Model UT513

OPERATING MANUAL

Model UT513: OPERATING MANUAL

Table of Contents
Title

Introduction
Unpacking the Meter
Safety Information
International Electrical Symbols
Battery Saver (Sleep Mode)
Battery Indication
The Meter Structure
Display
Key Functions
Measurement Operation
A. Measuring Voltages
B. Measuring Insulation Resistance
a) Continuous Measurement
b) Timed Measurement
c) Polarization Index (PI) Measurement
d) Compare Function
The Use of Power Adaptor

Page

5
5
6
8
8
8
9
12
14
17
17
19
20
20
21
22
23

Model UT513: OPERATING MANUAL

Title

USB Interface
Maintenance
A. General Service
B. Replacing the Battery
Specifications
Safety and Compliances
Physical Specifications
General Specifications
Feature Summary
Detailed Accuracy Specifications
A. Voltage Measurement
B. Insulation Resistance Measurement

Page

24
25
25
26
27
27
27
28
28
29
29
30

Model UT513: OPERATING MANUAL

List of Tables
Table

Title

1.
2.
3.
4.
5.
6.
7.

Unpacking Inspection
International Electrical Symbols
Battery Indication
Meter Front Description
Meter Side Description
Display Description
Key Description

Page

5
8
8
10
11
13
14

Model UT513: OPERATING MANUAL

List of Figures
Figure

1.
2.
3.
4.
5.
6.
7.
8.

Title

The Meter Front Structure


The Meter Side Structure
Display
Voltages Measurement
Insulation Resistance Measurement
The Use of Power Adaptor
USB Interface Connection
Battery Replacement

Page

9
11
12
17
19
23
24
26

Model UT513: OPERATING MANUAL

Table 1. Unpacking Inspection

Introduction
Uni-Trend Model UT513 insulation resistance tester
(hereafter, the Meter) is a handheld instrument
designed primarily to make resistance/ insulation
resistance measurement.

Item
1
2

Unpacking the Meter

The Meter includes the following items:

4
5
6
7
8
9

Description
English Operating Manual
One plug test lead to one alligator
clip (Black colour)
One plug test lead to one alligator
clip (Green colour)
Two plugs test lead to one
alligator clip (Red colour)
1.5V Battery (LR14)
Tool Box
USB Interface Cable
Software
Power adaptor (input voltage
230V, 50/60Hz, 75mA, output
DC14V, 1.0A)(optionally, available
at extra cost)

Qty
1 piece
1 piece
1 piece
1 piece
8 pieces
1 piece
1 piece
1 piece
1 piece

In the event you find any missing or damage, please


contact your dealer immediately.

Model UT513: OPERATING MANUAL

Safety Information
This Meter complies with the standards IEC61010 safety
measurement requirement: in pollution degree 2,
overvoltage category (CAT. III 600V) and double
insulation.
CAT II: Local level, appliance, PORTABLE EQUIPMENT
etc., with smaller transient voltage overvoltages than
CAT. III
Use the Meter only as specified in this operating manual,
otherwise the protection provided by the Meter may be
impaired.
Danger identifies conditions and actions that pose
hazard(s) to the user.
Warning identifies avoiding electric shock.
Caution identifies conditions and actions that may
damage the Meter and carrying out accurate
measurement.

Operating Caution identifies conditions that user


needs to take extra care during operating the Meter
International electrical symbols used on the Meter and
in this Operating Manual are explained on page 8.
Danger
Use of instrument in a manual not specifed by the
manufactuer may impair safety features/protection
provided by the equipment. Read the following
safety information carefully before using or
servicing the instrument.
Do not apply more than 600V.
Do not use the Meter around explosive gas,
vapor or dust.
Do not use the Meter in a wet environment.
When using the test leads, keep your figures
away from the lead contacts. Keep your
figures behind the finger guards on the leads.
Do not use the Meter with any parts or cover
removed.

Model UT513: OPERATING MANUAL

When carrying out insulation measurement,


do not contact the circuit under test.
Warning
Do not use the Meter if it is damaged or metal
part is exposed. Look for cracks or missing
plastic.
Be careful when working above 33V rms, 46.7V
ac rms or 70V DC. Such voltages pose a shock
hazard.
Discharge all loading of circuit under test after
measuring high voltage.
Do not change battery when the Meter is in
wet environment.
Place test leads in proper input terminals.
Make sure all the test leads are firmly
connected to the Meters input terminals.
Make sure the Meter is turned off when
opening the battery compartment.
Caution

When performing resistance tests, remove all


power from the circuit to be measured and
discharge all the power.
When servicing the Meter, use only the same
model number or identical electrical
specifications of test leads and power adaptor.
Do not use the Meter if the battery indicator
(
) shows a battery empty condition. Take
the battery out from the Meter if it is not used
for a long time.
Do not use or store the Meter in an
environment of high temperature, humidity,
explosive, inflammable and strong magnetic
field. The performance of the Meter may
deteriorate after dampened.
Soft cloth and mild detergent should be used
to clean the surface of the Meter when
servicing. No abrasive and solvent should be
used to prevent the surface of the Meter from
corrosion, damage and accident.
Dry the Meter before storing if it is wet.

Model UT513: OPERATING MANUAL

International Electrical Symbols

Battery Indication

International symbols on the Meter and in this manual


are explained in Table 2.

There is a battery indicator shows on the display upper


left hand corner. Below Table 3 is the explanation:

Table 2. International Electrical Symbols

Risk of electric shock


Equipment protected by double or
reinforced insulation.
DC Measurement
AC Measurement
Grounding
See Manual
Empty of Built-In Battery
Conforms to Standards of European Union

Battery Saver (Sleep Mode)


The Meter enters the Sleep Mode and blanks the display
if there is no button press for 15 minutes. This is done
to conserve battery power. The Meter comes out of
Sleep Mode when ON/OFF button is pressed and hold
for 1 second.

Table 3. Battery Indication

Battery
Indicator

Battery Voltage
10V or less. It means the battery is
empty, dont use the Meter as it cannot
guarantee accuracy.
10V~10.5V. It means the battery is
nearly empty, replacing battery is
necessary. At this status, the Meter can
still do 500V and 1000V output
measurement, accuracy will not be
affected.
10.6V~11.5V
11.6V or more

Model UT513: OPERATING MANUAL

The Meter Structure


Below Figure 1 and Table 4 shows the Meter front
structure and description

Green

Black

Red

Figure 1. The Meter Front Structure

Model UT513: OPERATING MANUAL

Table 4. Meter Front Description

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16

10

LCD
Scroll Button
Emergency Stop
Data Clear the Display Backlight Button
Down Button
On/Off Button
Compare Button
Insulation Resistance Button
DC Voltages measurement Button
Timer Button.
AC Voltages measurement Button
Test Button
USB Button
Data Store Button.
Data Recall Button
Scroll Button

17
18
19
20
21

22

Up Button
LINE: High Voltage output input terminal
(two plugs red test lead to one alligator clip)
High voltage line shielding input terminal
(two plugs red test lead to one alligator clip)
GUARD: Grounding protection input terminal
(one plug black test lead to one alligator clip)
EARTH: High resistance measurement input
terminal (one plug green test lead to one
alligator clip)
Testing leads:
Two plugs red test lead to one alligator clip.
One plug black test lead to one alligator clip.
One plug green test lead to one alligator
clip.

Model UT513: OPERATING MANUAL

Below Figure 2 and Table 5 shows the Meter side


structure and description

Table 5. Meter Side Description

1
2
3

Safety Shutter
Power adaptor Input Terminal
USB Port

Figure 2. The Meter Side Structure

11

Model UT513: OPERATING MANUAL

Display
Table 6 and Figure 3 describe the display.

Figure 3. Display

12

Model UT513: OPERATING MANUAL

Table 6. Display Description

Number
1
2
3
4
5
6
7
8
9
10
11

Meaning
Indicator for DC voltage
Indicator for data store full
Indicator for clearing
Indicator for AC voltage
Indicator for timer
Step symbol
Indicates selected pass/fail compare value
Indicates for negative reading
Timer 1 symbol
Timer 2 symbol
Data store is on

Number
12
13
14
15
16
17
18
19
20
21

Meaning
Data recall is on
Indicator for polarization index
Unit symbols
The continuity buzzer is on
Compare feature pass
Analogue bar graph
Risk of electric shock
Compare feature fail
Indicator for power adaptor
Battery life indicator

13

Model UT513: OPERATING MANUAL

Key Functions
ON/OFF

E-STOP

CLEAR /

SAVE

14

Table 7. Key Description

Turn on or off the Meter. Press and


hold the button for 1 second to turn the
Meter on. Press again to turn off the
Meter.
The Meter default range is 500V
insulation resistance continuous
measurement when turning on.
Emergency stop button. Press this
button when the Meter is hang and
cannot turn off the power.
Press to turn on or off the display
backlight
Press and hold to clear the stored data
Press to store the current measurement
value. The maximum number of stored
reading is 18. When the stored readings
memory is full, the Meter shows FULL
and stop storing. Press and hold
CLEAR /
to clear the stored value
in order to store the next measurement

value.
LOAD

Press once to recall the first stored


value.
Press again to exit Load feature.
Load feature can only be used when
there is no high voltage output.
When the insulation resistance
measurement has no testing voltage
output, press to select one voltage
range up.
Under load mode: press to recall the
previous stored value.
When the insulation resistance
measurement has no testing voltage
output, press to select one voltage
range down.
Under load mode: press to recall the
next stored value.

Model UT513: OPERATING MANUAL

Table 7. Key Description

When set the timer duration for the


measurement of insulation resistance
or polarization index, press to
decrement the time. The maximum
length of time is 15 minutes and 30
seconds, the Meter will automatically
carry out measurement.
When compare feature measuring
insulation resistance, press to
decrement a resistance comparing
value.
After polarization index
measurement, press to display
polarization index, TIME 2 insulation
resistance value and TIME 1
insulation resistance value in
sequence.
When set the timer duration for the
measurement of insulation resistance
or polarization index, press to

USB

increment the time. The maximum


length of time is 30 minutes and 30
seconds, the Meter will automatically
carry out measurement.
When use the compare feature
measuring insulation resistance,
press to increment a resistance
comparing value.
After polarization index
measurement, press to display
polarization index, TIME 2 insulation
resistance value and TIME 1
insulation resistance value in
sequence.
Press once to start the data
transferring to the computer via USB,
USB symbol shows on the display.
Press again to stop the data
transferring to the computer via USB,
USB symbol disappears.

15

Model UT513: OPERATING MANUAL

Table 7. Key Description

COMP
TIME

TEST
IR
DCV
ACV

16

Set a pass / fail limit for insulation tests.


The default value is 10M
Pres to step through continuous
measurement, timed measurement and
polarization index measurement in
sequence.
Press to stop or start an insulation
resistance test
Press to initiate insulation resistance
measurement
Press to initiate DC voltage
measurement
Pres to initiate AC voltage
measurement

Model UT513: OPERATING MANUAL

Measurement Operation

A. Measuring Voltages

Below section explains how to make measurements.


Press and hold ON/OFF to turn on the Meter, press
again to turn off the Meter. The Meter default range is
500V insulation resistance continuous measurement
when turning on.

Red

Green

Figure 4. Voltages Measurement

17

Model UT513: OPERATING MANUAL

Operating Catuion
To avoid harms to you or damages to the Meter,
please do not attempt to measure voltages
higher than 600V or 600V rms, although
readings may be obtained.
Special care should be taken when measuring
high voltage.
To measure voltages, set up the Meter as Figure 4 and
do the following:
1. Press DCV or ACV button to select DC voltage or
AC voltage measurement
2. Insert the red and green test lead into the tested
circuit.
3. When measuring DC voltage, if the red test lead is
negative voltage, - symbol will show on the display.
Note
When voltage measurement has been completed,
disconnect the connection between the testing
leads and the circuit under test and remove testing
leads away from the input terminals of the Meter.

18

Model UT513: OPERATING MANUAL

B. Measuring Insulation Resistance


Red

Green

Black

Groud

Figure 5. Insulation Resistance Measurement

Operating Caution
When performing insulation resistance tests,
remove all power from the circuit to be
measured and discharge all the power.
Operating the Meter must be very careful as it
outputs dangerous voltage during
measurement. Must make sure the tested object
is firmed clipped, hands are away from the
clips, then press TEST button to put high
voltage.
Do not short circuit the testing leads during
high voltages output or test insulation
resistance after high voltages output. This kind
of incorrect operating may cause sparking and
fire, which damages the Meter and harms to
you.
Do not measure over 10 seconds when:
500V measure resistance lower than 2M
1000V measure resistance lower than 5M
2500V measure resistance lower than 10M
5000V measure resistance lower than 20M

19

Model UT513: OPERATING MANUAL

To measure insulation resistance, set up the Meter as


Figure 5 and do the following:
1. Press IR button to select insulation resistance
measurement.
2. When there is no testing voltage output, press
and button to select voltages of 500V, 1000V,
2500V or 5000V.
3. When performing insulation resistance tests, remove
all power from the circuit to be measured and
discharged all the power.
4. Insert the red test lead into the LINE input terminal
and the black test lead into GUARD input terminal.
5. Connect the red and black alligator clip to the circuit
to be measured, negative voltage output from LINE
terminal.
6. Choose below insulation resistance measurement
mode.

20

Press TIME button to select continuous


measurement mode, there is no timer icon on the
LCD.
Press and hold TEST button for 1 second to
carry out continuous measurement. Output
insulation resistance testing voltage, TEST button
light up, blinks on every 0.5 seconds.
Press TEST button to close the insulation resistance
measurement voltage when measurement is
completed. TEST button lights off, disappears.
The LCD shows the current insulation resistance
measurement value.

Press TIME button to select timed measurement


mode, the LCD displays TIME 1 and symbols.
Press and buttons to set the time (00:10~15:00).
Within 1 minute, the time increment or decrement
by every 5 seconds. Afterward, the time increment
or decrement by every 30 seconds.

Model UT513: OPERATING MANUAL

Then press and hold TEST button for 2 second to


carry out timed measurement. TIME 1 and are
displayed and blinked on the LCD on every 0.5
seconds.
When the set time is reached, the insulation
resistance measurement voltage will be closed
and the measurement will be automatically stopped.
The LCD displays the insulation resistance reading.

Press TIME button to select timed measurement


mode, the LCD displays TIME 1 and symbols.
Press and buttons to set the time (00:10~15:00).
Within 1 minute, the time increment or decrement
by every 5 seconds. Afterward, the time increment
or decrement by every 30 seconds.
Press TIME button again. TIME 2, PI and symbols
appear on the LCD.
Press and buttons to set the time (00:15~15:30).
Within 1 minute, the time increment or decrement
by every 10 seconds. Afterward, the time increment

or decrement by every 30 seconds.


Then press and hold TEST button for 2 seconds to
carry out timed measurement.
TIME 1 and are displayed and blinked on the
LCD on every 0.5 seconds before TIME 1 set time
is reached.
TIME 2 and are displayed and blinked on the
LCD on every 0.5 seconds before TIME 2 set time
is reached.
When the two set time are reached, the insulation
resistance measurement voltage will be closed and
the measurement will be automatically stopped.
The LCD displays the polarization index reading.
Press , to set through the polarization index,
TIME 2 insulation resistance reading and TIME 2
insulation resistance reading.
Information:
PI = 3 minutes ~10 minutes reading / 30 seconds ~1
minute reading
PI
4 or more 4~2
2.0~1.0
Standard The best Good Warning

1.0 or less
Bad

21

Model UT513: OPERATING MANUAL

Press COMP button to select compare feature.


COMP symbol displays on the LCD..
Press and buttons to set the compare value
Below is the list in sequence of the compare value:
10M , 20M , 30M , 40M , 50M , 60M , 70M ,
80M , 90M ,100M , 200M , 300M , 400M ,
500M , 600M , 700M , 800M , 900M ,1G ,
2G , 3G , 4G , 5G , 6G , 7G , 8G , 9G ,
10G ,20G , 330G , 40G , 50G , 60G , 70G ,
80G , 90G , 100G ,200G , 300G , 400G ,
500G , 600G , 700G , 800G , 900G
Press and hold TEST button for 2 seconds to carry
out the measurement.
The NG symbol will display if the insulation
resistance value is smaller than resistance value.
Otherwise GOOD symbol will be displayed.

22

Model UT513: OPERATING MANUAL

The Use of Power Adaptor


The use of power adaptor, see figure 6

1. Open the side safey shutter, then you will see there
is a power adaptor input terminal.
2. Make sure the Meter is power off and insert the
UT513 power adaptor to the input terminal.
3. It is highly recommeded to take out all the batteries
when you are using the power adaptor.
4. Make sure the Meter is power off when you
disconnect the UT513 power adaptor from the Meter.
5. It is highly recommeded to use Uni-Trend supplied
UT513 power adaptor to avoid dangerous.

Figure 6. The Use of Power Adaptor

23

Model UT513: OPERATING MANUAL

USB Interface
Connecting the USB interface, see figure 7

Figure 7. USB Interface Connection

24

1. Install the included software, the installation guide


can be seen from the CD.
2. Open the side safety shutter, then you will see there
is a USB port.
3. Insert the included USB cable to the Meters USB
port and the other end to the computer.

Model UT513: OPERATING MANUAL

Maintenance

If the Meter is wet, dry it before use.

This section provides basic maintenance information


including battery replacement instruction.
Warning
Do not attempt to repair or service your Meter unless
you are qualified to do so and have the relevant
calibration, performance test, and service
information.
A. General Service
Periodically wipe the case with a damp cloth and
mild detergent. Do not use abrasives or solvents.
To clean the terminals with cotton bar with detergent,
as dirt or moisture in the terminals can affect
readings.
Turn the Meter to OFF when it is not in use.
Take out the battery when it is not using for a long
time.
Do not use or store the Meter in a place of humidity,
high temperature, explosive, inflammable and strong
magnetic field.

25

Model UT513: OPERATING MANUAL

B. Replacing the Battery

Operating Caution
Dont mix to use old and new batteries.
Be careful the polarity is correct when installing
batteries.
Do not use the Meter if the battery indicator
(
) shows a battery empty condition.
Do you carry out measuring during the battery
compartment is open.
Follow Figure 8 and proceed as follows to replace the
battery:

Figure 8. Battery Replacement


Warning
To avoid electric shock, remove all the test leads
from the Meter when replacing the batteries.

26

Turn the Meter to OFF and remove all connections


from the terminals.
Remove the screw from the battery compartment,
and separate the battery compartment from the case
bottom.
Replace with 8pcs of new 1.5V (LR14) batteries.
Rejoin the case bottom and battery compartment,
and reinstall the screw.

Model UT513: OPERATING MANUAL

Specifications
Safety and Compliances
Certification
Compliances

IEC 61010 CAT.III 600V overvoltage and double insulation standard

Physical Specifications
Display (LCD)
Operating Temperature
Storage Temperature
Relative Humidity
Battery Type

Dimensions (HxW xL)


Weight

Digital: 9999 counts


Analog bar graph.
-10 ~40 (14 ~104 )
-20 ~60 (-4 ~152 )
85% @ -10 ~40 below;
90% @ -20 ~60 :
8pcs of 1.5V (LR14) batteries or power adaptor (input voltage 230V, 50/60Hz,
75mA, input DC14V, 1.0A).
Power adaptor is optionally at extra cost.
202 x 155 x 94 mm
Approx. 2kg (including battery)

27

Model UT513: OPERATING MANUAL

General Specifications
Range
Overloading
Battery Indicator
Icon Display
Current Consumption

Auto
Display OL on insulation resistance range
Display
Equips with function and battery indicator icons.
Maximum: around 1.0A
Average: around 20mA

Feature Summary
Display Backlight
Computer connection
Data Logging and Recall
Autorange
Warning
Voltage
COMP Measurement
PI Measurement
TIME

28

Bright backlight for clear readings in poorly lighted areas.


Via USB interface.
18
The Meter automatically selects best range
and red light will on.
Auto release voltage
Use the Compare function to set a pass/fail compare level for the insulation
measurements.
Polarization Index is the ratio of insulation resistance. You can pre-set two
point of times and automatically carry out the measurement.
To carry out measurement by setting a specified time within 15 minutes.

Model UT513: OPERATING MANUAL

Detailed Accuracy Specifications


Accuracy: ([% of reading] + [number of least significant digits), guarantee for 1 year.
Operating temperature: 18 ~28
Relative humidity: 45~75%RH
A. Voltage Measurement

Measurement Range
Resolution
Accuracy

DC Voltage
30 ~ 600V

AC Voltage
30V~600V (50/60Hz)

1V
(2%+5) Among them 30~100V(50/60Hz) (2%+8)

29

Model UT513: OPERATING MANUAL

B. Insulation Resistance Measurement


Output Voltage

500V

1000V

2500V

5000V

Display Range

0.5M ~20G

2M ~40G

5M ~100G

10M ~1000G

DC1000V 0%~+ 20%

DC 2500V 0%~+ 20%

DC5000V 0%~+ 20%

1mA~1.2mA @ 1M

1mA~1.2mA @ 2.5M

1mA~1.2mA @ 5M

Open Circuit Voltage DC 500V 0%~+ 20%


Test Current
1mA~1.2mA @ 500k
Accuracy

Short Circuit

10.0M
~29.9M : (For
reference only)
30.0M ~99.9M : (3%+5)
~100G : (10%+5)
100M ~9.99G : (5%+5)
10.0G ~99.9G : (10%+5)
Above 100G : [ (20%+5)
Humidity:Below 50%]

0.50M

~99.9M : (3%+5) 2.0M

~99.9M : (3%+5)

5.0M

~99.9M : (3%+5)

100M

~9.99G : (5%+5)

100M

~9.99G : (5%+5)

100M

~9.99G : (5%+5)

10.0G

~20.0G : (10%+5) 10.0G

~40.0G : (10%+5) 10.0G

Less than 2.0mA

Operating Caution
At any output voltage, when the tested resistance is les than 10M , the testing time cannot exceed 10
seconds continuously.

30

Model UT513: OPERATING MANUAL

*END*
This operating manual is subject to change without notice.

31

Model UT513: OPERATING MANUAL

Copyright 2007 Uni-Trend Group Limited.


All rights reserved.
Manufacturer:
Uni-Trend Technology (Dongguan) Limited
Dong Fang Da Dao
Bei Shan Dong Fang Industrial Development District
Hu Men Town, Dongguan City
Guang Dong Province
China
Postal Code: 523 925
Headquarters:
Uni-Trend Group Limited
Rm901, 9/F, Nanyang Plaza
57 Hung To Road
Kwun Tong
Kowloon, Hong Kong
Tel: (852) 2950 9168
Fax: (852) 2950 9303
Email: [email protected]
http://www.uni-trend.com