1725 UV-Vis Glossary
1725 UV-Vis Glossary
1725 UV-Vis Glossary
A
Absorbance A measure of how much of the
incident radiation that is directed at a sample is
absorbed by the sample. Absorbance is defined
by the formula A = log10 (1/T), where T is the
fractional transmittance.
Absorbance Accuracy The agreement between
the measured value of the absorbance of a
standard and the true value of the absorbance of
the same standard.
Absorbance Reproducibility The agreement
(concordance) of a series of Absorbance values
obtained by repetition of the same measurement.
Absorptivity Called the Molar Absorptivity or
the Molar Extinction Coefficient or Molar
Absorption Coefficient. It is the absorbance at a
specified wavelength of a 1 mole per liter
solution of a compound measured over a 10 mm
pathlength. It has dimensions of M-1cm-1.
W Z
B
Bbackground see background spectrum.
Background correction The subtraction of the
background data from the sample data.
C
Calibration The process in which the software
analyzes a set of standards in order to calculate a
method model for predicting component
concentrations or classes from unknown samples.
Calibration curve See standard curve.
Cell See cuvette.
E
Electronic signature A computer data
compilation of any symbol or series of symbols
executed, adopted, or authorized by an
individual to be the legally binding equivalent of
the individuals handwritten signature.
Energy The instrument operates by measuring
the energy of light at the current wavelength
received by the detectors in the sample and
reference beams. The energy readings are stored
as raw %T data and used to calculate the
reported values in the selected Data Mode.
Entrance slit A thin opening in an opaque
screen through which light enters a
monochromator.
Enzyme A biological catalyst (a protein) that
facilitates a chemical reaction but which is not
consumed during the reaction.
F
Factor See concentration factor.
Fiber optic accessory A sampling accessory that
uses fiber optic technology to transport a beam of
light from the instrument through the accessory,
where the beam interacts with a sample, and then
back to the instrument. The instrument can then
use the changed beam to create a sample
spectrum.
Fiber optic cable A bundle of thin glass or
plastic transparent fibers that are enclosed by a
less refractive material. The fibers transmit light
by internal reflection.
G
Gain The ratio of the signal output of a system
to the signal input of the system. For example, a
gain of 10 would mean that a property of the
signal (usually voltage or power) had increased by
a factor of 10.
Galactic A file format for saving spectra in a
binary format that is compatible with other
applications and other types of data in the
industry, especially spectral and chromatographic
data. Galactic files include some file header
information and are saved with an extension of
.spc.
Goodness of Fit Provides a measure of the
correspondence between the Synthetic Spectrum
and the measured Sample Spectrum. When the
correspondence is exact, the value of the
Goodness of Fit will be zero.
In the MCA application, the Goodness of Fit is
the sum of the squares of the data points on the
Difference Spectrum divided by the number of
data points measured. The software does not
need to display the Difference Spectrum in order
to calculate the Goodness of Fit.
K
Kalmer filter The algorithm used in the
Multicomponent Analysis (MCA) application. It
enables spectra to be quantified for up to 20
components.
Kinetics test A test that is designed to measure
the absorbance of a sample over a period of time.
This mode may be used to determine the
absorbance of a reactant or the product of the
reaction at intervals during the reaction.
Kinetic data rate The rate at which data is
sampled during a kinetics test.
Kinetic experiment The processes of finding
out which model of enzyme behavior best fits a
set of rate data, and of determining the values of
the relevant kinetic parameters.
M
MCA see multicomponent analysis.
Menu bar The horizontal list of menu names,
typically near the top of a software application
window.
Method A collection of parameters defining the
state of the software and the instrument.
Molecular Spectroscopy Methods involving
analysis of the spectra formed by electronic
transitions in molecules.
Multicomponent analysis (MCA) The
simultaneous determination of two or more
substances in a sample. The sample is measured
at several wavelengths (one peak wavelength for
each substance) and the concentration of each
substance is calculated from the absorbance
values. This procedure is possible due to the
additive nature of absorbance (Atotal = A1 + A2)
N
Nanometer An X-axis unit used for wavelength.
One nanometer equals 1 x 10-9 meter.
Natural bandwidth of a peak The width (i.e.,
wavelength range) of the peak at half the peak
height, measured at infinite resolution.
Near infrared (NIR or near-IR) The region of
infrared radiation extending from approximately
12,000 cm-1 to 4,000 cm-1.
Noise The general term used to describe small,
irregular, random variations in the
spectrophotometer reading.
Noise, peak-to-peak The difference between
the maximum and minimum values of a noise
signal measured over a specified time period.
Non-linear regression Fits experimental data to
any equation that defines Y as a function of X
and one or more parameters. The non-linear
regression algorithms find the values of the
parameters that minimize the sum of the squares
of the vertical distances between the data points
and the curve. This process may also be described
as performing a Least Squares Fit to a curve.
O
Optical chopper A device for passing and then
interrupting a beam of light for a known brief
interval. Examples include rotating shutters or
rotating sector mirrors.
P
Parameter A property whose value determines
the characteristics or behavior of a software
application.
Partial Least Squares (PLS) A quantitative
method in TQ Analyst that uses a partial least
squares statistical analysis to find relationships
between the absorbance spectra and the
component concentrations of the corresponding
samples.
Pathlength The distance a beam of incident
energy travels within a sample. A longer
pathlength increases the absorption of infrared
energy by the sample. If the pathlength is too
great, totally absorbing bands will result. Since
the absorption depends on the pathlength (as
well as on concentration), if the pathlengths of
the samples and the standards used to quantify
them are not the same, the quantitative method
must account for the differences in pathlength. If
the pathlength values are known, they can be
entered in the quantitative method or specified
at run time. A quantitative method can also be
configured to predict or calculate the pathlength
values if the spectra contain a peak or region that
varies with pathlength.
PDF See portable document format.
Peak A region of a spectrum where the sample
absorbs radiation.
10
R
Ratioing The process of removing the effects of
the instrument and any water or carbon dioxide
absorptions (if these gases are not completely
purged from the instrument) from a sample
spectrum by dividing the spectrum by a
background spectrum, or a reference background
spectrum, at each data point.
Reactant A compound present at the start of,
and taking part in, a chemical reaction.
Read-only A specification when saving files.
When files are saved as read-only in Windows,
they may be opened and changed, but the altered
file cannot be resaved with the previous file
name. The file must be saved under a different
file name.
11
12
13
T
Temperature controller A component or
accessory used to monitor or control the
temperature of a system component. Typically
used to monitor or control the sampling
temperature of gases and liquids.
Temperature sensor A device that allows
software to read the temperature of an
instrument component, such as a gas cell or
block heater.
Template A template contains the default
information for a certain type of workflow event
or specification.
Thermoelectric cell holder An accessory into
which the sample cell is inserted that controls the
temperature to a known value.
Thermoelectric flow cell An accessory that
aspirates a sample into a cell that is temperatureregulated by an electronic means.
Three-point Net Absorbance Test A
measurement used to correct for the effects of a
broad sloping baseline on a peak height
determination by calculating a peak height
relative to a calculated baseline determined from
two "baseline" wavelengths.
Toolbar A long narrow strip at the top of a
software application window that contains action
buttons. Toolbars provide a convenient way to
initiate frequently used commands and other
functions in a single step.
Translucent The description of a material that
scatters part and transmits part of the incident
light; neither opaque or transparent.
14
U
Ultra-micro cuvette Cuvettes that
accommodate sample volumes between 5 and
100 microliters.
Ultraviolet (UV) The portion of the
electromagnetic spectrum from 10 to 380 nm.
Ultraviolet lamp A type of lamp (usually
deuterium) that emits a high quantity of
ultraviolet radiation. It is used for
spectrophotometric measurements in the UV
region.
Unsaturated molecule One in which there are
one or more pi (p) bonds. Unsaturated organic
compounds, such as ethene (ethylene), are able
to take part readily in addition reactions with
electrophilic reagents, such as bromine. This is
because the carbon atoms are forming "double
bonds", and are capable of increasing the number
of atoms to which they are bonded from three to
four. The absorption spectra of unsaturated
molecules may lie in the UV or visible,
depending on the size and structure of the
unsaturated system.
USB port A universal serial bus connector on a
computer. A USB port uses standard A
(toward the computer) and B (away from the
computer) connectors and a standard format that
works with any USB-compatible device such as a
printer, disk drive or mouse. The computer
operating system will automatically detect a USB
device after it is plugged in the first time; these
devices can be connected and disconnected at
any time.
15
W
Wavelength The distance between
corresponding points in consecutive light waves.
Wavelength is measured in micrometers or
nanometers.
Wavelength Accuracy The closeness of the
wavelength value reported by the instrument to
the actual value, determined by comparing
measured and known values of sharp lines. When
a continuum source (such as a tungsten or
deuterium lamp) is used, the monochromator
passes not just one wavelength, but a small
distribution of wavelengths. In this case, the
weighted center of the intensity distribution is
what should be used to calculate the accuracy.
Z
Z-Dimension The distance between the base of
the cuvette and the center of the light beam
incident on the cuvette.
Zero The act of setting the instrument to read
0.0 Abs on a reference material. See also blank.
16