Running UPLC
Running UPLC
Running UPLC
Running UPLC
,HPLC UPLC Preparing to Run
.
.UPLC HPLC -
1
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
General Operating Recommendations CORNING PYREX Type 1, Class A, Borosilicate Bottles
SCHOTT DURAN Borosilicate Glass 3.3 Bottles
Solvent Filters
Mobile Phase Preparation
Filtration on membrane:
Actions : Solvent degassing and filtration
2
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Bottle Caps Startup the System According to Instructions
http://www.forumsci.co.il/HPLC/UPLC_setup_21-9_09_Empower.pdf
3
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Injector design examples
ACQUITY fixed loop injector design
Flow-Through variable injector design
(Rheodyne)
loop
0.010
0.000 1. Full Loop Injection Method/Mode as the injection technique using
-0.010 Pressure Assist
5.605.705.805.906.006.106.206.306.40
Minutes o Injects 100% of actual loop volume
to column and
detector
needle loop
Possible contaminated
area
VDD
BSM
to column and
Sample Syringe detector
Wash block loop
VDD
BSM
Strong needle wash
Sample Syringe followed by weak needle
wash
For more details on solvents, see the section titled Selecting weak
wash and strong wash solvents in the ACQUITY Operators Guide
5
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Strong Wash Solvent Strong Wash Solvent
If you dilute the samples, match the weak wash solvent to the
sample diluent
6
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
ACQUITY UPLC Sample Manager Full Loop Injection
Injection Methods Example of a 20 L injection with a 4X overfill
Overfill Factor 1 to 4
7
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Full Loop Injection Full Loop Injection
Step 3 Position sample and overfill the loop Step 4 Sample Injection
8
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Partial Loop Injection Partial Loop (Pressure Assist)
1 Not Recommended
Sample
2 Not Recommended
Weak Wash Weak Wash
Solvent Solvent 5 Not Recommended
10 1.0 5.0
20 2.0 10.0
Air Gap
50 5.0 25.0
Sample Loop
Volume Injected
9
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Pressure Assist Injection Sequence Pressure Assist Injection Sequence
Step 3 Position the sample Step 4 Injection valve injects sample
10
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Summary of Partial Loop
Partial Loop Using Needle Overfill
using Pressure Assist
BENEFITS:
Short Cycle Time
Sample volume is conserved
Recommended for large sample loop injections
TRADEOFF:
Accuracy and Precision are lower than Full Loop Mode
Performance dependency on weak wash solvent matching to mobile
phase
Accuracy is generally lower compared to Needle Overfill
Partial Loop using Needle Overfill Partial Loop using Needle Overfill
Step 1 Aspirate Sample and Air Gap Step 2 Injection Sequence
11
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Partial Loop using Needle Overfill Partial Loop using Needle Overfill
Step 3 Injection Sequence Step 4 Injection Sequence
12
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Partial Loop using Needle Overfill
Step 2 Injection Sequence Partial Loop using Needle Overfill
Step 3 Injection Sequence
Partial Loop using Needle Overfill Partial Loop using Needle Overfill
Step 4 Injection Sequence Step 5 Injection Sequence
13
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Summary of Partial Loop Injection Parameters
using Needle Overfill
No weak wash injected onto column
Injection Parameters
mobile phase and sample injected
Partial loop
Low draw speed for low volume
Pre and post air gap are essential
Automatic parameters
o Pre and post air gap of 1 L
o 100 L per minute
14
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Strong Contaminations Main Parameters Summary
Air Gap
Pre and Post Air Gap Influence: Example
Question:
4 4 1 140787 132577 96240 57406 136252 95577 45260 16142
1) Should I use air gap? 0 0 1 90362 85040 62005 42553 92886 64077 30046 12199
Diffusion effect 55.80% 55.90% 55.21% 34.90% 46.69% 49.16% 50.64% 32.32%
o Yes or No?
4 4 1 140787 132577 96240 57406 136252 95577 45260 16142
o Which one? 0 4 1 89291 84372 63720 40764 85074 61209 29754 11578
Diffusion effect 57.67% 57.13% 51.04% 40.83% 60.16% 56.15% 52.11% 39.42%
Pre air, post air or both?
4 4 1 140787 132577 96240 57406 136252 95577 45260 16142
2) What volume? 4 0 1 120864 113953 81763 52170 122771 85128 39046 14816
Diffusion effect 16.48% 16.34% 17.71% 10.04% 10.98% 12.27% 15.91% 8.95%
Question:
Should I use?
o Full loop
15
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Air Gap Effects
pre air gap is more critical than post air gap Air Gap
automatic
Large air gap volume can effect the chromatogram.
4,4
4,0
0,0
0,4
Pre air gap Post air gap Peak area Peak height % area difference
Higher values:
o Can effect the chromatogram
Higher peak height RSD
Loss of resolution
16
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Designing for UPLC ACQUITY UPLC TUV and PDA
Features
AU
0.030
A peak which does not have enough data points will be difficult to
integrate and therefore have irreproducible peak areas and 0.020
heights. 0.010
0.000
0.50 0.52 0.54 0.56 0.58 0.60 0.62 0.64 0.66 0.68 0.70 0.72 0.74
Minutes
17
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Effect of Sampling Rate on Reproducibility Tunable Ultra Violet Specifications
Sampling Points Peak Peak Area Peak Peak Wavelength Range 190 700 nm
Rate Across Area %RSD Height Height
Increased data acquisition rates
Peak %RSD
Maximum 80 pts per second in single wavelength mode
1 pt/s 2 44125 2.436 27451 15.515
Maximum 2 pts per second in dual wavelength mode
2 pts/s 4 32822 1.790 41207 13.455
Optimized time constant to filter smaller peak widths
5 pts/s 7 31554 0.971 67355 3.962
Two flow cells
10 pts/s 13 31321 1.129 73638 1.015 Analytical
0.080
No Filtering
MassLynx v4.0/4.1
80 pts/sec maximum data rate
0.070 0.1s
0, 1, 2, or 3 second time constants
0.3s
0.060 3D mode only
0.5s
0.050 1.0s
Empower build 1154/2154
0.040 80 pts/sec maximum data rate
AU
0.010
Two flow cells
Analytical
0.000 High Sensitivity
0.50 0.52 0.54 0.56 0.58 0.60 0.62 0.64 0.66 0.68 0.70 0.72 0.74 0.76 0.78 0.80
Minutes
18
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Conditions: Influence of Detector Settings:
Influence of Detector Settings: Sampling Rate Sampling Rate
0.05
Chromatographic Conditions : Analytes (Caffeine and Metabolites):
Column: ACQUITY BEH C18 2.1 x 30 mm, 1.7 m 1. 1-methylxanthene
AU
Mobile Phase A: 0.1% Formic Acid in H2O 2. 1,3-dimethyluric acid 40 Hz
3. Theobromine 0.00
Mobile Phase B: 0.1% Formic Acid in ACN
4. 1,7-dimethyluric acid 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50
Flow Rate: As indicated
5. 1,7-dimethylxanthene Minutes
Isocratic: 97% A: 3% B 6. Caffeine 0.05
Injection Volume: 2.0 L
Sample Diluent: 0.2% Formic Acid in water
AU
Strong Needle Wash: 50 L 50 ACN: 50 H2O
0.00 5 Hz
Weak Needle Wash: 500 L 0.1% Formic Acid in
H 2O
0.05
Temperature: 32C
AU
Detection: UV @ 280 nm
Time Constant: 0.1
0.00 1 Hz
Sampling rate: as indicated 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50
Instrument: UPLC: Waters ACQUITY UPLC, with Minutes
TUV detector
Influence of Sampling Rate on Peak Height Chromatographic Conditions : Analytes (Caffeine and Metabolites):
Column: ACQUITY BEH C18 2.1 x 30 mm, 1.7 m 1. 1-methylxanthene
Sampling Rate 40 32 20 10 5 2 1
Mobile Phase A: 0.1% Formic Acid in H2O 2. 1,3-dimethyluric acid
1-methylxanthene 27145 27284 27525 26815 26694 21370 13530 3. Theobromine
Mobile Phase B: 0.1% Formic Acid in ACN
1,3-dimethyluric acid 41173 41191 41869 41173 41221 36798 26356 4. 1,7-dimethyluric acid
Flow Rate: As indicated
theobromine 56726 56624 57589 56673 56578 51930 38274 5. 1,7-dimethylxanthene
Isocratic: 97% A : 3% B
1,7-dimethyluric acid 25182 25015 25501 24917 25009 23586 20093 6. Caffeine
Injection Volume: 2.0 L
1,7-dimethylxanthene 18910 16717 18311 16739 17912 17829 15984
caffeine 14405 14103 14024 14367 13901 14201 13529 Sample Diluent: 0.2% Formic Acid in water
Strong Needle Wash: 50 L 50 ACN : 50 H2O
Weak Needle Wash: 500 L 0.1% Formic Acid in
Influence of Sampling Rate on Resolution H 2O
Sampling Rate 40 32 20 10 5 2 1 Temperature: 32C
1-methylxanthene
1,3-dimethyluric acid 6.9 6.98 6.91 6.96 6.7 5.84 4.79
Detection: UV @ 280 nm
theobromine 1.9 1.86 1.93 1.87 1.87 1.78 1.44 Time Constant: as indicated
1,7-dimethyluric acid 7.81 7.75 7.71 7.84 7.67 7.23 6.73 Sampling rate: 20 pts/sec
1,7-dimethylxanthene 2.36 2.29 2.36 2.42 2.37 2.43 2.09 Instrument: UPLC: Waters ACQUITY UPLC, with
caffeine 16.29 16.39 16.27 16.14 16.24 15.98 15.5
TUV detector
19
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Influence of Detector Settings:
Time Constant Influence of Detector Settings:
Time Constant
1-methylxanthene
1,3-dimethyluric acid 6.91 6.64 5.24 3.53
0.00 Tc = 2.5 Theobromine 1.93 1.87 1.53 1.08
0.00 1.00 2.00 3.00 4.00
1,7-dimethyluric acid 7.71 7.58 6.79 5.19 3.68
Minutes
1,7-dimethylxanthene 2.36 2.37 2.18 1.8 1.45 1.22 1.02
Caffeine 16.27 16.5 16.19 15.09 13.47 12.18 10.87
Zooming In
The TIR condition requires that a component of the incident wave
A light guiding flow cell is
Light Path penetrates the AF, typically to a small fraction of the wavelength of
essentially an optical fiber
incident light.
whose core material is a fluid
The optical fiber consists of a
core material surrounded by a
cladding layer (Teflon AF).
Light is guided through the core TeflonAF Teflon AF, n(w)
by the process of total internal Mobile Phase
reflection;
light rays that encounter the TeflonAF
interface between the core and
cladding are reflected back into 90- Mobile Phase, n(f)
the core with essentially 100%
efficiency
20
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Why is Flow Cell The Median Baseline Filter (MBF)
Maintenance Important?
06-Oct-2005 11:09:15 grds wp1114.arw
0.2 MBF Off
Teflon AF light guiding
0.15 MBF On
flow cells may, with time,
0.1
develop absorbance
0.05
baselines that exhibit
0
curvature.
-0.05
0 0.5 1 1.5 2 2.5 3 3.5 4
This is attributed to
changes in optical
characteristics of AF
surface due to
contaminants.
A processing tool for alleviating baseline curvature
These changes effect TIR
efficiency. The MBF operates during acquisition in real time on the chromatogram
whose baseline is to be corrected.
Example of baseline curvature. The top plot is a chromatogram of a The MBF automatically extracts the baseline, smoothes it, and then
gradient separation. The bottom plot is zoomed vertically to show
more clearly the magnitude of the baseline curvature. subtracts the smoothed baseline from the original chromatogram.
Result is a chromatogram with greatly reduced baseline curvature.
25 mm Pathlength 10 mm Pathlength
2400 nL Volume 500 nL Volume
21
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
ACQUITY PDA
Flow Cell Options Flow Cell Comparison
UPLC Conditions
System: ACQUITY UPLC with TUV High Sensitivity Flow Cell
Column: 2.1 X 50 mm ACQUITY UPLC BEH C18, 1.7 m
Mobile Phase 7:3 Water/ACN + 0.1% Formic
Flow Rate: 0.65 mL/min
High Sensitivity Analytical Injection Volume: 2 L
Strong Wash = 75% ACN/25% Water (250 uL)
Weak Wash = 5% ACN/95%Water (1000 uL)
Sample Temperature: 10 C
Column Temperature: 35 C
Detection: ACQUITY TUV @ 231nm (0-1.7 min)
280 nm (1.7 3.0 min) 20 pps, FTC=0.30
Sample = 080 mg/mL2,4-D in 50% ACN/50% Water
Data: Empower 2
25 mm Pathlength 10 mm Pathlength
2400 nL Volume 500 nL Volume
0.22
0.18
0.0055
Area 2995 6845 2.29 Area 140955 326881 2.319
0.16
0.14
AU
0.0050 Peak Width at 13.4% 2.02 sec 2.14 sec 1.059 0.08
0.06
0.04
-0.02
-0.04
0.0035
0.12
AU
0.0030
AU
0.10
0.0025
0.08
0.0020
Noise:
0.0015 High Sensitivity Flow Cell = 0.000015 AU 0.06
Analytical Flow Cell = 0.000012 AU
0.0010
0.04
0.0005
0.02
0.0000
0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 0.00
Minutes
4.40 4.45 4.50 4.55 4.60 4.65 4.70 4.75 4.80 4.85 4.90
Minutes
22
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Flow Cell Recommendations ACQUITY UPLC
High Brightness Lamp
High brightness deuterium lamp
Analytical Flow Cell Low noise characteristics
Better chromatographic resolution Nearly twice the brightness (energy) of ordinary deuterium
lamps
High Sensitivity Flow Cell
Benefits
For the highest sensitivity
Less Noise
Higher peak height (Beers law) Higher sensitivity
In general, better signal to noise Lamp Warranty
o Slightly higher noise because more flow perturbations to contribute to 2000 hours
the noise, especially with absorbing mobile phases Lamp Firmware through Console
Both Cells Hours of operation
Same linearity Ignitions
Cover the full flow rate range
Serial Number
23
Dr. Shula Levin, Medtechnica
Preparing to Run UPLC; Injection and Detection Paramters
Recommendations Sources of Potential Contaminants
Wash Process
Bad storage could induce possible Dedicated to sample to decrease carry over
24
Dr. Shula Levin, Medtechnica