Film-Screen Mammography QA: What You Need To Know: Objectives
Film-Screen Mammography QA: What You Need To Know: Objectives
Film-Screen Mammography QA: What You Need To Know: Objectives
Film-Screen
Mammography QA:
What You Need to
Know
Beth Schueler
Mayo Clinic Rochester
Medical Physicist
Responsibilities
Required for
New mammography units or processors
Disassembled and reassembled mammography
units and processors
After mammography units or processors have
undergone a major repair
ACR QC Manual
ACR QC Manual
MQSA 900.12(e)
1.Unit Assembly
1.Unit Assembly
2.Collimation
2.Collimation
3.System Resolution
3.System Resolution
4.AEC Performance
4.AEC Performance
6.Artifact Evaluation
6.Artifact Evaluation
(5)(vi) Dosimetry
8.kVp Accuracy/Reprod.
8.kVp Accuracy/Reprod.
10.Breast Dosimetry
10.Breast Dosimetry
11.Viewbox Luminance
11.Viewbox Luminance
(5)(xi) Decompression
(2) Image Quality Evaluation
ACR QC Manual
MQSA 900.12(e)
1.Unit Assembly
(5)(xi) Decompression
2.Collimation
(5)(vii) Collimation
3.System Resolution
4.AEC Performance
6.Artifact Evaluation
8.kVp Accuracy/Reprod.
10.Breast Dosimetry
11.Viewbox Luminance
1. Unit Assembly
Unit is mechanically stable
Cassettes do not slip when gantry is angled
Compressed breast thickness scale is
accurate to within 5 mm
Test with a phantom at 2, 4, and 6 cm thickness
10
2. Collimation
2. Collimation
11
2. Collimation
3. System Resolution
Action Limits:
X-ray to light:
L + R or anterior + chest deviations > 2% SID
X-ray to IR:
Compression paddle:
14
3. System Resolution
Repeat for all target
materials, focal spots,
film-screen combinations
View pattern on the film
with a 7-10X magnifier
Action limits:
15
16
ANTERIOR
13
CHEST WALL
ANODE-CATHODE
AXIS
Action limit:
18
19
20
Example 1: Thickness/kVp
Tracking
Example 1: Thickness/kVp
Tracking
2
Film Optical Density
2
1.9
1.8
1.7
1.6
1.5
1.4
2
1.7
1.6
1.5
21
22
Example 1: Thickness/kVp
Tracking
Example 2: Thickness/kVp
Tracking
2
1.9
24 kVp
1.8
25 kVp
25 kVp
2
Film Optical Density
1.8
1.4
1
28 kVp
1.7
1.6
1.5
26 kVp
1.4
1
23
1.9
1.9
1.8
1.7
1.6
1.5
1.4
24
Example 2: Thickness/kVp
Tracking
2
1.9
25 kVp 25 kVp
1.8
Action limit:
1.7
1.6
24 kVp
1.5
28 kVp
1.4
1
25
26 kVp
26
27
Action limit:
29
6. Artifact Evaluation
6. Artifact Evaluation
Repeat for
31
32
6. Artifact Evaluation
6. Artifact Evaluation
Processor artifacts
will be oriented on
both films in the
same direction
relative to the
direction of film
feed
33
34
Artifact Problem-solving
Artifact Problem-solving
35
36
6. Artifact Evaluation
ACR Mammography
Accreditation Phantom
with contrast disc
AEC exposure using
clinical technique factors
for a 4.2-cm thick breast
38
Action limit:
Additional recommendations:
Background density
< 1.4 OD
Background density
1.8 OD or higher
within 0.20 OD of facilitys target
Density difference
Density difference
< 0.40
0.55 or higher
within 0.05 of facilitys target
39
41
40
Count as 0.5
if 2 or 3 specks visible
Action limit:
< 4 fibers, < 3 speck
groups, < 3 masses
visible
Change in score of >
0.5 from previous
measurement
Count as 0.5
if visible and not generally
circular
44
kVp accuracy
Action limits:
kVp reproducibility
45
46
Breast entrance
exposure
Action limit:
MQSA: HVL < kVp/100 (mm Al)
ACR: HVL < kVp/100 + 0.3 (mm Al) or
47
AEC Reproducibility
Repeat 4 times
Action limit:
Dg = DgN x XESE
Action limit:
Dg > 3 mGy
50
Photometers
Vendors include:
UDT Instruments
Quantum Instruments
Various display monitor
luminance measuring
units (MSfit ACT, )
Action limit:
Radiation output rate < 7.0 mGy/sec
51
52
Report
Action limits:
kcd/m2
Corrective Action
56
Practical Tips
kVp first
Phantom technique
Dosimeter tests
57
58
Small bucky
Collimation, Resolution
AEC
Reading room
59
60
10
# of Failures
Found *
35
QC Test
Unit Assembly: Thickness readout
AEC: Thickness/kVp tracking
Uniformity Screen Speed
System Resolution
Image Quality Evaluation
kVp Accuracy/Reprod.
Beam Quality (HVL)
Breast Dosimetry
34
62
Mammography Equipment
Evaluation
63
64
Major Repairs
Required for
Item
# of Failures
Found *
31
24
5
0
0
0
0
0
Major
MP
Repair Involvement
Y
On-site testing
Y
On-site testing
Filter replacement
On-site testing
Collimator/blade replacement
Collimator adjustment
Y
N
On-site testing
Oversight
Oversight
Oversight
Optional
66
11
Can be
downloaded
as an Excel
file from ACR
website
www.acr.org
900.12(b)(5)
Illumination of not
less than 160 lux at
the maximum SID
Difference between
readings with and
without field light on
Photometer
67
68
Compress foam or
tennis balls to 111 N
(25 lbs)
Measure the height
of each corner of
the paddle above
the breast tray
Maximum deviation
< 1.0 cm
70
71
72
12
Information Resources
Information Resources
http://www.fda.gov/cdrh/mammography/index.html
http://www.geocities.com/djsimpkin/
http://www.acr.org/
Test data forms in Excel
74
13