2016 ECCMID Poster CMV Validation ELITe Ingenius
2016 ECCMID Poster CMV Validation ELITe Ingenius
2016 ECCMID Poster CMV Validation ELITe Ingenius
OBJECTIVES
The CMV ELITe MGB assay is a quantitative nucleic acids amplification assay for the detection and
quantification of Human Cytomegalovirus (CMV) based on MGB technology. The validation study was
performed with whole blood and plasma samples in combination with ELITe InGenius the first fully automated
sample-to-result solution introduced with a comprehensive quantitative transplant pathogen monitoring menu.
ELITe InGenius (ELITechGroup Molecular Diagnostics) automatically performs nucleic acid extraction, PCR
set-up, amplification and results analysis integrating in a single platform 12 extraction modules and 12
independently controlled unitary thermal cyclers equipped with 6 optical channels.
Nucleic acid extraction was performed in prefilled unitary cartridge, ELITe InGenius SP 200 (EMD), using a
magnetic bead technology. 200 L of samples (whole blood or plasma) were automatically dispensed in the
lysis buffer of the extraction cartridge with an internal control to check the process integrity. Nucleic acids
were eluted in 100 L in a dedicated storage tube. 20 L of extracted nucleic acid and 20 L of CMV ELITe
MGB Kit Real-Time PCR reagent were then automatically dispensed in a unitary PCR vessel: ELITe InGenius
PCR Cassette which is automatically capped by the system prior the PCR reaction. Extraction parameters,
thermal profiles information and result interpretation rules are all included in a specific pre-programmed assay
protocol.
CMV ELITe MGB kit consists of a quantitative monoreagent, based on MGB technology Real-Time PCR. The
verification and validation of CMV ELITe MGB Kit in combination with ELITe InGenius was based on:
(1) analytical studies to verify the PCR performances: efficiency, linearity, accuracy, repeatability, reproducibility;
and sensitivity; (2) analytical studies to verify the whole process (extraction and PCR) performance: linearity,
LoD, LoQ, conversion factor to International Unit with blood and plasma and reproducibility with certified
reference materials (Qnostics Ltd. and Acrometrix); (3) clinical study to evaluate the diagnostic sensitivity and
specificity assessed by testing clinical samples and negative donor samples with each matrix; (4) robustness
study: calibration stability, control chart consistency, cross contamination, whole system failure.
Sample characteristics and performance acceptance criteria are summarized in Table 1.
Performance Testing
Samples
Performance Criteria
Efficiency
3 x 100,000 gEq/reaction
Linearity
R2>0.99
Accuracy
Intra-run repeatability
Ct CV<2%
Inter-run repeatability
Ct CV<3%
Reproducibility
Ct CV<3%
18/18 replicates
were detected positive
Samples
Performance Criteria
Linearity
LoD
Sensitivity for IC
Whole Process Analytical
Performance Testing
Dilutions of the 1st WHO IS for Human CMV for NAT in whole blood and plasma tested in 10
replicates carrying out the whole analysis procedure
QCMD 2014 panel (CMVDNA14) and CMV
Molecular Q panel (Qnostics), AcroMetrix
CMVtc panel (ThermoFisher)
Performance Criteria
Clinical sensitivity
Clinical specificity
>90%
The LoD and the LoQs obtained with Whole Blood and Plasma matrix and determined in IU/ml and qEq/mL
is reported in the Table 5.
>90%
Robustness testing
Samples
Performance Criteria
Calibration stability
Ct CV% <5
Control stability
Cross contamination
Samples
Limit of Detection
Whole Blood
Plasma
*prepared by diluting the 1st WHO International Standard for Human Cytomegalovirus for Nucleic Acid Amplification Techniques in
CMV DNA - negative EDTA WB and Plasma
Clinical Studies
The clinical sensitivity and the clinical specificity obtained after discrepant sample analysis are presented in
the Table 6.
Whole Blood
RESULTS
CMV ELITe MGB Kit tested in combination with ELITe InGenius system passed all the performance acceptance
criteria established for the analytical and clinical studies regardless the sample matrix tested.
Analytical Studies
Plasma
Diagnostic Sensitivity
100% (60/60)
100% (54/54)
Diagnostic Specificity
93.2% (55/59)
98.3% (57/58)
The quantification reproducibility with CMV positive plasma samples versus the CE-IVD reference method
was assessed by regression analysis (Graph 1: Fitted Line Plot).
The results obtained with certified reference material to assess the integrated system performance are
described in Table 2, 3 and 4.
Panel
Member
Consensus
Value
Log IU/mL
Standard
Deviation
Positive /
Replicates
Mean Value
Result
Obtained
Log IU/ml
LOG
obtainedexpected
Log IU/mL
Acceptance
Criteria
CMVDNA14-01
2.468
0.343
2/2
2.256
-0.212
Passed
CMVDNA14-02
3.034
0.281
2/2
2.915
-0.119
Passed
CMVDNA14-03
3.383
0.368
2/2
3.185
-0.198
Passed
CMVDNA14-04
3.014
0.251
2/2
2.976
-0.038
Passed
CMVDNA14-05
1.859
0.462
2/2
1.706
-0.153
Passed
CMVDNA14-06
2.767
0.325
2/2
2.526
-0.241
Passed
CMVDNA14-07
4.030
0.280
2/2
3.924
-0.106
Passed
CMVDNA14-08
Negative
N.A.
0/2
Passed
CMVDNA14-09
2.065
0.512
2/2
1.273
< LoQ
Passed
CMVDNA14-10
3.947
0.278
2/2
3.946
-0.001
Passed
A R-Sq equal to 80.6% demonstrates a good correlation between the quantification of the two methods.
The Log difference (IU/ml) between both methods is shown on the Graph 2 (Scatter plot).
All the samples, except 3, had a Log difference (IU/ml) within 0.5 Log confirming the good reproducibility
between the two methods.
All samples (10) were correctly detected. Eight (8) out of nine positive samples were quantified within the
range defined by the EQA Consensus 1 Standard Deviation (SD) and one positive sample (CMVDNA14-09)
had a titer below the lower limit of quantification (LLoQ).
Sample
Acceptance Criteria
CMVMQP01-High
5.000
2/2
5.024
Passed
CMVMQP01-Medium
4.000
2/2
3.996
Passed
CMVMQP01-Low
3.000
2/2
3.060
Passed
CMVMQP01-Negative
0/2
Passed
Robustness Studies
Robustness studies had confirmed the total absence of cross-over and cross-contamination during repeated
runs and the whole system failure was assessed at 1.67%.
The stability of the standard curve has been established at 30 days since the quantification results obtained
at 30 days were comparable to the results obtained at time 0.
The Graph 3 shows the control chart verification obtained with 20 replicates of CMV ELITe Positive Control.
All positive samples were detected as positive with a titer within the expected value 0.5 Log.
Sample
Acceptance Criteria
6.477
2/2
6.386
Passed
5.477
2/2
5.444
Passed
4.477
2/2
4.473
Passed
3.477
2/2
3.441
Passed
2.477
2/2
2.575
Passed
All positive samples were detected as positive with a titer within the expected value 0.5 Log.
CONCLUSION
The excellent results obtained during the verification and validation study support the CE-IVD marking of
CMV ELITe MGB assay in combination with ELITe InGenius system for the detection and the quantification
of DNA extracted and amplified from whole blood and plasma.
EMD-CMV- ECCMID2016-550-2016/00EN