Agilent G1316 90011 TCC A B C Ebook

Download as pdf or txt
Download as pdf or txt
You are on page 1of 126

Agilent 1200 Series

Thermostatted Column
Compartment
G1316A/G1316B/G1316C

User Manual
1200 Series TCC User Manual

Agilent Technologies

Notices
Agilent Technologies, Inc. 1995,
1996-2008
No part of this manual may be reproduced
in any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.

Manual Part Number


G1316-90011

Edition
09/2008
Printed in Germany
Agilent Technologies
Hewlett-Packard-Strasse 8
76337 Waldbronn

Research Use Only


Not for use in Diagnostic Procedures.

Manual Structure
The User Manual G1316-90011 (English)
and its localized versions contain a subset
of the Service Manual and is shipped with
the Thermostatted Column Compartment in
printed matter.
Latest versions of the manuals can be
obtained from the Agilent web.

Warranty
The material contained in this document is provided as is, and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.

receive no greater than Restricted Rights as


defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.

Safety Notices

CAUTION
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could
result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated conditions are fully understood and
met.

Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend


If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
Commercial computer software as
defined in DFAR 252.227-7014 (June 1995),
or as a commercial item as defined in FAR
2.101(a) or as Restricted computer software as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation
or contract clause. Use, duplication or disclosure of Software is subject to Agilent
Technologies standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will

WA R N I N G
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice,
or the like that, if not correctly
performed or adhered to, could
result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated conditions are fully understood and met.

1200 Series TCC User Manual

In This Guide

In This Guide

This manual covers the Agilent 1200 Series Thermostatted Column


Compartments (TCC)
G1316A Agilent 1200 Series TCC
G1316B Agilent 1200 Series TCC SL
G1316C Agilent 1200 Series TCC SL Plus

1 Introduction to the Column Compartment


This chapter gives an introduction to the TCC, instrument overview and
internal connectors.

3 Installing the Column Compartment


This chapter describes the installation of the Thermostatted Column
Compartment.

4 How to optimize the Column Compartment


This chapter provides information on how to optimize the Thermostatted
Column Compartement.

5 Troubleshooting and Diagnostics


Overview about the troubleshooting and diagnostic features.

6 Test Functions
This chapter describes the TCCs built in test functions.

1200 Series TCC User Manual

In This Guide

7 Maintenance
This chapter describes the maintenance of the TCC.

8 Parts and Materials for Maintenance


This chapter provides information on parts for maintenance.

9 Appendix
This chapter provides addition information on safety, legal and web.

1200 Series TCC User Manual

Contents

Contents

1 Introduction to the Column Compartment

Main Features 10
System Overview 11
Column-Identification System 13
Column Switching Valve (Optional for G1316A/G1316B SL)
Electrical Connections 18
Instrument Layout 21
Agilent Diagnostic Software 22
2 Site Requirements and Specifications

23

Site Requirements and Specifications 24


Physical Specifications 27
Performance Specifications 28
Extended Specifications on G1316B SL/G1316C SL Plus
3 Installing the Column Compartment

15

30

33

Unpacking the Column Compartment 34


Optimizing the Stack Configuration 35
Installing the Column Compartment 38
Installing Valve Heads (G1316C SL Plus) 42
Flow Connections of the Column Compartment 44
Installation of Heater and Cooling Devices 47
Placing Columns 49
4 How to optimize the Column Compartment

51

Optimizing the Performance of your Column Compartment


Using Additional Heater and Cooling Devices 53

1200 Series TCC User Manual

52

Contents

5 Troubleshooting and Diagnostics

55

Overview of the Column Departments Indicators and Test Functions


Status Indicators 57
Available Tests depending on User Interfaces 59
Agilent Diagnostic Software 60
6 Test Functions

61

Thermostat Function Test 62


Pressure Test 65
Column Thermostat Temperature Calibration
7 Maintenance

56

66

71

Introduction to Maintenance and Repair 72


Warnings and Cautions 73
Overview of Maintenance 75
Cleaning the Column Compartment 76
Changing Column Identification Tags 77
Replacing Head Parts of Column Switching Valve (G1316A/G1316B SL)
Adding Heater and Cooling Devices (G1316B SL/G1316C SL Plus) 82
Correcting Leaks 84
Replacing the Column Compartments Firmware 85
Replacing Valve Heads (G1316C SL Plus) 86
Preparing the G1316C SL Plus for Transportation 89
8 Parts and Materials for Maintenance

79

91

Valve Options Overview 92


Heater and Cooling Devices for G1316B SL/G1316C SL Plus 93
Column Switching Valve 2 Position/6 Port (G1316A/G1316B SL) 95
Column Switching Valve 8 Position-9 Port (G1316C SL Plus) 97
Micro Column Switching Valve 2 Postion/6 Port (G1316A/G1316B SL) 99
Micro Column Switching Valve 2 Position/10 Port (G1316A/G1316B SL) 100
Accessory Kits 102
Plastic Parts 109
Leak Parts 110

1200 Series TCC User Manual

Contents

9 Appendix

113

Safety Symbols 114


The Waste Electrical and Electronic Equipment Directive
Lithium Batteries Information 118
Radio Interference 119
Sound Emission 120
Solvent Information 121
Agilent Technologies on Internet 123

1200 Series TCC User Manual

117

Contents

1200 Series TCC User Manual

1200 Series TCC User Manual

1
Introduction to the Column
Compartment
Main Features

10

System Overview

11

Column-Identification System

13

Column Switching Valve (Optional for G1316A/G1316B SL)


Electrical Connections 18
Serial Number Information
Instrument Layout

15

20

21

Agilent Diagnostic Software

22

This chapter gives an introduction to the TCC, instrument overview and


internal connectors.

Agilent Technologies

Introduction to the Column Compartment


Main Features

Main Features

The Agilent 1200 Series thermostatted column compartments are stackable


temperature-controlled column compartments for LC. They are available as
standalone modules or as a component of a Agilent 1200 Series system. They
are used for heating and cooling to meet extreme requirements of retention
time reproducibility.
The main features are:
Peltier heating and cooling from 10 degrees below ambient up to
80 C(G1316A) or 100 C (G1316B SL/G1316C SL Plus) with high heating
and cooling speeds for maximum application flexibility and stability,
holds up to three 30-cm columns and optimized design gives minimum dead
volumes and maximum efficiency,
two independently programmable heat exchangers contribute volumes of
only 3 and 6 l,
G1316B SL features additional heating and cooling devices for low flow
rates, which reduce the risk of additional dispersion,
G1316B SL and G1316C SL Plus can be supplemented by a kit to install a
small heat-exchanger with 1.6 l delay volume to reduce the delay volume.
In additon a cooling device with 1.5 l is available,
electronic column-identification module as standard for GLP
documentation of column type, and major column parameters,
optional high-quality Rheodyne column switching valves with ceramic
stator-face assemblies for prolonged lifetime.
For specifications, see Performance Specifications on page 28.

10

1200 Series TCC User Manual

Introduction to the Column Compartment


System Overview

System Overview

The Concept of Heating and Cooling


The design of this thermostatted column compartment uses column heating
and cooling devices with Peltier elements. The solvent entering the column
compartment is heated up or cooled down to a settable temperature with two
low-volume heat exchangers (3 l on left side, 6 l on right side), made of a
short piece of capillary 0.17 mm i.d. leading through a heat exchanger. The
heat exchanger is designed such that it can function simultaneously as an air
heater. The shape of the heat exchanger surface allows the area around the
column to be kept at a similar temperature level as the liquid running through
the column. This is done by thermal convection and radiation between the
heat exchanger fins. This design ensures that the column and the solvent
flowing through it are almost at the same temperature.
Actual temperature control is accomplished at the heat exchanger. The solvent
cools down or heats up on its transfer from the heating block to the column
inlet. This depends on several factors: flow rate, setpoint temperature,
ambient temperature and column dimensions.
In a flow-through temperature regulation system, there are necessarily slightly
different temperatures at different positions. If, for example, the temperature
set by the user is 40 C, then the heat exchanger is regulated to a temperature
40.8 C which is different by a certain offset (here. 0.8 C). The solvent
temperature at the column entry would be about 39 C.
The actual temperature displayed on the user interface is always the derived
temperature taken at the heat exchanger, corrected by the offset explained
above.
Any type of heated column compartment brings one important consequence
for column temperature equilibration. Before an equilibrium is reached, the
whole mass of column, column packing, and solvent volume inside the column
has to be brought to the selected temperature. This depends on several factors:
flow rate, setpoint temperature, ambient temperature and column dimensions.
The higher the flow rate, the faster the column equilibrates (due to
thermostatted mobile phase).

1200 Series TCC User Manual

11

Introduction to the Column Compartment


System Overview
Column Thermostat Temperature Calibration on page 66 shows a setpoint
temperature of 40 C. Some time after entering the setpoint the heat
exchanger has reached its temperature and the control activity starts. The
TEMPERATURE NOT READY signal will be cancelled 20 seconds after the sensed
temperature was within a range of 0.5 C of the setpoint (other values can be
set via the user interface). However this does not necessarily mean that the
column has already reached the correct temperature. The equilibration of the
column may take longer. Stability of the pressure signal is a good indication
for equilibrium.

IZbeZgVijg8

=ZViZmX]Vc\ZgiZbeZgVijgZ

8dajbciZbeZgVijgZ

:mVbeaZ[dgV[adlgViZd[*ba$b^cd[lViZg

I^bZb^c

Figure 1

Equilibration of Heat Exchanger and Column Temperature

The temperature calibration and verification is described in the Service


Manual.

12

1200 Series TCC User Manual

Introduction to the Column Compartment


Column-Identification System

Column-Identification System

The Agilent 1200 Series thermostatted column compartment is equipped with


a column-identification system. It allows to read and write column-specific
information to and from the column-identification tag.

6ciZccVh

8dajbc"^YZci^[^XVi^dciV\

8dajbcXa^e

Figure 2

Column-Identification System

Table 1 on page 13 shows the information that can be stored. The information
fields can be edited via the user interface.
Table 1

Column-Identification Module Information

Item

Example

Product number

79916OD-552

Serial number

950522

Batch number

1675

Geometry [mm]

100 2.1

Stationary phase

ODS Hypersil

Particle size

10 m

Number of injections

1267

Maximum pressure allowed [bar]

400

1200 Series TCC User Manual

Comment

Date of manufacturing

See Note below.

13

Introduction to the Column Compartment


Column-Identification System
Table 1

Column-Identification Module Information

Item

Example

Maximum temperature recommended [C]

70

Maximum pH recommended

12

Comment

Column void volume [ml]

The number of injections will be updated each run to create a column lifecycle
(history). The user interface allows to edit all information.

NOTE

14

If a column switching valve (see Column Switching Valve (Optional for G1316A/G1316B
SL) on page 15) is installed in the module, the update of the number of injections depends
on the position of the column switching valve. For example, if the left column is selected,
the right column is not updated, and vice versa. If no column switching valve is installed
both sides are updated at the same time.

1200 Series TCC User Manual

Introduction to the Column Compartment


Column Switching Valve (Optional for G1316A/G1316B SL)

Column Switching Valve (Optional for G1316A/G1316B SL)

Figure 3

1200 Series TCC User Manual

Location of Column Switching Valve

15

Introduction to the Column Compartment


Column Switching Valve (Optional for G1316A/G1316B SL)

Two Column Selection


The valve can select either column 1 or column 2. The offline column is sealed
by connecting head to tail. Switching should be done when the flow is off and
the pressure is zero.

NOTE

Before switching the valve, switch off the pump or set the flow to zero. Keeping the flow on
while the valve is switched can cause exceeding the maximum pressure. This will stop
method or sequence execution.

;gdbVjidhVbeaZg
=ZViZgVhhZbWan&

=ZViZgVhhZbWan'

8dajbc&

8dajbc'

IdYZiZXidg

Figure 4

Column 1 Active
;gdbVjidhVbeaZg
=ZViZgVhhZbWan'

=ZViZgVhhZbWan&

8dajbc'

8dajbc&

IdYZiZXidg

Figure 5

16

Column 2 Active

1200 Series TCC User Manual

Introduction to the Column Compartment


Column Switching Valve (Optional for G1316A/G1316B SL)

Precolumn Back-flushing
The sample is injected into series-connected precolumn and analytical
column. After the valve has switched, the analytical column flow continues in
normal direction. Only the precolumn is back-flushed, eluting highly retained
peaks directly to the detector.
;gdbVjidhVbeaZg
=ZViZgVhhZbWan'

=ZViZgVhhZbWan&

EgZXdajbc

6cVani^XVa
8dajbc

IdYZiZXidg

Figure 6

1200 Series TCC User Manual

Precolumn Back-flushing

17

Introduction to the Column Compartment


Electrical Connections

Electrical Connections

The CAN bus is a serial bus with high speed data transfer. The two
connectors for the CAN bus are used for internal Agilent 1200 Series
module data transfer and synchronization.
One analog output provides signals for integrators or data handling
systems.
The REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shut down, prepare, and so on.
With the appropriate software, the RS-232C connector may be used to
control the module from a computer through a RS-232C connection. This
connector is activated and can be configured with the configuration switch.
See your software documentation for further information.
The power input socket accepts a line voltage of 100 240 volts AC 10%
with a line frequency of 50 or 60 Hz. Maximum power consumption is
220 VA. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible
fuses, because automatic electronic fuses are implemented in the power
supply. The security lever at the power input socket prevents the module
cover from being taken off when line power is still connected.

NOTE

18

Never use cables other than the ones supplied by Agilent Technologies to ensure proper
functionality and compliance with safety or EMC regulations.

1200 Series TCC User Manual

Introduction to the Column Compartment


Electrical Connections

HZXjg^inaZkZg

6E<GZbdiZ

GH"'('8

86C

8dc[^\jgVi^dchl^iX]

EdlZg

hV[ZinhiVcYVgYh
Xdc[^\jgVi^dchl^iX]
hZii^c\h
kdaiV\ZgVc\Z
edlZgXdchjbei^dc$
[gZfjZcXn

Figure 7

1200 Series TCC User Manual

Rear View of Oven - Electrical Connections and Label

19

Introduction to the Column Compartment


Electrical Connections

Serial Number Information


The serial number information on the instrument labels provide the following
information:

20

CCYWWSSSSS

Format

CC

country of manufacturing
DE = Germany
JP = Japan
CN = China

YWW

year and week of last major manufacturing change, e.g. 820


could be week 20 of 1998 or 2008

SSSSS

real serial number

1200 Series TCC User Manual

Introduction to the Column Compartment


Instrument Layout

Instrument Layout

The industrial design of the module incorporates several innovative features.


It uses Agilents E-PAC concept for the packaging of electronics and
mechanical assemblies. This concept is based upon the use of expanded
polypropylene (EPP) layers foam plastic spacers in which the mechanical and
electronic boards components of the module are placed. This pack is then
housed in a metal inner cabinet which is enclosed by a plastic external
cabinet. The advantages of this packaging technology are:
virtual elimination of fixing screws, bolts or ties, reducing the number of
components and increasing the speed of assembly/disassembly,
the plastic layers have air channels molded into them so that cooling air can
be guided exactly to the required locations,
the plastic layers help cushion the electronic and mechanical parts from
physical shock, and
the metal inner cabinet shields the internal electronics from
electromagnetic interference and also helps to reduce or eliminate radio
frequency emissions from the instrument itself.

1200 Series TCC User Manual

21

Introduction to the Column Compartment


Agilent Diagnostic Software

Agilent Diagnostic Software

The Agilent Diagnostic Software is a standalone product that can be used with
or without data system. Agilent Lab Advisor helps to manage the lab for high
quality chromatographic results and can monitor in real time a single Agilent
LC or all the Agilent GCs and LCs configured on the lab intranet.
Agilent Lab Advisor provides diagnostic capabilities for all Agilent 1200 Series
HPLC modules. This includes tests and calibrations procedures as well as the
different injector steps to perform all the maintenance routines.
Agilent Lab Advisor also allows users to monitor the status of their LC
instruments. The Early Maintenance Feedback (EMF) feature helps to carry
out preventive maintenance. In addition, users can generate a status report for
each individual LC instrument. The tests and diagnostic features as provided
by the Agilent Lab Advisor Software may differ from the descriptions in this
manual. For details refer to the Agilent Lab Advisor help files.
This manual provides lists with the names of Error Messages, Not Ready
messages, and other common issues.

22

1200 Series TCC User Manual

1200 Series TCC User Manual

2
Site Requirements and Specifications
Site Requirements and Specifications
Power Consideration 24
Power Cords 25
Bench Space 26
Environment 26
Physical Specifications

24

27

Performance Specifications

28

Extended Specifications on G1316B SL/G1316C SL Plus

Agilent Technologies

30

23

Site Requirements and Specifications


Site Requirements and Specifications

Site Requirements and Specifications

A suitable environment is important to ensure optimal performance of the


instrument.

Power Consideration
The module power supply has wideranging capability (see Table 2 on page 27).
It accepts any line voltage in the range described in the above mentioned table.
Consequently there is no voltage selector in the rear of the module. There are
also no externally accessible fuses, because automatic electronic fuses are
implemented in the power supply.

WA R N I N G

Incorrect line voltage at the instrument


Shock hazard or damage of your instrumentation can result, if the devices are
connected to a line voltage higher than specified.
Connect your instrument to the specified line voltage.

WA R N I N G

Module is partially energized when switched off, as long as the power cord is
plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the
cover is opened and the module is connected to power.
Remove the power cable from the instrument before opening the cover.
Do not connect the power cable to the Instrument while the covers are removed.

24

1200 Series TCC User Manual

Site Requirements and Specifications


Site Requirements and Specifications

CAUTION

Unaccessable power plug.


In case of emergency it must be possible to disconnect the instrument from the power
line at any time.
Make sure the power connector of the instrument can be easily reached and

unplugged.
Provide sufficient space behind the power socket of the instrument to unplug the

cable.

Power Cords
Different power cords are offered as options with the module. The female end
of all power cords is identical. It plugs into the power-input socket at the rear
of the module. The male end of each power cord is different and designed to
match the wall socket of a particular country or region.

WA R N I N G

The absence of ground connection and the use of an unspecified power cord can
lead to electric shock or short circuit.
Electric Shock
Never operate your instrumentation from a power outlet that has no ground

connection.
Never use a power cord other than the Agilent Technologies power cord designed

for your region.

WA R N I N G

Use of unsupplied cables


Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
Never use cables other than the ones supplied by Agilent Technologies to ensure

proper functionality and compliance with safety or EMC regulations.

1200 Series TCC User Manual

25

Site Requirements and Specifications


Site Requirements and Specifications

Bench Space
The column compartment dimensions and weight (see Physical
Specifications on page 27 ) allow to place this module on almost any desk or
laboratory bench. It needs an additional 2.5 cm (1.0 inches) of space on either
side and approximately 8 cm (3.1 inches) in the rear for the circulation of air
and electric connections.
If the bench should carry a complete Agilent Series system, make sure that
the bench is designed to carry the weight of all the modules.
The module should be operated in a horizontal position.

Environment
Your column compartment will work within specifications at ambient
temperatures and relative humidity as described in Physical
Specifications on page 27.

26

1200 Series TCC User Manual

Site Requirements and Specifications


Physical Specifications

Physical Specifications

Table 2

Physical Specifications

Type

Specification

Weight

10.2 kg (22.5 lbs)

Dimensions
(width depth height)

410 435 140 mm (16.1 17 5.5


inches)

Line voltage

100 240 VAC, 10%

Line frequency

50 or 60 Hz, 5%

Power consumption

320 VA / 150 W / 512 BTU

Ambient operating
temperature

0 55 C (32 131 F)

Ambient non-operating
temperature

-4070 C (-4158 F)

Humidity

< 95%, at 2540 C (77104 F)

Operating Altitude

Up to 2000 m (6500 ft)

Non-operating altitude

Up to 4600 m (14950 ft)

For storing the module

Safety standards: IEC, CSA,


UL

Installation Category II, Pollution


Degree 2

For indoor use only. Research


Use Only. Not for use in
Diagnostic Procedures.

1200 Series TCC User Manual

Comments

Wide-ranging capability

Maximum

Non-condensing

27

Site Requirements and Specifications


Performance Specifications

Performance Specifications

Table 3

28

Performance Specifications Thermostatted Column Compartment

Type

Specification

Comments

Temperature
range

10 degrees below ambient to 80 C


10 degrees below ambient to 100 C

G1316A
G1316B (SL)/G1316C (SL Plus)

up to 80 C: flow rates up to 5 ml/min


up to 100 C: flow rates up to
2.5 ml/min

G1316A/G1316B (SL)/G1316C (SL Plus)


G1316B (SL)/G1316C (SL Plus)

Temperature
stability

0.15 C
0.05 C

G1316A
G1316B (SL)/G1316C (SL Plus)

Temperature
accuracy

0.8 C
0.5 C

With calibration

Column
capacity

Three 30 cm

Warm-up/co
ol-down time

5 minutes from ambient to 40 C


10 minutes from 40 20 C

Dead volume

3 l left heat exchanger


6 l right heat exchanger

i.d. 0.17 mm, see Extended


Specifications on G1316B SL/G1316C SL
Plus on page 30

Communicati
ons

Controller-area network (CAN), GPIB,


RS-232C, APG Remote: ready, start,
stop and shut-down signals, LAN via
other 1200 series module

no GPIB on G1316B (SL)/G1316C (SL


Plus)

Safety and
maintenance

Extensive diagnostics, error detection


and display (through control module
and Agilent ChemStation), leak
detection, safe leak handling, leak
output signal for shutdown of pumping
system. Low voltages in major
maintenance areas.

1200 Series TCC User Manual

Site Requirements and Specifications


Performance Specifications
Table 3

NOTE

Performance Specifications Thermostatted Column Compartment

Type

Specification

GLP features

Column-identification module for GLP


documentation of column type, see
Column-Identification System on
page 13

Housing

All materials recyclable.

Comments

All specifications are valid for distilled water at ambient temperature (25 C), set point at
40 C and a flow range from 0.25 ml/min.

1200 Series TCC User Manual

29

Site Requirements and Specifications


Extended Specifications on G1316B SL/G1316C SL Plus

Extended Specifications on G1316B SL/G1316C SL Plus

The 1200 series G1316B SL/G1316C SL Plus thermostatted column


compartment is usable from 10 C below ambient up to 80 C for flow ranges
up to 5 ml/min and up to 100 C for flow ranges up to 2.5 ml/min. Additional
heating and cooling devices are available for the G1316B SL/G1316C SL Plus
to reduce the risk of additional dispersion at low flow rates see Figure 8 on
page 30. These devices can be installed in any position in the column
compartment, see Installation of Heater and Cooling Devices (G1316B
SL) on page 47.
A"h]VeZYegZ"Xdajbc
]ZViZg=ZViZg
adc\"je

A"h]VeZYegZ"Xdajbc
]ZViZg=ZViZg
adc\"Ydlc
A"h]VeZYegZ"Xdajbc]ZViZg
KdajbZ/&#+a
bdjciZYdcXVgg^Zg
ide[dgXdajbc&!
Wdiidb[dgXdajbc'

Figure 8

J"h]VeZYedhi"XdajbcXddaZg
KdajbZ/&#*a
bdjciZYdcXVgg^Zg

New additional heater and cooling devices

The standard column compartment is equipped with a 3 l and 6 l heater or


cooler. Both can be set at the same or different temperature. To reduce the
delay volume, a kit (G1316B SL/G1316C SL Plus Capillary System Kit on
page 104) has been set up for installing a small heaters with 1.6 l internal
delay volume and also a new cooling device with 1.5 l internal volume is
available.

30

1200 Series TCC User Manual

Site Requirements and Specifications


Extended Specifications on G1316B SL/G1316C SL Plus

NOTE

If the additional heater and cooling devices are used as shown in Figure 8 on page 30, the
column identification system cannot be used. If the column identification system is
required, fix the heater and cooling devices in the upper or lower locations or fix them
right/left of the current location.

1200 Series TCC User Manual

31

32

Site Requirements and Specifications


Extended Specifications on G1316B SL/G1316C SL Plus

1200 Series TCC User Manual

1200 Series TCC User Manual

3
Installing the Column Compartment
Unpacking the Column Compartment
Delivery Checklist 34
Optimizing the Stack Configuration
Installing the Column Compartment

34
35
38

Installing Valve Heads (G1316C SL Plus)

42

Flow Connections of the Column Compartment


Installation of Heater and Cooling Devices
Placing Columns 49
Column-Identification Tag
Column Clip 50

44

47

49

This chapter describes the installation of the Thermostatted Column


Compartment.

Agilent Technologies

33

Installing the Column Compartment


Unpacking the Column Compartment

Unpacking the Column Compartment

If the delivery packaging shows signs of external damage, please call your
Agilent Technologies sales and service office immediately. Inform your service
representative that the module may have been damaged during shipment.

CAUTION

"Defective on arrival" problems


If there are signs of damage, please do not attempt to install the module. Inspection by
Agilent is required to evaluate if the instrument is in good condition or damaged.
Notify your Agilent sales and service office about the damage.
An Agilent service representative will inspect the instrument at your site and

initiate appropriate actions.

Delivery Checklist
Ensure all parts and materials have been delivered with the module. The
delivery checklist is shown below. Please report missing or damaged parts to
your local Agilent Technologies sales and service office.
Table 4

34

Column Compartment Delivery Checklist

Description

Quantity

Thermostatted column compartment

Power cable

CAN cable

Column switching valve

optional

User Manual

Accessory kit (see Accessory Kits on


page 102)

1200 Series TCC User Manual

Installing the Column Compartment


Optimizing the Stack Configuration

Optimizing the Stack Configuration

If your column compartment is part of a Agilent 1200 Series system, you can
ensure optimum performance by installing the following configuration. This
configuration optimizes the system flow path and ensures minimum delay
volume.
For installations of the G1316C SL Plus as part of the Method Development
Solution, please refer to the Method Development Solution User and
Installation Guide part number: G4230-90000.

1200 Series TCC User Manual

35

Installing the Column Compartment


Optimizing the Stack Configuration

HdakZciXVW^cZi

KVXjjbYZ\VhhZg
Ejbe
AdXVa
JhZg>ciZg[VXZ

6jidhVbeaZg

8dajbcXdbeVgibZci

9ZiZXidg

Figure 9

36

Recommended Stack Configuration (Front View)

1200 Series TCC User Manual

Installing the Column Compartment


Optimizing the Stack Configuration

68edlZg

GZbdiZXVWaZ

86C7jhXVWaZid
adXVajhZg^ciZg[VXZ

86C7jhXVWaZ

A6CidA8
8]ZbHiVi^dcadXVi^dc
YZeZcYhdcYZiZXidg
6cVad\YZiZXidg
h^\cVa
&dg'djiejiheZg
YZiZXidg

Figure 10

1200 Series TCC User Manual

Recommended Stack Configuration (Rear View)

37

Installing the Column Compartment


Installing the Column Compartment

Installing the Column Compartment

Parts required

Description

Column compartment

Power cord
For other cables see text below

Preparations

Locate bench space.


Provide power connections.
Unpack the Column compartment.

WA R N I N G

Module is partially energized when switched off, as long as the power cord is
plugged in.
Risk of stroke and other personal injury. Repair work at the module can lead to
personal injuries, e. g. shock hazard, when the module cover is opened and the
instrument is connected to power.
Never perform any adjustment, maintenance or repair of the module with the top

cover removed and with the power cord plugged in.


The security lever at the power input socket prevents that the module cover is taken

off when line power is still connected. Never plug the power line back in when cover
is removed.

CAUTION

Valve properties are read from the valve head RFID tag during initialization of the
module. Valve properties will not be updated, if the valve head is replaced while the
module is on.
Selection of valve port positions can fail, if the instrument does not know the
properties of the installed valve.
Always switch off the instrument when replacing the valve head.

38

1200 Series TCC User Manual

Installing the Column Compartment


Installing the Column Compartment

CAUTION

The valve actuator contains sensitive optical parts, which need to be protected from
dust and other pollutions. Pollution of these parts can impair the accurate selection of
valve ports and therefore bias measurement results.
Always install a valve head for operation and storage. For protecting the actuator, a

dummy valve head (part of transportation lock kit part number: G1316-67001) can be
used instead of a functional valve. Do not touch parts inside the actuator.

If the Thermostatted Column Compartment SL Plus (G1316C SL Plus only)


includes the valve drive option, it is shipped with a transportation lock, which
needs to be removed during installation.
1 Remove the 5 screws, which hold the lock in position (G1316C SL Plus
only).

2 Remove the dummy valve head by unscrewing the cap nut and removing it
from the valve drive (G1316C SL Plus only).

1200 Series TCC User Manual

39

Installing the Column Compartment


Installing the Column Compartment

3 Place the column compartment in the stack or on the bench in a horizontal


position.
4 Ensure the power switch at the front of the column compartment is OFF.
HiVijh^cY^XVidg
\gZZc$nZaadl$gZY

A^cZedlZghl^iX]
l^i]\gZZca^\]i

Figure 11

Front View of the Thermostatted Column Compartment

5 Connect the power cable to the power connector at the rear of the column
compartment.
6 Connect the CAN cable to other Agilent 1200 Series modules.
7 If Agilent ChemStation is the controller, connect either

40

1200 Series TCC User Manual

Installing the Column Compartment


Installing the Column Compartment

the LAN connection to the LAN interface board in the module or


the GPIB cable to the module.

NOTE

If a Agilent DAD/MWD/FLD is in the system, the LAN/GPIB should be connected to the


DAD/MWD/FLD (due to higher data load).

8 Connect the APG Remote cable (optional) for non-Agilent 1200 Series
instruments.
9 Turn ON power by pushing the button at the lower left side of the column
compartment. The status LED should be green.
HZXjg^inaZkZg
8dc[^\jgVi^dchl^iX]

86C
6E<GZbdiZ
GH"'('8

Figure 12

NOTE

EdlZg

Rear View of the Thermostatted Column Compartment

The column compartment is turned on when the line power switch is pressed and the green
indicator lamp is illuminated. The column compartment is turned off when the line power
switch is protruding and the green light is OFF.

1200 Series TCC User Manual

41

Installing the Column Compartment


Installing Valve Heads (G1316C SL Plus)

Installing Valve Heads (G1316C SL Plus)

Several optional valve heads are available for the G1316C SL Plus, which can
be installed and exchanged easily.
Parts required

CAUTION

Part number

Description

5067-4107

8pos/9prt valve head high pressure and/or

5067-4108

8pos/9prt valve head low pressure

Valve Damage
Using a low pressure valve on the high pressure side can damage the valve.
When using multiple column compartments as part of a method development

solution, make sure that the high pressure valve head is connected to the
autosampler and the low pressure valve head is connected to the detector.
For details, please refer to the Method Development Solution User and Installation

Guide (part number: G4230-90000).

CAUTION

Column Damage or Bias Measurement Results


Switching the valve to a wrong position can damage the column or bias measurement
results.
Fitting the lobe to the groove is essential for making sure the valve is switched to

the correct position.

42

1200 Series TCC User Manual

Installing the Column Compartment


Installing Valve Heads (G1316C SL Plus)

1 Put the valve head onto the valve drive such that the lobe 2 Screw the valve head onto the valve drive using the union
fits to the groove.

nut.

3 Install all required capillary connections to the valve head. 4 Push the valve head until it snaps in and stays in the rear
position.

1200 Series TCC User Manual

43

Installing the Column Compartment


Flow Connections of the Column Compartment

Flow Connections of the Column Compartment

Parts required

Description
Other modules
Parts from accessory kit, see Accessory Kits on page 102
Two wrenches 1/4 5/16 inch for capillary connections

Preparations

WA R N I N G

Install the column compartment

Toxic and hazardous solvents


The handling of solvents and reagents can hold health risks.
When working with solvents observe appropriate safety procedures (for example,

goggles, safety gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when toxic or
hazardous solvents are used.

1 Press release buttons and remove front cover to gain


access to heater area.

2 The column compartment is equipped with an


column-identification system that can read column tags.

6ciZccVh

8dajbciV\

8dajbcXa^e

44

1200 Series TCC User Manual

Installing the Column Compartment


Flow Connections of the Column Compartment

NOTE

NOTE

For more information on column identification, see


Column-Identification System on page 13 .

The internal volumes of the heat exchanger assemblies


comprise a volume of 3 l (left) and 6 l (right). The internal
capillary diameter is 0.17 mm.

3 Place the column on the left heat exchanger assembly

4 Or place the column on the right heat exchanger

and connect the capillaries to the column.


;gdbVjidhVbeaZg

assembly and connect the capillaries to the column.

8a^e

IdYZiZXidg

NOTE

5 Fix the column with the column clip from the accessory
kit.

See Column Switching Valve (Optional for G1316A/G1316B


SL) on page 15.
on how to connect the column selection valve.

1200 Series TCC User Manual

45

Installing the Column Compartment


Flow Connections of the Column Compartment

6 If the column compartment is not part of a Agilent 1200


Series system, or if an Agilent 1200 Series Autosampler is
located on top, connect the corrugated tubing to the
waste outlet.

7 Route tubings from modules above through the openings


in the funnel holder (top) and the plastic bottom part.
Remove small plastic plugs first.

8 Put the front cover back in place.

The installation of the column compartment has now been completed.

NOTE

46

Always operated the TCC with the front cover in place for proper thermostatting conditions
and to protect the column area against strong drafts from the ouside.

1200 Series TCC User Manual

Installing the Column Compartment


Installation of Heater and Cooling Devices

Installation of Heater and Cooling Devices

Installation of Heater and Cooling Devices (G1316B SL)


With the introduction of the 1200 series TCC SL (G1316B SL), the heater
elements were redesigned in order to allow the adding of small heater and
cooling devices.

NOTE

Depending on the application, these heater and cooling devices can be fixed at various
places. Information about the usage of these heater and cooling devices can be found in
Technical Notes or in the Agilent 1200 Series Rapid Resolution LC System manual (part
number: G1312-90300).

]daZhidViiVX]XVgg^Zg

A"h]VeZYegZ"Xdajbc]ZViZg
KdajbZ/&#+a
bdjciZYdcXVgg^Zg
ide[dgXdajbc&!
Wdiidb[dgXdajbc'

Figure 13

NOTE

J"h]VeZYedhi"XdajbcXddaZg
KdajbZ/&#*a
bdjciZYdcXVgg^Zg

Installation points for heat exchanger / cooling devices

If the additional heat exchanger and cooling devices are used as shown in this figure, the
column identification system cannot be used. If the column identification system is
required, fix the heater and cooling devices in the upper or lower locations or right/left of
the current location.

1200 Series TCC User Manual

47

Installing the Column Compartment


Installation of Heater and Cooling Devices

Installation of Heater and Cooling Devices (G1316C SL Plus)


For the G1316C SL Plus, additional heat exchanger and cooling devices can be
installed on the carrier part number: G1316-89200 using 3 screws (part number:
0515-1052), included to part number for carrier) as shown in figure below.

7j^aY"^c]ZViZmX]Vc\Zg

Figure 14

48

Installing the low dispersion heat exchangers

1200 Series TCC User Manual

Installing the Column Compartment


Placing Columns

Placing Columns

Column-Identification Tag
When correctly placed on the heat exchanger, the distance between the
column-identification tag and antenna is 12 mm. This is the optimum
distance for proper function. The identification tag can be easily removed from
the column.

NOTE

For columns with small diameter, a cable tie wrap should be used to fix the column
identification tag to the column. Assure that the tie wrap does not block the front cover.

NOTE

The tag needs to be placed differently, depending on whether the column is installed at the
left or right heat exchanger, see Figure 15 on page 49 and Figure 16 on page 50. The Agilent
logo should always be at front.

8dajbc"^YZci^[^XVi^dciV\

Figure 15

1200 Series TCC User Manual

Column-Identification Tag for Left Heat Exchanger

49

Installing the Column Compartment


Placing Columns

8dajbc"^YZci^[^XVi^dciV\

Figure 16

Column-Identification Tag for Right Heat Exchanger

Column Clip
For better positioning of the column on the heat exchanger a column clip is
available (see Accessory Kits on page 102).

Figure 17

50

Column Clip (part number 5063-6526, pack of 6)

1200 Series TCC User Manual

1200 Series TCC User Manual

4
How to optimize the Column
Compartment
Optimizing the Performance of your Column Compartment
Using Additional Heater and Cooling Devices

52

53

This chapter provides information on how to optimize the Thermostatted


Column Compartement.

Agilent Technologies

51

How to optimize the Column Compartment


Optimizing the Performance of your Column Compartment

Optimizing the Performance of your Column Compartment

For best performance results of the column compartment:


Use short connection capillaries and place them close to the heat
exchanger. This will reduce heat dissipation and external band-broadening.
Use the left heat exchanger for small volume columns, for example, 23 mm
i.d. columns at flow rates of less than 200 l/min.
For even lower band-broadening, the heat exchanger can be by-passed and
the column is placed well between the heat exchanger fins.
Keep the left and right heat exchanger temperature the same unless you do
specific applications.
Assure that the front cover is always closed.

52

1200 Series TCC User Manual

How to optimize the Column Compartment


Using Additional Heater and Cooling Devices

Using Additional Heater and Cooling Devices

The optimization, installation, interconnection, and specific settings when


using additional heating and cooling devices are described in the Agilent 1200
Series Rapid Resolution LC System manual (G1312-90300).

1200 Series TCC User Manual

53

54

How to optimize the Column Compartment


Using Additional Heater and Cooling Devices

1200 Series TCC User Manual

1200 Series TCC User Manual

5
Troubleshooting and Diagnostics
Overview of the Column Departments Indicators and Test Functions

56

Status Indicators 57
Power Supply Indicator 57
Module Status Indicator 57
Available Tests depending on User Interfaces
Agilent Diagnostic Software

59

60

Overview about the troubleshooting and diagnostic features.

Agilent Technologies

55

Troubleshooting and Diagnostics


Overview of the Column Departments Indicators and Test Functions

Overview of the Column Departments Indicators and


Test Functions

Status Indicators
The instrument is provided with two status indicators which indicate the
operational state (prerun, run, and error states) of the instrument. The status
indicators provide a quick visual check of the operation of the instrument.

Error Messages
In the event of an electronic, mechanical or hydraulic failure, the instrument
generates an error message in the user interface. The following pages describe
the meaning of the error messages. For each message, a short description of
the failure, a list of probable causes, and a list of suggested actions to fix the
problem are provided.

Thermostat Diagnostic Test


The thermostat diagnostic test evaluates the heating and cooling efficiency of
the two peltier elements.

Temperature Calibration and Verification


The temperature calibration and verification procedure enables the
instrument temperature to be measured against an external, calibrated
measuring device. Normally, temperature calibration is not required
throughout the lifetime of the instrument. However, in order to comply with
local regulatory requirements, calibration and verification may be required.
The following sections describe these functions in detail.

56

1200 Series TCC User Manual

Troubleshooting and Diagnostics


Status Indicators

Status Indicators

Two status indicators are located on the front of the module. The lower left
indicates the power supply status, the upper right indicates the instrument
status.
HiVijh^cY^XVidg
\gZZc$nZaadl$gZY

A^cZedlZghl^iX]
l^i]\gZZca^\]i

Figure 18

Location of Status indicators

Power Supply Indicator


The power supply indicator is integrated into the main power switch. When
the indicator is illuminated (green) the power is ON.

Module Status Indicator


The module status indicator indicates one of four possible instrument
conditions:
When the status indicator is OFF (and power switch indicator is on), the
instrument is in a prerun condition, and is ready to begin an analysis.

1200 Series TCC User Manual

57

Troubleshooting and Diagnostics


Status Indicators
A green status indicator, indicates the module is performing an analysis
(run mode).
A yellow indicator indicates a not-ready condition. The module is in a
not-ready state when it is waiting for a specific condition or action to be
completed (for example, immediately after changing a setpoint), or while a
self-test procedure is running.
An error condition is indicated by a red status indicator. An error condition
indicates the module has detected an internal problem which affects
correct operation of the instrument. Usually, an error condition requires
attention (for example, leak, defective internal components). The error
state is propagated through the system to all connected modules, so the
error might come from a different module. Check the error log of your user
interface for the originating module. For safety reasons, an error condition
always interrupts the analysis.

58

1200 Series TCC User Manual

Troubleshooting and Diagnostics


Available Tests depending on User Interfaces

Available Tests depending on User Interfaces

NOTE

Depending on the used interface, the available tests and the screens/reports may vary.
Preferred tool should be the Agilent Lab Advisor Software, see Agilent Diagnostic
Software on page 22.
Screenshots used in this document are based on the Agilent ChemStation. In future, the
user interface may not show the Diagnostics/Tests anymore. In this case use the Agilent
Lab Monitor Diagnostic (LMD) Software.
The Agilent ChemStation may not include any maintenance/test functions.

Table 5

Test Functions available vs. User Interface - TCC

Test

Agilent ChemStation

Instant Pilot G4208A

LMD Software

Thermostat Function Test

Yes

No

Yes

Temperature Calibration

Yes

Yes 1

Yes

section Maintenance

1200 Series TCC User Manual

59

Troubleshooting and Diagnostics


Agilent Diagnostic Software

Agilent Diagnostic Software

The Agilent Diagnostic Software is a standalone product that can be used with
or without data system. Agilent Lab Advisor helps to manage the lab for high
quality chromatographic results and can monitor in real time a single Agilent
LC or all the Agilent GCs and LCs configured on the lab intranet.
Agilent Lab Advisor provides diagnostic capabilities for all Agilent 1200 Series
HPLC modules. This includes tests and calibrations procedures as well as the
different injector steps to perform all the maintenance routines.
Agilent Lab Advisor also allows users to monitor the status of their LC
instruments. The Early Maintenance Feedback (EMF) feature helps to carry
out preventive maintenance. In addition, users can generate a status report for
each individual LC instrument. The tests and diagnostic features as provided
by the Agilent Lab Advisor Software may differ from the descriptions in this
manual. For details refer to the Agilent Lab Advisor help files.
This manual provides lists with the names of Error Messages, Not Ready
messages, and other common issues.

60

1200 Series TCC User Manual

1200 Series TCC User Manual

6
Test Functions
Thermostat Function Test
Pressure Test

62

65

Column Thermostat Temperature Calibration 66


Column Thermostat Temperature Calibration Procedure
Column Thermostat Calibration Problems 69
Installing the Temperature Sensor 69

68

This chapter describes the TCCs built in test functions.

Agilent Technologies

61

Test Functions
Thermostat Function Test

Thermostat Function Test

Heater Function Test Description


The heater function test is used to evaluate the cooling and heating
performance of the two peltier elements.
When the test is started, both heat exchangers are cooled initially to 25 C.
This temperature is held for 12 seconds, and then the setpoint is changed to
20 C. The time required to reach 20 C is a measure of the cooling efficiency
of the peltier elements. At 3.5 minutes, the setpoint is changed to 30 C, and
both elements begin heating. The time required to reach 30 C is a measure of
heating efficiency.

Heater Function Test


Thermostat Function Test Result
A typical thermostat function test profile is shown in Figure 19 on page 63.

62

1200 Series TCC User Manual

Test Functions
Thermostat Function Test

IZbeZgVijgZP8R

G^\]iEZai^ZgZaZbZci
AZ[iEZai^ZgZaZbZci

I^bZPb^cjiZhR

Figure 19

Typical Thermostat Function Test Profile

Evaluating the Thermostat Function Test


During the cooling phase, the Peltier elements should cool at a rate of >2 C
minute. During the heating phase, the temperature change should be
>3 C/minute. Defective thermostat components may cause cooling or heating
rates to fall outside these limits.

1200 Series TCC User Manual

63

Test Functions
Thermostat Function Test

Function Test Failed


Probable Causes
Column compartment cover not installed correctly (bad insulation).
Air intake blocked (insufficient air flow for cooling).
Poor peltier efficiency (if setpoint temperatures can still be reached, and
are stable, there is no requirement to exchange the heater assembly).
Defective sensors on flex board.
Defective heater assembly.

Suggested Actions
Ensure cover is installed correctly.
Ensure sufficient space is available for air circulation see Bench Space on
page 26.
Exchange the heater assembly.

64

1200 Series TCC User Manual

Test Functions
Pressure Test

Pressure Test

For running a pressure test, please refer to the corresponding pump manual.
The pressure test may be used for testing the tightness of a valve installed in
the TCC.

CAUTION

Wrong use of pressure test may damage valve.


The current implementation of the pressure test automatically uses the maximum
pressure generated by the pump used by that system.
Do not use the test for modules having a lower maximum pressure than the pump

as this will damage the valve. For example do not use 400 bar valve in a TCC in
combination with a 600 bar pump.

1200 Series TCC User Manual

65

Test Functions
Column Thermostat Temperature Calibration

Column Thermostat Temperature Calibration

Temperature Calibration Principle


The actual temperatures of the column heat exchangers (left and right) depend
on the column setpoint temperature. For setpoint temperatures above (36 C),
the heat exchangers are heated to a temperature slightly above the setpoint
temperature. Conversely, for setpoint temperatures below (36 C), the heat
exchangers are kept at a temperature slightly below the setpoint temperature.
This fine temperature correction compensates for the small amount of heat
exchange through the instrument housing, and ensures the column is always
kept at the setpoint temperature.
At (36 C), the column setpoint and heat-exchanger temperatures are equal
(temperature cross-over point). This is the temperature at which a calibrated
measuring device can be used to calibrate the column thermostat.

IZbeZgVijgZ
Y^[[ZgZcXZP8R

iZbeZgVijgZVi
bZVhjg^c\ed^ci

8dajbchZied^ciiZbeZgVijgZ

8Va^WgVi^dcViXgdhh"dkZged^ci(+8

IZbeZgVijgZP8R

Figure 20

66

1-Point Calibration at the Temperature Cross-Over Point

1200 Series TCC User Manual

Test Functions
Column Thermostat Temperature Calibration

The column thermostat is calibrated correctly when the measured


temperature (using the external measuring device, Column Thermostat
Temperature Calibration Procedure on page 68) and the cross-over
temperature (36 C) of both heat exchangers (left and right) are within a range
of 0.5 C.

1200 Series TCC User Manual

67

Test Functions
Column Thermostat Temperature Calibration

Column Thermostat Temperature Calibration Procedure


Tools required

Temperature measuring device (see note below)

Parts required

Description
Calibrated temperature measuring device

NOTE

For the measuring and calibration process Agilent Technologies recommends a


thermometer with 0.1 C precision. Contact the local Agilent Technologies support
representative for ordering information.

NOTE

The figures in this procedure refer to a specific type of temperature sensor (Heraeus,
Quat340, quartz surface-temperature measurement sensor). Other sensors may require a
different fixing.

1 Install the temperature sensor (Installing the Temperature Sensor on


page 69).
2 Select the column-compartment temperature calibration mode in the user
interface.
3 Wait for the temperature to stabilize at the calibration temperature (36 C).
4 Measure the temperature of the heat exchanger.
5 If the measured temperature deviates by more than 0.5 C from the actual
temperature, enter the measured value in the measured-temperature field
for the left heat exchanger.
6 Install the sensor at the measurement point on the right heat exchanger.
Repeat the calibration procedure for the right heat exchanger.

NOTE

Limits
After calibration, the measured temperature and the calibration temperature should be
within 0.5 C. The maximum deviation which can be adjusted is 1.6 C. If the measured
value and the calibration value differ by more than 1.6 C, this is an indication that a
problem exists, Column Thermostat Calibration Problems on page 69.

68

1200 Series TCC User Manual

Test Functions
Column Thermostat Temperature Calibration

Column Thermostat Calibration Problems


If the temperature cannot be calibrated, check the following:
Has the thermostat front cover been closed correctly?
Is the measuring device functioning correctly, and is calibrated according to
the manufacturers instructions.

Hardware Failures
Probable hardware failures that can lead to a failed calibration procedure are:
Defective or wrongly calibrated measuring device.
Defective heater assembly.
Defective ambient-temperature sensor.
Defective CCM board.

Installing the Temperature Sensor


Installation of the temperature sensor is required for the temperature
calibration and temperature verification procedures.

NOTE

The figures below refer to a specific type of temperature sensor (Heraeus, Quat340, quartz
surface-temperature measurement sensor). Other sensors may require a different fixing.

1200 Series TCC User Manual

69

Test Functions
Column Thermostat Temperature Calibration

1 Remove the front cover.

2 Install the temperature sensor at the measurement


position on the left heat exchanger.

3 Route the sensor wire through the slit in the leak tray.

70

4 Re-install the front cover.

1200 Series TCC User Manual

1200 Series TCC User Manual

7
Maintenance
Introduction to Maintenance and Repair
Warnings and Cautions
Overview of Maintenance

72

73
75

Cleaning the Column Compartment


Changing Column Identification Tags

76
77

Replacing Head Parts of Column Switching Valve (G1316A/G1316B


SL) 79
Adding Heater and Cooling Devices (G1316B SL/G1316C SL Plus)
Correcting Leaks

82

84

Replacing the Column Compartments Firmware


Replacing Valve Heads (G1316C SL Plus)

85

86

Preparing the G1316C SL Plus for Transportation

89

This chapter describes the maintenance of the TCC.

Agilent Technologies

71

Maintenance
Introduction to Maintenance and Repair

Introduction to Maintenance and Repair

Simple Repairs
The column compartment is designed for easy repair. The most frequent
repairs such as change of column and column switching valve head parts can
be done from the front of the column compartment without removing the TCC
from the HPLC stack. These repairs are described in Maintenance on
page 71.

Exchanging Internal Parts


Some repairs may require exchange of defective internal parts. Exchange of
these parts requires removing the column compartment from the stack,
removing the covers, and disassembling the column compartment. The
security lever at the power input socket prevents that the column
compartment cover is taken off when line power is still connected.
These repairs are described in the Service Manual.

72

1200 Series TCC User Manual

Maintenance
Warnings and Cautions

Warnings and Cautions

WA R N I N G

Module is partially energized when switched off, as long as the power cord is
plugged in.
Risk of stroke and other personal injury. Repair work at the module can lead to
personal injuries, e. g. shock hazard, when the module cover is opened and the
instrument is connected to power.
Never perform any adjustment, maintenance or repair of the module with the top

cover removed and with the power cord plugged in.


The security lever at the power input socket prevents that the module cover is taken

off when line power is still connected. Never plug the power line back in when cover
is removed.

WA R N I N G

Sharp metal edges


Sharp-edged parts of the equipment may cause injuries.
To prevent personal injury, be careful when getting in contact with sharp metal

areas.

WA R N I N G

Toxic and hazardous solvents


The handling of solvents and reagents can hold health risks.
When working with solvents observe appropriate safety procedures (for example,

goggles, safety gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when toxic or
hazardous solvents are used.

1200 Series TCC User Manual

73

Maintenance
Warnings and Cautions

CAUTION

Electronic boards and components are sensitive to electrostatic discharge (ESD).


ESD can damage electronic boards and components.
In order to prevent damage always use ESD protection when handling electronic

boards and components.

CAUTION

Hot heat exchangers


The column compartment has two heat exchanger assemblies that might be hot.
Allow them to cool down before starting repairs.

74

1200 Series TCC User Manual

Maintenance
Overview of Maintenance

Overview of Maintenance

The following pages describe maintenance procedures (simple repairs) that


can be done without opening the main cover.
Table 6

Simple Repairs

Procedure

Typical Frequency

Changing Column
Identification Tags on page 77

When column performance or new


application requires a change

Replacing Head Parts of


Column Switching Valve
(G1316A/G1316B SL) on
page 79

If the valve performance shows


indication of leakage or wear

Correcting Leaks on page 84

If a leak has occurred

1200 Series TCC User Manual

Notes

Check for leaks

75

Maintenance
Cleaning the Column Compartment

Cleaning the Column Compartment

The column compartment case should be kept clean. Cleaning should be done
with a soft cloth slightly dampened with water or a solution of water and a
mild detergent. Make sure not to let liquid drip into your module.

WA R N I N G

Liquid dripping into the electronic compartment of your module.


Liquid in the module electronics can cause shock hazard and damage the module.
Do not use an exessively damp cloth during cleaning.
Drain all solvent lines before opening any fittings.

76

1200 Series TCC User Manual

Maintenance
Changing Column Identification Tags

Changing Column Identification Tags

The column compartment is equipped with an column-identification system,


that stores column specific information. Two identification antennas are
incorporated in the heat exchanger assemblies.

6ciZccVh

8dajbc"^YZci^[^XVi^dciV\

8dajbcXa^e

Figure 21

Column-Identification System

When

If column is used on the opposite heat exchanger or a tag is added to a new column.

Parts required

Part number

Description

5062-8588

Column identification tag, pack of 3

1 The identification tag can be easily removed from the column.


2 The tag needs to be placed differently, depending on whether the column is
installed at the left or right heat exchanger, see Figure 15 on page 49 and
Figure 16 on page 50. The Agilent logo should always be at front.
When correctly placed on the heat exchanger, the distance between tag and
antenna is 12 mm. This is the optimum distance for proper function.

1200 Series TCC User Manual

77

Maintenance
Changing Column Identification Tags

8dajbc"^YZci^[^XVi^dciV\

Figure 22

Column-Identification Tag for Left Heat Exchanger

8dajbc"^YZci^[^XVi^dciV\

Figure 23

Column-Identification Tag for Right Heat Exchanger

3 For columns with small diameter, a cable tie wrap should be used to fix the
column identification tag to the column. Assure that the tie wrap does not
block the front cover.

78

1200 Series TCC User Manual

Maintenance
Replacing Head Parts of Column Switching Valve (G1316A/G1316B SL)

Replacing Head Parts of Column Switching Valve


(G1316A/G1316B SL)

>hdaVi^dchZVa
GdidghZVa
HiVidgg^c\
HiVidg[VXZ
HiVidg]ZVY
Heg^c\h^YZ
h]dlhid
gZVg

HiVidghXgZlh

Figure 24

Column Switching Valve Parts

When

If valve leaks.

Tools required

1/4 inch wrench


9/64 inch hex key

Parts required

Description
For parts refer to Valve Options Overview on page 92.

1 Remove capillaries from ports 1, 5, and 6.

1200 Series TCC User Manual

79

Maintenance
Replacing Head Parts of Column Switching Valve (G1316A/G1316B SL)

2 Loosen each fixing stator screw two turns at a time. Remove bolts from
head.

3 Remove the stator head and the ceramic stator face.

NOTE

Valve Head, 8 Position/9 Port, High Pressure part number: 5067-4107 has no stator face.

4 Remove the stator ring.


5 Remove the rotor seal (and isolation seal if damaged or contaminated).
6 Install the new isolation seal (if required). Ensure the metal spring inside
the ring faces towards the valve body.
7 Install the new rotor seal.
8 Replace the stator ring. Ensure the stator ring is flush with the valve body.
9 Place the new (if required) ceramic stator face in place on the stator head.
Reinstall the stator head.

80

1200 Series TCC User Manual

Maintenance
Replacing Head Parts of Column Switching Valve (G1316A/G1316B SL)

NOTE

Valve Head, 8 Position/9 Port, High Pressure part number: 5067-4107 has no stator face.

10 Insert the stator screws in the stator head. Tighten the screws alternately
two turns at a time until the stator head is secure.
11 Reconnect the pump capillaries to the valve ports. Slide the waste tube into
the waste holder in the leak tray.

CAUTION

Wrong use of pressure test may damage valve.


The current implementation of the pressure test automatically uses the maximum
pressure generated by the pump used by that system.
Do not use the test for modules having a lower maximum pressure than the pump

as this will damage the valve. For example do not use 400 bar valve in a TCC in
combination with a 600 bar pump.

12 Perform a pressure-tightness test to ensure the valve is pressure tight to


400 bar.

1200 Series TCC User Manual

81

Maintenance
Adding Heater and Cooling Devices (G1316B SL/G1316C SL Plus)

Adding Heater and Cooling Devices


(G1316B SL/G1316C SL Plus)

The addtional heater and cooling devices can be arranged in the G1316B SL in
various locations depending on the application needs . Some examples are
shown below.

Figure 25

NOTE

Arrangements of Heater and Cooling Devices (G1316B SL/G1316C SL Plus)

If the additional heater and cooling devices are used as shown above, the column
identification system cannot be used. If the column identification system is required, fix the
heater and cooling devices in the upper or lower locations or fix them right/left of the
current location.

The heater and cooling devices are mounted on a carrier that can be fitted to
the left and/or right heat exchangers.

82

1200 Series TCC User Manual

Maintenance
Adding Heater and Cooling Devices (G1316B SL/G1316C SL Plus)

=ZViZmX]Vc\Zg

8Vgg^Zg[dg]ZViZgdgXddaZg

=ZViZgYZk^XZ

HXgZlYg^kZgIdgmdgEdh^Yg^k

Figure 26

Fixing Heater or Cooling Devices (G1316B SL)

For the heater device inlet capillary choose fittings which are compatible to
your column.
HlV\ZadX`XdbeVi^WaZXdajbch*%+*"))*)!e`d[&%!l^i][ZggjaZh
;^ii^c\hXgZladc\

HlV\ZadX`XdbeVi^WaZXdajbch!gZbdkVWaZ%&%%"'%-+
CjihZVai^\]i

Figure 27

1200 Series TCC User Manual

Fitting type depends on column type

83

Maintenance
Correcting Leaks

Correcting Leaks

When

If a leakage has occurred at the heat exchanger or at the capillary connections or at the column
switching valve.

Tools required

Tissue, pipette
Wrench 1/4 5/16 inch for capillary connections

NOTE

Depending on the column position or the use of additional heat-exchanger assemblies, the
view of Figure 28 on page 84 may vary.

1 Remove the front cover.


2 Use a pipette and tissue to dry the leak sensor area.
3 Observe the capillary connections and the column switching valve for leaks
and correct, if required.
4 Re-install the front cover.
8dajbchl^iX]^c\kVakZ
8dajbc

AZV`hZchdgVhhZbWan
LVhiZ djiaZi

Figure 28

84

Possible Leak Areas

1200 Series TCC User Manual

Maintenance
Replacing the Column Compartments Firmware

Replacing the Column Compartments Firmware

The installation of older firmware might be necessary:


to keep all systems on the same (validated) revision, or
if third part control software requires a special version.
To upgrade/downgrade the TCCs firmware the following steps have to be
performed:
When

If new version solves problems of currently installed version or after exchange of the TCC main board
(CCM) the version on board is older than previous installed one.

Tools required

LAN/RS-232 Firmware Update Tool, or Instant Pilot G4208A or Control Module G1323B

Parts required

Description
Firmware, tools and documentation from Agilent web site

Preparations

Read update documentation provided with the Firmware Update Tool.

1 Download the modules firmware, the LAN/RS-232 FW Update Tool Version


2.10 or above and the documentation from the Agilent web
http://www.chem.agilent.com/scripts/cag_firmware.asp.
2 Load the firmware into the TCC as described in the documentation.

NOTE

The G1316B SL requires firmware revision A.06.02 or higher (main and resident).
The G1316C SL Plus requires firmware revision A.06.10 or higher (main and resident).

1200 Series TCC User Manual

85

Maintenance
Replacing Valve Heads (G1316C SL Plus)

Replacing Valve Heads (G1316C SL Plus)

Several optional valve heads are available for the G1316C, which can be
installed and exchanged easily.
Parts required

CAUTION

Part number

Description

5067-4107

8pos/9prt valve head high pressure and/or

5067-4108

8pos/9prt valve head low pressure

Valve Damage
Using a low pressure valve on the high pressure side can damage the valve.
When using multiple column compartments as part of a method development

solution, make sure that the high pressure valve head is connected to the
autosampler and the low pressure valve head is connected to the detector.
For details, please refer to the Method Development Solution User and Installation

Guide (part number: G4230-90000).

CAUTION

Column Damage or Bias Measurement Results


Switching the valve to a wrong position can damage the column or bias measurement
results.
Fitting the lobe to the groove is essential for making sure the valve is switched to

the correct position.

CAUTION

Valve properties are read from the valve head RFID tag during initialization of the
module. Valve properties will not be updated, if the valve head is replaced while the
module is on.
Selection of valve port positions can fail, if the instrument does not know the
properties of the installed valve.
Always switch off the instrument when replacing the valve head.

86

1200 Series TCC User Manual

Maintenance
Replacing Valve Heads (G1316C SL Plus)

CAUTION

The valve actuator contains sensitive optical parts, which need to be protected from
dust and other pollutions. Pollution of these parts can impair the accurate selection of
valve ports and therefore bias measurement results.
Always install a valve head for operation and storage. For protecting the actuator, a

dummy valve head (part of transportation lock kit part number: G1316-67001) can be
used instead of a functional valve. Do not touch parts inside the actuator.

1 Switch off the module.

3 Unscrew the valve head.

2 Push the valve head for bringing it to its outer position


and unscrew all capillary connections from the valve
head.

1200 Series TCC User Manual

87

Maintenance
Replacing Valve Heads (G1316C SL Plus)

4 Put the new valve head onto the valve drive such that the 5 Screw the valve head onto the valve drive using the union
lobe fits to the groove.

nut.

6 Install all required capillary connections to the valve head. 7 Push the valve head until it snaps in and stays in the rear
position.

8 Switch on the module.

88

1200 Series TCC User Manual

Maintenance
Preparing the G1316C SL Plus for Transportation

Preparing the G1316C SL Plus for Transportation

When

If the Thermostatted Column Compartment SL Plus G1316C shall be transported

Tools required

Screwdriver Pozidriv 1 PT3

Parts required

Part number

Description

G1316-67001

Transportation Lock Kit

The module has been shipped with transportation locks, which must be used
for transportation protection. A transportation lock kit can be re-ordered
using
1 Remove the valve head as described in Replacing Valve
Heads (G1316C SL Plus) on page 86.

1200 Series TCC User Manual

2 Replace the valve head by the transportation valve head.


Bring the transportation valve head to the outer position.

89

Maintenance
Preparing the G1316C SL Plus for Transportation

3 Use 4 screws M3x8 (0515-0897) for fixing the lock plate


(G1316-03701) to the heat exchangers in the outer
positions of the plate.

90

4 Use a screw (2680-0128) for fixing the lock plate to the


transportation valve head (G1316-40002) in the central
position of the plate.

1200 Series TCC User Manual

1200 Series TCC User Manual

8
Parts and Materials for Maintenance
Valve Options Overview

92

Heater and Cooling Devices for G1316B SL/G1316C SL Plus

93

Column Switching Valve 2 Position/6 Port (G1316A/G1316B SL)


Column Switching Valve 8 Position-9 Port (G1316C SL Plus)

95

97

Micro Column Switching Valve 2 Postion/6 Port (G1316A/G1316B


SL) 99
Micro Column Switching Valve 2 Position/10 Port (G1316A/G1316B
SL) 100
Accessory Kits 102
G1316A Accessory Kit (Standard) 102
G1316A Accessory Kit (2PS/10PT modules) 103
G1316B SL/G1316C SL Plus Accessory Kit (Standard) 103
G1316B SL/G1316C SL Plus Capillary System Kit 104
G1316B SL Micro Valve Kit 2 Position/10 Port 106
Column Regeneration Kit (G1316A/G1316B SL) 107
Accessories (G1316C SL Plus) 108
Plastic Parts
Leak Parts

109
110

This chapter provides information on parts for maintenance.

Agilent Technologies

91

Parts and Materials for Maintenance


Valve Options Overview

Valve Options Overview

This overview gives a summary of the main parts and assemblies. More details
are available with each valve option in this chapter.
Table 7

Valves for G1316A and G1316B SL

Module

Valve Description

Valve

Rotor Seal

Stator

G1316A (#055)
G1316-68700

Column Switching Valve 2 Position/6 Port


(G1316A/G1316B SL) on page 95

0101-0920 1

0100-1855 (Vespel)
0100-1854 (Tefzel)
0100-2233 (PEEK)

0100-1851 (Face)
0100-1850 (Head)
0100-1852 (Seal)

G1316A (#056)

Micro Column Switching Valve 2 Postion/6


Port (G1316A/G1316B SL) on page 99

0101-1051

0100-2087 (Vespel)

0101-2089

G1316A (#057)
G1316-68709

Micro Column Switching Valve 2 Position/10


Port (G1316A/G1316B SL) on page 100

0101-1343 2

0101-1360

0101-1362

G1316B (#055)

Column Switching Valve 2 Position/6 Port


(G1316A/G1316B SL) on page 95 (600 bar)

0101-1420

0101-1409

0101-1417

G1316B (#057)

Micro Column Switching Valve 2 Position/10


Port (G1316A/G1316B SL) on page 100
(600 bar)

0101-1419

0101-1415

0101-1421

re-build kit 0101-1258 includes 3-groove rotor seal, stator face assy, isolation seal, instructions.

re-build kit 0101-1360 includes PEEK rotor seal, PEEK stator face, hex key.

Table 8

Valves for G1316C SL Plus

Kit

Kit Description

Valve Head

Rotor Seal

Stator

G4230A

Method Development Valve Kit, low


pressure, includes 8pos/9port valve

5067-4108

5067-41131

5067-4112

G4230B

Method Development Valve Kit, high


pressure, includes 8pos/9port valve

5067-4107

5067-4111

5067-4110

92

kit with stator face and rotor seal

1200 Series TCC User Manual

Parts and Materials for Maintenance


Heater and Cooling Devices for G1316B SL/G1316C SL Plus

Heater and Cooling Devices for G1316B SL/G1316C SL Plus

The use of these heater and cooling devices is described in Technical Notes or
in the Agilent 1200 Series Rapid Resolution LC System (1200 RRLC System)
manual.
Table 9

Heater and Cooling Devices for G1316B SL

Item

Description

Part Number

Heater long-up (0.12 mm i.d., 1.6 l internal volume)


(G1316B SL/G1316C SL Plus)

G1316-80002

Part ofG1316B SL/G1316C SL Plus Capillary System Kit on


page 104.
Heater long-down (0.12 mm i.d., 1.6 l internal volume)
(G1316B SL/G1316C SL Plus)

G1316-80003

Part ofG1316B SL/G1316C SL Plus Capillary System Kit on


page 104.

Post-column cooler, (0.12 mm i.d., 1.5 l internal volume)


(G1316B SL/G1316C SL Plus)

G1316-80004

Part ofG1316B SL/G1316C SL Plus Capillary System Kit on


page 104.
Carrier for Heater and Cooler
(G1316B SL),

G1316-83200

includes 3 screws SKT-HD-CAP, M3 x 0.5, 12 mm long

0515-1052

Part of G1316B SL/G1316C SL Plus Capillary System Kit on


page 104.

1200 Series TCC User Manual

93

Parts and Materials for Maintenance


Heater and Cooling Devices for G1316B SL/G1316C SL Plus
Table 10

Consumables (G1316C SL Plus)

Description

Part Number

Column Clip Set, 8 Colors

5042-9918

Carrier for Heat Exchanger G1316C SL Plus

G1316-89200

Fitting Holder Assy, includes following items:


Fitting Fork
Fitting Clip
Screws (Pack of 4)

G1316-68706

A"h]VeZYegZ"Xdajbc]ZViZg
KdajbZ/&#+a
bdjciZYdcXVgg^Zg
ide[dgXdajbc&!Wdiidb[dg
Xdajbc'

J"h]VeZYEdhi"XdajbcXddaZg
KdajbZ/&#*a
bdjciZYdc8Vgg^Zg

]daZhidViiVX]XVgg^Zg
Figure 29

94

Heater and Cooling Devices for G1316B SL

1200 Series TCC User Manual

Parts and Materials for Maintenance


Column Switching Valve 2 Position/6 Port (G1316A/G1316B SL)

Column Switching Valve 2 Position/6 Port (G1316A/G1316B SL)

Table 11
Item

Column Switching Valve 2PS/6PT for G1316A/G1316B SL

Description

Part Number

Column switching valve kit, includes all parts required for


installation

G1353-68700

Column switching valve (complete assembly)

0101-0920

Cover plate (when switching valve is not installed)

G1316-44103

Capillary Kit Column Switching includes two capillaries (0.17mm


i.d., 180 mm) and three capillaries (0.17 mm i.d., 90 mm)

G1316-68708
G1313-87305
G1316-87300

Rhebuild kit for 7750-030 valve, includes: 3-groove rotor seal,


stator face assy, isolation seal, instructions

0101-1258

Stator screws

1535-4857

Stator Head

0100-1850

Stator face

0100-1851

Stator ring

Rotor seal 3 grooves (Tefzel)


Rotor seal 3 grooves (Vespel)
Rotor seal 3 grooves (PEEK)

0100-1854
0100-1855
0100-2233

Isolation seal

0100-1852

1200 Series TCC User Manual

95

Parts and Materials for Maintenance


Column Switching Valve 2 Position/6 Port (G1316A/G1316B SL)

+
*
)
(
'
&

Figure 30

96

Heg^c\h^YZ
h]dlhid
gZVg

Column Switching Valve Parts

1200 Series TCC User Manual

Parts and Materials for Maintenance


Column Switching Valve 8 Position-9 Port (G1316C SL Plus)

Column Switching Valve 8 Position-9 Port (G1316C SL Plus)

Column Switching Valve Parts High Pressure (G1316C SL Plus)


Table 12
Item

Valve 8 Postion/9 Port, High Pressure (G1316C SL Plus)


Description

Part Number

Valve Head, 8 Position/9 Port, High Pressure

5067-4107

Stator Screws

1535-4857

Stator Head

5067-4110

Rotor Seal High Pressure

5067-4111

Isolation Seal

0100-1852

)
(

'
&

Figure 31

1200 Series TCC User Manual

Column Switching Valve Parts (5067-4107)

97

Parts and Materials for Maintenance


Column Switching Valve 8 Position-9 Port (G1316C SL Plus)

Column Switching Valve Parts 400 bar (G1316C SL Plus)


Table 13
Item

Valve 8 Postion/9 Port, 400 bar (G1316C SL Plus)


Description

Part Number

Valve Head, 8 Position/9 Port, 400 bar

5067-4108

Stator Screws

1535-4857

Stator Head

5067-4112

3, 4

Stator Face/Rotor Seal 400 bar

5067-4113

Isolation Seal

0100-1852

*
)
(

'
&

Figure 32

98

Column Switching Valve Parts (5067-4108)

1200 Series TCC User Manual

Parts and Materials for Maintenance


Micro Column Switching Valve 2 Postion/6 Port (G1316A/G1316B SL)

Micro Column Switching Valve 2 Postion/6 Port


(G1316A/G1316B SL)

Table 14
Item

Description

Part Number

Micro Column Switching Valve 2PS/6PT, 400 bar

0101-1051

Stator face

0100-2089

Rotor seal 3 grooves

0100-2087

Table 15
Item

Micro Column Switching Valve 2PS/6PT for G1316A/B, 400 bar

Micro Column Switching Valve 2PS/6PT for G1316B SL, 600 bar

Description

Part Number

Micro Column Switching Valve 2PS/6PT, 600 bar

0101-1420

Stator, 600 bar

0101-1417

Rotor seal 3 grooves, 600 bar

0101-1409

1200 Series TCC User Manual

99

Parts and Materials for Maintenance


Micro Column Switching Valve 2 Position/10 Port (G1316A/G1316B SL)

Micro Column Switching Valve 2 Position/10 Port


(G1316A/G1316B SL)

NOTE

Technical details can be found in the Technical Note that is supplied with the kit.

Table 16
Item

Description

Part Number

2PS/10 PT Valve kit, includes all parts required for installation

G1316-68709

Rebuild kit, includes PEEK rotor seal, PEEK stator face, hex key

0101-1360

2PS/10 PT valve

0101-1343

Rotor seal (Vespel)

0101-1361

Stator face

0101-1362

Capillary kit, see Table 18 on page 100

G1316-68711

Table 17
Item

Table 18
From

Micro Column Switching Valve 2PS/10PT for G1316A/B, 400 bar

Micro Column Switching Valve 2PS/10PT 600 bar for G1316B SL, 600 bar

Description

Part Number

Column Switching Valve 2PS/10PT for -LC System, 600 bar

0101-1419

Stator, 600 bar

0101-1421

Rotor seal 5 grooves, 600 bar

0101-1415

Capillary Kit (G1316-68711)


To

ID [mm]

Length [mm]

Qty

Part number

Valve (port 2)

0.17

700

5065-9932

Remark

Capillaries
ALS 1

100

1200 Series TCC User Manual

Parts and Materials for Maintenance


Micro Column Switching Valve 2 Position/10 Port (G1316A/G1316B SL)
Table 18

Capillary Kit (G1316-68711)

From

To

ID [mm]

Length [mm]

Qty

Part number

Remark

Valve (port 3)

TCC 3 l (In)

0.17

105

5021-1816

TCC 2 3 l (Out)

Column 1

0.17

105

5021-1816

Column 1

Valve (port 6)

0.17

105

5021-1816

for long column

Column 1

Valve (port 6)

0.17

200

5065-9931

for short column

Valve (port 7)

Detector (In)

0.17

280

5021-1818

Valve (port 1)

TCC 6 l (In)

0.17

105

5021-1816

TCC ** 6 l (Out)

Column 2

0.17

105

5021-1816

Column 2

Valve (port 8)

0.17

105

5021-1816

for long column

Column 2

Valve (port 8)

0.17

200

5065-9931

for short column

Valve (port 5)

Valve (port 10)

0.17

105

5021-1816

Regeneration pump

Valve (port 4)

0.25

800

5065-9930

Valve (port 9)

Waste

0.6

2000

5062-2463

PTFE

Ferrules, screws, fingertight fittings, etc.


1/16 fittings and ferrules

5062-2418

10/pk

fingertight fitting long

5062-8541

10/pk

fitting screw long

10

5065-4454

10/pk

fitting screw extra long

10

5065-9967

10/pk

front ferrule

5180-4108

10/pk

back ferrule

5180-4114

10/pk

0890-1763

Plastic tubing cutter

8710-1930

Hex key 3/32

8710-2462

Rheotool Socket wrench 2 x 1/4

8710-2391

Peek tubing 1/16

0.18

1500

ALS - Autosampler

TCC - Thermostatted Column Compartment (heat exchanger: 3 l left or 6 l right)

1200 Series TCC User Manual

101

Parts and Materials for Maintenance


Accessory Kits

Accessory Kits

The accessory kits (for G1316A, G1316B SL or G1316C SL Plus) contain


accessories and tools needed for the installation and maintenance.

G1316A Accessory Kit (Standard)


Table 19
Item

102

G1316A Accessory Kit (Standard)

Description

Part Number

G1316A Accessory Kit (Standard)

G1316-68705

Column identification tag (blank)


for re-ordering use (pack of 3)

5062-8588

Column clip, qty=2, for re-order use (pack of 6)

5063-6526

Corrugated tubing (to waste), re-order 5 m

5062-2463

CAN cable 0.5 m

5181-1516

Wrench open end 1/4 5/16 inch

8710-0510

Wrench open end, 5/16 3/8 inch

8710-2409

Capillary column-heat exchanger 90 mm lg, 0.17 i.d. (not


assembled) contains items 2, 3 and 4

G1316-87300

Ferrule front SST, qty=2

Ferrule back SST, qty=2

Fitting SST, qty=2

re-order 5062-2418, pack of 10 each of fittings, front- and back ferrule

1200 Series TCC User Manual

Parts and Materials for Maintenance


Accessory Kits

G1316A Accessory Kit (2PS/10PT modules)


Table 20
Item

G1316A Accessory Kit (2PS/10PT modules)

Description

Part Number

G1316A Accessory Kit 2PS/10PT

G1316-68725

Column identification tag (blank)


for re-ordering use (pack of 3)

5062-8588

Column clip, qty=2, for re-order use (pack of 6)

5063-6526

Corrugated tubing (to waste), re-order 5 m

5062-2463

CAN cable 0.5 m

5181-1516

Wrench open end 1/4 5/16 inch

8710-0510

Wrench open end, 5/16 3/8 inch

8710-2409

PEEK Capillary 280 mm lg, 50 i.d., qty=4

G1316-87309

MIC Valve Fitting, qty=2

5022-2186

Column holder for -LC columns, qty=2

5001-3702

G1316B SL/G1316C SL Plus Accessory Kit (Standard)


Table 21
Item

Accessory Kit (Standard) G1316B SL and G1316C SL Plus

Description

Part Number

G1316B SL/G1316C SL Plus Accessory Kit (Standard)

G1316-68735

Column identification tag (blank)


for re-ordering use (pack of 3)

5062-8588

Column clip, qty=2, for re-order use (pack of 6)

5063-6526

Corrugated tubing (to waste), re-order 5 m

5062-2463

CAN cable 0.5 m

5181-1516

Wrench open end 1/4 5/16 inch

8710-0510

Wrench open end, 5/16 3/8 inch

8710-2409

1200 Series TCC User Manual

103

Parts and Materials for Maintenance


Accessory Kits
Table 21
Item

Accessory Kit (Standard) G1316B SL and G1316C SL Plus

Description

Part Number

Screwdriver Torx TX8

8710-2509

Screwdriver Hexagonal 2.5 mm

5965-0028

Capillary column-heat exchanger 90 mm lg, 0.17 i.d. (not


assembled) contains items 2, 3 and 4

G1316-87300

Capillary column-heat exchanger 115 mm lg, 0.17 i.d. (not


assembled) contains items 2, 3 and 4

G1316-87321

Capillary column-heat exchanger 170 mm lg, 0.17 i.d. (not


assembled) contains items 2, 3 and 4

G1316-87323

Ferrule front SST, qty=2

Ferrule back SST, qty=2

Fitting SST, qty=2

re-order 5062-2418, pack of 10 each of fittings, front- and back ferrule

G1316B SL/G1316C SL Plus Capillary System Kit


G1316B SL/G1316C SL Plus Capillary System Kit
Table 22

G1316B SL Capillary System Kit G1316-68744 (G1316B#060)

Item

Description

Part Number

2 Carriers for Heater or Cooling Devices

G1316-83200

Heater long-up (0.12 mm i.d., 1.6 l internal volume)

G1316-80002

Heater long-down (0.12 mm i.d., 1.6 l internal volume)

G1316-80003

Post-column cooler (0.12 mm i.d., 1.5 l internal volume)

G1316-80004

Capillary System Kit, see Table 23 on page 105 for details

G1316-68716

For items * see also Heater and Cooling Devices for G1316B SL/G1316C SL
Plus on page 93.

104

1200 Series TCC User Manual

Parts and Materials for Maintenance


Accessory Kits
Table 23
Item

Capillary System Kit G1316-68716

Description

Part Number

Seat Capillary 100 mm x 0.12 mm, 0.8 OD

G1367-87303

DAD Heat Exchanger Capillary 310 mm x 0.12 mm

G1315-87339

SST Capillary 340 mm x 0.12 mm, m/m

G1316-87319

SST Capillary 300 mm x 0.12 mm, m/m

G1316-87318

SST Capillary 210 mm x 0.12 mm, m/m

G1316-87317

SST Capillary 170 mm x 0.12 mm, m/m

G1316-87316

SST Capillary 130 mm x 0.12 mm, m/f

G1316-87315

SST Capillary 90 mm x 0.12 mm, m/f

G1316-87314

SST Capillary 70 mm x 0.12 mm, m/f

G1316-87313

SST Capillary 50 mm x 0.12 mm, m/f

G1316-87312

SST Capillary 170 mm x 0.12 mm, m/f

G1316-87327

SST Capillary 500 mm x 0.12 mm, m/m

G1316-87309

SST Capillary 500 mm x 0.12 mm, m/m

G1315-87307

G1316C SL Plus Capillary System Kits


Table 24

G1316C Capillary System Kits

Description1

Part Number

Solvent selection tubing kit, 4 solvents

5067-4601

Method Development Capillary Kit, low dispersion, short column

5067-1595

Method Development Capillary Kit, low dispersion, long column

5067-1596

Method Development Capillary Kit, general purpose

5067-1597

RRHT Selectivity Method Development Kit, 2.1 mm ID

5190-1431

RRHT pH Method Development Kit, 2.1 mm ID

5190-1432

RRHT Selectivity Method Development Kit, 4.6 mm ID

5190-1433

RRHT pH Method Development Kit, 4.6 mm ID

5190-1434

1200 Series TCC User Manual

105

Parts and Materials for Maintenance


Accessory Kits
Table 24

G1316C Capillary System Kits

Description1

Part Number

Rapid Resolution Selectivity Method Development Kit

5190-1435

Rapid Resolution pH Method Development Kit

5190-1436

for capillary kit contents, please refer to the Method Development Solution User and
Installation Guide (G4230-90000)

G1316B SL Micro Valve Kit 2 Position/10 Port


Table 25
Item

106

G1316B SL Micro Valve Kit 2PS/10PT

Description

Part Number

G1316B SL Micro Valve Kit 2PS/10PT

G1316-68745

Column identification tag (blank)


for re-ordering use (pack of 3)

5062-8588

Column clip, for re-order use (is pack of 6)

5063-6526

Corrugated tubing (to waste), re-order 5 m

5062-2463

Wrench open end 1/4 5/16 inch

8710-0510

Wrench open end, 5/16 3/8 inch

8710-2409

Screwdriver Torx TX8

8710-2509

Screwdriver Hexagonal 2.5 mm

5965-0028

2 Column holders for -LC columns

5001-3702

Column Regeneration Kit for -LC columns, see Table 26 on


page 107

G1316-68721

Capillary column-heat exchanger 90 mm lg, 0.17 i.d. (not


assembled) contains items 2, 3 and 4

G1316-87300

Capillary column-heat exchanger 115 mm lg, 0.17 i.d. (not


assembled) contains items 2, 3 and 4

G1316-87321

Capillary column-heat exchanger 170 mm lg, 0.17 i.d. (not


assembled) contains items 2, 3 and 4

G1316-87323

1200 Series TCC User Manual

Parts and Materials for Maintenance


Accessory Kits

Column Regeneration Kit (G1316A/G1316B SL)


Refer to Figure 33 on page 108 for connection diagram.
Table 26

Column Regeneration Kit

Description

where used

Column Regeneration Kit

Part Number
G1316-68721

SST Capillary, 700 mm x 0.17 mm, 1/32 - 1/32

column to cell

G1312-87304

SST Capillary, 100 mm x 0.12 mm, 1/32 - 1/32

switching capillary

G1316-27301

SST Capillary, 100 mm x 0.12 mm, male/female 1/32 - 1/16

adapter capillary

G1316-87304

SST Capillary, 340 mm x 0.12 mm, male/female 1/32 - 1/16

WPS to valve

G1316-87305

SST Capillary, 70 mm x 0.12 mm, male/female 1/32 - 1/16, qty=2

valve to heatexchanger

G1316-87306

SST Capillary 50 mm x 0.12 mm, male/female

column to cell

G1316-87312

SST Capillary 70 mm x 0.12 mm, male/female

column to cell

G1316-87313

SST Capillary, 75 mm x 0.12 mm, male/female 1/32 - 1/16

valve to detector

G1316-87326

Seat Capillary, 100 mm x 0.12 mm, qty=2

G1367-87303

PEEK fitting, special for Chip-LC

G4240-43200

Flexible PEEK Tubing, 450 mm x 0.4 mm, =1

1200 Series TCC User Manual

valve to waste

5022-6503

107

Parts and Materials for Maintenance


Accessory Kits

Figure 33

Connection Diagram for Column Regeneration

Accessories (G1316C SL Plus)


Table 27

108

Accessories (G1316C SL Plus)

Description

Part Number

Transportation Lock Kit G1316C SL Plus, includes following items:


Lock Plate
Screw M4
Screw M3x8 (pack of 4)
Spring Washer
Transportation Valve Head

G1316-67001

1200 Series TCC User Manual

Parts and Materials for Maintenance


Plastic Parts

Plastic Parts

Table 28

Plastic Parts

Item

Description

Part Number

Front cover G1316A (1200 Series)

G1316-68714

Front cover G1316B SL (1200 Series)

G1316-68724

Front cover G1316C SL Plus (1200 Series)

G1316-68754

Name plate Agilent (1200 Series)

5042-8901

'

Figure 34

1200 Series TCC User Manual

Plastic Parts

109

Parts and Materials for Maintenance


Leak Parts

Leak Parts

Table 29

Leak Parts G1316A/G1316B SL

Item

Description

Part Number

Leak funnel

5041-8388

Leak funnel holder

G1316-42300

Waste assembly, includes complete Y-tubing assembly with leak


funnel

G1316-60002

Corrugated waste tube (reorder pack), 5 m

5062-2463

&
'
(

Figure 35

110

Leak Parts

1200 Series TCC User Manual

Parts and Materials for Maintenance


Leak Parts
Table 30

Leak Parts G1316C SL Plus

Item

Desription

Part Number

Leak Panel Kit, includes following items:


Leak Panel Top
Leak Panel Base

G1316-68722

Leak Tube Kit, includes following items:


Funnel Holder G1316C SL Plus
Tubing-flex polyethylene
Leak Funnel

G1316-67000

Capillary Guide

G1316-42303

&

Figure 36

1200 Series TCC User Manual

Capillary Guide

111

112

Parts and Materials for Maintenance


Leak Parts

1200 Series TCC User Manual

1200 Series TCC User Manual

9
Appendix
Safety Symbols

114

The Waste Electrical and Electronic Equipment Directive


Lithium Batteries Information
Radio Interference
Sound Emission

117

118

119
120

Solvent Information

121

Agilent Technologies on Internet

123

This chapter provides addition information on safety, legal and web.

Agilent Technologies

113

Appendix
Safety Symbols

Safety Symbols

Safety Symbols
Table 31

Safety Symbols

Symbol

Description
The apparatus is marked with this symbol when the user should refer to the instruction manual
in order to protect risk of harm to the operator and to protect the apparatus against damage.
Indicates dangerous voltages.
Indicates a protected ground terminal.
Indicates eye damage may result from directly viewing the light produced by the deuterium lamp
used in this product.
The apparatus is marked with this symbol when hot surfaces are available and the user should
not touch it when heated up.

WA R N I N G

A WARNING
alerts you to situations that could cause physical injury or death.
Do not proceed beyond a warning until you have fully understood and met the

indicated conditions.

CAUTION

A CAUTION
alerts you to situations that could cause loss of data, or damage of equipment.
Do not proceed beyond a caution until you have fully understood and met the

indicated conditions.

114

1200 Series TCC User Manual

Appendix
Safety Symbols

General Safety Information


The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Agilent
Technologies assumes no liability for the customers failure to comply with
these requirements.

WA R N I N G

Ensure the proper usage of the equipment.


The protection provided by the equipment may be impaired.
The operator of this instrument is advised to use the equipment in a manner as

specified in this manual.

General
This is a Safety Class I instrument (provided with terminal for protective
earthing) and has been manufactured and tested according to international
safety standards.

Operation
Before applying power, comply with the installation section. Additionally the
following must be observed.
Do not remove instrument covers when operating. Before the instrument is
switched on, all protective earth terminals, extension cords,
auto-transformers, and devices connected to it must be connected to a
protective earth via a ground socket. Any interruption of the protective earth
grounding will cause a potential shock hazard that could result in serious
personal injury. Whenever it is likely that the protection has been impaired,
the instrument must be made inoperative and be secured against any intended
operation.
Make sure that only fuses with the required rated current and of the specified
type (normal blow, time delay, and so on) are used for replacement. The use of
repaired fuses and the short-circuiting of fuse holders must be avoided.

1200 Series TCC User Manual

115

Appendix
Safety Symbols
Some adjustments described in the manual, are made with power supplied to
the instrument, and protective covers removed. Energy available at many
points may, if contacted, result in personal injury.
Any adjustment, maintenance, and repair of the opened instrument under
voltage should be avoided whenever possible. When inevitable, this has to be
carried out by a skilled person who is aware of the hazard involved. Do not
attempt internal service or adjustment unless another person, capable of
rendering first aid and resuscitation, is present. Do not replace components
with power cable connected.
Do not operate the instrument in the presence of flammable gases or fumes.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
Do not install substitute parts or make any unauthorized modification to the
instrument.
Capacitors inside the instrument may still be charged, even though the
instrument has been disconnected from its source of supply. Dangerous
voltages, capable of causing serious personal injury, are present in this
instrument. Use extreme caution when handling, testing and adjusting.
When working with solvents please observe appropriate safety procedures
(e.g. goggles, safety gloves and protective clothing) as described in the material
handling and safety data sheet by the solvent vendor, especially when toxic or
hazardous solvents are used.

116

1200 Series TCC User Manual

Appendix
The Waste Electrical and Electronic Equipment Directive

The Waste Electrical and Electronic Equipment Directive

Abstract
The Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC), adopted by EU Commission on 13 February 2003, is
introducing producer responsibility on all electric and electronic appliances
starting with 13 August 2005.

NOTE

This product complies with the WEEE Directive (2002/96/EC) marking requirements. The
affixed label indicates that you must not discard this electrical/electronic product in
domestic household waste.
Product Category:
With reference to the equipment types in the WEEE Directive Annex I, this product is
classed as a Monitoring and Control Instrumentation product.

NOTE

Do not dispose off in domestic household waste


To return unwanted products, contact your local Agilent office, or see
www.agilent.com for more information.

1200 Series TCC User Manual

117

Appendix
Lithium Batteries Information

Lithium Batteries Information

WA R N I N G

Lithium batteries may not be disposed-off into the domestic waste. Transportation of
discharged Lithium batteries through carriers regulated by IATA/ICAO, ADR, RID,
IMDG is not allowed.
Danger of explosion if battery is incorrectly replaced.
Discharged Lithium batteries shall be disposed off locally according to national

waste disposal regulations for batteries.


Replace only with the same or equivalent type recommended by the equipment

manufacturer.

WA R N I N G

Lithiumbatteri - Eksplosionsfare ved fejlagtig hndtering.


Udskiftning m kun ske med batteri af samme fabrikat og type.
Lever det brugte batteri tilbage til leverandren.

WA R N I N G

Lithiumbatteri - Eksplosionsfare.
Ved udskiftning benyttes kun batteri som anbefalt av apparatfabrikanten.
Brukt batteri returneres appararleverandoren.

NOTE

118

Bij dit apparaat zijn batterijen geleverd. Wanneer deze leeg zijn, moet u ze niet weggooien
maar inleveren als KCA.

1200 Series TCC User Manual

Appendix
Radio Interference

Radio Interference

Cables supplied by Agilent Technoligies are screened to provide opitimized


protection against radio interference. All cables are in compliance with safety
or EMC regulations.

Test and Measurement


If test and measurement equipment is operated with unscreened cables, or
used for measurements on open set-ups, the user has to assure that under
operating conditions the radio interference limits are still met within the
premises.

1200 Series TCC User Manual

119

Appendix
Sound Emission

Sound Emission

Manufacturers Declaration
This statement is provided to comply with the requirements of the German
Sound Emission Directive of 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB.
Sound Pressure Lp < 70 dB (A)
At Operator Position
Normal Operation
According to ISO 7779:1988/EN 27779/1991 (Type Test)

120

1200 Series TCC User Manual

Appendix
Solvent Information

Solvent Information

Flow Cell
To protect optimal functionality of your flow-cell:
Avoid the use of alkaline solutions (pH > 9.5) which can attack quartz and
thus impair the optical properties of the flow cell.
If the flow cell is transported while temperatures are below 5 degree C, it
must be assured that the cell is filled with alcohol.
Aqueous solvents in the flow cell can built up algae. Therefore do not leave
aqueous solvents sitting in the flow cell. Add a small % of organic solvents
(e.g. acetonitrile or methanol ~5%).

Use of Solvents
Observe the following recommendations on the use of solvents.
Brown glass ware can avoid growth of algae.
Small particles can permanently block capillaries and valves. Therefore
always filter solvents through 0.4 m filters.
Avoid the use of the following steel-corrosive solvents:
Solutions of alkali halides and their respective acids (for example,
lithium iodide, potassium chloride, and so on),
High concentrations of inorganic acids like sulfuric acid and nitric acid,
especially at higher temperatures (if your chromatography method
allows, replace by phosphoric acid or phosphate buffer which are less
corrosive against stainless steel),
Halogenated solvents or mixtures which form radicals and/or acids, for
example:
2CHCl3 + O2 > 2COCl2 + 2HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs
quickly with dried chloroform if the drying process removes the
stabilizing alcohol,

1200 Series TCC User Manual

121

Appendix
Solvent Information
Chromatographic grade ethers, which can contain peroxides (for
example, THF, dioxane, di-isopropylether) such ethers should be filtered
through dry aluminium oxide which adsorbs the peroxides,
Solvents containing strong complexing agents (e.g. EDTA),
Mixtures of carbon tetrachloride with 2-propanol or THF.

122

1200 Series TCC User Manual

Appendix
Agilent Technologies on Internet

Agilent Technologies on Internet

For the latest information on products and services visit our worldwide web
site on the Internet at:
http://www.agilent.com
Select Products/Chemical Analysis
It will provide also the latest firmware of the Agilent 1200 Series modules for
download.

1200 Series TCC User Manual

123

Index

Index

Agilent
Diagnostic Software 22, 60
Lab Advisor Software 22, 60
on internet 123
algae 121, 121
ambient operating temperature 27
ambient non-operating temperature 27

dead volume 28
delivery checklist 34
Diagnostic
software 22, 60
dimensions 27

heaterexchanger
pre-column 30
heating concept 11
humidity 27

B
battery
safety information
bench space 26

118

C
calibration
temperature 56, 66
column switching valve (optional)
description 15
precolumn backflushing 17
column switching valve (optional)
two column selection 16
column clip 50
column identification 13
tag 49
column
changing column and tags 77
cool-down time 28
cooling concept 11
cooling device
post column 30

I
E
electrical connections
descriptions of 18
electronic waste 117
electrostatic discharge (ESD)
environment 26
error messages 56
external measuring device
information 68

74

F
features
GLP 28, 29
instrument layout 21
safety and maintenace 28
flow cell 121
solvent information 121
frequency range 27
function test
failed 64
result 62

G
GLP features

28

installation
bench space 26
capillaries and waste tubings 46
column 45, 45
flow connections 44
module 38
power cords 25
temperature sensor 69
unpacking 34
instrument layout 21
internet 123
introduction
concept of heating and cooling 11
system overview 11

L
Lab Advisor
software 22, 60
leaks, correcting 84
line frequency 27
line voltage 27
lithium batteries 118

N
non-operating altitude 27
non-operating temperature 27

124

1200 Series TCC User Manual

Index

operating Altitude 27
operating temperature 27
optimization 51
optimizing the performance 52

safety class I 115


safety information
lithium batteries 118
safety
general information 115
standards 27
symbols 114, 114
serial number
information 20
site requirements
environment 26
solvent information 121
solvents 121
sound emission 120
specification
physical 27
specifications 28
stack configuration 35
status indicators 56, 57
system overview 11

P
parts identification
leak panels 110
parts identification
plastics 102
performance
Optimization 51
optimizing 52
specifications and features
physical specifications 27
physical specifications 27
post-column
cooling device 30
power consideration 24
power consumption 27
power cords 25
pre-column
heatexchanger 30

28

R
radio interference 119
repairs
changing the column 77
cleaning the instrument 76
column switching valve parts 79
correcting leaks 84
introduction 72
overview 75
types simple/internal 72
warnings and cautions 72

1200 Series TCC User Manual

T
tag

thermostat function test result 62


thermostat function test 56
troubleshooting
available tests vs interface 59
error messages 56
status indicators 56, 57

U
unpacking 34
User Interfaces
diagnosic tests

59

V
verification
temperature 56
voltage range 27

W
warm-up time 28
waste
electrical and electronic
equipment 117
WEEE directive 117
weight 27

installation 49
temperature verification
principle 69
temperature calibration
description 66
problems 69
procedure 68
temperature verification
external measuring device 68
temperature
range 28
tests
available tests vs interface 59
test
thermostat function test failed 64

125

www.agilent.com

In This Book

This manual contains technical reference


information about the Agilent 1200 Series
thermostatted column compartment.
The manual describes the following:
introduction and specifications,
installation,
using and optimizing,
troubleshooting and diagnose,
maintenance and repair,
parts identification,
hardware information,
safety and related information.

Agilent Technologies 1995, 1996-2008

Printed in Germany
09/2008

*G1316-90011*
*G1316-90011*
G1316-90011

Agilent Technologies

You might also like