MS-3 Laser Scanner User Manual: P/N 84-000003 Rev C
MS-3 Laser Scanner User Manual: P/N 84-000003 Rev C
User Manual
Disclaimer
The information and specifications described in this manual are subject to change without notice.
Technical Support
For technical support, e-mail: [email protected].
Warranty
For current warranty information, see: www.microscan.com/warranty.
Table of Contents
Chapter 1 Quick Start
Step 1 Check Hardware .......................................................................... 1-2
Step 2 Connect the System ..................................................................... 1-3
Step 3 Position Symbol and Scanner ...................................................... 1-4
Step 4 Install ESP.................................................................................... 1-5
Step 5 Select Scanner Model .................................................................. 1-6
Step 6 Autoconnect ................................................................................. 1-7
Step 7 Test for Read Rate ....................................................................... 1-8
Step 8 Calibrate the Scanner .................................................................. 1-9
Step 9 Save Calibration Settings for Power-On .................................... 1-10
Step 10 Configure the Scanner ............................................................. 1-11
Chapter 2 Using ESP
Setup Mode ............................................................................................. 2-2
Application Mode ..................................................................................... 2-3
Dropdown Menus .................................................................................... 2-4
Connect Menu ........................................................................................ 2-9
View ...................................................................................................... 2-11
Navigating in ESP.................................................................................. 2-12
Send/Receive Options ........................................................................... 2-13
Chapter 3 Communications
Communications by ESP......................................................................... 3-2
Communications by Serial Command ..................................................... 3-2
Password Protection................................................................................ 3-3
RS-232/422 Host Port ............................................................................. 3-4
RS-232 Auxiliary Port ............................................................................ 3-11
Preamble ............................................................................................... 3-20
Postamble.............................................................................................. 3-21
LRC Status ............................................................................................ 3-22
Intercharacter Delay .............................................................................. 3-22
Chapter 4 Read Cycle
Read Cycle by ESP ................................................................................. 4-2
Read Cycle by Serial Command ............................................................. 4-2
Multisymbol.............................................................................................. 4-3
Number of Symbols ................................................................................. 4-4
Serial Trigger ......................................................................................... 4-12
End of Read Cycle................................................................................. 4-14
Decodes Before Output ......................................................................... 4-16
Scanner Setup ....................................................................................... 4-17
Laser Setup ........................................................................................... 4-21
Chapter 5 Symbologies
Symbologies by ESP ............................................................................... 5-2
Symbologies by Serial Command ........................................................... 5-2
MS-3 Laser Scanner User Manual iii
Table of Contents
Code 39 ................................................................................................... 5-3
Code 128 ................................................................................................. 5-6
Interleaved 2 of 5................................................................................... 5-10
Codabar................................................................................................. 5-13
UPC/EAN .............................................................................................. 5-16
Code 93 ................................................................................................. 5-19
Pharmacode .......................................................................................... 5-20
Narrow Margins ..................................................................................... 5-22
Symbology ID ........................................................................................ 5-23
Background Color.................................................................................. 5-24
Autodiscriminate .................................................................................... 5-24
Symbol Ratio Mode ............................................................................... 5-25
Chapter 6 I/O Parameters
Output Conditions by ESP Menu............................................................. 6-2
I/O Parameters by Serial Command ....................................................... 6-3
Symbol Data Output ................................................................................ 6-4
Message Output ...................................................................................... 6-7
No Read Message................................................................................... 6-8
Bad Symbol Message ............................................................................. 6-9
No Symbol Message ............................................................................. 6-10
Beeper................................................................................................... 6-11
Partial Output ........................................................................................ 6-12
Serial Verification .................................................................................. 6-13
EZ Button .............................................................................................. 6-15
Output 1................................................................................................. 6-18
Output 2................................................................................................. 6-24
Output 3................................................................................................. 6-25
Quality Output ....................................................................................... 6-26
Chapter 7 Matchcode
Matchcode by ESP .................................................................................. 7-2
Matchcode by Serial Command .............................................................. 7-2
Overview of Matchcode ........................................................................... 7-3
Using Master Symbols ............................................................................ 7-3
Matchcode Type ...................................................................................... 7-4
New Master Pin ....................................................................................... 7-9
Master Symbol Database ...................................................................... 7-10
Chapter 8 Diagnostics
Diagnostics by ESP ................................................................................. 8-2
Diagnostics by Serial Command ............................................................. 8-2
Diagnostic Messages Overview .............................................................. 8-3
Counts ..................................................................................................... 8-4
Hours Since Last Reset........................................................................... 8-6
Laser High ............................................................................................... 8-7
Laser Low ................................................................................................ 8-8
Scanner Communication
There are five ways to program the scanner:
1. Microscans Windows-based ESP, which offers point-and-click ease of use and
visual responses to user adjustments.
2. Serial commands such as <K100,1> can be sent from ESPs Terminal or another
terminal program.
3. Embedded firmware (onboard menus).
4. Symbol configuration.
5. The EZ Button on the top of the scanner.
Note: You can learn the current setting of any parameter by inserting a question mark after
the command number, as in <K100?> To see all K commands, send <K?>.
Laser Aperture
The MS-3 has been tested for compliance with FCC (Federal Communications Commission) regulations
and has been found to conform to all applicable FCC Rules and Regulations.
To comply with FCC RF exposure compliance requirements, this device must not be co-located or operate
in conjunction with any other antenna or transmitter.
Changes or modifications not expressly approved by the party responsible for compliance could void the
users authority to operate the equipment.
The MS-3 conforms to the following requirements:
FCC 15.107:2015 Class B Conducted Emissions
FCC 15.109(g):2015 Class B Radiated Emissions
The MS-3 has been tested for compliance with CE (Conformit Europenne) standards and guidelines,
and has been found to conform to applicable CE standards, specifically the following requirements:
EN 55022:2010/AC:2011 Class B Radiated Emissions
EN 55022:2010/AC:2011 Class B Conducted Emissions
EN 55032:2012 Class B Radiated Emissions
EN 55032:2012 Class B Conducted Emissions
The MS-3 has been tested by an independent electromagnetic compatibility laboratory in accordance with
the applicable specifications and instructions.
The MS-3 is compliant with FDA performance standards for laser products except for deviations pursuant
to laser notice no. 50, dated June 24, 2007.
Canada
The MS-3 conforms to the following requirement:
ICES-003:2012 updated 2014 Class B Radiated and Conducted Emissions
Japan
The MS-3 conforms to the following requirement:
VCCI:2015-04 Class B Radiated and Radiated and Conducted Emissions
This declaration is based upon information obtained from sources which Microscan believes to be reliable, and
from random sample testing; however, the information is provided without any representation of warranty,
expressed or implied, regarding accuracy or correctness. Microscan does not specifically run any analysis on our
raw materials or end product to measure for these substances.
The information provided in this certification notice is correct to the best of Microscans knowledge at the date of
publication. This notice is not to be considered a warranty or quality specification. Users are responsible for
determining the applicability of any RoHS legislation or regulations based on their individual use of the product.
In regards to RoHS Directive 2011_65_EU Microscan produces Monitoring and Control Instruments as well as
Industrial Monitoring & Control Instruments as defined within the directive. Microscan has developed and is
implementing a RoHS2 compliance plan with the intention of bringing all active products listed in our current
marketing literature within full compliance as per the directive deadlines.
Key milestones for the transition plan are as follows:
Complete internal product audit by July 2014.
Initial Monitoring and Control Instruments RoHS2 compliant products available by December 2014
Initial Industrial Monitoring & Control Instruments RoHS2 compliant products available by July 2015
All new products introduced in 2015 are expected to be WEEE & RoHS2 compliant.
Microscan will mark the products with the CE marking that complies with the RoHS2 process to acquire CE certification
per the example given: Example >> Machinery directive + EMC directive + RoHS2 = Declaration of Conformity.
Chapter Contents
Step 1 Hardware Required ..........................................................................1-2
Step 2 Connect the System.........................................................................1-3
Step 3 Position Symbol and Scanner ..........................................................1-4
Step 4 Install ESP .......................................................................................1-5
Step 5 Select Reader Model ........................................................................1-6
Step 6 Autoconnect .....................................................................................1-7
Step 7 Test for Read Rate...........................................................................1-8
Step 8 Calibrate the Reader........................................................................1-9
Step 9 Save Calibration Settings for Power-On .........................................1-10
Step 10 Configure the Reader ...................................................................1-11
This section is designed to get the scanner up and running quickly so the user can get a
sense of its capabilities and test sample bar code symbols. Detailed setup information
for configuring the scanner for your specific application can be obtained in the subse-
quent chapters.
If connecting to a host with an IB-131 Interface box, you will need the following:
Itema Part Number Description
1 FIS-0003-XXXXG MS-3 laser scanner
2 FIS-0001-0035G IC-332 Interface converter
3 99-000018-01 IB-131 Interface box
4 user-supplied Host computer
5 61-300026-03 Null modem configuration cable
97-000012-02
6 Power supply
(90-264 VAC, 24VDC)
7 99-000017-01 Optional object detector
a. See figure 1-1 for diagram of system.
2
3
4 Scanner
Network
Host
6 7
Network
the IB-131 HOST 25-pin connec-
Host
5
tor via the null modem cable (5).2
(Refer to IB-131 Interface on
page A-8 for pin connections.) 6
7
4. Connect power supply (6) to the
IB-131 POWER connector.
5. Apply power to the system.
Figure 1-2 Hardware Connections
1. Since power supply is included in the single cabling assembly, the reader cable should not
exceed 3 feet. RS232 cabling from the IB-131 to the host can be up to 47 feet provided it does
not include power input.
2. If using your own null modem RS232 host cable, be certain that the hosts TxD connects to the
readers RxD and the readers TxD connects to the hosts RxD.
Skew
Bar code axis
symbol
Tilt
axis
Scanner
Pitch
axis
Scan line
Note: If using an I 2/5 symbol, verify that the number of characters in the symbol being
scanned matches the symbol length enabled for the I 2/5 symbol type. (Default is 10 and 6.)
See Interleaved 2 of 5 on page 5-13.
Note: If you need to select another model later, you can find it in the App Mode under
Model on the menu bar.
1. Select MS-3 laser from the menu and click OK.
If you do not want to make this selection every time you load ESP, uncheck Show
this window at Startup.
2. Select the default name, for example MS-3 Laser-1, or type in a file name of
your choice and click OK.
3. Click Yes when the connect to the reader dialog appears.
Step 6 Autoconnect
1. In the Connecting to... dialog, if your communications port is not the default
COM1, use the pull down arrow to change your communications port.
3. If connection fails, enable a different Com port, check connections, and try again.
Tip: If you do not see either the CONNECTED or DISCONNECTED message at the bot-
tom of your dialog, try expanding the ESP window horizontally.
Important Note: When you connect to the reader, the readers settings will be loaded
into ESP.
By ESP
After connecting to the scanner, ESP will open in Easy Setup Mode.
1. Click the EZ Button in Easy Setup Mode to begin the read rate test.
By the EZ Button
EZ Button
1. Press and hold the EZ Button
on the MS-3 until you hear one
Laser
T ES T
beep and see o Read rate
ne LED momen-tarily turn performance
MS-3
100% GR
amber. This will signal the LEDs 80% ST
beginning
routine.1 of the read rate 60%
release. PWR
1 1- 0000 63-0 1
1. This assumes that the default functions of the EZ Button have not been re-programmed.
The test will end automatically when the optimum combination of settings has been
achieved.
By ESP
Click the Calibrate button in Easy Setup Mode to begin the read rate test.
You can observe the progress of the calibration routine on the Calibration popup.
Note: You can also calibrate the scanner in the Calibration menu in the Apps Mode.
By the EZ Button
EZ Button
Press and hold the EZ Button
until you hear two beeps and see
LASER
TEST
two amber performance LEDs turn
amber to indicate that the calibra- Read rate MS-3
performance
100% GR
tion is in progress.
The scanner will beep once at the LEDs 80% ST
60%
end of calibration. 40%
20%
PWR
11-000063-01
By ESP
Caution: If you have settings in the scanner that you have not yet loaded into ESP, the
ESP settings will overwrite the scanners settings when you save. Under these condi-
tions, it is highly recommended that you first do Receive Reader Settings before you
save for power-on.
After testing and/or calibrating the scanner, you can save the settings for power-on by
clicking the Save button.
LASER
you hear three beeps and see three T ES T
MS-3
performance 100% GR
LEDs 80% ST
60%
40%
20%
PWR
1 1- 0000 63-0 1
1. This assumes that the default functions of the EZ Button have not been re-programmed.
By ESP
To change reader settings, or to access the utilities or terminal window, click on the
App Mode button.
By Serial Commands
From your terminal program or the terminal screen in ESP, you can enter serial string
commands and configuration and utility commands as described in this manual.
See Serial Configuration Commands on page A-13 and Summary of Utilities
Commands on page 11-3.
Note: You can learn the current setting of any parameter by inserting a question mark
after the number, as in <K100?> To see all K commands, send <K?>.
This section is designed to help you understand the structure, elements, and application
of the ESP (Easy Setup Program).
When you start up ESP, unless otherwise specified, you will enter the Setup Mode for
initial setup. From there, you move easily into the App Mode (application mode) where
you can access several configuration and utilities menus.
Setup Mode
Setup Mode
In Setup Mode you are presented with the Test option and if appropriate for your
application, Calibration. After connecting to your scanner (or reader) the Setup Mode
will first appear. This will provide you with instructions specific to your model that will
help you in positioning, testing, and if appropriate, calibrating.
Test
Click the Test button to start the read rate test for a quick indication of the read capa-
bilities and the limits of your application. When Decodes per Second is unchecked,
the test will count the percentage of decodes relative to the number of actual scans.
Click Stop to end the test.
Calibrate
Some models include a calibration routine that will optimize reads by comparing read
rates at various settings in focal lengths, scan speeds, and gain settings.
Test button
Calibration is avail-
able on some models
Application Mode
From Setup Mode, you can click on the App Mode button to access specific configura-
tion menus, utilities, and a terminal window where serial commands can be entered.
Note that the App Mode and Setup Mode buttons appear in the same position to allow
easy switching between these primary modes.
The icons on this Click here to open Click here to
toolbar are for operations a terminal window access read rate
and other utilities
Pulldown menus
Note: For specific information on any of the icons shown above in the operations bar or
configuration bar, see specific chapters in this document.
Pulldown Menus
Pulldown Menus
File
New
Whenever New is selected, the default configuration of
ESP is loaded.
Open/Save
When Save or Save As is selected, the ESP configura-
tion is saved to the host computers hard drive and
available whenever the same file is selected under
Open.
Important: When you save menu changes to your
hard drive, these changes are not saved to your reader.
Figure 2-1 shows how settings can be saved and
received between ESP and the reader and ESP and the
host hard drive.
Host hard
drive
Import/Export
Import converts the ASCII settings from a text file to ESP configuration settings.
Export converts the active ESP configuration settings to an ASCII text file.
Model
In Model you can select between models. When you choose
another model, your current connection with your present model
will be terminated.
To connect to another model, select New Model, choose a new
model and click OK.
Note that all the models you have enabled by selecting will con-
tinue to appear in the Model menu and that the same menu is
repeated when clicking the Switch Model icon.
When you save your ESP file, you will be saving the settings of all the models defined in
a single ESP file.
Pulldown Menus
Options
You can use the Options menu
save memos and set up ESP
Preferences.
Note: Preferences will be saved
and be loaded into ESP when
ESP is opened next, whether or
not you save the ESP file to the
computer.
Preferences
General tab
Reload Last File
At startup, reloads the last file
saved to the computer.
Show Model Prompt
At startup, remembers the last connected model and displays it in the Connecting...
dialog whenever you attempt to connect.
Skip Easy Mode
At startup, Skips the Easy Mode and opens directly in the application mode.
Show Connect Prompt
At startup, displays the Would you like to connect... prompt.
Receive After Connect
At startup, loads the readers settings into ESP. (This is not recommended if you want to
preserve your ESP settings for future use.)
Enable Send and Save as Factory Settings
At startup, enables the Send and Save as Factory option in the Send/Recv com-
mand.
Pulldown Menus
Bar Code Options tab
Sets up the sizing, font, and
caption parameters for the bar
codes that you can create and
print from the Bar Code
Options Dialog under View.
Sizing Information
Sets the bar height (in inches)
and bar widths (in thousands of
an inch) of the symbols. A bar
width of 13 is 0.013 inches.
Caption
Sets the characteristics of cap-
tion font and its placement in
relation to the bar code symbol.
Advanced tab
As indicated.
Document Memo
Whatever you type into the Document Memo will appear in a text box whenever your
cursor hovers over the Document Memo option.
Model Memo
Memos created in Model Memo are specific to the model enabled when the message
was created.
Connect Menu
Autoconnect
Generally Autoconnect will be the quickest way that ESP
can get connected to a reader. Autoconnect will try con-
necting at the most common communications settings and
step through the various settings until they match up with
the hosts settings.
Connect
When you select Connect, you will need to manually select the communications set-
tings from a popup dialog.
Configure Multidrop
To connect to the reader by Multidrop, you will need a multidrop concentrator such as
the MS-Connect 5100 and the required power supplies and cabling to communicate
with your scanner(s).
1. From the Communications drop down menu in
ESP, select Configure Multidrop to bring up the
Multidrop Settings dialog.
2. If necessary, change the default address to
match the address of your multidropped scanner
and click OK.
If the host serial port is not connected in ESP,
you will see the following popup message:
ESP is not currently connected to the multidrop
concentrator. Do you wish to establish a connec-
tion now?
3. Click Yes.
You will see a Serial Communication Parameters dialog as shown on the next
page.
Connect Menu
4. Select the concentrators host port communications
settings.
When you click Connect, you will be connecting to
your concentrator, which can then relay commands to
the scanner whose address was set in the Multidrop
Settings dialog.
5. Click Connect.
6. You should see the CONNECTED message in green
at the bottom of the window along with the scanners
Multidrop address.
View
View indicates the current view (with a dot before the item)
and allows you to quickly move to other views which are also
accessed by clicking the icons on the toolbars.
It also allows you to access the Barcode Dialog.
Barcode Dialog
In the Barcode Dialog you can create bar code symbols by directly typing in the text bar
code text you want to appear in Code 128 bar code symbols. This is a very useful tool
for creating configuration symbols which allows you to configure your reader by reading
the bar code symbols you print out. To configure the size, text and caption parameters,
see Bar Code Options tab on page 2-8.
Navigating in ESP
Navigating in ESP
To change reader settings, or to access the utilities or terminal window, click on the
App Mode button.
Send/Receive Options
To access Receive, Save and Default options, click the Send/Recv
button: You can also access this selector by right-clicking in any of the
configuration views.
Receiving
From the Send/Recv selector, select Receive Reader Settings.
Caution: Do no select this option if you do not want to upload the readers settings. For
example, if your ESP file has a number of custom settings that you want to maintain
and download into the reader, these settings would be lost by choosing Yes.
This is useful if you want to receive (upload) the readers settings and save them as a
computer file for later retrieval. For example, if your reader has settings that you do not
want to change, choosing Yes would allow you to load those settings to ESP and save
as an ESP file for later retrieval.
Receiving the readers settings will also assure that you will not be subsequently saving
any unwanted changes that you or someone else previously made in ESP.
See Send/Receive Options on page 2-13 for more detail.
Saving
1. Send, No Save.
(same as <A>)
This saves ESP set-
tings to current
memory.
2. Send and Save.
(same as <Z>)
This activates all changes in current memory and saves to the reader for
power-on.
3. Send and Save as Customer Defaults.
(same as <Zc>)
(This option will be visible only if you had checked Enable Send and Save as
Customer Defaults in ESP Preferences in the Options menu.)
Use this to save your own set of default settings that you can quickly retrieve with
a <Zrc> command.
For more on defaulting and saving settings, see Defaulting/Saving/Resetting on
page A-20.
Send/Receive Options
Defaulting
When you select Default Current... or Default all ESP... you are only defaulting the
ESP settings.
Advanced Options
Send Current View
This is the same as Save
to Reader, Send No Save
except that only the com-
mands in the current menu
tree are sent.
Send Current Command
This is the same as Send
Current View above, but
only saves the command
that is currently selected.
Add Exceptions
After you perform a Receive from Reader com-
mand1 and you click on the Add Exception option
you may see a list of serial commands that looks
like the example to the right. These are serial com-
mands that may be in your readers firmware, but
not included or different from your current ESP
software.
You can edit these commands by double-clicking on
them and changing as needed.
When exceptions are present, you can also access
them from an Exceptions button that will appear
on the Applications button bar.
It is important to note that these
commands will be saved to your reader whenever you do a Save to
Reader command or send an <A> or a <Z> command.
Also, if there is a corresponding ESP menu item, the ESP Value column
for that item will be blank following a Receive from Reader com-
mand.
1. From the Send/Recv button or right-clicking from within the tree menus.
Chapter Communications
3
Chapter Contents
Communications by ESP.............................................................................3-2
Communications By Serial Command.........................................................3-2
Password Protection ...................................................................................3-3
RS-232/422 Host Port .................................................................................3-4
RS-232 Auxiliary Port ................................................................................3-11
Preamble ...................................................................................................3-20
Postamble .................................................................................................3-21
LRC Status ................................................................................................3-22
Intercharacter Delay ..................................................................................3-22
The ESP (Easy Setup Program), configuration changes can be made in the ESP
menus, then sent and saved to your reader. The user can also send serial commands to
the reader via the ESPs Terminal window.
This section includes connecting parameters and options for communicating by the aux-
iliary port and various interfaces.
Note: When assigning characters in user-defined fields, the characters NULL <> and ,
can only be entered through embedded menus, not through ESP or serial commands.
Note: You can learn the current setting of any parameter by inserting a question mark
after the number, as in <K100?>. To see all K commands, send <K?>.
Note: Default settings for establishing communications are:
Baud = 9600
Parity = Even
Stop Bits = One
Data Bits = Seven
Flow Control = None
Note: You can learn the current setting of any parameter by inserting a question mark
after the number, as in <K100?> To see all K commands, send <K?>.
Communications by ESP
Communications by ESP
To change a setting,
double-click the set-
ting and use your cur-
sor to scroll through
the options.
Note: You can learn the current setting of any parameter by inserting a question mark
after the number, as in <K100?> To see all K commands, send <K?>.
Chapter 3 Communications
Password Protection
Note: Password protection options are entered serially and are not available in ESP
menus.
Password Status
Definition: If Password Status is set to Disabled, then no parameters are off lim-
its.
If Password Status is set to Enabled, then this option cannot be subse-
quently disabled until the password defined in <K733> is entered. This
parameter is non-volatile and will be included in any <Z> or <Zc>
parameter save.
Serial Cmd: <K732,status>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
New Password
Definition: To change the password from the current password, enter the current
password, a comma, and the new password. If successfully entered, the
new password will be displayed and the command change protection
temporarily removed.
Serial Cmd: <K733 password,new password>
Note: Follow this with a <Z> or <Zc> to save for power-on.
Default: MICRO
Options: Any ASCII string up to 8 characters.
Chapter 3 Communications
Point-to-Point (standard)
Usage: Used only with RS232 or RS422.
Definition: Standard Point-to-Point requires no address and sends data to the host
whenever it is available, without any request or handshake from the host.
Serial Cmd: <K140,0>
Polling Mode D
Usage: When in Polling Mode D, an address of 1 is automatically displayed on
the configuration screen. However, during transmission, a 1C hex poll
address (FS) and a 1D hex select address (GS) are substituted for the 1.
Definition: Like Point-to-Point, Polling Mode D requires a dedicated connection to
the host; but unlike Point-to-Point, it requires an address and must
wait for a poll from the host before sending data.
Serial Cmd: <K140,4>
Chapter 3 Communications
Multidrop
See also Quick Start Help for Multidrop setup.
From Host
Definition: This option allows the handshaking protocol to be initiated from the host,
if not configured in an unpolled mode. Messages sent to the host will
include the scanners defined protocol. The status of From Host deter-
mines if messages sent from the host to the scanner must include the
defined protocol. If From Host is disabled, the defined protocol is not
included. If From Host is enabled, the defined protocol must be
included.
Serial Cmd: <K140,6,RES,address,REQ,EOT,STX,ETX,ACK,NAK,from host>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Chapter 3 Communications
Chapter 3 Communications
Usage: These commands set the communication parameters with the auxiliary
port which can be used to configure menus, send data to the host, dis-
play data transmissions originating from the host of the scanner, and
relay data from other scanners set in tandem (daisy chained). If the
scanners host port needs to be dedicated to the host, but configuration
must be done on the fly, the auxiliary port can be set to accept configura-
tion changes by Command Processing (page 3-17).
Definition: An auxiliary port connects the scanner to a remote display or to other
scanners that can display or transfer data.
Chapter 3 Communications
Transparent Mode
Usage: A common application, in conjunction with handheld scanners, is one that
employs an auxiliary readout to detect mis-applied bar code symbols.
Definition: In Transparent mode data is passed between the auxiliary port and the
host. The scanner buffers data from the auxiliary port and echoes the
keyed data on the auxiliary port.
Data initiated from the Auxiliary Port
Auxiliary port data is passed through to the host whenever a return
key is pressed at the auxiliary port or symbol data is sent.
Whenever aux port data is sent with symbol
data, the aux port data will appear between the Host Aux
preamble and the symbol data. Port
Scanner
Chapter 3 Communications
Scanner
Scanner
Scanner
Chapter 3 Communications
Chapter 3 Communications
Daisy Chain ID
Usage: Used in a daisy chain setup in cases where the host needs to know which
scanner sent the data.
Definition: A one or two character prefix which identifies the particular daisy chain
scanner from which the data is being sent.
Serial Cmd: <K101,aux port mode,baud rate,parity,stop bits,data bits,daisy chain ID
status,daisy chain ID>
Default: 1/
Options: Any one or two ASCII characters.
Preamble
Preamble
Preamble Status
Usage: Useful for identifying and controlling incoming data. For example, defin-
ing the preamble as a carriage return and a line feed causes each
decoded message to be displayed of on its own line.
Definition: Defines a one to four character data string that can be added to the front
of the decoded data.
Serial Cmd: <K141,preamble status,preamble character(s)>
Default: Disabled
Options: 0 = Disabled 1 = Enabled (within any protocol)
Preamble Characters
Serial Cmd: <K141,preamble status,preamble characters>
Default: ^M corresponds to: carriage return/null/null/null.
Options: Up to four user-defined ASCII characters, including control characters.
Chapter 3 Communications
Postamble
Postamble Status
Usage: Useful for identifying and controlling incoming data. For example, defin-
ing the postamble as a carriage return and a line feed causes each
decoded message to be displayed of on its own line.
Definition: Defines a one to four character data string that can be added to the end
of the decoded data.
Serial Cmd: <K142,status,postamble character(s)>
Default: Enabled
Options: 0 = Disabled 1 = Enabled (within any protocol)
Postamble Characters
Serial Cmd: <K142, status,postamble characters>
Default: ^M^J. Corresponds to carriage return/line feed/null/null, as displayed in
the menu.
Options: Up to four user-defined ASCII characters, including control characters.
LRC Status
LRC Status
(Longitudinal Redundancy Check)
Intercharacter Delay
Usage: Used only when a host cannot receive data quickly enough and there is
enough time between symbols to allow data to be completely transferred.
It is rarely used since any setting other than zero will slow down commu-
nications. For example, a 200 setting will result in a 1/5 second delay
between each character that is transmitted.
Definition: The time interval in milliseconds between individual characters transmit-
ted from the scanner to the host.
Serial Cmd: <K144,intercharacter delay>
Default: 0
Options: 0 to 255 (in milliseconds). Zero (0) causes no delay between
characters.
After youve established communications and completed basic read rate testing, you will
need to address the spatial and timing parameters associated with your application. In a
typical operation a bar coded item moves along a line past a scanner. A trigger or timer
activates a read cycle during which the scanner actively searches for symbols. You will
need to decide how to initiate the read cycle and how and when to end it.
Note: When assigning characters in user-defined fields, the characters NULL <> and , can
only be entered through embedded menus, not through ESP or serial commands.
Note: You can learn the current setting of any parameter by inserting a question mark after
the number, as in <K100?>. To see all K commands, send <K?>.
To change a setting,
double-click the
setting and use your
cursor to scroll
through the options.
Multisymbol
Usage: Commonly used in shipping applications where a shipping symbol contains
individual symbols for part number, quantity, etc. This feature allows one
trigger to pick up all the symbols. AIAG and UCC/EAN-128 are two application
standards that address this need.
Definition: Allows the user to set up conditions that will allow more than one symbol to
be decoded and output.
Conditions The following conditions apply:
1. All noread messages are posted at the end of the data string.
2. If more than one symbol is within the scan beam at the same time,
symbol data may not be displayed in the order of appearance.
3. If Matchcode Type is set to Sequential or if Trigger is set to Continuous
Read 1 Output, Number of Symbols will default to 1 (if set to any number
greater than 1).
4. The maximum number of characters in any one bar code (other than
PDF417) is 64.
5. The maximum number of characters in a single scan line is 102 (Code
39).
6. The maximum number of characters for all symbols is 392, including
preamble, separators, and LRC.
Number of Symbols
Definition: Number of Symbols allows the user to define up to 6 symbols that can be
read in a single read cycle.
Serial Cmd: <K222,number of symbols,multisymbol separator>
Default: 1
Options 1 to 6
Multisymbol Separator
Usage: Used to delimit or separate data fields with a user defined character.
Definition: The character thats inserted between each symbol scanned when Multi-
symbol is set to any number greater than 1.
ESP: To select a new multisymbol separator, double-click on Separator and
select a character in the popup window.
Serial Cmd: <K222,number of symbols,multisymbol separator>
Note: If Multisymbol Separator has been changed to any character other
than the default comma and you wish to re-define the separator as a
comma, use ESP or the embedded menu.
Default: , (comma)
Options: Any available ASCII character, except < > NUL.
Trigger
Trigger Mode
Definition: The type of trigger event that will initiate the read cycle.
Serial Cmd: <K200,trigger mode,trigger filter duration>
Default: Continuous Read
Options: 0 = Continuous Read 3 = External Edge
1 = Continuous Read 1 Output 4 = Serial Data
2 = External Level 5 = Serial Data & External Edge
Continuous Read
Usage: Continuous Read is useful in testing symbol readability or scanner func-
tions. It is not recommended for normal operations.
Definition: In Continuous Read, trigger input options are disabled, the scanner is
always in the read cycle, and it will attempt to decode and output every
scan crossing a symbol.When To Output and Noread options have no
affect on Continuous Read.
Serial Cmd: <K200,0>
Positive
1 Initiate Read Cycle:
Object # 1, moving in front of the
detector beam, causes a change in
0
the trigger state, initiating the read
Object cycle. Associated waveforms assume
detector External Trigger State is set
End Read Cycle: to Positive.
The same object, moving out
1 of the detector beam, causes Positive
another change in the trigger
Object state, ending the read cycle.
detector 0
Trigger Level
Definition: External Level allows the read cycle to begin when a trigger (change of
state) from an external sensing device is received. The read cycle endures
until the object moves out of the sensor range and the active trigger state
changes again.
Serial Cmd: <K200,2>
Important: Level and Edge apply to the active logic state (Positive or Negative) that
exists while the object is in a read cycle, between the rising edge and falling edge. Rising
edge is the trigger signal associated with the appearance of an object. Falling edge is the
trigger signal associated with the subsequent disappearance of the object.
Trigger Edge
Usage: Users can select the trigger polarity that will operate with their systems.
Definition: Determines whether a positive or negative transition will initiate the read
cycle.
Serial Cmd: <K202,external trigger state>
Default: Positive
Options: 0 = Negative 1 = Positive
Serial Trigger
Usage: Allows the user to define the trigger character that initiates the read cycle.
Definition: An on-line host command composed of a single ASCII host serial trigger
character that initiates or ends the read cycle.
Trigger characters are of two types: delimited and non-delimited. A delimited
character is entered within angle brackets delimiters < >.
A non-delimited trigger character is one that either starts or ends the read
cycle and is NOT enclosed by delimiters such as < >.
Definition: The read cycle is the time during which the scanner will attempt to read and
decode a symbol. At the end of a read cycle the scanner stops reading
symbols and sends the symbol data or noread message to the host.
Timeout
Usage: Typically used with Serial or Edge and Continuous One.
It is effective in highly controlled applications when the maximum length of
time between objects can be predicted. It assures that a read cycle ends
before the next bar coded object appears, giving the system extra time to
decode and transmit the data to the host.
Definition: Timeout ends the read cycle when the time set in Read Cycle Timeout
elapses or times out.
If in Continuous Read 1 Output, a timeout initiates a new read cycle and
allows the same symbol to be read again.
With either External Edge, Serial Data, or Serial Data & Edge enabled, a
timeout ends the read cycle and symbol data or a noread message is sent
to the host.
With External Level enabled, the read cycle does not end until the falling
edge trigger or a timeout occurs. The next read cycle does not begin until
the next rising edge trigger.
Scanner Setup
Scanner Setup includes Gain, Scan Speed, Laser Power, and AGC.
Gain Level
Usage: Can be useful for fine-tuning gain or when portions of analog signals spike,
as in the case of specular reflection or extraneous noise.
Definition: Sets a voltage amplitude value for the analog circuitry.
Serial Cmd: <K504,gain level,AGC sampling,AGC min,AGC max>
Default: 90 (high density scanner) or 110 (low density scanner).
Note: Gain Level is optimized at the factory before shipment; the default
value may not be 90.
Options: 40 to 255
Scan Speed
Usage: To ensure a minimum number of scans, faster scan speeds are typically
used for fast moving symbols and/or longer symbols with larger bar widths.
See Formulas for Number of Decodes on page A-27.
Note: Scan beams move faster across symbols further out in the scan
range since the moving beam is being projected from a spinning mirror.
Definition: Allows the user to set the number of scans per second by controlling the
spinning mirror motor speed.
Serial Cmd: <K500,scan speed>
Default: 50 (x 10) Low density scanner
35 (x 10) High density scanner
Options: 300 to 1000
Laser Power
Usage: Used to optimize read performance at different distances. Typically higher
power is used at greater distances.
Definition: Allows the user or calibration routine to select laser power setting.
Serial Cmd: <K700,laser on/off status,laser framing status,laser on position,laser off
position,laser power>
Default: High Power (Low density scanner)
Medium Power (High density scanner)
Options: 0 = Low 1 = Medium 2 = High
AGC Minimum
(Read Only)
AGC Maximum
(Read Only)
Maximum Element
Usage: Caution: Do not change this parameter unless instructed by a your techni-
cal representative.
Definition: Maximum Element represents the maximum bar element width, as mea-
sured in micro seconds, thats allowed before a video reset occurs (default
to white).
Serial Cmd: <K502,maximum element>
Default: 1500 (45.75S)
Options: 1 to 65535 (.035S to 2294S increments)
Laser Setup
Laser Setup includes Laser On/Off and Laser Framing.
Note: See also Auto Frame on page 9-6 for more details on laser framing.
Laser On Position
Definition: Percentage of the full scan arc that the
scan beam moves through before the
laser turns ON. For example, if set for
30, the laser will be off during the first
30 percent of the scan.
When Laser Framing is disabled, the
laser is ON for 100 percent of the full scan.
When Laser Framing is enabled, the
laser turns ON at the specified percentage
of the scan.
Laser On Position
Laser Power
Definition: This parameter allows the user to select a laser power setting as follows:
Serial Cmd: <K700,laser on/off status,laser framing status,laser on position,laser off
position,laser power>
Default: High
Options: 0 = Low (~0.6mW)
1 = Medium (~1.0mW)
2 = High (~1.5mW)
Chapter Symbologies
5
Chapter Contents
Symbology by Serial Command ..................................................................5-2
Code 39 .......................................................................................................5-3
Code 128 .....................................................................................................5-6
Interleaved 2 of 5.......................................................................................5-13
Codabar.....................................................................................................5-17
Code 93 .....................................................................................................5-24
Narrow Margins .........................................................................................5-27
Symbology ID ............................................................................................5-28
Background Color......................................................................................5-29
Autodiscriminate ........................................................................................5-29
This section describes the various symbol types that can be read and decoded by the
reader.
Code 39 is enabled by default. To enable all codes, send a <P> serial command to the
scanner from the Terminal window.1
See http://www.aimusa.org/standards/aimpubs.htm for additional information
about symbologies.
Note: When assigning characters in user-defined fields, the characters NULL <> and ,
can only be entered through embedded menus, not through ESP or serial commands.
1. If using an I 2/5 symbol, verify that the number of characters in the symbol being scanned
matches the code length enabled for the I 2/5 symbol type (default is 10 and 6).
Symbologies by ESP
To open nested
options, single-click
the +.
To change a setting,
double-click the setting
and use your cursor to
scroll through the
options.
Code 39
Usage: Code 39 is considered the standard for non-retail bar code applications.
Definition: An alphanumeric symbol with unique start/stop symbol patterns, composed
of 9 black and white elements per character, of which 3 are wide.
Code 39 Status
Serial Cmd: <K470,status,check digit status,check digit output,large intercharacter
gap,fixed symbol length status,symbol length,full ASCII set>
Default: Enabled
Note: This is the only symbol type enabled by default.
Options: 0 = Disabled 1 = Enabled
Code 128
Usage: Code 128 is useful in applications with tight spots and high security needs.
Definition: A very dense alphanumeric symbology. It encodes all 128 ASCII charac-
ters; it is continuous, has variable length, and uses multiple element widths
measured edge to edge.
EAN-128 Status
Note: Code 128 must be Enabled for EAN-128 to function.
Usage: Used as a standard for shipping symbols.
Definition: A subset of Code 128, with extended features. (For more information, see
GS1 US at:
http://www.gs1us.org/gs1us.html
Or see AIM, at:
http://www.aimglobal.org/standards/aimpubs.asp.
Interleaved 2 of 5
Usage: It is has been popular because it is the most dense symbol for printing
numeric characters less than 10 characters in length; however we do not
recommend this symbology for any new applications because of inherent
problems such as truncation.
Definition: A dense, continuous, self-checking, numeric symbology. Characters are
paired together so that each character has five elements, two wide and
three narrow, representing numbers 0 through 9, with the bars representing
the first character and the interleaved spaces representing the second char-
acter. (A check digit is highly recommended.)
Important: You must set Symbol Length in order to decode I 2/5 symbols.
Interleaved 2 of 5 Status
Serial Cmd: <K472,status,check digit status,check digit output status,symbol length
#1,symbol length #2,unused,range mode>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Codabar
Usage: Used in photo-finishing and library applications. Formerly used in some
medical applications but not typically used in newer applications.
Definition: Codabar is a 16-character set (0 through 9, and the characters $, :, /, ., +,
and ) with start/stop codes and at least two distinctly different bar widths.
Codabar Status
Serial Cmd: <K471,status,start & stop match status,start & stop output,large interchar-
acter gap,fixed symbol length status,symbol length,check digit type,check
digit output status>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
UPC/EAN
Usage: Used primarily in POS application in the retail industry. It is commonly used
in applications in combination with Matchcode when there is a need to ver-
ify that the right product is being placed in the right packaging.
Definition: UPC (Universal Product Code) is a fixed length numeric, continuous sym-
bology. UPC can have two- or five-digit supplemental symbol data following
the normal code. The U.P.C., Version A (U.P.C., A) symbol is used to
encode a 12 digit number. The first digit is the number system character,
the next five are the manufacturer number, the next five are the product
number, and the last digit is the checksum character.
When enabled, the scanner will read UPC version A and UPC version E
only.
UPC Status
Serial Cmd: <K473,UPC status,EAN status,supplementals status,separator status,sep-
arator character,unused,UPC-E output as UPC-A>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
EAN Status
Usage: EAN is the European version of the UPC symbology and is used in Euro-
pean market applications.
Definition: EAN is a subset of UPC. When enabled, the scanner will read UPC version
A, UPC version E, EAN 13, and EAN 8. It also appends a leading zero to
UPC version A symbol information and transmits 13 digits. If transmitting 13
digits when reading UPC version A symbols is not desired, disable EAN.
Note: The extra character identifies the country of origin.
Serial Cmd: <K473,UPC status,EAN status,supplementals status,separator status,sep-
arator character,unused,UPC-E output as UPC-A>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Disabled
UPC Supplementals will not be decoded.
Enabled
When enabled, the scanner will try to decode a main and a supplemental. However, if a supplemen-
tal is not decoded, at the end of the read cycle, the main will be sent by itself. The UPC main and
supplemental symbols are considered to be one symbol and will be treated as such.
Required
When set to Required and either the main or supplemental symbol is not read, a single
noread condition results. The UPC main and supplemental symbols are treated as one
symbol.
For example, if Supplementals is set to Required, Separator is enabled, and an asterisk
is defined as the UPC separator character, then the data will be displayed as: MAIN *
SUPPLEMENTAL.
Note: Under no circumstances will supplemental symbol data be sent without a main sym-
bol.
Note: If additional symbolsother than the main or supplementalwill be read in the
same read cycle, Number of symbols should be set accordingly.
Code 93
Usage: Used occasionally in clinical industry.
Definition: Code 93 is a variable length, continuous symbology employing four element
widths. Each Code 93 character has nine modules that may be either black
or white. Each character contains three bars and three spaces.
Code 93 Status
Serial Cmd: <K475,status,fixed symbol length status,fixed symbol length>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Pharmacode
Usage: Pharmacode is used mostly with packaging for the pharmaceuticals industry.
Definition: The symbol encodes up to five different numbers, each with its own color
which may be entered in decimal or binary format with a 1 represented by a
thick bar and a 0 represented by a thin bar. Bar width is independent of
height.
In decimal format, each part can be up to 999999.
In binary format, each input have up to 19 ones and zeros.
Note: It is recommended that you disable AGC before enabling Pharmacode.
Pharmacode Status
Serial Cmd: <K477,status, fixed bar count status,fixed bar count, min. no. of bars,bar
widths,direction,fixed threshold value>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Direction (Pharmacode)
Definition: Specifies the direction that a bar can be read.
Serial Cmd: <K477,status,fixed bar count status,fixed bar count,min. no. of bars,
bar widths,direction,fixed threshold value>
Default: Forward
Options: 0 = Forward 1 = Reverse
Narrow Margins
Usage: Used when the leading and trailing edges of the symbols are smaller than
the standard margin or when other objects encroach into the margins.
Definition: Allows the scanner to read symbols with quiet zones less than 8 times the
width of the narrow bar element. Quiet zone is the space at the leading
and trailing ends of a symbol. Each quiet zone can be as narrow as only
five times the width of the narrow bar element when Narrow Margins is
enabled.
Serial Cmd: <K450,narrow margins status,symbology id status>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Note: Do not use Narrow Margins with Large Intercharacter Gap enabled in Code 39 or
Codabar.
Symbology ID
Usage: Used when the symbology type and how its decoded needs to be known.
Definition: Symbology ID is an AIM standard prefix set of characters that identify the
symbol type.
When enabled, the scanner analyzes and identifies the symbology and
adds a three character identifying prefix to the data:
1. ] (close bracket character) indicating the presence of a symbology ID
2. A, C, E, F, G, I, L, p, M
A = Code 39; C = Code 128 or EAN-128; E = UPC/EAN; F = Codabar;
G = Code 93; I = Interleaved 2 of 5; L = PDF417 (special option); p =
Pharmacode; M = MSI Code
3. Modifier (see the explanation of modifiers below)
Serial Cmd: <K450,narrow margins,symbology identifier status>
Serial Cmd: <K450,0,symbology identifier status>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Explanation of Modifiers for Code 39, Codabar, and I2/5
For Code 39, Codabar, and Interleaved 2 of 5, the modifier indicates Check Digit and
Check Digit Output Status.
For Code 39 only, Full ASCII needs to be enabled to see modifiers 4, 5, and 7.
Check Digit Full ASCII conversion performed
Modifier Check Digit
Output (Code 39 only)
0 Disabled N/A No
1 Enabled Enabled No
3 Enabled Disabled No
4 Disabled N/A Yes
5 Enabled Enabled Yes
7 Enabled Disabled Yes
Example: ]A5 indicates a Code 39 symbol with Check Digit and Check Digit Output Status
enabled and Full ASCII conversion performed.
For Other Symbologies
For Code 128, a 1 indicates EAN-128; otherwise the modifier is a 0.
For all other codes, the modifier is 0.
Background Color
Usage: If the background is darker than the symbol, then enable black background.
Typically the background is white; but on PCBs for example, they can be
black.
Definition: Allows the user to choose which symbol background (spaces and surround-
ing medium) the scanner can read.
Serial Cmd: <K451,background color>
Default: White
Options: 0 = White 1 = Black
Autodiscriminate
Although Autodiscriminate is not a configuration command, but it is included here as a
convenient tool for enabling most symbol types.
Usage: Commonly used for quick setup mode to detect symbol type. This is partic-
ularly useful for users who might be unfamiliar with their applications sym-
bology.
Note: It does not alter individual fields such as Start/Stop, Fixed Length,
etc. These need to be setup individually.
Definition: Enables all available symbology types except PDF417, UCC/EAN 128. The
user may also individually disable/enable each symbology type.
Serial Cmd: <P>
Default: Code 39 (only)
Options: <P> Enables all codes except noted above.
<Q> Enable Code 39 only
<R> Enable Codabar only
<S> Enable I 2/5 only
Note: If using an I 2/5 symbol, verify that the number of characters in the symbol being
scanned matches the symbol length enabled for the I 2/5 symbol type (default is 10 and 6).
Code 39
Serial Cmd: <K452,Code 39,Codabar,Interleaved 2 of 5,Code 93>
Default: Standard
Options: 0 = Tight
1 = Standard
2 = Aggressive
Tight
Maximum ratio is set to 3.5:1. A Validate function, which compares the ratio between the
maximum and minimum bars in the symbol, is used for additional security. A Validate Bars
function minimizes the acceptance of skewed symbols.
Standard
Maximum ratio is set to 4.0:1. A Validate function, which compares the ratio between the
maximum and minimum bars in the symbol, is used for additional security.
Aggressive
Maximum ratio is set at 5.6:1. A Validate function is not used in this mode. This method
may be useful on long symbols where the spot velocity can change throughout the symbol
and effect the minimum and maximum bar values.
Codabar
Serial Cmd: <K452,Code 39,Codabar,Interleaved 2 of 5,Code 93>
Default: Standard
Options: 0 = Tight
1 = Standard
2 = Aggressive
Tight
Maximum ratio is set to 3.5:1. A Validate function, which compares the ratio between the
maximum and minimum bars in the symbol, is used for additional security. A Validate Bars
function minimizes the acceptance of skewed symbols.
Standard
Maximum ratio is set to 4.0:1. A Validate function, which compares the ratio between the
maximum and minimum bars in the symbol, is used for additional security.
Aggressive
Maximum ratio is set at 5.9:1. A Validate function is not used in this mode. This method
may be useful on long symbols where the spot velocity can change throughout the symbol
and effect the minimum and maximum bar values.
Interleaved 2 of 5
Serial Cmd: <K452,Code 39,Codabar,Interleaved 2 of 5,Code 93>
Default: Standard
Options: 0 = Tight
1 = Standard
2 = Aggressive
Tight
Maximum ratio is set to 3.5:1. A Validate function, which compares the ratio between the
maximum and minimum bars in the symbol, is used for additional security. A Validate Bars
function minimizes the acceptance of skewed symbols.
Standard
Maximum ratio is set to 4.0:1. A Validate function, which compares the ratio between the
maximum and minimum bars in the symbol, is used for additional security.
Aggressive
Maximum ratio is set at 6.2:1. A Validate function is not used in this mode. This method
may be useful on long symbols where the spot velocity can change throughout the symbol
and effect the minimum and maximum bar values.
Code 93
Serial Cmd: <K452,Code 39,Codabar,Interleaved 2 of 5,Code 93>
Default: Standard
Options: 0 = Tight
1 = Standard
2 = Aggressive
Tight
Maximum ratio is set to 2.95:1. A Validate function, which compares the ratio between the
maximum and minimum bars in the symbol, is used for additional security. A Validate Bars
function minimizes the acceptance of skewed symbols.
Standard
Maximum ratio is set to 2.8:1. A Validate function, which compares the ratio between the
maximum and minimum bars in the symbol, is used for additional security.
Aggressive
Maximum ratio is set at 2.65:1. A Validate function is not used in this mode. This method
may be useful on long symbols where the spot velocity can change throughout the symbol
and effect the minimum and maximum bar values.
This section includes instructions for setting up conditions for beeper and EZ Button
controls and for changing input/output of data messages and electrical transitions for
control of internal and external devices.
Note: The characters NULL <> and , can only be entered through embedded menus,
not through ESP or serial commands.
Note: You can learn the current setting of any parameter by inserting a question mark
after the number, as in <K100?>. To see all K commands, send <K?>.
To change a setting,
double-click the set-
ting and use your cur-
sor to scroll through
the options.
Disabled
Usage: Useful when an application only needs to use the discrete outputs and
can allow the scanner to do the decision-making. When Disabled, the
host does not need the symbol data and the communication lines are
used only for setup and status checks.
Definition: When set to Disabled, the scanner will not transmit any data that is gen-
erated during a read cycle (symbols, noreads, etc.).
Match
Usage: Match is used in an application that requires specific symbol information
and needs to sort, route or verify based on matching the specific symbol
data.
Definition: When set to Match, the scanner transmits symbol data whenever a sym-
bol matches a master symbol. However, if Matchcode Type is Disabled,
it transmits on any good read.
Note: A noread can still be transmitted if Enabled.
Good Read
Usage: Good Read is used when an application requires all symbol data to be
transmitted. Its typically used in tracking applications in which each
object is uniquely identified.
Definition: With Good Read enabled, the scanner transmits symbol data on any
good read regardless of Matchcode Type setting.
Note: A noread can still be transmitted if enabled.
As Soon As Possible
Usage: As Soon As Possible is useful in applications in which symbol data
needs to be moved quickly to the host, typically when the host is making
decisions based on symbol data.
Definition: Enabling As Soon As Possible causes symbol data to be sent to the host
immediately after a symbol has been successfully decoded.
Note: More than one decode might in fact be required to qualify as a
good decode, depending on how Decodes Before Output is set.
Usage: End of Read Cycle is useful in timing-based systems in which the host is
not ready to accept data at the time it is decoded.
Definition: Enabling End of Read Cycle means that symbol data does not get sent
to the host until the read cycle ends with a timeout or new trigger.
Message Output
The following flow chart explains the logic paths taken for Noread, Bad Symbol and
No Symbol outputs.
Notes:
Start of When Symbol Detect is enabled, the scanner will test for a
Read Cycle minimum number of bar and space transitions as set in
Transition Counter Threshold. If that minimum thresh-
old is met, a Bad Symbol message is sent; otherwise a No
Scans until end Symbol message is sent.
of read cycle
If conditions for a Bad Symbol, No Symbol, or No Object
messages are met but the corresponding output is not
enabled, a Noread message will be output.
Does the
scanner decode the Yes
symbol?
No
Noread
Is Symbol
No message sent
Detect enabled?
(if enabled)
Yes
Is
Transition Bad Symbol
Counter Threshold Yes message sent
met? (if enabled)
No
No Symbol
message sent
(if enabled)
End of
read cycle
Noread Message
Noread Message
Usage: Used in applications where the host needs serial verification that a sym-
bol has not been read and especially useful in new print verification.
Definition: When enabled, and if no symbol has been decoded before timeout or the
end of the read cycle, the noread message will be transmitted to the
host.
Noread Status
Serial Cmd: <K714,status,noread message>
Default: Enabled
Options: 0 = Disabled 1 = Enabled
Noread Message
Definition: Any combination of ASCII characters can be defined as the noread
message.
Serial Cmd: <K714,status,message>
Default: NOREAD
Options: 1 to 7 ASCII characters.
Note: Noread Message will only be transmitted if Symbol Output (Symbol Data
Output Status on page 6-4) is set to Match, Mismatch or Good Read.
Noread Message can be set to any ASCII characters except NULL <> and , (comma).
Usage: Useful in verifying the presence of a symbol that has not been decoded.
Definition: When enabled, a message is sent to the host when a symbol is detected
but not decoded. See Transition Counter Threshold on page 6-8.
The Bad Symbol output is tied to the transition counter. If during a read
cycle no symbol is decoded and the required setting for the Transition
Sample Threshold is met, a user defined message will be sent to the
host. See Output Message Flow on page 9-8.
No Symbol Message
No Symbol Message
Note: No Symbol Message does not apply to multisymbol operations.
No Symbol Status
Serial Cmd: <K716,no symbol status,no symbol message>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
No Symbol Message
Serial Cmd: <K716,no symbol status,no symbol message>
Default: NO_SYMBOL
Options: Up to 10 ASCII characters (except NUL)
Beeper
Usage: Can be used as an audible verification that either a good read or a noread
has occurred.
Definition: A beep is emitted either after each good read of a symbol or after each
noread, according to the Beeper Output Condition selection.
Note: Beeper will also sound if any of the following occur:
the scanner is defaulted a send/save command from ESP or an Exit
command from any embedded menu.
a <Z>, <Zp>, <Zd>, or <K,1> command is sent
Partial Output
Partial Output
Definition: When enabled, allows you to pre-select portions of symbols to be trans-
mitted by the scanner.
Usage: In Multisymbol mode, partial output is performed on each separate
symbol. For example, if Start Position is set to 3 and Partial Length is
set to 5, the following symbols are transmitted as follows:
1234567890 as 34567
1234 as 34
123456789,abcde as 34567,cde (two symbols with separator)
12 as [blank]
Status
Serial Cmd: <K703,partial output status,start position,length>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Serial Verification
Usage: These commands are useful in removing any doubt about the scanners
acceptance and response to host configuration commands.
For example, if the current preamble is SOM and <Kd1, START> is
entered, the scanner will echo back <Kd1, SOM> since the attempted
entry START exceeds the four character limit for that command. There-
fore it is rejected and the existing SOM message is echoed back and
remains the preamble message.
Definition: Allows the user to verify configuration command status.
Function: If a command with multiple fields is processed, some of the fields may
have been processed properly while others were not. The changes will
appear in the string echoed back so that the user will know which fields
did or did not change.
Serial Cmd: <K701,serial command echo status,serial command beep status,con-
trol/hex output>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Serial Verification
Control/Hex Output
Usage: Useful for viewing settings with binary characters when using serial com-
mand on a terminal.
Definition: Determines the response to a Serial Command Echo or status request
command.
When set to Control, two characters are transmitted to represent a non-
displayable character. For example, a carriage return will be shown as the
two characters: ^M.
When set to Hex, the output is the hex character.
Serial Cmd: <K701,serial command echo status,serial command beep status,con-
trol/hex output>
Default: Control
Options: 0 = Control 1 = Hex
EZ Button
The EZ Button provides a convenient way to perform programmable operations both
globally and for each of the three programmable button positions.
The EZ Button has three positions, each of which can be user-programmed to perform
a variety of functions.
Global Status
Definition: Determines the basic function of the EZ Button.
Serial Cmd: <K770,global status,default on power-on>
Default: Enabled
Options: 0 = Disabled 1 = Enabled 2 = Trigger
3 = Unlatch Output
Disabled: When selected, the EZ Button is disabled.
Enabled: When the EZ Button is enabled, the function of each button position can be
determined by the EZ Button Mode command.
Trigger: When selected, the EZ Button will act as a trigger for the scanner. All other
button operations will not be active. The trigger will operate the same way as an exter-
nal trigger.
Trigger mode Operation
The read cycle endures for as long as the EZ Button is pressed,
External Level unless a timeout occurs and Timeout is enabled for End Of
Read Cycle.
As with Level, Edge allows a read cycle to be initiated by press-
External Edge ing the EZ Button, but unlike Level mode, the read cycle ends
with a good read output, a timeout, or a new trigger.
Unlatch Output. When selected, will unlatch any outputs that have been latched via
Latch Mode 1 (Unlatch on New Master Change) on page 6-20.
Default on Power-on
Definition: When enabled, if the EZ Button is held down during power-on, the scan-
ner will default to customer defaults and save for power-on. This is the
same as sending a <Zrc> command.
Serial Cmd: <K770,global status,default on power-on>
Default: Enabled
Options: 0 = Disabled 1 = Enabled
EZ Button Modes
Definition: Provides six user-selectable functions for each of the three EZ Button positions.
Serial Cmd: <K771,position 1 mode,position 2 mode,position 3 mode>
Defaults: As indicated in the table below.
Options: As indicated in the table below.
The positions are selected by the length of time the button is held. When you want to
select position one, you press and holds the button until one beep is emitted from the
scanner and the 20% LED is turned on. You then release the button to select position one.
If you want to select position two, you continue to hold until you hear two short beeps and
the 40% LED is turned on. Finally, to select position three, you continue to hold the button
until you hear three short beeps and the 60% LED is turned on.
For example, if you want to select the function for position three, you will hear one short
beep, then two short beeps, and finally three short beeps and the 20%, 40%, and 60%
LEDs turn on in progression.
The following table lists the options for the three positions with the default position in bold text:
EZ Button Options
Single Beep Two Beeps Three Beeps
0 = Disabled 0 = Disabled 0 = Disabled
1 = Read Rate 1 = Read Rate 1 = Read Rate
2 = Autocalibration 2 = Autocalibration 2 = Autocalibration
3 = Save for Power-on 3 = Save for Power-on 3 = Save for Power-On
4 = Auto Framing 4 = Auto Framing 4 = Auto Framing
5 = Load New Master 5 = Load New Master 5 = Load New Master
6 = Sleep Mode 6 = Sleep Mode 6 = Sleep Mode
Disabled: When set to Disabled, the button position will have no function associated with
it. The scanner will still respond when the button position is reached.
Read Rate: Read rate will perform decodes/second and is the same as sending a <C>
from the terminal. To exit read rate mode, press the EZ Button once quickly.
Calibration: The calibration routine will be initiated. To abort calibration, press the EZ Button
once quickly.
Save for Power-on: When when this button position is selected, all scanner settings will
be saved to non-volatile memory to be recalled when scanner is powered-on the next
time. This is the same as sending the <Z> in the terminal.
Output 1
Usage: This option provides switching to host software to control external devices
such as PLCs and relays. It is useful for routing and sorting and to prevent
mis-packaging and mis-routing.
Definition: Sets the discrete output functions for specific user-selected conditions.
Output On
Definition: Allows the user to set the conditions under which an output (or outputs)
will be activated.
Serial Cmd: <K810,output on,polarity,pulse width,output mode>
Default: Mismatch or Noread
Options: 0 = Mismatch or Noread
4 = Trend Analysis
1 = Match
5 = Validation
2 = Mismatch
6 = Diagnostic Warning
3 = Noread
Note: If Output On is set to any mode containing Match or Mismatch, the transition
(switching) will only occur if Matchcode Type is enabled and Master Symbol(s) are
loaded into memory.
Mismatch or Noread
Activates discrete output when the symbol data does not match that of the master symbol
or the symbol has not been decoded before the end of the read cycle.
Match
Activates a discrete output when the symbol data matches the master symbol.
Mismatch
Activates a discrete output whenever the symbol data does not match that of the master
symbol.
Noread
Activates a discrete output whenever the symbol data is not decoded before the end of the
read cycle.
Validation
(Not yet implemented)
Diagnostic Warning
The output remains active as long as one of the diagnostic conditions is met.
Polarity
Definition: Sets the polarity of the discrete output.
When set to Negative polarity, and an output condition from the scanner
is met, the output-1 pin on the scanners 15 pin connector go low for the
time specified in Pulse Width. (See Table A-3, MS-3 Laser Connec-
tor, 15-pin Socket, on page A-4.)
Serial Cmd: <K810,output on,polarity,pulse width,output mode>
Default: Negative
Options: 0 = Negative 1 = Positive
Pulse Width
Definition: Sets the time in 1 mS increments that the discrete output remains active.
Serial Cmd: <K810,output on,polarity,pulse width,output mode>
Default: 50 (50 mS)
Options: 0 to 65535 (0 to 65.535 seconds). Divide the number entered on the
command line by 1000 for time in seconds.
Output 1
Output Mode
Definition: Sets the condition in which the discrete output is de-activated.
Serial Cmd: <K810,output on,polarity,pulse width,output mode>
Default: Pulse
Options: 0 = Pulse 2 = Latch Mode 2
1 = Latch Mode 1 3 = Latch Mode 3
Pulse
Definition: This is the default mode of operation in which the programmable output
is activated when the Output On condition has been met and held active
for the duration of the selected pulse width.
Mismatch
Definition: Output will be activated when the number of Mismatches equals the
value entered for Number to Output On within the trigger window
selected in Trigger Evaluation Period.
Noread
Definition: The output will be activated when the number of noreads equals the
value entered for Number to Output On within the trigger window
selected in Trigger Evaluation Period.
Output 1
Definition: Sets the threshold for decodes required per trigger (read cycle) that
must occur before an output can be activated.
Example: Trend analysis mode = Decodes per Trigger
Number to Output On = 4
Number of Triggers = 25
Decodes per Trigger Threshold = 10
In this example, the scanner will activate an output whenever the num-
ber of decodes falls below the decodes per trigger threshold (10) for 4
trigger (read cycle) events.
Serial Cmd: <K780,trend analysis mode,number of triggers,number to output
on,decodes/trigger threshold>
Default: 0
Options: 0 to 65535
Diagnostics (output 1)
Applies warning to Output 1.
Note: When Diagnostics is enabled (Output On on page 6-18), the output mode
configuration has no effect. The output will remain active as long as one of the diagnos-
tic conditions is met. The output will go inactive as soon as it detects no active diagnos-
tic warning.
Service Threshold
Definition: Activates the output whenever the service threshold has been met. Will
output upon every occurrence of the threshold.
Serial Cmd: <K790,usused,service threshold,unused,laser current high,laser cur-
rent low>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Output 2
Output 2
Serial Cmd: <K811,output on,polarity,pulse width,output mode>
Output 3
Serial Cmd: <K812,output on,polarity,pulse width,output mode>
Quality Output
Quality Output
Usage: Host system can monitor scanning performance (printing process, scan-
ning environment changes, etc.) by setting thresholds in their processing
or displaying the outputs for visual verification.
Definition: Allows the host to track the number of good reads per trigger by append-
ing a five digit numeric field to the end of symbol data.
When enabled, symbol data is followed by a separator, followed by the
five digits in the form of 00000. For example, a certain symbol that has
been averaging 00100 reads per read cycle suddenly drops to 00012, this
would be a good indication that a problem has arisen even though good
reads are still occurring.
Reads/Trigger Status
Serial Cmd: <K704,quality output separator,reads/trigger status,decode direction
status>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Chapter Matchcode
7
Chapter Contents
This section explains the matchcode output functions and the master symbol database
setup.
Note: When assigning characters in user-defined fields, the characters NULL <> and ,
can only be entered through embedded menus, not through ESP or serial commands.
Note: You can learn the current setting of any parameter by inserting a question mark
after the number, as in <K100?>. To see all K commands, send <K?>.
Matchcode by ESP
Matchcode by ESP
To change a setting,
double-click the set-
ting and use your cur-
sor to scroll through
the options.
Chapter 7 Matchcode
Overview of Matchcode
Usage: Matchcode is used in applications to sort, route, or verify data based on
matching the specific symbol in a variety of ways as defined in this sec-
tion. For example, a manufacturer might sort a product based on dates
that are embedded in the symbol.
Definition: With Matchcode you can store master symbol data in the scanners
memory, compare this data against the scanned symbols, and define how
symbol data and/or discrete signal output will be directed. A master sym-
bol database can be set up for up to 10 master symbols. See Master
Database on page 11-8.
Matchcode Type
Matchcode Type
Note: When Matchcode is set to Enabled, Wild Card or Sequential, Number of
Symbols will default back to 1 (if set to any number greater than 1).
Definition: Allows the user to choose the way that master symbols will be compared
with subsequently scanned symbols.
Note: First set Triggering Mode to External or Serial.
Serial Cmd: <K223,matchcode type,sequential matching,match start position,
match length,wild card character,sequence on noread,sequence on mis-
match>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
2 = Wild Card 3 = Sequential
Disabled: Has no effect on operations.
Enabled: Instructs the scanner to compare symbols or portions of symbols with the
master symbol.
Sequential Instructs the scanner to sequence after each match (numeric only) and
Matching: compare symbols or portions of symbols for sequential numbers.
Note: If Matchcode Type is set to Sequential, Number of Symbols
will default back to 1 (if set to any number greater than 1).
Wild Card: Allows the user to enter user defined wild card characters in the master
symbol.
Sequential Matching
Usage: Useful in tracking product serial numbers that increment or decrement
sequentially.
Definition: With Sequential enabled, Sequential Matching determines if a count is
in ascending (incremental) or descending (decremental) order.
Serial Cmd: <K223, matchcode type,sequential matching,match start position,
match length,wild card character,sequence on noread,sequence on mis-
match>
Default: Increment
Options: 0 = Increment 1 = Decrement
Chapter 7 Matchcode
Matchcode Type
Match Length
Usage: For example, if Match Length is set to 6 in a 10 character symbol, and
Match Start Position is set for 2, only the 2nd through 7th characters
(from left to right) will be compared.
Definition: Defines the length of the character string that will be compared with that
of the master symbol when Match Start Position is set to 1 or greater.
When Match Start Position is set to 0, no comparison will occur.
Serial Cmd: <K223,matchcode type,sequential matching,match start position,
match length,wild card character,sequence on noread,sequence on mis-
match>
Default: 1
Options: 1 to 64
When more than one master symbol is enabled, the maximum number for each addi-
tional symbol diminishes as shown here:
# of Symbols: 1 2 3 4 5 6 7 8 9 10
Maximum
64 64 42 32 25 21 18 16 14 12
characters:
Chapter 7 Matchcode
Sequence On Noread
Usage: Sequence On Noread is useful when the scanner needs to stay in
sequence even if no decode occurs.
Definition: When Sequence On Noread is Enabled and Matchcode is set to
Sequential, the scanner sequences the master symbol on every match
or noread. When disabled, it does not sequence on a noread.
Serial Cmd: <K223,matchcode type,sequential matching,match start position,
match length,wild card character,sequence on noread,sequence on
mismatch>
Default: Enabled
Options: 0 = Disabled 1 = Enabled
As an example of Sequence on Noread Enabled, consider the following series of
decodes:
Master sym-
Decoded symbol Master symbol after decode
bol
001 001 002
002 002 003
003 noread 004 (sequenced on noread)
004 004 005
005 noread 006 (sequenced on noread)
006 noread 007 (sequenced on noread)
007 007 008
As an example of Sequence on Noread Disabled, consider the following series of
decodes:
Master sym-
Decoded symbol Master symbol after decode
bol
001 001 002
002 002 003
003 noread 003 (not sequenced)
003 003 004
004 noread 004 (not sequenced)
004 noread 004 (not sequenced)
004 004 005
Matchcode Type
Sequence On Mismatch
Note: Matchcode must be set to Sequential for this command to function.
Usage: Enable this parameter if every trigger event should have a decode and
more than one consecutive mismatch may occur.
Disable this parameter if every trigger event should have a decode but no
more than one consecutive mismatch may occur.
Definition: When set to Enabled, the master symbol sequences on every decode,
match or mismatch.
When set to Disabled, the master symbol will not sequence whenever
consecutive mismatches occur.
Serial Cmd: <K223,matchcode type,sequential matching,match start position,
match length,wild card character,sequence on noread,sequence on
mismatch>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
The scanner will sequence the master to one more/less than the decoded symbol. As an
example of Sequence On Mismatch Enabled, consider the following decodes:
Master sym-
Decoded symbol Master symbol after decode
bol
001 001 002
002 002 003
003 abc 004 (sequenced on mismatch)
004 004 005
005 def 006 (sequenced on mismatch)
006 ghi 007 (sequenced on mismatch)
007 007 008
As an example of Sequence On Mismatch Disabled, consider the following decodes:
Master sym- Decoded sym-
Master symbol after decode
bol bol
001 001 002
002 002 003
003 abc 004 (sequenced because of previous match)
004 004 005
005 def 006 (sequenced because of previous match)
006 ghi 006 (not sequenced because of previous mismatch)
006 006 007
Chapter 7 Matchcode
After New Master Pin has been enabled and the pin activated, decodes will be saved
in the master database beginning with master symbol #1. If the scanner is configured
for a multisymbol read cycle (Number of Symbols is greater than 1), the remaining
decodes will be saved in each consecutive master symbol location. For example, if
Number of Symbols is set to 3 and New Master Pin is then activated, at the end of
the next read cycle, the decoded symbols will be saved as master symbols 1, 2, and 3.
Usage: Useful where more than one master symbol is required, as in a Multi-
symbol setup, for matching and other matchcode operations.
Definition: Allows you to define up to 10 master symbols as the master symbol data-
base, which can be entered by keyboard, scanned in, displayed, or
deleted by serial or ESP commands.
Chapter Diagnostics
88
Chapter Contents
This section describes warning and operating messages and their settings.
Important: Since Warning Messages and NOVRAM Messages are not enabled by
default, these messages will not be displayed. However, if you suspect that erroneous
defaults are occurring, enable Warning Messages and NOVRAM Messages by send-
ing <K400,1,,,1>.
Note: When assigning characters in user-defined fields, the characters NULL <> and ,
can only be entered through embedded menus, not through ESP or serial commands.
Note: You can learn the current setting of any parameter by inserting a question mark
after the number, as in <K100?>. To see all K commands, send <K?>.
To change a setting,
double-click the set-
ting and use your cur-
sor to scroll through
the options.
Chapter 8 Diagnostics
Counts
Counts
(These commands are READ ONLY.)
Sending <K406?> returns a four fields of data, as follows:
<K406, power-on,resets,power-on saves,customer default saves>
Power-on Count
Usage: Useful for detecting unwanted resets caused by power supply problems
or ESD transients.
Definition Returns the number of times the scanner has been re-powered.
Serial Cmd: Send <K406?>
Returns < (read only)
Read Only 0 to 65,535 powerups, 0 to 65,535 resets.
Ranges:
Resets
Usage: Useful for detecting unwanted resets caused by power supply problems
or ESD transients.
Definition Returns Resets for all the warm resets, including <A>, <Ard>, <Arp>
and <Arc>.
Serial Cmd: Send <K406?>
Returns <K406, power-on,resets,power-on saves,custom default
saves>
Read Only 0 to 65,535 powerups, 0 to 65,535 resets.
Ranges:
Chapter 8 Diagnostics
Power-on Saves
Usage: Useful for detecting unwanted resets caused by power supply problems
or ESD transients.
Definition Returns the number of times the scanner saves for the power-on resets,
including <Z>, <Zc>, <Zrd>.
Serial Cmd: Send <K406?>
Returns <K406, power-on,resets,power-on saves,custom default
saves>
Read Only 0 to 65,535 powerups, 0 to 65,535 resets.
Ranges:
Usage: Useful as a troubleshooting tool that can help pinpoint the cause of a
reset.
Definition: Records the number of hours and minutes of operation since the last sys-
tem reset.
Serial Cmd: Send <K407?>
Returns <K407hours,minutes>
Read Only 0 to 23 hours, 0 to 59 minutes.
Ranges:
Chapter 8 Diagnostics
Laser High
When enabled, a message is transmitted whenever the laser current exceeds a factory-
calibrated reference value which cannot be altered. The message repeats once every 30
minutes until the condition is corrected.
Laser Low
Laser Low
When enabled, a message is transmitted whenever the laser current falls below a fac-
tory-calibrated reference value which cannot be altered. The message repeats once
every 30 minutes until the condition is corrected.
Chapter 8 Diagnostics
Service Message
Definition: The service timer is reset on power-on, thus the timer only records the
time that has elapsed since the last reset.
The message is sent every time the timer expires.
Service Message
Serial Cmd: <K409,status,service message,threshold,resolution>
Default: SERVICE
Options: Any 1 to 10 ASCII characters except NULL < , or >.
Threshold
Definition: Sets the number of hours or minutes that will transpire before the Ser-
vice message is output.
Serial Cmd: <K409,status,service message,threshold,resolution>
Default: 300 (5 minutes)
Options: 1 to 65,535
Resolution
Definition: Records time in seconds or minutes.
Serial Cmd: <K409,status,service message,threshold,resolution>
Default: Seconds
Options: 0 = Seconds 1 = Minutes
Service Message
Chapter Calibration
9
Chapter Contents
Calibration ...................................................................................................9-2
Auto Frame..................................................................................................9-6
The purpose of calibration is to optimize certain scanner settings to obtain the highest
read rates in one or several label positions as required by the users application. This
section includes the overall steps in calibration and specific steps for calibration by ESP,
serial command, and embedded menus. For most applications, calibration is the only
optical setup required. This section also includes an explanation of Auto Frame com-
mands, the routine for constraining the width of the scan beam.
Calibration
Calibration
When a Calibrate command is sent, all symbologies are enabled except PDF4171 and
Pharmacode.2 The scanner looks for readable bar code symbols, and records the high-
est read rates it encounters while stepping through various settings for motor speed,
laser power, fixed gain, and AGC level. If the The software evaluates the results and
selects the best combination.
Calibration Methods
There are three ways to start calibration:
ESP Calibration menu (page 9-3).
Serial command <@CAL> (page 9-3).
From the embedded calibration menu <@> (page 9-4).
1. Calibration does not function with PDF417 symbols. However, you can substitute a Code 128
symbol with the same mil size as your PDF417. When you do this, enable Code 128 and after
calibration, switch back to PDF417.
2. Note: If using an I 2 of 5 label, verify that the number of characters in the label being scanned
matches the code length enabled for the I 2 of 5 code type (default is 10 and 6). See Inter-
leaved 2 of 5 on page 5-13.
Chapter 9 Calibration
Calibration By ESP
1. In the Apps Mode, click the Calibration button on the icon bar
at the top of ESP to open the calibration menu.
2. In the Fixed Parameter Calibration area you can select one or
two parameters that you do NOT want to change. (Selecting all
three will produce an error.)
For example, you want to fix Scan Speed and Laser Power.
Check the appropriate boxes and make your settings, as shown here:
Start the calibration procedure by click- Click on any one or two Fixed boxes
ing either Calibrate button. if you want those parameters to not
change during calibration.
Use the Test button to
start a read rate test.
See Auto
Frame on
page 9-6.
Uncheck
this box to
disable
Framing.
Click Save to save for Autodiscriminate is used with Test to enable most symbol types. You do
power-on. not have to click this when doing calibration since it is part of the routine.
3. Wait for the calibration to cycle through the various settings looking for the best
combination. A Calibration Successful message should eventually appear.
4. Click Save to save for power-on (same as a <Z> command).
Calibration
Chapter 9 Calibration
4. Exit the Calibration Setup menu.
To exit the Calibration Setup menu, press the ESC key. The following prompt
will be displayed:
DO YOU WANT TO USE CALIBRATION SETTINGS? (NOT SAVED) Y/N.
Typing Y will cause the current settings in the Calibration Setup menu will cause
the scanner to be reset without saving the settings for power-on.
You will then be asked,
DO YOU WANT TO SAVE THESE SETTINGS FOR POWER-ON? Y/N
These settings include Motor Speed, Gain, AGC Level, Laser Power and Laser
Frame.
Typing N will cause the scanner to be reset without saving the settings for power-
on.
Auto Frame
Auto Frame
Auto Frame by ESP
In the Calibration menu you can also use the Auto Frame command to constrain the
width of the scan beam. This is useful if extraneous markings interfere or slow down
decoding. It works by setting the length of time the laser is ON during each scan so that
only a selected portion of the scan width is effectively scanned.
1. In the Apps Mode, click the Calibration button on the icon bar at
the top of ESP to open the calibration menu.
In the Auto Frame portion of the display, note that in the default
mode is Framing Enabled the Off position is set to 95 and the On
position is set to 10.
2. To adjust the scan width automatically, click on the Auto Frame button.
The scanner will locate the start and stop characters and adjust the scan beam
timing so that the scan width is constrained but still covers the entire symbol.
Figure 9-9 Auto Frame Defaults Figure 9-10 Constrained Scan Beam Wid
You can also adjust the framing manually by moving the Off and On selectors while
visually observing the size of the scan beam.
Chapter 9 Calibration
Auto Frame
Terminal Window.......................................................................................10-2
Find Function.............................................................................................10-3
Macros.......................................................................................................10-4
Terminal Window Functions ......................................................................10-5
Note: The characters NULL <> and , can only be entered through embedded menus,
not through ESP or serial commands.
Note: You can learn the current setting of any parameter by inserting a question mark
after the number, as in <K100?>. To see all K commands, send <K?>.
Terminal Window
Terminal Window
Type in text
Copy, paste, here to find
and save in terminal
window
Clear
screen
Terminal Macro
screen List on
this bar
The terminal screen allows you to enter serial commands (in red) from the macro box,
by copying, or directly from your keyboard.
The terminal screen also displays bar code data or information from the reader (in
blue).
You can also right click in the terminal screen to bring up a handy option box.
Find Function
The Find box allows you to enter text strings to be searched for in the terminal window.
For example a series of bar codes have been scanned into the terminal view and you
want to determine if a particular bar code starting with ABC has been read.
1. Type ABC into the Find box.
2. Press Enter.
The first instance of ABC will be high-lighted in the terminal window.
3. Press the F3 key to search again for the next instance of ABC.
4. Press Shift-F3 to search for the previous instance of ABC.
Macros
Macros
Macros can be stored in a macro selection bar, edited in a separate window, and exe-
cuted by clicking on the macro name.
Click on Next Click on the first arrow here to Click here to edit a
Row to see Add Macro or Default Macros. macro.
the next row When you default macros, the
of macros entire macros set is restored to
their original macro commands.
When you click on the macro name, the macro is executed in the terminal window.
If this is a command, the command is also sent to the reader at the same time it is dis-
played.
Editing a Macro
When you click the arrow next to a any macro and select Edit, the following appears:
You can edit an existing macro or type in a new macro name and define it in the Macro
Value text box. Click OK.
Chapter Utilities
11
Chapter Contents
Utilities by ESP Menu ................................................................................11-2
Summary of Utilities Commands ...............................................................11-3
Read Rate .................................................................................................11-5
Counters ....................................................................................................11-6
Master Database .......................................................................................11-8
Digital Bar Code ......................................................................................11-11
Firmware .................................................................................................11-12
Device Control .........................................................................................11-13
Symbol Type ...........................................................................................11-15
Defaulting/Saving/Resetting ....................................................................11-16
Microscan Grading ..................................................................................11-17
Reader Status Requests .........................................................................11-19
Utility commands are generally commands that are performed during reader operations
to check read rates, determine read rates or perform miscellaneous operations on
reader hardware. Serial utility commands are not prefaced with a K and a numeric
code, nor do they require an initialization command (<A> and <Z>). They can be
entered from within any terminal program or from within ESP in the Terminal window.
Note: You can learn the current setting of any parameter by inserting a question mark
after the number, as in <K100?>. To see all K commands, send <K?>.
Note: The characters NULL <> and , can only be entered through embedded menus,
not through ESP or serial commands.
Chapter 11 Utilities
Chapter 11 Utilities
Read Rate
When in a read rate mode, the scanner will output a number preceding the symbol
data. This number will represent either the number of decoded scans per second or the
percentage of successful decoder per total number of scans.
Make a note of the decode rate. It will be a useful reference when calculating the num-
ber of decodes (see Formulas for Number of Decodes on page A-46).
Counters
Counters
Counter commands can be a numeric value from 00000 to 65,535. After reaching the
maximum numeric limit of 65,535, the counter will automatically roll-over and start
counting again at 00000. To obtain the cumulative total of counts after the roll-over has
occurred, add 65,536 per each roll-over (the scanner does not keep track of the num-
ber of roll-overs) to the current count.
Note: All counter values will be lost if power is recycled to the scanner or the scanner
receives a reset or save command.
Counter by ESP
You can access Counters from the Utilities menu.
Right-click the appropriate counter option and select Request to display count or
Clear to set counter to zero. Or, right-click on Counters and select Request All.
Chapter 11 Utilities
Trigger Counter
Sending <T> displays the total number of triggers since the last reset.
Mismatch Counter
With Matchcode enabled, sending <X> displays the number of decoded symbols since
the last reset that do not match the master symbol.
Master Database
Master Database
Important: Master Symbol Database is used only for comparing entire bar codes,
when Sequential and Wild Card are NOT enabled, and Start Position is equal to 0.
Usage: Useful where more than one master symbol is required, as in a Multi-
symbol setup, for matching and other matchcode operations.
Definition: Allows you to define up to 10 master symbols as the master symbol data-
base, which can be entered by keyboard, scanned in, displayed, or
deleted by serial or ESP commands.
To access in ESP
1. Click the Utilities button and click the Master Database tab.
2. Click the Matchcode Type checkbox to enable Matchcode.
3. Set the number of master symbols you want to include.
Chapter 11 Utilities
Master Database
Table 11-3 Maximum Characters for Master Symbol Database
Master Symbol Number Maximum Characters
#1 127
#2 62
#3 39
#4 28
#5 20
#6 15
Note: See Chapter 7, Master Symbol Database for information on master symbol
database and related commands.
Definition: After youve set the size in the database, you can order the scanner to
scan-in the next symbol as master symbol 1.
Serial Cmd: <Gmaster symbol number>
To store the next symbol scanned as master symbol #1 send: <G> or
<G1>. For all symbols numbers except 1, the number must be included.
Options: 1 to 10
Chapter 11 Utilities
Firmware
Firmware
Firmware Update is used to download application code to the reader.
Application code versions are specific to your reader. Consult with your sales represen-
tative before downloading application code. If needed, an application code will be sent
to you in the form of a *.mot file.
To download application code:
1. First make sure the host is connected to your reader.
2. Apply power to the reader.
3. In the Firmware Update pulldown window, select App Code.
This will open a file dialog box.
4. Navigate to the appropriate file (a *.mot file) and open the file.
5. Allow a minute or so for the firmware to download.
As application code begins to download to the reader, the reader will go silent, the
readers RDY and GD/RD LEDs will flash intermittently, and a progress indicator at the
bottom of the ESP window will let you know when the download is complete.
Caution: Do not interrupt power or disconnect the host cable while download is in
progress.
Firmware/Checksum Verification
From Firmware Verification you can request the part number and checksum from by
selecting App Code or Boot Code and clicking the accompanying Request... button.
Chapter 11 Utilities
Device Control
Outputs
You can access Device Control from the Utilities menu.
Output # 1 Pulse
Definition: Activates the link between Output 1 (+) and Output 1 () of the host con-
nector for the duration set by Pulse Width on page 6-19. (regardless
of Master Symbol or Output 1 status).
ESP: Click Output # 1 Pulse to activate.
Serial Cmd: <L1>
Output # 2 Pulse
Definition: Activates the link between Output 2 (+) and Output 2 () of the host con-
nector for the duration set by Pulse Width on page 6-19. (regardless
of Master Symbol or Output 2 status).
ESP: Click Output # 2 Pulse to activate.
Serial Cmd: <L2>
Output # 3 Pulse
Definition: Activates the link between Output 3 (+) and Output 3 () of the host con-
nector for the duration set by Pulse Width on page 6-19. (regardless
of Master Symbol or Output 3 status).
ESP: Click Output # 3 Pulse to activate.
Serial Cmd: <L3>
Device Control
Extras
You can access Device Control from the Utilities menu
Disable Reader
Usage: This feature is useful during extended periods of time when no symbols
are being scanned or the reader is being configured.
Definition: Sending <I> will turn the laser off and end the current read cycle.
ESP: Click Disable Reader button.
Serial Cmd: <I>
Disabling laser scanning will not affect any downloaded commands to the reader.
Enable Reader
Definition: Causes the laser scanner to be on continuously.
ESP: Click Enable Reader button.
Serial Cmd: <H>
Note: Enable/Disable Reader does not relate to Laser On/Off command.
Send Motor On
Usage: This feature is useful during extended periods of time when no symbols
are being scanned or the reader is being configured.
Definition: Turns the spinning mirror on (if not already running).
ESP: Click Send Motor On button.
Serial Cmd: <KE>
Chapter 11 Utilities
Symbol Type
Sending <P> enables the scanner to decode all available symbol types except Pharma-
code and PDF417 without changing scanner configuration.
Sending <Q> enables the scanner to decode Code 39 symbols without changing scan-
ner configuration.
Sending <R> enables the scanner to decode Codabar symbols without changing scan-
ner configuration.
Sending <S> enables the scanner to decode I 2 of 5 symbols without changing scanner
configuration.
Defaulting/Saving/Resetting
Defaulting/Saving/Resetting
Chapter 11 Utilities
Microscan Grading
Microscan grading informs the user how the scanner perceives a symbol it is
scanning. Except for special circumstances, the slower the scan speed and the closer
the symbol is to the focal point, the more accurate the results.
The grading feature is available for the following symbologies: Code 39, Codabar, Code
128, I-2/5, and Code 93.
Command Format
<GRADE,xx.xx> where xx.xx is the symbol length in inches with up to two digits in
front of the decimal point and two digits behind the decimal point.
Example: Send <GRADE,01.50> for a symbol 1.5 inches in length.
Operation
Once the command has been sent, the scanner will enter a user test mode for up to two full
seconds. During this time the scanner will attempt to complete ten decodes of the same-
symbol. If this occurs, the scanner will output the grading information for this symbol.
Microscan Grading
Decodability - Calculates the decodability of the symbol using the video acquired
by the reader. The calculations are consistent with ANSI X3.182-1990 however the
it is important to note the measurements used to acquire the symbol information
are proprietary and are not ANSI compliant.
W/N Ratio - Calculates the wide to narrow ratio of the symbol using the video
acquired by the reader. The calculations are consistent with ANSI X3.182-1990
however the it is important to note the measurements used to acquire the barcode
information are proprietary and are not ANSI compliant.
Average Narrow Bar - Calculates the wide to narrow ratio of the symbol using the
video acquired by the reader. The calculations are consistent with ANSI X3.182-
1990 however the it is important to note the measurements used to acquire the
barcode information are proprietary and are not ANSI compliant.
Average Bar Error - Calculates the average bar error in relation to the spaces and
the average narrow element. This is a good representation of the amount of over or
under printing.
Intercharacter Gap - Measures the largest intercharacter gap within the symbol.
If the symbol does not have intercharacter gaps then N/A is output.
Margin - Measures the size of the smallest margin around the symbol.
Chapter 11 Utilities
These represent: the status byte discussed above, the boot part number (35-338200-
12), the application part number (35-338201-14), the fpga part number (5-559001-
18), flash checksum (476C), current RAM parameter checksum (33), and flash param-
eter checksum (33).
Optics
Semiconductor, visible laser diode
(650nM nominal); CDRH Class II
Connector
3 ft. (914mm) cable terminated with 15-pin D-sub socket
connector
Electrical
5VDC 5%, 200mV p-p max. ripple, 260mA @ 5VDC
(typical)
Discrete I/O
Trigger Input: 3 to 24V rated (1mA @ 5 VDC)
New Master: 3 to 24V rated (1mA @ 5VDC)
Outputs (1, 2, 3): 5V TTL compatible, can sink 10 mA and
source 2 mA.
Symbologies
Standard: Code 39, Code 128, Interleaved 2 of 5,
Codabar, Code 93, UPC/EAN, Code 93, Pharmacode MS-3 Integral Right
Special option: PDF417 Angle Laser Scanner
FCC
The MS-3 conforms to the following requirements:
FCC 15.107:2015 Class B Conducted Emissions
FCC 15.109(g):2015 Class B Radiated Emissions
CE Mark
The MS-3 conforms to the following requirements:
EN 55022:2010/AC:2011 Class B Radiated Emissions
EN 55022:2010/AC:2011 Class B Conducted Emissions
EN 55032:2012 Class B Radiated Emissions
EN 55032:2012 Class B Conducted Emissions
Canada
ICES-003:2012 updated 2014 Class B Radiated and Conducted Emissions
Japan
VCCI:2015-04 Class B Radiated and Radiated and Conducted Emissions
Australia and New Zealand
AS/NZS CISPR 22:2009 + A1:2010 Class B Radiated and Conducted Emissions
AS/NZS CISPR 32:2013 Class B Radiated and Conducted Emissions
Safety Certifications
CDRH, FCC, UL/cUL, CE, BSMI (compliant)
RoHS/WEEE Compliant
2016 Microscan Systems, Inc.
All rights reserved. Specifications subject to change.
Product specifications are given for typical performance at 25C (77F) using grade A labels. Performance
characteristics may vary at high temperatures or other environmental extremes. WarrantyOne year limited
warranty on parts and labor. Extended warranty available.
High Density
Low Density
Pin Host RS-232 Host & Aux. RS-232 Host RS-422/485 In/Out
1 Power +5VDC In
2 TxD TxD TxD (-) Out
3 RxD RxD RxD (-) In
4 Power/Signal Ground
5 NC
6 RTS Aux. TxD TxD (+) Out
7 Output 1 TTL (can sink 10mA and source 2mA) Out
8 Default Configuration (NPN) In
9 Trigger (NPN) In
10 CTS Aux. RxD RxD (+) In
11 Output 3 TTL (can sink 10mA and source 2mA) Out
12 New Master Pin (NPN) In
13 Chassis Ground
14 Output 2 TTL (can sink 10mA and source 2mA) Out
15 NC
Caution:
Appendix F Defaulting/Saving/Resetting
Understanding and controlling your scanners active, saved, and default settings is critical
to the operation of your scanner.
<A> Reset (does not save for power-on)
<Ard> Reset and recall Microscan defaults
<Arp> Reset and recall power-on parameters
<Arc> Reset and recall customer default parameters
<Z> Save current settings for power-on
<Zc> Save current settings as customer default parameters
<Zrd> Recall Microscan default parameters and save for power-on
<Zrc> Recall customer default parameters and save for power-on
Power-on Parameters
Power-on parameters (saved by the <Z> command) are saved to NOVRAM and recalled
and loaded into current parameters when power is cycled or the <Arp> command is
issued.
Software Default
Microscan default parameters can be recalled (loaded into current settings) with <Ard>
command or recalled and saved for power-on with the <Zrd> command.
Hardware Default
If a software default reset is not possible, it may be necessary to reset the scanner by
shorting (connecting) specific pins. This procedure has the same effect as the <Zrd> soft-
ware command.
Important: For this reset to occur, this command must be executed within 60 seconds
after a power-on or a reset.
Caution: Be certain that the correct pins are located. Connecting the wrong pins could
cause serious damage to the unit.
3. Momentarily connect these wires (or pins) and listen for a series of short beeps.
4. Within 3 seconds, connect them again. A longer beep should be heard. If not, repeat
the process.
Default/Reset Definitions
Defaults are original or saved customer settings that can be recalled, either by software
reset <Zrd> or hardware reset.
Also when in multidrop, defaulting might be required if the scanner has been assigned a
polling address and you wish to access the scanners menu.
Defaulting the scanner by <Zrd> resets the configuration parameters to Microscan default
values. Defaulting might be necessary if you have make temporary changes, communica-
tions between the scanner and another device are lost or interrupted, or you are using
incompatible equipment (for example, a terminal that is set at 38.4K baud communicating
with a scanner that is set at 115.2K baud).1
Resets (A commands) affect only the current settings (active memory) and are not
saved for power on or recall.
Active Memory is where the scanners active settings are stored during use. These are
not available on power-on unless saved by a Z command.
1. There are no menu options or host commands for resetting the configuration program.
If the scanner is able to read the start symbol, it will begin to beep and the performance LEDs
will be activated.
3. When the scanner starts to beep, remove the configuration symbol from the scan-
ners field of view.
The beeping will end and the scanner will be in the configuration mode.
During the configuration mode the LEDs will emit a side-to-side pattern to let you know
you are currently ready to configure.
You can create your own start symbol by encoding the following in Code 128:
FNC4 <_CFG>.
This will terminate symbol configuration mode. Changes made in configuration will remain
in effect just as would changes made in the menus or by serial commands. A-16
Object
Detector
Scanner
Object Detector
SH
-------- DR 3 = ND (number of decodes) 1
SS
Scan Line
SH (Symbol Height) (ladder formula only) is a measurement of the
height of individual bars.
SS (Scan Speed) is the distance per second that a symbol moves
through the scan.
DR (Decode Rate) is the number of decodes per second that a given
symbol receives (as derived from a stationary scan test). Ladder
Scan line
Picket Fence
1. The -3 component in the formula is included to allow for AGC acquisition, an incomplete first scan, and an
incomplete last scan. This applies only if the number inside the parentheses equals 4 or more. If the number
equals 3, then only subtract 2 to derive 1 good scan.
2 SS -
SR = --------------------------
( SW SL )
SR = Sweep Rate, the number of passes, up or down, that are described by the arc of the raster
mirror.
SS = Scan Speed, the distance per second that a symbol moves through the scan.
SW = Scan Width, the width across the scan beam, at a given distance from the scanner, in which a
symbol can be read (picket fence formula only).
SL = Symbol Length, the length of the longest printed symbol to be read plus the length of the quiet
zones (picket fence formula only).
Example:
SW = 5
22
SS = 2 inches per second SR = ---------------- = 1
(5 1)
SL = 1
Important: The maximum sweep rate that your can set the scanner to is 80. If your solu-
tion is greater than 80, you must change another parameter such as scan speed, etc. and
recalculate.
Since SR can only be entered into the software as whole numbers, round off fractions of
SR to the next higher number, for example enter 0.4 as 1 and 3.5 as 4.
In setting up the scanner, raster height depends on the raster arc and the symbol distance
(range) from the scanner. Both the top offset and bottom offsets can be adjusted indepen-
dently in 255 increments and the resulting arc can vary from 0 to 30.
1. The number 2 in the SR formula ensures that each symbol receives two full raster sweeps.
Scan Line
Scan Speed
Scan speed is a function of motor speed and is adjustable. A slower scan speed may allow
greater symbol range and/or higher decode rates, but at the cost of fewer scans per symbol.
Applies to both picket fence and ladder oriented symbols. If your application allows it, slowing
scan speed (the time in seconds that a symbol is fully within the scan width of the scanner) is
an effective way to increase the number of decodes.
Range
Adjusting the symbols range, if possible, is one of the quickest and most effective ways to
improve decode rates. However, in some applications you may need to select a less than
optimum range, or one that is beyond the fringes of the ranges.
Scan Width
Increasing scan width will increase the number of scans in a picket fence oriented applica-
tion. Scan width is linked with scan range and changing one will usually require a change
in the other.
Do Not:
Aim the scanner into direct light or sunlight.
Aim the scanner into an external object detector or other light-emitting device.
Obstruct the scanner window with mounting hardware or other objects.
Connect chassis of scanner and host to different ground potentials.
Operate the scanner in excessive temperature environments.
RS232
RS232 defines an interface between two devices, such as the scanner and host. It differs
from the other interfaces by dedicating individual pins to specific functions and by requiring
both devices to share a common ground line. Since both device chassis are connected to
a common ground, a ground loop potential and the possibility of noise interference exists;
therefore, cable lengths are limited to a maximum of 50 feet (19.7m). Despite being the
most limited, this interface is used frequently because of the large installed base of RS232
equipment.
RS422
RS422, unlike RS232, measures signals deferentially; that is, the receiver looks at the
potentials between the two receive (or transmit) wires rather than the potential between
signal and ground. As a result, cables, if shielded, can be up to 4000 feet (1219m) in
length. Like RS232, RS422 communication is designed for only two devices on a single
line and must have a common ground. It can be used wherever RS232 is used.
RS485
RS485, like RS422, can transmit up to 4000 feet (1219 m) using differential voltages, but
unlike RS422, its transmitters are turned off until a request for data is received from the
host. RS485 is used exclusively in multidrop protocol.
USB
The Universal Serial Bus can connect up to 127 devices to a host at a maximum of 6
megabits per second (Mbps) of bandwidth.
Scanner 02
Scanner 03
Polling Sequence
Data that is transmitted to the host (symbol data, noread messages, counters, etc.) via
concentrators is solicited by poll requests from the host.
The polling sequence example is by poll address 1E (ASCII hex value for Scanner 02)
and a REQ (request). The scanner responds by first transmitting its own address, 1E,
followed by a STX (start of text) character, and then the data. Next it transmits an ETX
(end of text) character and an LRC (longitudinal redundancy check) character.
If the concentrator (or controller) receives the data from the scanner and is able to validate it
with an LRC calculation, it responds with an ACK (acknowledgment). If the scanner in turn
receives the ACK, the scanner ends this exchange with a RES (reset).
Polling Sequence
Polling Reset
If the scanner has no information, it responds to a poll request by transmitting a RES
(reset).
If the scanner receives a NAK instead of the ACK after transmitting its data string, it will
re-attempt to send the data string up to three times. If the scanner still does not receive
an ACK, it will send a RES (reset) and discard the data in its buffers.
If the scanner transmits data to the concentrator and the concentrator responds with an
ACK or NAK, but the scanner doesnt receive the concentrators response, the scanner
will timeout and send a REQ to the concentrator and request another response. If after
three retries (the number of times it transmits a REQ to the concentrator) the scanner
receives no response, it ends the transmission with a RES (reset).
Select Sequence
Unlike poll requests, select commands always originate from the host and consist of serial
configuration or operation commands to devices that are configured in Multidrop. The
scanner complies with the command when it is polled during the cycle.
Polling Sequence
A RES (reset) is the first command in the select sequence. The 1F hex is the select
address associated with Scanner 02 address. It is followed by a REQ (request). The scan-
ner responds with its own select address, 1F hex, and an ACK (acknowledge). The con-
centrator then transmits an STX (start of text), the data (in this case a <T>), an ETX (end
of text), and an LRC character.
The scanner replies by transmitting its own address, followed by an ACK, acknowledging
receipt of the command. Upon receipt of an ACK, the concentrator concludes the success-
ful exchange with a RES.
In the example above, the scanner only acknowledges a trigger counter request from the
concentrator. It does not respond to the trigger counter request until a subsequent poll. For
example, if the scanners trigger count was 12 at the time the trigger counter request was
received, on a subsequent poll it would send 02T/00012. (The 02 at the beginning of the
string is the scanners address.)
Select Reset
If the scanner receives bad data from the concentrator, it transmits a SEL (its select
address) and a NAK to the concentrator. The concentrator re-transmits the data up to
three times. The concentrator will end the sequence with a RES (reset) if no ACK is
received.
Multidrop Addresses
Poll Select Poll Select
Multidrop Character Character Multidrop Character Character
Address Address
ASCII HEX ASCII HEX ASCII HEX ASCII HEX
01 ^\ 1C ^] 1D 26 N 4E O 4F
02 ^^ 1E ^- 1F 27 P 50 Q 51
03 SP 20 ! 21 28 R 52 S 53
04 " 22 # 23 29 T 54 U 55
05 $ 24 % 25 30 V 56 W 57
06 & 26 ' 27 31 X 58 Y 59
Appendix M Troubleshooting
Symptom Cause Solution
Power supply is lower than required voltage. Use a shorter cable or provide required input
voltage.
Motor won't spin up and Default pin held low at power-up. Boot mode
scan. enabled. Bad cable. Fix or replace cable.
Missing or faulty Application Code; Boot Enter boot mode with power-up and default
mode enabled. pin. Download new Application Code.
LEDs are flashing but Boot mode enabled. Verify Application Code. Release default line
MS-3 is not scanning. and re-power unit.
The symbology being used in the application Enable the symbology type being used in the
is not enabled in the scanner. application.
MS-3 is not decoding.
Length field does not match actual length. Disable fixed length option or change length
field to match actual symbol length.
High frequency ambient lighting. Shade bar code read area, especially the
scan path between the unit and the bar code.
Try aggressive ratios. (Tight ratios decode
high-quality symbols. Standard ratios
Poor bar code printing, ratio errors, or erroneous decode most symbols. Aggressive ratios
bar widths. are loosened to allow decoding of poor quality
symbols. See Symbol Ratio Mode for more
information.) Improve bar code print quality.
Symbol is not in the focal area of laser. Re-position symbol or scanner for optimal
focus position.
Try aggressive ratios. (Tight ratios decode
high-quality symbols. Standard ratios
decode most symbols. Aggressive ratios
Poor decode performance. Excessive curvature of symbol. are loosened to allow decoding of poor quality
symbols. See Symbol Ratio Mode for more
information.)
Gain value not optimal. Try autocalibration.
Other objects in field of view are misleading Try to eliminate shiny objects from the field of
the gain control system. view.
Scan rate is too fast, exceeding MS-3s video Slow down the scan rate.
bandwidth.
Do not violate the 10x rule for quiet zone;
Quiet zone not large enough. enable narrow quiet zone setting.
Clean window with non-abrasive cleaner and
Contaminants on window. wiping cloth; refer to the User Manual for
cleaning instructions.
A preceding bright object is increasing recovery Remove shiny objects that may enter the
Reads well with static scanners field of view before the symbol
bar codes, but once bar time for gain control. enters the field of view. Try fixed gain.
codes are in motion
there are many No Remove shiny objects that may enter the
Reads. Specular reflection preventing AGC stability. scanners field of view before the symbol
enters the field of view. Try fixed gain.
Try aggressive ratios. (Tight ratios decode
high-quality symbols. Standard ratios
Different segments of decode most symbols. Aggressive ratios
bar code decode 100%; Poor printing, ratio errors, or erroneous bar are loosened to allow decoding of poor quality
others decode 0%. widths. symbols. See Symbol Ratio Mode for more
information.) Correct printing variations and
improve bar code print quality.
MS-3 has defaulted settings. No longer set to Connect using default settings instead of
Cannot communicate user communication settings. user settings.
with MS-3. Unit is in boot mode and communication settings Set communication parameters to 57.6K, 8, 1,
are not the same as application mode. None. Exit boot mode.
81% - 100% <Cp> or <C> the % of scans to decodes is presented as a percentage and shown by the LEDs.
In ESP Mode.
No power.
0% <Cp> or <C> the % of scans to decodes is presented as a percentage and shown by the LEDs.
1% - 20% <Cp> or <C> the % of scans to decodes is presented as a percentage and shown by the LEDs.
21% - 40% <Cp> or <C> the % of scans to decodes is presented as a percentage and shown by the LEDs.
41% - 60% <Cp> or <C> the % of scans to decodes is presented as a percentage and shown by the LEDs.
61% - 80% <Cp> or <C> the % of scans to decodes is presented as a percentage and shown by the LEDs.
Scanner
Pitch
Pitch
axis
Scanner Scan line