First Name Last Name : LPF Stands For

Download as doc, pdf, or txt
Download as doc, pdf, or txt
You are on page 1of 35

Eddy Current Inspection 50 Question Quiz

~ First name ~ Last name


~ ~

LPF stands for:

Low Pulse Frequency

Low Pass Filter


1
Last Pass Filter

Low Pass Frequency

In almost all cases, eddy current inspection procedures require the equipment
2 to be calibrated to:

A reference standard

An identified defect

A crack which is ? the rejection criteria

A crack twice the rejection criteria

The HPF allows:


3

Low frequencies to pass and filters out the high frequencies

High frequencies to pass and filters out the low frequencies

For the combined frequencies to be adjusted to a harmonic balance

The shape of the waveform to be clipped beyond it frequency balance

When testing for subsurface flaws, the frequency should be:


As high as possible

As low as possible
4
Calculated to produce a 90o difference between the liftoff and flaw signals

None of the above

The coil in an eddy current probe is most often made from:


5

Iron

Copper

Silver

Platinum

The main purpose of an iron core in a probe is to:


6

Shift the transmission of eddy currents in order to decrease penetration

Allow the probe to operate at a higher frequency

Concentrate the magnetic field near the center of the probe

Allow testing of very dense materials

Probes for inspection of pipe and tubing are typically of the:


7

Surface probe variety

Bolt hole variety

Bobbin (ID) variety

All of the above are correct

Eddy current testing can be used to:


8

Detect surface and near surface cracks

Measure electrical conductivity

Measure the thickness of nonconductive coatings on plastics

Both A and B
Sliding probes usually operate in the:
9

Reflection mode

Through-transmission mode

Pulsed mode

Differential mode

Scanning speed must be controlled:


10

When using a small transducer

When using a large transducer

When using a high pass filter

When using a large low frequency probe

Which type of probe is most often used to inspect the inside diameter of a
11 machined hole?

Pencil probes

Surface probes

Bolt hole probes

Bobbin probes

Inductance is identified by the letter:


12

Probe shielding is used to:


13
Shape the eddy currents to the curvature of the part

Reduce the inductive coupling of the probe and part

Increase the probe impedance

Reduce the effects of nonrelevant features in close proximity to the probe

Use of the HPF is not recommended:


14

On thin parts

On thick parts

When scanning manually

On ferrous parts

Eddy currents are generated when:


15

A conductive material is placed in a changing magnetic field

When a conductive material is moved through a static magnetic field

When a static magnetic field is moved across the surface of a conductive


material

All of the above

The HPF is used to:


16

Eliminate low frequencies which are produced by slow changes, such as a


conductivity shift

Adjust the bandwidth to a neutral frequency in order to maximize depth of


penetration

Remove any standing waves in the output signal

Shift the waveform from positive to negative when a rejectable defect is


identified

Wide surface probes are used when scanning:


17

Large areas for very small cracks

Small areas for delaminations


Large areas for relatively large defects

None of the above

The principles of operation of the most commonly used eddy current


18 instruments are based on:

Roentgen's formulas

Maxwell's inductance bridge

Reciprocity

The Inverse Square Law

Which type of probe has a long slender housing to permit inspection in


19 restricted spaces?

Pancake probes

Pencil probes

Encircling probes

Sliding probes

A probes that can be used to inspect the entire circumference of test objects
20 are:

Encircling or bobbin probes

Circumference probes

Pencil probes

None of the above

Sliding probes are used to test:


21

Large flat surfaces

Inside bolt holes

In radiuses

Around the edges of fasteners


Pencil probes are prone to:
22

Energy spikes

Low frequency noise

Wobble

Both A and C

Since eddy current signals are affected by many different variables, it is


23 particularly important to use what when setting up the equipment?

Couplant

Fluorescent particles

Reference standards

Non abrasive cleaners

Some common classifications of probes include:


24

Surface probes

Bolt hole probes

ID probes

All of the above

When an absolute probe is brought near a ferromagnetic material, the


25 inductive reactance of the coil will:

Remain unchanged

Increase because the material will concentrate the magnetic field in and
around the coil

Decrease because the material will concetrate the magnetic field in and
around the coil

Decrease because the magnetic field of the coil will be weakened by the
material

Which type of probe is most commonly sued to inspect solid products such as
26 bar stock?

Bobbin probes

Surface coils

Encircling coils

Pencil probes

Filtering is applied to the received signal and, therefore:


27

It should be added to the base signal

It is not directly related to the probe drive frequency

Should be added to the pick-up coil

Should be subtracted from the amplitude of the dB

Probes with iron cores tend to:


28

Be more sensitive than air core probes and less affected by probe wobble

Be more difficult to use

Increase the background noise of the signal

Both B and C

When using the liftoff trace of an impedance plane instrument to distinguish


29 between several materials with high electrical conductivity, it is best to test
using a:

High frequency

Variable frequency

Low frequency

None of the above

The higher the inductance of a coil at a given frequency:


30

The more penetrating the eddy currents will be


The less sensitive the coil will be

The more sensitive the coil will be

None of the above

HPF stands for:


31

High Pulse Filter

Harmonic Pulse Filter

High Pass Filter

High Pulse Factor

The higher the frequency of the current used to drive the probe, the:
32

More effective shielding will be due to skin effect

Deeper the eddy currents will penetrate

Stronger the probe's magnetic field will be

Both A and C

A probe that is often intended to be used in contact with the test surface is
33 called a:

Reference probe

Surface probe

Transmission probe

Reflection probe

Inductance is caused by:


34

The interaction of a changing magnetic field with a conductor

Direct current

Resistance in the coil

None of the above


When testing for surface flaws, the probe drive frequency used:
35

Should be as high as possible

Should be as low as possible

Depends on the conductivity and permeability of the material

Depends only on the material conductivity

Drilled holes are commonly used to represent:


36

Pitting

Cracks

Delaminations

None of the above

Narrow EDM notches and saw cuts:


37

Are never used because they are too wide

Are never used due to their heat affected zones

Are commonly used to represent cracks

Both A and B

Most surface probe coils are wound so that:


38

They transmit a frequency that will slightly resonate the part surface

They create a static magnetic field

The axis of the coil is perpendicular to the test surface

Both B and C

Which of the following are a common eddy current reference standard?


39

Conductivity standards
Tube discontinuity standards

Hole discontinuity standards

All of the above

Filters are adjusted in:


40

Hz

KHz

MHz

THz

The main function of the LPF is to:


41

Control probe wobble

Adjust the machine to the proper conductivity standard

Shift the waveform to the left of the screen

Remove high frequency interference noise

When making a conductivity measurement, the thickness of the material


42 should be at least ___ times the standard depth of penetration.

When a probe is brought near a conductive but nonmagnetic material, the


43 coil's inductive reactance will:

Increase

Decrease

Remain the same

Remain the same until the probe touches the material


The depth of penetration is affected by:
44

Magnetic permeability

Electrical resistivity

Probe drive frequency

All of the above

Relative permeability is:


45

The permeability in a vacuum

A unitless value

The ratio of the measured permeability and the permeability in a vacuum

Both B and C

What is the relationship between electrical conductivity and electrical


46 resistivity?

They are directly proportional

They are not related

One is the inverse of the other

It depends on the test frequency

What material(s) is/are commonly used to shield an eddy current probe?


47

Ferrite

Aluminum

Lead

Both A and B

When using eddy currents to measure the thickness of a nonconductive


48 coating applied to a conductive base, the measurement is based on:
A frequency change due to liftoff

An impedance change due to a change in conductivity

An impedance change due to liftoff

Both A and C

Phase lag:
49

Increase with discontinuity depth

Decrease with discontinuity depth

Is the same as phase angle

Is the same as the angle separating the liftoff and flaw signals on an
impedance plane

A bolt hole probe and scanner is used to inspect:


50

Flat surfaces

Radiuses

Holes

Both B and C

Copyright ? The Collaboration for NDT Education..

Eddy Current Inspection 35 Question Quiz

~ First name ~ Last name


~ ~

Filters are adjusted in:


Hz

KHz
1
MHz

THz

In almost all cases, eddy current inspection procedures require the equipment
2 to be calibrated to:

A reference standard

An identified defect

A crack which is ? the rejection criteria

A crack twice the rejection criteria

An eddy current test circuit will have:


3

Resistance

Inductive reactance

A small amount of capacitance

All of the above

HPF stands for:


4

High Pulse Filter

Harmonic Pulse Filter

High Pass Filter

High Pulse Factor

Scanning speed must be controlled:


5

When using a small transducer

When using a large transducer


When using a high pass filter

When using a large low frequency probe

Some common classifications of probes include:


6

Surface probes

Bolt hole probes

ID probes

All of the above

The HPF is used to:


7

Eliminate low frequencies which are produced by slow changes, such as a


conductivity shift

Adjust the bandwidth to a neutral frequency in order to maximize depth of


penetration

Remove any standing waves in the output signal

Shift the waveform from positive to negative when a rejectable defect is


identified

A probes that can be used to inspect the entire circumference of test objects
8 are:

Encircling or bobbin probes

Circumference probes

Pencil probes

None of the above

Sliding probes usually operate in the:


9

Reflection mode

Through-transmission mode

Pulsed mode

Differential mode
A probe that is often intended to be used in contact with the test surface is
10 called a:

Reference probe

Surface probe

Transmission probe

Reflection probe

What is the relationship between electrical conductivity and electrical


11 resistivity?

They are directly proportional

They are not related

One is the inverse of the other

It depends on the test frequency

What material(s) is/are commonly used to shield an eddy current probe?


12

Ferrite

Aluminum

Lead

Both A and B

Which of the following are a common eddy current reference standard?


13

Conductivity standards

Tube discontinuity standards

Hole discontinuity standards

All of the above

Use of the HPF is not recommended:


14
On thin parts

On thick parts

When scanning manually

On ferrous parts

Probes with iron cores tend to:


15

Be more sensitive than air core probes and less affected by probe wobble

Be more difficult to use

Increase the background noise of the signal

Both B and C

Which type of probe is most commonly sued to inspect solid products such as
16 bar stock?

Bobbin probes

Surface coils

Encircling coils

Pencil probes

Probe shielding is used to:


17

Shape the eddy currents to the curvature of the part

Reduce the inductive coupling of the probe and part

Increase the probe impedance

Reduce the effects of nonrelevant features in close proximity to the probe

Narrow EDM notches and saw cuts:


18

Are never used because they are too wide

Are never used due to their heat affected zones

Are commonly used to represent cracks

Both A and B
Discontinuities, such as delaminations, that are in a plane that is parallel with
19 the test surface will likely:

Be easily detected with a surface probe

Be easily detected with an internal probe

Be easily detected with an external probe

None of the above

Which component features should be similar to the reference standard?


20

Material thickness

Material geometry

Material conductivity

All of the above

When a probe is brought near a conductive but nonmagnetic material, the


21 coil's inductive reactance will:

Increase

Decrease

Remain the same

Remain the same until the probe touches the material

The depth of penetration is affected by:


22

Magnetic permeability

Electrical resistivity

Probe drive frequency

All of the above

The main purpose of an iron core in a probe is to:


23
Shift the transmission of eddy currents in order to decrease penetration

Allow the probe to operate at a higher frequency

Concentrate the magnetic field near the center of the probe

Allow testing of very dense materials

The main function of the LPF is to:


24

Control probe wobble

Adjust the machine to the proper conductivity standard

Shift the waveform to the left of the screen

Remove high frequency interference noise

The higher the frequency of the current used to drive the probe, the:
25

More effective shielding will be due to skin effect

Deeper the eddy currents will penetrate

Stronger the probe's magnetic field will be

Both A and C

The principles of operation of the most commonly used eddy current


26 instruments are based on:

Roentgen's formulas

Maxwell's inductance bridge

Reciprocity

The Inverse Square Law

When maintaining constant liftoff is a problem, what type of probe should be


27 used?

A absolute probe

A differential probe

A reflection probe
Both B and C

Relative permeability is:


28

The permeability in a vacuum

A unitless value

The ratio of the measured permeability and the permeability in a vacuum

Both B and C

Filtering is applied to the received signal and, therefore:


29

It should be added to the base signal

It is not directly related to the probe drive frequency

Should be added to the pick-up coil

Should be subtracted from the amplitude of the dB

LPF stands for:


30

Low Pulse Frequency

Low Pass Filter

Last Pass Filter

Low Pass Frequency

Probes for inspection of pipe and tubing are typically of the:


31

Surface probe variety

Bolt hole variety

Bobbin (ID) variety

All of the above are correct

The coil in an eddy current probe is most often made from:


Iron

Copper
32
Silver

Platinum

Inductance is identified by the letter:


33

Inductance is caused by:


34

The interaction of a changing magnetic field with a conductor

Direct current

Resistance in the coil

None of the above

Most surface probe coils are wound so that:


35

They transmit a frequency that will slightly resonate the part surface

They create a static magnetic field

The axis of the coil is perpendicular to the test surface

Both B and C

Copyright ? The Collaboration for NDT Education..

Eddy Current Inspection 20


Question Quiz
~ First name ~ ~ Last name ~

In almost all cases, eddy current inspection procedures require the


equipment to be calibrated to:

A reference standard
An identified defect
1 A crack which is ? the rejection criteria
A crack twice the rejection criteria

An eddy current test circuit will have:


2
Resistance
Inductive reactance
A small amount of capacitance
All of the above

Drilled holes are commonly used to represent:


3
Pitting
Cracks
Delaminations

None of the above

When maintaining constant liftoff is a problem, what


4 type of probe should be used?

A absolute probe
A differential probe
A reflection probe
Both B and C
Which type of probe has a long slender housing to permit
5 inspection in restricted spaces?

Pancake probes
Pencil probes
Encircling probes
Sliding probes

Inductance is identified by the letter:


6
L
M
Z
X

Filters are adjusted in:


7
Hz
KHz
MHz
THz

When a probe is brought near a conductive but


8 nonmagnetic material, the coil's inductive reactance will:

Increase
Decrease
Remain the same
Remain the same until the probe touches the material

Sliding probes usually operate in the:


9
Reflection mode
Through-transmission mode
Pulsed mode
Differential mode
Wide surface probes are used when scanning:
10
Large areas for very small cracks
Small areas for delaminations
Large areas for relatively large defects
None of the above

When using eddy currents to measure the thickness of a


11 nonconductive coating applied to a conductive base, the
measurement is based on:

A frequency change due to liftoff


An impedance change due to a change in conductivity
An impedance change due to liftoff
Both A and C

Most surface probe coils are wound so that:


12
They transmit a frequency that will slightly resonate the
part surface
They create a static magnetic field
The axis of the coil is perpendicular to the test surface
Both B and C

The HPF is used to:


13
Eliminate low frequencies which are produced by slow
changes, such as a conductivity shift
Adjust the bandwidth to a neutral frequency in order to
maximize depth of penetration
Remove any standing waves in the output signal
Shift the waveform from positive to negative when a
rejectable defect is identified

Which type of probe is most commonly sued to inspect


14 solid products such as bar stock?
Bobbin probes
Surface coils
Encircling coils
Pencil probes

Probes for inspection of pipe and tubing are typically of


15 the:

Surface probe variety


Bolt hole variety
Bobbin (ID) variety
All of the above are correct

Some common classifications of probes include:


16
Surface probes
Bolt hole probes
ID probes
All of the above

The HPF allows:


17
Low frequencies to pass and filters out the high
frequencies
High frequencies to pass and filters out the low
frequencies
For the combined frequencies to be adjusted to a
harmonic balance
The shape of the waveform to be clipped beyond it
frequency balance

When testing for subsurface flaws, the frequency should


18 be:

As high as possible
As low as possible
Calculated to produce a 90o difference between the liftoff
and flaw signals
None of the above

The main purpose of an iron core in a probe is to:


19
Shift the transmission of eddy currents in order to
decrease penetration
Allow the probe to operate at a higher frequency
Concentrate the magnetic field near the center of the
probe
Allow testing of very dense materials

A probes that can be used to inspect the entire


20 circumference of test objects are:

Encircling or bobbin probes


Circumference probes
Pencil probes
None of the above

Copyright ? The Collaboration for NDT Education..

|Quiz Main Page|

Click on the results button when you're done

Questions & Answers


- Electromagnetic Testing (ET) Level -
1-2-3
by Edward A. Ginzel *
Acknowledgement:
Materials Research Institute Canada are thanked for permission to use their
Q&A of the NDT TestMaker software CD-ROM.
|Quiz Main Page|

Click on the results button when you're done


Level Question [Ref]
1
1 Eddy current testing requires that the material tested be [ECM]
a) magnetic
b) non-magnetic
c) conductive
d) both b and c
An oscillator used in generating time varying current in an ECT
2 [ECM]
instrument usually supplies currents at frequencies from
a) 1 Hz to 20 kHz
b) 50 Hz to 60 Hz
c) 1 kHz to 2 MHz
d) 0.5 MHz to 4 GHz
3 The eddy current probe is most simply a [ECM]
a) coil of copper wire
b) semiconductor
c) capacitor
d) ceramic resistor
Critical to the principles of eddy current testing is the understanding
4 [ECM]
that
a) all materials conduct electricity
b) all current flow has associated with it a magnetic field
c) only ferro electric materials conduct eddy currents
d) copper wire is used for all probe coils
The magnitude of magnetic flux density associated with current
5 flowing in a conductor varies with position and current. Magnetic flux [ECM]
density is therefore a(n)
a) vector quantity
b) scalar quantity
c) square wave quantity
d) unknown quantity
The magnetic flux density associated with a current carrying conductor
6 [ECM]
_________ with increasing distance from the conductor.
a) increases
b) decreases
c) wobbles
d) remains constant
7 The Weber (Wb) is the unit of [ECM]
a) magnetic flux
b) magnetic flux density
c) eddy current intensity
d) the Hall effect
When a magnetic flux near a conductor is caused to change (as when a
8 [Burns]
magnet is moved past a coil) the result is
a) an increased resistance in the coil wire
b) and induced voltage
c) de-magnetization of the magnet
d) phase reversal
According to Ohm's Law, in any electrical circuit, the voltage divided
9 [ECM]
by the circuit impedance is equal to
a) magnetic flux density
b) capacitance
c) current flow
d) phase angle
When an alternating voltage induces an EMF, the induced EMF in
10 [Burns]
opposition to the change causing it. This is a statement of
a) Faraday's Law
b) Lenz's Law
c) Snell's Law
d) Eddy's Law of Voltage
11 Eddy currents are so named because [ECM]
a) of the Canadian match manufacturer that discovered them
b) of the circular probes required to induce them
c) of their circular paths
d) small swirl patterns form on the surface domains and crystal
structure
A device which stores electrical energy within a specified volume is
12 [Ramshaw]
a(n)
a) inductor
b) capacitor
c) conductor
d) resistor
The field that couples an eddy current probe and a test piece causes
13 [ECM]
eddy currents to flow by
a) resonance
b) induction
c) friction
d) the Bernoulli effect
The region around a magnet where its magnetic effects can be detected
14 [Jacobowitz]
is called the
a) near zone
b) induction zone
c) magnetic field of force
d) electro-magnetic region of influence
15 In a fixed magnetic field, a compass needle would arrange itself [Jacobowitz]
a) parallel to the lines of force
b) perpendicular to the lines of force in the plane of the field
c) perpendicular to the lines of force at right angles to the plane of the
field
d) 45 to the field lines
The location where magnetic lines of flux are considered to leave a
16 [Jacobowitz]
magnet is
a) the north pole of that magnet
b) the south pole of that magnet
c) dependent on the position of the earth's magnetic pole
d) called Domain Centre
The location where magnetic lines of flux are considered to enter a
17 [Jacobowitz]
magnet is
a) the north pole of the magnet
b) the south pole of the magnet
c) dependent on the position of the sun
d) called domain centre
18 Direct current can be used to induce eddy currents if [Handbook]
a) the test piece is moved cyclically near the current carrying conductor
b) the current carrying conductor is moved cyclically near the test piece
c) either a or b
d) the diameter of the conductor is less than 1/10 the thickness of the
test piece
19 The rate of flow of electric charge defines [Jacobowitz]
a) voltage
b) coulomb
c) electric current
d) impedance
20 1 mA can also be written as [Jacobowitz]
a) 0.01 A
b) 0.001 A
c) 100 A
d) 1000 MA
21 A sample which has a very low resistance will also have very high [Jacobowitz]
a) resistivity
b) conductivity
c) conductance
d) impedance
22 Compared to conductive wires, insulators are considered to have [Jacobowitz]
a) very high resistance
b) very low resistance
c) about the same resistance
d) high conductance
Resistance of a conductor is usually given at a specified temperature.
23 [Jacobowitz]
Correction for resistance at other temperatures is obtained by using
a) liquid nitrogen
b) the temperature coefficient of resistance
c) vacuum or bell chambers
d) matching circuits
Increasing the temperature of an aluminium test specimen form 20C.
24 [ECM]
to 40C. will result in ___________ specimen resistivity.
a) no change to
b) doubling
c) halving
d) a slight increase in
25 Eddy currents are normally induced in a sample such that they are [ECM]
a) parallel to the coil windings
b) parallel to the sample surface
c) limited to the area of inducing field
d) all of the above
Decrease in eddy current flow with increasing depth from the surface is
26 [ECM]
known as
a) absorption
b) the Hall effect
c) the skin effect
d) the edge effect
27 The pickup (receiver) coil of an eddy current probe is [Handbook]
a) always the same coil as the magnetizing coil
b) identical in diameter to the magnetizing coil
c) coincident in position to the magnetizing coil
d) of any size, position or even configuration
Heat generation associated with eddy current testing of magnetic
28 [Handbook]
materials is a result of
a) dispersion and diffraction
b) ohmic or resistive losses
c) hysteresis losses
d) both b and c
What is the difference between electric currents and eddy currents on
29 [Handbook]
the atomic level?
a) there is no difference
b) direction of electron flow
c) no magnetic field is associated with electric current
d) eddy currents do not occur in semi conductors
30 Current flowing in the eddy current transducer would be considered [Handbook]
a) eddy current
b) magnetizing currents
c) back EMF
d) all of the above

References (NDT TestMaker software)

1. [Handbook] ASNT; Nondestructive Testing Handbook, vol. four,


Electromagnetic Testing, Second Edition, ASNT, 1986
2. [Jacobowitz] Jacobowitz,H.; Electricity Made Simple, Doubleday &
Company, 1959
3. [Mix] Mix, P.E.,; Introduction to Non-Destuctive Testing, John Wiley
& Sons, 1987
4. [ECM] V.S.Cecco, G.van Drunen and F.L.Sharp, Eddy Current Manual
vol. 1, Chalk River Nuclear Laboratories, 1983
5. [Burns] D.M.Burns & S.G.G.MacDonald Physics for Biology and
Premedical Students, Addison-Wesley Publishing Company, 1970
6. [Ramshaw] R.S.Ramshaw & R.G.van Heeswajk, Electromagnetic
Fields and Forces University of Waterloo, Standford Educational Press,
1979
7. [Blitz] J.Blitz, Electrical and Magnetic Methods of Nondestructive
Testing, Adam Hilger, IOP Publishing Ltd., 1991
8. [Techniques] D.E.Bray and D.McBride, Nondestructive Testing
Techniques John Wiley & Sons, 1992
9. [Gardner] W.E.Gardner, Editor, Improving the Effectiveness and
Reliability of Nondestructive Testing Pergamon Press, 1992
10. [ASME] American Society of Mechanical Engineers, ASME Section
V, The American Society of Mechanical Engineers, 1992
|NDTnet Forum| |Quiz Main Page|
|NDTnet|
NDTnet - [email protected]

Level Question [Ref]


1
Which of the following conditions is not necessary for eddy
1 [ECM]
current testing?
a) electrical contact
b) specimen conductivity
c) an alternating magnetic field
d) induced electrical current
Which of the following is not a mandatory component in a
2 [ECM]
"basic" eddy current test apparatus?
a) an AC source
b) a coil (probe)
c) an impedance plane
d) a volt meter
Which of the following is not a probe configuration used in
3 [ECM]
eddy current testing?
a) self inductance (single coil)
b) send-receive (2 coils)
c) magnetic reaction (coil & Hall detector)
d) semi-conductor reaction (2 Hall detectors)
The sense or direction of a magnetic field around a conductor
4 [ECM]
is most commonly determined using
a) Lenz's Law
b) Ohm's Law
c) a Rowland Ring
d) the right hand rule
5 Tesla or Webers per square metre (Wb/m) are units of [ECM]
a) eddy current
b) impedance
c) reluctance
d) magnetic flux density
If the electric current in a coil is doubled the magnetic flux
6 [ECM]
density
a) halves
b) remains unchanged
c) doubles
d) quadruples
An increase in which of the following would result in the
7 [Burns]
increase of magnetic flux density (B) in a solenoid?
a) core permeability
b) number of coil turns
c) current in the coil
d) all of the above
A voltage in induced in a region of space when there exists a
8 [ECM]
changing magnetic field. This is a statement of
a) Faraday's Law
b) Oersted's Law
c) Helmholtz's Theorem
d) Ohm's Law
9 Lenz's Law states [Burns]
a) an alternating magnetic field induces an alternating voltage
b) the magnitude of induced current is a function of magnetic
flux through a circuit
c) the induced EMF is opposite to the change causing it
d) = B A cos where B= flux density, A = circuit area and
= the angle between B and the circuit area A
10 The back EMF opposing the inducing EMF is a result of [ECM]
a) the Hall effect
b) eddy current flow
c) geo-magnetic reversals
d) weak nuclear forces
Which of the following is not an example of
11 [Ramshaw]
electromechanical energy conversion devices?
a) a dry cell battery
b) a generator or alternator
c) a microphone
d) an electric motor
An eddy current test system closely approximates a
12 transformer. In this approximation, what would the [ECM]
"secondary coil" be represented by?
a) the induced eddy currents
b) the eddy current probe
c) the test sample
d) a Hall detector used as a receiver
By convention, the direction of a magnetic line of force is
13 represented by an arrow on a line. The arrow would point in [Jacobowitz]
the direction
a) in which a unit north pole would be moved
b) in which a unit south pole would be moved
c) perpendicular to the plane of the line
d) indicated by the thumb in the left hand rule
14 The magnetic line of force is [Jacobowitz]
a) a mythical quantity
b) a imaginary but useful concept
c) equal to 1gm mass when converted by Einstein's equation
d) 1 micron diameter and 10 microns long
The principal cause of magnetism in a naturally magnetic
15 [Jacobowitz]
substance is
a) hysteresis
b) the weak nuclear force
c) uncompensated electron spin
d) graviton concentration in the Domain wall
The number of lines of magnetic flux divided by a unit area is
16 [Jacobowitz]
the
a) permeability
b) flux density
c) pole strength
d) field intensity
Moving a direct current carrying conductor up and down near
17 [Handbook]
a conductive test piece will result in
a) no current flow in the test piece
b) DC being induced in the test piece
c) AC being induced in the test piece
d) a short circuit
The relationship between electric current flow, electromotive
18 [Jacobowitz]
force and resistance to electric current flow is described by
a) Lenz's law
b) Ohm's law
c) Faraday's rule
d) the ampere-ohm equation
19 Another term for voltage is [Jacobowitz]
a) electromotive force
b) magnetomotive force
c) potential drop
d) both a and c
In order to use a galvanometer (which normally measures
currents in the range of milliamps) as an ammeter measuring
20 [Burns]
10 to 20 amps you would put a _________ in ___________
with the galvanometer.
a) resistor, series
b) resistor, parallel
c) capacitor, series
d) capacitor, parallel
Conductance is an electrical quantity which can also be
21 [Jacobowitz]
defined as the reciprocal of
a) inductance
b) resistance
c) resistivity
d) reluctance
When an eddy current probe is brought near a conductive
22 [ECM]
sample the net magnetic flux in the system
a) increases
b) decreases
c) remains unchanged
d) drops to zero when the part is contacted
23 Eddy current density in a sample is [ECM]
a) proportional to the conductivity of the sample
b) proportional to the permeability of the sample
c) inversely proportional to the depth from the surface of the
sample
d) all of the above
Strictly speaking, the standard skin depth equation; J/Jo = (e^-
24 [ECM]
) sin (wt-), is true for only
a) thick material and planar magnetic fields
b) tubular products
c) thin plate inspection
d) all of the above
At 2 standard depths of penetration, eddy current density is
25 [ECM]
about ____ that at the surface.
a) 0,66
b) 0,37
c) 0,14
d) 0,09
When inspecting a rod with an encircling coil the eddy
26 [ECM]
current density at the centre of the rod is zero for
a) rods with diameters greater than 2
b) rods with radius greater than 2
c) all conditions
d) no condition, a slight current density will always exist
27 Phase lag of eddy currents in a sample is dependent on [ECM]
a) depth into the sample
b) resistivity of the test piece
c) relative magnetic permeability of the sample
d) all of the above
28 Why are eddy current coils not made using iron wire? [Blitz]
a) to avoid hysteresis effects
b) to make mathematical calculations easier
c) to prevent excessive heat build-up
d) for cathodic breakdown considerations
The higher the value of inductance for a given frequency the
29 [Blitz]
greater the degree of
a) balance ability
b) sensitivity
c) Q factor
d) capacitive reactance
30 The transmit-receive or transformer style probe provides [Blitz]
a) improved s/n ratio
b) increased sensitivity to deeper defects
c) both a and b
d) no advantage over single coil probes

You might also like