HANDLING MICROORGANISMS AND
PATHOGENS
SafeFAST Top
Class II Microbiological Safety Cabinets
THE USER-FRIENDLY
PRACTICAL KEYBOARD
APPLICATIONS
SafeFAST Top Class II Microbiological Safety Cabinets have
been adopted worldwide for product, personnel and environmental protection while handling
harmful agents pathogenic to
human beings and/or animals as
defined in the appropriate International Standards, in a wide
range of applications such as:
Microbiology, Virology, Haematology, Cell culture, Genetics,
Handling of hazardous agents to
human beings or animals.
STRATIFIED SAFETY-GLASS
Stratified safety-glass front sash
window.
DIRECT DUCTING
As standard supplied with 200mm
diameter collar on the top of the
cabinet for optional direct ducting to
facility exhaust system.
SafeFAST Top
Class II Microbiological Safety Cabinets
OPERATIONAL
PRINCIPLES
UV STERILIZING LAMP
UV sterilizing lamp (optional) installed on the front closure-panel of the
SafeFAST Top complete with two
switch-off countdown timers, one
variable on a 03 hours scale (1 minute steps), the other set to 3 fixed
hours.
CABINET ELEMENTS
1 automatic safety service connection for gas, 1 for vacuum and 1
electrical socket fitted as standard
in each size module.
QUALITY
ASSURANCE
DEPARTMENT
LIGHTINGTEST
VIBRATION TEST
ELECTRICAL TEST
TECHNICAL SPECIFICATIONS
Each size available with single motorblower (S series) or Double motorblower (D series)
Description
Unit
SafeFAST Top
212 S/D
215 S/D
218 S/D
Overall Dimensions
1320
1625
1930
mm
WxHxD (1)
1470x785
1470x785
1470x785
Usefull Dimensions
1190
1495
1800
mm
WxHxD
660x580
660x580
660x580
Working aperture
mm
200
Maximum front aperture
mm
460
Weight
kg
185
215
260
300
Exhaust flow rate
m3/h
290
390
485
585
Noise level (1)
dB(A)
<57
<58
<59
<60
Lighting level
lux
>1000
>1100
>1200
>1200
Electrical Data
1/Ph+E
1/Ph+E
1/Ph+E
1/Ph+E
230V 50Hz
230V 50Hz
230V 50Hz
230V 50Hz
Current consumption S series (1)(2) A
1,9
2,1
3,4
3,6
Current consumption D serie (1)(2) A
2,2
2,4
3,9
4,2
Electrical class / IP
1 / 20
Internal electrical outlet
The electrical outlets have a total load capacity of 6A and are protected with one T6A fuse
Heat emission
W
175
240
295
360
209 S/D
1015
1470x785
885
660x580
Solid Work Surface Single Piece Work Surface UV light Additional Tap (Fuel Gas / Gas / Vacuum) Additional
Electrical Outlet Stainless Steel Hanging Bar Movable Stainless Steel Armrest Anti Blow Back Damper Thimble
Duct Exhaust Transition Additional Exhaust HEPA Filter (only D version) Additional Exhaust Carbon Activated Filter
(only D version) Pre-Filter Grid Floor Stand 900 mm Working Height With Footrest (other heights on request) Electric
Adjustable Floor Stand 800 to 1100 mm working height Floor Stand With Castors
Faster S.r.l.
Via Merendi, 22 20010 Cornaredo (MI) Italy
Tel +39 02 93 991 92 Fax +39 02 93 991 608
www.faster-air.com [email protected]
a
company
FA S T E R
F O R A
GREEN
FUTURE
Striving everyday to improve our environmental
performance, Faster developed environmental policies and procedures are
founded on three guiding
principles:
Protect the Environment
for present and future generations manufacturing
low energy consumption
equipments
Reduce risks and improve
efficiencies
Introduce improved technology and processes
Faster/safefast top/2009/01 In this brochure all pictures and specifications are purely indicative and vary without notice