0% found this document useful (0 votes)
1K views140 pages

Analytical SFC System Service Manual

SFC Service manual

Uploaded by

Edwin Guevara R
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
1K views140 pages

Analytical SFC System Service Manual

SFC Service manual

Uploaded by

Edwin Guevara R
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 140

Agilent 1260 Infinity

Analytical SFC System

Service Manual

Agilent Technologies
Notices
Agilent Technologies, Inc. 2010 - 2012 Warranty receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
No part of this manual may be reproduced
The material contained in this docu- 1987). U.S. Government users will receive
in any form or by any means (including elec-
ment is provided as is, and is sub- no greater than Limited Rights as defined in
tronic storage and retrieval or translation
into a foreign language) without prior agree-
ject to being changed, without notice, FAR 52.227-14 (June 1987) or DFAR
ment and written consent from Agilent in future editions. Further, to the max- 252.227-7015 (b)(2) (November 1995), as
Technologies, Inc. as governed by United imum extent permitted by applicable applicable in any technical data.
States and international copyright laws. law, Agilent disclaims all warranties,
either express or implied, with regard Safety Notices
Manual Part Number to this manual and any information
contained herein, including but not CAUTION
G4301-90100 Rev. C limited to the implied warranties of
merchantability and fitness for a par- A CAUTION notice denotes a
Edition ticular purpose. Agilent shall not be hazard. It calls attention to an
05/2012 liable for errors or for incidental or
consequential damages in connection
operating procedure, practice, or
Printed in Germany the like that, if not correctly per-
with the furnishing, use, or perfor-
Agilent Technologies mance of this document or of any formed or adhered to, could
Hewlett-Packard-Strasse 8 information contained herein. Should result in damage to the product
76337 Waldbronn Agilent and the user have a separate
written agreement with warranty
or loss of important data. Do not
This product may be used as a com-
ponent of an in vitro diagnostic sys- terms covering the material in this proceed beyond a CAUTION
tem if the system is registered with document that conflict with these notice until the indicated condi-
the appropriate authorities and com- terms, the warranty terms in the sep- tions are fully understood and
plies with the relevant regulations. arate agreement shall control.
met.
Otherwise, it is intended only for gen-
eral laboratory use. Technology Licenses
The hardware and/or software described in WA R N I N G
this document are furnished under a license
SFC Fusion and Aurora SFC Fusion are and may be used or copied only in accor- A WARNING notice denotes a
trademarks of Aurora SFC Systems, Inc. All dance with the terms of such license. hazard. It calls attention to an
rights reserved.
Restricted Rights Legend operating procedure, practice,
The Agilent 1260 Infinity Analytical SFC
System includes a special designed Agilent If software is for use in the performance of a
or the like that, if not correctly
1260 Infinity system including a G4301 A U.S. Government prime contract or subcon- performed or adhered to, could
Aurora SFC Fusion A5 module. tract, Software is delivered and licensed as result in personal injury or
Commercial computer software as
defined in DFAR 252.227-7014 (June 1995),
death. Do not proceed beyond a
or as a commercial item as defined in FAR WARNING notice until the indi-
2.101(a) or as Restricted computer soft- cated conditions are fully under-
ware as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation stood and met.
or contract clause. Use, duplication or dis-
closure of Software is subject to Agilent
Technologies standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will

1260 Infinity Analytical SFC System SVC Manual


In This Book

In This Book

This Manual describes the Agilent 1260 Infinity SFC System.

1 Introduction to Supercritical Fluid Chromatography (SFC)


This chapter provides an overview of the history, theory and benefits of SFC.

2 Site Requirements and Specifications


This chapter provides information on environmental requirements, physical
and performance specifications only for the G4309 Agilent 1260 Infinity
Analytical SFC System.

3 Installing the G4309A Agilent 1260 Infinity SFC System


This chapter provides an overview of the installation and setup of the
hardware and software

4 Configuring the System


How to configure the Agilent 1260 Infinity SFC Analytical system and G4301 A
Aurora SFC Fusion A5 in ChemStation with Aurora add-on drivers.

5 Using the Aurora SFC Fusion A5


This chapter provides information and hints on the use of the SFC System.

6 Maintenance and Repair


In this chapter only the SFC specific procedures are described. For procedures
similar to the Agilent module procedures, please refer to the single module
manuals (G1312C, G1329B, G1316C, G1315/65C, G1322A)

7 Parts for Maintenance


This chapter provides information on parts for maintenance and repair.

1260 Infinity Analytical SFC System SVC Manual 3


In This Book

8 Identifying Cables
This chapter provides information on cables used with the Agilent 1200
Infinity Series modules.

9 Appendix
This chapter provides addition information on safety, legal and web.

4 1260 Infinity Analytical SFC System SVC Manual


Contents

Contents

1 Introduction to Supercritical Fluid Chromatography (SFC) 7


History of SFC 8
Theory of SFC 9
Benefits of SFC 10
Common flow path overview for packed column SFC instrumentation 11
The Agilent 1260 Infinity Analytical SFC System 12

2 Site Requirements and Specifications 17


Side Requirements 18
Specifications 22

3 Installing the G4309A Agilent 1260 Infinity SFC System 33


Hardware Installation 34
Software Installation 68

4 Configuring the System 71


Configuring SFC interface in ChemStation 72
Setting up the Method 76
Status 78
Control 79
Testing the System 80

5 Using the Aurora SFC Fusion A5 81


Powering up the Module 82
Power-up Sequence and Operational Control States 83
Operational Control States 84
Controlling the Aurora SFC Fusion A5 through the Agilent ChemStation 87
Running a method on the SFC system 89
Shutting Down the SFC System 92

1260 Infinity Analytical SFC System SVC Manual 5


Contents

6 Maintenance and Repair 95


Inspection and Preventative Maintenance Intervals 96
General Maintenance procedures 99

7 Parts for Maintenance 111


G4301 A Aurora SFC Fusion A5 Parts 112
Agilent Module Parts 114

8 Identifying Cables 119


A5 Cables 120
Overview 121
BCD Cables 123
External Contact Cable 125
CAN/LAN Cables 126
Auxiliary Cable 127
RS-232 Cables 128

9 Appendix 129
General Safety Information 130
The Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC) 133
Radio Interference 134
Sound Emission 135
Solvent Information 136
Agilent Technologies on Internet 137

6 1260 Infinity Analytical SFC System SVC Manual


1260 Infinity Analytical SFC System SVC Manual

1
Introduction to Supercritical Fluid
Chromatography (SFC)
History of SFC 8
Theory of SFC 9
Benefits of SFC 10
Common flow path overview for packed column SFC instrumentation 11
The Agilent 1260 Infinity Analytical SFC System 12
The SFC Fusion A5 (G4301A) 13
HPLC-SFC binary pump (G4302A) 13
SFC-Autosampler (G4303A) 14
The Column Compartment (G1316C) 14
UV-detection (DAD G1315C and MWD G1365C) 15
Applications 15
Columns 16

This chapter provides an overview of the history, theory and benefits of SFC.

Agilent Technologies 7
1 Introduction to Supercritical Fluid Chromatography (SFC)
History of SFC

History of SFC

Supercritical fluid chromatography (SFC) was first introduced by Klesper et


al. in 1962 (Klesper, E.; Corwin, A. H.; Turner, D. A. J. Org. Chem. 1962,
27,700.) for the separation and analysis of a porphyrin mixture using open
tubular SFC. The first commercial instruments using packed columns were
available from Hewlett-Packard (HP) in 1982. Since then, several vendors have
developed and commercialized packed column SFC instrumentation for
analytical as well as for preparative separation. SFC is widely accepted for the
separation of chiral compounds and increased user interest has been observed
for a wide spectrum of small to medium sized molecules due to the analysis
speed achieved and the low solvent consumption. The latest introduction of
analytical SFC instrumentation is based on the Aurora A5TM Fusion module
from Aurora (founded by Terry Berger, one of the SFC pioneers at HP) coupled
to an Agilent 1260 Infinity Rapid Resolution LC system optimized for SFC.

8 1260 Infinity Analytical SFC System SVC Manual


Introduction to Supercritical Fluid Chromatography (SFC) 1
Theory of SFC

Theory of SFC

EgZhhjgZ

HjeZgXg^i^XVa
;aj^YGZ\^dc

8g^i^XVa A^fj^Y

EgZhhjgZ
8g^i^XVaEd^ci

Hda^Y

<Vh

8g^i^XVa IZbeZgVijgZ
IZbeZgVijgZ
Figure 1 State of a solvent

The superior separation properties achieved by SFC can be explained best by


the thermodynamics of liquids and gases (see the phase diagram in Figure 1
on page 9). Above a critical pressure, liquids can no longer enter the gaseous
state; similarly, above a critical temperature, gases cannot be converted to
liquids. Above both the critical pressure and temperature (characterized by
the critical point), solvents are in the supercritical state. Under these
conditions, the mobile phases exhibit gaseous as well as liquid-like properties.
The major advantages of this state related to chromatography are improved
diffusion characteristics and mass transfer and low viscosity, which result
in high separation efficiency and fast separation capability.

1260 Infinity Analytical SFC System SVC Manual 9


1 Introduction to Supercritical Fluid Chromatography (SFC)
Benefits of SFC

Benefits of SFC

SFC is widely accepted for the analysis and separation of chiral compounds. In
addition, it gains increasing acceptance as a complementary liquid-based
separation technique to HPLC for high-throughput and high-resolution
analysis of complex mixtures. This is due to the thermodynamic properties of
supercritical fluids, which can be exploited for high throughput and high
efficiency. In addition, the mild thermal conditions also allow for the analysis
of thermally labile compounds. Typically, analysis times and column
re-equilibration are decreased by a factor of 35 compared to standard HPLC.
With the increasing costs of organic solvents and the environmental
awareness to minimize toxic waste, production SFC is increasingly accepted
as the green alternative to normal phase or reversed phase chromatography,
gaining popularity in method development and UV- and MS-based separation
and purification. A variety of parameters, such as stationary phase selection,
mobile phase composition, modifier type and concentration, column
temperature and system pressure, can be easily manipulated to fulfill
separation requirements by influencing, optimizing and exploiting selectivity
in SFC.

10 1260 Infinity Analytical SFC System SVC Manual


Introduction to Supercritical Fluid Chromatography (SFC) 1
Common flow path overview for packed column SFC instrumentation

Common flow path overview for packed column SFC


instrumentation

In commercially available SFC systems, CO2 is initially pumped in liquid state


and is brought into the supercritical state by heating it above the critical
temperature before it enters the high-pressure area of the LC instrument.
After high-pressure mixing with a modifier, the mobile phase passes through
the injection loop, where the sample is introduced into the supercritical
stream, and further transported to the separation column. The high pressure
of the mobile phase must be maintained downstream of the detector outlet
using a backpressure regulator to keep the mobile phase in its supercritical
condition over the complete flowpath.

1260 Infinity Analytical SFC System SVC Manual 11


1 Introduction to Supercritical Fluid Chromatography (SFC)
The Agilent 1260 Infinity Analytical SFC System

The Agilent 1260 Infinity Analytical SFC System

The Agilent 1260 Infinity Analytical SFC System (G4309A) consists of a binary
HPLC-SFC pump (G4302A), a degasser (G1322A), an SFC Autosampler
(G4303A), a thermostatted column compartment (G1316C), and either a DAD
(G1315C) or MWD (G1365C) equipped with a high pressure SFC flow cell and
the SFC Fusion A5 module (G4301A). The SFC Fusion A5 module,
redestills and preconditions the CO2 by boosting the pressure to just below the
column head pressure, relieving the HPLC-SFC pump of any compression
requirements. This results in low detector noise and significantly higher
sensitivity. Therefore, the Agilent HPLC-SFC pump receives pre-conditioned
CO2, and acts only as metering device for the mobile phase flow and to form
the gradient with the second pump head by adding the appropriate amount of
modifier solvent. Downstream of the detector, the mobile phase is redirected
back into the SFC Fusion A5 to an integrated back pressure regulator that
maintains the backpressure over the system. The Agilent 1260 Infinity SFC
system with the Aurora SFC Fusion A5 module is completely controlled by
Agilent ChemStation software. The system diagram is shown below:

>cejiIZZ
LVh]ejbe [gdbYZiZXidg

7EG id6AH

[gdblVh]hdakZciWdiiaZ

idlVhiZ
id6\^aZciH;8ejbe
DjiejiIZZ

Figure 2 Agilent 1260 Infinity Analytical SFC System

12 1260 Infinity Analytical SFC System SVC Manual


Introduction to Supercritical Fluid Chromatography (SFC) 1
The Agilent 1260 Infinity Analytical SFC System

The SFC Fusion A5 (G4301A)


The SFC Fusion A5 is responsible for all tasks connected to pre- and
post-conditioning of the mobile phase. In contrast, flow rate, mobile phase
composition, detection, column temperature and data analysis are controlled
by the modules of the Agilent 1260 Infinity SFC system in combination with
the ChemStation software. This includes metering the carbon dioxide flow and
mixing the modifier into the mobile phase by the HPLC-SFC binary pump.
In detail, the SFC Fusion A5 uses vapor-phase carbon dioxide, redestills it to
the liquid state and boosts its pressure to just under the column head
pressure. Since the CO2 gas is a very poor solvent, most contaminants in the
carbon dioxide are left in the source, which allows for the use of inexpensive,
beverage-grade CO2, unlike in any other commercially available instrument.
SFC Fusion A5 further recollects the effluent from the UV (or other)
detector and controls the backpressure up to 400 bar over the complete
system. In addition, SFC Fusion A5 delivers the wash solvent to flush the
fixed (sample) loop of the SFC Autosampler.

HPLC-SFC binary pump (G4302A)


The HPLC-SFC binary pump is equipped with passive inlet valves and with
special seals and pistons to allow for CO2 pumping in channel A while channel
B adds organic modifier for either isocratic or gradient performance.
Pumphead B is also equipped with a purge valve to allow for quick changeover
of the organic modifier.

1260 Infinity Analytical SFC System SVC Manual 13


1 Introduction to Supercritical Fluid Chromatography (SFC)
The Agilent 1260 Infinity Analytical SFC System

SFC-Autosampler (G4303A)
In SFC, the complete solvent flow path needs to be pressurized under all
conditions to avoid expansion of the supercritical fluid. This excludes the use
of a variable injection loop design that is generally used in other Agilent
autosamplers. Therefore, the autosampler used in the Agilent 1260 Infinity
analytical SFC system has been converted to a fixed loop injector containing a
5 L sample loop.
The injection loop of the autosampler is installed between two ports of the 2
position/6 port injection valve. The total delay volume of the injector is about
3.3 L. As with any fixed loop injector, overfill of sample is necessary to inject
5 L reproducibly. To completely fill the sample loop, an excess of sample is
required. This is about 3 loop volumes (15 L of the installed loop) of sample
to achieve 95 % of the maximum loop volume. For smaller injection volumes,
the loop needs to be partially filled by sandwiching the sample between two
air bubbles, one on either side, followed by a plug of modifier or other solvent
behind the sample. Default methods for full loop filling are provided, and
should be used as initial injection conditions. The default methods are
automatically loaded in ChemStation.

The Column Compartment (G1316C)


The temperature of the mobile phase prior to detection is a critical parameter
for minimizing baseline noise recorded in the detector flow cell. The heat
exchanger on the right side of the column compartment is used to pre-heat the
mobile phase before it enters the column, indirectly heating the column. The
heat exchanger on the left side is used to change the temperature of the mobile
phase to achieve minimum noise. This is of crucial importance, since the
refractive index of carbon dioxide responds up to 50 times stronger than
water-based mobile phases; thus, even small changes in temperature can
significantly affect noise levels. Temperature changes of mobile phase with the
left side heat exchanger e.g. between 38 C and 49 C resulted in a variation in
peak-to-peak noise of over an order of magnitude.

14 1260 Infinity Analytical SFC System SVC Manual


Introduction to Supercritical Fluid Chromatography (SFC) 1
The Agilent 1260 Infinity Analytical SFC System

UV-detection (DAD G1315C and MWD G1365C)


The system can be equipped either with a DAD (G1315C) or an MWD (G1365C)
using a high pressure detector flow cell suitable for SFC (10 mm path length,
13 L volume), with short transfer tubing to minimize peak broadening.
Electronic temperature control provides highest baseline stability and stable
sensitivity values under fluctuating temperature and humidity conditions.
This feature aids greatly in minimizing detector noise, and now enables
impurity and EE (enantiomeric excess) analysis by SFC, particularly when
using elevated temperatures.

Applications
SFC has gained a wide interest and acceptance in many small molecule
applications because of its high separation speed and efficiency, selectivity,
low operating costs, and due to low generation of organic solvent waste.
Important applications have been developed for the analysis of
pharmaceutical drugs, natural products, fatty acids, vitamins, pesticides,
lipids and chiral compounds. See Figure 3 on page 16.

1260 Infinity Analytical SFC System SVC Manual 15


1 Introduction to Supercritical Fluid Chromatography (SFC)
The Agilent 1260 Infinity Analytical SFC System

Columns
In contrast to reversed phase separation, there is no universal stationary
phase available for SFC separations. Most typically used stationary phases are
ethyl pyridine, diol, cyano, amino, Silica and SCX columns. This usually leads
to additional effort to screen different columns in order to achieve optimum
separation. On the other hand, it provides a valuable tool for achieving
different selectivities for a given analyte mixture. A review of column
developments for SFC was recently published by T.Berger, B.Berger &
R.E.Majors in LCGC North America, May 1, 2010

L]ZgZH;8;^ih>c/[gdbcdc"edaVgid]^\]an"X]Vg\ZY
6aXd]dah >cdg\Vc^X
:i]Zg 6cVa^cZh 6X^YhEg^bVgn ^dchAVg\Z
EZciVcZ :hiZgh 6b^YZh 6a^e]Vi^XVb^cZh6be]diZg^Xh EZei^YZh egdiZ^ch9C6$GC6

H;8/ejgZ8D '

H;8/ejgZ8D' bdY^[^Zg

H;8/ejgZ8D' bdY^[^Zg VYY^i^kZh

H;8/ejgZ8D' bdY^[^Zg VYY^i^kZh lViZg

>dceV^g^c\

GZkZghZYE]VhZ=EA8

CdgbVaE]VhZ=EA8 >dcX]gdbVid\g\gV]n

>dcZmX]Vc\Z

Figure 3 Where SFC Fits In: from non-polar to highly-charged

16 1260 Infinity Analytical SFC System SVC Manual


1260 Infinity Analytical SFC System SVC Manual

2
Site Requirements and Specifications
Side Requirements 18
Power Considerations 18
Power Cords 19
Bench space 20
Environment 21
Ventilation 21
Specifications 22
System Specifications 22
Specifications 22
Agilent 1260 Infinity System 26

This chapter provides information on environmental requirements, physical and


performance specifications only for the G4309 Agilent 1260 Infinity Analytical
SFC System.

Agilent Technologies 17
2 Site Requirements and Specifications
Side Requirements

Side Requirements

A suitable environment is important to ensure optimal performance of the


instrument.

Power Considerations
The module power supply has wide ranging capability. It accepts any line
voltage in the range described in Table 6 on page 27. Consequently there is no
voltage selector in the rear of the module. There are also no externally
accessible fuses, because automatic electronic fuses are implemented in the
power supply.

Hazard of electrical shock or damage of your instrumentation


WA R N I N G
can result, if the devices are connected to a line voltage higher than specified.
Connect your instrument to the specified line voltage only.

The module is partially energized when switched off, as long as the power cord is
WA R N I N G
plugged in.
Repair work at the module can lead to personal injuries, e.g. electrical shock, when
the cover is opened and the module is connected to power.
Always unplug the power cable before opening the cover.
Do not connect the power cable to the instrument while the covers are removed.

18 1260 Infinity Analytical SFC System SVC Manual


Site Requirements and Specifications 2
Side Requirements

Inaccessible power plug.


CAUTION
In case of emergency it must be possible to disconnect the instrument from the power
line at any time.
Make sure the power connector of the instrument can be easily reached and
unplugged.
Provide sufficient space behind the power socket of the instrument to unplug the
cable.

Power Cords
Different power cords are offered as options with the module. The female end
of all power cords is identical. It plugs into the power-input socket at the rear
of the module. The male end of each power cord is different and designed to
match the wall socket of a particular country or region.

Absence of ground connection or use of unspecified power cord


WA R N I N G
The absence of ground connection or the use of unspecified power cord can lead to
electric shock or short circuit.
Never operate your instrumentation from a power outlet that has no ground
connection.
Never use a power cord other than the Agilent Technologies power cord designed
for your region.

Use of unsupplied cables


WA R N I N G
Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.

1260 Infinity Analytical SFC System SVC Manual 19


2 Site Requirements and Specifications
Side Requirements

Bench space
G4301 Aurora SFC Fusion A5 module
The Aurora SFC Fusion A5 Module requires approximately 1 foot of linear
bench space immediately adjacent to the target Agilent 1260 Infinity system
stack. Approximately 5 inches of free space is required behind the instrument
for cable access and adequate air flow for ventilation. Similar access to the
rear of the 1260 Infinity system is also required to install cables and interface
cards. As mentioned earlier, for optimal performance, the rear air space
should not be heated significantly above room temperature by the exhaust of
other instrumentation in the lab; rather, hot exhaust should be vented or
directed upward from the instrument.
The module is designed to be installed on either side of the 1260 Infinity
system stack with sufficient high pressure transfer tubing to attach to a
double-stacked system. If the 1260 Infinity system is attached to a split-flow
detector such as Mass Spec or ELSD, the Aurora SFC Fusion A5 module
should be positioned on the opposite side of the stack. Shelves overhanging
the Aurora SFC Fusion A5 module should provide a minimum of 6 inches of
clearance to allow access to the rear power switch. Finally, the PC system
interface to the Aurora SFC Fusion A5 module is USB 2.0. A six-foot cable is
supplied with the system. The CPU must be placed within range of this cable.
Alternately, the user may supply an extended length USB cable not to exceed
16 feet.
While the Aurora SFC Fusion A5 module can exist on either side of the 1260
Infinity system, it is often easier to locate it on the left side. For more
specification details see Aurora SFC Fusion A5 on page 22

Agilent modules
The dimensions and weight of your module (see Agilent 1260 Infinity
System on page 26) allow it to be placed on almost any laboratory bench. It
needs an additional 2.5 cm (1.0 inches) of space on either side and
approximately 8 cm (3.1 inches) at the rear for the circulation of air and
electric connections.
The module should be operated in a horizontal position.

20 1260 Infinity Analytical SFC System SVC Manual


Site Requirements and Specifications 2
Side Requirements

If a Thermostatted Autosampler is installed, an additional 25 cm (10 inches)of


space on either side for the circulation of air, and approximately 8 cm
(3.1 inches) at the rear is required for electrical connections.
If a complete 1260 Infinity system is to be installed on the bench, make sure
that the bench is designed to carry the weight of all the modules. For a system
including the Thermostatted Autosampler it is recommended to position the
modules in two stacks.

Environment
Condensation within the module
CAUTION
Condensation will damage the system electronics.
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.

Ventilation

Waste tube has to be connected to hood or vent


WA R N I N G
The effluent from a supercritical fluid chromatograph may contain vaporized, toxic
solvents. Never vent into an enclosed, occupied space. Always vent into a fume
hood or vent to the outside.

1260 Infinity Analytical SFC System SVC Manual 21


2 Site Requirements and Specifications
Specifications

Specifications

System Specifications
Type Specifications

Flow range 0.1 mL/min to 5 mL/min (settable),


1 mL/min to 5 mL/min (recommended)

Maximum operating pressure 600 bar

Upgrade possibility of exisiting Yes


1100/1200/1260 LC

Option for SFC/UHPLC in one system Yes

Unattended operation Leak sensors, diagnostic software features

Aurora SFC Fusion A5


Select the laboratory bench space before your system arrives. Pay special
attention to the total height requirements. Avoid bench space with
overhanging shelves. Pay special attention to the total weight of the modules
and solvents you have in addition to the Aurora SFC Fusion A5. Make sure
that your laboratory bench can support this weight.

Personal injury
WA R N I N G
The G4301 A Aurora SFC Fusion A5 module is heavy.
Enlist the aid of a co-worker to share the lifting load to avoid personal injury.

Table 1 Physical Specifications for the G4301 A Aurora SFC Fusion A5

Type Specification

Weight 26 kg
56 lbs

Dimensions (height width depth) 60 cm x 26 cm x 48 cm


23 in x 10 in x 18 in

22 1260 Infinity Analytical SFC System SVC Manual


Site Requirements and Specifications 2
Specifications

Table 1 Physical Specifications for the G4301 A Aurora SFC Fusion A5

Type Specification

Line voltage 100 - 240 VAC, 10 %

Line frequency 50 - 60 Hz, 5 %

Power consumption 700 VA Max

Operating temperature 15 - 30 C

Non-operating temperature -40 - 70 C

Humidity <95 %, at 25 - 40 C Non-condensing

Laboratory ventilation minimum 6 air exchanges/hr for lab air; CO2 monitor
recommended w/ alarm @ 5000 ppm

Exhaust vent capacity >20 L/min with sustained negative pressure

Operating altitude up to 2000 m


6500 ft

Non-operating altitude up to 4600 m


14950 ft

Safety standards IEC, NRTL

The Fusion module is heavy (approximately 26 kg or 56 lbs). Enlist the aid of a


co-worker to share the lifting load in order to avoid possible injury. It should
be positioned on a sturdy bench capable of holding the total weight of the
Fusion module plus the Agilent 1260 Infinity system.

Table 2 Chemical Specifications for the G4301 A Aurora SFC Fusion A5

Type Specifications

Inlet CO2 bulk purity >99.99 % vapor; >99.999 % liquid

Inlet CO2 phase vapor from non-dip-tube high pressure cylinder; liquid
from commercial CO2 delivery system

Inlet CO2 supply pressure 40 - 70 bar


580 - 1000 psi

Inlet CO2 temperature 15 - 30 C

1260 Infinity Analytical SFC System SVC Manual 23


2 Site Requirements and Specifications
Specifications

Table 2 Chemical Specifications for the G4301 A Aurora SFC Fusion A5

Type Specifications

Wash solvent HPLC grade alcohol

Liquid coolant 30 % propylene glycol in deionized water; proprietary


antioxidants; red dye added for safety

Coolant volume < 280 mL

Table 3 Wetted Materials Specifications for the G4301 A Aurora SFC Fusion A5

Type Specifications

High pressure flow path 300 and 400 series stainless steel PEEK, carbon filled
PEEK PTFE, PTFE, FEP, CTFE UHMW PE ruby, sapphire,
ceramic

Low pressure flow paths [waste, wash 316 stainless steel PEEK PTFE, PTFE, FEP, CTFE CPE;
pump, leak tray] LDPE Tygon PVC

Vapor exhaust Tygon PVC

Performance Specifications
Table 4 Performance Specifications of the G4301 A Aurora SFC Fusion A5

Type Specifications

Hydraulic system Single piston with proprietary motor control

Total hydraulic volume <5 mL @ pressure <70 bar


<25 mL @ pressure up to 400 bar

Chiller system Thermoelectric cooling with secondary air/liquid


cooling circuit

Back Pressure Regulation (BPR) system Low volume diaphragm type with proprietary drive
control; replaceable BPR head assy; No recalibration
required after head replacement

Chiller temperature -20 9 C

Booster pump speed range 0 6000 steps/sec average step rate

Booster pump pressure range 100 400 bar up to 5 mL/min demand

24 1260 Infinity Analytical SFC System SVC Manual


Site Requirements and Specifications 2
Specifications

Table 4 Performance Specifications of the G4301 A Aurora SFC Fusion A5

Type Specifications

Pressure pulsation <2 % amplitude at pump speed >300 steps/sec and


outlet pressure >100 bar

BPR thermal range 40


- 70 C
104 158 F

BPR thermal precision 1 C

BPR pressure range 100 400 bar

Backpressure accuracy 1 %

Backpressure precision 0.5 bar (0.2 bar typical)

Backpressure thermal precision 1 C

Control and data evaluation Agilent ChemStation for LC with SFC Fusion A5 driver;
Aurora A5 Diagnostic Program

Analog in pressure monitoring 1 V FS; one input; range set by calibration to host pump

Communications USB 2.0; APG Remote: ready, start, stop and


shut-down signals; relay contact closure [wash pump
only]

1260 Infinity Analytical SFC System SVC Manual 25


2 Site Requirements and Specifications
Specifications

Agilent 1260 Infinity System


Unspecified Conditions
WA R N I N G
Operating the instrumentation under conditions other than its intended use might
result in a potential safety hazard or might damage the instrumentation.
Never operate your instrumentation under conditions other than those specified by
the vendor.

G1322A Vacuum Degasser


Table 5 Physical Specifications of the G1322A Vacuum Degasser

Type Specification Comments

Weight 7 kg (15.4 lbs)

Dimensions 345 435 80 mm


(width depth height) (13.5 17 3.1 inches)

Line voltage 100 120 or 220 240 VAC, 10 % Wide-ranging


capability

Line frequency 50 or 60 Hz, 5 %

Power consumption 30 W Maximum

Ambient operating 0 55 C (32 131 F)1


temperature
Ambient non-operating -40 70 C (-4 158 F)
temperature

Humidity <95 %, at -25 40 C (-77 104 F) Non-condensing

Operating altitude Up to 2000 m (6500 ft)

Non-operating altitude Up to 4600 m (14950 ft) For storing the


instrument

Safety standards: IEC, CSA, Installation Category II, Pollution Degree 2


UL

1
This temperature range represents the technical specifications for this instrument. The tempera-
tures mentioned may not be suitable for all applications and all types of solvent.

26 1260 Infinity Analytical SFC System SVC Manual


Site Requirements and Specifications 2
Specifications

The G1322 Vacuum Degasser has been tested for evaporation of solvents into the
NOTE atmosphere by an independent institute with approved methods. The tests were performed
with Methanol (BIA Nr. 7810) and Acetonitrile (NIOSH, Nr. 1606). Evaporation of these
solvents into the atmosphere when operating the degasser was below the limits of
detection.

G1315C/D Diode Array Detector and MWD G1365C


Table 6 Physical Specifications

Type Specification Comments

Weight 11.5 kg (26 lbs)

Dimensions (height 140 x 345 x 435 mm (5.5 x 13.5 x 17 inches)


width depth)

Line voltage 100 240 VAC, 10 % Wide-ranging capability

Line frequency 50 or 60 Hz, 5 %

Power consumption 160 VA / 160 W / 546 BTU Maximum

Ambient operating 055 C (32131 F)


temperature

Ambient non-operating -40 70 C (-4 158 F)


temperature

Humidity < 95 %, at 25 40 C (77 104 F) Non-condensing

Operating altitude Up to 2000 m (6562 ft)

Non-operating altitude Up to 4600 m (15091 ft) For storing the module

Safety standards: IEC, CSA, Installation category II, Pollution degree 2 For indoor use only.
UL

ASTM: Standard Practice for Testing Variable Wavelength Photometric Detectors Used in
NOTE Liquid Chromatography.
Reference conditions: cell path length 10 mm, time constant 1 s (equal to response time
2 s), flow 1 mL/min LC-grade Methanol, slit width 4 mm.
Linearity measured with caffeine at 265 mm.

1260 Infinity Analytical SFC System SVC Manual 27


2 Site Requirements and Specifications
Specifications

Table 7 Performance Specifications

Type Specification

Wave length range 190 950 nm

Wavelenghth accuracy 1 nm

Linearity range 2.0 AU at 265 nm

Noise, wet 0.05 mAU under SFC conditions at 254 nm, 20 %


MetOH/80 % CO2
0.007 mAU under LC condition

Drift 0.9 mAU/h

Data rate 80 Hz

Cell pressure limit 400 bar

Cell path length 10 mm

Cell volume 13 L

28 1260 Infinity Analytical SFC System SVC Manual


Site Requirements and Specifications 2
Specifications

G4303A SFC Autosampler


Table 8 Physical Specifications

Type Specification Comments

Weight 15.5 kg (34.2 lbs)

Dimensions (height 200 x 345 x 440 mm (8 x 13.5 x 17 inches)


width depth)

Line voltage 100 240 VAC, 10 % Wide-ranging capability

Line frequency 50 or 60 Hz, 5 %

Power consumption 300 VA / 200 W / 683 BTU Maximum

Ambient operating 455 C (41131 F)


temperature

Ambient non-operating -40 70 C (-4 158 F)


temperature

Humidity < 95 %, at 25 40 C (77 104 F) Non-condensing

Operating altitude Up to 2000 m (6562 ft)

Non-operating altitude Up to 4600 m (15091 ft) For storing the module

Safety standards: IEC, CSA, Installation category II, Pollution degree 2 For indoor use only.
UL

Table 9 Performance Specifications

Type Specification

Number of samples 100 x 2 mL vials, 30 x 6 mL vials

Injection volume 5 L for full loop injections (5 L loop as default) larger


loops possible (10 L, 20 L), partial loop filling for
smaller injection volumes

Injection principle Fixed loop

Injection precision <0.3 % RSD for 5 L

Sample temperature control 4 40 C with sample thermostat

Sample carryover 0.05 %

1260 Infinity Analytical SFC System SVC Manual 29


2 Site Requirements and Specifications
Specifications

G4302A Binary SFC Pump


Table 10 Physical Specifications

Type Specification Comments

Weight 15.5 kg (34 lbs)

Dimensions (height 180 x 345 x 435 mm (7 x 13.5 x 17 inches)


width depth)

Line voltage 100 240 VAC, 10 % Wide-ranging capability

Line frequency 50 or 60 Hz, 5 %

Power consumption 220 VA, 74 W / 253 BTU Maximum

Ambient operating 055 C (32131 F)


temperature

Ambient non-operating -40 70 C (-4 158 F)


temperature

Humidity < 95 %, at 25 40 C (77 104 F) Non-condensing

Operating altitude Up to 2000 m (6562 ft)

Non-operating altitude Up to 4600 m (15091 ft) For storing the module

Safety standards: IEC, CSA, Installation category II, Pollution degree 2 For indoor use only.
UL

For use with flow rates below 500 L/min, or for use without damper and mixer, a vacuum
NOTE degasser is required.
All specification measurements are done with degassed solvents.

30 1260 Infinity Analytical SFC System SVC Manual


Site Requirements and Specifications 2
Specifications

Table 11 Performance Specifications

Type Specification

Number of co-solvents 1, with SSV pump option 3, with external SSV 12

Precision of flow rate Same as G1312B, ?0.07 %0.07 RSD or ?0.02 min SD,
whatever is greater based on retention time at
constant room temperature

Precision of composition Same As G1312B, <0.15 % RSD or <0.04 min SD


whatever is greater

G1316C Thermostatted column compartment


Table 12 Physical Specifications

Type Specification Comments

Weight 11.2 kg (22 lbs)

Dimensions (height 140 x 345 x 435 mm (5.5 x 13.5 x 17 inches)


width depth)

Line voltage 100 240 VAC, 10 % Wide-ranging capability

Line frequency 50 or 60 Hz, 5 %

Power consumption 320 VA / 150W / 512 BTU Maximum

Ambient operating 055 C (32131 F)


temperature

Ambient non-operating -40 70 C (-4 158 F)


temperature

Humidity < 95 %, at 25 40 C (77 104 F) Non-condensing

Operating altitude Up to 2000 m (6562 ft)

Non-operating altitude Up to 4600 m (15091 ft) For storing the module

Safety standards: IEC, CSA, Installation category II, Pollution degree 2 For indoor use only.
UL

1260 Infinity Analytical SFC System SVC Manual 31


2 Site Requirements and Specifications
Specifications

Table 13 Performance Specifications

Type Specification

Column Up to 300 mm length x 4.6 mm ID (or less)

Column capacity Up to 3, with additional 2 x TCCs up to 9

Switching valve Optional (2/6, 2/10, 8/9, 6 column selector)

Temperature range Ambient -10 C to 100 C

Column compartement temperature 0.5 C


accuracy

Column compartement temperature 0.05 C


stability

Active solvent pre-heating and As standard


post-conditioning

Automated method development Optional

32 1260 Infinity Analytical SFC System SVC Manual


1260 Infinity Analytical SFC System SVC Manual

3
Installing the G4309A Agilent 1260
Infinity SFC System
Hardware Installation 34
General Procedures 34
Installing G4301 A Aurora SFC Fusion A5 37
Preparing the HPLC 55
Software Installation 68
Installing Aurora SFC Fusion A5 Driver Software 68

This chapter provides an overview of the installation and setup of the hardware
and software

Agilent Technologies 33
3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Hardware Installation

General Procedures
Proper use of wrenches
Some of the plumbing connections require a nut to be tightened onto a fitting.
There are often two sets of flats next to each other. Attempting to tighten the
nut without securing the other part of the fitting with a second wrench can
result in loosening yet another connection upstream or downstream. It is best
practice to always hold the fitting with one wrench while tightening or
loosening another connection.

Figure 4 Proper tightening of Fittings

34 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Compression (Swaged Fittings)


The fittings used in the Aurora SFC Fusion A5 are Valco. Fittings used in all
Agilent modules are Swagelok. Use the appropriate fitting as recommended by
the equipment manufacturer. The recommended tightening procedure to
install new fittings is to tighten the nut finger tight, then an additional 1/4 to
1/2 turn to seal. In general, previously swaged fittings need only an additional
1/8th turn once finger tight.
In Supercritical Fluid Chromatography, the fluid has 1/10th the viscosity of
water, so this may not be tight enough. All connections should be checked for
leaks and tightened further if necessary. Soapy water or Snoop make it easy to
find leaks if carbon dioxide is in the fluid. Tiny bubbles appear in the liquid
around the fitting.
Each fitting should be individually and carefully installed. The depth of the
tube inside the fitting is very important. If the tube pilot (length beyond the
ferrule end) is too long, the fitting can leak or, after excessive tightening, bind
permanently. If the pilot is too short, a poorly swept volume can be created.
This poorly swept volume will create noticeable chromatographic tailing. If the
pilot is much too short, the fitting could fail under use. Pilot depths are not
always interchangeable between fittings. It is a best practice to swage a tube in
the fitting in which it will be used. It is best to provide some light force to hold
the tube in the fitting and prevent the tube from exiting while tightening the
fitting.
Excessive force can result in breakage of some components, and should
obviously be avoided. It may be more expedient to replace the whole fitting if
one of the connections fails to seal. You may notice that some of the more
expensive components (such as a pressure transducer) have a less expensive
fitting mounted to them to act as a sacrificial fitting. Connections should be
made to the less expensive component, and repetitive removal and
replacement to one of the more expensive fittings should be avoided.

1260 Infinity Analytical SFC System SVC Manual 35


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Compression fittings used in the A5


Cji E^adi9Zei] use a nut and ferrule. As the nut is
;ZggjaZ tightened the ferrule compresses
on the tubing creating a leak-tight
seal
IjW^c\

When correctly installed, there is


minimal additional volume. All
volume is well swept.

When improperly set, the pilot does


not fill the fitting, leading to
additional poorly swept volumes.

If over-tightened, the ferrule can


compress the tubing and cause
sufficient deformation to prevent
removal of the fitting.

Figure 5 Compression fittings

36 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Installing G4301 A Aurora SFC Fusion A5


Early connection may damage the instrument
CAUTION
Do not connect AC power or interconnection cables or gas tubing to the Fusion
instrument until these installation procedures direct you to do so.

Preparation
Locate all modules, devices and supporting equipment before continuing.
Ensure that the supply tubing can reach a physically secured source of CO2.
Ensure that adequate venting is available and within reach of supplied waste
systems.
This document describes a particular order of plumbing the system, with
plumbing and electrical connections described last. These operations are
performed at the rear of the systems. Depending upon your individual
installation, you may wish to perform operations at the rear of the
instruments first. This is perfectly acceptable, provided you can maintain
access to supply connections to ensure integrity and leak tightness of fittings
and connections.

Unpacking the G4301 A Aurora SFC Fusion A5

Damaged Packaging
When you receive your G4301 A Aurora SFC Fusion A5 module, inspect the
shipping boxes for any signs of damage. If the shipping container or
cushioning material is damaged, notify the carrier and save the shipping
material for inspection. Save all materials until the contents have been
checked for completeness and the instrument has been mechanically and
electrically checked.

Signs of damage
CAUTION
If there are signs of damage to the G4301 A Aurora SFC Fusion A5 Module, please
do not attempt to install or use the instrument.

1260 Infinity Analytical SFC System SVC Manual 37


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Delivery Checklist
Compare the delivery checklist with the contents of the shipping boxes to
ensure completeness of the shipment.
For parts identification see Parts for Maintenance on page 111. Please
report missing or damaged parts to your local Agilent Technologies
representative.

Connecting the Waste system to the Fusion BPR Outlet


The Aurora SFC Fusion A5 has a waste bottle located outside of the cabinet.
It can be located anywhere easily accessible and visible within the range of the
supplied tubing. The waste bottle serves multiple purposes and collects liquid
waste from multiple sources. The primary purpose is to separate the gaseous
and liquid waste from the outlet of the BPR (system) in such a manner that the
gaseous waste can be appropriately vented outside of the lab environment.
The waste bottle has input and output ports located above any collected
liquids. The mixed stream enters the waste bottle and the gaseous stream exits
from the spout.

Proper system operation requires adequate space in the waste bottle to allow gaseous exit.
NOTE It is the responsibility of the operator to ensure that the waste bottle is empty before
beginning operation of the A5, and to monitor and empty the waste bottle as needed during
usage. This is not a warning.

38 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

IjW^c\In\dc
($-&$)>9
6*LVhiZA^cZ KZciZYLVhiZA^cZ
&$&+^cX]iZ[adc

6*LVhiZ7diiaZ

Figure 6 Installing the Waste Container

Exposure to toxic substances


WA R N I N G
The vapor exiting the G4301 A Aurora SFC Fusion A5 module may contain several
percent organic solvent. The effluent should NEVER be vented directly into an
enclosed space occupied by humans because of the potential for long-term
exposure to toxic substances.

Locate and assemble the waste bottle and Tygon vent tubing. The vent tubing
can be placed over the spout on the top of the waste bottle. Route the Tygon
tubing to an appropriate vent. The system must be actively vented.
Locate the A5 Waste line. Insert the free end through a hole in the top of the
waste bottle cap. Insert the tube half-way into the waste bottle. Connect the
fitting end to the outlet union on the BPR. This union uses a 10-32 CPI fitting.
Tighten snugly.

1260 Infinity Analytical SFC System SVC Manual 39


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Installing the Flowcell


The back pressure regulator exists after any detectors in the HPLC system
Thus, the detector flow cells (or splitter in the case on an ELSD or Mass
Spectrometer), operate at an elevated pressure relative to HPLC.
Agilent Technologies offers a Diode Array Detector (DAD) flow cell that has
been extensively optimized for use in Supercritical Fluid Chromatography
(SFC). This cell is pressure-rated and tested to 400 bar. It contains extensive
thermal conditioning not found in standard HPLC flow cells. Agilent
Technologys cell is highly recommended for SFC usage.
The flow cell should already be installed in the DAD/MWD. Carefully examine
the inlet and outlet ports of the cells to ensure that flow is in the correct
direction. In the DAD/MWD, the outlet port is normally located below the inlet
port on the connection block.
The inlet port of the DAD/MWD flow cell enters the stationary portion of the
handle. This stationary bar acts as an initial thermal conditioning zone.
Normally, this port has a male fitting. The outlet port connects directly to the
cylindrical portion of the flow cell. Normally, this connection has a female
fitting.

Figure 7 DAD SFC Flowcell

40 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Damage to the flowcell


CAUTION
Verify that the cell installed in the detector is capable of the high pressures used in
Supercritical Fluid Chromatography (SFC). Exposing a standard 1260 Infinity flow
cell to high pressures will result in leakage or damage to the cell.

Connecting the BPR to the 1260 Infinity stack


Connect the Aurora SFC Fusion A5 return transfer tube to the outlet block
of the detector. This tube can then be routed out the concave opening in the
bottom of the detector behind the detector cover. The return transfer tube
should then be routed to the space between the 1260 Infinity stack and the
Aurora SFC Fusion A5. Move the tube upward between the units and
through the upper tee-slot on the side cover of the Aurora SFC Fusion A5.
The return transfer tube can then be fastened in the right port of the tee in the
lower center of the BPR drawer.

6*GZijgcigVch[Zga^cZ

Figure 8 Connecting the Aurora SFC Fusion A5 return Line

1260 Infinity Analytical SFC System SVC Manual 41


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Optimized Agilent 1290 Infinity Thermostatted column compartment


Plumbing
Supercritical Fluid Chromatography (SFC) is susceptible to increased noise
due to poor thermal matching of components within the Agilent 1260 Infinity
stack. (The refractive index of carbon dioxide is 50 times more susceptible to
temperature changes than water. Consequently, thermal control in SFC is
extremely important). The Agilent 1290 Infinity Thermostatted Column
Compartment (TCC) contains two thermal conditioning zones that can greatly
increase system performance by matching temperatures of the mobile phase
to the modules being used.
Each of these zones contains internal transfer lines that can be used to
thermally condition the fluid flowing through them. The two zones exist on the
left and right side blocks within the TCC.
The block on the right side has a 6 L internal conditioning volume that is
used to precondition the mobile phase before it enters the column. When using
150 mm or shorter columns, place them in the right side of the oven. This zone
is used to precondition the fluid to column temperature and provide thermal
control of the column.
The left side block is plumbed with the effluent from the column. The purpose
of the left side block is to independently match the temperature of the mobile
phase to the optimum temperature for the detector.

AZ[i]ZViZmX]Vc\ZgVY_jhiZYid G^\]i]ZViZmX]Vc\Zg^hjhZYid
dei^b^hZYZiZXidgcd^hZedhiXdajbc# VY_jhiiZbeZgVijgZegZ"Xdajbc#
I]^hiZbeZgVijgZVY_jhibZci]Vh
Z[[ZXihdchZaZXi^k^in
Figure 9 Plumbing the Agilent 1290 Infinity Thermostatted column compartment

42 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Connecting the SFC Autosampler


Agilent Autosamplers need to be converted for use with SFC. If this has not
already been done, refer to the installation and upgrade section of the
Autosampler compatibility kit for specific instructions.

[gdbejbe

cZZYaZhZVi

XdciV^cZYlVhiZ idXdajbc

Figure 10 Plumbing the G4203A SFC Autosampler

If the autosampler has been upgraded, the plumbing connections can be made
to the injection valve.
In the external loop operation of the autosampler, all connections are made on
the injection valve as shown. Port 1 is connected to the inlet of the right side
(column pre-heater) thermal conditioning block of the oven.
Port 6 is connected to the output flow from the binary pump.

1260 Infinity Analytical SFC System SVC Manual 43


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Connecting the Booster to the Agilent 1260 Infinity system


Locate the stainless steel booster transfer line. Connect one end to the top
port of the output tee on the center right side of the booster drawer. Tighten
finger tight with an additional 1/8th turn as needed.

6*jbeigVch[Zga^cZVhhZbWan

Figure 11 Connecting the Booster outlet line

This tube can be routed through the tee slots on the side of the Aurora SFC
Fusion A5. The tube is routed upward between the Aurora SFC Fusion A5
and the Agilent 1260 Infinity stack to the bottom side of the binary pump. The
tube is then routed horizontally to the bottom center of the binary pump
where it can enter the concave opening behind the cover. The tube should then
be routed to the left side of the pump beneath the Channel A (left side). The
end of the booster transfer tube can then be installed with the adapter in the
passive inlet valve.
Any spare tubing can be located between the A5 and HPLC stack.

44 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

The injector wash pump


The wash pump is required only with autosamplers normally operated as
broken loop autosamplers converted to external loop autosamplers. The
injector wash pump requires a source of wash solvent. This wash solvent is
used to prime the injector system to ensure proper operation while aspirating
sample. If the metering device in the autosampler is not filled with solvent, it
can cavitate and yield anything between poor area reproducibility to no peaks
(no actual injection). The wash pump also washes the injection system (needle
interior, needle seat, injection valve, and sample loop) before and after each
injection.
The injector wash pump is connected to a bottle filter (supplied) through the
wash pump inlet line. The wash pump inlet line is connected to the lower port
on the injector wash pump located on the left side of the BPR drawer. The
other end should be connected to the bottle filter. The filter can then be placed
in a user-supplied wash solvent bottle.
The wash solvent bottle can be left on the bench or placed in the 1200 Infinity
Series solvent cabinet. The wash pump is connected to the metering device
(syringe pump) through a check valve intended to prevent siphoning. The
operation of the check valve should be verified to ensure it is not leaking,
because this can cause a loss of injection precision.

1260 Infinity Analytical SFC System SVC Manual 45


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

LVh]ejbe^caZiVhhZbWan
&$-IZ[adcIjW^c\

7diiaZ[^aiZg

Figure 12 Connecting the wash pump inlet

The outlet of the injector wash pump is connected to the metering device in
the Agilent autosampler. To plumb the outlet side of the injector wash pump,
locate the was pump transfer line with check valve and install it in the upper
port of the injector wash pump. Flow through the wash pump is in the upward
direction. The injection wash system will not function if the check valve is
improperly installed. An arrow embossed on the check valve body indicates
the direction of flow; verify that the check valve is firmly installed.
Connect the wash pump transfer line with the spring loaded check valve. A
union is provided on the other side of the of the wash pump transfer line.
Follow instructions below for priming the injector wash pump before
connecting the union and the autosampler metering device.

46 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

BZiZg^c\YZk^XZ
CZZYaZ

LVh]ejbeigVch[Zga^cZ

LVh]ejbeigVch[Zga^cZ Jc^dc&%"('

EZZ`ijW^c\

Figure 13 Connecting the Wash Line and Check Valve

Priming the Injector Wash Pump


The wash pump is NOT self priming; it must be filled with wash solvent. See
the plumbing scheme, Figure 13 on page 47. There is an extra in-line check
valve downstream of the wash pump. This check valve contains a ball pushed
into a seat using a spring. The purpose of the spring is to prevent siphoning of
the wash solvent from the container, through the injection valve to waste. With
the check valve in place, it is easy to prime the injector wash pump without
siphoning.
Once the wash pump transfer line and spring loaded check valve are
connected to the injector wash pump, you can fill the pump through the
injector wash pump transfer line. A syringe and several Luer adapters are
included in the ship kit (see Figure 14 on page 48). The syringe can be
connected to the wash pump transfer line (once the union is removed) using
the Luer adapter. Retracting the syringe pulls solvent through the system,
check valves, and tubing. This effectively primes the injector wash pump. After
priming, remove the Luer adapter, reinstall the union, and connect to the
autosampler metering device.

1260 Infinity Analytical SFC System SVC Manual 47


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Hng^c\Z
LVh]ejbeigVch[Zga^cZ

AjZg[^ii^c\
8]ZX`kVakZiZbedgVg^angZbdkZY

Figure 14 Priming the wash line

Connecting Aurora SFC Fusion A5 to a Source of Carbon Dioxide


The Aurora SFC Fusion A5 has a 1/8th inch tube inlet connection on the
lower left side on the back of the module. This connection is actually part of a
very high surface area filter intended to intercept catastrophic levels of
particulates. You need to connect this input to a source of carbon dioxide. The
most common source of carbon dioxide is liquefied carbon dioxide from a
room-temperature cylinder. At room temperature, the pressure in the cylinder
could change from a little above 50 bar to just below 70 bar. Unlike most SFCs,
Aurora SFC Fusion A5 is immune to the variations in flow resulting from
cylinder pressure.

Creating severe frostbite in a short time


WA R N I N G
Expanding carbon dioxide can become extremely cold, capable of creating severe
frostbite in a short time. Avoid contact with expanding gases. Do not vent
substantial quantities into the laboratory.

Use the system in a well ventilated area


WA R N I N G
Carbon dioxide is poisonous at high concentrations and should only be used in well
ventilated areas. The system effluent should be vented into a fume hood or to the
outside. Evacuate if a large spill occurs. A carbon dioxide sensor/alarm is
recommended.

48 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Individual cylinders
CO2 cylinders can be dangerous if handled improperly
WA R N I N G
Carbon dioxide in cylinders is partially liquefied under high pressure and contains a
great deal of energy. If containment is breached (a break in the line or cylinder) the
entire contents will vaporize and quickly expand up to 500 times in volume and
create very forceful high velocity gas streams. Cylinders must be properly
constrained, and proper tubing used, to avoid damage that could generate
projectiles.

Any industrial grade of carbon dioxide is acceptable provided it is supplied in


a cylinder without a DIP tube. Drawing off the vapor phase leaves non-volatile
contaminants behind in the cylinder. Using cylinders with a DIP tube subjects
the chromatograph to contaminants soluble in the dense, liquid layer.
Larger tanks are more convenient in that they require to be changed less
frequently. Cylinders can contain up to 35 kg of CO2. Generally, 4.6 mm
columns are run at 3 5 mL/min, which is approximately 2.5 4 g/min of
carbon dioxide. This is equivalent to 150 250 g/h; 1.2 6 kg/day. Thus, a
15 kg cylinder should last 2.2 to 11 days; a 25 kg cylinder would last 4 to 19
days; and a 35 kg cylinder could last 5.3 to 27 days - all depending on use (3
5 mL/min; 8 24 h/day). For individual users, particularly new users, the use
of cylinders is perfectly acceptable. Larger groups should consider installing a
gas delivery system and a bulk storage tank.
Locate the cylinder as close to the instrument as possible. In the past, the
cylinders were stored at much higher temperature than the lab temperature,
which resulted in vaporization in the supply line coming into the lab. Most
SFC pumps cannot condense this vapor and therefore, cannot deliver CO2.
Fusion has a very powerful condenser designed to accept vapor phase CO2.
Nevertheless, it is always advisable to not stress any equipment.
Facilities and safety personnel often wish to store and mount the cylinders
outside the lab sometimes quite far from the intended location of the
instrument. They should recognize that the transfer lines can hold large
volumes, equivalent to a large fraction of a cylinder, particularly if tubing with
large ID is used. Shut-off valves at both ends of a transfer line are not
recommended, unless one or both has a pressure relief valve or burst disk.

1260 Infinity Analytical SFC System SVC Manual 49


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Be sure the cylinder is properly constrained and cannot tip over. Suitable
chains or cylinder straps are required.
Cylinders in the USA and Canada use a CGA 320 cylinder adapter. One is
included in the Fusion USA ship kit, along with a 1/4 inch FNPT to 1/8 inch
compression fitting, and 6 feet of 1/8 inch OD stainless steel tubing. The filter
fitting sticking out the back of the Aurora SFC Fusion A5 module contains
1/8 inch nut and ferrule(s), which could be used with the Aurora-supplied
6 feet tube, or a longer, user-supplied tube, to connect a cylinder to the Fusion
module.
There are at least four different European standards for the connection of
carbon dioxide cylinders. They do not appear to change at national borders
(some users in one country have different supply connections from other
users in the same country). Aurora recommends that customers in Europe
contact their gas supply companies and ask them how to mate the cylinders
they supply with an American 1/8 OD supply line. This should require a
cylinder connector and a reducing union down to a 1/8 inch compression
fitting.

Leaks will not be sensed or protected


WA R N I N G
Carefully check supply cylinders and inlet fittings for leaks. Any leaks present in the
supply line and inlet fitting will not be sensed or protected by safety features in the
Aurora SFC Fusion A5.

50 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

8<6('%l^i]<Vh`Zi
Cdgi6bZg^XVdcan#
8djcignheZX^[^XVYVeeiZgcdi^cXajYZY#
8DHjeeana^cZ6hhZbWan
'
&$-^cX]
HlVY\ZadX`

Figure 15 Figure 3.25

Improper plumbing can cause leaks


WA R N I N G
It is imperative to use two wrenches to install the supply fitting in the bulkhead filter
(entry connection) on the back of the Aurora SFC Fusion A5. Turning or twisting
the bulkhead supply filter could cause failures or leaks in tubing within the Aurora
SFC Fusion A5.
All supply fittings need to be thoroughly checked for leaks. Any leaks in the supply
fittings can vent the CO2 supply.

1260 Infinity Analytical SFC System SVC Manual 51


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Overtightening the fitting could damage the filter


CAUTION
Use two wrenches when installing the CO2 supply line to the bulkhead filter (entry
connection) on the back of the Aurora SFC Fusion A5. Although the filter is
captured and should not rotate, there is a remote possibility that fittings inside the
Fusion module could be loosened by severe stress on the nut on the supply line.
Leakage inside the cabinet requires service by trained personnel.

Gas Delivery Systems (GDS)


Anyone performing semi-prep SFC has probably been convinced that
operation without a GDS is problematic. Thus, many laboratories are now
plumbed with carbon dioxide boosted to 70 80 bar outlet pressure. Even
though an analytical system does not need such a GDS, it is perfectly adapted
to its use.
Aurora SFC Fusion A5 has an inlet safety shut-off valve rated to 1500 psi
(>100 bar). While this is rating is well above the outlet pressure of any typical
GDS, past experience suggests that it is wise to allow for some extra margin.
Agilent Technologies suggests setting the local output of any GDS between
60 bar and 70 bar, through local outlet pressure regulators.
Under these conditions, the GDS will almost always provide liquid carbon
dioxide to the chromatograph. Unlike some earlier systems, the Fusion module
easily condenses any fluid that is present as a vapor, and prevent pump
cavitations.

Cable connections to the HPLC


Personal injury
WA R N I N G
Ensure that the AC power cord is NOT yet connected to the Aurora SFC Fusion A5
instrument.

Cable connections to the Aurora SFC Fusion A5 are dependent upon the
installed configuration. The primary decision to be made is whether or not an
Agilent autosampler is present. When an Agilent 1260 Infinity SFC
Autosampler is present, the Aurora SFC Fusion A5 remote line and relay
contacts lines are connected to the autosampler.

52 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

The Binary pump analog output signal should be connected to the Aurora SFC
Fusion A5 Reference in terminal.

GZ[ZgZcXZh^\cVa

GZbdiZh^\cVa

GZaVnh^\cVa

Figure 16 Reference, Remote and Relay Signal Connections

Lastly, a cable should be connected to the Aurora SFC Fusion A5 USB port.
This can be run to any convenient, compatible USB port on the PC.
CAN cabling between the Agilent 1260 Infinity modules is unchanged from the
normal, recommended means of interconnecting HPLC devices.
The Agilent 1260 infinity stack requires a LAN connection to the PC. This LAN
connection requires all the normal HPLC/ChemStation properties such as
BOOTP, Firewall, and IP Address settings. These remain unchanged in an
Aurora SFC Fusion A5 installation.

1260 Infinity Analytical SFC System SVC Manual 53


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Connecting and operating multiple instruments on a single PC is not


supported. Both the Aurora SFC Fusion A5 USB and Agilent 1260 Infinity
LAN connections must be made on the PC on which ChemStation is installed
and which is used for instrument control.

Connecting the Leak Tray Waste Line


The Aurora SFC Fusion A5 contains a leak tray on the bottom of the
instrument to collect and sense any liquid spills that may occur in the Aurora
SFC Fusion A5 cabinet. In the bottom of this leak tray is an active sensor
that continuously monitors for the presence of liquid.
The drip tray contains an overflow drain to divert any large amounts of
collected liquid to an external collection container. The overflow tube
incorporates a simple push-to-connect fitting. It is connected to the port on
the bottom center on the front of the instrument. Pushing on the outside ring
of the port allows this line to be removed.
Since this liquid may be organic solvents, you should supply an appropriate
collection container.

6*AZV`igVnlVhiZa^cZ
,$&+In\dc*$&+>9

Figure 17 Leak Tray Waste Line

54 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Preparing the HPLC


Preferred Stack Arrangement
To minimize delays and broadening caused by excessive tube lengths, we
recommend the following stack layout. The SFC Fusion A5 needs to be
immediately adjacent to the LC stack, but can be placed on either side. In the
preferred arrangement, the degasser and solvent tray are located on the top of
the stack. These feed into the binary pump, and the A5 injection wash pump.
The binary pump is located below the degasser and directly above the
autosampler. The thermostatted column compartment is located below the
autosampler, with the detector at the bottom of the stack.

9Z\VhhZg

7^cVgnH;8Ejbe

H;86jidhVbeaZg

I]ZgbdhiViiZYXdajbcXdbeVgibZci

969dgBL9

Figure 18 Preferred order of Agilent HPLC Components

It is recommended to plumb the system with small diameter tubing. For


general use 0.17 mm or 0.007 inch tubing should be used. In more demanding
applications where pressure drops are not excessive, 0.12 mm or 0.005 inch

1260 Infinity Analytical SFC System SVC Manual 55


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

tubing can be selectively used. Use ferrules and tubing connections as


recommended by the equipment manufacturer.
All four degasser channels will be available for solvent usage. It is highly
recommended that one channel be reserved for neat methanol (no additives).
Reserving this channel will minimize the time needed for flushing the degasser
as solvents are changed. Failure to adequately flush the degasser when
switching solvents will greatly impact sensitivity.

Agilent 1260 Infinity SFC Binary Pump


It is highly recommended that you familiarize yourself with standard
maintenance functions and terminology used in the binary pump. This
information is available in the Agilent Binary pump reference manual and
user guide.

9VbeZgl^i]Wj^aY^cEgZhhjgZHZchdg!egZhhjgZa^b^i+%%WVg!VcY8a^e[dgb^mZg

8Ve^aVgnEK
KVakZ=daYZg
KVakZ]daYZg[dgi]^gYD7K

8Ve^aaVgn

6YVeiZgbVaZ$[ZbVaZ

8d^aZY8Ve^aaVgn
Figure 19 Diagram of the Agilent Binary Pump

The G4303A SFC Pump is a binary high pressure mixing pump, optimized for
the usage in a SFC System. The Purge Valve is mounted on the right side for
purging the Modifier Channel (Channel B) only. An additional third Outlet
Ball Valve (OBV) is installed with a Valve Holder on Pumphead B, to allow the
Channel B to be flushed, even if CO2 tank pressure is supplied to the System.

56 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

On the Passive Inlet valve of the CO2 Channel (Channel A) is an Adapter to


allow a Swagelok fitting to be connected to the valve. The damper includes the
pump pressure sensor and allows a pressure up to 600 bar. Different set of
piston seals are installed in the two pump heads. Normal phase seals (yellow
PE) in pump head channel A, standard seals PTFE, carbon filled in pump head
channel AB.

Toxic, flammable and hazardous solvents, samples and reagents


WA R N I N G
The handling of solvents, samples and reagents can hold health and safety risks.
When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor, and follow good
laboratory practice.
The volume of substances should be reduced to the minimum required for the
analysis.
Do not operate the instrument in an explosive atmosphere.

The BNC coaxial cable can be used to connect the Analog Output on the back
panel of the Binary pump to the SFC Fusion A5 Reference In
terminal/connector on the back of the Fusion module. Be aware that the
output of similar appearance on the DAD and VWD detectors should NOT be
used.

GZVgd[H;87^cEjbe

Figure 20 Connecting the Analog Reference Signal

1260 Infinity Analytical SFC System SVC Manual 57


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Modifications to the Agilent Autosampler

KVakZ7dYn

>hdaVi^dcHZVaHeg^c\id7VX`

6jgdgV(<gddkZGdidgHZVa

HiVidg=ZVY

.$+)]ZmhXgZlh

Figure 21 Exploded View of the Autosampler Injection Valve

Replacing a rotor seal is well documented in Agilent manuals, and service videos. Please
NOTE refer to these guides for further assistance.

58 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

1 The Autosampler should be plumbed in accordance with Figure 22 on


page 59.

BZiZg^c\9Zk^XZ
CZZYaZ

[gdbejbe

CZZYaZHZVi

8]ZX`KVakZ

8dciV^cZYLVhiZ
idXdajbc
Figure 22 Autosampler Fluidic connections

2 Connect one end of the wash pump transfer line with the check valve to the
outlet port of the injection wash pump.
3 Before connecting the union end of the wash pump transfer line, use a
syringe and Luer Adapter to prime (pull sample through) the injection wash
pump
4 The existing line (blue capillary) from the inlet of the Agilent 1260 Infinity
SFC Autosampler metering device can be plumbed into the union on the
wash pump transfer line

1260 Infinity Analytical SFC System SVC Manual 59


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Positioning the waste line


In normal operation, the sample loop is switched from mainpass i(n the high
pressure flow stream) to bypass (out of the flow stream and connected to the
seat capillary and waste). As the valve switches, the fluid in the loop expands
and rushes out of the loop through the waste line. If the waste line is
positioned to empty into a liquid reservoir such as a WPA flush port, the
exiting fluid from the sample loop could splash the flush port fluid across the
ALS. The preferred positioning of the waste line is through the leak vent. In
this position, the expanding fluid exits harmlessly into the leak tubing.
Alternatively, a waste line can be fabricated that empties into the Aurora SFC
Fusion A5 gas liquid separator (waste bottle).
Under no circumstances should the waste line be allowed to be unconstrained.
During operation of the injection wash pump, this is the final exit of the wash
fluid.

Installing the Agilent BCD Board


The BCD Interface card (Interface board (BCD) with external contacts and
BCD outputs (G1351-68701)) and relay cable from the shipping kit are used to
control the wash pump.

789>ciZg[VXZXVgY

Figure 23 BCD Interface Card

Agilent HPLC autosamplers have a small compartment located on the back


panel near the top, with a cover held in place by two captured knurled nuts.
Loosen the nuts and remove the cover, taking care to retain it for future use.

60 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Take the BCD Interface card from the shipping kit, taking care to prevent
damage from static electricity, solvent, etc.

Electronic boards are sensitive to electrostatic discharge (ESD) and should be handled
CAUTION
with care so as not to damage them. Touching electronic boards and components can
cause electrostatic discharge.
ESD can damage electronic boards and components.
Be sure to hold the board by the edges and do not touch the electrical components.
Always use an ESD protection (for example, an ESD wrist strap) when handling
electronic boards and components.

1 Carefully slide the board into the slot. When it is fully in place, gently push
the bezel until the connectors engage. Tighten the captive knurled nuts.
2 With the BCD card in place, connect the 15-pin relay contacts connector to
the relay input connector on the back of the Aurora SFC Fusion A5.
3 Connect the 9-pin Remote connector on the back left of the autosampler to
the 9-pin Remote connector on the back of the Aurora SFC Fusion A5.

Figure 24 Autosampler Signal Cable Connections

1260 Infinity Analytical SFC System SVC Manual 61


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Installing an Injection Program in the Method


The injector program is delivered as part of the default setup, but you can
enter it manually, if necessary.

Table 14 Agilent 1260 Infinity SFC Autosampler Injector Program

Injection

Injection Volume 15.0 L

Injection Mode Injector Program

Time

Stoptime No Limit

Posttime Off

High Throughput

Autom. Delay Vol. Reduction Off

Overlapped Injection disabled

Minimized Carry Over Off

Needle Wash

Wash Mode Wash in Flushport

Wash Time 1.0 s

Wash Location Vial 10

Repeat 1 times

Injector Cleaning

Injection Valve Cleaning / Valve Switching

Time 1 Bypass off

Time 2 Mainpass/Bypass off

62 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Table 14 Agilent 1260 Infinity SFC Autosampler Injector Program

Time 3 Mainpass/Bypass off

Time 4 Mainpass/Bypass off

Valve Movements 1

Auxiliary

Draw Speed 100 L/min

Eject Speed 100 L/min

Draw Position 0.0 mm

Equilibration Time 2.0 s

Sample Flush-Out Factor 5.0 times Injection Volume

Vial/Well Bottom Sensing No

Store Temperature No

Injector program table

Row Action

1 NEEDLE down

2 CONTACT B CLOSED

3 WAIT 0.10 min

4 CONTACT B OPEN

5 VALVE bypass

6 DRAW 1.5 L from air, def. speed

7 DRAW def. amount from sample, def. speed, def.


offset

8 DRAW 5 L from air, def. speed

9 NEEDLE wash as method

10 EJECT max. amount into seat, def. speed

11 VALVE mainpass + start pulse

12 CONTACT B CLOSED

1260 Infinity Analytical SFC System SVC Manual 63


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

Table 14 Agilent 1260 Infinity SFC Autosampler Injector Program

13 WAIT 0.10 min

14 CONTACT B OPEN

Agilent Contacts Option

Contact 1 Open

Contact 2 Open

Contact 3 Open

Contact 4 Open

Using the Agilent SFC Autosampler as a fixed loop to (external) loop autosampler has
NOTE several minor consequences that need to be understood to achieve adequate performance.

64 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Dead Volume-Partial Loop Injections


Plumbed as a loop injector, the Agilent 1260 Infinity SFC Autosampler has a
dead volume between the high pressure needle seat and the groove on the
rotor of the injection valve. Two different diameter tubes are used to make this
connection. Thus, the actual dead volume can have several different values,
depending on the ID of the tubing that is used. This dead volume can be
several micro liters or more. When attempting to make low-volume injections,
it may be possible that no sample enters the loop unless precautions are taken.
Use an air bubble on each side of the sample, insert a plug of modifier or other
solvent behind the sample, or any other method to ensure that the sample
reaches the loop.

Gjc L^i]YgVl
& '

LVhiZ LVhiZ
CZZYaZ CZZYaZ

AdVY 8dajbc Ejbe


8dajbc Ejbe ( >c_ZXi"GdiViZ
)

LVhiZ
LVhiZ lVh

'#*"'#* A
* A

8dajbc Ejbe
* gdiViZkVakZ 8dajbc Ejbe

Figure 25 Full Loop Injection

1260 Infinity Analytical SFC System SVC Manual 65


3 Installing the G4309A Agilent 1260 Infinity SFC System
Hardware Installation

9gVlV^gWjWWaZ 9gVl'cYV^gWjWWaZ

CZZYaZ CZZYaZ CZZYaZ


LVhiZ LVhiZ LVhiZ

8dajbc Ejbe 8dajbc Ejbe 8dajbc Ejbe

CZZYaZ CZZYaZ LVhiZ


LVhiZ

8dajbc Ejbe 8dajbc Ejbe


Figure 26 Partial Loop Injection

Some General Injection Rules


20 L loops are commonly used in HPLC. In SFC, the sample is often dissolved
in a solvent that is stronger than the mobile phase. Injecting large volumes,
such as 20 L, of such a solvent will cause peak distortion.
As a rule of thumb, no more than approximately 5 L of a polar solvent such
as methanol should be injected onto a 4.6 mm ID column, as shown in
Figure 27 on page 67.
This should NOT have a major impact on area reproducibility but should
destroy efficiency as indicated in Figure 27 on page 67.

66 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Hardware Installation

Use Small Loops


The loop should not be much larger than the maximum desired injection (or
=<5 L). Bear in mind that the loop is washed with a strong solvent, and is
usually filled with that solvent. The sample displaces some or all of this
solvent.
If the loop is too large, too much strong solvent is injected, regardless of
whether the loop is filled with sample or wash solvent. Peaks will be distorted
and efficiency will degrade.
b6J

>c_ZXi^dcKdajbZ

Figure 27 Plot showing loss of efficiency with large injection volume

If a large loop is used, air bubbles can be used to displace the wash solvent.

Less Polar Sample Solvents Help


Replacing the sample solvent with a much less polar solvent can allow
injection of much larger volumes. However, the new solvent must be
significantly less polar than the modifier used. Replacing methanol with
ethanol or even isopropanol has minimal effect. Solvents such as chloroform
or methylene chloride tend to cause significant focusing (NO broadening), but
since they are chlorinated should probably be avoided.

1260 Infinity Analytical SFC System SVC Manual 67


3 Installing the G4309A Agilent 1260 Infinity SFC System
Software Installation

Software Installation

The software installation consists of the following steps:


1 Install Agilent ChemStation, for more details see Agilent ChemStation for
LC and CE Systems Installation Manual, Chapter Installing the Agilent
ChemStation, Initial Installation.
2 Install SFC USB driver.
3 Install the SFC Fusion Driver Software, see Installing Aurora SFC Fusion
A5 Driver Software on page 68

Installing Aurora SFC Fusion A5 Driver Software


The driver is installed as an add-on product for ChemStation.
The following screenshots show the installation steps. It is important to select
the ChemStation installation directory as destination path!
1 Start the Sfc Fusion ChemStation Interface Setup Wizard

68 1260 Infinity Analytical SFC System SVC Manual


Installing the G4309A Agilent 1260 Infinity SFC System 3
Software Installation

2 Select installation folder.

3 Confirm installation.

1260 Infinity Analytical SFC System SVC Manual 69


3 Installing the G4309A Agilent 1260 Infinity SFC System
Software Installation

4 Install the Sfc Fusin ChemStation Interface.

5 Complete the installation.

70 1260 Infinity Analytical SFC System SVC Manual


1260 Infinity Analytical SFC System SVC Manual

4
Configuring the System
Configuring SFC interface in ChemStation 72
Setting up the Method 76
Status 78
Control 79
Testing the System 80

How to configure the Agilent 1260 Infinity SFC Analytical system and G4301 A
Aurora SFC Fusion A5 in ChemStation with Aurora add-on drivers.

Agilent Technologies 71
4 Configuring the System
Configuring SFC interface in ChemStation

Configuring SFC interface in ChemStation

The Aurora SFC Fusion A5 ChemStation interface works with the ChemStation Modular
NOTE LC System, not with Modular LC System (Classic). The Modular LC System (Classic)
instrument does not support RC.Net drivers, which are necessary for the LC modules.

1 Select Modular LC System as instrument type.

2 Start the Agilent ChemStation.


3 Configure the devices in the Agilent ChemStation, see Agilent ChemStation
for LC and CE Systems Installation Manual, Chapter Configuring the
Instruments, Configuring a Modular 3D LC System or Modular LC System.

72 1260 Infinity Analytical SFC System SVC Manual


Configuring the System 4
Configuring SFC interface in ChemStation

4 Depending on the current configuration, Agilent ChemStation startup


phase asks if the configuration shall be modified.

5 In the Instrument Configuration editor Auto Configuration sets up the Agilent


system as usual.

Conflict with SFC macro implementation


CAUTION
It is important that the Aurora SFC is listed below the Agilent pump; otherwise, the
SFC will become pump 1 of the system, which conflicts with the current SFC macro
implementation. The SFC macro always checks the first pump of the system for
error or standby.

1260 Infinity Analytical SFC System SVC Manual 73


4 Configuring the System
Configuring SFC interface in ChemStation

6 To set up the SFC module, scroll down the Configurable Modules slider until
you see the SFC icon. Select it with a double-click or right arrow to move it
to the Selected Modules panel.

7 Select Configure to open the SFC configuration dialog box.

74 1260 Infinity Analytical SFC System SVC Manual


Configuring the System 4
Configuring SFC interface in ChemStation

8 In the Aurora SFC Configuration, enter the device name or use the default to
identify the module within Agilent ChemStation. Select a COM Port in the
Aurora COM Port field.

9 Press OK to accept the settings.


10 Click OK in the Agilent ChemStation Instrument Configuration to apply all
configurations.

1260 Infinity Analytical SFC System SVC Manual 75


4 Configuring the System
Setting up the Method

Setting up the Method

1 In the Agilent ChemStation, the method parameters for all RC.Net devices,
including SFC settings, are combined into a single tabbed dialog box:

76 1260 Infinity Analytical SFC System SVC Manual


Configuring the System 4
Setting up the Method

2 The Instrument Curves tab shows all instrument curves (monitor traces)
available for storing with the acquired data file:

1260 Infinity Analytical SFC System SVC Manual 77


4 Configuring the System
Status

Status

1 To start Aurora SFC Status in Agilent ChemStation go to View > Instrument


Actuals. The Instrument Actuals have to be enabled.

2 The available status information depends on the configuration (e.g. Aux


Pressure).
3 The button in the upper left corner of the Status user interface allows you to
either switch the device ON or put it into Standby mode.

78 1260 Infinity Analytical SFC System SVC Manual


Configuring the System 4
Control

Control

The following steps show you how you can control your SFC instrument in the
Agilent ChemStation.
1 In Instrument > More SFC you find more menu items that allow you to control
the Aurora device directly.

2 To change the status of the SFC device and to Depressurize, go to Instrument


> More SFC > Control. This displays the Aurora SFC Control dialog box.

1260 Infinity Analytical SFC System SVC Manual 79


4 Configuring the System
Testing the System

Testing the System

The SFC driver lists all available configuration, method and status values in
ChemStation registers, available by macros.

Additionally, it works with the ChemStation


SendModule$-function. Using this command, you can send commands
directly to the SFC module. There are special commands, allowing you to
enable and disable diagnostic mode (
print SendModule$(SFC1, EnableDiag) /
print SendModule $(SFC1, DisableDiag).
Enabling Diagnostic immediately shows additional status information in the
Status window. Additionally, a separate configuration menu item shows up in
Instrument > More, and Additional instrument curves are available in the Online
plot.

Enabling Diagnostic mode requires a restart of Agilent ChemStation, to show the


NOTE Additional configuration item and Instrument curves.

80 1260 Infinity Analytical SFC System SVC Manual


1260 Infinity Analytical SFC System SVC Manual

5
Using the Aurora SFC Fusion A5
Powering up the Module 82
Power-up Sequence and Operational Control States 83
Operational Control States 84
Controlling the Aurora SFC Fusion A5 through the Agilent
ChemStation 87
Running a method on the SFC system 89
Shutting Down the SFC System 92
Partial Shutdown 92

This chapter provides information and hints on the use of the SFC System.

Agilent Technologies 81
5 Using the Aurora SFC Fusion A5
Powering up the Module

Powering up the Module

The Aurora SFC Fusion A5 module is powered on by pressing the top of the
rocker-type power switch located on the upper right rear of the module. Once
the rocker is pressed, the Aurora SFC Fusion A5 module responds by
entering its power-up sequence. The power button of the Aurora SFC Fusion
A5 unit must remain accessible at all times. Never arrange equipment so that
the switch cannot be accessed.

The power switch has to be reachable for emergency


WA R N I N G
The power switch of the G4301A Aurora SFC Fusion A5 unit must remain
accessible at all times. Never arrange equipment so that the switch cannot be
accessed.

82 1260 Infinity Analytical SFC System SVC Manual


Using the Aurora SFC Fusion A5 5
Power-up Sequence and Operational Control States

Power-up Sequence and Operational Control States

When power is applied to the Aurora SFC Fusion A5 module, a series of


events is initiated. The order of these events is designed to safely initialize and
test individual component functionality. The module power-up sequence
executes the following steps:
1 Power is automatically applied to the processor and the two module fans.
2 The processor initializes:
a Any temporary configuration or calibration data that has not been
stored in flash memory is lost.
b A checksum validation is made of data stored in flash memory.
c A self test is run to test power supply voltage levels and sensor readings
for in-range values
d Stored calibration and configuration data are downloaded to RAM.
e The event logbook is updated
3 The booster pump drive is rotated to find its index pulse.
4 Index pulses are tested for module fans and the coolant pump.
5 The BPR is homed to its fully open (depressurized) position.
6 If installed, the wash pump is rotated to its index pulse.
At the completion of a successful power-up sequence, the processor places the
Aurora SFC Fusion A5 module in the OFF operational state, described in
Operational Control States on page 84. If an error is encountered and
unresolved after multiple attempts, the module is placed into the ERROR state
and a notation is stored in the event log.

1260 Infinity Analytical SFC System SVC Manual 83


5 Using the Aurora SFC Fusion A5
Operational Control States

Operational Control States

The Aurora SFC Fusion A5 module has three defined operational control
states: OFF, STANDBY and ON. You control the three states by selecting the A5
Control option from either the ChemStation SFC Fusion icon, by pressing
various icons of the ChemStation GUI, or from the Aurora A5 Diagnostic
Program status tab.
Some components of the Aurora SFC Fusion A5 module are not governed by
the three described states but are continuously on. These include:
The processor.
Continuously records and transmits sensor data to the host control
system.
Handles status and command requests from the host controller.
Monitors sensors for safety-related parameters.
Updates the event log.
Pressure and temperature sensors are continuously powered and sensed.
Coolant and electronics bay fans are continuously powered.
The wash pump is activated by contact closure of the external contacted
board placed in the SFC Autosampler, independent of the Aurora SFC
Fusion A5 control state.

84 1260 Infinity Analytical SFC System SVC Manual


Using the Aurora SFC Fusion A5 5
Operational Control States

The OFF State


The OFF state is characterized as follows:
The CO2 supply valve is closed (unpowered).
The booster pump drive is unpowered.
The BRP drive is unpowered.
The BRP heater is unpowered.
The chiller is unpowered.
The secondary cooling circuit pump is unpowered.
The OFF state is always entered after a successful power-up sequence. It can
also be entered by selecting the Off option in the control window, by a timeout
from the STANDBY state or by pressing Off in the Agilent ChemStation
graphical user interface twice in succession.
When the Aurora SFC Fusion A5 Module is in the OFF state, the top power
status light is constant and the bottom ready status light is off.

The STANDBY State


The STANDBY state is characterized as follows:
The CO2 supply valve is closed (unpowered).
The booster pump drive is unpowered.
The BRP drive is powered.
The BRP heater is powered.
The chiller is powered.
The secondary cooling circuit pump is powered.
The STANDBY state can also be entered by selecting the Standby option in the
control window, by pressing the power button associated with the SFC icon of
the GUI, or by pressing Off in the ChemStation graphical user interface once
while the system is running.
When the STANDBY state is entered from the OFF state, the BPR must be
homed. Once this is accomplished, the BPR drive is active but in a hold state.

1260 Infinity Analytical SFC System SVC Manual 85


5 Using the Aurora SFC Fusion A5
Operational Control States

When the Aurora SFC Fusion A5 Module is in the STANDBY state, the two
status lights flash alternately. The STANDBY state remains active for up to
three hours. If no user-initiated action is taken to change or renew the state
within this time, a timeout occurs and the processor automatically enters the
OFF state.

The ON State
The ON state is characterized as follows:
The CO2 supply valve is open (powered).
The booster pump drive is powered and begins to pump CO2 to the pressure
target.
The BRP drive is powered and the BRP begins the process of regulating to
its setpoint.
The BRP heater is powered.
The chiller is powered.
The secondary cooling circuit pump is powered.
The ON state can also be entered by selecting the On option in the control
window, by pressing the power button associated with the SFC icon of the
GUI, or by pressing On in the ChemStation graphical user interface.
When the Aurora SFC Fusion A5 Module is in the ON state, the top status
light is continuously lit. The bottom status light is lit when the system reaches
a ready state, indicating that backpressure and booster pump pressure are
under control and stabilized within their control band.
Modules that start from the OFF state generally must first perform an
initialization routine before moving to ON. In the case of the Aurora SFC
Fusion A5 module, initialization causes the BPR first to home, then move to
a default initialization position. The booster pump delays operation until the
chiller passes below a threshold temperature value.

86 1260 Infinity Analytical SFC System SVC Manual


Using the Aurora SFC Fusion A5 5
Controlling the Aurora SFC Fusion A5 through the Agilent ChemStation

Controlling the Aurora SFC Fusion A5 through the Agilent


ChemStation

Exploring the A5 module Graphical User Interface (GUI)

Figure 28 Setting Control State

Figure 29 Editing the Aurora Fusion A5 Method Parameters

1260 Infinity Analytical SFC System SVC Manual 87


5 Using the Aurora SFC Fusion A5
Controlling the Aurora SFC Fusion A5 through the Agilent ChemStation

Editable method parameters for the Aurora SFC Fusion A5 are the Nozzle
Pressure (system back-pressure) and the Nozzle Temperature (temperature of
the back-pressure regulator).

Controlling the wash pump

When the wash pump icon is clicked, the wash pump starts pumping for 30
seconds and flushes the autosampler metering, loop, needle and needle seat
with solvent, Typically, modifier solvent is used to flush the autosampler. The
wash pump is controlled by the Agilent SFC autosampler via the built-in BCD
board, and works independently of the current state of the Aurora Fusion A5.

88 1260 Infinity Analytical SFC System SVC Manual


Using the Aurora SFC Fusion A5 5
Running a method on the SFC system

Running a method on the SFC system

Loading the SFC_Def analysis method


When the Aurora SFC Driver add-on for ChemStation is installed, the method
SFC_Def.M is copied to the default method folder. This method contains
parameters most commonly used in SFC. You should develop your SFC
methods from this default method.
SFC_DEF.M contains an injector program for full loop injection.

1260 Infinity Analytical SFC System SVC Manual 89


5 Using the Aurora SFC Fusion A5
Running a method on the SFC system

Adjusting the method for use


The Injector timetable includes an additional flush of the Autosampler for
0.1 min by the wash pump. By default, the injection volume is 15 L, which
corresponds to an overfill of the sample loop by approximately three times.

Default Pump setting is 80 % CO2 : 20 % modifier. Typical Flow rate in SFC is


3 mL/min; however, when
SFC_DEF.M is loaded, the flow rate is set to 0 mL/min. Solvent compressibility
should be always set to no compensation for the CO2 Channel, and to the
corresponding value for the modifier.

90 1260 Infinity Analytical SFC System SVC Manual


Using the Aurora SFC Fusion A5 5
Running a method on the SFC system

1260 Infinity Analytical SFC System SVC Manual 91


5 Using the Aurora SFC Fusion A5
Shutting Down the SFC System

Shutting Down the SFC System

The manner of shutting down the SFC depends on the requirement for rapid
equilibration of the system on the next startup, and the duration of the
shutdown. If the system will be shut down for some time, it is probably best to
shut down all components including the DAD and column oven. These
components tend to take longer to reach their stable operating conditions than
the other components in the system. You should always shut down both the
Aurora SFC Fusion A5 and the binary pump if the system is to be idle for a
long time.

Partial Shutdown
Injuries from pressurized CO2
WA R N I N G
Setting the pump to STANDBY does not depressurize the system. Do not attempt to
loosen fittings or perform maintenance under these conditions. Serious skin and
eye injuries can occur as the result of sudden release of CO2 in the liquid or
supercritical state. Always wear gloves and eye protection when maintaining the
Aurora SFC Fusion A5 unit.

Leaving the system pressurized


If the system is to be shut down for less than two hours, a partial pressurized
shutdown is recommended. Press the control buttons (for example, button 2) to
switch the Aurora SFC Fusion A5 and binary pump states to STANDBY. In
this case, the system remains pressurized, and slowly bleeds pressure through
the nozzle. A residual pressure remains in the system when the nozzle closes
fully at lower pressure. The booster remains chilled, and much of the startup
CO2 is preserved. You should be aware that the system is pressurized and not
to attempt maintenance under these conditions.
The detector and column oven are left in the ON state, to maintain their
readiness. You may elect to exit the ChemStation in this state, and should
answer NO to the Shutdown Lamps query that appears during shutdown.
Restarting the ChemStation brings the system to this same state. If more than

92 1260 Infinity Analytical SFC System SVC Manual


Using the Aurora SFC Fusion A5 5
Shutting Down the SFC System

two hours elapse in STANDBY mode, the Aurora SFC Fusion A5 enters the
OFF state, and pressure may be lost at a more rapid rate.

Depressurizing the system


When maintenance is required on the SFC system, such as replacing the
column, the system should be depressurized. It is not necessary to shut down
all modules, but only the pumps and any other devices undergoing
maintenance. To depressurize the system, stop the Aurora SFC Fusion A5
and the binary pump. In the Control menu of the Aurora SFC Fusion A5
booster (the standby state will be selected) check the Depressurize box and
click OK.

If the Aurora SFC Fusion A5 unit itself is to be serviced, select OFF in the Control window.
NOTE This shuts off the BPR heater and booster chiller and allows them to move toward ambient
temperature.

This causes the BPR to home and fully open the CO2 path to depressurize. The
Aurora SFC Fusion A5 unit contains approximately 25 mL of stored CO2.
This amount of CO2 should be vented properly, which takes several minutes.
You should allow the system to drain below 40 bar before cracking any fittings.
At this point the, CO2 is in the vapor state, and represents a small expanded
volume. However, do not inhale vapor directly from a cracked fitting. The
concentration of CO2 emerging from a flow line, even at low pressure, can be
dangerous or even lethal.

Avoid inhaling high concentration of CO2


WA R N I N G
Never inhale vapor issuing from an SFC flow line. Exposure to concentrations of CO2
over 5 % in air can be lethal. Always keep tubes directed away your face. CO2 is
ubiquitous in the atmosphere, but at high levels should be treated with the same
care as other toxic chemicals. Always wear gloves and eye protection for safety.
Avoid inhaling venting gas near open fittings.

Alternatively, if the column oven contains a column switching valve, one path
may be jumpered without a column. The reduced restriction will allow the
system to depressurize much faster. Further, the isolated column can be
exchanged immediately since the contained volume of CO2 is small.

1260 Infinity Analytical SFC System SVC Manual 93


5 Using the Aurora SFC Fusion A5
Shutting Down the SFC System

System contains always 25 ml liquid CO2 under pressure


WA R N I N G
The Aurora SFC Fusion A5 unit contains approximately 25 mL of liquid CO2. The
CO2 must be vented properly since the expanded volume will allow local
concentrations exceeding the OSHA PEL. Always allow the system to depressurize
to below 40 bar before cracking any fittings. Always keep fittings directed away
from the face.

94 1260 Infinity Analytical SFC System SVC Manual


1260 Infinity Analytical SFC System SVC Manual

6
Maintenance and Repair
Inspection and Preventative Maintenance Intervals 96
Daily Inspection and Maintenance 96
Every 3 months 97
Annually 97
As Needed (Corrective) 98
General Maintenance procedures 99
Booster Drawer 99
Replacing Fuses 107
Standard Decontamination 108
Plugged BPR Decontamination 109
Preparing for storage or shipping 109

In this chapter only the SFC specific procedures are described. For procedures
similar to the Agilent module procedures, please refer to the single module
manuals (G1312C, G1329B, G1316C, G1315/65C, G1322A)

Agilent Technologies 95
6 Maintenance and Repair
Inspection and Preventative Maintenance Intervals

Inspection and Preventative Maintenance Intervals

Inspection and maintenance are critical elements of long term reliability and
performance of the Agilent 1260 Infinity Analytical SFC System. Maintenance
falls into two categories: preventative and corrective. Preventative
maintenance intervals vary based on system use. The intervals offered in this
section are for systems with average use of approximately 30 hours per week.
The preventive maintenance intervals can be extended for infrequently used
systems, while heavily used systems may require more frequent preventative
maintenance. Most service can be performed directly by the user or by
in-house maintenance technicians.

Daily Inspection and Maintenance


Verify that power and signal cables are firmly connected and not under
strain.
Inspect all user-serviceable high pressure tubes and transfer lines for
crimping or very tight bends. Replace as necessary.
Wipe up any visible liquid spills or condensation on or near the instrument.
Verify that all covers are securely fastened to the frame.
Inspect all reservoirs to ensure an adequate solvent supply.
Prime the wash pump and modifier pumps. Check that the purge valve
reseals without leaking
Empty all waste containers
Check that the exhaust line is attached to a ventilation system, and that the
ventilation system is drawing.
If an inlet step-down regulator is used with a gas supply system, check that
the inlet pressure is between 40 and 70 bar.
Check the integrity of the SFC Flow path (that is, that column, flow cell etc
have not been removed).
With the system running, visually inspect unions and tees for leaks. Run
diagnostic leak test for added sensitivity.
Check CO2 air monitor, if available, for suitable exposure level [< 5000 ppm
CO2]
96 1260 Infinity Analytical SFC System SVC Manual
Maintenance and Repair 6
Inspection and Preventative Maintenance Intervals

Every 3 months
Run nozzle diagnostic test.
Run system leak test.
Evaluate system calibration.
Check chiller efficiency curve.
If CO2 cylinders are used as supply, change the cylinder seal (approximately
every 10 cylinders) at the next tank change.
Remove visible dust accumulation in the area of the module.

Annually
Have service evaluation and maintenance by a qualified Agilent service
technician.
Perform 3-month maintenance.
Change coolant.
Clean fan access paths.
Upgrade firmware with any recommended safety or performance
enhancements.
Change wash line check valve.
Change injection valve 3-groove rotor.
Inspect high pressure flow cell.
Inspect leak sensor function.
Change pump seals SFC Pump.
Change purge valve frit.
Change needle and needle seat in the autosampler.
Change boaster pump seal.

1260 Infinity Analytical SFC System SVC Manual 97


6 Maintenance and Repair
Inspection and Preventative Maintenance Intervals

As Needed (Corrective)
Change booster and CO2 pump check valves.
Change high pressure transfer lines with metal ferrules or PEEK end
fittings after 10-20 reseals or when leaking.
Change Booster piston (rare).
Exchange BPR head (rare).

98 1260 Infinity Analytical SFC System SVC Manual


Maintenance and Repair 6
General Maintenance procedures

General Maintenance procedures

Booster Drawer
8]ZX`kVakZXVgig^\Z
DjiaZi8K]daYZg

E^hidc

E^hidcgZiV^c^c\g^c\

8]^aaZgeaViZVcYheVXZg

E^hidc

Ejbe

GZiV^c^c\cji

8]ZX`kVakZXVgig^Y\Z

>caZi8K]daYZg

Figure 30 Exploded view of Booster Components

1260 Infinity Analytical SFC System SVC Manual 99


6 Maintenance and Repair
General Maintenance procedures

Removing the vapor shield


The vapor shield is not shown in the graphic Figure 30 on page 99.
Most maintenance procedures require removal of the vapor shield to access
the underlying pump unit.

Tools required A 3/16 hex-drive wrench (mounted on the back of the front cover)

1 Set the Control state to OFF on the Aurora SFC Fusion A5 module.
2 Wait for the chiller temperature to approach room temperature to prevent
significant condensation on the chiller assembly and pump head.
3 Turn the power OFF on the Aurora SFC Fusion A5 module. Unplug the
power cord.
4 Remove the front cover of the Aurora SFC Fusion A5 unit by pulling
gently at the upper left and right indents to the rear of the cover. The cover
will release from its magnetic catch. Lift the cover upward to clear the two
mounting pins at the base and set it aside.

The 3/16 hex wrench used to remove the vapor shield and pump head is stored inside the
HINT removable front cover.

5 While holding the vapor shield with one hand, use the 3/16 hex wrench to
loosen the four cap screws attaching each corner of the shield until they
each disengage from the front panel. The screws are captured in the shield;
do not try to remove them completely.
6 Remove the shield and store it in a safe location. Do not use a container for
disassembled parts. This will scratch the plastic and impair visibility of the
pump head during operation.

100 1260 Infinity Analytical SFC System SVC Manual


Maintenance and Repair 6
General Maintenance procedures

Replacing the vapor shield


1 Locate the vapor shield approximately over the mounting holes in the
booster drawer front panel.
2 Engage each screw approximately one turn.
3 Inspect the border of the vapor shield to make sure it is in sealing contact
with the foam seal of the drawer face. Adjust as necessary.
4 Tighten the mounting screws to ensure at least 50 % compression of the
foam seal by the shield.
5 Replace the front cover by aligning the two base mounting pins and tilting
forward to engage the magnetic catches.

Replacing Booster Pump check valves

Tools required A 1/4 open-end wrench


A 9/16 open-end wrench

Parts required # p/n Description


1 G4301-00200 Plastic cover with screws

1 Remove the vapor shield.


2 Using the 1/4 and 9/16 wrenches, loosen and remover the inlet or outlet
capillary tube.
3 Using the 9/16 wrench, loosen and remove the desired check valve holder.
The check valve may or may not be extracted with the holder.
4 The inlet check valve assembly includes a PEEK gasket. Set this gasket aside
for reuse. (Inlet CV only)
5 Remove the defective check valve cartridge.
Do not mix up check valve directions
CAUTION
The orientation of the CV cartridge is critical. An arrow on the side of the cartridge
indicates the direction of flow. Make sure the cartridge is installed to allow flow in
the proper direction when installed into the pump head.

1260 Infinity Analytical SFC System SVC Manual 101


6 Maintenance and Repair
General Maintenance procedures

6 Insert the new check valve cartridge into the holder oriented correctly for
the direction of flow (arrow up). The inlet check valve is inserted with the
non-filter end of the cartridge showing. The outlet check valve is inserted
with the filter end of the cartridge showing.
7 Replace the PEEK gasket on the top of the inlet check valve cartridge with
the flat side of the gasket facing the cartridge. (Inlet CV only)
8 Insert the CV holder into the pump head and tighten with a 9/16 wrench.
9 Refasten the inlet or outlet capillary line, holding the CV holder with a
9/16 wrench, and tightening the fitting with the 1/4 wrench to seal.
10 Replace the vapor shield

Removing the Pump head

Tools required A 1/4 open-end wrench


A 9/16 open-end wrench
A 3/16 hex-drive wrench
seal insertion/removal tool
Ultrasonic bath
Isopropanol
Deionized water

Parts required # p/n Description


1 G4301-60220 Boaster pump piston kit
1 Piston seal

Each time the booster pump head is removed, the piston seal should be exchanged, since
NOTE the seal surface may be easily scratched or distorted during removal. For this reason, the
procedures are bundled. Cleaning the pump head is optional after visual inspection. Sealing
surfaces of the pump head are critical to successful operation. Never use metal tools or
paper toweling to wipe, probe or contact these surfaces.

1 Set the Aurora SFC Fusion A5 Control mode to OFF.


2 Wait for the pump head to reach room temperature.
3 Power off the Aurora SFC Fusion A5 module.
4 Remove the vapor shield

102 1260 Infinity Analytical SFC System SVC Manual


Maintenance and Repair 6
General Maintenance procedures

5 Using the 9/16 and 1/4 wrenches, remove the inlet line from the inlet
check valve holder and the outlet line from the outlet check valve holder.
Danger of piston breakage
CAUTION
Be careful not to break the piston when removing the pump head. Twisting the
pump head can cause the piston to break.

6 Using the 3/16 hex-drive, carefully remove the two knurled nuts at the
front of the pump head.
7 Carefully separate the pump head from the pump. Move the pump head
straight out from the pump and remove it from the piston. Be careful not to
break or damage the piston. Also remove the seal from the piston if it did
not stay in the pump head.
8 If the seal remains with the pump head, insert the flanged end of the seal
insertion/removal tool into the seal cavity. Tilt it slightly so that flange is
under the seal and pull out the seal.

Inspecting and Cleaning the Pump Head


1 Visually inspect the piston seal cavity in the pump head. Use magnification
if necessary. Remove any foreign material using a cotton swab, or
equivalent, and avoid scratching the sealing surfaces. Be sure no fibers
from the cleaning swab remain in the components.
2 The pump head, may be further cleaned as follows:
a Remove inlet and outlet check valves.
b Clean with 50 % isopropanol in water in an ultrasonic bath for at least
30 min, followed by rinsing for at least 10 min in 100 % isopropanol. Be
sure that all particles loosened by the above procedures have been
removed from the components before re-assembly.
c Replace the check valves.
3 Wipe off any residual liquid from external (non-sealing) surfaces with a soft
cloth such as a microfiber towel.

1260 Infinity Analytical SFC System SVC Manual 103


6 Maintenance and Repair
General Maintenance procedures

Replacing the Piston Seal


1 Sonicate or soak the new seal in isopropanol for 15 min to clean and
provide lubrication for installing.
2 Place the replacement seal on the rod-shaped end of the seal
insertion/removal tool so that the spring is visible when the seal is fully
seated on the tool.
3 Insert the tool into the pump head so that the open side of the seal enters
first, facing the high-pressure cavity of the pump head.
4 Be careful to line up the seal with the cavity while inserting. Withdraw the
tool, leaving the seal in the pump head.
When you look into the pump head cavity, only the polymer portion of the
seal should be visible.

Replacing the Pump Head


1 Fill the pump head cavity about one third full with isopropyl alcohol.
2 Wet the piston tip with a few drops of isopropyl alcohol.
3 Holding an absorbent towel beneath the pump head assembly, line up the
pump head and carefully slide it into place. Be sure that the inlet valve is on
the bottom and the outlet valve is on the top. Do not force the pump head
into place.
4 Finger tighten both knurled nuts into place. To tighten firmly, alternately
turn nuts 1/4 turn while gently wiggling the pump head to center it.
5 Re-attach the inlet and outlet lines.

104 1260 Infinity Analytical SFC System SVC Manual


Maintenance and Repair 6
General Maintenance procedures

Cleaning or Replacing Booster Pump Piston

Tools required Tools for removing the vapor shield (see Removing the vapor shield on page 100) and pump
head (see Removing the Pump head on page 102)
A 9/64 hex-drive wrench

Parts required # Description


1 Piston replacement kit

In most cases, this procedure will be used only to replace a broken piston. Pumping CO2
NOTE
does not tend to leave deposits on the piston. Development of such deposits warrants
examination of the CO2 supply system and correcting the source of the deposited materials.
Release of extraneous materials into the CO2 supply system may cause contamination of
the Agilent 1260 Infinity Analytical SFC System.

1 Remove the Vapor Shield.


2 Remove the pump head.
3 Clean the pump head.
Take care not to break coolant tubes
CAUTION
Use care removing the chiller assembly from the mounting posts. The assembly is
connected to a circulation pump behind the drawer panel. Do not pull the flow lines
hard as this may loosen or crimp the tubes and cause the chiller to lose efficiency or
cause leaks in the secondary cooler system.

4 With a gentle rocking motion, loosen the chiller plate assembly and
carefully slide it forward off the pump head mounting posts. Carefully twist
the assembly out of the way.
5 Use the 9/64 hex wrench to unscrew the two cap screws attaching the
spacer and very carefully remove the spacer by pulling straight back. This
fully exposes the piston and retaining ring.
6 Remove the retaining ring by prying it out with a small blunt instrument or
tweezers at the slot provided.
7 Grasp the metal base of the piston assembly so that you avoid exerting any
side load on the sapphire rod, and remove the piston from the slot in the
carrier by sliding it up.

1260 Infinity Analytical SFC System SVC Manual 105


6 Maintenance and Repair
General Maintenance procedures

8 Grasp the metal base of the replacement piston assembly, and insert it into
the slot in the piston carrier until it bottoms in the slot.
9 Replace the retaining ring and spacer. Reattach the spacer mounting
screws. If properly positioned, the spacer should be pressed into the foam
wall seal.
10 Gently slide the chiller back onto the pump mounting posts and firmly
press it onto the spacer. If properly positioned, the chiller heat exchanger
should now be pressed into the foam wall seal.
11 Replace the piston seal
12 Replace the pump head
13 Replace the vapor shield

Replacing the CO2 Inlet Filter


1 Unscrew the filter closure from the filter housing.
2 Use a seal insertion/removal tool or a non-metallic object (such as a wooden
toothpick) to remove the large seal that remains in the housing
3 Unscrew the old filter and remove the small seal from the filter closure.
4 Place one of the small seals included in the replacement element kit over
one of the new filters from the kit. Screw the new filter into the filter
closure (finger tight).
5 Place one of the large seals from the replacement kit on the filter closure.
Insert the filter closure into the housing and tighten.

106 1260 Infinity Analytical SFC System SVC Manual


Maintenance and Repair 6
General Maintenance procedures

Replacing Fuses
The power entry module of the Aurora SFC Fusion A5 unit contains an
external fuse drawer that is user serviceable.
To replace fuses: (before replacing fuses, first try to determine cause of fuse
activation and repair)

Parts required # Description


1 Fuses with 8A 250V
see shipkit

1 Power down the unit.


2 Disconnect the power cable from the power entry module.
3 Depress the release lever of the fuse drawer and pull the drawer straight
back to remove.
4 Replace blown fuses with 8 A250 V Time Delay fuses of matching size. (A
set of replacement fuses is included shipping kit).
5 Replace the fuse drawer by sliding it into the power entry module until it
locks into place.

1260 Infinity Analytical SFC System SVC Manual 107


6 Maintenance and Repair
General Maintenance procedures

Standard Decontamination
Cleaning
External surfaces of the enclosure can be wiped with a damp soft cloth. More
stubborn marks can be removed with a 50 % isopropanol:water mixture or
mild cleanser such as Soft Scrub. The latter may also be used to remove
surface paint blemishes that may result from normal use.
The vapor shield of the booster drawer should be wiped only with a very soft
cloth such as a microfiber polypropylene cloth, otherwise the surface may be
scratched. Other user-accessible internal surfaces can be cleaned with a damp
cloth.

BPR
The BPR head contacts CO2, modifiers and sample material. To
decontaminate, rinse with 50 % modifier flow at 5 mL/min for 15 min followed
by pure CO2 for 5 min.

Wash Pump
1 Drain the inlet line of old solvent.
2 Flush the inlet line and filter from a small intermediate reservoir to rinse
contaminated residual fluid from the lines.
3 Insert the inlet line into a fresh supply of pure solvent
4 Prime the wash pump for 2 min (four consecutive presses of the 30 s timer)
to clear the remaining flow path.

108 1260 Infinity Analytical SFC System SVC Manual


Maintenance and Repair 6
General Maintenance procedures

Plugged BPR Decontamination


Decontamination of plugged BPR heads may require more aggressive solvents.
In this case use the following procedure:
1 Depressureize the A5 unit completely.
2 Disconnect the BPR inlet and outlet tubes from the BPR drawer.
3 Attach the Inlet tube via a transfer line to waste.
4 Attach a solvent pump to the outlet tube of the BPR head
5 Prime the pump with a suitable solvent for the obstructing material.
6 Flush backwards with strong solvent at 1 mL/min for 20 minutes. Do not
exceed a pressure of 400 bar.
7 If the pump cannot transfer fluid at less than 400 bar discontinue the
operation and perform steps to exchange the BPR head.
8 If the backflush is successful, rinse the BPR head with Isopropanol for 10
minutes at 1 mL/min to clear the strong solvent.
9 Reconnect the BPR inlet and outlet lines.
10 Complete the standard decontamination procedure listed above

Preparing for storage or shipping


If the A5 module needs to be stored in other than its operational location, it is
best to store it in the original factory packaging. This packaging can also be
used to reship the device to a secondary location. If the original packaging is
unavailable, the unit should be stored upright and preferably covered in a
plastic bag or wrap to prevent exposure to dust.
To prepare the unit for storage use the following procedure:
1 Follow the standard decontamination procedure.
2 Depressurize the SFC system completely.
3 Power off the unit.
4 Remove the front panel.
5 Disconnect the wash pump transfer line from the autosampler.
6 Drain the wash pump lines of fluid.
7 Coil the lines to fit in the A5 module behind the removable front panel.

1260 Infinity Analytical SFC System SVC Manual 109


6 Maintenance and Repair
General Maintenance procedures

8 Disconnect the booster pump transfer line from the binary pump at the
pump inlet check valve.
9 Disconnect the BPR return line from the detector.
10 Coil both lines to fit inside the A5 unit behind the removable front panel.
11 Replace the front panel
12 Disconnect the power cord and all signal cables from both ends of the
connection.
13 Store cables and cords in a large plastic zip-lock bag.
14 Cover the unit with a large plastic bag.
15 If the original container is available, place the unit with its left side down in
the packaging. Otherwise, store the unit upright in the storage area.
16 If the HPLC will also be stored or shipped, and will be reconfigured as an
SFC system, the upgrade components can remain in the system.
17 If the two systems are to be permanently separated, uninstall the check
valves, modifier purge valve, 3-groove rotor and high pressure flow cell by
reversing the installation procedures in Hardware Installation on page 34.
Store the components along with the original software disc and any
upgrades with the module.

110 1260 Infinity Analytical SFC System SVC Manual


1260 Infinity Analytical SFC System SVC Manual

7
Parts for Maintenance
G4301 A Aurora SFC Fusion A5 Parts 112
Agilent Module Parts 114

This chapter provides information on parts for maintenance and repair.

Agilent Technologies 111


7 Parts for Maintenance
G4301 A Aurora SFC Fusion A5 Parts

G4301 A Aurora SFC Fusion A5 Parts

p/n Description

G4301-65000 Mainboard

G4301-65010 Motor Drive Board

G4301-60110 Wash Line Check Valve

G4301-60120 (2x) Check Valve Wash Pump

G4301-60130 Wash pump piston

G4301-60140 Wash pump seal

G4301-60500 Nozzle

G4301-60210 Check valve booster pump

G4301-60220 Booster Pump Kit

G4301-60230 CV adapter booster pump

G4301-00200 Plastic cover with screws

G4301-69100 BPR Drawer

G4301-69200 Booster Drawer

112 1260 Infinity Analytical SFC System SVC Manual


Parts for Maintenance 7
G4301 A Aurora SFC Fusion A5 Parts

1260 Infinity Analytical SFC System SVC Manual 113


7 Parts for Maintenance
Agilent Module Parts

Agilent Module Parts

Agilent 1260 Infinity SFC Binary Pump


,
(

&%
.

&
)
+
-

'

*
Figure 31 Agilent 1260 Infinity SFC Binary Pump specific SFC Parts

114 1260 Infinity Analytical SFC System SVC Manual


Parts for Maintenance 7
Agilent Module Parts

p/n Description

G4302-20000 Adapter OV SFC

5023-1803 Adapter male/female

G1311-60065 Damper (600bar with pressure Sensor)

G1312-87306 (2x) Capillary 105x0.17 mm

G1312-87300 Absorber capillary

G1312-60061 Purge valve

G1312-60067 Outlet check valve

G1312-60066 Passive inlet valve

G1312-04100 Mixer Clamp

G1312-87330 Mixer

0905-1420 Pump seals, PE (PumpHead A)

0905-1503 Pump seals, PTFE (PumpHead B)

For all other Part Numbers please refer to the G1312B User manual.
NOTE

1260 Infinity Analytical SFC System SVC Manual 115


7 Parts for Maintenance
Agilent Module Parts

Agilent 1260 Infinity SFC Autosampler

Figure 32 Autosampler Valve

p/n Description

0101-1409 3 Groove rotor seal

0101-1248 Sample Loop 5l

For all other Part Numbers please refer to the G1329B User manual.
NOTE

116 1260 Infinity Analytical SFC System SVC Manual


Parts for Maintenance 7
Agilent Module Parts

DAD SFC Parts

Figure 33 DAD Flowcell SFC

p/n Description

G4301-60100 SFC Flow Cell

79883-68700 High Pressure Flow cell refurbishment Kit

For all other Part Numbers please refer to the G1315/65C User manual.
NOTE

1260 Infinity Analytical SFC System SVC Manual 117


7 Parts for Maintenance
Agilent Module Parts

118 1260 Infinity Analytical SFC System SVC Manual


1260 Infinity Analytical SFC System SVC Manual

8
Identifying Cables
A5 Cables 120
Overview 121
BCD Cables 123
External Contact Cable 125
CAN/LAN Cables 126
Auxiliary Cable 127
RS-232 Cables 128

This chapter provides information on cables used with the Agilent 1200 Infinity
Series modules.

Agilent Technologies 119


8 Identifying Cables
A5 Cables

A5 Cables

p/n Description

00-84-2040 Relay Cable, HD15M/F (VGA type), 6'

00-84-2020 BNC Cable

00-84-2030 Remote Cable, DB9 M/M

00-84-2090 USB Cable - 6'

120 1260 Infinity Analytical SFC System SVC Manual


Identifying Cables 8
Overview

Overview

Never use cables other than the ones supplied by Agilent Technologies to ensure proper
NOTE functionality and compliance with safety or EMC regulations.

BCD cables

p/n Description

G1351-81600 Agilent module to general purpose

Auxiliary

p/n Description

G1322-61600 Agilent vacuum degasser

CAN cables
p/n Description

5181-1516 CAN cable, Agilent module to module, 0.5 m

5181-1519 CAN cable, Agilent module to module, 1 m

RS-232 cables

p/n Description

G1530-60600 RS-232 cable, 2 m

1260 Infinity Analytical SFC System SVC Manual 121


8 Identifying Cables
Overview

LAN cable

p/n Description

5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)

5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection)

122 1260 Infinity Analytical SFC System SVC Manual


Identifying Cables 8
BCD Cables

BCD Cables

One end of these cables provides a 15-pin BCD connector to be connected to


the Agilent modules. The other end depends on the instrument to be
connected to

Agilent Module to General Purpose

p/n G1351-81600 Wire Color Pin Agilent Signal Name BCD Digit
module

Green 1 BCD 5 20

Violet 2 BCD 7 80

Blue 3 BCD 6 40

Yellow 4 BCD 4 10

Black 5 BCD 0 1

Orange 6 BCD 3 8

Red 7 BCD 2 4
Brown 8 BCD 1 2

Gray 9 Digital ground Gray

Gray/pink 10 BCD 11 800

Red/blue 11 BCD 10 400

White/green 12 BCD 9 200

Brown/green 13 BCD 8 100

not connected 14

not connected 15 +5V Low

1260 Infinity Analytical SFC System SVC Manual 123


8 Identifying Cables
BCD Cables

Agilent Module to 3396 Integrators

p/n 03396-60560 Pin 3396 Pin Agilent Signal Name BCD Digit
module

1 1 BCD 5 20

2 2 BCD 7 80

3 3 BCD 6 40

4 4 BCD 4 10

5 5 BCD0 1

6 6 BCD 3 8

7 7 BCD 2 4

8 8 BCD 1 2

9 9 Digital ground

NC 15 +5V Low

124 1260 Infinity Analytical SFC System SVC Manual


Identifying Cables 8
External Contact Cable

External Contact Cable

5 1
10 6
15 11

One end of this cable provides a 15-pin plug to be connected to


Agilent modules interface board. The other end is for general purpose.

Agilent Module Interface Board to general purposes

p/n G1103-61611 Color Pin Agilent Signal Name


module

White 1 EXT 1

Brown 2 EXT 1

Green 3 EXT 2

Yellow 4 EXT 2

Grey 5 EXT 3

Pink 6 EXT 3

Blue 7 EXT 4
Red 8 EXT 4

Black 9 Not connected

Violet 10 Not connected

Grey/pink 11 Not connected

Red/blue 12 Not connected

White/green 13 Not connected

Brown/green 14 Not connected

White/yellow 15 Not connected

1260 Infinity Analytical SFC System SVC Manual 125


8 Identifying Cables
CAN/LAN Cables

CAN/LAN Cables

Both ends of this cable provide a modular plug to be connected to


Agilent modules CAN or LAN connectors.
CAN Cables

p/n Description

5181-1516 CAN cable, Agilent module to module, 0.5 m

5181-1519 CAN cable, Agilent module to module, 1 m

LAN Cables

p/n Description

5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)

5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection)

126 1260 Infinity Analytical SFC System SVC Manual


Identifying Cables 8
Auxiliary Cable

Auxiliary Cable

One end of this cable provides a modular plug to be connected to the


Agilent vacuum degasser. The other end is for general purpose.

Agilent Vacuum Degasser to general purposes

p/n G1322-81600 Color Pin Agilent Signal Name


1100

White 1 Ground

Brown 2 Pressure signal

Green 3

Yellow 4

Grey 5 DC + 5 V IN

Pink 6 Vent

1260 Infinity Analytical SFC System SVC Manual 127


8 Identifying Cables
RS-232 Cables

RS-232 Cables

p/n Description

G1530-60600 RS-232 cable, 2 m

RS232-61600 RS-232 cable, 2.5 m


Instrument to PC, 9-to-9 pin (female). This cable has special pin-out,
and is not compatible with connecting printers and plotters. It's also
called "Null Modem Cable" with full handshaking where the wiring is
made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.

5181-1561 RS-232 cable, 8 m

128 1260 Infinity Analytical SFC System SVC Manual


1260 Infinity Analytical SFC System SVC Manual

9
Appendix
General Safety Information 130
The Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC) 133
Radio Interference 134
Sound Emission 135
Solvent Information 136
Agilent Technologies on Internet 137

This chapter provides addition information on safety, legal and web.

Agilent Technologies 129


9 Appendix
RS-232 Cables

General Safety Information

The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Agilent
Technologies assumes no liability for the customers failure to comply with
these requirements.

Ensure the proper usage of the equipment.


WA R N I N G
The protection provided by the equipment may be impaired.
The operator of this instrument is advised to use the equipment in a manner as
specified in this manual.

Safety Standards
This is a Safety Class I instrument (provided with terminal for protective
earthing) and has been manufactured and tested according to international
safety standards.

Operation
Before applying power, comply with the installation section. Additionally the
following must be observed.
Do not remove instrument covers when operating. Before the instrument is
switched on, all protective earth terminals, extension cords,
auto-transformers, and devices connected to it must be connected to a
protective earth via a ground socket. Any interruption of the protective earth
grounding will cause a potential shock hazard that could result in serious
personal injury. Whenever it is likely that the protection has been impaired,
the instrument must be made inoperative and be secured against any intended
operation.

130 1260 Infinity Analytical SFC System SVC Manual


Appendix 9
RS-232 Cables

Make sure that only fuses with the required rated current and of the specified
type (normal blow, time delay, and so on) are used for replacement. The use of
repaired fuses and the short-circuiting of fuse holders must be avoided.
Some adjustments described in the manual, are made with power supplied to
the instrument, and protective covers removed. Energy available at many
points may, if contacted, result in personal injury.
Any adjustment, maintenance, and repair of the opened instrument under
voltage should be avoided whenever possible. When inevitable, this has to be
carried out by a skilled person who is aware of the hazard involved. Do not
attempt internal service or adjustment unless another person, capable of
rendering first aid and resuscitation, is present. Do not replace components
with power cable connected.
Do not operate the instrument in the presence of flammable gases or fumes.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
Do not install substitute parts or make any unauthorized modification to the
instrument.
Capacitors inside the instrument may still be charged, even though the
instrument has been disconnected from its source of supply. Dangerous
voltages, capable of causing serious personal injury, are present in this
instrument. Use extreme caution when handling, testing and adjusting.
When working with solvents please observe appropriate safety procedures
(e.g. goggles, safety gloves and protective clothing) as described in the material
handling and safety data sheet by the solvent vendor, especially when toxic or
hazardous solvents are used.

1260 Infinity Analytical SFC System SVC Manual 131


9 Appendix
RS-232 Cables

Safety Symbols
Table 15 Safety Symbols

Symbol Description

The apparatus is marked with this symbol when the user should refer to the
instruction manual in order to protect risk of harm to the operator and to
protect the apparatus against damage.

Indicates dangerous voltages.

Indicates a protected ground terminal.

Indicates eye damage may result from directly viewing the light produced
by the deuterium lamp used in this product.

The apparatus is marked with this symbol when hot surfaces are available
and the user should not touch it when heated up.

A WARNING
WA R N I N G
alerts you to situations that could cause physical injury or death.
Do not proceed beyond a warning until you have fully understood and met the
indicated conditions.

A CAUTION
CAUTION
alerts you to situations that could cause loss of data, or damage of equipment.
Do not proceed beyond a caution until you have fully understood and met the
indicated conditions.

132 1260 Infinity Analytical SFC System SVC Manual


Appendix 9
The Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC)

The Waste Electrical and Electronic Equipment (WEEE) Directive


(2002/96/EC)

Abstract
The Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC), adopted by EU Commission on 13 February 2003, is
introducing producer responsibility on all Electric and Electronic appliances
from 13 August 2005.

NOTE

This product complies with the WEEE Directive (2002/96/EC) marking requirements. The
affixed label indicates that you must not discard this electrical/electronic product in
domestic household waste.
Product Category: With reference to the equipment types in the WEEE Directive Annex I,
this product is classed as a Monitoring and Control instrumentation product.
Do not dispose off in domestic household waste
To return unwanted products, contact your local Agilent office, or see www.agilent.com for
more information.

1260 Infinity Analytical SFC System SVC Manual 133


9 Appendix
Radio Interference

Radio Interference

Never use cables other than the ones supplied by Agilent Technologies to
ensure proper functionality and compliance with safety or EMC regulations.

Test and Measurement


If test and measurement equipment is operated with equipment unscreened
cables and/or used for measurements on open set-ups, the user has to assure
that under operating conditions the radio interference limits are still met
within the premises.

134 1260 Infinity Analytical SFC System SVC Manual


Appendix 9
Sound Emission

Sound Emission

Manufacturers Declaration
This statement is provided to comply with the requirements of the German
Sound Emission Directive of 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB.
Sound Pressure Lp < 70 dB (A)
At Operator Position
Normal Operation
According to ISO 7779:1988/EN 27779/1991 (Type Test)

1260 Infinity Analytical SFC System SVC Manual 135


9 Appendix
Solvent Information

Solvent Information

Flow Cell
To protect optimal functionality of your flow-cell:
Avoid the use of alkaline solutions (pH > 9.5) which can attack quartz and
thus impair the optical properties of the flow cell.

Use of Solvents
Observe the following recommendations on the use of solvents.
Brown glass ware can avoid growth of algae.
Avoid the use of the following steel-corrosive solvents:
Solutions of alkali halides and their respective acids (for example,
lithium iodide, potassium chloride, and so on),
High concentrations of inorganic acids like sulfuric acid and nitric acid,
especially at higher temperatures (if your chromatography method
allows, replace by phosphoric acid or phosphate buffer which are less
corrosive against stainless steel),
Halogenated solvents or mixtures which form radicals and/or acids, for
example:
2CHCl3 + O2 2COCl2 + 2HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs
quickly with dried chloroform if the drying process removes the
stabilizing alcohol,
Chromatographic grade ethers, which can contain peroxides (for
example, THF, dioxane, di-isopropylether) such ethers should be filtered
through dry aluminium oxide which adsorbs the peroxides,
Solvents containing strong complexing agents (e.g. EDTA),
Mixtures of carbon tetrachloride with 2-propanol or THF.

136 1260 Infinity Analytical SFC System SVC Manual


Appendix 9
Agilent Technologies on Internet

Agilent Technologies on Internet

For the latest information on products and services visit our worldwide web
site on the Internet at:
http://www.agilent.com
Select Products/Chemical Analysis
It will provide also the latest firmware of the modules for download.

1260 Infinity Analytical SFC System SVC Manual 137


Index

Index

A external contact P
cable 125
Agilent physical specifications 27, 29, 30, 31
on internet 137 power consideration 18
algae 136
F
power consumption 26, 26, 27, 29,
ambient non-operating temperature 26, flow cell 136 30, 31
27, 29, 30, 31 solvent information 136 power cords 19
ambient operating temperature 26, 27, frequency range 26, 27, 29, 30, 31
29, 30, 31 R
auxiliary H RS-232C
cable 121, 127 humidity 26, 27, 29, 30, 31 cable 128
RS-232
B I cable 121
BCD installation
cable 121, 123 power cords 19 S
site requirements 18 safety class I 130
C internet 137 safety
cable general information 130
Auxiliary 121, 127 L standards 30, 31, 27, 29
BCD 121, 123 laboratory bench 20 symbols 132
CAN 126 site requirements 18
LAN
external contact 125 solvents 136
cable 122, 126
LAN 122, 126
line frequency 26, 27, 29, 30, 31 specification
RS-232 121, 128
line voltage 26, 27, 29, 30, 31 physical 27, 26, 31, 29, 30
cables
Specifications 22
overview 121
CAN
N
non-operating altitude 26, 27, 29, 30,
V
cable 126
condensation 21 31 vacuum degasser 30
non-operating temperature 27, 29, 30, voltage range 26, 27, 29, 30, 31
31
D
W
dimensions 26, 26, 27, 29, 30, 31 O weight 26, 26, 27, 29, 30, 31
operating Altitude 26, 27, 29, 30, 31
E
operating temperature 27, 29, 30, 31
electrostatic discharge (ESD) 61

138 1260 Infinity Analytical SFC System SVC Manual


www.agilent.com

In This Book

The manual contains the following information:


Introduction
Site Requirements and Specifications
Installing the System
Configuring the System
Using the Aurora SFC Fusion A5
Maintenance and Repair
Parts for Maintenance
Identifying Cables
Appendix

Agilent Technologies 2010 - 2012


Printed in Germany
05/2012

*G4301-90100*
*G4301-90100*
G4301-90100
Rev. C

Agilent Technologies

You might also like