Temperature Sensors Types MBT 5250, 5260 and 5252: Data Sheet
Temperature Sensors Types MBT 5250, 5260 and 5252: Data Sheet
Temperature sensors
types MBT 5250, 5260 and 5252
Features
MBT 5250/5260
For temperature measurement and regulation in piping systems and refrigeration plants
on ships or points where reliable, robust and
accurate equipment is required
Gaseous or liquid media, e.g. air, gas, vapour,
water or oil.
Up to +200C media temperatures
Pt100 or Pt1000 resistance element
Can be used with 2- or 3-wire connections
Gold plated male and female connector
MBT 5250 with interchangeable measuring
insert
MBT 5260 with xed measuring insert
Approvals
- Lloyds Register of Shipping, LR
- Germanischer Lloyd, GL
- Bureau Veritas, BV
- Det Norske Veritas, DNV
- Nippon Kaiji Kyokai, ClassNK
- Registro Italiano Navale, RINA
- American Bureau of Shipping, ABS
- Korean Register of Shipping, KRS
Electrical connection
Insertion length [mm]
Process connection
PG 9
PG 11
PG 13.5
Code no.
Code no.
Code no.
50
G 1/2 A
084Z8011
084Z8036
100
G 1/2 A
084Z8012
084Z8039
150
G 1/2 A
084Z8010
084Z8008
200
G 1/2 A
084Z8022
084Z8043
50
G 3/4 A
084Z8037
084Z8058
100
G 3/4 A
084Z8006
084Z8013
150
G 3/4 A
084Z8041
084Z8014
200
G 3/4 A
084Z8044
084Z8218
50
1/2 14 NPT
084Z8066
80
1/2 14 NPT
084Z8019
100
1/2 14 NPT
084Z8067
150
1/2 14 NPT
084Z8065
200
1/2 14 NPT
084Z8068
INDUSTRIAL CONTROLS
IC.PD.P30.I1.02- 520B1912
Data sheet
Electrical connection
Insertion length [mm]
PG 9
Process connection
PG 11
Code no.
Code no.
50
G 1/2 A
084Z8033
084Z8229
100
G 1/2 A
084Z8021
084Z8132
150
G 1/2 A
084Z8034
084Z8096
200
G 1/2 A
084Z8238
Technical data
Response times
Type
Protection tube
Air 1 m/s
t0.5
t0.9
t0.5
t0.9
8 1 mm
9s
33 s
95 s
310 s
3s
10 s
90 s
300 s
10 2 mm
12 s
42 s
111 s
391 s
4s
14 s
96 s
323 s
Solid drilled
12 s
36 s
220 s
900 s
5s
15 s
210 s
850 s
8 1 mm
2s
6s
82 s
260 s
Solid drilled
4s
13 s
225 s
850 s
Materials
Protection tube in contact with media
Process connection
Extension length
Union
Gasket
Silicone
Vibration stability
t = temperature of
medium, numerical
value
Shock: 100 g in 6 ms
Vibrations: 4g sine function 5 - 200 Hz, measured acc. to IEC 68-2-6
Enclosure
PG 9, PG 11 or PG 13.5
Electrical connections
2-wire, 3 terminals
IC.PD.P30.I1.02- 520B1912
Data sheet
Technical data
L = Insertion length
Tm = Media temperature
Tp = Temperature for electric plug
TA = Ambient temperature
Permissible
media velocity
Air
25 m/s
Water
3 m/s
Process
connection
G 1/4 A G 1/2 A
G 3/8 A M18
G 3/4 A
M24
Max. tightening
torque
50 Nm
100 Nm
Dimensions
Without extension length
Plugs
PG 9, PG 11
Gasket
PG 13.5
Please note:
1)
IC.PD.P30.I1.02- 520B1912
A = Extension length
B = Insertion length
C = Protection tube
D = Process connection 1)
E = Union
Process connection
G 1/4 A
HEX 22
HEX 27
M24 2, G 3/4 A
HEX 32
Data sheet
1)
2)
3)
Dimensions
Extension length
Insertion length
Union nut
Welded
Protection tube
10 x 2
Type: BM
Process connection
Transmitter type MBT 9110, as
terminal block
Connection head, type B
PG 16
Process
connection
HEX 27
G 3/4 A
HEX 32
IC.PD.P30.I1.02- 520B1912
Data sheet
Temperature
range [C]
Insertion
length [mm]
50 to 200
50
80
100
150
200
250
50
80
100
150
200
250
Extension length:
Element:
Connection head:
Protection tube:
50 mm
Pt 100, EN 60751, Class B
B-head
Low temperature range: 10 x 2 mm
High temperature range: 11 x 1 mm
Connection
Transmitter
output
Transmitter
setting
2-wire, 3 terminals
None
None
2-wire
4 - 20 mA,
standard
0 to +100C
G A
Code no.
084Z8210
084Z6140
084Z82111)
084Z82121)
084Z82131)
084Z6139
084Z8214
084Z6142
084Z82151)
084Z82161)
084Z82171)
084Z6143
Process connection
G 3/4 A
-14 NPT
Code no.
Code no.
084Z8230
084Z6165
084Z6164
084Z6166
084Z82311)
084Z61671)
1)
084Z8232
084Z61681)
1)
084Z8233
084Z61691)
084Z6141
084Z6170
084Z8234
084Z6171
084Z6144
084Z6172
084Z82351)
084Z61731)
084Z82361)
084Z61741)
084Z82371)
084Z61751)
084Z6145
084Z6176
1)
Preferred versions
400 C version is available on request
Electrical connection
With temperature transmitter
IC.PD.P30.I1.02- 520B1912
Data sheet
Technical data
Weight
Insertion length
[mm]
Electrical
connection
Process connection
Transmitter
setting
G 1/2 A
G 3/4 A
Weight [g]
Weight [g]
Weight [g]
430
480
430
460
510
460
490
540
490
520
570
520
250
550
600
550
50
420
470
420
450
500
450
480
530
480
510
560
510
540
590
540
50
100
Sensors are
without transmitter
2-wire,
3 terminals
150
200
4 - 20 mA,
2-wire universal
temperature
transmitter
100
150
200
0 -> +100C
250
-14 NPT
11 1, 15 3
8 1, 10 2
L = Insertion lenght
Protection tube
10 2
50 Nm
Permissible
media velocity
Air
25 m/s
Steam
25 m/s
Water
3 m/s
Materials
Protection tube in contact with media
Extension length
Union nut
Connection head
Transmitter:
t = temperature of medium,
numerical value
Shock:
100 g in 6 ms
Vibrations:
Enclosure
PG 16
Cable entry BM
PG 9
Temperature transmitter
MBT 9110
Supply voltage:
8 - 35V d.c.
Output:
4 - 20 mA
1)
Ambient:
The temperature of the temperature transmitter is inuenced by media temperature, ambient temperature and ventilation in the engine room. If the temperature of the temperature transmitter exceeds the max. allowed temperature the
temperature transmitter must be placed in a separated enclosure, as described in the separate data sheet for MBT 9110.
IC.PD.P30.I1.02- 520B1912
Data sheet
MBT programme
Danfoss offers a complete range of MBT temperature sensors for all types of applications. The
range consists of exhaust gas sensors, universal
sensors, bearing sensors, cargo sensors, stern tube
sensors as well as general purpose sensors.
General features
Further information
IC.PD.P30.I1.02- 520B1912
Data sheet
IC.PD.P30.I1.02- 520B1912