MRM Aoac 2007 01 PDF
MRM Aoac 2007 01 PDF
MRM Aoac 2007 01 PDF
5 g anhydrous
Pesticide Residues in Foods by Acetonitrile MgSO4/NaOAc (4/1, w/w) per g sample are added to a centrifuge
Extraction and Partitioning with Magnesium Sulfate
tube or bottle, which is shaken and centrifuged. A portion of the
Gas Chromatography/Mass Spectrometry and
Liquid Chromatography/Tandem Mass Spectrometry MeCN extract (upper layer) is added to anhydrous MgSO4/PSA
First Action 2007 sorbent (3/1, w/w; 200 mg per 1 mL extract), mixed, and
[Applicable for the following pesticides in grapes, lettuces, and centrifuged. This final extract is transferred to autosampler vials for
oranges: atrazine, azoxystrobin, bifenthrin, carbaryl, chlorothalonil, analysis by gas chromatography/mass spectrometry (GC/MS) and
chlorpyrifos, chlorpyrifos-methyl, l-cyhalothrin (incurred in liquid chromatography/tandem mass spectrometry (LC/MS/MS) to
lettuces), cyprodinil, o,p-DDD, dichlorvos, endosulfan sulfate,
identify and determine a wide range of pesticide residues. To
ethion (inc urred in oranges), imazalil , imidaclopri d,
kresoxim-methyl (incurred in grapes), linuron, methamidophos, achieve <10 ng/g detection limits in modern GC/MS, large volume
methomyl, permethrins (incurred in lettuces) procymidone, injection (LVI) of 8 mL is typically needed, or the final extract can be
pymetrozine, tebuconazole, thiabendazole (incurred in oranges), concentrated and solvent exchanged to toluene (4 g/mL), in which
tolylfluanid (degraded in lettuces), and trifluralin. These were
case 2 mL splitless injection is used.
representative pesticide analytes chosen in representative matrixes,
and the method is expected to be applicable to many other similar Both GC/MS and LC/MS/MS techniques are prone to matrix
pesticides and matrixes. Limits of quantitation were demonstrated
effects in pesticide residue analysis, albeit for different reasons
to be <10 ng/g.]
[Erney, D.R., Gillespie, A.M., Gilvydis, D.M., & Poole, C.F. (1993)
See Tables 2007.01AE for the results of the interlaboratory J. Chromatogr. 638, 5763; Hajslova, J., & Zrostlikova, J. (2003) J.
study supporting acceptance of the method. Chromatogr. A 1000, 181197; Alder, L., Luderitz, S., Lindtner, K.,
A. Principle & Stan, H.J. (2004) J. Chromatogr. A 1058, 6779]. To account for
these effects, matrix-matched calibration was conducted
The QuEChERS (quick, easy, cheap, effective, rugged, and safe) (calibration standards in solvent solution may also be used if matrix
method uses a single-step buffered acetonitrile (MeCN) extraction effects are shown not to occur). Due to the situation that some
and salting out liquidliquid partitioning from the water in the laboratories had LVI capability and others did not, the necessary
sample with MgSO 4 . Dispersive-solid-phase extraction amounts of matrix blank(s) and final extract volume was different
(dispersive-SPE) cleanup is done to remove organic acids, excess for some laboratories than others. Depending on the water content of
the matrix, a 15 g sample typically yields 1114 mL of initial MeCN
water, and other components with a combination of primary
extract after centrifugation. In dispersive-SPE, roughly half of the
secondary amine (PSA) sorbent and MgSO4; then the extracts are
extract is lost to the powders, thus about 67 mL of final extract can
analyzed by mass spectrometry (MS) techniques after a be expected for a 15 g sample. Two options were provided in the
chromatographic analytical separation. Figure 2007.01 outlines the protocol to account for the different situations among the
protocol in a box format. In brief, a well-chopped food sample along laboratories.
Table 2007.01A. Interlaboratory study results for incurred pesticides (and chlorpyrifos-methyl)
a b c d Outlier
sr RSDr , % SR , ng/g RSDR , % e
Analyte Matrix Avg. concn Rec., % HorRat No. of labs labs
Chlorpyrifos-methyl Grapes 165 14 8.5 35 83 21 1.00 11 6-C, 4-C
Lettuces 178 20 11 30 89 17 0.81 10 11-SG
Oranges 174 25 14 36 87 20 0.98 12
f f
Kresoxim-methyl Grapes 9.2 1.9 21 3.2 NA 35 1.09 12
g
Cyprodinil Grapes 112 NA NA 18 NA 16 0.73 13
l-Cyhalothrin Lettuces 58 6.1 11 11 NA 20 0.80 9 11-C
f
Permethrins Lettuces 112 9.8 8.7 41 NA 36 1.63 9 6-C, 1-C
Imidacloprid Lettuces 12 NA NA 1.6 NA 14 0.44 11
Ethion Oranges 198 23 12 36 NA 18 0.89 11 11-C
Thiabendazole Oranges 53 3.8 7.2 7.6 NA 14 0.58 12
f
Imazalil Oranges 13 NA NA 4.7 NA 35 1.15 8 7-SG
a
sr = Standard deviation for repeatability (within laboratory).
b
RSDr = Relative standard deviation for repeatability.
c
sR = Standard deviation for reproducibility (among laboratories).
d
RSDR = Relative standard deviation for reproducibility.
e
C = Cochran outlier; SG = single Grubbs outlier.
f
RSDr >15%; 120% < Rec. < 70%; RSDR >25%; HorRat >1.2; and fewer than 8 laboratories in an assessment.
g
NA = Not applicable.
In Option A, if the laboratory had LVI capability, then 1 or 2 mL 75C for MeCN extracts or 100C for toluene ramped to 150C at
extracts were taken for dispersive-SPE (the volume depended on the 25C/min, then to 280C at 10C/min, and hold for 10 min. All
analyst preference and the type of centrifuge and tubes available in collaborators had much experience in pesticide residue analysis and
the laboratory). The final extract volume was 0.5 mL if 1 mL was were free to use their own analytical conditions provided that peak
taken for dispersive-SPE, and 1 mL if 2 mL underwent the cleanup shapes were Gaussian, peak widths at half heights were <5 s, and
step. In either case, two 15 g blank samples were used for the matrix signal-to-noise ratio (S/N) of the quantitative ion for the pesticides
blank (0-standard) and 6 matrix-matched calibration standards (5, at 10 ng/g equivalent concentrations in the sample were >10. For
10, 50, 100, 250, and 1000 ng/g equivalent concentrations). For qualitative purposes (which were not the focus of this study), at least
dispersive-SPE of the matrix blanks, either 7 separate tubes using 3 ions yielding relative abundances that reasonably match a
the same 12 mL extract volumes as the test samples could have contemporaneously analyzed reference standard are typically
been used, or 12 dispersive-SPE tube(s) with 7-fold greater extract needed to make an analyte identification.
volume(s).
In Option B, if LVI is not available for GC/MS, then 30 mL of (b) Liquid chromatograph/tandem mass spectrometer. A triple
matrix blank extract was needed after dispersive-SPE cleanup to quadrupole, ion trap, or other LC/MS/MS instrument may be used
provided it is capable of electrospray ionization (ESI) in the positive
prepare the matrix-matched calibration standards (or $60 mL initial
mode with computerized instrument control/data collection and has
extract). In this case, 6 matrix blanks of 15 g each were extracted
along with the test samples to provide enough blank extract volume, an AS. An injection volume (5100 mL) will be determined for each
which were combined, and seven 8 mL aliquots were distributed to instrument to achieve S/N > 10 for the quantitation ion for a 10 ng/g
7 dispersive-SPE tubes containing 0.4 g PSA + 1.2 g anhydrous equivalent sample concentration. As in GC/MS, the collaborators
MgSO4. had much experience in the analysis of pesticides and were free to
use their own conditions. Suggested LC conditions, however,
B. Apparatus and Conditions include a 15 cm long, 3.0 mm id, 3 mm particle size C18 column, flow
Note: Tables 4 and 5 of the collaborative study [J. AOAC Int. 90, rate of 0.3 mL/min, and gradient elution with an initial condition of
485(2007)] list the analytical instrumentation and sources of sample 25% MeOH in 5 mM formic acid solution taken linearly in 15 min to
preparation materials used by each laboratory in the study. Further 90% MeOH in 5 mM formic acid solution and held for 15 min. A
information appears in the full report. Since the time of the short C18 guard column must be used to protect the analytical
collaborative study, at least 3 vendors, United Chemical column, and a bypass valve must be used before the MS instrument
Technologies (Bristol, PA, USA), Restek (Bellefonte, PA, USA) and to avoid introduction of the early and late eluting nonanalyte
Supelco (Bellefonte, PA, USA) have introduced commercial components into the detector. The MS/MS conditions were
dispersive-SPE products for QuEChERS and other applications. See optimized in each laboratory using direct infusion into the ESI
Table 4 [J. AOAC Int. 90, 485(2007)] for sources of analytical source to provide highest S/N for the quantitation ion of each
instruments. LC-type analyte from a single MS/MS transition. A second
(a) Gas chromatograph/mass spectrometer.An ion trap, transition with reasonably matching relative abundance ratios vs a
quadrupole, time-of-flight (TOF), or other GC/MS instrument may contemporaneously analyzed reference standard is typically needed
be used with electron impact (EI) ionization, an autosampler (AS), for qualitative purposes.
and computerized instrument control/data collection. Either LVI of
(c) Centrifuge(s).Capable of holding the 50 mL centrifuge
8 mL for a 1 g/mL MeCN extract (e.g., 75C ramped to 275C at
tubes or bottles used for extraction and 1015 mL graduated
200C/min) or 2 mL splitless injection of 4 g/mL extracts in toluene centrifuge tubes or 2 mL mini-tubes used in dispersive-SPE.
at 250C may be used. A 35 m, 0.25 mm id, Determine the rpm settings that yield a given relative centrifugal
phenylmethyl-deactivated guard column must be used as a retention force (RCF), and ensure that maximum ratings of the centrifuge,
gap in either case. The analytical column is a 30 m, 0.25 mm id, tube/bottles, and rotors for the instrument are not exceeded.
0.25 mm film thickness (5%phenyl)-methylpolysiloxane (low
bleed) analytical column (DB-5ms or equivalent). Set He head (d) Balance(s).Capable of accurately measuring weights from
pressure on the column to be 10 psi or constant flow to be 0.05 to 100 g within 0.01 g.
1.0 mL/min with systems capable of electronic pressure/flow
(e) Freezer.Capable of continuous operation <20C.
control. After an appropriate time for solvent delay, use an
appropriate oven temperature program, for example, starting at (f) Furnace/oven.Capable of 500C operation.