Keytek Ce Master System Users
Keytek Ce Master System Users
Immunity Tester
Operations Manual
CEMASTER
Immunity Tester
Operations Manual
MA-95-016-005-00 REV D
Thermo KeyTek
One Lowell Research Center
Lowell, MA 01852-4345
Phone: 978-275-0800
Fax: 978-275-0850
E-mail: [email protected]
Thermo KeyTek 1997-2001 All rights reserved
The copyright protection afforded to Thermo KeyTek includes all forms of media
material and information including, without limitation, the software*, generated icons
and logos etc. and all associated documentation. This operating manual and associated
software contained within may not be duplicated in any way without the express
permission of Thermo KeyTek.
Printable products of the software such as test reports and graphs, and associated files
of the same, are the property of the purchaser and may be subject to their own
copyright restrictions.
*The CEMASTER software described in this manual may be copied by, strictly for the
use of, the purchaser in accordance with the furnished license agreement.
Trademark Credits
CEMASTER is a registered trademark of Thermo KeyTek
IBM is a registered trademark of International Business Machines Corporation. Microsoft
Windows and MS-DOS are registered trademarks of Microsoft Corporation.
Notice
In no event will Thermo KeyTek be liable for any damages, consequential, incidental or
indirect (including loss of business profits, business interruption, loss of business
information, and the like) arising out of the use or inability to use the hardware or
software supplied.
The author and publisher of the operating manual make no warranty of any kind,
expressed or implied, with regard to the instructions and suggestions contained within.
The author and publisher of this operating manual shall not be liable in any event of
incidental or consequential damages in connection with, or arising out of, the furnishing,
undertaking or use while following verbal or written instructions and suggestions in any
interpretation of its contents.
Thermo KeyTek products undergo continuous development and the company reserves
the right to change their specifications without obligation or prior notification.
Thermo KeyTek One Lowell Research Center Lowell, MA 01852-4345 USA Tel: 978-275-0800
The information on IEC requirements -- figures and diagrams, tables and data, and test levels -- are
accurate as of the date of printing of this manual. However, the IEC and EN documents are subject
to review and change. The information in this manual is intended as a guideline only.
Product Safety
The following safety instructions have been included in compliance with safety standard
regulations. Please read them carefully.
Read Instructions -- Read all safety and operating instructions before operating the
instrument.
Retain Instructions -- Retain all safety and operating instructions for future reference.
Heed Warnings - Adhere to all warnings on the instrument and in the operating instructions.
Carts and Stands -- Use the instrument only with a cart or stand that is recommended or
included as part of the %'/CUVGT system by the manufacturer.
Ventilation -- The instrument should be situated so that its location or position does not
interfere with its proper ventilation. Do not install in a cabinet or other situation that may
impede the flow of air through the ventilation openings.
Heat -- The instrument should be situated away from heat sources such as heat registers or
other instruments which produce heat.
Power Sources -- Connect the instrument only to the type of power source described in the
operating instructions or as marked on the instrument.
Power Cord Protection -- Place power supply cords so that they are not likely to be walked
on or pinched by items placed on them or against them. Pay particular attention to cords at
plugs, convenience receptacles, and the point where they enter and exit the instrument. The
cordsets used should be the right-angle cordsets supplied with the unit, to prevent damage to
the cordsets when moving the%'/CUVGT.
CEMASTER System 1
Lifting and Carrying -- When moving or installing a %'/#56'4follow the instructions
given in the Installation section of this manual. Moving a %'/#56'4 requires two persons.
Object and Liquid Entry -- Take care that objects do not fall and that liquids are not spilled
into the enclosure through openings.
Defects and Abnormal Stress -- Whenever it is likely that the normal operation has been
impaired, make the equipment inoperable and secure it against further operation. Normal
operation is likely to be impaired if, for example, the instrument:
Sitting or Climbing -- Do not sit or climb upon the instrument or use it as a step or ladder.
CEMASTER System 2
Surge Testing Guidelines
DO NOT WORK ALONE.
Do not use the equipment in conditions other than reasonable laboratory conditions.
There should be no condensing humidity or water standing on the floor or work surfaces;
there shouldnt be significant dust or other contamination.
Ensure that NO ONE is touching the equipment under test (EUT) during the test or
immediately after the test until AC power to the EUT has been turned off.
Ensure that there is a barrier to act as protection in case the equipment under test
explodes. This may happen due to power-follow1 after a failure. The barrier should be
interlocked to prevent surging and to disconnect all AC if the barrier is removed.
The equipment under test must be surrounded by sufficient insulating material to
withstand twice the surge voltage. Consider distance to the floor or table and walls if air
is the insulating material.
Ensure that the proper supply mains voltages are applied to both Thermo KeyTek
equipment and to the equipment under test, and that the AC branch circuit is capable of
supplying the current.
The ground (protective earth), neutral and phase lines of the AC supply to the equipment
under test supply must be connected properly. Do not defeat the protective earth
connection.
The ground (protective earth), neutral and phase lines of the AC supply to the Thermo
KeyTek equipment must be connected properly. Do not defeat the protective earth
connection.
When surging a powered EUT, the mains supply to the equipment under test must be
capable of handling the potential AC fault current (e.g. do not use a UPS to power the
EUT).
Never surge an AC mains line other than through the EUT output connector and the
%'/#56'4internal filter.
Use only equipment which is designed to be safe for the test being performed.
Do not test in a potentially explosive atmosphere (e.g. gas fumes).
Never use equipment that is operating in a strange manner, or that shows clear indication
of abuse.
If probes are in use, be sure they are differential probes which have no ground connection
to the surged ground or to the equipment under test.
1
Power-follow is a condition where the Surge event causes a low-impedance path which full mains current may then flow
through. One example of this is a gas-tube arrestor, which maintains a high impedance until a surge event switches it to a low-
mpedance state; the low-impedance state is maintained until the mains voltage drops below a critical threshold. Similar effects
can be seen due to electrical arcs or exploded components.
CEMASTER System 3
Safety Concerns During Surge Testing
Surge testing is hazardous. The equipment under test (EUT) can ignite, possibly explosively.
Noxious, toxic and sometimes fatal fumes can be generated by the burning equipment.
Accumulated gases may ignite explosively (i.e., flashover).
In an environment where surge testing takes place, it is absolutely crucial that these minimum
safety precautions be taken:
Surge testing should be performed only by properly trained test personnel who are
experienced in conducting such test, or be observed and supervised by such experienced
personnel. No person subject to heart or neurological conditions should be allowed to
conduct surge tests. Persons with pacemakers should not be allowed in or near the area
where testing is conducted.
All personnel working in the area must be shielded with appropriate eye protection, body
protection and electrical protection. They should not be allowed to work in a direct line
of a possible explosion of the equipment under test.
The test area should be a clear and unobstructed environment dedicated to such tests.
The test area should be equipped with ventilating hoods and blowers to remove gases that
may be caused by exploding or burning components.
The test area should have nonflammable walls and floors plus shielding to contain
exploding parts and flames.
There must be fire extinguishers certified for use in electrical and chemical fires readily
available at the test site. DO NOT USE WATER TO EXTINGUISH AN
ELECTRICAL FIRE.
All flammable materials and debris must be outside the test area, and the area must be
well marked, preferably by physical barriers, to prevent accidental intervention by non-
test personnel while a test is in progress.
CEMASTER System 4
Inappropriate Uses
The unit must be kept within the environmental limits of the operating requirements. This
includes not using the %'/#56'4 in condensing humidity.
The unit must not be operated by wearers of electronic life-support equipment. Personnel
with pace-makers, heart problems, nervous disorders, and similar problems should not be
allowed to operate the %'/#56'4 or be present during ESD testing.
The ESD probe is capable of giving an unpleasant shock -- it is not intended for use as a
shock prod, "cattle prod", or as any device designed to shock people or animals.
CEMASTER System 5
Description
The %'/#56'4 is a single-box test system for those in need of the CE Mark. The
%'/#56'4 performs 6 different immunity tests, in accordance with European Norm
requirements and the controlling, Basic EMC immunity standards of:
The %'/#56'4 is the only combination EMC immunity tester that includes built in dip and
interrupt, internal IEC 1000-4-8 capability, computer-controlled ESD tester and simplified,
predefined test routines. The CEMASTER is also the only EMC immunity tester that customers
configure to meet their own testing needs.
Because of the %'/#56'4 's custom configurability, growing companies are now able to
afford in-house compliance testing and larger manufacturers may purchase multiple units for
various departments. The %'/#56'4 is a perfect solution for test houses seeking to
complement their high-end test systems and offer their customers on-site testing capabilities as
an additional service. As a user-friendly, portable immunity tester, the %'/#56'4 is the ideal
solution for rental companies.
CEMASTER System 6
Installation
Assembly Instructions
%'/#56'4Assembly
The %'/#56'4 requires two people for unpacking. Check the packaging materials and
follow any special unpacking instructions shipped with your unit. The %'/#56'4comes
fully assembled.
Save all shipping materials (shipping boxes, foam pieces, bags, and special instructions) for
possible future shipments. The original shipping materials must be used whenever the
%'/#56'4 is returned for calibration or service.
Carefully check all shipping materials for accessories which may be packaged with the
%'/#56'4.
Computer Assembly
If your system includes a computer: Unpack the computer and accessories. Verify that the
voltage settings of computer and monitor match your mains voltage.
The optional ESD probe connector attaches to the front panel. This connector is polarized
and can only be inserted in one orientation: align the flat of the connector plug with the flat of
the front panel opening. Do not force the connector.
2
Some computers use a low-power RS-232 driver with a reduced drive voltage. If the FiberCom fails to work with a
computer, an external supply should be used [Thermo KeyTek part #02-701-695-00 (USA) and 62-050-415-00
(European)].
CEMASTER System 7
The CEMASTER can accept a Direct Current (DC) power source at the EUT input and
provide it to the EUT output. When connecting to DC it is important to note the assignment
of EUT inlet and outlet connectors.
EUT Mains Inlet, Panel Wiring EUT Mains Outlet, Panel Wiring
The IEC Inlet connector is wired, Regardless of nationality, the outlet is always
L1 left, L2 right, PE center. wired, L2 left, L1 right, PE center.
For DC use L1 and L2 should be used for voltages or voltage and return; the PE connection
should be used for return only.
Interlock connector
The optional interlock allows the user to connect a remote switch which, when open, removes all
AC and high-voltage from the %'/#56'4 front panel outlets, and disables the ESD head.
A spare interlock connector is provided in the accessories kit for the %'/#56'4. To use this
connector, wire your cable from the connector TIP and RING contacts, and provide a simple
switch or contact closure at the other end of the cable.
The interlock operates from an internal, +5V, 20mA, logic level signal. Do not connect the
interlock to a power source.
Use a cable with voltage rating of twice the voltage rating of your EUT.
Route the interlock cable away from the EUT output and the EUT.
DO NOT connect the interlock within the EUT, or to circuitry tied to the EUT.
The CM-CCLC cable clamp includes an interlocked safety cover. By connecting the stereo
phono plug cable between CEMASTER and either end of the cable clamp, the CEMASTER will
CEMASTER System 8
operate only if the cover of the cable clamp is properly seated. The unused interlock jack on the
opposite end of the cable clamp can be used to wire to additional interlock switches.
Other Cables
The %'/#56'4 accessories kit includes safety socket connectors and an interlock connector
which the user may use to fabricate special test cables. When fabricating cables, always use wire
rated for twice the peak, pulse voltage present during testing.
The AC power source for the EUT must have adequate capacity for full power-follow3 in the
event that testing causes an equipment failure. Using a supply without sufficient capacity may
result in a failure being masked.
The problem is due to the coupling network mandated by IEC 1000-4-5; this network
connects the surge generator to the AC mains during the surge pulse, and creates a short pulse
of mains current in PE, tripping the GFI. This problem only occurs during Surge events with
a PE coupling mode selected.
3
Power-follow is a condition where the Surge event causes a low-impedance path which full mains current may then
flow through. One example of this is a gas-tube arrestor, which maintains a high impedance until a surge event
switches it to a low-impedance state; the low-impedance state is maintained until the mains voltage drops below a
critical threshold. Similar effects can be seen due to electrical arcs or exploded components.
CEMASTER System 9
There are two solutions to this problem:
The first solution is to power the EUT from a source that does not have a GFI device
installed. This solution may not be practical in some locations due to building codes or safety
regulations.
The second solution is to install an isolation transformer between the AC mains and
%'/#56'4. The isolation transformer should be rated for -- at least -- 1920VA for
120VAC mains, or 3840VA for 220/240VAC mains. The transformer should be wired as
diagrammed below.
Note
A solid-state regulated line source such as an uninterruptable
power supply (UPS) or a power frequency amplifier will generally
not have the instantaneous capacity or full power-follow required
for Surge and PQF testing, and therefore should not be used as the
AC power source for any equipment being tested.
CEMASTER System 10
Physical Environment
%'/#56'4 systems are intended for operation in a laboratory environment, protected from
excess dust, humidity, and temperature.
Operating limits
Temperature: 15 - 40 C
Humidity: 10 - 85%, non-condensing
Altitude: 10,000 feet max.
Storage limits
Temperature: 0 - 60 C
Humidity: 10 - 90%, non-condensing
Altitude: 10,000 feet max.
Note:
If your %'/#56'4 system has been subjected to temperatures or humidity outside of the
normal operating limits for a period exceeding four hours, place the %'/#56'4 in an
environment of the proper temperature and humidity, and allow the %'/#56'4 to stabilize in
that environment for a period of 24 hours; during stabilization, the unit should be unpacked, but
not powered nor operating.
The %'/#56'4 CM-ESD option may require as much as 48 hours stabilization time if the
period outside the humidity limits exceeds four hours.
CEMASTER System 11
Test Area Considerations and Site Preparation
Conduct pulsed EMI testing under sound laboratory conditions. Verify the following:
Allow at least six inches clearance from the wall or other equipment on both sides of the
instrument for proper air flow. The system must be mounted on a level surface.
For EFT testing the %'/#56'4 should be mounted on the floor, on the ground reference
plane. For other testing, the %'/#56'4 may be placed on a table that can support the
weight of the unit. It may also be placed on a sturdy cart. However, since the unit is heavy,
make certain that the instrument and cart combination is stable, and cannot be easily tipped -
particularly if the wheels strike a bump on the floor such as the edge of a rug, a pebble or a
small piece of hardware.
CEMASTER System 12
Special Considerations for EFT Testing
Note
While the possibility of component destruction is less of a
concern during EFT testing, it should not be ignored.
EFT testing involves the use of very high frequency pulses. For this reason the testing standards
specify the use of a ground reference plane in the area where EFT testing is performed. In
addition, the standards specify the minimum and maximum spacings among the various
equipment present in the test area.
The floor of the test area is to be covered with copper or aluminum with a minimum thickness of
0.01 inch, which must be bonded to building Protective Earth (PE) ground. In addition, the
cabinet of the %'/#56'4must be directly tied to the ground reference plane. Provision for a
mounting bracket is made on the rear and bottom of the %'/#56'4 and the bracket and
screws are provided.
CEMASTER System 13
EFT Connection to Equipment Under Test for Floor-standing EUT
When used with table-top equipment, the equipment should be placed on an insulating table of
0.8 meters height.
CEMASTER System 14
Data and signal cables are tested using the EFT Capacitive Coupling Clamp (CM-CCL). When
performing these tests, the EUT and any auxiliary equipment should be powered from the
%'/#56'4 EUT output, to provide isolation between the test setup and building power mains.
The minimum spacing required between the clamp and the equipment in the test area is shown in
the figure below.
CEMASTER System 15
Special Considerations for ESD Testing
ESD testing involves the use of high-voltage, fast-rising pulses. For this reason the testing
standards specify the use of a ground reference plane in the area where ESD testing is performed.
IEC 1000-4-2 requires the use of a ground plane which must be bonded to building Protective
Earth (PE) ground. The minimum size of the ground plane is 1 m2, the exact size depending on
the dimensions of the EUT; the ground plane shall extend beyond the EUT by at least 0.5 m on
all sides. The ground plane shall be sheet copper or aluminum of 0.25 mm minimum thickness,
or 0.65 mm minimum thickness if another metal is used.
For ESD testing, the %'/#56'4need not be connected to the ground plane.
The EUT shall be positioned at least 1 m from walls and any metallic structures.
For table-top equipment, a non-conducting table 0.8 m high is placed on the ground plane. A
ground plane (HCP) 1.6 m by 0.8 m will be put on the table, and the EUT and any cables shall be
placed on an insulating support 0.5mm thick placed on top of the HCP. The HCP shall connect to
the ground reference plane with a resistive cable with two, 470,000 ohm resistors, one resistor at
each end of the cable.
For Magnetic Field testing, the %'/#56'4 need not be connected to the ground plane.
The EUT is placed on the ground reference plane, on an insulating support 0.1 meter thick.
The coil and EUT shall be positioned at least 1 meter from walls and metallic objects.
CEMASTER System 16
Handling, Transportation, Storage
Save all shipping materials (shipping boxes, foam pieces, bags, and special instructions) for
possible future shipments.
The %'/#56'4 requires two people for unpacking. Follow the instructions in the
INSTALLATION section of this manual and any special instructions shipped with the unit.
Issues related to transportation of %'/#56'4 models are identical to those for installation;
the weight of the unit requires that two people help in packing and movement. When packing
the unit for transportation the original shipping materials should be used, or new materials of
equivalent strength and durability.
For long storage intervals, unplug the power cords of the instrument and cover the instrument
to protect from dust and liquid spills.
CEMASTER System 17
This page left blank intentionally.
CEMASTER System 18
Introduction
All features of the %'/#56'4, and all connections to the Equipment Under Test, are made
from the front panel of the %'/#56'4. The front panel controls and features are broken into
three major groups: the System Interconnections, the Equipment Under Test Connections, and
the User Interface.
CEMASTER System 19
Connections to the Equipment Under Test
On the top, right side of the %'/#56'4 front panel are the connections to the Equipment
Under Test. These include:
For safety, a positive, operator-action is required to energize the EUT; i.e. if the EUT
power switch is ON when the CEMASTER is first turned on, the switch must be toggled
OFF and then ON to energize the EUT Power connector.
CEMASTER System 20
User Interface
When purchased with the CM-FP option, the %'/#56'4 front panel includes a liquid crystal
display with seven function keys, and a 15-key keypad to the right of the display.
CEMASTER System 21
Programming the CEMASTER from the Front Panel
When purchased with the CM-FP option, the front panel display shows status and the display and
keypad allow convenient programming of the %'/#56'4 via screen menus. Programming is
intuitive and easy. There are four basic menu types.
The Main Menu is the 'top' menu; this is the menu seen after the %'/#56'4initializes, and
this is the menu returned to after exiting all lower menus. The Main Menu allows selecting the
test type to perform: Surge, EFT, ESD, etc. You select a test type either by moving a highlight
bar down the screen using the arrow function keys, then pressing ENTER, or by pressing the
number key associated with the test type.
1. EFT/B 5. PQF
2. Surge 6. Pulse Magnetic
3. Power Frequency 7. Calibration
4. ESD 8. System Setup
The Main Menu shows all possible options, however, you can only select an option if it is
installed in the %'/#56'4system. Attempting to select an option which is not installed will
display an error message that the option is not available.
After selecting a test type, a Test Menu opens. The Test Menu offers a user-defined option, and
usually one or more pre-programmed tests to speed setup and testing per European Norms. The
user-defined option gives access and control of every test parameter; this is the most powerful
option, and allows engineering and debugging tests of EUT. The pre-programmed tests are a fast
way to perform certification testing to European Norms, as most parameters default to the proper
values and cannot be changed -- by locking these parameters the user can be certain that the tests
are performed according to the European Norms, and performed the same way every time. Many
pre-programmed tests perform a sequence of tests; for example, at several voltage levels, at
several mains synchronizations, and in several coupling modes. As with the Main Menu, you
select a test either by moving a highlight bar down the screen using the arrow function keys, then
pressing ENTER, or by pressing the number key associated with the test.
EXIT
CEMASTER System 22
After selecting either the User-Defined test option, or a pre-programmed test, a Parameter
Definition Menu opens. The Parameter Definition Menu has edit fields, for waveform,
voltage, etc. To change a field you first move the highlight bar to that field, using the function
keys. After selecting a field there are different editing steps, depending on the type of field.
List fields have a limited number of specific values. An example is the line synch field
which can accept "L1" or "RAND". With a list field selected, the value is changed by
using the keypad arrow keys to scroll through the options.
Numeric fields are edited by typing a value at the keypad. The keypad input fills the
available field from right to left (If the field allows a decimal place, such as "##.##" and
thus a value between 0 and 99.99, to enter a value of 10, press "1", "0", "0", "0"). Press
"+/-" after the number to make it negative. Press ENTER or move the highlight bar to
another field to accept the value; the value must be accepted before a test can be run.
Some fields combine Numeric and List fields. When selected, the keypad arrow keys
change the list portion (L1,RND) and the keypad number keys replace the numeric value.
When all entries have been made and accepted the test can be started by pressing the F1 key --
the RUN function key. To exit this menu without starting a test, press the F7 key -- the EXIT
function key.
When a pre-programmed test has been selected, many parameters assume default values and
cannot be changed by the user; these fields cannot be selected or highlighted. The fields for these
parameters are replaced with the an AUTO symbol.
When a test is running, a RUN Menu is displayed. The RUN Menu shows the test type (Surge,
EFT, etc.) and the EUT status (EUT ON, EUT OFF) on the top line of the display. The right side
of the display shows the test parameters you programmed: waveform, voltage, coupling mode,
line synch, number of tests, and time remaining in this test step (For example, the Surge Run
Menu will display the test steps -- or test states -- on the STATUS line of CHARGING, FIRING,
CEMASTER System 23
and WAITING. During each of these states, the time remaining field will count down and display
the time remaining in that state.). Some tests, such as ESD, will not display time remaining as the
time remaining depends on how the operator uses the ESD probe.
The left side of the display shows status; from top to bottom this is; Status (of test: charging,
running, etc.), Time (secs remaining in this test; if this test is one of a sequence this is the time
for the individual test and not the time for the entire sequence), Test (the present test in a
sequence, or test n of total tests), and Comm (or communications status: local, remote).
SURGE: RUN EUT OFF
12 Ohm
Status: Charging +2000 V
Time: 37s MAINS:L1,L2/PE
Test: 1/90 L1:180 deg
Comm: Local 3
60 s
STOP PAUSE
In the RUN Menu, the STOP function key -- the F1 key -- terminates a test immediately and
returns to the Parameter Definition Menu. The PAUSE function key -- the F3 key -- causes a
temporary suspension of the test. Once a test has paused, press the F3 key a second time to
resume the test (not all tests can be paused and restarted; if critical timing makes this impossible,
the system will treat a pause as a stop, and the test must be restarted from the beginning). Pause
is useful for checking EUT status.
Selections 7 and 8 of the Main Menu -- Calibration and System Menus -- do not run tests, but
perform special functions. The Calibration Menu allows calibration of the CEMASTER by a
Thermo KeyTek authorized service representative. The System Menu provides troubleshooting
and diagnostic tools to the user, and provides access to run-time options.
For more information on the Calibration and System Menus see the TROUBLESHOOTING
chapter.
CEMASTER System 24
STATUS: INITIALIZING EUT OFF At power-up a standard message screen is
shown while the system initializes and
Hazardous voltages and emissions are performs self-tests.
present by design when this equipment is
operating. Take safety precautions at Press any key to show the Main Menu screen
all times. below.
<PRESS ANY KEY TO CONTINUE>
MAIN EUT OFF The Main Menu allows selecting a test type;
move the highlight bar using the arrow
1. EFT/B 5. PQF function keys and press ENTER, or press the
2. Surge 6. Pulse Magnetic number key associated with the test type. As
3. Power Frequency 7. Calibration example, press 2 to open the Surge menu.
4. ESD 8. System Setup
SURGE: TESTS EUT OFF The Test Menu offers pre-defined tests to
1. User defined 6. Class 3 Fast speed setup and a user defined mode for
2. Pre-Qual A 7. Class 1 Full control of every parameter. Press 1 to open
3. Pre-Qual B 8. Class 2 Full the user defined menu.
4. Class 1 Fast 9. Class 3 Full
5. Class 2 Fast The EXIT function key moves back one menu
level.
EXIT
SURGE: USER DEF EUT OFF Parameter Definition Menus have edit fields,
Waveform .......... 12 Ohm for waveform, voltage, etc. Move the high-
Voltage ........... +2000 V light bar using the function keys to select a
Output:Coupling ... MAINS:L1,L2/PE field. Edit a field by typing in a number, or
Phase Sync:Angle .. L1:180 deg use the keypad arrow keys to scroll through a
Number of Tests ... 500 list of options. The RUN function key starts
Time Between Tests 60 s the test.
RUN EXIT
.
SURGE: RUN EUT OFF The RUN Menu identifies the test and shows
12 Ohm status. The screen shows test parameters, and
Status: Charging +2000 V EUT status (ON,OFF), unit status (charging,
Time: 37s MAINS:L1,L2/PE waiting, firing), test status (time remaining,
Test: 1/90 L1:180 deg and test 1 of 90), and communications status
Comm: Local 3 (Local, remote). Stop and Pause keys end the
60 s test or causes a temporary hold.
STOP PAUSE
Front Panel Screens showing the Initializing screen and Four Menu Types
CEMASTER System 25
The Status bar indicates EUT status shows the front
the active test menu; panel outlet is ON or OFF
Surge
The bottom line of The RUN / STOP The arrow function The EXIT function key
text maps to the function key starts or keys select a field to moves to the previous
seven function keys stops tests edit menu
Numeric fields are selected by the function keys Some fields combine Numeric and
and edited by typing a value at the keypad. List fields. When selected, the
Press ENTER to accept the value. keypad arrow keys change the list
portion (L1,RND) and the keypad
number keys replace the numeric
value.
CEMASTER System 26
The left column shows test
information: The Status bar indicates EUT status shows the front
the active test; Surge, panel outlet is ON or OFF
Status (charging, running, etc.) and the menu; RUN.
Time (secs left in this test step)
Comm (local, remote)
Test (test n of total tests)
The STOP -F1- function The PAUSE -F3- function The right column shows pro-
key halts the test key pauses the test. Press grammed test parameters:
F3 again to resume waveform, voltage, coupling
mode, line synch, number of
tests, and time between tests
(or length of test).
CEMASTER System 27
Map of Front Panel Menu Structure
The next two pages show the structure of the front panel menus.
Selection Flow, from Main Menu through Test Menu, Parameter Definition Menu, and Run
Menu
CEMASTER System 28
Map of Front-Panel Menus, Page 1 of 2
CEMASTER System 29
Map of Front-Panel Menus, Page 2 of 2
CEMASTER System 30
Programming the CEMASTER from a Personal Computer
When purchased with the CM-SW option, the %'/#56'4may be programmed and controlled
from a Personal Computer (PC) running Windows.
The PC software gives full access to all the features and capabilities of the %'/#56'4,
including running a single test, running a pre-defined test sequence, and running diagnostics.
Full instructions for the software are described in the %'9CTG section.
CEMASTER System 31
Running Tests
STANDARDS OVERVIEW
The European Unions EMC Directive requires that electronic and electrical products be tested
for immunity to both man-made and natural phenomena and to insure that products do not emit
unintentional signals that may interfere with the continued, reliable operation of other products.
Compliance with the EMC Directive requires that products be tested in accordance with
European Norms, or ENs, issued by CENELEC. These ENs are not developed by CENELEC but
are IEC and CISPR standards redesignated as ENs.
Product EMC Standards relate to a particular type of product, system or installation for which
specific conditions must be considered. Product EMC Standards include:
Resistance Welding equipment
Measuring Relays and Protection equipment
Product Standards have priority over Product Family and Generic Standards; however, where no
Product Standard or Product Family Standard exists for a particular product, the relevant Generic
Standard will apply.
Product Family Standards have priority over Generic Standards; however, where no Product
Family EMC Standard exists for a particular product family, the relevant Generic Standard will
apply.
Generic EMC Standards are a special type of Product Family Standard which apply to products
operating in a particular environment for which no dedicated Product or Product Family Standard
exists. They specify a set of essential requirements, test procedures and generalized performance
criteria applicable to products or systems operating in this environment.
CEMASTER System 32
The Generic EMC Standards include: EN 50082-1, EN 50082-2, EN 50081-1, and EN 50081-2
which cover immunity and emissions testing for commercial, residential and light industrial
environments, as well as for industrial environments.
Basic EMC Publications may be standards or technical reports which, by definition, are NOT
dedicated to specific product families, products, systems or installations. They may concern:
Terminology
Descriptions of electromagnetic phenomena
Specification of compatibility levels (NOT compliance levels)
General requirements for the limitation of emissions
Recommendations for test levels for immunity
Measurement techniques
Descriptions and classification of the environment
Basic EMC Publications include Basic EMC standards such as the IEC 1000-4-X Series, CISPR
22, CISPR 11, and others
TEST REQUIREMENTS
The Basic EMC Publications (IEC 1000-4-2, IEC 1000-4-4, etc.) set the waveforms, test
environment, and measurement techniques. The CEMASTER was designed to meet the
requirements of these publications.
Test levels are set by the Generic EMC Standards and Product Family Standards. The
preprogrammed tests stored in the CEMASTER are designed to meet the test levels of:
Most of the remainder of a test procedure is set by the Basic EMC Publications (IEC 1000-4-2,
IEC 1000-4-4, etc.) which define the test setup and environmental limits, in detail.
The test procedure should also describe how the EUT is setup and connected to test equipment
and ancillary equipment. If the EUT involves software, a description of which software and
software revisions should be included.
CEMASTER System 33
An important part of both the test procedure and the test report is to establish how the EUT is
exercised during the test. Most test procedures recommend that the EUT be operated in "the most
sensitive mode". Preliminary testing will usually be required to determine what the most
sensitive mode of operation is, which may vary with configuration, software, and operations
performed.
Finally, the test procedure should consider what level of disturbances of operation is allowed.
Ideally the EUT should operate through testing without disturbance, however there may be
disturbance which are considered acceptable, in that they do not result in damage. As example,
the following are four levels of system disturbance:
In no case is it acceptable that the equipment become dangerous or unsafe as a result of testing.
The test report should present all of the information of the test procedure, report the test results,
and give explanation of the results, as required. For acceptance tests, the test report is a required
result of testing.
The remainder of this section covers the basic EMC publications in detail; explains the
waveforms and testing requirements; and explains how to use the %'/#56'4 to meet the
requirements of the Basic EMC Publications.
The information on IEC requirements -- figures and diagrams, tables and data, and
test levels -- are accurate as of the date of printing of this manual. However, the
IEC and EN documents are subject to review and change. The information in this
manual is intended as a guideline only.
CEMASTER System 34
IEC 1000-4-2 -- ESD Testing
What is Electrostatic Discharge (ESD) Testing?
These tests relate to equipment, systems, sub-systems, and peripherals which may be involved in
static electricity discharges (either from persons or adjacent equipment), owing to environmental
and installation conditions, such as relative low humidity, use of low-conductivity (artificial
fiber) carpets, vinyl garments, etc.
Two different types of tests can be performed with the ESD head - Contact or Air. Contact
discharges are tests in which a sharp metal ESD tip physically touches the EUT. This method is
regarded as the most repeatable method, and is the only acceptable method for waveform
verification and calibration. Air discharges are tests in which a rounded ESD tip is quickly thrust
toward the EUT until an ESD event occurs by "leaping" from the ESD tip to the EUT through the
air. This method does not yield as repeatable results, but can be used if a contact discharge
cannot be performed for some reason or another.
The ESD waveform is characterized by the current in Contact-mode discharge, as defined in the
figure and table below:
Level Set Voltage Peak Current ( 10%) Current (30%) at 30nS Current (30%) at 60nS
1 2 kV 7.5A 4 2
2 4 kV 15 A 8 4
3 6 kV 22.5 A 12 6
4 8 kV 30 A 16 8
CEMASTER System 35
General Guidelines for Testing
Place the air-discharge or contact-mode output tip into the ESD probe output tip receptacle.
Connect the ESD probe to the %'/#56'4using the 8-pin DIN cable. Connect the discharge
return cable from the ESD probe to ground (an indicator on the ESD probe will light if the
discharge return cable is not connected to the ESD probe). The bargraph indicator on the ESD
probe is intended as a rough guide to the actual test voltage, and can serve in many applications
as a pulse -- discharge -- indicator. The bargraph display is not calibrated, and can be affected by
high humidity; for precise readings of probe tip voltage or periodic calibrations, Thermo
KeyTek's DCA-2, high-voltage divider should be used.
The ESD probe can be operated in manual triggering mode -- normal -- or automatic triggering
mode -- tripod. Normal triggering is intended for handheld applications; in normal mode the
probe will fire once for each time the probe button is pressed if the repetition is set to one, or fire
continuously at 1pps or 20pps while the button is depressed if the repetition is set higher. Tripod
triggering mode is intended for tripod operation; in this mode the probe will start triggering when
the probe button is pressed and released and operation continues until either: the test ends, the
operator ends the test by pressing the STOP -- F1 function -- key, or the probe button is pressed a
second time.
For safety, in all modes of operation, the operator must select a program and press run from the
%'/#56'4 front panel or application software, THEN press the probe button to start the test.
For Contact discharges the TPC-3 contact tip should physically touch the EUT. For Air
discharge the rounded, TPA-2 tip is thrust toward the EUT at a constant rate until an ESD event
occurs by "leaping" from the ESD tip to the EUT through the air. The probe is then pulled back -
away from the EUT - in preparation for the next test. Some practice is required to obtain
consistent results with air discharge.
Certification testing per European Norms uses 1pps for compliance testing. Use 20 pulses per
second mode for quick scans to determine the most sensitive areas; then switch to 1pps for
compliance testing. In 20 pps mode the pulse-detection and pulse counting features of the
%'/#56'4 are disabled -- after the first pulse event, the ESD probe will free-run at 20pps
whether or not there is a discharge.
CEMASTER System 36
Testing Requirements per IEC 1000-4-2
Testing to IEC 1000-4-2 has the following requirements:
Ambient temperature: 15 C to 35 C
Relative Humidity: 30 % to 60 %
Atmospheric Pressure: 86 kPa (860 mbar) to 106 kPa (1060 mbar)
The discharge return cable should, where possible, be placed off the ground reference plane,
and should not come closer than 0.2 m to other conductive parts of the test setup.
Vertical Coupling Planes (VCP) and Horizontal Coupling Planes (HCP) may be used for
indirect application of the ESD pulse. These planes shall be constructed from the same
material and thickness as the ground plane, and should be connected to the ground plane by a
cable with two, 470,000 ohm resistors, one resistor at each end of the cable.
For table-top equipment, a non-conducting table 0.8 m high is placed on the ground plane. A
ground plane (HCP) 1.6 m by 0.8 m will be put on the table, and the EUT and any cables
shall be placed on an insulating support 0.5mm thick placed on top of the HCP. The HCP
shall connect to the ground reference plane with a resistive cable as in the previous
paragraph.
IEC 1000-4-2 requires the use of a ground plane which must be bonded to building Protective
Earth (PE) ground. The minimum size of the ground plane is 1 m2, the exact size depending on
the dimensions of the EUT; the ground plane shall extend beyond the EUT by at least 0.5 m on
all sides. The ground plane shall be sheet copper or aluminum of 0.25 mm minimum thickness,
or 0.65 mm minimum thickness if another metal is used.
For Floor-standing EUT the EUT is placed on a non-conducting mat of 0.1 mm minimum
thickness placed on the ground plane. ESD testing may be direct -- applied to the EUT -- or
indirect -- applied to the ground plane or to a vertical plane bonded to the ground plane.
CEMASTER System 37
Typical ESD Test Setup for Floor-standing EUT
During testing, the EUT is operated in a normal fashion. Compliance testing is typically
performed with the EUT continually operating in the most sensitive mode, as determined by
preliminary testing.
Pulses are applied to points and surfaces which are accessible to personnel during normal usage.
The voltage is increased from the minimum to the selected test level, in order to determine the
threshold of failure (if any). The final test level is set by the product specification and the test
class. The test is applied with single discharges, at preselected points. At least ten discharges of
the most sensitive polarity shall be applied, and a rest interval of at least one second is
recommended (or longer, as required to verify continued operation of the EUT).
CEMASTER System 38
The %'/#56'4 supports 1000-4-2 testing to class 3 in Air-discharge mode and class 2 in
contact discharge mode. The following table lists the test capabilities of the ESD hardware:
Peak Voltage
Class kV
Contact Air
1 2 2
2 4 4
3 N/A 8
There are two tests mandated by the specifications - a 4 kV contact discharge and an 8 kV air
discharge. Repetition is set for 1 pps and polarity is set by the user to the most sensitive polarity.
The user can select mode -- normal or tripod -- and the number of pulses.
CEMASTER System 39
Waveform Verification
IEC 1000-4-2 requires that the ESD output be verified periodically. Verification requires an ESD
target (Thermo KeyTek CTC-3), a target plane (Thermo KeyTek TP-1), and an oscilloscope of
1GHz minimum bandwidth. The ESD probe with contact-mode tip is placed on a tripod, the
%'/#56'4set to TRIPOD mode, 1pps.
The rise time, peak amplitude, and amplitude at 30 ns and 60 ns should be monitored and
recorded and checked against the figure and table at the start of this section.
Performing ESD calibration is not trivial. The high-speed oscilloscope is expensive; the
extremely fast risetime and high-bandwidth required by specification can require that a
dedicated test site be built, with Faraday-shields for the oscilloscope, and EMI damping
materials on floors and walls to reduce reflections.
Thermo KeyTek offers a calibration service for all equipment, including the ESD probe.
Contact factory for details.
CEMASTER System 40
IEC 1000-4-4 -- EFT Testing
For safety reasons, the precautions taken during EFT testing should be the same as
for Surge testing -- read and follow the guidelines of the Safety Concerns During
Surge Testing section, Page 4.
It is important that the Equipment Under Test be properly connected -- read and
follow the guidelines of the Test Area Considerations and Site Preparation
section and of the Special Considerations for EFT Testing section, Page 12, as
well as the Testing Requirements per IEC 1000-4-4 section which follows.
The test is intended to demonstrate the immunity of electrical and electronic equipment when
subjected to types of transient disturbances such as those originating from switching transients
(interruption of unsuppressed inductive loads, relay contact bounce, etc.).
EFT Waveform
CEMASTER System 41
General Guidelines for Testing
The EFT test voltage is referenced to the chassis of the %'/#56'4 The EFT waveform is
coupled into the EUT via the AC Mains and the mains coupler built into the %'/#56'4, or via
signal lines (or both). An optional Capacitive Cable Clamp is available for coupling to signal
lines or to power lines where the CEMASTER internal coupler cannot be used.
When applied to the EUT Mains, the EFT voltage is applied between ground -- the Ground
Reference Plane (GRP) -- and each of the power supply terminals and between the GRP and the
protective earth of the EUT.
Power-line -- Mains -- testing is performed by plugging the EUT into the %'/#56'4 EUT
outlet, establishing the EUT in a normal operating mode, and applying the EFT waveform to,
first the line (L1), then neutral (L2), then PE.
Data-line testing starts with the EUT and auxiliary equipment plugged into the %'/#56'4
EUT outlet -- to prevent EFT waveforms from contaminating the AC Mains -- and couples the
EFT waveform from the CLAMP outlet to a Capacitive Cable Clamp where the waveform is
coupled into the signal lines. The Capacitive Cable Clamp should be adjusted to fit snugly about
the cables to be tested.
IEC 1000-4-4 requires the use of a ground plane which must be bonded to building Protective
Earth (PE) ground. The minimum size of the ground plane is 1 m2, the exact size depending on
the dimensions of the EUT and %'/#56'4; the ground plane shall extend beyond the EUT
and %'/#56'4 by at least 0.1 m on all sides. The ground plane shall be sheet copper or
CEMASTER System 42
aluminum of 0.25 mm minimum thickness, or 0.65 mm minimum thickness if another metal is
used.
The cabinet of the CEMASTER must be directly tied to the ground reference plane. Provision
for a mounting bracket is made on the rear of the CEMASTER and the bracket and screws are
provided.
For floor mounted equipment, the EUT shall be placed on the GRP and shall be insulated from
the GRP by an insulating support 0.1 meter 0.01 meter thick.
For table-top equipment, the EUT should be placed on a table of height 0.8 meter 0.08 meter
above the GRP.
CEMASTER System 43
EFT Connection to Equipment Under Test for Table-top EUT
Data and signal cables are tested using the EFT Capacitive Coupling Clamp (CM-CCL). When
performing these tests, the EUT and any auxiliary equipment should be powered from the
%'/#56'4EUT output, to provide isolation between the test setup and building power mains.
The minimum spacing required between the clamp and the equipment in the test area is shown in
the figure below (Note: if the Auxiliary equipment is not considered part of the EUT, the
separation distance is 5 meters as shown. If the Auxiliary equipment is considered part of the
EUT, the separation drops to 1 meter).
The voltage test levels to be applied vary with the test class, and are listed below:
The level used in testing depends on the European Norm (EN) chosen to test against. As
example, EN50082-1 (for Residential, commercial, and light industry) specifies a test level of
1000 volts. The EFT voltage is applied for one minute, for each test.
CEMASTER System 44
User Defined EFT Test
The user-defined test option allows individual control of all test parameters required in EFT
testing; this allows testing at different levels, modes, and repetition rates.
On being triggered by pressing the start key, a user defined test produces a series of output
events with the parameters defined below:
CEMASTER System 45
Waveform Verification
IEC 1000-4-4 requires that the EFT output be verified periodically. For verification, the output of
the generator is connected to an oscilloscope through a 50-ohm, high-voltage coaxial attenuator.
The rise time, duration, and repetition rate of the impulses within one burst should be monitored
and recorded.
The bandwidth of oscilloscope and attenuator must exceed 400MHz. For digital oscilloscopes a
sampling rate exceeding 1GSa/s is recommended. The attenuator must have a voltage rating of
1250V minimum.
The test setup shown below is usable for Waveform Verification. The BARTH Model
2237HFNFP attenuator4 has sufficient voltage and power rating to connect directly to the
%'/#56'4 output. The Tektronix5 attenuators drop the voltage to within the range of typical
oscilloscope inputs. The circuit below provides a total attenuation of 1000:1.
The circuit above is for measurement of the EFT waveform at the HN-coax connector;
this circuit is not rated for use across the AC Mains outlet. Connecting this circuit
across the AC Mains with Mains voltage present will result in damage of the
attenuator.
If measuring EFT waveform at the AC Mains outlet: Check that the EUT switch is off
and no AC is applied to the EUT Mains Inlet before connecting the attenuator; make the
measurement and remove the attenuator immediately.
4
Barth Electronics, Inc., 1300 Wyoming St., Boulder City, NV 89005 USA
5
Tektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070-1000 USA
CEMASTER System 46
IEC 1000-4-5 -- SURGE Testing
Surge testing is hazardous -- read and follow the guidelines of the Safety
Concerns During Surge Testing section, Page 4.
It is important that the Equipment Under Test be properly connected -- read and
follow the guidelines of the Test Area Considerations and Site Preparation
section and of the Special Considerations for Surge Testing section, Page 11.
Surges occur on the AC power mains as a result of switching operations in the power grid and
from nearby lightning strikes, either directly to the power distribution system or to nearby
ground. Radiated coupling of surges into I/O lines can also occur.
Electronic products are tested for Surge immunity to insure their continued reliable operation if
subjected to realistic levels of surge voltages.
The waveform used for IEC testing is the 1.2/50 s open circuit voltage (8/20 s short circuit
current) combination wave. This waveform is separately characterized by the open circuit voltage
and short circuit current.
CEMASTER System 47
Short-Circuit Surge Current
The %'/#56'4 includes a built-in coupler which can place surges on AC or DC Mains for
application to the EUT. The angle of application onto the AC Mains waveform can be precisely
controlled; application at 90 and 270 cause the surge to add to the peak line voltage for worst-
case voltage; application at other angles can be used to find sensitive components -- such as
diodes, SCRs, or solid-state switches -- which are most sensitive at the time they turn on or off.
The %'/#56'4 also has front panel, safety socket connections to apply surges to the magnetic
field test coil (CM-HCOIL) or for component testing. The CM-I/OCD optional accessory
provides convenient coupling of surges onto data and I/O lines.
When surging a DC Mains or the EUT with no mains applied, the Line Coupling Phase Angle
selection must be set to RND -- random -- as the test will be unable to synchronize to a DC
mains and will not run.
CEMASTER System 48
Probing the EUT during Surge testing should be attempted only by experienced test engineers.
When connecting probes to the EUT, follow these guidelines:
Use balanced, differential, high-voltage probes rated for the surge voltage (such as the
Thermo KeyTek PK1001D probe set). Most oscilloscope probes are rate for 600Vac or less
and should NOT be used in Surge testing.
Do not connect probe grounds to the EUT! This will create a path for Surge and fault
currents which can endanger the operator and test equipment.
If connecting a voltmeter or other high-impedance test equipment to the EUT, carefully check
the common-mode voltage rating of the equipment. This is the voltage rating of both probes
above the instrument ground. Surge testing will typically exceed the common-mode voltage
rating of most Mains-powered equipment, and can exceed the rating of battery-powered
equipment. Use of an instrument above its common-mode voltage rating may result in
improper operation or damage.
6
Pearson Electronics, Inc., 1860 Embarcadero Rd., Palo Alto, CA 94303 USA
CEMASTER System 49
Testing Requirements per IEC 1000-4-5
The Basic EMC Standards for Surge define the methods of generating consistently reproducible
surge voltages for test purposes. They specify generator and coupler/decoupler design and
performance in sufficient detail to produce correlatable results between test sites. These are
issues which are controlled by the design of the CEMASTER and need not be considered by the
operator or user.
While the Basic EMC Standard specifies how to perform Surge testing, the Generic, Product and
Product Family Standards specify the test levels and pass/fail performance criteria. IEC 1000-4-5
also includes a climatic specification, as follows:
The test level, polarity, and source impedance are set by the European Norms (EN), as a function
of use and exposure class (See the Class Definition Table and Test Levels Table, below). The
%'/#56'4 is capable of testing to Classes 0, 1, 2, and 3 test levels.
CEMASTER System 50
Class definitions
Class 0 Well-protected electrical environment, often within a special room.
Class 1 Partly protected electrical environment.
Class 2 Electrical environment where the cables are well separated, even at short runs.
Class 3 Electrical environment where cables run in parallel.
Class 4 Electrical environment where the interconnections are running as outdoor cables
along with power cables, and cables are used for both electronic and electric
circuits.
Class 5 Electrical environment for electronic equipment connected to tele-
communication cables and overhead power lines in a non-densely populated
area.
Class x Special conditions specified in the product specification.
CEMASTER System 51
User Defined Surge Test
The User Defined menu allows individual control of all test parameters required in surge testing;
this allows testing at different levels, modes, and repetition rates.
A User defined Surge Test is controlled by a set of test parameters and results in a surge event, or
number of surge events, whose parameters are defined in the table below.
Pre-Qualification test sequences offer test levels below the Class 1 test level. These tests are
useful for tracking down problems in units which fail to meet the Class 1 test levels.
Fast test sequences perform tests to one and only one Class level, while Full test sequences
follow the directions of IEC 1000-4-5, testing to a Class test level and all Classes below that
level. A Full test sequence is required to certify equipment. A Fast test sequence allow
quickly tracking down problems; e.g. if a unit always passes Class 1 only to fail Class 2, the Fast
test can test only the Class 2 level.
CEMASTER System 52
After selecting a preprogrammed test you have the options of selecting Line Coupling, and
Interval between tests.
To meet the specifications of IEC 1000-4-5 the Interval between tests should be 60 seconds (the
default) or greater. Use a larger interval, if needed, to verify operation of the EUT between
pulses. While you can use a shorter interval to speed pre-qualification testing, note that the surge
event can cause stress to the EUT; applying surges at too fast a rate may be overly stressful and
could result in damage to the EUT. The time to complete column in the following table assumes
the default Interval between tests is used.
When Line Coupling is set to MAINS, the %'/#56'4 cycles through each line coupling
mode: L1-to-L2, L1-to-PE , and L2-to-PE.
When surging a DC Mains or the EUT with no mains applied, the Line Coupling Phase Angle
selection must be set to RND --random -- as the test will be unable to synchronize to a DC
mains and will not run.
CEMASTER System 53
Waveform Verification
IEC 1000-4-5 requires that the simulator output be verified periodically. High voltage differential
surge probes are required for verifying the open-circuit voltage, and a suitable current
transformer is required for verification of the short-circuit current. A digital or storage
oscilloscope with 100MHz bandwidth is sufficient for measuring the surge voltage and current
waveforms and peaks.
The probes are connected across the lines to be verified (e.g. L1 and L2), the %'/#56'4 is
programmed to apply a pulse to these lines, and the signal is captured by the oscilloscope. The
peak value, rise, and duration is measured and calculated and compared to the waveforms at the
beginning of this section.
When measuring current at the EUT outlet, the EUT outlet must be switched off and the EUT
Mains power cord should be removed from the inlet for safety. When measuring voltage across
the EUT output, the AC Mains should be switched off and the line synchronization set to
RANDOM; this simplifies measurements, as the Surge Voltage waveform will not be
superimposed on the Mains waveform.
CEMASTER System 54
IEC 1000-4-8
POWER-FREQUENCY MAGNETIC FIELD Testing
What is Power Frequency Magnetic Field Testing?
Electronic products are often subjected to magnetic fields at AC mains frequencies. These fields
are frequently produced in the vicinity of power transformers and can cause problems with
CRTs, Hall effect sensors, and other electronic products having a sensitivity to magnetic fields.
Electronic Products are tested for immunity to power frequency magnetic fields to insure their
continued reliable operation when placed in service.
The induction coil is characterized by a test volume, and an Induction Coil Factor; the coil factor
is the ratio between the magnetic field strength generated by an induction coil and the current
value required to obtain that field strength. For a standard coil, the coil factor is determined by
the manufacturer of the coil, and can be verified by laboratory measurements before carrying out
the tests.
The %'/#56'4 can source 5 amperes; this is sufficient to drive a coil with a coil-factor of 0.8
to 4 A/m. With lesser coil factors, the current will increase, and this may cause distortion of the
sinusoidal current. When the programmed magnetic field strength, divided by the coil factor,
exceeds 5 amperes, the current should be checked for indications of distortion.
The size and shape of the induction coil may vary, but one standard coil is the one-meter square
coil available from Thermo KeyTek as CM-HCOIL. This coil is capable of providing a constant
field-strength over a volume of 0.6 by 0.6 by 0.5 meter; for EUT larger than this, a larger coil
will be required. The Coil Factor for this coil is 0.87.
Since the Equipment Under Test may show sensitivity to the direction of magnetic field, the coil
or EUT is typically rotated to apply the field in three orthogonal directions.
The unit of magnetic field measurement is A/m. In free-space, 1 A/m is equivalent to 1.26 micro-
Teslas or 0.0126 Gauss.
CEMASTER System 55
Example of the three Orthogonal Axes for Magnetic Field Testing
IEC 1000-4-8 requires the use of a ground plane which must be bonded to building Protective
Earth (PE) ground. The minimum size of the ground plane is 1 m2, the exact size depending on
the dimensions of the EUT; the ground plane shall extend beyond the EUT by at least 0.5 m on
all sides. The ground plane shall be sheet copper or aluminum of 0.25 mm minimum thickness,
or 0.65 mm minimum thickness if another metal is used.
The EUT is placed on the ground reference plane, on an insulating support 0.1 meter thick. The
EUT is operated normally and the magnetic field is applied. The coil is rotated twice and the test
is repeated.
IEC 1000-4-8 requires that one terminal of the test generator be attached to the ground reference
plane; the %'/#56'4 does this internally. You should not make an external connection
between coil and ground reference plane.
CEMASTER System 56
A User-defined test is controlled by a the parameters in the table below:
Waveform Verification
IEC 1000-4-8 requires that the simulator output be verified periodically. This is to ensure that the
EUT is being tested at the proper test level. Waveform Verification takes place with the same
setup as for EUT testing, but with the EUT removed.
The %'/#56'4 is set for the desired field strength. The magnetic-field strength meter is
placed in the center of the coil and oriented to achieve a maximum reading. The reading of the
field meter is compared to the desired field-strength, and the coil factor adjusted accordingly:
lower if the measured field strength is too low, higher if the measured field strength is too high.
The measurement and adjustment may be repeated until the measured field strength is within
acceptable tolerances of the desired field strength.
CEMASTER System 57
IEC 1000-4-9
PULSED MAGNETIC FIELD Testing
What is PULSED MAGNETIC FIELD Testing?
Pulse Magnetic fields are produced as a result of a large current impulse through a conductor.
An example is lightning current flowing through a grounding conductor at a power sub-station.
Pulse magnetic fields can also occur in heavy industrial areas where very large current impulses
are used in a manufacturing process.
Electronic products are tested for immunity pulse magnetic fields to insure their continued
reliable operation when placed in service in a very harsh environment.
The induction coil is characterized by a test volume, and an Induction Coil Factor; the coil factor
is the ratio between the magnetic field strength generated by an induction coil and the current
value required to obtain that field strength. For standard dimension coil, the coil factor is
determined by the manufacturer of the coil, and can be verified by laboratory measurements
before carrying out the tests.
The size and shape of the induction coil may vary, but one standard coil is the one-meter square
coil available from Thermo KeyTek as CM-HCOIL. This coil is capable of providing a constant
field-strength over a volume of 0.6 by 0.6 by 0.5 meter; for EUT larger than this, a larger coil
will be required. The Coil Factor for this coil is 0.87.
Since the Equipment Under Test may show sensitivity to the direction of magnetic field, the coil
or EUT is typically rotated to apply the field in three orthogonal directions.
The unit of magnetic field measurement is A/m. In free-space, 1 A/m is equivalent to 1.26 micro-
Teslas or 0.0126 Gauss.
CEMASTER System 58
Example of the three Orthogonal Axes for Magnetic Field Testing
IEC 1000-4-9 requires the use of a ground plane which must be bonded to building Protective
Earth (PE) ground. The minimum size of the ground plane is 1 m2, the exact size depending on
the dimensions of the EUT; the ground plane shall extend beyond the EUT by at least 0.5 m on
all sides. The ground plane shall be sheet copper or aluminum of 0.25 mm minimum thickness,
or 0.65 mm minimum thickness if another metal is used.
The EUT is placed on the ground reference plane, on an insulating support 0.1 meter thick. The
EUT cabinet is connected to the ground plane via the EUT earth terminal (if any). The EUT is
operated in a normal fashion and pulses are applied. The coil is rotated twice and the test is
repeated.
IEC 1000-4-9 requires that one terminal of the test generator be attached to the ground reference
plane; the CEMASTER does this internally. You should not make an external connection
between coil and ground reference plane.
CEMASTER System 59
User Defined Pulsed Magnetic Field Immunity Test
The User Defined menu allows individual control of all test parameters required in Pulsed
Magnetic Field Immunity Testing; this allows testing at different levels, number of pulses, and
time between tests.
There are two pre-stored IEC Standard test sequences, which test to different magnetic field
strengths. The only adjustment the user can make is to the time between tests.
Class 3 Test: This test sets the magnetic field strength to 100 A/m and programs 5 pulses.
Class 4 Test: This test sets the magnetic field strength to 300 A/m and programs 5 pulses.
CEMASTER System 60
Waveform Verification
IEC 1000-4-9 requires that the simulator output be verified periodically, to ensure that the EUT
is being tested at the proper test level. Waveform Verification takes place with the same setup as
for EUT testing, but with the EUT removed.
The current probe is placed over one of the coil leads. The %'/#56'4 is set for Power
Frequency Magnetic Field testing, and programmed for a field strength of three A/m. The
magnetic-field strength meter is placed in the center of the coil and oriented to achieve a
maximum. This reading is recorded. The rms-reading of the current is recorded. The ratio of
magnetic field to current is calculated and recorded.
The %'/#56'4 is set for Pulsed Magnetic Field testing, and programmed for the proper field
strength specified by the test level. A pulse is programmed and the peak current recorded via the
oscilloscope. The ratio of magnetic-field to current -- calculated in the prior step -- is multiplied
by the peak current -- measured in this step -- to obtain the actual magnetic field applied during
test. This result is compared to the test level, and the coil factor adjusted accordingly: lower if the
measured field strength is too low, higher if the measured field strength is too high. The
measurement and adjustment may be repeated until the measured field strength is within
acceptable tolerances of the desired field strength.
Alternatively, a peak reading magnetic field meter of bandwidth greater than 10MHz may be
used. In this case the magnetic field may be measured directly (the current need not be
measured).
The current waveform obtained in Pulsed Magnetic Field Testing is identical to that
obtained in Surge testing. IEC 1000-4-9 refers to the wave as a 6.4x16uS wave; this is the
result of a difference in definition of risetimes and duration. Per IEC 1000-4-9, the
risetime is simply the time from 10% to 90% with no multiplier. The duration is simply
the time from 50% of rise to 50% of fall.
7
Pearson Electronics, Inc., 1860 Embarcadero Rd., Palo Alto, CA 94303 USA
CEMASTER System 61
IEC 1000-4-11
POWER QUALITY FAILURE Testing
What is Power Quality Failure Testing?
Dips and interrupts can occur on the AC power mains as a result of a fault in the distribution
system such as an open circuit breaker or a sudden large load being turned on in the immediate
vicinity. A power distribution system fault can cause a switch in the distribution grid to open and
close a number of times, resulting in multiple interrupts to electrical and electronic equipment.
Electronic products are tested for immunity to dips and interrupts to insure their continued
reliable operation if subjected to dips and/or interrupts on the AC power mains.
IEC 1000-4-11 requires the generator be capable of supplying peak inrush currents of 500A for
220V to 240V mains, and 250A for 100V to 120V mains. The %'/#56'4is designed to meet
these levels, however -- due to building wiring capabilities -- the site where the %'/#56'4 is
being used may not meet the requirement.
The peak inrush current should be measured at the test site, using the circuit and method
described in the Waveform Verification section.
If the %'/#56'4plus test site meets the inrush current requirement (500A peak for 220V to
240V mains, or 250A for 100V to 120V mains), it is not necessary to measure the EUT peak
inrush current.
If the CEMASTER plus test site does not meet the inrush current requirement, the test can still
be performed as long as the EUT peak inrush current is less than the peak inrush current of the
%'/#56'4 plus test site. In this case, it is necessary to measure the EUT peak inrush current.
CEMASTER System 62
The circuit and method for measuring EUT peak inrush current is described in the Waveform
Verification section.
CEMASTER System 63
Each sequence has the following, common characteristics:
Default/Standby Test Level = 100%
Starting Phase Angle = 0 (and also 180 for Test Duration < 0.02 s)
Interval Between Tests = 10 s
Number of Tests = N/A
Each of the IEC Standard test sequences, except Combination, is performed at the Test Level
and using the Test Duration defined in the table above, as follows:
Do not allow a new test sequence to start until Interval Between Tests has elapsed
End of test sequence
CEMASTER System 64
Waveform Verification
IEC 1000-4-11 requires that the simulator output be verified periodically. For Dip & Interrupt
test simulators, it is necessary to verify the voltage transition levels, transition times to 100%,
and the inrush current capability. Most modern oscilloscopes are capable of observing the voltage
levels and transition times. Transition times are measured with a 100-ohm, 600-watt load. For
verifying the inrush current, a bridge rectifier, suitably rated 1700uF capacitor, and appropriate
current transformer (Pearson8 model 110 or equivalent) are required.
The Line Qualify test sequences meet the IEC 1000-4-11: 1994 standard for qualifying the Test
generator peak inrush current drive capability". These test sequences require the use of a standard
line qualification test load, an uncharged capacitor whose value is 1700 F in series with a
suitable rectifier (from IEC 1000-4-11: 1994, Clause 6.1.2). The two preprogrammed test
sequences for verifying the peak inrush drive current capability of the test generator are:
Line Qualify 90
Line Qualify 270
And the Line Qualify test sequence consists of the following steps:
Start at Default/Standby Test Level (open circuit)
Switch to Test Level at the Starting Phase Angle defined by the test sequence name
Maintain Test Level for Test Duration
Switch to Default/Standby Test Level (open circuit)
End test
Warning!
The Line Qualification Load will draw significant current; it should be plugged into the EUT
outlet only when no voltage is present -- EUT switch OFF. DO NOT plug the Line Qualification
Load into a live outlet as an arc may result which might damage the outlet.
The charged capacitor requires a time to discharge after each test. The time allowed should
exceed 7 time constants (seven times value-resistor {in ohms} times value-capacitor {in farads}).
Failure to allow sufficient time to discharge may give invalid readings. Allow the Line
Qualification Load to discharge -- EUT switch OFF -- before handling.
8
Pearson Electronics, Inc., 1860 Embarcadero Rd., Palo Alto, CA 94303 USA
CEMASTER System 65
Line Qualification Test Setup -- Mains Peak Current
CEMASTER System 66
%'/#56'4 %'9CTG
A Help file for the %'/#56'4%'9CTG is provided with the software which can be opened by
clicking Help|Contents on the menu bar. It provides complete operating instructions for the
program; using the Help file while experimenting with the program is the recommended method
for learning the software. These written instructions are largely duplicated in the Help file.
The CM-SW option, available as a purchased %'/#56'4 option, includes the %'9CTG
program. It may be used on a Personal Computer (PC) running either Windows 3.1 or
Windows95. The computer must have at least 16MB of RAM and a 80486/66MHz or faster
processor. Screen transitions are slower on 486 processors, thus Pentium computers are
recommended. %'9CTG is designed for an 800x600 display driver. Some drivers require
selection of small fonts on the Windows 95 display settings. You should select such a driver (or
one with greater resolution) in your Windows configuration. On a driver with a smaller display,
scrolling is necessary to see the entire application screen.
The CM-SW option includes an RS-232 to optical converter (the FC-11), a short RS-232 cable
with a 9-pin D-shell connector, and a fiber optic cable with a duplex modular connector. The
RS-232 connector connects to a Com port on your computer. The fiber optic cable connects to
the %'/#56'4 front panel; this connector is polarized and can only be inserted in one
orientation: with the small tab on the connector facing right. Do not force the connector. The
FiberCom option derives power from the RS-232 port of the computer -- no external power
source is required.
%'9CTGversion 2.0 (and higher) also supports '/%2TQ and has some additional controls.
References to EMCPro and hardware selection in this chapter apply to that %'9CTG version
only. %'9CTGconfiguration for %'/#56'4or '/%2TQ is automatic when a tester is
connected
CEMASTER System 67
Basic Operations
The following steps will take you through major operating modes of the program and will
acquaint you with the screens. Refer to the next section for full description of all the controls
and displays on the screens. These steps can be run in simulation mode or with a %'/#56'4
connected. If the %'/#56'4 is connected and powered on, the application will send
commands to it by default.
Start the %'/#56'4 by powering it on and pressing a key to move beyond the warning
screen.
Start the %'9CTG program by double-clicking on the ICON in the %'/#56'4 group on
Windows 3.1, or running the program from the Start menu (PROGRAMS | %'/#56'4
Software | %'9CTG) on Windows95.
Defining a Test
Click on one of the New Sequence buttons at the left of the Main Screen to create a new test.
Select "User-defined" from the box on the New Sequence Definition screen. The other items
on that screen are defaults (the same ones as on the %'/#56'4 front panel) and can be left
alone or changed, as you wish.
Click "OK" on the Custom Sequence Wizard screen to accept the Wizard defaults. This screen
allows you to create several tests of the same type quickly.
Click "OK" on the Sequence Editor Screen to accept the test sequence. This screen allows you
to manually edit the order of the tests or the test parameters.
Back on the Main Screen, you now have one sequence defined and shown in the grid. You can
return to the Sequence Editor at any time by double-clicking on the row or row number of the
sequence you wish to edit.
On the Menu bar at the top of the screen, select FILE | SAVE if you want to save your sample
test.
Running a Test
Click on the RUN button to start the test. The grids on the Run Screen, and the Status Block at
the lower right, will show the status of the executing tests.
CEMASTER System 68
Hit the ABORT button to return to the Main Screen. After test completion, a notice is
displayed, and return to the Main Screen is automatic on acknowledgment of the notice.
Exit the application by choosing FILE | EXIT from the main application menu.
Remote mode. While the PC %'9CTG program controls the %'/#56'4, the normal idle
screen contains a "Comm: Remote; Status: Idle" indication if no test is running. When a test is
active, a Run Screen (the same as from front panel control) will be shown on the front panel
display. If the EXIT button on the front panel is pressed when in "Remote Idle," the
%'/#56'4 will no longer be controlled by the PC. This is not recommended. Termination of
tests and remote control from the %'9CTG menu is preferable. (ESD test operation is a special
case; see the next item.)
ESD operation. The ESD test function is designed for maximum control at the ESD head.
%'/#56'4operation puts the unit in a "Test Run" state, where it will stay until the user hits
EXIT on the front panel. The same front-panel control is in place when running from %'9CTG.
Leave the ESD run state by hitting EXIT on the front panel when all ESD pulsing is done. You
may also select ABORT from %'9CTG, but this will terminate the entire sequence.
If your %'/CUVGT does not have a front panel, the ABORT from %'9CTG is the only
mechanism for stopping an ESD test. For this reason, putting an ESD test in a sequence with
other test types is not recommended in that configuration.
During ESD tests, the user may perform more tests than are defined in the sequence definition;
for this reason, the %'9CTG screen may show tests executed greater than the total tests, such as
"140 of 100" pulses.
Mouse operations. If running Windows 3.1 with the "snap-to" option enabled on the mouse
driver (see the Mouse driver selections on the Windows Control Panel), the mouse cursor may
jump when the mouse button is pressed. Turn off "snap-to," if it is enabled. Not all mouse
drivers support snap-to.
Editing pre-programmed surge tests. Some of the pre-programmed surge tests (pre-
qualification and Level 1) include voltages which are below the normal limit for surge user-
defined tests. These tests will run as pre-programmed tests, but not as user-defined tests. If you
edit these pre-programmed tests and attempt to run without changing the voltages, %'9CTG will
report errors at each voltage below the normal limit. If these low-voltage, pre-programmed tests
are edited, be certain to remove or change all the voltages below the 200V limit.
Test timing counts. The test count displayed on the %'9CTG screen is a combination of
values from %'9CTG timing and firmware timing. It may sometimes change quickly, faster than
CEMASTER System 69
a normal clock, or lag the front panel count, due to communication delays. The test is being
executed by the firmware, and the front panel has the most current information for each test.
Only %'9CTG supports user-defined sequences and combinations of sequences; that timing
information is summarized in the Sequence List Status block, as described in other sections of
the manual.
CEMASTER System 70
Class 1 Full sequence is the same as the %'/CUVGT sequence Surge (1.2/50) Class 1
Fast, and therefore a direct substitution can be made. Note, though, that the new suite
will not run on an '/%2TQ.
Program settings - defaults - affecting communications, data logging, and user-defined messages,
may be changed from the menu bar; the setup option. See the `Menu, Toolbar, and Status Bar
section for more detail.
Viewing, editing, and printing test reports outside %'9CTG. Test logs may also be
viewed, edited, and printed using a word processor. The following document format controls are
appropriate for MicroSoft Word. Equivalent margin and font selections are likely to be necessary
from other word processors.
1. Start MS Word and select the text log file you wish to print from FILE
OPEN.
CEMASTER System 71
%'9CTGin Detail
Main Screen -- The overview screen; used to create and run tests.
New Sequence Definition Screen -- Used to define a test sequence, selecting pre-programmed
or custom (user-defined) tests.
New Custom Sequence Wizard Screen -- If a User Defined test sequence is selected, the
Custom Sequence Wizard will help define the sequence of tests.
Run Screen -- Shows Test Status information while tests are running.
Main Screen
CEMASTER System 72
Main Screen
The Main Screen is the screen you use to create tests, load or save tests, and run tests.
To create a new test, click on one of the seven test buttons, left-center on the screen. To load an
existing test, click on the Toolbar or Menubar at the top of the screen. To Run a test click on the
RUN button at the bottom left of the screen. The following sections describe the information and
the controls on the Main Screen.
Toolbar
The toolbar at the top of the screen supports some system-level operations. It is described in
detail in the section on Menu, Toolbar, and Status Bar, which appear on several screens.
When on the Run Screen, the Test Status Block shows which test is currently executing, and
shows duration and elapsed times. While displayed on the Main Screen, this block is updated
when a test sequence is selected or deselected on the Sequence List grid. It is described in detail
in the section on the Run Screen.
The test control block has control buttons to start, stop, and control tests.
RUN Button -- The Run button starts the Test. Once the test is running, the Run button
becomes the Pause button.
PAUSE Button (Only enabled on Run Screen) -- Click the Pause button to temporarily
halt the test.
RESUME Button (Only on Run Screen, when in Paused state) -- Click to resume a
paused test.
When you click the Break button, if the Present Sequence Status has a repeat count
greater than one, it drops to one. Any test presently executing continues to completion,
but at completion the present test ends.
If the Sequence List Status has a repeat count greater than one, it drops to one. Any test
presently executing continues to completion, but at completion the present test ends; at
completion of all tests in the present sequence, the present sequence ends.
CEMASTER System 73
COMMENT Button -- The Comment button allows definition of a user comment while
tests are running. The comment can then be inserted into the log file at any instance the
user chooses by selecting the OK button on the comment entry window.
When checked, Auto Pause will enable any pauses defined in the Sequence Editor. At
each pause the operator must click the OK button on a Windows prompt box to continue
the test. When un-checked, any pauses defined in the Sequence Editor are ignored. When
checked, Repeats will allow any repeats defined in the Sequence Editor and in the main
menu. When un-checked, any repeats are ignored; each test and sequence will execute
exactly one time. (This is the mechanism for single-stepping, used in coordination with
the Pause column on the Sequence Editing screen.)
EUT POWER STATUS Button -- The EUT Power status button shows the EUT status.
The EUT power may be turned off from the computer by clicking this button. When off,
the EUT Power can only be turned on from the %'/#56'4, by turning the EUT switch
ON, or toggling OFF and ON.
The Sequence List Grid shows which test sequences have been defined (or loaded with the
current file). The grid allows editing the sequence during the definition and editing phases. A
yellow highlight bar identifies the test to be edited when using CUT, COPY, and PASTE.
To Edit a Test Sequence: Double-click on the gray, numbered, row-header cell to the
left of the test you want to edit, or anwhere in the row of that test. The Sequence editor
will open.
To Add a Test Sequence to the end of the sequence list: Deselect any selected Test
Sequences by clicking the button in the upper-left corner of the grid. Then, click a test
type button.
To Insert a Test Sequence: Click the gray numbered cell to the left of the test you want
to insert the new sequence in front of. Then select a test type by clicking a test type
button.
To Delete a Test Sequence: Click the gray numbered cell to the left of the test you
want to delete. Then click the Cut button from the toolbar.
To Move a Test Sequence: Click the gray numbered cell to the left of the test you want
to move. Click the Cut button from the toolbar to remove this test. Click the gray
numbered cell to the left of the test you want to insert the cut sequence in front of. Then
click the paste button from the toolbar to insert the test.
CEMASTER System 74
Test Type Controls
These 6 buttons at the left of the Main Screen select a test type to create. There are 8 different
immunity tests available, in accordance with European Norm requirements and the controlling,
basic EMC immunity standards of:
The tests available on your system will depend on the options installed in your %'/#56'4. If
a test is not available, the corresponding test button will be grayed out, and cannot be selected.
The legend on the last button will depend on the surge option installed in the system.
Repeat Control
The repeat control sets how many times the sequence shown in the Sequence List grid is
executed. Click this control and enter a number.
Test Description
If a test is loaded from a file, the description box will contain the text description filed with the
test. If a new test is created, enter a description of the test in the description box for inclusion in
reports and logfiles, and to be saved with the test.
Tool Bar
The toolbar holds icons for common tasks, allowing you to select that task with a single mouse-
click. See the section below on the Menu, Toolbar, and Status Bar.
File Icons -- The file icons create a new test sequence, load an existing test sequence,
or save the current test sequence. If you attempt to create a new sequence or load an
existing sequence, you will be prompted to save any existing work.
Edit Icons -- The CUT, COPY, and PASTE icons allow you to edit the test sequence
grid. These icons are also found in the Sequence Editor.
Starting a Run
CEMASTER System 75
When you wish to execute the presently defined sequence list, select the Run button in the Test
Control area. A Windows message box will appear to inform you if you are doing a
Local/Simulation run. If you are about to do a Remote run (connected to a %'/#56'4) no
warning is given.
If Logging is enabled, a Run-Time Logging Information screen will appear where you will see
the header information that will appear in the log report(s) and can edit this information/
You may edit the 60 character log report header line or the 40 character operator name. There is a
60 character field for a description of the equipment under test and a text entry field where you
can make any comments regarding the run before you begin. The comment will appear after the
header on the first page of the log text report.
Enter up to 8 characters for the file name for either or both of the log report types selected. This
must be a valid 8 character DOS filename. The file paths and extensions are shown for reference
only and are not editable in this window.
CEMASTER System 76
New Sequence Definition Screen
The New Sequence Definition Screen is opened when you create a new test by clicking on one of
the test type buttons from the Main screen.
The list box to the left of the screen shows sequences that you can choose. The User Defined
sequence gives you the most control of the test, allowing you to define all of the Sequence
Defaults shown on the right side of the screen. The other options shown -- which vary by test
type -- are test sequences which the %'/#56'4 has stored and available. These sequences
speed set-up and execution of commonly performed IEC test sequences by setting and locking
many of the defaults -- the locked defaults will be grayed, and cannot be selected. The standard
test sequences for each test type are described in the %'/#56'4 manual.
To continue, click one of the sequences shown. If the Sequence Defaults to the right of the screen
are enabled, you can select one or more of the Defaults. Finally, click OK.
CEMASTER System 77
New Custom Sequence Wizard Screen
The New Custom Sequence Wizard screen quickly creates test sequences. The Wizard is run
automatically whenever you select a User Defined Sequence in the New Sequence Definition
Screen. The Wizard allows you to create lists of one or more parameters, then forms all
combinations of these parameters to quickly create a large number of tests. When you click OK,
the Wizard closes and the Sequence Editor opens for fine editing of test parameters.
To use the Wizard, select one of the radio button options at the bottom left of the screen, or click
within any parameter definition box to the right of the screen. The defaults in the combo boxes
are those you selected from the New Sequence Definition Screen. Make a parameter selection,
then click the Insert button (to the left of the parameter definition boxes) to copy the selection
into the list box. You can change list boxes using the radio buttons below the list box, and edit
the list box using the CUT, COPY, and PASTE buttons above the list box.
CEMASTER System 78
As example, if you click Voltages, and insert 250V, 500V, and 1000V, then click Output and
select L1-L2 and L1-PE, when you exit the Wizard, the Wizard will create six tests:
L1-L2 at 250V
L1-L2 at 500V
L1-L2 at 1000V
L1-PE at 250V
L1-PE at 500V
L1-PE at 1000V
If you want to create a User-Defined test sequence with a single test, simply click the OK button
to exit the Wizard and open the Sequence Editor.
Sequence Editor
The Sequence Editor is opened after selecting a test sequence from the New Sequence Definition
Screen, or by double-clicking an existing test from the test grid of the Main Screen. The
sequence editor gives you fine control over what tests are performed and all options. If you wish
to execute a predefined test with no changes, you should simply click on the OK button. If you
make any changes in a predefined test, the test will be re-defined as a User Defined test.
CEMASTER System 79
Sequence Description Grid
The Sequence Description Grid shows which tests have been defined in a sequence.. It supports
the editing process on the Sequence Editing Screen as well as allowing the data to be changed in
any individual cell. Basic test editing is allowed on any test in the grid. Text cells can be
changed by selecting a cell, deleting the current entry, and typing the new one. Cells that support
a list of legal values have a standard Windows drop-down list controlled by the drop-down
button at the right of the cell. Open the list and select a new valve to change it. Additional
editing controls that affect more than one cell are also supported and are described in more detail
below.
The PAUSE column of this grid allows single-stepping or automatic stopping at any test
in the sequence. Check the box in the row of the test where you wish to pause. (The
pause will occur before the test executes.) Make sure that the Auto Pause box is checked,
on the Main Screen, before the test starts. The PAUSE box for all tests in a sequence
may be set with the technique described in this section for changing an entire column.
To set or reset the PAUSE box, use a single left button click of the mouse to select the
PAUSE cell then use single left button clicks to set or reset the check-box in the cell. A
message can be displayed during a PAUSE - see the Message Selection section.
A yellow highlight bar identifies the test to be edited when using CUT, COPY, and PASTE,
which act on entire test rows.
Click the gray, numbered cell to the left of the test you want to delete. The entire row will
highlight. Then click the CUT button from the toolbar.
Click the gray, numbered cell to the left of the test you want to move. Click the CUT
button above the grid to remove this test. Click the gray numbered cell to the left of the
test that will follow the moved test. Then click the PASTE button from the toolbar to
insert the test.
Click the gray, numbered box to the left of the test you want to copy; the entire row will
highlight. Click the COPY button.
CEMASTER System 80
After a Cut or Copy operation, click the gray, numbered box to the left of the test you
want to insert in front of; the entire row will highlight. Click the PASTE button. NOTE:
To select a block of rows, hold [Shift] key when left-clicking; to select multiple single
rows, hold [Cntrl] key when left-clicking.
The Sequence Editor allows several special editing operations in addition to the normal Cut and
Paste Operations. These operations take place over all or part of a column. This can change many
tests with a single editing step.
Click one cell within the column you want to change. Then, click the gray column header
at the top of the grid. All cells in the column will be highlighted.
Press ENTER. The cell you highlighted will open for editing. Select a new value and
press ENTER (if this is a numeric cell, type in a new value and press ENTER; if this is a
check box, press ENTER to toggle).
All cells in the column -- and thus all tests within this sequence -- will take on the new
value you entered.
Click the first cell in the range within the column you want to change. Then, hold down
the shift key and click the last cell in the range you want to change. All cells between the
first and last cells will be highlighted.
Press ENTER. The last cell you highlighted will open for editing. Select a new value and
press ENTER (if this is a numeric cell, type in a new value and press ENTER; if this is a
check box, press ENTER to toggle).
All cells in the range will take on the new value you entered.
When you change a column, other columns may change as a result. These changes will
always be 'legal' -- that is, the test will run with the resulting change, and you will not get
an error. But the changes may not be uniform. Check the other columns to be sure the
tests will run the way you want them to.
Keyboard Editing
The grid can be edited using the keyboard cursor keys, Tab, and ENTER. Press Tab until the
grid is selected -- the highlighted cell within the grid will have a dotted selection-box drawn
inside it. Use the cursor keys to select a cell to edit. Press ENTER to open the cell. If the cell is
CEMASTER System 81
a drop-down box, use the cursors to select an option. If the cell is numeric, type in a new value.
Press ENTER to accept the change.
To assign a message to be displayed during a run-time pause, double-click the left mouse button
on the desired PAUSE cell. This will bring up the Message Selection window where a single 80-
character message may be selected from 200 available messages.
Select the desired message by clicking on the message number button to the right of the message,
then click OK.
CEMASTER System 82
When OK is selected, the Message Selection Window closes, and the message number is pasted
in the PAUSE cell of the Sequence Editor Grid.
To remove a message from being displayed during pause, double-click the left mouse button on
the desired PAUSE cell. This will bring up the Message Selection window where the selected
message will be displayed highlighted.
Deselect the message by clicking on the upper left corner button on the Message Selection grid,
then click OK.
NOTE: The Message Selection grid does allow editing of message text but does not allow Cut,
Copy and Paste operations.
CEMASTER System 83
The Run Screen
The Run Screen is displayed while tests are executing. A Present Sequence grid control is added
to the screen, along with the Sequence List grid control. Together they describe the current set of
tests being run.
This list of sequences is the same as shown on the Main Screen, and shows the presently defined
set of test sequences. A yellow highlight bar indicates the test sequence that is presently
executing.
The Present Sequence Grid is displayed when running tests. The grid shows all the parameters of
the present test sequence, which consists of one or more tests. It is essentially the same
information presented on the Sequence Edit screen. A horizontal highlight bar across the grid
identifies which test of the sequence is running.
CEMASTER System 84
The Test Status Block at the lower right of the Run Screen shows which test is currently
executing, test duration, and elapsed times. It contains two display areas: one for the Sequence
List, and one for Present Sequence -- the one that is currently executing. The two sets of data are
similar: the Sequence List Status summarizes statistics for the entire set of sequences, and the
current sequence within that list; the Present Sequence summarizes information about the
current test within the present sequence.
Sequence -- The sequence that is running, out of the total number of sequences.
Duration -- The duration shown is the estimated time required to complete the entire
sequence, excluding time for pauses. The time format is hours:minutes:seconds.
Elapsed Time -- The elapsed time is the total time from the start of the test to the present
time.
Repeat Count -- These fields show the present loop count, and the total list count for the
sequence. The total reflects the value from the List Repeats field on the Main
Screen.
Test -- These fields identify which test is running out of the total number of tests in the
present sequence.
Duration -- The duration shown is the estimated time required to complete the Present
Test Sequence. The time format is hours:minutes:seconds.
Elapsed Time -- The elapsed time is the total time from the start of the present sequence
to the present time.
Repeat Count -- These fields show the present loop count, and the total list count for the
sequence. The total reflects the value from the Sequence Editor, Sequence Repeats
field.
This set of buttons is the same as on the Main Screen and is described in that section. Buttons
may be enabled or not, depending on the state of the system and the software.
Menubar
The menu bar performs actions, sets options and preferences, and accesses the help file. The
menu options are described below.
CEMASTER System 85
Menubar Selections
File Menu
The File menu supports loading and saving of test files, and exit from the application. The
default extension used for test sequence files is .SEQ.
File | New
Delete any present tests and prepare for creating a new test. If you have tests defined and
have not saved them you will be prompted to save the tests before continuing.
File | Open
Load an existing test file. The File Common-Dialog box will open, allowing you to click
on a file to load. If you have tests defined and have not saved them you will be prompted
to save the tests before continuing.
The Save command will save the present test to a disk file. If the test was loaded from
disk, the test will be written to the same file. If the test is a new test, you will be prompted
for a path and filename to use in saving the test.
The Save As command will prompt you for a path and filename to use in saving the test.
Use this command to create file copies of tests, or new tests based on the old ones that
you have edited.
The Print Sequence command will print the present Sequence List.
The Print Setup command controls the printer, selecting which printer is used, which font
is used, and paper orientation. Print Log will allow the user to select and print a Log file.
This section holds the last four files you opened / saved. You can quickly load a prior file
by clicking one of these selections or typing the corresponding number
File | Exit
CEMASTER System 86
This selection exits from the application.
Edit
The Edit menu allows keyboard selection of the CUT, COPY, and PASTE selections that are
also accessible from the Toolbar. In addition, there is a DELETE selection which does a
destructive cut of selected data.
Setup
The Setup menu is where the user may select and set parameters for run time operation or report
generation. The items on the setup menu are:
Setup menu
Communications
The user may select the COM port to set up communications with a C'/CUVGT.
%'9CTG allows selections (COM1 - COM4) and automatically detects which ports are
available on the system. The default selection is COM1. Other communication
parameters, although shown, are fixed.
CEMASTER System 87
Communication Parameters Screen
Log Reports
Log Reports allows the user to select the type of logging reports they which to produce:
Text Reports or Export Reports. For each type of report, you may define your own file
extension and path or you may accept the defaults (you will be asked to enter a log report
file name at run time). The file extension must be a (up to) 3 character valid DOS file
extension. The path must be a valid DOS path of up to 60 characters in length.
You may also define your own header line that will appear as the first line in the header
of all Log reports. A default report header is supplied. You may enter a new report header
of up to 60 characters.
A 40 character field is supplied where you may enter an operator name that will also
appear on the Log reports.
CEMASTER System 88
Log Reports Screen
Sequence Reports
Sequence Reports allows the user to define his own header line that will appear as the
first line in the header of all Sequence List reports. All other data for the Sequence List
report is fixed. A default report header is supplied. You may enter a new report header of
up to 60 characters.
CEMASTER System 89
Message Editor
The Message Editor allows you to define up to 200 messages that may later be attached to
Auto-Pause controls on the test edit screens for display to an operator at run time. Use the
scroll bar control on the right side of the message grid to scroll through the 200 messages.
Each message may be up to 80 characters (maximum) in length. A set of default messages
is supplied for use with several optional components available for use with the
%'/#56'4. These default messages may be redefined if you desire.
To define a message, click within a message cell to highlight it (green highlight), then
click a second time to enter edit mode for the cell. Enter the desired message text then
press Enter.
To select messages for CUT, COPY, and PASTE operations, click the message number
cells on the left side of the message grid. This will highlight the message in yellow.
Multiple messages may be selected by holding the Ctrl key while clicking, or a block of
messages may be selected by holding the Shift key while clicking after the first message
in the block has been selected.
To deselect all messages, click the button in the upper left corner of the message grid.
Show Versions
CEMASTER System 90
Show Versions is a display of the version number of the %'9CTG application, plus the
version of all major software components the %'9CTG application uses. When
communicating remotely with a %'/#56'4, the firmware version plus the version of
all major firmware components of the %'/#56'4 is also shown.
HW Selection
This selection is necessary only when running %'9CTG in local mode (with simulated
hardware). The menu supports selection of test hardware type (%'/CUVGT or '/%2TQ) and also
of a second surge type (10/700 Telecom wave or Ring wave) if '/%2TQis selected. Selection
of this configuration information is automatic in Remote mode, with hardware connected.
HW Selection Screen
CEMASTER System 91
Enable
There are two selections under Enable: Remote Communications and Logging. Both of these
selections are ON/OFF selections that are indicated by a check mark when they are ON and no
check mark when they are OFF.
Remote Communications
This selection turns ON or OFF the remote (PC-based) control of the %'/CUVGT by the
%'9CTG application. The selection is disabled if the communications cannot be
established with a %'/CUVGT: (communications cable not connected or %'/CUVGT
power is OFF). When communication with a %'/CUVGT is established, this selection is
enabled and %'9CTG will automatically try to turn remote control ON.
The status of communications is shown on the Status Bar at the bottom of the %'9CTG
screen. When communication is not established the first item next to the
Communication: label says DISCONNECTED. When communication is established,
the label says CONNECTED.
Logging
This selection turns ON or OFF the logging feature for producing run time log reports. At
least one of the two Log report types must have been enabled in Setup | Log Reports to
enable logging or an error message will appear when you try to enable it. Logging may
only be set ON or OFF when the Main Screen is visible: Logging may not be turned ON
or OFF once a run is started.
The status of Logging is also shown on the Status Bar at the bottom of the %'9CTG
screen. When Logging is not enabled the item next to the Logging: label says OFF.
When Logging is enabled, the item says ON.
System
These are debugging and field service functions that do not need to be accessed by the user for
any normal operation.
CEMASTER System 92
Help
The Help menu brings up the Help file and shows the About box.
Toolbar
Toolbar
The toolbar holds icons for common tasks, allowing you to select that task with a single mouse-
click.
File Icons -- The file icons create a new test sequence, load an existing test sequence,
or save the current test sequence. If you attempt to create a new sequence or load an
existing sequence, you will be prompted to save any existing work.
Edit Icons -- The CUT, COPY, and PASTE icons allow you to edit the test sequence
grid. These icons are also found in the Sequence Editor.
The Printer icon will print the present sequence list, i.e. it is the same as the File Print
Status Bar
The status bar at the bottom of the screen gives a quick synopsis of the CEMASTER and test
status. In some cases, the status bar will be replaced with help prompts. It contains the following
information.
CEMASTER System 93
Communication
DISCONNECTED/CONNECTED
LOCAL/REMOTE
When communications with the %'/#56'4 are established, this indicates whether
control is from the front panel (LOCAL) or from the PC (REMOTE). The program will
automatically go Remote; select from the menu bar (EnableRemote Communications)
to change to simulation mode when a %'/#56'4 is connected and powered on.
Run State
Logging
Indicates if run time test information is being written to disk (ON) or not (OFF).
CEMASTER System 94
Maintenance
Annual Calibration
Periodic calibration of the %'/#56'4 is required. The recommended calibration interval is
dependent on use. If periodic waveform verification is performed, calibration need be
performed only if the waveform verification fails; if periodic waveform verification is not
performed the recommended calibration interval is one year. Factory calibration is available;
consult factory for terms and conditions.
Updating Firmware
There may be periodic updates of the %'/#56'4firmware or application software. Each
update is provided with instructions for installation and use. An update of the %'/#56'4
firmware requires the CM-SW option be installed. If this option is not installed a Thermo
KeyTek authorized service representative is required to upgrade the firmware in the field.
If you have questions about what is the latest versions of firmware, contact Customer Service.
The latest firmware revisions are also listed on the Thermo Voltek web page, at:
http://www.thermovoltek.com
Decommission Issues
There are no special issues in decommissioning %'/#56'4 equipment.
CEMASTER System 95
Troubleshooting
This section aids in finding and solving any problems you may experience. As a first step, the
%'/#56'4 includes diagnostic tools, accessed through the System Menu, and described
below.
System Menu
The System Menu gives access to several useful functions to aid in troubleshooting.
SYSTEM SETUP EUT OFF The System menu selects special system
1. Parameters
2. Access Code operations; move the highlight bar using
3. Diagnostics the arrow function keys and press ENTER,
4. Versions or press the number key associated with the
sub-menu.
EXIT
The Parameters Menu offers control of a number of system parameters, allowing customization
of the %'/#56'4 system. Check the menu for options available in your system; possible
options are: control of the beeper; default coil factor to use; etc.
The Access Code Menu is not available at this time. In the future, this menu will allow locking
features of the %'/#56'4system -- preventing unauthorized use.
The Versions Menu shows internal revisions for the code and databases used by the
%'/#56'4. This information is important and should be checked whenever contacting
Customer Service.
VERSIONS EUT OFF
Firmware ...... Vx.xx The Version Menu shows a Build date and
Kernel ...... Vx.xx
Sequence Database .... Vx.xx Sep 24, 96 time, five arguments with revisions, and a
Rule Database .... Vx.xx 16:15 code indicating the installed options.
String Database .... Vx.xx
Option Code .... AA
EXIT
CEMASTER System 96
The Diagnostics Menu performs a quick check of basic %'/#56'4 functions, and can often
localize a problem and allow a field service repair rather than a factory repair.
DIAGNOSTICS:TESTS EUT OFF
1. Line Sync 1 5. ESD The DIAGNOSTICS: TESTS menu
2. Line Sync 2 6. RELAY
3. High Volt selects special system operations; move the
4. Coil highlight bar using the arrow function keys
and press ENTER, or press the number key
EXIT associated with the test.
Before running any diagnostic, remove all connections from the front panel EUT connector, the
coaxial connector, and the safety socket connectors.
To use the diagnostic tests, follow the requirements from the list below, choose a test, and follow
any on-screen directions. If the test halts, record any message displayed and report the message to
Customer Service.
1. Line Sync 1 -- General Purpose check of Mains circuitry. Checks that Mains contactor is
operational, thermal breaker is operational, basic wiring OK, and line sync is operational.
Requires: EUT Mains present, EUT Switch ON, no EUT connected, no interlock fault.
2. Line Sync 2 -- Checks PQF circuitry. Checks PQF transformer and contactors are
operational, PQF switches are operational, and Aux line sync circuit is operational.
Requires: EUT Mains present, EUT Switch ON, no EUT connected, no interlock fault.
3. High Volt -- Checks high-voltage supply, energy storage capacitors, and discharge circuitry
CEMASTER System 97
Troubleshooting Guide
In the event of a problem with the %'/#56'4, please follow the following procedure.
CEMASTER System 98
Error Messages
The following are possible error messages, their causes, and possible solutions.
No AC Phase Reference Detected The Surge parameters specify line sync, and no
Check EUT Power switch and EUT AC mains was detected for synchronization.
power To surge without AC mains present: Change
the line sync setting from L1 to RND -
random.
To surge with AC mains: Check that EUT
power cord is plugged in, EUT power is
present, and the %'/#56'4 EUT switch is
on; check that EUT ON is shown on the front
panel display.
AC Required At Mains Output During a PQF test, AC was not present (PQF
Check EUT Power switch and EUT requires AC to be present).
power Check that EUT power cord is plugged in,
EUT power is present, and the %'/#56'4
EUT switch is on; check that EUT ON is
shown on the front panel display.
ESD Simulator Not Connected An ESD test is selected, and no ESD simulator
head was attached.
Attach an ESD head; check that the cable is
properly connected (remove ground-return
cable; if the ground-return indicator does not
light on the head, there is a poor connection).
CEMASTER System 99
Mains Fault: Disconnect and check EUT During a PQF test a high-current was detected,
If fault recurs contact Thermo KeyTek possibly due to a short in the EUT.
Service Disconnect the EUT and check it for possible
shorts line-to-line or line-to-chassis. Retest.
Test with a second EUT if possible.
If this message persists factory service is
required. Consult Customer Service for detail.
Unspecified Fault Record the fault number and retry the failed
#<fault number> test. If this fault persists, report it to Customer
Service.
Timeout on Remote Control: Check PC The %'/#56'4 while running under control
Application and Connection of a PC, lost communications with the PC. For
safety, a loss of communications halts all tests.
Check connections between the %'/#56'4
and the PC, and check that the PC application
is running.
If your CEMASTER is operational, access the SYSTEM menu, VERSIONS screen and copy all
the information shown to include in your trouble report.
VERSIONS EUT OFF
Firmware ...... Vx.xx The Version Menu shows a Build date and
Kernel ...... Vx.xx
Sequence Database .... Vx.xx Sep 24, 96 time, five arguments with revisions, and a
Rule Database .... Vx.xx 16:15 code indicating the installed options.
String Database .... Vx.xx
Option Code .... AA
EXIT
Customer Service
Thermo KeyTek
One Lowell Research Center
Lowell, MA 01852-4345 USA
Voice: 978-275-0800 FAX: 978-275-0850
http://www.thermokeytek.com
MC USA
U.S Headquarters
5560 Skylane Blvd.
Santa Rosa, CA 95403
http://www.multi-contact-usa.com