Lab Report
Lab Report
3. Set the power supply to each voltage listed in the table 1. Measured each voltage
using the above steps as a guide. The first line of the table has been completed as an
example. To obtained accurate readings with the oscilloscope, it is necessary to
selected the VOLT/DIV that gives several division of change between the ground
reference and the voltage to be measured. The reading on the oscilloscope and meter
should agree with each other within approximately 3%.
4. Before viewing ac signals, it is a good idea to checked the probe compensation for
your oscilloscope. To checked compensation, set the VOLT/DIV control to 0.1V/DIV,
the AC-GND-DC coupling control to DC, and the SEC/DIV control 2ms/div. Touched
the probe tip to the PROBE COMP connecter. Should observed a square wave with a
square wave with a flat top and square corners. If necessary, adjusted the
compensation to achieved a good square wave.
5. Set the function generator for an ac waveform with a frequency of 1.0kHz. Adjusted
the amplitude of the function generator for 1.0V rms as a read on your DMM. Set the
SEC/DIV control to 0.2ms/div and the VOLTS/DIV to 0.5V/div. Connected the scope
probe and its grounded to the function generator. Adjusted the vertical POSITION
control and trigger LEVEL control for a stable display near the center of the screen.
Should observed approximately two cycles of an ac waveform with a peak-to-peak
amplitude of 2.8V. This represents 1.0Vrms as shown in figure 1.1.
6. Used the DMM to set the function generator amplitude to each value listed in the
table 1.2. Repeated the ac voltage measurement as outline in step 4. The first line of
the table has been completed as an example.
VOLT/DIV
setting
1.0V
2.5V
4.5V
8.3V
0.2volt/div
1volt/div
2volt/div
5volt/div
Power
generator
amplitude
1.0V
2.5V
4.5V
7.2V
VOLT/DIV
setting
0.5volt/div
2volt/div
2volt/div
2volt/div
Number of
divisions of
deflection(div)
5.0div
2.4div
2.25div
1.6div
Table 1.1
Oscilloscope
(measured
voltage)
1.0V
2.4V
4.5V
8V
DMM
(Measured
voltage)
1.0V
2.49V
4.54V
8.29V
Number of
divisions of
deflection(div)
5.6div
2.8div
4.5div
7.8div
Table 1.2
Oscilloscope
measured
(peak-to-peak)
2.8Vpp
5.6Vpp
9Vpp
15.6Vpp
Oscilloscope
measured
(rms)
1.0Vrms
2.0 Vrms
3.2 Vrms
5.5 Vrms
Graph Measured DC Voltage (V) over Power Supply (V) for Digital Multimeter and Oscilloscope
Graph Measured AC Voltage (V) over Power Supply (V) for Oscilloscope.
A 1
Relative accuracy,
Yn X n
Yn
1. Yn = 1.0V, Xn = 1.0V
% Accuracy =
= 100 %
2. Yn = 2.49 V, Xn = 2.4V
% Accuracy =
= 96.39 %
3. Yn = 4.54 V, Xn = 4.5V
% Accuracy =
= 99.1%
4. Yn = 8.29 V, Xn = 8V
% Accuracy =
= 96.5%
(b)
The purpose of this group is used to display and adjust the signal for
optimal viewing. It includes display screen, the intensity-control knob, the
beam-find button, the focus-control, and the power switch.
ii.
iii.
iv.
b) Vertical control
i.
This group is used to adjust the vertical components (Y-axis) and the
vertical positon of the signal.
ii.
It consists of the vertical-position knobs, channel-selector switch, voltsper-division selector knobs, input-coupling switch, the channel-modeselector switch, the channel-2-invert switch.
iii.
iv.
v.
vi.
A division is one block on the screen and each tick mark is one fifth of a
division or 0.2. This knob is marked in both volts and millivolts.
vii.
c) Triggering control
i.
ii.
The trigger group is made up of the trigger level knob, the rising/falling
edge switch, the trigger mode switch, the hold off knob, the trigger source
switches, and the trigger coupling switch.
iii.
The trigger-level knob sets the voltage level at which the oscilloscope
will trigger. When the signal is not stable, the trigger level may be too
high or too low for the oscilloscope to recognize the signal.
iv.
v.
vi.
Then, the trigger-mode switch will normally be set to AUTO but it can
be changed to NORM.
vii.
viii.
ix.
d) Horizontal control
i.
The Horizontal Control Group have various knobs and switch which are
coarse and fine position knobs, the horizontal magnifications switch, the
second-per-division-selector knob, and the magnification-scale selector
switch.
ii.
iii.
iv.
v.
vi.
vii.
The seconds-per-division knob sets the time base for the horizontal scale. It
is marked in seconds, milliseconds, and microseconds.
6. If you having difficulty obtaining a stable display, which group of controls should you
adjust?
Trigger control. This control sets an internal voltage which is compared to the voltage
of the input signal. As the input signal same as the trigger voltage, the scope trigger.
In order to get a stable image, the level knob need to be turn back and forth slowly
until the stable image obtained.
7. (a) If an ac waveform has 3.4 divisions from peak-to-peak and the VOLT/DIV control
is set to 5.0V/Div, what is the peak-to-peak voltage?
Vpp = 5.0x3.4
=17.5V
(b)
Vrms = Vpp / 22
= 17.5 / 22
= 6.187V
The RMS voltage is the effective value of a varying voltage. It is same to steady DC
constant value so they have same effect. As example, a lamp connected to a 6V RMS
AC supply would have same brightness when connected to a steady 6V DC supply.
Others example, a sin wave with an RMS value of 8V RMS, required 8V DC to
supply same power via DC voltage. This is because the RMS wave value is equivalent
to steady DC value.
8. If you want to view the amplitude of an ac waveform that is 20.0Vrms, what setting of
the volt/div is best?
By assuming the best number of division of deflection is 4, then the best setting is 20
volts/div. This is because the larger the portion of display, the more accurate the
measurement.
9. The most accurate way to measure a waveform on an oscilloscope is to use a large
portion of the display area. Why?
This is because if a volts/div setting is 5 volts, then each of the vertical major division
represents 5 volts. However, if the setting is 0.5 volts/div, then each of the vertical
major division represents 0.5 volts. Therefore, by having large portion of display area,
the vertical minor division would be able represent smaller changes of voltage in
waveform.