r i t h r n t o p r o c e s s t h e
AZTEC 5.2 7.5
t - o r ; > r e s s z dd a t a is a p p l i e d , which uses t h e i o l l o w i n p
,.ormu l a : SAPA-1 5.0 7.5
SAPA-2 5.8 7.6
:( i) &(-?S( i - j ) - j x ( i - 2 ) - 5 ~ (~ - L ) - I ~ s ( i )
30
ADAPTIVE 4 .9 8.0
-6S( i - l ) - j S ( i - ? ) - ? S ( i - j ) ) (7)

./U :.c : i i ) is nev d a t a p o i n t , while X ( i - n )


'1 31-c the
HYURID 1 7.2 I 7.6

: i:,i.ial data points.


fi
R-__
esults I' ORIGINAL. MTA

: ; ! c d c s c r i b e d mc'tiicd w i t h v a r i a b l e -ipproximat ion


.rror and e r r o r and T h r e s h o l d v a l u e s has been a p p l i e d
t o t h e compression of d i f f e r e n t s i n g l e l e a d ECG s i g n a l s
sampled a t 500 Hz w i t h a 12-bit A/D c o n v e r t e r . The app-
roximation and t h r e s h o l d v a l u e s a r e l i m i t e d t o h e b e t -
ween 1.5-30 and 2-13 and t h e c o n s t a n t s a , and a 2 have
been determined e m p i c a l l y between 0,7-1,l and 0,05-0,2.
Other a l g o r i t h m s have been a l s o a p p l i e d t o t h e same ECG
s i g n a l s i n o r d e r t o compare on t h e same b a s e s w i t h t h e
proposed a l g o r ithm.

I n a d d i t i o n to v i s u a l i n s p e c t , t h e performance
index P R D ( t h e p e r c c n t root-mean square d i f f e r e n c e ) i s
computed a s f o l l o w s ;

I
ADAppNE CR: 10.0
Pw: 8.8
1
4
wticrt. X ( i ) and XreC(i) a r e samples of t h e o r i g i n a l and
r e c o n s t r u c t e d d a t a sequences. The compression r a t i o
(CR) i s given by;

CR Number of samples of o r i g i n a l d a t a
Number of samples of compressed d a t a
(9)
)L '-

F i g u r e 1. O r l g i n a l and r e c o n s t r u c t e d ECG s i g n a l s

\
62 I
To v i s u a l i n s p e c t , the o r i g i n a l and r e c o n s t r u c t e d
ECG s i g n a l s a r e given i n Figure 1. As shown i n Figure 1
t h e new hybrid technique g i v e s the h i g h e s t CR and t h e
lowest PRD compared with o t h e r technique r e s u l t s . By
changing t h e t h r e s h o l d parameters of t h e p r e s e n t e d hyb-
r i d compression algorithm, d i f f e r e n t CR and PRD r e s u l t s
of t h e a l g o r i t h m a p p l i e d t o same ECG d a t a a r e obtained
and the r e s u l t s a r e shown i n Figure 2.

ORIGINAL wm

Figure 2. A p p l i c a t i o n r e s u l t s of p r e s e n t e d hybrid
compression a l g o r i t h m f o r d i f f e r e n t t h r e s h o l d
parameters.

bnc 1 7 s ion

An a d a p t i v e real-time algorithm f o r on-line com-


p r e s s i o n and transmission of ECG d a t a i s presented. The
algorithm combined the b e s t f e a t u r e s of two o t h e r tech-
niques c a l l e d SAPA-2 and Adaptive compression technique.
On a database of 650 electrocardiograms sampled a t 500
Hz, 1 2 b i t p r e c i s i o n , we obtained compression r a t i o s
between 8-13 and PRD's below 10, and without l o s i n g c l i -
n i c a l important content.

References

I11 S.M.S. J a l a l e d d i n e , C.G.Hutchens e t . a l . , "ECG data


compression techniques -A u n i f i e d approach", IEEE
T r . on BME, Vo1.37, pp. 329-340, April 1990.

121 B . Furht and A.Perez, "An Adaptive Real-time ECG


cornpre s s ion a l g o r i t h m with v a r i a b l e threshold" ,
IEEE T r . on BME, Vo1.35, pp.489-493, June 1988.

I31 M.Ishijima, S.B. Shin e t . a l . , "Scan -Along poly-


gonal approximation f o r d a t a compression on
electrocardiograms", IEEE T r . on BME, V01.30, pp.
723-728, nov. 1983.

Acknowledzement

This work i s supported by Turkish S t a t e Plan-


nig Organization.

63