B. Spectrophotometer Cells, 1 Cm. C. Filtering Apparatus and Filter: See 2120B.2c
B. Spectrophotometer Cells, 1 Cm. C. Filtering Apparatus and Filter: See 2120B.2c
B. Spectrophotometer Cells, 1 Cm. C. Filtering Apparatus and Filter: See 2120B.2c
tristimulus (ordinate) values, preferably by using the weighted- facturers directions. Express results as prescribed in 2120C.6c
ordinate method.* Calibrate calculation algorithm software for both original and pH-adjusted samples.
against platinum-cobalt standard reference. Alternatively, obtain ADMI weighted-ordinate values for
b. Spectrophotometer cells, 1 cm. color by a published computation method.2
c. Filtering apparatus and filter: See 2120B.2c.
4. Quality Control
2130 TURBIDITY*
2130 A. Introduction
if good measurement techniques are used and the characteristics Its precision, sensitivity, and applicability over a wide
of the particles in the measured suspensions are similar. Poor turbidity range make the nephelometric method preferable to
measurement technique can have a greater effect on measure- visual methods. Report nephelometric measurement results as
ment error than small differences in instrument design. Turbi- nephelometric turbidity units (NTU).
dimeters of nonstandard design, such as forward-scattering de-
vices, may be more sensitive than nephelometers to the presence
of larger particles. While it may not be appropriate to compare 3. Storage of Sample
their output with that of instruments of standard design, they still
Determine turbidity as soon as possible after the sample is
may be useful for process monitoring.
An additional cause of discrepancies in turbidity analysis is taken. Gently agitate all samples before examination to ensure a
the use of suspensions of different types of particulate matter for representative measurement. Sample preservation is not practi-
instrument calibration. Like water samples, prepared suspen- cal; begin analysis promptly. Refrigerate or cool to 4C, to
sions have different optical properties depending on the particle minimize microbiological decomposition of solids, if storage is
size distributions, shapes, and refractive indices. A standard required. For best results, measure turbidity immediately without
reference suspension having reproducible light-scattering prop- altering the original sample conditions such as temperature or
erties is specified for nephelometer calibration. pH.
larger than 0.1 m;* the usual membrane filter used for bacte- turbidity of the standard after calibrating the instrument with a
riological examinations is not satisfactory. Rinse collecting flask fresh formazin or microsphere suspension. If there is any doubt
at least twice with filtered water and discard the next 200 mL. about the integrity or turbidity value of any secondary standard,
Some commercial bottled demineralized waters have a low check instrument calibration first with another secondary stan-
turbidity. These may be used when filtration is impractical or a dard and then, if necessary, with user-prepared formazin. Most
good grade of water is not available to filter in the laboratory. secondary standards have been carefully prepared by their man-
Check turbidity of bottled water to make sure it is lower than the ufacturer and should, if properly used, give good agreement with
level that can be achieved in the laboratory. formazin. Prepare formazin primary standard only as a last
b. Stock primary standard formazin suspension: resort. Proper application of secondary standards is specific for
1) Solution IDissolve 1.000 g hydrazine sulfate, each make and model of nephelometer. Not all secondary stan-
(NH2)2H2SO4, in distilled water and dilute to 100 mL in a dards have to be discarded when comparison with a primary
volumetric flask. CAUTION: Hydrazine sulfate is a carcinogen; standard shows that their turbidity value has changed. In some
avoid inhalation, ingestion, and skin contact. Formazin suspen- cases, the secondary standard should be simply relabeled with
sions can contain residual hydrazine sulfate. the new turbidity value. Always follow the manufacturers di-
2) Solution IIDissolve 10.00 g hexamethylenetetramine, rections.
(CH2)6N4, in distilled water and dilute to 100 mL in a volumetric
flask. 4. Procedure
3) In a flask, mix 5.0 mL Solution I and 5.0 mL Solution II.
Let stand for 24 h at 25 3C. This results in a 4000-NTU a. General measurement techniques: Proper measurement tech-
suspension. Transfer stock suspension to an amber glass or other niques are important in minimizing the effects of instrument vari-
UV-light-blocking bottle for storage. Make dilutions from this ables as well as stray light and air bubbles. Regardless of the
stock suspension. The stock suspension is stable for up to 1 year instrument used, the measurement will be more accurate, precise,
when properly stored. and repeatable if close attention is paid to proper measurement
c. Dilute turbidity suspensions: Dilute 4000 NTU primary techniques.
standard suspension with high-quality dilution water. Prepare Measure turbidity immediately to prevent temperature
immediately before use and discard after use. changes and particle flocculation and sedimentation from chang-
d. Secondary standards: Secondary standards are standards ing sample characteristics. If flocculation is apparent, break up
that the manufacturer (or an independent testing organization) aggregates by agitation. Avoid dilution whenever possible. Par-
has certified will give instrument calibration results equiva- ticles suspended in the original sample may dissolve or other-
lent (within certain limits) to the results obtained when the wise change characteristics when the temperature changes or
instrument is calibrated with the primary standard, i.e., user- when the sample is diluted.
prepared formazin. Various secondary standards are available Remove air or other entrained gases in the sample before
including: commercial stock suspensions of 4000 NTU measurement. Preferably degas even if no bubbles are visible.
formazin, commercial suspensions of microspheres of sty- Degas by applying a partial vacuum, adding a nonfoaming-type
rene-divinylbenzene copolymer, and items supplied by in- surfactant, using an ultrasonic bath, or applying heat. In some
strument manufacturers, such as sealed sample cells filled cases, two or more of these techniques may be combined for
with latex suspension or with metal oxide particles in a more effective bubble removal. For example, it may be neces-
polymer gel. The U.S. Environmental Protection Agency1 sary to combine addition of a surfactant with use of an ultrasonic
designates user-prepared formazin, commercial stock forma- bath for some severe conditions. Any of these techniques, if
zin suspensions, and commercial styrene-divinylbenzene sus- misapplied, can alter sample turbidity; use with care. If degas-
pensions as primary standards, and reserves the term sec- sing cannot be applied, bubble formation will be minimized if
ondary standard for the sealed standards mentioned above. the samples are maintained at the temperature and pressure of the
Secondary standards made with suspensions of microspheres water before sampling.
of styrene-divinylbenzene copolymer typically are as stable as Do not remove air bubbles by letting sample stand for a period of
concentrated formazin and are much more stable than diluted time because during standing, turbidity-causing particulates may
formazin. These suspensions can be instrument-specific; there- settle and sample temperature may change. Both of these conditions
fore, use only suspensions formulated for the type of nephelom- alter sample turbidity, resulting in a nonrepresentative measure-
eter being used. Secondary standards provided by the instrument ment.
manufacturer (sometimes called permanent standards) may be Condensation may occur on the outside surface of a sample
necessary to standardize some instruments before each reading cell when a cold sample is being measured in a warm, humid
and in other instruments only as a calibration check to determine environment. This interferes with turbidity measurement. Re-
when calibration with the primary standard is necessary. move all moisture from the outside of the sample cell before
All secondary standards, even so-called permanent stan- placing the cell in the instrument. If fogging recurs, let sample
dards, change with time. Replace them when their age exceeds warm slightly by letting it stand at room temperature or by
the shelf life. Deterioration can be detected by measuring the partially immersing it in a warm water bath for a short time.
Make sure samples are again well mixed.
b. Nephelometer calibration: Follow the manufacturers op-
erating instructions. Run at least one standard in each instrument
* Nuclepore Corp., 7035 Commerce Circle, Pleasanton, CA, or equivalent.
AMCO-AEPA-1 Standard, Advanced Polymer Systems, 3696 Haven Ave., range to be used. Make certain the nephelometer gives stable
Redwood City, CA, or equivalent. readings in all sensitivity ranges used. Follow techniques out-
ODOR (2150)/Introduction 2-11
lined in s 2b and 4a for care and handling of sample cells, ties and discrepancies in turbidity measurements make it
degassing, and dealing with condensation. unlikely that results can be duplicated to greater precision
c. Measurement of turbidity: Gently agitate sample. Wait than specified.
until air bubbles disappear and pour sample into cell. When
possible, pour well-mixed sample into cell and immerse it in 6. Reference
an ultrasonic bath for 1 to 2 s or apply vacuum degassing,
causing complete bubble release. Read turbidity directly from 1. U.S. ENVIRONMENTAL PROTECTION AGENCY. 1993. Methods for Deter-
instrument display. mination of Inorganic Substances in Environmental Samples. EPA-
d. Calibration of continuous turbidity monitors: Calibrate con- 600/R/93/100 - Draft. Environmental Monitoring Systems Lab., Cin-
tinuous turbidity monitors for low turbidities by determining tur- cinnati, Ohio.
bidity of the water flowing out of them, using a laboratory-model
nephelometer, or calibrate the instruments according to manufac- 7. Bibliography
turers instructions with formazin primary standard or appropriate
secondary standard. HACH, C.C., R.D. VANOUS & J.M. HEER. 1985. Understanding Turbidity
Measurement. Hach Co., Technical Information Ser., Booklet 11,
5. Interpretation of Results Loveland, Colo.
KATZ, E.L. 1986. The stability of turbidity in raw water and its
Report turbidity readings as follows: relationship to chlorine demand. J. Amer. Water Works Assoc.
78:72.
Report to the MCCOY, W.F. & B.H. OLSON. 1986. Relationship among turbidity,
Turbidity Range Nearest particle counts and bacteriological quality within water distribution
NTU NTU lines. Water Res. 20:1023.
BUCKLIN, K.E., G.A. MCFETERS & A. AMIRTHARAJAH. 1991. Penetration
01.0 0.05
of coliform through municipal drinking water filters. Water Res.
110 0.1
25:1013.
1040 1
HERNANDEZ, E., R.A. BAKER & P.C. CRANDALL. 1991. Model for evalu-
40100 5
ating turbidity in cloudy beverages. J. Food Sci. 56:747.
100400 10
HART, V.S., C.E. JOHNSON & R.D. LETTERMAN. 1992. An analysis of
4001000 50
low-level turbidity measurements. J. Amer. Water Works Assoc.,
1000 100
84(12):40.
LECHEVALLIER, M.W. & W.D. NORTON. 1992. Examining relationship
When comparing water treatment efficiencies, do not esti- between particle counts and Giardia, Cryptosporidium, and turbid-
mate turbidity more closely than specified above. Uncertain- ity. J. Amer. Water Works Assoc. 84(12):54.
2150 ODOR*
2150 A. Introduction