UT513 Eng Manual UNI-T PDF
UT513 Eng Manual UNI-T PDF
UT513 Eng Manual UNI-T PDF
OPERATING MANUAL
Table of Contents
Title
Introduction
Unpacking the Meter
Safety Information
International Electrical Symbols
Battery Saver (Sleep Mode)
Battery Indication
The Meter Structure
Display
Key Functions
Measurement Operation
A. Measuring Voltages
B. Measuring Insulation Resistance
a) Continuous Measurement
b) Timed Measurement
c) Polarization Index (PI) Measurement
d) Compare Function
The Use of Power Adaptor
Page
5
5
6
8
8
8
9
12
14
17
17
19
20
20
21
22
23
Title
USB Interface
Maintenance
A. General Service
B. Replacing the Battery
Specifications
Safety and Compliances
Physical Specifications
General Specifications
Feature Summary
Detailed Accuracy Specifications
A. Voltage Measurement
B. Insulation Resistance Measurement
Page
24
25
25
26
27
27
27
28
28
29
29
30
List of Tables
Table
Title
1.
2.
3.
4.
5.
6.
7.
Unpacking Inspection
International Electrical Symbols
Battery Indication
Meter Front Description
Meter Side Description
Display Description
Key Description
Page
5
8
8
10
11
13
14
List of Figures
Figure
1.
2.
3.
4.
5.
6.
7.
8.
Title
Page
9
11
12
17
19
23
24
26
Introduction
Uni-Trend Model UT513 insulation resistance tester
(hereafter, the Meter) is a handheld instrument
designed primarily to make resistance/ insulation
resistance measurement.
Item
1
2
4
5
6
7
8
9
Description
English Operating Manual
One plug test lead to one alligator
clip (Black colour)
One plug test lead to one alligator
clip (Green colour)
Two plugs test lead to one
alligator clip (Red colour)
1.5V Battery (LR14)
Tool Box
USB Interface Cable
Software
Power adaptor (input voltage
230V, 50/60Hz, 75mA, output
DC14V, 1.0A)(optionally, available
at extra cost)
Qty
1 piece
1 piece
1 piece
1 piece
8 pieces
1 piece
1 piece
1 piece
1 piece
Safety Information
This Meter complies with the standards IEC61010 safety
measurement requirement: in pollution degree 2,
overvoltage category (CAT. III 600V) and double
insulation.
CAT II: Local level, appliance, PORTABLE EQUIPMENT
etc., with smaller transient voltage overvoltages than
CAT. III
Use the Meter only as specified in this operating manual,
otherwise the protection provided by the Meter may be
impaired.
Danger identifies conditions and actions that pose
hazard(s) to the user.
Warning identifies avoiding electric shock.
Caution identifies conditions and actions that may
damage the Meter and carrying out accurate
measurement.
Battery Indication
Battery
Indicator
Battery Voltage
10V or less. It means the battery is
empty, dont use the Meter as it cannot
guarantee accuracy.
10V~10.5V. It means the battery is
nearly empty, replacing battery is
necessary. At this status, the Meter can
still do 500V and 1000V output
measurement, accuracy will not be
affected.
10.6V~11.5V
11.6V or more
Green
Black
Red
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
10
LCD
Scroll Button
Emergency Stop
Data Clear the Display Backlight Button
Down Button
On/Off Button
Compare Button
Insulation Resistance Button
DC Voltages measurement Button
Timer Button.
AC Voltages measurement Button
Test Button
USB Button
Data Store Button.
Data Recall Button
Scroll Button
17
18
19
20
21
22
Up Button
LINE: High Voltage output input terminal
(two plugs red test lead to one alligator clip)
High voltage line shielding input terminal
(two plugs red test lead to one alligator clip)
GUARD: Grounding protection input terminal
(one plug black test lead to one alligator clip)
EARTH: High resistance measurement input
terminal (one plug green test lead to one
alligator clip)
Testing leads:
Two plugs red test lead to one alligator clip.
One plug black test lead to one alligator clip.
One plug green test lead to one alligator
clip.
1
2
3
Safety Shutter
Power adaptor Input Terminal
USB Port
11
Display
Table 6 and Figure 3 describe the display.
Figure 3. Display
12
Number
1
2
3
4
5
6
7
8
9
10
11
Meaning
Indicator for DC voltage
Indicator for data store full
Indicator for clearing
Indicator for AC voltage
Indicator for timer
Step symbol
Indicates selected pass/fail compare value
Indicates for negative reading
Timer 1 symbol
Timer 2 symbol
Data store is on
Number
12
13
14
15
16
17
18
19
20
21
Meaning
Data recall is on
Indicator for polarization index
Unit symbols
The continuity buzzer is on
Compare feature pass
Analogue bar graph
Risk of electric shock
Compare feature fail
Indicator for power adaptor
Battery life indicator
13
Key Functions
ON/OFF
E-STOP
CLEAR /
SAVE
14
value.
LOAD
USB
15
COMP
TIME
TEST
IR
DCV
ACV
16
Measurement Operation
A. Measuring Voltages
Red
Green
17
Operating Catuion
To avoid harms to you or damages to the Meter,
please do not attempt to measure voltages
higher than 600V or 600V rms, although
readings may be obtained.
Special care should be taken when measuring
high voltage.
To measure voltages, set up the Meter as Figure 4 and
do the following:
1. Press DCV or ACV button to select DC voltage or
AC voltage measurement
2. Insert the red and green test lead into the tested
circuit.
3. When measuring DC voltage, if the red test lead is
negative voltage, - symbol will show on the display.
Note
When voltage measurement has been completed,
disconnect the connection between the testing
leads and the circuit under test and remove testing
leads away from the input terminals of the Meter.
18
Green
Black
Groud
Operating Caution
When performing insulation resistance tests,
remove all power from the circuit to be
measured and discharge all the power.
Operating the Meter must be very careful as it
outputs dangerous voltage during
measurement. Must make sure the tested object
is firmed clipped, hands are away from the
clips, then press TEST button to put high
voltage.
Do not short circuit the testing leads during
high voltages output or test insulation
resistance after high voltages output. This kind
of incorrect operating may cause sparking and
fire, which damages the Meter and harms to
you.
Do not measure over 10 seconds when:
500V measure resistance lower than 2M
1000V measure resistance lower than 5M
2500V measure resistance lower than 10M
5000V measure resistance lower than 20M
19
20
1.0 or less
Bad
21
22
1. Open the side safey shutter, then you will see there
is a power adaptor input terminal.
2. Make sure the Meter is power off and insert the
UT513 power adaptor to the input terminal.
3. It is highly recommeded to take out all the batteries
when you are using the power adaptor.
4. Make sure the Meter is power off when you
disconnect the UT513 power adaptor from the Meter.
5. It is highly recommeded to use Uni-Trend supplied
UT513 power adaptor to avoid dangerous.
23
USB Interface
Connecting the USB interface, see figure 7
24
Maintenance
25
Operating Caution
Dont mix to use old and new batteries.
Be careful the polarity is correct when installing
batteries.
Do not use the Meter if the battery indicator
(
) shows a battery empty condition.
Do you carry out measuring during the battery
compartment is open.
Follow Figure 8 and proceed as follows to replace the
battery:
26
Specifications
Safety and Compliances
Certification
Compliances
Physical Specifications
Display (LCD)
Operating Temperature
Storage Temperature
Relative Humidity
Battery Type
27
General Specifications
Range
Overloading
Battery Indicator
Icon Display
Current Consumption
Auto
Display OL on insulation resistance range
Display
Equips with function and battery indicator icons.
Maximum: around 1.0A
Average: around 20mA
Feature Summary
Display Backlight
Computer connection
Data Logging and Recall
Autorange
Warning
Voltage
COMP Measurement
PI Measurement
TIME
28
Measurement Range
Resolution
Accuracy
DC Voltage
30 ~ 600V
AC Voltage
30V~600V (50/60Hz)
1V
(2%+5) Among them 30~100V(50/60Hz) (2%+8)
29
500V
1000V
2500V
5000V
Display Range
0.5M ~20G
2M ~40G
5M ~100G
10M ~1000G
1mA~1.2mA @ 1M
1mA~1.2mA @ 2.5M
1mA~1.2mA @ 5M
Short Circuit
10.0M
~29.9M : (For
reference only)
30.0M ~99.9M : (3%+5)
~100G : (10%+5)
100M ~9.99G : (5%+5)
10.0G ~99.9G : (10%+5)
Above 100G : [ (20%+5)
Humidity:Below 50%]
0.50M
~99.9M : (3%+5)
5.0M
~99.9M : (3%+5)
100M
~9.99G : (5%+5)
100M
~9.99G : (5%+5)
100M
~9.99G : (5%+5)
10.0G
Operating Caution
At any output voltage, when the tested resistance is les than 10M , the testing time cannot exceed 10
seconds continuously.
30
*END*
This operating manual is subject to change without notice.
31
32