Asme Sec V A-2 RT PDF
Asme Sec V A-2 RT PDF
Asme Sec V A-2 RT PDF
T-210 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
T-220 General Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
T-221 Procedure Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
T-222 Surface Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
T-223 Backscatter Radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
T-224 System of Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
T-225 Monitoring Density Limitations of Radiographs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
T-226 Extent of Examination. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
T-230 Equipment and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
T-231 Film . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
T-232 Intensifying Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
T-233 Image Quality Indicator (IQI) Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
T-234 Facilities for Viewing of Radiographs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
T-260 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
T-261 Source Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
T-262 Densitometer and Step Wedge Comparison Film . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
T-270 Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
T-271 Radiographic Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
T-272 Radiation Energy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
T-273 Direction of Radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
T-274 Geometric Unsharpness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
T-275 Location Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
T-276 IQI Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
T-277 Use of IQIs to Monitor Radiographic Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
T-280 Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
T-281 Quality of Radiographs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
T-282 Radiographic Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
T-283 IQI Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
T-284 Excessive Backscatter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
T-285 Geometric Unsharpness Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
T-286 Evaluation by Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
T-290 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
T-291 Radiographic Technique Documentation Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
T-292 Radiograph Review Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Figure
T-275 Location Marker Sketches. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Tables
T-233.1 Hole-Type IQI Designation, Thickness, and Hole Diameters . . . . . . . . . . . . . . . . . . . . 22
T-233.2 Wire IQI Designation, Wire Diameter, and Wire Identity . . . . . . . . . . . . . . . . . . . . . . 23
T-276 IQI Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
T-283 Equivalent Hole-Type IQI Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
16
Mandatory Appendices
Appendix I In-Motion Radiography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I-210 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I-220 General Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I-223 Backscatter Detection Symbol Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I-260 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I-263 Beam Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I-270 Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I-274 Geometric and In-Motion Unsharpness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I-275 Location Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I-277 Placement and Number of IQIs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
I-279 Repaired Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
I-280 Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
I-285 Geometric and In-Motion Unsharpness Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Appendix II Real-Time Radioscopic Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-210 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-220 General Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-221 Procedure Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-230 Equipment and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-231 Radioscopic Examination Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-235 Calibration Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-236 Calibrated Line Pair Test Pattern and Step Wedge . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-237 Equivalent Performance Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-260 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-263 System Performance Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
II-264 Measurement With a Calibration Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
II-270 Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
II-278 System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
II-280 Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
II-286 Factors Affecting System Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
II-290 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
II-291 Radioscopic Technique Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
II-292 Evaluation by Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Appendix III Digital Image Acquisition, Display, and Storage for Radiography and
Radioscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-210 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-220 General Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-221 Procedure Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-222 Original Image Artifacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-230 Equipment and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-231 Digital Image Examination Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-234 Viewing Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-236 Calibrated Optical Line Pair Test Pattern and Optical Density Step Wedge . . . . . . 35
III-250 Image Acquisition and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-255 Area of Interest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-258 System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
III-260 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
III-263 System Performance Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
III-280 Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
III-286 Factors Affecting System Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
III-287 System-Induced Artifacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
III-290 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
17
III-291 Digital Imaging Technique Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
III-292 Evaluation by Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Appendix VI Digital Image Acquisition, Display, Interpretation, and Storage of Radiographs for
Nuclear Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
VI-210 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
VI-220 General Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
VI-221 Supplemental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
VI-222 Written Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
VI-223 Personnel Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
VI-230 Equipment and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-231 System Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-232 System Spot Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-240 System Performance Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-241 Spatial Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-242 Contrast Sensitivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-243 Dynamic Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-244 Spatial Linearity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-250 Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-251 Spatial Resolution Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-252 Contrast Sensitivity Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-253 Dynamic Range Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-254 Spatial Linearity Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
VI-260 Demonstration of System Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-261 Procedure Demonstration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-262 Processed Targets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
18
VI-261 Procedure Demonstration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-262 Processed Targets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-263 Changes in Essential Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-264 Frequency of Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-265 Changes in System Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-270 Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-271 System Performance Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-272 Artifacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-273 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-280 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-281 Process Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-282 Interpretation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-283 Baseline. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-290 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-291 Reporting Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
VI-292 Archiving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
VI-A-210 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
VI-A-220 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
VI-A-221 Reference Film . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
VI-A-230 Equipment and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
VI-A-231 Reference Targets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
VI-A-232 Spatial Resolution Targets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
VI-A-233 Contrast Sensitivity Targets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
VI-A-234 Dynamic Range Targets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
VI-A-235 Spatial Linearity Targets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
VI-A-240 Miscellaneous Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
VI-A-241 Material. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
VI-A-242 Film Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
VI-A-243 Spatial Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
VI-A-244 Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
VI-A-245 Linearity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Figures
I-263 Beam Width Determination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
VI-A-1 Reference Film . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Nonmandatory Appendices
Appendix A Technique Sketches for Pipe or Tube Welds. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
A-210 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
19 A02
Single-Wall Radiographic Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Double-Wall Radiographic Techniques. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Figures
C-210-1 Side and Top Views of Hole-Type IQI Placements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
C-210-2 Side and Top Views of Hole-Type IQI Placements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
C-210-3 Side and Top Views of Hole-Type IQI Placements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
C-210-4 Side and Top Views of Hole-Type IQI Placements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
D-210-1 Complete Circumference Cylindrical Component. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
D-210-2 Section of Circumference 240 deg. or More Cylindrical Component. . . . . . . . . . . . . . 58
D-210-3 Section(s) of Circumference Less Than 240 deg. Cylindrical Component. . . . . . . . . . 58
D-210-4 Section(s) of Circumference Equal to or More Than 120 deg. and Less Than 240
deg. Cylindrical Component. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
D-210-5 Complete Circumferential Welds Spherical Component . . . . . . . . . . . . . . . . . . . . . . . . . . 59
D-210-6 Welds in Segments of Spherical Component . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
D-210-7 Plan View A-A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
D-210-8 Array of Objects in a Circle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
20
ARTICLE 2
RADIOGRAPHIC EXAMINATION
01 T-221.1 Radiographic examination shall be per- A lead symbol B, with minimum dimensions of
1
A02 formed in accordance with a written procedure. Each 2 in. (13 mm) in height and 116 in. (1.6 mm) in
procedure shall include at least the following informa- thickness, shall be attached to the back of each film
tion, as applicable: holder during each exposure to determine if backscatter
(a) material type and thickness range, radiation is exposing the film.
(b) isotope or maximum X-ray voltage used,
(c) minimum source-to-object distance,
(d) distance from source side of object to the film T-224 System of Identification
at the minimum source-to-object distance, A system shall be used to produce permanent identi-
(e) maximum source size, fication on the radiograph traceable to the contract,
(f) film brand and designation, and component, weld or weld seam, or part numbers, as
(g) screens used. appropriate. In addition, the Manufacturers symbol or
name and the date of the radiograph shall be plainly
T-221.2 Procedure Demonstration. Demonstration
and permanently included on the radiograph. This identi-
of the density and IQI image requirements of the written
fication system does not necessarily require that the
procedure on production or technique radiographs shall
information appear as radiographic images. In any case,
be considered satisfactory evidence of compliance with
this information shall not obscure the area of interest.
that procedure.
21
T-226 2001 SECTION V T-233.2
T-226 Extent of Examination T-233 Image Quality Indicator (IQI) Design A02
The extent of radiographic examination shall be as T-233.1 Standard IQI Design. IQIs shall be either
specified by the referencing Code Section. the hole type or the wire type. Hole-type IQIs shall
be manufactured and identified in accordance with the
requirements or alternates allowed in SE-1025. Wire-
type IQIs shall be manufactured and identified in accord-
T-230 EQUIPMENT AND MATERIALS ance with the requirements or alternates allowed in
T-231 Film SE-747, except that the largest wire number or the
identity number may be omitted. ASME standard IQIs
A02 T-231.1 Selection. Radiographs shall be made using shall consist of those in Table T-233.1 for hole type
industrial radiographic film. and those in Table T-233.2 for wire type.
A02 T-231.2 Processing. Standard Guide for Controlling
T-233.2 Alternative IQI Design. IQIs designed and
the Quality of Industrial Radiographic Film Processing,
manufactured in accordance with other national or
SE-999, or paragraphs 23 through 26 of Standard Guide
international standards may be used provided the re-
for Radiographic Examination SE-94 shall be used as
quirements of either (a) or (b) below, and the material
a guide for processing film.
requirements of T-276.1 are met.
(a) Hole Type IQIs. The calculated Equivalent IQI
Sensitivity (EPS), per SE-1025, Appendix X1, is equal
T-232 Intensifying Screens
to or better than the required standard hole type IQI.
Intensifying screens may be used when performing (b) Wire Type IQIs. The alternative wire IQI essential
radiographic examination in accordance with this Ar- wire diameter is equal to or less than the required
ticle. standard IQI essential wire.
22
T-234 ARTICLE 2 RADIOGRAPHIC EXAMINATION T-262.3
A02 TABLE T-233.2 (a) X-Ray Machines. For X-ray machines operating
WIRE IQI DESIGNATION, WIRE DIAMETER, at 500 kV and less, the focal spot size may be determined
AND WIRE IDENTITY by the pinhole method,1 or in accordance with SE-
Set A Set B 1165, Standard Test Method for Measurement of Focal
Spots of Industrial X-Ray Tubes by Pinhole Imaging.
Wire Wire Wire Wire
Diameter, in. (mm) Identity Diameter, in. (mm) Identity
(b) Iridium-192 Sources. For Iridium-192, the source
size may be determined in accordance with SE-1114,
0.0032 (0.08) 1 0.010 (0.25) 6 Standard Test Method for Determining the Focal Size
0.004 (0.01) 2 0.013 (0.33) 7 of Iridium-192 Industrial Radiographic Sources.
0.005 (0.13) 3 0.016 (0.41) 8
0.0063 (0.16) 4 0.020 (0.51) 9
0.008 (0.20) 5 0.025 (0.64) 10
0.010 (0.25) 6 0.032 (0.81) 11 T-262 Densitometer and Step Wedge
Comparison Film
Set C Set D
T-262.1 Densitometers. Densitometers shall be cali-
Wire brated at least every 90 days during use as follows:
Diameter, Wire Wire Wire
in. (mm) Identity Diameter, in. (mm) Identity (a) A national standard step tablet or a step wedge
calibration film, traceable to a national standard step
0.032 (0.81) 11 0.100 (2.54) 16 tablet and having at least 5 steps with neutral densities
0.040 (1.02) 12 0.126 (3.20) 17
from at least 1.0 through 4.0, shall be used. The step
0.050 (1.27) 13 0.160 (4.06) 18
0.063 (1.60) 14 0.200 (5.08) 19 wedge calibration film shall have been verified within
0.080 (2.03) 15 0.250 (6.35) 20 the last year by comparison with a national standard
0.100 (2.54) 16 0.320 (8.13) 21 step tablet.
(b) The densitometer manufacturers step-by-step in-
structions for the operation of the densitometer shall
be followed.
(c) The density steps closest to 1.0, 2.0, 3.0, and
4.0 on the national standard step tablet or step wedge
T-234 Facilities for Viewing of Radiographs
calibration film shall be read.
Viewing facilities shall provide subdued background (d) The densitometer is acceptable if the density
lighting of an intensity that will not cause troublesome readings do not vary by more than 0.05 density units
reflections, shadows, or glare on the radiograph. Equip- from the actual density stated on the national standard
ment used to view radiographs for interpretation shall step tablet or step wedge calibration film.
provide a variable light source sufficient for the essential
IQI hole or designated wire to be visible for the T-262.2 Step Wedge Comparison Films. Step wedge
specified density range. The viewing conditions shall comparison films shall be verified prior to first use,
be such that light from around the outer edge of the unless performed by the manufacturer, as follows:
radiograph or coming through low-density portions of (a) The density of the steps on a step wedge compari-
the radiograph does not interfere with interpretation. son film shall be verified by a calibrated densitometer.
(b) The step wedge comparison film is acceptable
if the density readings do not vary by more than
0.1 density units from the density stated on the step
T-260 CALIBRATION wedge comparison film.
T-261 Source Size T-262.3 Periodic Verification
T-261.1 Verification of Source Size. The equipment (a) Densitometers. Periodic cablibration verification
manufacturers or suppliers publications, such as techni- checks shall be performed as described in T-262.1 at
cal manuals, decay curves, or written statements docu- the beginning of each shift, after 8 hr of continuous
menting the actual or maximum source size or focal use, or after change of apertures, whichever comes
spot, shall be acceptable as source size verification. first. The densitometer is acceptable if the density
23
T-262.3 2001 SECTION V T-275
readings are within 0.05 of the calibration readings separate the images of the source-side and film-side
determined in T-262.1(c). portions of the weld so that there is no overlap of the
(b) Step Wedge Comparison Films. Verification areas to be interpreted. When complete coverage is
checks shall be performed annually per T-262.2. required, a minimum of two exposures taken 90 deg.
to each other shall be made for each joint.
T-262.4 Documentation
(2) As an alternative, the weld may be radio-
(a) Densitometer calibration readings required by T-
graphed with the radiation beam positioned so that the
262.1(c) shall be recorded in an appropriate calibra-
images of both walls are superimposed. When complete
tion log.
coverage is required, a minimum of three exposures
(b) Periodic verification readings required by T-262.3
taken at either 60 deg. or 120 deg. to each other shall
do not have to be recorded.
be made for each joint.
(3) Additional exposures shall be made if the
required radiographic coverage cannot be obtained using
T-270 EXAMINATION
the minimum number of exposures indicated in (b)(1)
T-271 Radiographic Technique2 or (b)(2) above.
A single-wall exposure technique shall be used for
radiography whenever practical. When it is not practical T-272 Radiation Energy 01
to use a single-wall technique, a double-wall technique
The radiation energy employed for any radiographic
shall be used. An adequate number of exposures shall
technique shall achieve the density and IQI image
be made to demonstrate that the required coverage has
requirements of this Article.
been obtained.
T-271.1 Single-Wall Technique. In the single-wall
T-273 Direction of Radiation
technique, the radiation passes through only one wall
of the weld (material), which is viewed for acceptance The direction of the central beam of radiation should
on the radiograph. be centered on the area of interest whenever practical.
T-271.2 Double-Wall Technique. When it is not
practical to use a single-wall technique, one of the T-274 Geometric Unsharpness
following double-wall techniques shall be used.
Geometric unsharpness of the radiograph shall be
(a) Single-Wall Viewing. For materials and for welds
determined in accordance with:
in components, a technique may be used in which the
radiation passes through two walls and only the weld
Ug p Fd /D
(material) on the film-side wall is viewed for acceptance
on the radiograph. When complete coverage is required
for circumferential welds (materials), a minimum of where
three exposures taken 120 deg. to each other shall be Ugp geometric unsharpness
made. Fp source size: the maximum projected dimension
of the radiating source (or effective focal spot)
(b) Double-Wall Viewing. For materials and for welds
in the plane perpendicular to the distance D
in components 312 in. (89 mm) or less in nominal
from the weld or object being radiographed, in.
outside diameter, a technique may be used in which
Dp distance from source of radiation to weld or
the radiation passes through two walls and the weld
object being radiographed, in.
(material) in both walls is viewed for acceptance on
dp distance from source side of weld or object be-
the same radiograph. For double-wall viewing, only a
ing radiographed to the film
source-side IQI shall be used. Care should be exercised
to ensure that the required geometric unsharpness is NOTE: Refer to Standard Guide for Radiographic Testing SE-94 for
not exceeded. If the geometric unsharpness requirement a method of determining geometric unsharpness. Alternatively, a
nomograph as shown in Standard Guide for Radiographic Testing
cannot be met, then single-wall viewing shall be used. SE-94 may be used.
(1) For welds, the radiation beam may be offset
from the plane of the weld at an angle sufficient to
T-275 Location Markers
2
Sketches showing suggested source, film, and IQI placements for pipe Location markers (see Fig. T-275), which are to
or tube welds are illustrated in Article 2, Nonmandatory Appendix A. appear as radiographic images on the film, shall be
24
T-275 ARTICLE 2 RADIOGRAPHIC EXAMINATION T-277.1
placed on the part, not on the exposure holder /cassette. T-276 IQI Selection
Their locations shall be permanently marked on the
T-276.1 Material. IQIs shall be selected from either A02
surface of the part being radiographed when permitted,
the same alloy material group or grade as identified
or on a map, in a manner permitting the area of interest
in SE-1025, or SE-747, as applicable, or from an alloy
on a radiograph to be accurately traceable to its location
material group or grade with less radiation absorption
on the part, for the required retention period of the
than the material being radiographed.
radiograph. Evidence shall also be provided on the
radiograph that the required coverage of the region T-276.2 Size. The designated hole IQI or essential
being examined has been obtained. Location markers wire shall be as specified in Table T-276. A thinner
shall be placed as follows. or thicker hole-type IQI may be substituted for any
section thickness listed in Table T-276, provided an
T-275.1 Single-Wall Viewing
equivalent IQI sensitivity is maintained. See T-283.2.
(a) Source-Side Markers. Location markers shall be
(a) Welds With Reinforcements. The thickness on
placed on the source side when radiographing the
which the IQI is based is the nominal single-wall
following:
thickness plus the estimated weld reinforcement not to
(1) flat components or longitudinal joints in cylin- exceed the maximum permitted by the referencing Code
drical or conical components; Section. Backing rings or strips shall not be considered
(2) curved or spherical components whose concave as part of the thickness in IQI selection. The actual
side is toward the source and when the source-to- measurement of the weld reinforcement is not required.
material distance is less than the inside radius of the (b) Welds Without Reinforcements. The thickness
component; on which the IQI is based is the nominal single-wall
(3) curved or spherical components whose convex thickness. Backing rings or strips shall not be considered
side is toward the source. as part of the weld thickness in IQI selection.
(b) Film-Side Markers
(1) Location markers shall be placed on the film T-276.3 Welds Joining Dissimilar Materials or
side when radiographing either curved or spherical Welds With Dissimilar Filler Metal. When the weld
components whose concave side is toward the source metal is of an alloy group or grade that has a radiation
and when the source-to-material distance is greater attenuation that differs from the base material, the IQI
than the inside radius. material selection shall be based on the weld metal
(2) As an alternative to source-side placement in and be in accordance with T-276.1. When the density
T-275.1(a)(1), location markers may be placed on the limits of T-282.2 cannot be met with one IQI, and the
film side when the radiograph shows coverage beyond exceptional density area(s) is at the interface of the
the location markers to the extent demonstrated by Fig. weld metal and the base metal, the material selection
T-275, sketch (e), and when this alternate is documented for the additional IQIs shall be based on the base
in accordance with T-291. material and be in accordance with T-276.1.
(c) Either Side Markers. Location markers may be
placed on either the source side or film side when
radiographing either curved or spherical components T-277 Use of IQIs to Monitor Radiographic
whose concave side is toward the source and the Examination
source-to-material distance equals the inside radius
of the component. T-277.1 Placement of IQIs A02
(a) Source-Side IQI(s). The IQI(s) shall be placed
T-275.2 Double-Wall Viewing. For double-wall on the source side of the part being examined, except
viewing, at least one location marker shall be placed for the condition described in T-277.1(b).
adjacent to the weld (or on the material in the area When, due to part or weld configuration or size, it
of interest) for each radiograph. is not practical to place the IQI(s) on the part or weld,
the IQI(s) may be placed on a separate block. Separate
T-275.3 Mapping the Placement of Location Mark- blocks shall be made of the same or radiographically
ers. When inaccessibility or other limitations prevent similar materials (as defined in SE-1025) and may be
the placement of markers as stipulated in T-275.1 and used to facilitate IQI positioning. There is no restriction
T-275.2, a dimensioned map of the actual marker on the separate block thickness, provided the IQI /area-
placement shall accompany the radiographs to show of-interest density tolerance requirements of T-282.2
that full coverage has been obtained. are met.
25
Fig. T-275
Source side
acceptable
Film side Source side
unacceptable acceptable
Film side
Film side Source side unacceptable
unacceptable acceptable
Flat component or longitudinal seam Curved components with radiation source to Curved components with convex surface
[See T-275.1(a)(1)] film distance less than radius of component towards radiation source
[See sketch (e) for alternate] [See T-275.1(a)(2)] [See T-275.1(a)(3)]
(a) (b) (c)
26
Either side
location marker
is acceptable
D
2001 SECTION V
x Mf x
Source side
unacceptable Film side
Source side marker alternate
acceptable
Flat component or logitudinal seam
Curved components with radiation source to x = (t / D) (Mf / 2) Curved components with radiation source
film distance greater than radius of curvature x = additional required coverage at center curvature
[See T-275.1(b)(1)] beyond film side location marker [See T-275.1(c)]
(d) t = component thickness (f)
Mf = film side location marker interval
LEGEND: Radiation source D = source to component distance
[See T-275.1(b)(2)]
Location marker (e)
Component center
TABLE T-276
IQI SELECTION
IQI
(1) The IQI on the source side of the separate when used, the lead letter F, may be placed in the
block shall be placed no closer to the film than the area of interest.
source side of the part being radiographed.
(2) The separate block shall be placed as close T-277.2 Number of IQIs. When one or more film
as possible to the part being radiographed. holders are used for an exposure, at least one IQI
(3) When hole-type IQIs are used, the block dimen- image shall appear on each radiograph except as outlined
sions shall exceed the IQI dimensions such that the in (b) below.
outline of at least three sides of the IQI image shall (a) Multiple IQIs. If the requirements of T-282 are
be visible on the radiograph. met by using more than one IQI, one shall be representa-
(b) Film-Side IQI(s). Where inaccessibility prevents tive of the lightest area of interest and the other the
hand placing the IQI(s) on the source side, the IQI(s) darkest area of interest; the intervening densities on
shall be placed on the film side in contact with the the radiograph shall be considered as having acceptable
part being examined. A lead letter F shall be placed density.
adjacent to or on the IQI(s), but shall not mask the (b) Special Cases3
essential hole where hole IQIs are used. (1) For cylindrical components where the source
(c) IQI Placement for Welds Hole IQIs. The is placed on the axis of the component for a single
IQI(s) may be placed adjacent to or on the weld. The exposure, at least three IQIs, spaced approximately 120
identification number(s) and, when used, the lead letter deg. apart, are required under the following conditions:
F, shall not be in the area of interest, except when (a) When the complete circumference is radio-
geometric configuration makes it impractical. graphed using one or more film holders, or;
(d) IQI Placement for Welds Wire IQIs. The (b) When a section or sections of the circumfer-
IQI(s) shall be placed on the weld so that the length ence, where the length between the ends of the outermost
of the wires is perpendicular to the length of the weld. sections span 240 or more deg., is radiographed using
The identification numbers and, when used, the lead one or more film holders. Additional film locations
letter F, shall not be in the area of interest, except may be required to obtain necessary IQI spacing.
when geometric configuration makes it impractical.
(e) IQI Placement for Materials Other Than Welds.
3
The IQI(s) with the IQI identification number(s), and, Refer to Nonmandatory Appendix D for additional guidance.
27
T-277.2 2001 SECTION V T-282.2
(2) For cylindrical components where the source (8) In order to maintain the continuity of records
is placed on the axis of the component for a single involving subsequent exposures, all radiographs exhib-
exposure, at least three IQIs, with one placed at each iting IQIs that qualify the techniques permitted in
end of the span of the circumference radiographed and accordance with (1) through (6) above shall be retained.
one in the approximate center of the span, are required
T-277.3 Shims Under Hole IQIs. For welds, a shim
under the following conditions:
of material radiographically similar to the weld metal
(a) When a section of the circumference, the
shall be placed between the part and the IQI, if needed,
length of which is greater than 120 deg. and less than
so that the radiographic density throughout the area of
240 deg., is radiographed using just one film holder, or; interest is no more than minus 15% from (lighter than)
(b) When a section or sections of the circumfer- the radiographic density through the IQI.
ence, where the length between the ends of the outermost The shim dimensions shall exceed the IQI dimensions
sections span less than 240 deg., is radiographed using such that the outline of at least three sides of the IQI
more than one film holder. image shall be visible in the radiograph.
(3) In (1) and (2) above, where sections of longitu-
dinal welds adjoining the circumferential weld are
radiographed simultaneously with the circumferential
T-280 EVALUATION
weld, an additional IQI shall be placed on each longitu-
dinal weld at the end of the section most remote T-281 Quality of Radiographs
from the junction with the circumferential weld being All radiographs shall be free from mechanical, chemi-
radiographed. cal, or other blemishes to the extent that they do not
(4) For spherical components where the source is mask and are not confused with the image of any
placed at the center of the component for a single discontinuity in the area of interest of the object being
exposure, at least three IQIs, spaced approximately 120 radiographed. Such blemishes include, but are not lim-
deg. apart, are required under the following conditions: ited to:
(a) When a complete circumference is radio- (a) fogging;
graphed using one or more film holders, or; (b) processing defects such as streaks, watermarks,
(b) When a section or sections of a circumfer- or chemical stains;
ence, where the length between the ends of the outermost (c) scratches, finger marks, crimps, dirtiness, static
sections span 240 or more deg., is radiographed using marks, smudges, or tears;
one or more film holders. Additional film locations (d) false indications due to defective screens.
may be required to obtain necessary IQI spacing.
(5) For spherical components where the source is
placed at the center of the component for a single T-282 Radiographic Density
exposure, at least three IQIs, with one placed at each T-282.1 Density Limitations. The transmitted film
end of the radiographed span of the circumference density through the radiographic image of the body of
radiographed and one in the approximate center of the the appropriate hole IQI or adjacent to the designated
span, are required under the following conditions: wire of a wire IQI and the area of interest shall be
(a) When a section of a circumference, the 1.8 minimum for single film viewing for radiographs
length of which is greater than 120 deg. and less than made with an X-ray source and 2.0 minimum for
240 deg., is radiographed using just one film holder, or; radiographs made with a gamma ray source. For com-
(b) When a section or sections of a circumfer- posite viewing of multiple film exposures, each film
ence, where the length between the ends of the outermost of the composite set shall have a minimum density of
sections span less than 240 deg. is radiographed using 1.3. The maximum density shall be 4.0 for either single
more than one film holder. or composite viewing. A tolerance of 0.05 in density
is allowed for variations between densitometer readings.
(6) In (4) and (5) above, where other welds are
radiographed simultaneously with the circumferential T-282.2 Density Variation
weld, one additional IQI shall be placed on each (a) General. If the density of the radiograph anywhere
other weld. through the area of interest varies by more than minus
(7) When an array of components in a circle is 15% or plus 30% from the density through the body
radiographed, at least one IQI shall show on each of the hole IQI or adjacent to the designated wire of
component image. a wire IQI, within the minimum /maximum allowable
28
T-282.2 ARTICLE 2 RADIOGRAPHIC EXAMINATION T-291
29
T-291 2001 SECTION V T-292
(e) X-ray machine focal spot size or isotope physical T-292 Radiograph Review Form
sources size
The Manufacturer shall prepare a radiograph review
(f) base material type and thickness, weld thickness, form. As a minimum, the following information shall
weld reinforcement thickness, as applicable be provided.
(g) minimum source-to-object distance (a) a listing of each radiograph location
(h) distance from source side of object to the film (b) the information required in T-291, by inclusion
at the minimum source-to-object distance or by reference
(i) film manufacturer and Manufacturers type/desig- (c) evaluation and disposition of the material(s) or
nation weld(s) examined
(d) identification (name) of the Manufacturers repre-
(j) number of film in each film holder/cassette
sentative who performed the final acceptance of the
(k) single- or double-wall exposure radiographs
(l) single- or double-wall viewing (e) date of Manufacturers evaluation
30
ARTICLE 2
MANDATORY APPENDIX
(a) For longitudinal welds the lead symbol B shall I-274.1 Geometric Unsharpness. Geometric un-
be attached to the back of each film cassette or at sharpness for an in-motion radiograph shall be deter-
approximately equal intervals not exceeding 36 in. (914 mined as specified in T-274.
mm) apart, whichever is smaller. I-274.2 In-Motion Unsharpness. In-motion un-
(b) For circumferential welds, the lead symbol B sharpness of the radiograph shall be determined in
shall be attached to the back of the film cassette in accordance with:
each quadrant or spaced no greater than 36 in. (914
mm), whichever is smaller. wd
UM p
D
where
I-275 Location Markers
wp beam width at the source side of the weld mea-
sured in the direction of motion Location markers shall be placed adjacent to the
ap slit width in diaphragm in direction of motion weld at the extremity of each film cassette and
31
I-275 2001 SECTION V I-285
also at approximately equal intervals not exceeding 15 Wire IQIs, when used, shall be placed on the weld
in. (381 mm). seam so that the length of the wires is perpendicular
to the length of the weld and spaced as indicated above
for hole IQIs.
I-277 Placement and Number of IQIs
(a) For longitudinal welds, hole IQIs shall be placed I-279 Repaired Area
adjacent to and on each side of the weld seam, or on When radiography of a repaired area is required, the
the weld seam at the beginning and end of the weld length of the film used shall be at least equal to the
seam, and thereafter at approximately equal intervals length of the original location marker interval.
not exceeding 36 in. (914 mm) or for each film cassette.
Wire IQIs, when used, shall be placed on the weld
seam so that the length of the wires is perpendicular I-280 EVALUATION
to the length of the weld and spaced as indicated above
for hole IQIs. I-285 Geometric and In-Motion
Unsharpness Limitations
(b) For circumferential welds, hole IQIs shall be
placed adjacent to and on each side of the weld seam Neither the geometric nor in-motion unsharpness of
or on the weld seam in each quadrant or spaced no the radiograph shall exceed the values specified in
greater than 36 in. (914 mm) apart, whichever is smaller. T-285.
32
ARTICLE 2
MANDATORY APPENDIX
33
II-263 2001 SECTION V II-292
34
ARTICLE 2
MANDATORY APPENDIX
35
III-260 2001 SECTION V III-292
36
ARTICLE 2
MANDATORY APPENDIX
APPENDIX IV INTERPRETATION, the unprocessed original full image and the digitally
EVALUATION, AND DISPOSITION processed image, shall be recorded and stored on video
OF RADIOGRAPHIC AND tape, magnetic tape, or optical disk.
RADIOSCOPIC EXAMINATION TEST
RESULTS PRODUCED BY THE
IV-221 Procedure Requirements
DIGITAL IMAGE ACQUISITION
AND DISPLAY PROCESS A written procedure is required and shall contain,
as a minimum, the following system performance param-
IV-210 SCOPE eters:
The digital image examination test results produced (a) image digitizing parameters modulation trans-
in accordance with Article 2, Mandatory Appendix II, fer function (MTF), line pair resolution, contrast sensi-
and Article 2, Mandatory Appendix III, may be interpre- tivity, dynamic range, and pixel size;
ted and evaluated for final disposition in accordance (b) image display parameters monitor size includ-
with the additional provisions to Article 2 as indicated ing display pixel size, luminosity, format, contrast, and
herein. magnification;
The digital information is obtained in series with (c) signal processing parameters including density
radiography and in parallel with radioscopy. This data shift, contrast stretch, log transform, and any other
collection process also provides for interpretation, evalu- techniques that do not mathematically alter the original
ation, and disposition of the examination test results. digital data, e.g., linear and area measurement, pixel
sizing, and value determination;
(d) storage identification, data compression, and
IV-220 GENERAL REQUIREMENTS media (including precautions to be taken to avoid data
loss). The non-erasable optical media should be used
The digital image shall be interpreted while displayed for archival applications. This is frequently called the
on a cathode ray tube (soft display). The interpretation WORM (Write Once Read Many) technology. When
may include density and contrast adjustment, quantifica- storage is accomplished on magnetic or erasable optical
tion, and pixel measurement, including digital or optical media, then procedures must be included that show
density values and linear or area measurement. trackable safeguards to prevent data tampering and
The interpretation of a digitized image is dependent guarantee data integrity.
upon the same subjective evaluation by a trained inter-
preter as the interpretation of a radiographic or radio-
scopic image. Some of the significant parameters consid- IV-222 Original Image Artifacts
ered during interpretation include: area of interest, image
Any artifacts that are identified shall be noted or
quality, IQI image, magnification, density, contrast,
annotated on the digital image.
discontinuity shape (rounded, linear, irregular), and
artifact identification.
The digital image interpretation of the radiographic
and radioscopic examination test results shall be per- IV-230 EQUIPMENT AND MATERIALS
formed in accordance with the modified provisions to
IV-231 Digital Image Examination Record
Article 2 as indicated herein.
After the interpretation has been completed, the inter- The digital image examination data shall be recorded
pretation data and the digital image, which shall include and stored on video tape, magnetic disk, or optical disk.
37
IV-234 2001 SECTION V IV-292
38
ARTICLE 2
MANDATORY APPENDIX
39
V-230 2001 SECTION V V-230
image quality indicator sensitivity the smallest discernible detail and /or
hole type a rectangular plaque, made of material contrast change (e.g., IQI hole or wire) in a radiographic
radiographically similar to that of the object being image
radiographed, with small diameter holes (1T, 2T, and shim a material, radiographically similar to the
4T) used to check the image quality of the radiograph object being radiographed, that is placed between a
wire type a set of small diameter wires, made hole-type IQI and the object in order to reduce the
of material radiographically similar to that of the object radiographic density through the image of the hole-
being radiographed, used to check the image quality type IQI
of the radiograph source side that surface of the area of interest
image storage system a system that can store being radiographed for evaluation nearest the source
digital image data for future use of radiation
IQI image quality indicator step wedge calibration film a processed film
line pair resolution the number of line pairs per with discrete density steps that have been verified by
unit distance that are detectable in an image comparison with a national standard step tablet
log transform a function that applies a logarithmic step wedge comparison film a processed film
mapping to all density /greyscale values in an image; with discrete density steps that have been verified by
this operation is often performed when the resulting use of a calibrated densitometer, which is used to
distribution is normal, or if the resulting relationship determine if production radiographs meet density limits
with another variable is linear WORM (write once read many) a term relating
modulation transfer function (MTF) a measure to a type of digital storage media where the data can
of spatial resolution as a function of contrast; a plot be stored only once but accessed (nondestructively)
of these variables (spatial resolution and contrast) yields many times
a curve representing the frequency response of the (c) The following Code terms are used in conjunction
system with Article 2, Appendix VI:
national standard step tablet an x-ray film with contrast sensitivity the size of the smallest detect-
discrete density steps produced and certified by a nation- able change in optical density
ally recognized standardizing body dynamic range the extent of measurable optical
nonerasable optical media (optical disk) a storage density obtained in a single scan
media that prevents the erasure or alteration of digital image the digital representation of a target on
data after it is stored the reference film used to evaluate both the digitization
optical density step wedge a radiographic image and display aspects of a film digitization system
of a mechanical step wedge with precise thickness reference film a single industrial radiographic film
increments and may be used to correlate optical film that encompasses the targets necessary for the evaluation
density to the thickness of material, also known as a and quantification of the performance characteristics of
step tablet a film digitization system
penetrameter no longer used in Article 2; see spatial linearity the accuracy to which a digitiza-
image quality indicator tion system reproduces the physical dimensions of
quantification the act of determining or expressing information on the original film [both in the horizontal
a quantity (i.e., giving a numerical value to a measure- (along a single scan line) and vertical (from one scan
ment of something) line to another) directions]
radiographic examination a nondestructive method spatial resolution the size of the smallest detectable
for detecting discontinuities in materials and components element of the digitized image
using penetrating radiation to produce an image on a target a physical pattern on a reference film used
recording medium to evaluate the performance of a film digitization system
40
ARTICLE 2
MANDATORY APPENDIX
APPENDIX VI DIGITAL IMAGE (d) spot size(s) of the film scanning system;
ACQUISITION, DISPLAY, (e) image display pixel size as defined by the vertical/
INTERPRETATION, AND STORAGE horizontal resolution limits of the monitor;
OF RADIOGRAPHS FOR NUCLEAR (f) luminance of the video display; and
APPLICATIONS (g) data storage medium.
41
VI-230 2001 SECTION V VI-254
42
VI-254 ARTICLE 2 MANDATORY APPENDIX VI-291
43
VI-291 2001 SECTION V VI-A-234
graphic reader sheet may be digitized to fulfill this VI-A-232 Spatial Resolution Targets
requirement);
The reference film shall contain spatial resolution
(f) condition of original radiographs (VI-281);
targets as follows:
(g) procedure demonstration (VI-261);
(h) spatial linearity (VI-244); VI-A-232.1 Converging Line Pair Targets. Con- 01
(i) system performance parameters (VI-241); and verging line pairs shall consist of 3 identical groups
(j) personnel performing the digital imaging process of no less than 6 converging line pairs (6 light lines
(VI-223). and 6 dark lines). The targets shall have a maximum
resolution of no less than 20 line pairs per millimeter
(lp/mm) and a minimum resolution of no greater than
VI-292 Archiving 1 lp/mm. The 3 line pair groups shall be oriented in
When the final report and digitized information are the vertical, horizontal, and the last group shall be 45
used to replace the analog radiograph as the permanent deg. from the previous two groups. The maximum
record as required by the referencing Code Section, resolution shall be oriented toward the corners of the
all information pertaining to the original radiography film. Reference marks shall be provided to indicate
shall be documented in the final report and processed spatial resolution at levels of no less than 1, 2, 3, 4,
as part of the digital record. A duplicate copy of the 5, 6, 7, 8, 9, 10, 15, and 20 lp/mm. The spatial
WORM storage media is required if the radiographs resolution targets shall be located in each corner of
are to be destroyed. the needed film sizes.
VI-A-232.2 Parallel Line Pair Targets. Parallel line 01
pairs shall consist of parallel line pairs in at least the
ARTICLE 2 vertical direction on the reference film. It shall have
MANDATORY APPENDIX VI a maximum resolution of at least 20 lp/mm and a
SUPPLEMENT A minimum resolution of no less than 0.5 lp/mm. It shall
have distinct resolutions of 0.5, 1, 2, 3, 4, 5, 6, 7, 8,
VI-A-210 SCOPE 9, 10, 15, and 20 lp/ mm and have the corresponding
The reference film described in this supplement pro- reference marks. It shall be located near the middle
vides a set of targets suitable for evaluating and quantify- of the reference film.
ing the performance characteristics of a radiographic
digitizing system. The reference film is suitable for
evaluating both the radiographic film digitization process VI-A-233 Constrast Sensitivity Targets 01
and the electronic image reconstruction process. Contrast sensitivity targets shall consist of approxi-
The reference film shall be used to conduct perform- mately 0.4 in. by 0.4 in. (10 mm by 10 mm) blocks
ance demonstrations and evaluations of the digitizing centered in 1.6 in. by 1.6 in. (40 mm by 40 mm)
system to verify the operating characteristics before blocks of a slightly lower density. Two series of these
radiographs are digitized. The reference film provides step blocks shall be used with an optical density of
for the evaluation of spatial resolution, contrast sensitiv- approximately 2.0 on a background of approximately
ity, dynamic range, and spatial linearity. 1.95, an optical density change of 0.05. The second
block series will have an optical density of approxi-
mately 3.5 on a background of approximately 3.4, an
01 VI-A-220 GENERAL optical density change of 0.10. The relative density
change is more important than the absolute density.
VI-A-221 Reference Film
These images shall be located near the edges and the
The reference film shall be specified in VI-A-230 center of the film so as to test the contrast sensitivity
and VI-A-240. throughout the scan path.
44
VI-A-234 ARTICLE 2 MANDATORY APPENDIX VI-A-245
on the density strip (at approximately 1.0, 2.0, 3.0, VI-A-242 Film Size
and 4.0 optical densities), there shall be optical density
The film size shall be sufficient to accommodate the
changes of 0.02 which shall also be used to test the
largest area of interest to be digitized.
contrast sensitivity. These stepped density targets shall
be located near the edges of the film and near the
center so as to test the dynamic range throughout the VI-A-243 Spatial Resolution
scan path.
The spatial resolution shall be a minimum of 20
lp/mm.
01 VI-A-235 Spatial Linearity Targets
Measurement scale targets shall be located in the VI-A-244 Density 01
horizontal and vertical dimensions. The measurement The relative densities stated in VI-A-233 and VI-A-
scale targets shall be in English and/or metric divisions. 234 shall be within 0.005 optical density.
(a) The tolerance for the optical density changes
stated in VI-A-233 and VI-A-234 shall be 0.005.
01 VI-A-240 MISCELLANEOUS (b) The measured densities shall be within 0.15 of
REQUIREMENTS the values stated in VI-A-233 and VI-A-234. The actual
densities shall be recorded and furnished with the
Manufacturing specifications shall be minimum re- reference film.
quirements necessary for producing the reference film. (c) Density requirements shall be in accordance with
The reference film shall have a unique identification ANSI IT-2.19.
which appears as an image when digitized. (d) The background density, where there are no
images located, shall have a 3.0 optical density 0.5.
VI-A-241 Material
VI-A-245 Linearity
The reference film shall be a fine grain, industrial
type film. The film used will be of high quality so The measurement scale targets shall be accurately
the required specifications in VI-A-230 are met. electronically produced to 0.05 in. (1.3 mm).
45
Fig. VI-A-1 2001 SECTION V
46
ARTICLE 2
MANDATORY APPENDIX
APPENDIX VII RADIOGRAPHIC (a) Casting Areas Prior to Finish Machining. The
EXAMINATION OF METALLIC IQI shall be based on a thickness that does not exceed
CASTINGS the finished thickness by more than 20% or 14 in. (6.4
mm), whichever is greater. In no case shall an IQI
VII-210 SCOPE size be based on a thickness greater than the thickness
Metallic castings, due to their inherent complex con- being radiographed.
figurations, present examination conditions that are (b) Casting Areas That Will Remain in the As-Cast
unique to this product form. Condition. The IQI shall be based on the thickness
Radiographic examination may be performed on cast- being radiographed.
ings when the modified provisions to Article 2, as
indicated herein, are satisfied.
VII-280 EVALUATION
VII-220 GENERAL REQUIREMENTS VII-282 Radiographic Density
VII-224 System of Identification VII-282.1 Density Limitations. The transmitted film
A system shall be used to produce permanent identi- density through the radiographic image of the body of
fication on the radiograph traceable to the contract, the appropriate hole IQI or adjacent to the designated
component, or part numbers, as appropriate. In addition, wire of a wire IQI and the area of interest shall be
each film of a casting being radiographed shall be 1.5 minimum for single film viewing. For composite
plainly and permanently identified with the name or viewing of multiple film exposures, each film of the
symbol of the Material Manufacturer, Certificate Holder, composite set shall have a minimum density of 1.0.
or Subcontractor, job or heat number, date, and, if The maximum density shall be 4.0 for either single or
applicable, repairs (R1, R2, etc.). This identification composite viewing. A tolerance of 0.05 in density is
system does not necessarily require that the information allowed for variations between densitometer readings.
appear as radiographic images. In any case, this informa-
tion shall not obscure the area of interest.
VII-290 DOCUMENTATION
VII-270 EXAMINATION
VII-293 Layout Details1
VII-271 Radiographic Technique
To assure that all castings are radiographed consist-
VII-271.2 Double-Wall Viewing Technique. A ently in the same manner, layout details shall be
double-wall viewing technique may be used for cylindri- provided. As a minimum, the layout details shall include:
cal castings 312 in. (89 mm) or less in O.D. or when (a) sketches of the casting, in as many views as
the shape of a casting precludes single-wall viewing. necessary, to show the approximate position of each
location marker; and
(b) source angles if not perpendicular to the film.
VII-276 IQI Selection
VII-276.3 Additional IQI Selection Requirements. 1
Sample layout and technique details are illustrated in SE-1030,
The thickness on which the IQI is based is the single- Appendix (Nonmandatory Information) X1, Fig. X1.1, Radiographic
wall thickness. Standard Shooting Sketch (RSS).
47
ARTICLE 2
NONMANDATORY APPENDIX
APPENDIX A TECHNIQUE
SKETCHES FOR
PIPE OR TUBE WELDS
A-210 SCOPE
The sketches in the Appendix illustrate techniques
used in the radiographic examination of pipe or tube
welds.
Other techniques may be used.
48
ARTICLE 2 NONMANDATORY APPENDIX
Film Source
Single-
Wall Side
T-271.1 T-277.1(a)
T-276 Either
Single- and Side
Any
Wall Table T-275.3
Source T-275.1(c)
T-276
Film
Side
T-277.1(b)
Exposure Arrangement A
Single- Source
Wall Side
Source
T-271.1 T-277.1(a)
T-276 Film
and Side
Any Single-
Table T-275.1
Wall
T-276 (b)(1)
Film
Film
Side
Exposure Arrangement B T-277.1(b)
Source
Source
Side
Single- T-277.1(a)
Wall
T-271.1 T-276 Source
Film and Side
Any Single-
Wall Table T-275.1
T-276 (a)(3)
Film
Side
Exposure Arrangement C T-277.1(b)
49
2001 SECTION V
Optional
source
location Source
Double- Side T-
Wall: T- 277.1(a)
271.2(a) at
T-276 Film
Least 3
Single- and Side
Any Exposures
Wall Table T- T-275.1
120 deg. to
276 (b)(1)
Each Other
for Complete
Coverage Film Side
Film T-277.1(b)
Exposure arrangement D
Optional
source
location Source
Double- Side T-
Wall: T- 277.1(a)
271.2(a) at
T-276 Film
least 3
Single- and Side T-
Any Exposures
Wall Table T- 275.1
120 deg. to
276 (b)(1)
Each Other
for Complete
Coverage Film Side
T-277.1(b)
Film
Exposure arrangement E
Source
Double-Wall Double-
T- Wall
271.2(b)(1) (Ellipse):
T-276
at Least 2 Read Source Either
312 in. (89 mm) and
Exposures at Offset Side T- Side T-
or Less Table T-
90 deg. to Source 277.1(a) 275.2
276
Each Other Side and
for Complete Film Side
Coverage Images
Film
Exposure arrangement F
(continued)
50
ARTICLE 2 NONMANDATORY APPENDIX
Source
Double-
Wall: T- Double-
271.2(b)(2) Wall: Read
at Least 3 Super- T-276
Source Either
312 in. (89 mm) Exposures at imposed and
Side T- Side T-
or Less 60 deg. or Source Table T-
277.1(a) 275.2
120 deg. to Side and 276
Each Other Film Side
for Complete Images
Coverage
Film
Exposure arrangement G
51
ARTICLE 2
NONMANDATORY APPENDIX
APPENDIX C HOLE-TYPE
IQI PLACEMENT SKETCHES
FOR WELDS
C-210 SCOPE
The figures in this Appendix demonstrate typical IQI
(hole type) placement for welds. These sketches are
tutorial to demonstrate suggested locations of IQIs and
are not intended to cover all configurations or applica-
tions of production radiography. Other IQI locations
may be used provided they comply with the require-
ments of Article 2. Wire IQIs shall be placed in
accordance with the requirements of Article 2.
52
ARTICLE 2 NONMANDATORY APPENDIX Fig C-210-1
53
Fig C-210-2 2001 SECTION V
54
ARTICLE 2 NONMANDATORY APPENDIX Fig C-210-3
55
Fig C-210-4 2001 SECTION V
56
ARTICLE 2
NONMANDATORY APPENDIX
APPENDIX D NUMBER OF
IQIs (SPECIAL CASES)
D-210 SCOPE
The figures in this Appendix illustrate examples of
the number and placement of IQIs that may be used
in the special cases described in T-277.2(b). These
figures are not intended to cover all configurations or
applications of production radiography.
57
Fig. D-210-1 2001 SECTION V Fig. D-210-4
58
Fig. D-210-5 ARTICLE 2 NONMANDATORY APPENDIX Fig. D-210-8
FIG. D-210-7 PLAN VIEW A-A FIG. D-210-8 ARRAY OF OBJECTS IN A CIRCLE
[T-277.2(b)(7)]
59