AVO Measurement PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 30

Chapter-5: MEASUREMENT OF ELECTRICAL

CURRENT VOLTAGE AND RESISTANCE

5.1 INTRODUCTION
5.1.1 Principles of Current and Voltage Measurements
A device called the ammeter measures the current in
RT
A an electric circuit. It is connected in series with the
+
IL circuit element in which the current is to be determined.
V
VT RL The voltage is measured by the voltmeter. It is
- VL
connected in parallel with the circuit element to

Figure 5.1. Connections for an determine the voltage across. Eventually, the ammeter
ammeter and a voltmeter. requires breaking the current loop to place it into the
circuit. The voltmeter connection is rather easy since it is connected without disturbing the
circuit layout. Therefore, most electrical measurements prefer determination of the voltage
rather than the current due the ease of measurement. Connections of ammeters and voltmeters
are illustrated in figure 5.1.

Resistance is defined by the Ohms law as the ratio of voltage and current in a circuit
element. The device that measures the resistance is called the ohmmeter. It applies a voltage
from a constant (DC) voltage source (usually from an antennal battery) and measures the
current passing through using an ammeter.

5.1.2 Instrument Loading


Ideal ammeter has zero internal resistance and no voltage drop across it. Ideal voltmeter has
infinite internal (meter) resistance and draws no current from the circuit. The practical
ammeter can be symbolized by an ideal ammeter with an added series resistance that
represents the meter
RM IM I=0 resistance. Similarly, the
Ideal
A +
IM practical voltmeter can be
+ - +
-
V MC 0V VM RM V denoted by an ideal
VM voltmeter in parallel with
+ RM - -
the meter resistance. These
Practical ammeter Practical voltmeter two models are illustrated
Figure 5.2. Models of practical ammeters and voltmeters. in figure 5.2. Eventually,
Measurement of Voltage and Current / 83

the practical ammeter has a voltage drop across and the practical voltmeter has a current
drawn from the circuit.

All measuring instruments draw energy from the source of measurement. This is called
the loading effect of the instrument. Hence, all measurements include errors due to
instrument loading. If the energy extracted by the instrument is negligibly small compared to
the energy that exists in the source, then the measurement is assumed to be close to perfect,
and the loading error is ignored.

5.1.2.1 Loading Errors in Ammeters

RM Any electrical circuit can be modeled by a voltage


RT
A source VT and a series resistance RT as illustrated in
+ figure 5.3. The circuit is completed when the load
IL
VT RL
- resistance RL is connected across the output
terminals and a load current IL flows through the
Figure 5.3. Ammeter connection load. An ammeter can be placed in series with the
and its loading effect.
load to measure this current. The current in the
circuit can be calculated as

VT
IL = ................................................................................................... (5.1)
RT + RL + RM

In the ideal condition, RM = 0 and the true value of the current is

VT
I LT = ........................................................................................................... (5.2)
RT + RL

The error is the difference between the measured value and the true value, and generally
expressed as the percentile error which is:

measured value true value


% loading error = x100 ............................................... (5.3)
true value

Hence, the loading error due to the ammeter can be found as:

VT VT

R + RL + RM RT + RL 100 RM
% loading error for ammeter = T x100 = ..... (5.4)
VT RT + RL + RM
RT + RL

Loading error can be ignored if RM<<(RT+RL) which is satisfied in most applications.


Introduction / 84

5.1.2.2 Loading Errors in Voltmeters


In voltage measurement, the meter is connected in
RT RLeff
parallel with load resistor as shown in figure 5.4. The

+ true value of the voltage across the resistor is


VT RL RM V (without the meter)
-
VT RL
VLT = ................................................ (5.5)
RT + RL
Figure 5.4. Voltmeter connection
and its loading effect. As the meter is connected, RM becomes in
parallel with RL and effective load resistance
becomes
RL RM
RLeff = ........................................................................................................ (5.6)
RL + RM
RLeff RL if RM>>RL. The voltage measured by the meter is
RL RM
VT
RL + RM
VL = VLind = ....................................................................................... (5.7)
RL RM
RT +
RL + RM
VLind VLT
% loading error = x100 ........................................................................... (5.8)
VLT

5.2 PERMANENT MAGNET MOVING COIL (PMMC) TYPE DEVICES


5.2.1 Principle of Operation

50 Scale

0 100

Backing
Pointer
Mirror Observers
The parallax error

Figure 5.5. An analog meter display

Many measuring instruments make use of analog meters in determining the value of current,
voltage or resistance. An analog meter indicates the quantity to be measured by a pointer and
scale as shown in figure 5.5. The user interprets the reading from the scale. The screen is
Measurement of Voltage and Current / 85

calibrated in a curvilinear fashion it has a mirror-backed scale to


Right-hand rule I
identify the position of the pointer. The reading must be done
under reasonable lighting conditions and just above the pointer. B
Thumb
Otherwise, there will be parallax errors in the reading as shown
in figure 5.5. The user can interpret the reading by within Fingers
small (minor) scale division, under the best measurement Figure 5.6. The right-
hand rule.
conditions.

The permanent-magnet moving-coil (PMMC) is the most popular type of analog meters.
It responds to a direct current (DC) applied to its coil, and moves the pointer against a

X X

Current into plane Applied field Resultant field Force


Figure 5.7. A current carrying conductor in an external magnetic field.

calibrated scale by an amount proportional to the current. Basic concepts related to the
principle of operation of PMMC devices and their utilization in measuring instruments are
discussed in the sections below.

5.2.1.1 Magnetic Field Established by a Current Carrying Conductor


A circular magnetic field is established around a straight current carrying conductor as
illustrated in figure 5.6. The direction of the field is identified according to the right-hand
rule. If we place the thumb in the direction of the current, then the fingers indicate the
direction of the magnetic field.

If the current carrying conductor is placed into a uniform magnetic field as shown in
figure 5.7, the field lines interact and exert a force perpendicular to the directions of both the
magnetic field and the current.
Index Field
The relationship is expressed by Flemings (B)
finger
Current (I)
left-hand rule as illustrated in figure 5.8. As the
three fingers are kept perpendicular to each other, Middle finger
Thumb Force
index finger is in the direction of the magnetic (F)
Figure 5.8. Flemings left-hand rule
PMMC Type Devices / 86

field (B), middle finger is in the direction of the current (I), and then the thumb indicates

D
Force
Magnetic
Force
A field

x I
Magnetic C
I Force
field
Force Force
B
Figure 5.9. Forces exerted on the current carrying coil.

direction of the force (F).

5.2.1.2 A Current-Bearing Coil in an External Magnetic Field


When a current-bearing coil is placed in a magnetic field, the forces exerted on the coil
produce a torque as illustrated in figure 5.9. If the current-bearing loop is free to move, it
rotates until the maximum number of lines of magnetic flux pass through the loop. This is the
principle of electric motors.

The force F (Newton) acting on a conductor of length L (meter), current I (ampere) and
external magnetic field strength B (Weber/m2 or Tesla) (Bsin if 90) is:

F = BIL ...................................................................................................................... (5.9)

The forces produced along the vertical portions of the current loop (coil) (sections A-B and C-
D) are equal in magnitude but opposite in direction. Therefore, they produce a torque
(Newton-meter) that causes rotation of the coil if it is free to do so. By definition

Torque = force x distance

Hence, the electromagnetic torque effective on the coil is

TEM = FxW = BILW = BIA .................................................................................... (5.10)

Where W is the width of the loop (B-C). LW = A, where A is the cross sectional area of the
loop. If the coil has N turns, then the total electromagnetic torque produced is

TEM = NBIA, Newton-meter .................................................................................. (5.11)

Example 5.1
Find the total torque on the coil and power dissipated by the coil if I = 1 mA, A = 1.75 cm2, B
= 0.2 Tesla (2,000 Gauss), N= 48 turns and coil resistance is 88 .
Measurement of Voltage and Current / 87

TEM = 1.68x10-6 N-m, power dissipation = I2R = 88 W

5.2.2 Moving Coil in Measuring Instruments


5.2.2.1 Balancing the Electromagnetic Torque by a Spring Torque
The coil is suspended in a uniform
Control
0 Spiral spring magnetic field and rotates due to the
spring
torque electromagnetic torque TEM. This torque
Electro-
is opposed by spiral control springs
magnetic
torque (figure 5.10) mounted on each end of the
coil. The torque put forth on the control
Figure 5.10. Compensating electromagnetic
torque by the torque of control springs. spring is
TSP = k ......................... (5.12)
where is the angle of rotation (degrees) and k is spring constant (N-m/degree). At
equilibrium (at balance) TEM = TSP yielding
NBIA = k .......................................................................................................... (5.13)
Equation 5.13 can be rearranged for :
NAB
= I = SI .............................................................................................. (5.14)
k
where S is the sensitivity
NAB deg ree
S= =
I k Amp
.............................................................................. (5.15)

The sensitivity S is constant for a specific equipment provided that the external magnetic field

Linear S Constant
Input Output
Moving Coil
instrument
I I I
Uniform scale Uniform scale
Figure 5.11. Model of a moving coil instrument.

strength B is constant. In this respect, the moving coil instrument can be considered as a
transducer that converts the electrical current to angular displacement. The linear relation
between and I indicate that we have a linear (uniform) scale as shown in figure 5.11.

Example 5.2
A moving coil has following parameters: Area A= 2 cm2, N=90 turns, B= 0.2 Tesla, coil
resistance = 50 , current I= 1 mA. Calculate:
PMMC Type Devices / 88

a. Power dissipated by the coil;


P = I2xRm = 50 W.
b. The electromagnetic torque established;
TEM=NBAI = 90x0.2x2x10-4x10-3 = 3.6x10-6 N-m
c. Assume that the electromagnetic torque of the coil is compensated by a spring torque
and the spring constant k = 3.6x10-8 N-m/degrees. Find the angle of deflection of the coil
at equilibrium. Ans.: = TEM / k = 100

5.2.2.2 The DArsonval Meter Movement


A PMMC meter that consists of a moving coil suspended between the poles of a horseshoe
type permanent magnet is called the DArsonval meter as shown in figure 5.12. Shoe poles
are curved to have a uniform magnetic field through the coil. The coil is suspended between
to pivots and can rotate easily as illustrated in figure 5.13. The permanent magnet and the iron
core inside the coil are fixed. Coil axes and the pointer are the moving parts.

Point contact

Axle

Jeweled pivot to
minimize friction
Figure 5.13. The
point contact
Figure 5.12. DArsonval movement.

The principle of operation is similar to the general moving coil instrument explained
above. There are mechanical stops at both ends to limit the movement of the pointer beyond
the scale. The amount of the DC current that causes maximum allowable deflection on the
screen is called the full-scale deflection current IFSD and it is specified for all meters by their
manufacturers.

The moving coil instrument provides a unidirectional movement of the pointer as the coil
moves against the control springs. It can be used to display any electrical variable that can be
converted to a DC current within the range of IFSD.
Measurement of Voltage and Current / 89

Primary stage Intermediate stage Final stage


Coil terminal
TEM
Measurement
One turn
medium Angular
Current carrying
N turn coil Spring displacement
conductor +
Signal Signal
magnetic field
manipulation conversion Pointer and scale
Electric Primary sensing
element element Signal
current element
TEM presentation
Total torque element
Permanent
Rotatable coil with
magnet Scale
pivoted axis
External power reading
Signal conversion
source Angular
element
velocity Observer

Figure 5.14. Functional block diagram of a moving coil type instrument.

5.2.2.3 Block Diagram of MC Instrument


Block diagram of a moving coil instrument viewed as a general measurement system is shown
in figure 5.14. All functional elements are indicated in the figure.

5.2.2.4 The Galvanometer

0
IFSD +

0 -

Basic Moving Coil Galvanometer type


Figure 5.15. Moving coil and galvanometer type displays.

The galvanometer is a moving coil instrument in which position of the pointer can be biased
so that it stays in the middle of the scale to indicate zero current as shown in figure 5.15. It
can deflect in both directions to show the negative and positive values. It is commonly used in
bridge measurements where zeroing (balancing null) of the display is important for a very
accurate measurement of the variable. It is also used in mechanical recorders in which a pen
assembly is attached to the tip of the pointer and it marks on the paper passing underneath.

Neither the standard moving coil instrument nor the galvanometer can be used for AC
measurement directly since the AC current produces positive deflection with the positive
PMMC Type Devices / 90

alternate and negative deflection with the negative alternate. Thus, a stable position on the
scale cant be obtained to indicate the magnitude of the current.

5.3 MC BASED MEASURING INSTRUMENTS


5.3.1 MC in Analog Electrical Measuring Instruments
The standard MC instrument indicates positive DC currents (IMC)
IFSD
RMC as deflection on the scale. The moving coil is usually made up of
a very thin wire. The current through the moving coil IMC is
+
V MC -
limited by the full-scale deflection current IFSD. IFSD is in the order

Figure 5.16. Model of 0.1to 10 mA and coil resistance RMC 10 to 1000 . The
maximum deflection angle is about 100. A voltage drop across
the coil is VMC = IMCRMC. The moving coil can be represented by IFSD and RMC as shown in
figure 5.16.

5.3.2 Ammeters
5.3.2.1 Basic DC Ammeter (Ampermeter)
The current capacity of the meter can be expended by adding a resistor
RMC in parallel with the meter coil as shown in figure 5.17. The input
IFSD + current is shared between the coil resistance RMC and the parallel
V MC -
resistance that is called the shunt RSH. As the maximum input current
RSH
IT flows in, the coil takes IFSD and remaining (IT - IFSD) is taken by the
IT (IT - I FSD ) shunt resistor. Voltage developed across the meter is
RM
VMC = I FSD RMC = (IT I FSD )RSH ................................... (5.16)
Figure 5.17. Basic
DC Ammeter. The meter resistance RM seen between the input terminals is

VMC
RM = = RMC // RSH ...................................................................................... (5.17)
IT

Example 5.3
Calculate the multiplying power of a shunt of 200 resistance used with a galvanometer of
1000 resistance. Determine the value of shunt resistance to give a multiplying factor of 50.
Ifsdx1000 = (IT Ifsd)x200 yielding IT = 6xIfsd.
For IT=50xIfsd, 1000xIfsd=(50-1)xIfsdxRsh yielding Rsh =1000/49 = 20.41
Measurement of Voltage and Current / 91

5.3.2.2 The Multi-Range Ammeter


Switch
poles The parallel resistance (shunt) can be changed to suit
different full-scale current requirements as indicated in
Rotary
switch the previous example. Using a set of resistors and
arm selecting them one by one can accommodate the
Figure 5.18. Make-before-
break type switch. function. The switch however must be of make-before-
break type (figure 5.18) that makes the contact with the
new position before it breaks the old connection. This eliminates the chance of forcing the full
input current through the moving coil during changing the position of the switch.

Example 5.4
Design a multi-range DC ammeter using the basic movement with an internal resistance RMC=
50 and full-scale deflection current IMC= IFSD= 1 mA. The ranges required 0-10 mA, 0-50
mA, 0-100 mA and 0-500 mA.
VMC = IMCxRMC = 50 mV
For range-1 (0-10 mA) RSH1= 50/9 =5.56

IT IFSD
RMC

RSH1 0 500 mA
0 500 mA 0
0
100 mA
RSH2 0 50 mA
Rotary 10 mA
0 100 mA
selector
RSH3 0 50 mA
switch
RSH4 0 10 mA Multi-range ammeter scale
Multi-range ammeter circuit
Figure 5.19. A multi-range ammeter circuit and scale for example 5.4

For range-2 (0-50 mA) RSH2= 50/49 =1.02


For range-3 (0-100 mA) RSH3= 50/99 =0.505
For range-4 (0-500 mA) RSH4= 50/499 =0.1
MC Based Measuring Instruments / 92

5.3.3 Voltmeters
RS IFSD RMC
5.3.3.1 Basic DC Voltmeter
+ - +
- The moving coil can be used as a voltmeter by adding a
VS V MC
series resistance RS as illustrated in figure5.20. The input
VM
+ RM - voltage is divided between the coil resistance RMC and RS.
Current passing through both resistors is IMC which is
Figure 5.20. Basic DC limited by the full-scale deflection current IFSD of the coil.
voltmeter.
The full-scale input voltage

VM = IFSD(RS+RMC) ............................................................................................ (5.18)

The input impedance seen is

RM = RS + RMC ................................................................................................... (5.19)

However, with RS>>RMC, RM is approximately equal to RS and VM IFSDRS.

Example 5.5
The coil of a moving coil voltmeter is 4 cm long and 3 cm wide and has 100 turns on it. The
control spring exerts a torque of 2.4x10-4 N-m when the deflection is 100 divisions on the full
scale. If the flux density of the magnetic filed in the air-gap is 0.1 Wb/m2, estimate the
resistance that must be put in series with the coil to give one volt per division. The resistance
of the voltmeter coil may be neglected.
TEM = TSP 2.4x10-4 = 100x0.1x12x10-4xIFSD IFSD =20 mA. Therefore, current per
division is 0.2 mA. Assuming that RMC is negligibly small compared to RS : RS = 5 k
Measurement of Voltage and Current / 93

5.3.4 Multi-Range Voltmeter


The series resistance can be changed to suit different full-scale voltage requirements as shown
in figure 5.21. Resistors are organized either in parallel fashion (conventional connection) as

Voltage to be measured RS2 RS1 RMC


RS4
IFSD 0 1000 V 1
RMC
RS3
0 100 V RS3 2
VM
RS2
0 50 V 3
Rotary 4
selector RS1 0 10 V RS4 0 1000 V
switch
Multi-range voltmeter circuit Multi-range voltmeter circuit
Parallel connection Series connection
Figure 5.21. Parallel and series resistance connections for a multi-range voltmeter.

in the case of ammeter and selecting them one by one or all connected in series like a voltage
divider (modified connection). The switch however must be of break-before-make type
(figure 5.22) that breaks the contact with the old position before it makes it with the new
position. This eliminates the chance of forcing a current larger than the full-scale current
through the moving coil during changing the position of the switch.

The resistors are also called the multiplier resistors.


Switch Resistance seen by the input terminals of the device
poles
RM = VM/IFSD ...................................... (5.20)
Rotary
switch and written on the face of the scale as /V. The
arm
contribution of the coil resistance RMC can be ignored if
Figure 22. A break-before-
make type switch it is too small compared to RM. The following example
illustrates the selection of multiplier resistors.

Example 5.6
A multi-range DC voltmeter is designed using a moving coil with full-scale deflection current
10 mA and coil resistance 50 . Ranges available: 0 10V, 0 50V, 0 100V, 0 - 1000V.
Determine the multiplier resistors and input resistance of the meter using:
a. Conventional connection
b. Modified connection
In conventional connection, resistors are selected one-by-one to satisfy
MC Based Measuring Instruments / 94

Zero Internal MC meter 10


adjust battery
RMC 100 2
0

Basic series ohmmeter circuit Series ohmmeter scale

Figure 5.23. Circuit and scale of a basic ohmmeter.

VM = IFSD (RMC + RS) = VMC + IFSDRS where VM is the full-scale voltage of the selected range.
VMC = (10 mA)(50) = 0.5V. Hence, RS = (VM 0.5)/10 k. Meter resistance seen between
the input terminals is RM = RMC + RS
Range 1 (0 10V): RS1 = 9.5/10 = 0.95 k = 950 ; RM1 = 950 + 50 = 1000
Range 2 (0 50V): RS2 = 49.5/10 =4.95 k; RM2 = 4.95 k +0.05 k = 5 k
Range 3 (0 100V): RS3 = 99.5/10 =9.95 k; RM3 = 9.95 k +0.05 k = 10 k
Range 4 (0 1000V): RS4 = 999.5/10 =99.95 k; RM4 = 99.95 k +0.05 k = 100 k
For the alternative modified arrangement, the resistor for the lowest range is
determined and others calculated as added to the total of the previous value. The total
resistance seen from the input in all ranges will be the same as those in the previous case.
Resistors between stages can be computed as RSn = RMn RM(n-1)
Range 1 (0 10V): RM1 = 1000 ; RS1 = 1000 - 50 = 950
Range 2 (0 50V): RM2 = 5 k; RS2 = 5 k - 1 k = 4 k;
Range 3 (0 100V): RM3 = 10 k; RS3 == 10 k - 5 k = 5 k;
Range 4 (0 1000V): RM4 = 100 k; RS4 = 100 k - 10 k = 90 k;

5.3.5 Ohm and VOM Meters


5.3.5.1 Analog Ohmmeter
Analog ohmmeter can be designed simply by adding a battery and a variable resistor in series
with the moving coil instrument as shown in figure 5.23. The unknown resistance is
connected to the terminals of the device to complete the electrical circuit. The output
terminals are shorted together with the leads (wires) used in connecting the external resistor.
The variable resistance is adjusted until the full-scale deflection current passes through the
coil. This is marked as the 0 resistance. When the leads are separated from each other, no
current flows indicating an open-circuit, which means infinite - resistance. Hence, the
scale is non-linear with resistance and increasing values are marked on the right-hand side
Measurement of Voltage and Current / 95

(opposite to ammeter). Multi-range ohmmeters can be obtained by combining the circuits of a


series ohmmeter and a multi-range ammeter.

5.3.5.2 VOM Meter


The functions of ammeter, voltmeter and ohmmeter can be combined in a multipurpose meter
called a VOM (volt-ohm-milliampere) meter, or shortly the VOM. It has several multiple
scales, usually color-coded in some way to make it easier to identify and read. Generally, it
has a single multipurpose switch to select the function and the range.

Example 5.7
A moving coil has 100 turns, 5 cm2 coil area, and air-gap magnetic flux density of 0.1 Tesla
(Wb/m2). TSP = 5x10-6 N-m at the full-scale deflection of 90. The potential difference across
the coil terminals at the full-scale deflection is 100 mV. Design a multi-range DC ammeter
with ranges 0-50 mA, 0-1 A and multi-range DC voltmeter with ranges 0-10 V and 0-200 V.
IFSD=TSP/NBA = 1 mA, therefore RMC= VMC / IFSD =100
For ammeter ranges: RSH1= 100 mV/ (50-1) mA = 2.04 and RSH2 = 100/999 = 0.1
For voltmeter ranges: RS1 = (10-0.1)V/1mA = 9.9 k and RS2 = 199.9 k
The Digital Voltmeter (DVM) / 96

5.4 THE DIGITAL VOLTMETER (DVM)


5.4.1 Utilization and Advantages
It is a device used for measuring the magnitude of DC voltages. AC voltages can be measured
after rectification and conversion to DC forms. DC/AC currents can be measured by passing
them through a known resistance (internally or externally connected) and determining the
voltage developed across the resistance (V=IxR). Similarly, the digital voltmeter shows the
numerical value for the signal magnitude.
The result of the measurement is displayed on a digital readout in numeric form as in the
case of the counters. Most DVMs use the principle of time-period measurement. Hence, the
voltage is converted into a time interval tg first. No frequency division is involved. Input
range selection automatically changes the position of the decimal point on the display. The
unit of measure is also highlighted in most devices to simplify the reading and annotation.

The DVM has several advantages over the analog type voltmeters as:
Input range: from 1.000 000 V to 1,000.000 V with automatic range selection.
Absolute accuracy: as high as 0.005% of the reading.
Stability
Resolution: 1 part in 106 (1 V can be read in 1 V range).
Input impedance: RI 10 M ; CI 40 pF
Calibration: internal standard derived from a stabilized reference voltage source.
Output signals: measured voltage is available as a BCD (binary coded decimal) code and
can be sent to computers or printers.

5.4.2 Basic Operation and Functional Block Diagram


Several techniques are utilized to obtain the voltage to time conversion and the respective
DVMs are named accordingly as the ramp type, integrating type, continuous balance type,
and successive approximation type. The ramp type is the simplest one and it will be discussed
below.

Functional block diagram of a positive ramp type DVM is shown in figure 5.24. The
timing diagram is given in figure 5.25. An internally generated ramp voltage is applied to two
comparators. The first comparator compares the ramp voltage into the input signal and
produces a pulse output as the coincidence is achieved (as the ramp voltage becomes larger
than the input voltage). The second comparator compares the ramp to the ground voltage (0
volt) and produces an output pulse at the coincidence. The input voltage to the first
Measurement of Voltage and Current / 97

DC input
Input
voltage
Ranging Comp.
& - 1.275 V
Polarity
Readout (Display)
Attenuator
tg
Decade
Gate
AND counters
Ramp control
Generator
Ground tg
Comp. Time-base
Sample rate oscillator
Tb or Tc
oscillator
fb or fc

Figure 5.24. Functional block diagram of a single-ramp type digital voltmeter.

Vm
2nd coincidence
(+10 V)
stop
Vi
1st coincidence
start time

tg
T
Ground comparator
Input comparator
-Vm
(- 10 V) Count gate
tg
(time interval)
Tc=1/fc
Clock pulses

Sample interval
Figure 5.25. Timing diagram of the digital voltmeter.

comparator must be between Vm. The ranging and attenuation section scales the DC input
voltage so that it will be within the dynamic range. The decimal point in the output display
automatically positioned by the ranging circuits.

The outputs of the two comparators derive the gate control circuit that generates and
output pulse that starts with the first coincidence pulse and ends with the second. Thus, the
duration of the pulse tg can be computed from the triangles as
The Digital Voltmeter (DVM) / 98

Vi t g T
= t g = Vi .......................................................................................... (5.21)
Vm T Vm

Hence, the voltage to time conversion is done yielding tg to Vi with T and Vm constant.
Number of time intervals (clock pulses) counted during this interval become:

T fc
N = t g f c = Vi ........................................................................................ (5.22)
Vm

For the ramp voltage with a fixed slope and time base running at a fixed rate (fc), N is
directly proportional to Vi. T.fc/Vm that is set to a constant factor of 10.

The polarity of the voltage is indicated if it is -. With no indication, it is understood that


the polarity is +. The polarity circuit with the help of comparator pulses detects the polarity.
For positive slope ramp type voltmeter, the first coincidence of the ramp is with the ground
voltage if the input is positive. With a negative input voltage however, the first coincidence
will be with the input voltage.

The display stays for sometimes (around three seconds) and than it is refreshed by the
sample rate oscillator. A trigger pulse is applied to the ramp generator to initiate a new ramp.
Meanwhile a reset (initialize) pulse is applied to the decade counters to clear the previously
stored code.

The display indicates the polarity as well as the numbers in decimal and a decimal point.
The first digit contains the polarity sign and the number displayed can be only 1 or 0 for
most voltmeters. Therefore, this is called half digit. Hence, a three and a half digit display
can have up to 1999 and a four and a half digit one can go up to 19999.
Measurement of Voltage and Current / 99

5.5 AC VOLTMETERS
5.5.1 Measurement of AC Voltages
The voltmeter based on the permanent magnet moving coil (PMMC or DArsonval) can not
be directly used to measure the alternating voltages. The instruments that are used for
measuring AC voltages can be classified as:
1. Rectifier DArsonval meter
2. Iron Vane (Moving Iron) type meter
3. Electrodynamometer
4. Thermocouple meter
5. Electrostatic voltmeter
The rectifier type (Rectifier DArsonval) meter is the extension of the DC voltmeter.
This type and the thermocouple based true rms meter will be explained below.

5.5.1.1 Average and RMS Values


i(t)=Imsin t
The moving coil instrument reads the average of an AC
waveform. The average of the current waveform i(t) shown in
Time
figure 5.26 is:
Figure 5.26. Alternating
T
current (AC) waveform. 1
T 0
I AV = I m sin tdt = 0 .......................................... (5.23)

where T is the period and = 2/T = radial frequency (rad/sec). However, if this current is
applied to a resistor R, the instantaneous power on the resistor

p(t) = i2(t)R .............................................................................................................. (5.24)

The average power over the period T becomes:


T 2
R 2 I R
PAV = I m sin tdt = m ............................................................................... (5.25)
T0 2

Hence, the average power is equivalent to the power that would be generated by a DC current
called the effective current that is

1 T2 I
I eff = I RMS =
T 0
i (t )dt = m = 0.707 I m
2
........................................................... (5.26)

Due to squaring, averaging (mean) and square-rooting operations, this is called the RMS.
value of the current and IRMS is the true value of the current that we want to measure.
AC Voltmeters /100

If the voltage is applied to the resistor


Vm
vo(t) through a diode as shown in figure 5.27, the
vi(t)=Vm sint
VAV negative half cycle is chopped off since the
Time
Time diode can conduct current only in positive
direction. This is called the half-wave
Figure 5.27. AC to DC conversion. rectifier. The average value of the current in
the resistor becomes:
T
1 2 V
VAV = Vm sin tdt = m = 0.318Vm .................................................................... (5.27)
T0

5.5.1.2 The Full-Wave Rectifier


The half-wave rectifier is used in some voltmeters, but the mostly adapted one uses the full
wave rectifier shown in figure 5.28. Here, a bridge-type full-wave rectifier is shown. For the

D2
D1
Im
A Rm C
+ + + +

D4 D3

D
Ii + - + -
+ alternate
+ Input -
- alternate
Figure 5.28. Bridge type full-wave rectifier.

positive (+) half cycle the current follows the root ABDC. For the half cycle root CBDA is
used. The current through the meter resistor Rm is the absolute value of the input current as
shown in the inset. The voltage waveform on the meter resistance Rm has the same shape as
the current. The average value of the voltage becomes:
Measurement of Voltage and Current /101
T
2 2 2V
VAV = Vm sin tdt = m = 0.636Vm .............................................................. (5.28)
T0

VAV is the DC component of the voltage and it is the value read by the moving coil
instruments. Hence, the inherently measured value (IM) is the average value, while the
true value is the RMS value. The voltage reading will contain reading error (unless it is
corrected) as

Vindicated Vtrue VAverage VRMS


%error = ( ) 100% = ( ) 100% = 10% .............. (5.29)
Vtrue VRMS

and the indicated voltage will be 10% less then the true value.

5.5.2 Form Factor and Waveform Errors


5.5.2.1 For Sinusoidal Waveforms
The ratio of the true value to the measured value is called the form factor or safe factor (SF).
For sinusoidal signals the form factor is

SF = (VRMS/VAV).................................................................................................. (5.30)

In AC voltmeters, the reading is corrected by a scale factor = safe factor (SF) = 1.11. This can
be done either at the calculation of the series resistance or setting the divisions of the scale.
Eventually, the error is eliminated as:

Vindicated Vtrue 1.11 VAverage VRMS


%error = ( ) 100% = ( ) 100% = 0% ........ (5.31)
Vtrue VRMS

This is of course true for sinusoidal signals. For other waveforms, the error may be nonzero
indicating erroneous readings.

5.5.2.2 For Triangular Waveform


A triangular voltage waveform v(t) with amplitude Vm and
v(t)
Vm period T is shown in figure 5.29. The negative portion is
t
converted to positive after the full-wave rectification. Due to
-Vm T
the symmetry of the signal, interval from 0 to T/4 can be used
Figure 5.29. A triangular for integration in finding the average (DC) and RMS values.
waveform
In this interval, the signal can be expressed as

v(t) =( 4Vm/T)t ................................................................................................... (5.32)

Thus
AC Voltmeters /102

4 T 4 4Vm V
VAV =
T 0 T
dt = m = 0.5Vm
2
...................................................................... (5.33)

This is the inherently measured (IM) value. A meter corrected for sinusoidal waveforms will
indicate

Vind = 1.11x0.5Vm= 0.555 Vm ............................................................................. (5.34)

The RMS value can be computed as:

2
4 T 4 16Vm Vm
T 0 T
VRMS = 2
dt = = 0.577Vm .......................................................... (5.35)
3

Hence, the form factor for the triangular waveform is 1.155 and 1.11Vaverage VRMS .The
percentile measurement error:
Vindicated Vtrue 1.11 VAverage VRMS 0.555 0.577
%error = ( ) 100% = ( ) 100% = 100% = 3.81%
Vtrue VRMS 0.577

5.5.3 The Correction Factor


A correction factor (CF) is used to multiply the reading indicated by the meter to correct the
measured value. The correction factor must be determined for every specific waveform
individually as:

V
( SF ) waveform ( RMS ) waveform
VIM
CF = = .............................................................. (5.36)
( SF )sin usoidal V
( RMS )sin usoidal
VIM

The voltage indicated for the triangular waveform using a meter adjusted for a sinusoidal
waveform can be written as:

VRMS
Vind = SFx(VIM ) waveform = ( ) sin usoidal x(V AV ) waveform ......................................... (5.37)
V AV

Eventually,

(Vind )(CF ) = (SF ) wave (VIM ) wave = (VRMS ) wave = Vtrue ......................................... (5.38)

The error without the correction:


1 CF
%error = 100% ................................................................................... (5.39)
CF
Measurement of Voltage and Current /103

5.55
5 AC
readings
DC
0 10 11.1
readings

v(t)
v(t)=Vmsint
VIM

Time Time
AC Full-wave Unidirectional DArsonval meter VRMS
Rectifier (SF = 1.11)
Voltage Voltage

Figure 5.30. Illustration of an AC voltmeter corrected for sinusoidal signals.

0.577
For the triangular wave shown in the above example CF = 0.5 = 1.154 = 1.0396
0707 1.11
0.636
yielding the percentile error of 3.81%, same as the one found before.

Figure 5.30 shows a pictorial presentation of the scale calibrated for sinusoidal voltage
waveforms; model of the AC voltmeter based on the basic DArsonval meter with samples of
input and output waveforms.

Example 5.8 Vm(t)


A DArsonval 10 V
(moving coil)
movement based AC 120 k
10 k Vm t
voltmeter is calibrated 0
Vs =
1 3 6
to read correctly the
8V -5 V
RMS value of applied
sinusoidal voltages. Circuit for example 5.8. Waveform for example 5.8.
The meter resistance
is 10 k/V and it is used in 0 10 V range.

a. Find Vm measured by the meter and the percentile loading error.


AC Voltmeters /104

True value of the voltage Vtrue= 8x120/130 = 7.38 V; Rm= 100 k leading to RL=
100x120/220 = 54.5 k. Therefore Vm = 8x54.5/64.5 = 6.76 V. Percentile loading error =
-8.4%.
b. A different periodic waveform is applied and the waveform Vm(t) shown appears
across the meter.
i. Calculate VRMS for this waveform,
1 1 3 250
2
VRMS = [ 100t 2 dt + 25dt = ; VRMS = 5.27V,
3 0 1 9
ii. How much is the voltage indicated by the meter (Vindicated)?
1 1 3
V ( AV )
= [ 10tdt + 5dt = 5V Therefore, Vind = 1.11x5 = 5.55 V
3 0 1

iii. Find the waveform error in this measurement.


% waveform error = 100x(5.55 5.27)/5.27 = 5.3%.

5.5.4 The Thermocouple-Based True RMS Meter


Alternating electrical currents and

Vm voltages that can be represented by

VRMS pure sinusoidal waveforms can be


rectified and measured by a
DArsonval movement based-meter.
The corrected meter displays the rms
Time
value of the applied waveform.
Figure 5.31. A complex waveform. Waveforms that follow other well-
known geometric shapes can also be
used if the correction factors can be computed easily. The rms value for a complex waveform
similar to the one shown in figure 5.31 can not be determined accurately by this technique. It
can be measured most accurately by an rms-responding meter.

The power generated by a waveform as applied to a resistor varies with the square of the
rms value of the waveform as

Ii PAV = I RMS
2
xR = VRMS
2
R ......................................... (5.40)
+
Vi ET This power indicates the rate of the heat energy added into
- the resistor. Eventually, the case temperature of the resistor
varies with the power. Hence, the case temperature changes
Figure 5.32.
Measurement of Voltage and Current /105

Measuring
Indicating
Thermocouple
Meter
AC input Input AC + DC
Voltage Ranging Amplifier - Amplifier
- +
Balancing
Thermocouple
Feedback
Current

Figure 5.33. Block diagram of a true rms-responding meter.

proportionally with the square of the rms value of the applied voltage or current. A
thermocouple that is placed into the same thermal environment with the resistor as shown in
figure 5.32 produces a DC output voltage ET related to the temperature.

The thermocouple voltage is a nonlinear function of the rms value of input voltage.
Figure 5.33 illustrates a true-rms reading voltmeter that uses two thermocouples. An AC
amplifier amplifies the input signal coming form the ranging circuit. Two resistor-
thermocouple sets are identical. The first set is connected to the AC input voltage and is called
the measuring one. The second thermocouple forms a bridge with the first one. A DC
amplifier amplifies the output of the bridge. At balance, the voltages generated by both
thermocouples are identical. Hence, the resistor connected to the balancing thermocouple
produces the same heat as the measuring one indicating that the feedback current form the
amplifier is equivalent to the rms value of the AC input current.

Ratio of the peak value (Vm) to rms value of a waveform is called the crest factor. The
meter can successfully display the rms value of the waveform provided that the peak value of
the input voltage does not saturate the AC amplifier. A smaller fraction of the full-scale meter
deflection is used in measuring waveforms with high crest factor to minimize the risk of
saturation of the AC amplifier. The frequency of the waveform that can be handled depends
upon the bandwidth of the input ranging circuits and AC amplifier, and it can go up to a few
MHz.
Problems /106

5.6 PROBLEMS
1. A moving coil instrument has the following data: # of turns of the coil = 100, width of the
coil = 2 cm, length of the coil = 3 cm, flux density in the air gap = 0.1 Wb/m2 (Tesla).
Calculate the deflection torque when carrying a current of 10 mA. Also calculate the
deflection (angle) if the control spring constant is 20x10-7 N-m/degree.
2. Design a multi-range DC ammeter using the basic movement with an internal resistance
RMC= 50 and full-scale deflection current IMC= IFSD= 10 mA. The ranges required 0-0.1
A, 0-1 A, 0-10 A and 0-100 A.

3. A moving coil instrument gives full-scale deflection of 10 mA when the potential


difference across its terminals is 100 mV. Calculate:
a. The shunt resistance for a full scale corresponding to 100 mA;
b. The resistance for full scale reading with 1000 V;
c. The power dissipated by the coil and by the external resistance in each case.
4. A basic DArsonval meter movement with an internal resistance RMC= 100 , full scale
current IFSD= 1 mA, is to be converted into a multi-range DC voltmeter with ranges 0-10
V, 0-50 V, 0-250 V and 0-500 V. Find the values of multiplier resistors using the potential
divider arrangement.
5. A 150-V DC voltage source is coupled to a 50 k
load resistor through a 100 k source resistance.
20 k
Two voltmeters (A) and (B) are available for the 20 k
V
measurement. Voltmeter-A has a sensitivity 1000
/V, while voltmeter-B has a sensitivity 20000 /V.
10 V

Both meters have 0 50 V range.


Figure for problem 6.
a. Calculate reading of each voltmeter.
b. Calculate error in each reading expressed in a percentage of the true value.
6. A voltmeter with a resistance of 20 k/V is used to measure the voltage on the shown
circuit on a 0 - 10 V range. Find the percentage loading
error.
100 k V
7. A generator produces 100 volts DC and has an internal
resistance of 100 k as shown in the figure. The output 100 V
voltage is measured using several voltage indicating
devices. Calculate the output voltage and the percentage Figure for problem 7.
Measurement of Voltage and Current /107

loading error for each of the following cases:


a. An ideal voltmeter (Ri ) Vo = 100 V,
b. A digital voltmeter with Ri = 10 M;
c. An oscilloscope (Ri = 1 M);
d. A moving coil type analog voltmeter with 1 k/V in 0 100 volt range

8. A DArsonval movement gives full-scale deflection of


10 k
1 mA when a voltage of 50 mV is applied across its 1 k V
terminals. Calculate the resistance that should be added
in series with this movement to convert it into a 0 100 90 V

V voltmeter. The above 0 100 V voltmeter is used to


Figure for problem 8.
measure the voltage across the 10 k resistor in the
shown circuit. Determine the percentage loading error.
9. The voltage waveform shown has a magnitude 50 V and it is applied to an AC voltmeter
composed of a full-wave
rectifier and a moving coil V (t) V2(t) =
1 Full-wave DArsonval meter
(DArsonval) meter. It is Rectifier (SF = 1.11)
calibrated to measure voltages V1(t)
with sinusoidal waveforms
Model for problem 9.
correctly.
a. Find the average and RMS values of V1(t)
V1(t) 50 V
b. Sketch the waveform for V2(t) -2 3
c. Find the average and RMS values of -1 0 1 2 t

V2(t). Ans. The RMS value of V2(t) is -50 V


the same as that of V1(t) which is 28.87
volts. The average value can be V2(t) = V1(t)
calculated from the area of the triangle 50 V
easily as 50/2 = 25 volts.
d. Find the voltage indicated by the meter. -2 -1 0 1 2 3 t
Ans. 25x1.11= 27.75 volts -50 V
e. Calculate the error due to the waveform
Waveforms for problem 9.
and find the correction factor.
Problems /108

10. A generator with 500 internal resistance has a sawtooth v(t)


Vm
output voltage as shown. The RMS value of this output is
to be measured by a moving coil instrument whose
t
0 T 2T
internal resistance is 10 k. The instrument has a full
Signal for problem 10.
wave rectifier and is calibrated for sinusoidal waveforms.
Calculate the error due to the waveform and also the loading error.
11. A moving coil has 100 turns, 3 cm2 coil area, and air-gap magnetic flux density of 0.1
Tesla (Wb/m2). The control spring exerts a torque of 3x10-7 N-m at the full-scale
deflection of 100. The potential difference across the coil terminals at the full-scale
deflection is 5 mV. Using the above movement:
a. Find the full scale deflection current and coil resistance;
b. Design a DC ammeter with a range 0-50 mA;
c. Design a multi-range DC voltmeter with ranges 0-10 V and 0-200 V.
d. What would be the deflection angle for an input voltage of 7 V in 0-10 V range?
12. A moving coil has 80 turns, 4 cm2 coil area, and air-gap magnetic flux density of 0.1 Tesla
(Wb/m2). The control spring exerts a torque of 4x10-7 N-m at the full-scale deflection of
90. The potential difference across the coil terminals at the full-scale deflection is 10 mV.
Using the above movement:
a. Find the full scale deflection current and coil resistance;
b. Design a DC ammeter with a range 0-100 mA;
c. Design a multi-range DC voltmeter with ranges 0-100 V and 0-200 V.
d. What would be the deflection angle for an input voltage of 65 V in 0-100 V range?
13. A DArsonval (moving coil) movement based AC
voltmeter is calibrated to read correctly the RMS
5 k 20 k
Vm
value of applied sinusoidal voltages. The meter
resistance is 4000/V and it is used in 0 50 V Vs
range.
Vm(t)
a. Find Vs if it is sinusoidal and Vm = 36 V 100 V
(RMS)
b. The periodic waveform vm(t) shown is
t
applied to the meter. 0
1 3 6
i. Calculate VRMS for this waveform,
-50 V
Figures for problem 13.
Measurement of Voltage and Current /109

ii. How much is the voltage indicated by the meter (Vindicated)?

iii. Find the waveform error in this measurement.

14. An AC voltmeter calibrated for sinusoidal voltages is used to measure both the input (V1)
and output (V2) voltages. It has a scale with 100 divisions and measurement ranges: (0
50) mV; (0 100) mV; (0 500) mV; (0 1) V; (0 2) V; (0 5) V and (0 10) V
a. Determine the range that would yield the most accurate reading for V1, the value
indicated by the meter for V1 and percentage reading uncertainty (assume that the
reading uncertainty is 0.5 division).
b. Repeat (a) for V2.

15. An average reading full-wave rectifier moving coil


v(t)
AC voltmeter is calibrated to read correctly the 5V
RMS value of applied sinusoidal voltages. The t
0
periodic waveform v(t) shown is applied to the 1 2 3
meter. Calculate VRMS for this waveform, Vindicated -5 V
and the waveform error in it. Waveform for problem 15.
16. Draw the circuit diagram and explain the operation
of the full-wave rectifier bridge circuit used to convert DArsonval movement into an AC
voltmeter.
a. What is the VRMS for a zero averaged square waveform of peak to peak value = 10
V? What is the value indicated for it by the AC voltmeter calibrated to read
applied sinusoidal voltages correctly? What is the percentage waveform error in
that value?

V(t) Vr(t)
5V 5V
Full-wave
t t
Rectifier
V(t) Vr(t)
-5 V

Figure for problem 16.

b. Repeat (a) if the square wave accepts amplitude values between 0 and 10 volts.

17. Explain the operation of one circuit through which the DArsonval movement can be
used as a meter for measuring periodic signals. What is the scale factor for calibrating
such a meter?
Problems /110

18. What is the VRMS for the waveform shown?


v(t)
What is the value indicated by an AC 10 V
voltmeter calibrated for sinusoidal 5 7 t
0
waveforms? What is the percentage waveform 1 3
error in that value? -10 V
Waveform for problem 18.
19. A digital voltmeter uses 3 digit display (it
can display up to 1999).
a. It is used to measure a voltage across a standard cell whose value is 1.234 volt 5
times and following readings are obtained: 1.2202, 1.2115, 1.2456, 1.2218.
Determine the accuracy, the precision and the bias of the voltmeter.
b. The digital voltmeter is of positive ramp type. The clock (time-base) runs at 1
MHz. The slope of the ramp is 1000 volt/s. The voltage applied for the
measurement is 1.5 volt DC. Draw the block diagram of the digital voltmeter and
sketch the diagram for voltage to time conversion. Then, determine the duration of
the gate signal produced as a result of the voltage-to-time conversion and number
of clock pulses applied to the counter.

20. Draw a simplified block diagram of ramp-type digital voltmeter and label each block
clearly. Show sample signals at various stages. State the advantages of voltage
measurement using a digital voltmeter.
21. For a ramp-type digital voltmeter:
a. Explain the function of the time-base oscillator.
b. Explain the voltage to time conversion.
c. How the polarity of the voltage is identified?
d. Assume that the number displayed is -10.025 V. How much is the uncertainty in
the voltage reading?
e. What is the significance of the sample rate?
f. What are the factors affecting the accuracy of the measurement?
g. What are the similarities and differences between electronic counters and digital
voltmeters?
Measurement of Voltage and Current /111

Chapter-5: MEASUREMENT OF ELECTRICAL CURRENT VOLTAGE AND


RESISTANCE .......................................................................................................................... 82
5.1 INTRODUCTION .................................................................................................... 82
5.1.1 Principles of Current and Voltage Measurements............................................ 82
5.1.2 Instrument Loading .......................................................................................... 82
5.2 PERMANENT MAGNET MOVING COIL (PMMC) TYPE DEVICES ............... 84
5.2.1 Principle of Operation ...................................................................................... 84
5.2.2 Moving Coil in Measuring Instruments ........................................................... 87
5.3 MC BASED MEASURING INSTRUMENTS ........................................................ 90
5.3.1 MC in Analog Electrical Measuring Instruments ............................................ 90
5.3.2 Ammeters ......................................................................................................... 90
5.3.3 Voltmeters ........................................................................................................ 92
5.3.4 Multi-Range Voltmeter .................................................................................... 93
5.3.5 Ohm and VOM Meters ......................................................................................... 94
5.4 THE DIGITAL VOLTMETER (DVM) ................................................................... 96
5.4.1 Utilization and Advantages .............................................................................. 96
5.4.2 Basic Operation and Functional Block Diagram.............................................. 96
5.5 AC VOLTMETERS ................................................................................................. 99
5.5.1 Measurement of AC Voltages .......................................................................... 99
5.5.2 Form Factor and Waveform Errors ................................................................ 101
5.5.3 The Correction Factor .................................................................................... 102
5.5.4 The Thermocouple-Based True RMS Meter ...................................................... 104
5.6 PROBLEMS ........................................................................................................... 106

You might also like