An Introduction To Pound-Drever-Hall Laser Frequency Stabilization
An Introduction To Pound-Drever-Hall Laser Frequency Stabilization
An Introduction To Pound-Drever-Hall Laser Frequency Stabilization
stabilization
Eric D. Black
LIGO Project, California Institute of Technology
Mail Code 264-33, Pasadena CA, 91125
(October 4, 1999)
Abstract
LIGO-P990042-00-D
I. INTRODUCTION
II. BACKGROUND
What we now call the Pound-Drever-Hall technique was proposed in 1980 by Drever
specifically for gravitational wave detection [8], and early development was done at the
University of Glasgow. Detailed experiments on the capabilities and performance of the
technique were carried out in a collaboration with Jan Halls group at JILA, the Joint
Institute for Laboratory Astrophysics in Boulder, Colorado [9].
The technique seems to have been proposed independently slightly later by Bjorklund
[10], for use in FM spectroscopy. The idea is similar to an older technique used in microwave
applications, invented in the forties by R. V. Pound [11].
Suppose we have a laser that we want to use for some experiment, but we need better frequency stability than the laser provides out of the box. Many modern lasers are
tuneable: They come with some input port into which you can feed an electrical signal and
adjust the output frequency. If we have an accurate way to measure the lasers frequency,
then we can feed this measurement into the tuning port, with appropriate amplification and
filtering, and hold the frequency (roughly) constant.
One good way to measure the frequency of a lasers beam is to send it into a FabryPerot cavity and look at what gets transmitted (or reflected). Recall that light can only
pass through a Fabry-Perot cavity if twice the length of the cavity is equal to an integer
number of wavelengths of the light. Another way to say this is that the frequency of the
lights electromagnetic wave must be an integer number times the cavitys free spectral range
f sr c/2L, where L is the length of the cavity and c is the speed of light. The cavity
acts as a filter, with transmission lines, or resonances spaced evenly in frequency every
free spectral range. Figure 1 shows a plot of the fraction of light transmitted through a
Fabry-Perot cavity versus the frequency of the light.
If we were to operate just off to one side of one of these resonances, but near enough
that some light gets transmitted (say, half the maximum transmitted power), then a small
change in laser frequency would produce a proportional change in the transmitted intensity.
We could then measure the transmitted intensity of the light and feed this signal back to
the laser to hold this intensity (and hence the laser frequency) constant.
This was often how laser locking was done before the invention of the Pound-DreverHall method, and it suffers from one major flaw: The system cannot distinguish between
fluctuations in the lasers frequency, which changes the intensity transmitted through the
cavity, and fluctuations in the intensity of the laser itself.
We could build a separate system to stabilize the lasers intensity, which was apparently
done with some limited success in the late seventies [9], but a better method would be to
measure the reflected intensity and hold that at zero, which would decouple intensity and
frequency noise. The only problem with this scheme is that the intensity of the reflected
beam is symmetric about resonance. If the laser frequency drifts out of resonance with the
cavity, you cant tell just by looking at the reflected intensity whether the frequency needs
to be increased or decreased to bring it back onto resonance. The derivative of the reflected
intensity, however, is antisymmetric about resonance. If we were to measure this derivative,
we would have an error signal that we can use to lock the laser. Fortunately, this is not
too hard to do: We can just vary the frequency a little bit and see how the reflected beam
responds.
Above resonance, the derivative of the reflected intensity with respect to laser frequency
is positive. If we vary the lasers frequency sinusoidally over a small range, then the reflected
intensity will also vary sinusoidally, in phase with the variation in frequency. (See Figure 2.)
Below resonance, this derivative is negative. Here the reflected intensity will vary 180
out of phase from the frequency. On resonance the reflected intensity is at a minimum, and
a small frequency variation will produce no change in the reflected intensity.
By comparing the variation in the reflected intensity with the frequency variation we
can tell which side of resonance we are on. Once we have a measure of the derivative of the
4
reflected intensity with respect to frequency, we can feed this measurement back to the laser
to hold it on resonance.
Fig. 3 shows a basic Pound-Drever-Hall setup. Here the frequency is modulated with
a Pockels cell [12], driven by some local oscillator. The reflected beam is picked off with
an optical isolator and sent into a photodetector, whose output is compared with the local
oscillators signal via a mixer. The output of the mixer is the product of its inputs, and
it contains signals at both dc and twice the the modulation frequency. It is the signal at
dc that we are interested in, since that is the component of the photodiodes output at the
modulation frequency. A low-pass filter on the output of the mixer isolates this dc signal,
which then goes through a servo amplifier and into the tuning port on the laser, locking the
laser to the cavity.
The Faraday isolator keeps the reflected beam from getting back into the laser and
destabilizing it. This isolator is not necessary for understanding the technique, but it is
essential in a real system. In practice, the small amount of reflected beam that gets through
the optical isolator is usually enough to ruin your lock. Similarly, the phase shifter is not
essential in an ideal system, but is necessary in practice to compensate for unequal delays in
the two signal paths. (It could just as easily go between the local oscillator and the Pockels
cell.)
This conceptual model is really only good if you are dithering the laser frequency slowly.
If you dither the frequency too fast, the light resonating inside the cavity wont have time
to completely build up or settle down, and the output will not follow the curve shown in
Fig. 2. However, the technique still works at higher modulation frequencies, and both the
noise performance and bandwidth of the servo are typically improved. Before we address
a conceptual picture that does apply to the high-frequency regime, we must establish a
quantitative model.
To describe the behavior of the reflected beam quantitatively, we can pick a point outside
the cavity and measure the electric field over time [13]. The electric field of the incident
beam can be written
Einc = E0 eit .
The electric field of the reflected beam (measured at the same point) is
Eref = E1 eit.
We account for the relative phase between the two waves by letting E0 and E1 be complex.
The reflection coefficient F () is the ratio of Eref and Einc , and for a symmetric cavity with
no losses it is given by
F () = Eref /Einc
r e
=
f sr
1 r2 e
(4.1)
f sr
where r is the amplitude reflection coefficient of each mirror, and f sr = c/2L is the free
spectral range of the cavity of length L.
The beam that reflects from a Fabry-Perot cavity is actually the coherent sum of two
different beams: the promptly reflected beam, which bounces off the first mirror and never
enters the cavity; and a leakage beam, which is the small part of the standing wave inside
the cavity that leaks back through the first mirror, which is never perfectly reflecting. These
two beams have the same frequency, and near resonance their intensities are almost the same
as well. Their relative phase, however, depends strongly on the frequency of the laser beam.
If the cavity is resonating perfectly, i.e. the lasers frequency is exactly an integer multiple
of the cavitys free spectral range, then the promptly reflected beam and the leakage beam
6
have the same amplitude and are exactly 180 out of phase. In this case the two beams
interfere destructively, and the total reflected beam vanishes.
If the cavity is not quite perfectly resonant, that is the lasers frequency is not exactly
an integer multiple of the free spectral range but close enough to build up a standing wave,
then the phase difference between the two beams will not be exactly 180 , and they will
not completely cancel each other out. (Their intensities will still be about the same.) Some
light gets reflected off the cavity, and its phase tells you which side of resonance you are
on. Figure 4 shows a plot of the intensity and phase of the reflection coefficient around
resonance.
It is useful to look at the properties of F () in the complex plane. (See Fig. 5.) It
is not too hard to show (see Appendix A) that the value of F always lies on a circle in
the complex plane, centered on the real axis, with being the parameter that determines
where on this circle F will be. |F () |2 gives the intensity of the reflected beam, and it
is given by the familiar Airy function. F is symmetric around resonance, but its phase is
different depending on which side of resonance you are on. As you increase , F advances
counterclockwise around the circle. For the symmetric, lossless cavity we are considering,
this circle intersects the origin, with F = 0 on resonance. Very near resonance, F is nearly
on the imaginary axis, being in the lower half plane below resonance and in the upper half
plane above resonance.
We will use this graphical representation of F in the complex plane when we try to
understand the results of our quantitative model.
We can measure the phase of the reflected beam by placing sidebands on the incident
beam, reflecting them off the cavity, and allowing them to interfere with the main reflected
beam. The sidebands will have a definite phase relationship with the incident beam. When
combined with the reflected beam, they will produce a measurable beat pattern that will
tell you the relative phase between the incident and reflected beams.
I talked about varying the frequency of this beam in the qualitative model, but in practice
it is easier to modulate the phase. The results are essentially the same, but the math
that describes phase modulation is easier than the math for frequency modulation. Phase
modulation is also easy to implement with a Pockels cell, as shown in Fig. 3. After the beam
has passed through the Pockels cell, its electric field is [14]
Einc = E0ei(t+ sin t)
E0 [J0 () + 2iJ1 () sin t] eit
h
(4.2)
I have written it in this form to show that there are actually three different beams incident
on the cavity: a carrier, with (angular) frequency , and two sidebands with frequencies
. Here, is the phase modulation frequency, and is known as the modulation
depth. If P0 |E0 |2 is the total power in the incident beam, then the power in the carrier
is (neglecting interference effects for now)
Pc = J02 () P0 ,
and the power in each first-order sideband is
Ps = J12 () P0 .
When the modulation depth is small ( 1), almost all of the power is in the carrier
and the first order sidebands.
Pc + 2Ps P0
We will use these results shortly.
8
To calculate the reflected beams field when there are several incident beams, we can
treat each beam independently and multiply each one by the reflection coefficient at the
appropriate frequency. In the Pound-Drever-Hall setup, where we have a carrier and two
sidebands, the total reflected beam is
Eref = E0 [F () J0 () eit
+F ( + ) J1 () ei(+)t F ( ) J1 () ei()t].
What we really want is the power in the reflected beam, since that is what we measure
with the photodetector. This is just Pref = |Eref |2, or after some algebra
Pref = Pc |F () |2 + Ps {|F ( + ) |2 + |F ( ) |2 }
q
+2 Pc Ps {Re [F () F ( + ) F () F ( )] cos t
+Im [F () F ( + ) F () F ( )] sin t}
+(2 terms).
(4.3)
We have added three waves of different frequencies, the carrier, at , and the upper
and lower sidebands at . The result is a wave with a nominal frequency of , but
with an envelope displaying a beat pattern with two frequencies. The terms arise from
the interference between the carrier and the sidebands, and the 2 terms come from the
sidebands interfering with each other.
We are interested in the two terms that are oscillating at the modulation frequency
because they sample the phase of the reflected carrier. There are two terms in this expression:
a sine term and a cosine term. Usually, only one of them will be important. The other will
vanish. Which one vanishes and which one survives depends on the modulation frequency.
In the next section we will show that at low modulation frequencies (slow enough for the
internal field of the cavity to have time to respond, or f sr /F ), F () F ( + )
F () F ( ) is purely real, and only the cosine term survives. At high ( f sr /F )
near resonance it is purely imaginary, and only the sine term is important.
9
We measure the reflected power given in Equation 4.3 with a high-frequency photodetector, as shown in Figure 3. The output of this photodetector includes all terms in Equation 4.3, but we are only interested in the sin(t) or cos(t) part, which we can isolate using
a mixer. A mixer is a nonlinear device whose output is the product of its inputs. Recall
that the product of two sine waves is
sin(t) sin(0 t) =
1
{cos [( 0 )t] cos [( + 0 )t]}.
2
If we feed the modulation signal (at ) into one input of the mixer and some other signal
(at 0 ) into the other input, the output will contain signals at both the sum ( + 0) and
difference ( 0 ) frequencies. If 0 is equal to , as is the case for the part of the signal
we are interested in, then the cos(( 0 )t) term will be a dc signal, which we can isolate
with a low-pass filter, as shown in Figure 3.
Note that if we mix a sine and a cosine signal, rather than two sines, we get
sin(t) cos(0 t) =
1
{sin [( 0 )t] sin [( + 0)t]}.
2
In this case, if = 0 our dc signal vanishes! If we want to measure the error signal at low
modulation frequencies, where it is the cosine term in Equation 4.3 that we want, we need
to convert the cos(t) term into a sin(t) term before feeding it into the mixer. Turning a
sine into a cosine is a simple matter of introducing a 90 phase shift, which we can do with
a phase shifter (or delay line), as shown in Figure 3.
In practice, you need a phase shifter even when the modulation frequency is high. There
are almost always unequal delays in the two signal paths that need to be compensated for to
produce two pure sine terms at the inputs of the mixer. The output of the mixer when the
10
phases of its two inputs are not matched can produce some odd-looking error signals (see
Bjorklund [10]), and when setting up a Pound-Drever-Hall lock you usually scan the laser
frequency and empirically adjust the phase in one signal path until you get an error signal
that looks like Figure 7.
Lets see how the quantitative model compares with our conceptual model, where we
slowly dithered the laser frequency and looked at the reflected power. For our phase modulated beam, the instantaneous frequency is
(t) =
d
(t + sin t)
dt
= + cos t.
The reflected power is just Pref = P0 |F () |2 , and we might expect it to vary over time
as
dPref
cos t
d
d|F |2
cos t.
() + P0
d
In the conceptual model, we dithered the frequency of the laser adiabatically, slowly
enough that the standing wave inside the cavity was always in equilibrium with the incident
beam. We can express this in the quantitative model by making very small. In this regime
the expression
F () F ( + ) F () F ( ) 2Re{F ()
d|F |2
,
d
d
F ()}
d
which is purely real. Of the terms, only the cosine term in Eqn. 4.3 survives.
11
If we approximate
d|F |2
cos t
d
+ (2terms) ,
in agreement with our expectation from the conceptual model.
The mixer will filter out everything but the term that varies as cos t. (We may have
to adjust the phase of the signal before we feed it into the mixer.) The Pound-Drever-Hall
error signal is then
d|F |2
d
q
d|F |2
2 Pc Ps
.
d
= P0
When the carrier is near resonance and the modulation frequency is high enough that the
sidebands are not, we can assume that the sidebands are totally reflected, F ( ) 1.
Then the expression
F () F ( + ) F () F ( ) i2Im{F ()},
(5.1)
is purely imaginary. In this regime, the cosine term in Equation 4.3 is negligible, and our
error signal becomes
q
= 2 Pc Ps Im{F () F ( + ) F () F ( )}.
Fig. 7 shows a plot of this error signal.
Near resonance the reflected power essentially vanishes, since |F () |2 0. We do want
to retain terms to first order in F (), however, to approximate the error signal.
12
Pref 2Ps
q
= 2N +
f sr
f sr
where is the deviation of the laser frequency from resonance. It is useful at this point to
make the approximation that the cavity has a high finesse F / (1 r2 ). The reflection
coefficient is then
F
i
,
4q
Pc Ps
That the error signal is linear near resonance allows us to use the standard tools of
control theory to suppress frequency noise. We will use this linear behavior later on to
examine some fundamental noise limits. It will be useful for us to write the error signal
in terms of the regular frequency f = /2, instead of , and define the proportionality
constant between and f.
= Df,
where the proportionality constant,
8 Pc Ps
D
,
13
(5.2)
There is a way to think about this technique conceptually at high modulation frequencies.
When the modulation frequency was low, we pictured the reflected power in the time domain
and compared it to the modulation of the laser. Here we will do the same thing, only the
visualization will be somewhat different.
We can represent the electric fields of each beam by time-varying vectors in a complex
plane that rotates along with the carrier at frequency . We can choose this moving
reference frame such that the incident carriers electric field component always lies along
the real axis. The part of the carrier that gets reflected from the Fabry-Perot cavity is also
represented by a vector in this plane, as shown in Fig. 5, and near resonance it is given by
i P0 /.
The sidebands have different frequencies than the carrier, so they are represented by
vectors that spin around in this reference frame. The upper ( + ) sideband has a higher
frequency than the carrier, so its vector rotates counterclockwise in the plane with a frequency of . The lower sideband has a lower frequency and rotates clockwise at . (See
Fig. 8.) The sum of the two sidebands, when they are both completely reflected off the
cavity, is a single electric field that oscillates up and down along the imaginary axis. (See
2
q
q
i2 P0 J1 () sin t
i P0
!2
P0
The cross term proportional to sin t represents the beating of the sidebands with the
reflected carrier, and its sign tells you which side of resonance you are on. The 2 term is
14
VI. NOISE AND FUNDAMENTAL LIMITS: HOW WELL CAN YOU DO?
I have only talked about laser frequency so far, but it is a straightforward exercise to
extend this analysis in terms of both frequency and cavity length. A little algebra shows
that the laser frequency and the cavity length are on equal footings near resonance. For
high modulation frequencies,
q
2LF
= 8 Pc Ps
f
L
,
+
f
L
where L is the deviation of the cavity length from resonance, analogous to f. All along I
have been talking about measuring the frequency noise of the laser and locking it to a quiet
cavity, but we could have just as easily measured the length noise in the cavity (provided
the laser was relatively quiet) and locked the cavity to the laser.
Note that it is not possible to distinguish laser frequency noise from cavity noise just by
looking at the error signal. One naturally wonders what other noise sources contribute to
the error signal. It is a straightforward exercise to show that none of the following contribute
to the error signal to first order: variation in the laser power, response of the photodiode
15
used to measure the reflected signal, the modulation depth , the relative phase of the two
signals going into the mixer, and the modulation frequency . The system is insensitive to
each of these because we are locking on resonance, where the reflected carrier vanishes. This
causes all of these first order terms to vanish in a Taylor expansion of the error signal about
resonance.
The system is first order sensitive to fluctuations in the sideband power at the modulation
frequency . Most noise sources fall off as frequency increases, and you can usually reduce
them as much as you want by going to a high enough modulation frequency. There is
one noise source, however, that does not trail off at high frequencies, and that is the shot
noise in the reflected sidebands. Shot noise has a flat noise spectrum, and for high enough
modulation frequencies it is the dominant noise source.
Any noise in the error signal itself is indistinguishable from noise in the lasers frequency.
There is a fundamental limit to how quiet the error signal can be, due to the quantum nature
of light.
On resonance, the reflected power, and hence the light power falling on the photodiode,
will be Pref = 2Ps . The shot noise in this signal has a flat spectrum with spectral density of
s
S =
hc
(2Ps ).
Dividing the error signal spectrum by D gives us the apparent frequency noise,
hc3
1
Sf =
.
8 F L Pc
Since you cant resolve the frequency any better than this, you cant get it any more
stable than this by using feedback to control the laser. Note that the shot noise limit does
not explicitly depend on the power in the sidebands, as you might expect. It only depends
on the power in the carrier [15].
16
Its worth putting in some numbers to get a feel for these limits. For this example we
will use a cavity that is 20cm long and has a finesse of 104 , and a laser that operates at
500mW with a wavelength of 1064nm. If the cavity had no length noise and we locked the
laser to it, the best frequency stability we could get would be
5
Sf = 1.2 10
Hz
Hz
104 20cm
F L
1064nm 500mW
.
Pc
The same shot noise would limit your sensitivity to cavity length if you were locking the
cavity to the laser. In this case, the apparent length noise would be
L
Sf
f
hc
.
=
8 F Pc
SL =
(6.1)
For the example cavity and laser we used above, this would be
21
SL = 8.1 10
Hz
104
F
500mW
.
1064nm Pc
Say you want to lock a cavity to a laser using a low-frequency servo and measure highfrequency length changes by looking at the error signal.
Mizuno has estimated the quantum limited sensitivity of any interferometer from very
general principles, and his result is known as the sensitivity theorem [16]. Since shot noise
is the quantum mechanism that limits the sensitivity of a Fabry-Perot cavity when used as
an interferometer, our shot noise limit ought to give the same limit as Mizunos sensitivity
theorem.
The sensitivity theorem is usually stated in terms of strain, Sh = SL /L. According to
this theorem, quantum effects must limit your strain sensitivity to [16]
s
Sh
1h
,
c
17
where is the energy stored in the light circulating in the interferometer. The energy stored
in a Fabry-Perot cavity on resonance is the circulating power Pcirc Pc F / [17] multiplied
by the time it takes for light to circulate once, 2L/c.
If we write our shot noise limit (Eqn. 6.1) in terms of these parameters, we get a strain
resolution limit of
s
Sh =
1h
,
2 c
There are lots of other kinds of noise, of course, but most of these can be suppressed
by feedback. Tim Day has done a nice analysis of technical noise sources in laser frequency
stabilization [18]. The closed loop frequency noise, in terms of the other noise sources, is
[18]
q
Sf,cl =
where Sf is the free running frequency noise in the laser, Sv,servo is the voltage noise in the
servo amplifier, and Sv,pd is the voltage noise in the photodetector. The terms K, G, and
Dv are the gains of the actuator, the servo amp, and the photodetector, respectively. The
units of K are Hz/V , Dv has units of V/Hz, and G is unitless. The photodetector gain Dv
is just the slope of the error signal D (the discriminant) multiplied by the voltage response
of the photodetector.
If you make the servo gain very large, G 1, the noise spectrum limits out to
Sf,cl
18
Sv,pd
.
Dv
The closed loop cavity noise is dominated by the noise in the photodetector. If the noise in
the photodetector is mainly shot noise, then this reduces to the limit derived in the previous
section.
If you want to get close to the shot noise limit, you need to concentrate on the photodetector (to reduce Sv,pd ). High servo gain takes care of all the other noise sources.
If you want to minimize the closed loop noise, you need to maximize D. This wont
affect the shot noise limit much, but it will help suppress technical noise sources. D depends
on the cavity finesse, the laser wavelength, and the power in the sidebands and the carrier.
Experimental details usually restrict your finesse and wavelength choices, but you often have
quite a bit of freedom in adjusting the sideband power. The question I want to address in
this section is this: How does D (and hence the closed loop noise) depend on the sideband
power?
D is proportional to the square root of the product of the sideband and carrier power.
(See Eqn. 5.2.) This has a very simple form when Pc + 2Ps P0 , i.e. when negligible power
goes into the higher order sidebands.
q
Pc Ps
P0
2
s
Pc
1
P0
Pc
P0
A plot of D against Pc /P0 traces out the top half of a circle, with a maximum at Pc /P0 = 1/2.
(See Fig. 11.) It is useful to express the power in the sidebands relative to the power in the
carrier, and this gives
Ps
1
= .
Pc
2
D is maximized when the power in each sideband is half the power in the carrier, and this
maximum is fairly broad.
19
If you want do a more careful analysis, write D in terms of the Bessel functions of the
modulation depth and find its maximum. Youll find the optimum modulation depth to be
= 1.08, and youll come up with essentially the same answer as with the simple estimate:
Ps
= 0.42.
Pc
VIII. ACKNOWLEDGMENTS
This work was supported by the National Science Foundation as part of the LIGO project.
For the general case of a Fabry-Perot cavity with lossy mirrors, the reflection coefficient
is
F =
r1 + r2 (r12 + t21 ) e
1 r1 r2 e
i
f sr
f sr
Here, r1 and t1 are the amplitude reflection and transmission coefficients of the input mirror,
and r2 is the amplitude reflection coefficient of the end mirror. (Note that the reflection
coefficient does not depend on the transmission of the end mirror!) It is straightforward
algebra to show that F satisfies the equation of a circle:
|F () Z0 |2 = R2
for all . Z0 and R are both real and are given by
Z0 =
h
i
r1
2
2
2
1
r
r
+
t
,
2
1
1
1 r12 r22
and
t21 r2
R=
.
1 r12 r22
If this circle intersects the origin (i.e. the reflected intensity vanishes on resonance) then
the cavity is said to be critically coupled. The requirement for critical coupling is
20
i
= t21 r2.
(A1)
t21 +
Other cases, known as under coupling and over coupling, are illustrated along with critical
coupling in Figure 12. Over coupling plays a central role in interferometric gravitationalwave detectors and can be achieved by making the end mirror much more reflective than
the near mirror (t2 t1).
21
REFERENCES
[1] C.E. Wieman and L. Hollberg. Using diode lasers for atomic physics. Rev. Sci. Instrum., 62(1):120, (1991).
[2] A. Abromovici et al. LIGO: The laser interferometer gravitational-wave observatory.
Science, 256:325333, (1992).
[3] S. Seel, R. Storz, G. Ruoso, J. Mlynek, and S. Schiller. Cryogenic optical resonators:
A new tool for laser frequency stabilization at the 1 Hz level. Phys. Rev. Lett.,
78(25):47414744, (1997).
[4] L.A. Rawley et al. Millisecond Pulsar PSR 1937+21: A Highly Stable Clock. Science,
238:761765, (1987).
[5] R. A. Boyd, J. L. Bliss, and K. G. Libbrecht. Teaching physics with 670-nm diode
lasersexperiments with fabry-perot cavities. Am. J. Phys., 64:11091116, (1996).
[6] G. F. Franklin. Feedback Control of Dynamic Systems. Addison-Wesley, Reading, Massachusetts, (1987).
[7] H. Kogelnik and T. Li. Laser beams and resonators. Applied Optics, 5:15501567,
(1966).
[8] R. W. P. Drever et al. A Gravity-Wave Detector Using Optical Cavity Sensing. In
E. Schmutzer, editor, Proceedings of the 9th International Conference on General Relativity and Gravitation, Jena, July 1980, pages 265267. (Cambridge University Press,
1983).
[9] R. W. P. Drever. Laser phase and frequency stabilization using an optical resonator.
Appl. Phys. B, (1983).
[10] G. C. Bjorklund. Frequency Modulation (FM) Spectroscopy: Theory of Lineshapes
and Signal-to-Noise Analysis. Appl. Phys. B, (1983).
22
eit eit
i
[15] The shot noise limit does depend implicitly on the power in the sidebands, since Pc =
P0 Ps , but this is a relatively minor effect.
[16] Jun Mizuno. Comparison of optical configurations for laser-interferometer gravitationalwave detectors. PhD thesis, Max-Plank-Institute f
ur Quantenoptik, (1995).
[17] A. E. Siegman. Lasers. University Science Books, Mill Valley, California, (1986).
[18] T. Day. Frequency Stabilized Solid State Lasers for Coherent Optical Communications.
PhD thesis, Stanford University, (1990).
23
FIGURES
FIG. 1. Transmission of a Fabry-Perot cavity versus frequency of the incident light. This cavity
has a fairly low finesse, about 12, to make the structure of the transmission lines easy to see.
FIG. 2. The reflected light intensity from a Fabry-Perot cavity as a function of laser frequency,
near resonance. If you modulate the laser frequency, you can tell which side of resonance you are
on by how the reflected power changes.
FIG. 3. The basic layout for locking a cavity to a laser. Solid lines are optical paths, and
dashed lines are signal paths. The signal going to the laser controls its frequency.
FIG. 4. Magnitude and phase of the reflection coefficient for a Fabry-Perot cavity. As in Fig. 1,
the finesse is about 12. Note the discontinuity in phase, caused by the reflected power vanishing
at resonance.
FIG. 5. The reflection coefficient in the complex plane. As the laser frequency (or equivalently,
the cavity length) increases, F () traces out a circle (counterclockwise). Most of the time, F is
near the real axis at the left edge of the circle. Only near resonance does the imaginary part of F
become appreciable. Exactly on resonance, F is zero.
FIG. 8. Sidebands at .
24
FIG. 9. The sum of the two sidebands shown in Fig. 8. This is the actual electric field produced
when the two sidebands interfere with each other. Note that the intensity oscillates at 2.
FIG. 10. Sidebands and the reflected carrier near resonance. The actual reflected beam is
the coherent sum of these two fields. The small reflected carrier introduces an asymmetry in the
intensity over its 2 period, which produces a component that varies at . This component is
the beat signal between the carrier and the sidebands, and its sign tells you whether you are above
resonance or below it.
FIG. 11. An approximate plot of D, the slope of the error signal near resonance, versus Ps /Pc .
The optimum value is at Ps /Pc = 1/2, and the maximum is very broad.
FIG. 12. Plots of F in the complex plane for various couplings. Only for the impedance matched
case does the reflected intensity vanish on resonance.
25
1.0
Transmission
0.8
0.6
0.4
0.2
1.5
FIG 1
BLACK
AM. J. PHYS.
2.5
Reflected Intensity
frequency
FIG 2
BLACK
AM. J. PHYS.
Laser
Faraday
Isolator
Pockels Cell
Optical
Isolator
Cavity
Local
Oscillator
Photodetector
Phase
Shifter
Servo
Amp
Low-Pass
Filter
FIG 3
BLACK
Mixer
AM. J. PHYS.
1.0
Intensity
0.8
0.6
0.4
0.2
0.6
0.8
1.2
1.4
0.6
0.8
1.2
1.4
180
120
Phase
60
-60
-120
-180
frequency (free spectral ranges)
FIG 4
BLACK
AM. J. PHYS.
Im{E}
Reflected Beam
Incident Beam
Re{E}
FIG 5
BLACK
AM. J. PHYS.
0.6
0.4
0.2
0.996
0.998
1.002
-0.2
-0.4
-0.6
FIG 6
BLACK
AM. J. PHYS.
1.004
1.0
0.5
0.9
0.95
1.05
-0.5
-1.0
frequency (free spectral ranges)
FIG 7
BLACK
AM. J. PHYS.
1.1
Im{E}
Re{E}
FIG 8
BLACK
AM. J. PHYS.
Im{E}
Re{E}
FIG 9
BLACK
AM. J. PHYS.
Im{E}
Re{E}
FIG 10
BLACK
AM. J. PHYS.
0.5
1.0
Ps /P c
FIG 11
BLACK
AM. J. PHYS.
Im{F}
Overcoupled
Re{F}
Undercoupled
FIG 12
BLACK
Impedance Matched
AM. J. PHYS.