OPERATING MANUAL
Model UT513: OPERATING MANUAL
Table of Contents
Title Page
Introduction 5
Unpacking the Meter 5
Safety Information 6
International Electrical Symbols 8
Battery Saver (Sleep Mode) 8
Battery Indication 8
The Meter Structure 9
Display 12
Key Functions 14
Measurement Operation 17
A. Measuring Voltages 17
B. Measuring Insulation Resistance 19
a) Continuous Measurement 20
b) Timed Measurement 20
c) Polarization Index (PI) Measurement 21
d) Compare Function 22
The Use of Power Adaptor 23
1
Model UT513: OPERATING MANUAL
Title Page
USB Interface 24
Maintenance 25
A. General Service 25
B. Replacing the Battery 26
Specifications 27
Safety and Compliances 27
Physical Specifications 27
General Specifications 28
Feature Summary 28
Detailed Accuracy Specifications 29
A. Voltage Measurement 29
B. Insulation Resistance Measurement 30
2
Model UT513: OPERATING MANUAL
List of Tables
1. Unpacking Inspection 5
2. International Electrical Symbols 8
3. Battery Indication 8
4. Meter Front Description 10
5. Meter Side Description 11
6. Display Description 13
7. Key Description 14
3
Model UT513: OPERATING MANUAL
List of Figures
4
Model UT513: OPERATING MANUAL
5
Model UT513: OPERATING MANUAL
6
Model UT513: OPERATING MANUAL
When carrying out insulation measurement, When performing resistance tests, remove all
do not contact the circuit under test. power from the circuit to be measured and
discharge all the power.
Warning When servicing the Meter, use only the same
Do not use the Meter if it is damaged or metal model number or identical electrical
part is exposed. Look for cracks or missing specifications of test leads and power adaptor.
plastic. Do not use the Meter if the battery indicator
Be careful when working above 33V rms, 46.7V ( ) shows a battery empty condition. Take
ac rms or 70V DC. Such voltages pose a shock the battery out from the Meter if it is not used
hazard. for a long time.
Discharge all loading of circuit under test after Do not use or store the Meter in an
measuring high voltage. environment of high temperature, humidity,
Do not change battery when the Meter is in explosive, inflammable and strong magnetic
wet environment. field. The performance of the Meter may
Place test leads in proper input terminals. deteriorate after dampened.
Make sure all the test leads are firmly Soft cloth and mild detergent should be used
connected to the Meters input terminals. to clean the surface of the Meter when
Make sure the Meter is turned off when servicing. No abrasive and solvent should be
opening the battery compartment. used to prevent the surface of the Meter from
corrosion, damage and accident.
Caution Dry the Meter before storing if it is wet.
7
Model UT513: OPERATING MANUAL
9
Model UT513: OPERATING MANUAL
1 LCD 17 Up Button
2 Scroll Button 18 LINE: High Voltage output input terminal
3 Emergency Stop (two plugs red test lead to one alligator clip)
4 Data Clear the Display Backlight Button 19 High voltage line shielding input terminal
5 Down Button (two plugs red test lead to one alligator clip)
6 On/Off Button 20 GUARD: Grounding protection input terminal
7 Compare Button (one plug black test lead to one alligator clip)
8 Insulation Resistance Button 21 EARTH: High resistance measurement input
9 DC Voltages measurement Button terminal (one plug green test lead to one
10 Timer Button. alligator clip)
11 AC Voltages measurement Button 22 Testing leads:
12 Test Button Two plugs red test lead to one alligator clip.
13 USB Button One plug black test lead to one alligator clip.
14 Data Store Button. One plug green test lead to one alligator
15 Data Recall Button clip.
16 Scroll Button
10
Model UT513: OPERATING MANUAL
Below Figure 2 and Table 5 shows the Meter side Table 5. Meter Side Description
structure and description
1 Safety Shutter
2 Power adaptor Input Terminal
3 USB Port
11
Model UT513: OPERATING MANUAL
Display
Table 6 and Figure 3 describe the display.
Figure 3. Display
12
Model UT513: OPERATING MANUAL
13
Model UT513: OPERATING MANUAL
Key Functions
Table 7. Key Description
14
Model UT513: OPERATING MANUAL
When set the timer duration for the increment the time. The maximum
measurement of insulation resistance length of time is 30 minutes and 30
or polarization index, press to seconds, the Meter will automatically
decrement the time. The maximum carry out measurement.
length of time is 15 minutes and 30 When use the compare feature
seconds, the Meter will automatically measuring insulation resistance,
carry out measurement. press to increment a resistance
When compare feature measuring comparing value.
insulation resistance, press to After polarization index
decrement a resistance comparing measurement, press to display
value. polarization index, TIME 2 insulation
After polarization index resistance value and TIME 1
measurement, press to display insulation resistance value in
polarization index, TIME 2 insulation sequence.
resistance value and TIME 1 USB Press once to start the data
insulation resistance value in transferring to the computer via USB,
sequence. USB symbol shows on the display.
When set the timer duration for the Press again to stop the data
measurement of insulation resistance transferring to the computer via USB,
or polarization index, press to USB symbol disappears.
15
Model UT513: OPERATING MANUAL
16
Model UT513: OPERATING MANUAL
Green
17
Model UT513: OPERATING MANUAL
Operating Catuion
To avoid harms to you or damages to the Meter,
please do not attempt to measure voltages
higher than 600V or 600V rms, although
readings may be obtained.
Special care should be taken when measuring
high voltage.
To measure voltages, set up the Meter as Figure 4 and
do the following:
1. Press DCV or ACV button to select DC voltage or
AC voltage measurement
2. Insert the red and green test lead into the tested
circuit.
3. When measuring DC voltage, if the red test lead is
negative voltage, - symbol will show on the display.
Note
When voltage measurement has been completed,
disconnect the connection between the testing
leads and the circuit under test and remove testing
leads away from the input terminals of the Meter.
18
Model UT513: OPERATING MANUAL
19
Model UT513: OPERATING MANUAL
20
Model UT513: OPERATING MANUAL
Then press and hold TEST button for 2 second to or decrement by every 30 seconds.
carry out timed measurement. TIME 1 and are Then press and hold TEST button for 2 seconds to
displayed and blinked on the LCD on every 0.5 carry out timed measurement.
seconds. TIME 1 and are displayed and blinked on the
When the set time is reached, the insulation LCD on every 0.5 seconds before TIME 1 set time
resistance measurement voltage will be closed is reached.
and the measurement will be automatically stopped. TIME 2 and are displayed and blinked on the
The LCD displays the insulation resistance reading. LCD on every 0.5 seconds before TIME 2 set time
is reached.
When the two set time are reached, the insulation
resistance measurement voltage will be closed and
Press TIME button to select timed measurement the measurement will be automatically stopped.
mode, the LCD displays TIME 1 and symbols. The LCD displays the polarization index reading.
Press and buttons to set the time (00:10~15:00). Press , to set through the polarization index,
Within 1 minute, the time increment or decrement TIME 2 insulation resistance reading and TIME 2
by every 5 seconds. Afterward, the time increment insulation resistance reading.
or decrement by every 30 seconds.
Press TIME button again. TIME 2, PI and symbols Information:
appear on the LCD. PI = 3 minutes ~10 minutes reading / 30 seconds ~1
Press and buttons to set the time (00:15~15:30). minute reading
Within 1 minute, the time increment or decrement
PI 4 or more 4~2 2.0~1.0 1.0 or less
by every 10 seconds. Afterward, the time increment
Standard The best Good Warning Bad
21
Model UT513: OPERATING MANUAL
22
Model UT513: OPERATING MANUAL
The Use of Power Adaptor 1. Open the side safey shutter, then you will see there
is a power adaptor input terminal.
The use of power adaptor, see figure 6 2. Make sure the Meter is power off and insert the
UT513 power adaptor to the input terminal.
3. It is highly recommeded to take out all the batteries
when you are using the power adaptor.
4. Make sure the Meter is power off when you
disconnect the UT513 power adaptor from the Meter.
5. It is highly recommeded to use Uni-Trend supplied
UT513 power adaptor to avoid dangerous.
23
Model UT513: OPERATING MANUAL
24
Model UT513: OPERATING MANUAL
Warning
Do not attempt to repair or service your Meter unless
you are qualified to do so and have the relevant
calibration, performance test, and service
information.
A. General Service
Periodically wipe the case with a damp cloth and
mild detergent. Do not use abrasives or solvents.
To clean the terminals with cotton bar with detergent,
as dirt or moisture in the terminals can affect
readings.
Turn the Meter to OFF when it is not in use.
Take out the battery when it is not using for a long
time.
Do not use or store the Meter in a place of humidity,
high temperature, explosive, inflammable and strong
magnetic field.
25
Model UT513: OPERATING MANUAL
26
Model UT513: OPERATING MANUAL
Specifications
Safety and Compliances
Certification
Compliances IEC 61010 CAT.III 600V overvoltage and double insulation standard
Physical Specifications
Display (LCD) Digital: 9999 counts
Analog bar graph.
Operating Temperature -10 ~40 (14 ~104 )
Storage Temperature -20 ~60 (-4 ~152 )
Relative Humidity 85% @ -10 ~40 below;
90% @ -20 ~60 :
Battery Type 8pcs of 1.5V (LR14) batteries or power adaptor (input voltage 230V, 50/60Hz,
75mA, input DC14V, 1.0A).
Power adaptor is optionally at extra cost.
Dimensions (HxW xL) 202 x 155 x 94 mm
Weight Approx. 2kg (including battery)
27
Model UT513: OPERATING MANUAL
General Specifications
Range Auto
Overloading Display OL on insulation resistance range
Battery Indicator Display
Icon Display Equips with function and battery indicator icons.
Current Consumption Maximum: around 1.0A
Average: around 20mA
Feature Summary
Display Backlight Bright backlight for clear readings in poorly lighted areas.
Computer connection Via USB interface.
Data Logging and Recall 18
Autorange The Meter automatically selects best range
Warning and red light will on.
Voltage Auto release voltage
COMP Measurement Use the Compare function to set a pass/fail compare level for the insulation
measurements.
PI Measurement Polarization Index is the ratio of insulation resistance. You can pre-set two
point of times and automatically carry out the measurement.
TIME To carry out measurement by setting a specified time within 15 minutes.
28
Model UT513: OPERATING MANUAL
A. Voltage Measurement
DC Voltage AC Voltage
Measurement Range 30 ~ 600V 30V~600V (50/60Hz)
Resolution 1V
Accuracy (2%+5) Among them 30~100V(50/60Hz) (2%+8)
29
Model UT513: OPERATING MANUAL
Operating Caution
At any output voltage, when the tested resistance is les than 10M , the testing time cannot exceed 10
seconds continuously.
30
Model UT513: OPERATING MANUAL
*END*
31
Model UT513: OPERATING MANUAL
Manufacturer:
Uni-Trend Technology (Dongguan) Limited
Dong Fang Da Dao
Bei Shan Dong Fang Industrial Development District
Hu Men Town, Dongguan City
Guang Dong Province
China
Postal Code: 523 925
Headquarters:
Uni-Trend Group Limited
Rm901, 9/F, Nanyang Plaza
57 Hung To Road
Kwun Tong
Kowloon, Hong Kong
Tel: (852) 2950 9168
Fax: (852) 2950 9303
Email: [email protected]
http://www.uni-trend.com
32