1012cei Handpiece Rev9 PDF
1012cei Handpiece Rev9 PDF
1012cei Handpiece Rev9 PDF
3 CE credits
This course was
written for dentists,
dental hygienists,
and assistants.
Anatomy of a Handpiece:
Understanding Handpiece
Maintenance and Repairs
A Peer-Reviewed Publication
Written by Glenn Williams, BS
This course has been made possible through an unrestricted educational grant. The cost of this CE course is $59.00 for 3 CE credits.
Cancellation/Refund Policy: Any participant who is not 100% satisfied with this course can request a full refund by contacting PennWell in writing.
Educational Objectives
The overall goal of this article is to provide the reader with
information on current air-driven and electric handpieces.
Upon completion of this course the reader will be able to do
the following:
1. List and describe the major components of air-driven
high-speed handpieces and electric handpieces
2. List and describe the main differences between air-driven
high-speed handpieces and electric handpieces
3. List and describe the signs of failure for air-driven and
electric handpieces
4. List and describe the maintenance steps for air-driven
handpieces
5. List and describe the options for repair/rebuild of airdriven handpieces and considerations in selecting one.
Abstract
Todays clinician is extremely dependent on the handpiece to
sustain a smooth-running practice. The handpiece is an incredibly sophisticated device that requires a diligent maintenance protocol to keep it running properly and safely; routine,
repeated heat sterilization has the most adverse effect on the
dental handpiece. Perhaps due to this dependency on handpieces, combined with the damage resulting from repeated
routine sterilization and the need for consistent maintenance,
the handpiece has earned an undeserved reputation for excessive breakdowns. The dental team can keep handpieces
functioning smoothly longer, and maximize the return on
the significant investment the dentist has made in handpiece
technology, through appropriate maintenance procedures.
Careful selection from repair options is also required.
Introduction
The dental handpiece of today is a sophisticated combination of precision parts moving in perfect synchronization at
extremely high speed. This combination delivers a smooth,
powerful cut that allows the clinician to remove tooth structure in an efficient manner with as little discomfort to the
patient as possible. In contrast, up until the 1950s all dentistry
was performed with electric belt-driven instruments with top
rotational speeds of 50,000 rpm. This relatively slow speed
increased the time spent on tooth preparation and was also extremely uncomfortable for the patient, who experienced much
pressure and vibration. The belts that ran the handpieces were
heavy and restrictive and tangled with any nearby hair. The
leap from belt-driven to air-driven handpieces in the United
States was made by Dr. John Borden during the 1950s, and
pioneering work for the air-driven high-speed handpiece was
performed by Sir John Walsh in New Zealand.1,2
Figure 1. First-generation air-driven high-speed handpiece
10.2 mm
11.9mm
9.8 mm
Stylus ATC 990
13.0 mm
14.5mm
Stylus ATC 890
Stylus ATC 890
and produce a very quiet hum as opposed to the highpitched whine of an air-driven high-speed handpiece. Fiber
optics and multiport water spray are delivered through the
attachments.7
Figure 5. Electric handpiece unit and attachment examples
Electric Handpieces
The single biggest difference in comparing an electric
handpiece to an air-driven handpiece is its constant speed.
With an electric handpiece, there is no difference between
free-running speed and active speed, due to the amount of
power generated by the electric motor driving the handpiece, and no power is lost while cutting tooth structure.
Another advantage of the new electric handpieces is that
they are significantly quieter than air-driven handpieces.
An electric handpiece consists of an entire system. A control
box must be mounted on the delivery unit, and a conventional delivery tubing plugs into the box, which provides
signal air and water, enabling the motor to be operated from
the foot control already in use. The control box has digital
settings so the operator can literally dial in the exact speed
at which he or she would like the bur to rotate. Lightweight
tubing extends from the control box to the motor. Electric
motors look and feel like cylindrical low-speed air motors
4
Any number of attachments with various gearing combinations will connect onto an electric motor. The most
commonly used attachment for operative dentistry is a 1:5
step-up referred to as a high speed attachment. Most
electric motors operate at 40,000 rpm; adding a 1:5 speedincreasing attachment provides 200,000 rpm at the bur.
This speed remains constant no matter how aggressively
the clinician is cutting, and the advantage is much faster
preparation time. There is a learning curve associated with
mastering this increased power. Electric attachments are
generally universal, meaning that any brand will work with
any motor. One exception is the new Comfort Drive by
Kavo, a more compact design.
There are a few trade-offs to be aware of with electric
handpieces. Together, the attachment and the motor are
similar to working with a low-speed handpiece in your
hand, versus the shorter and lighter high-speed air-driven
devices. This may be of particular concern to operators with
a smaller hand size. On the other hand, many users report
that since the actual cutting time is reduced, this offsets any
limitations based on size and weight. Electric handpieces
have been found to be more efficient than air-driven highspeed handpieces.8 Head size is also typically larger than
with the average high-speed air-driven handpiece. A major
investment is associated with equipping an office with
electric technology, with the cost for a complete system
beginning at around $3,500, and there are additional costs
for each attachment required for sterilization redundancy.
Electric handpieces consist of several complex parts the
attachment is more similar to a gear-driven low-speed than
to an air-driven high-speed handpiece. A series of drive
shafts and transmissions increase the rpm of the handpiece
from 40,000 where it connects to the motor up to 200,000
in the head. This intricate array of gears and bearings
makes repair more costly than for air-driven high-speed
handpieces.
www.ineedce.com
Unlike an air-driven handpiece, which loses torque as the turbine components wear, the electric motor is so powerful that it
will continue to drive the attachment as internal parts fail. Even
after bearings or gears disintegrate, the motor will continue to
operate the handpiece, and this increased friction leads to heat
that can burn a patient. On September 23, 2010, the FDA issued a second safety alert regarding reports of serious patient
tissue damage due to overheated electric attachments during
dental treatment procedures.9 The operator must be aware of
any sensation of heat in the handpiece or the head. It is imperative that you stop using an electric handpiece immediately if
you feel any heat during use. This simple precaution will, most
importantly, prevent patient injury and will also reduce the
cost of repair by not causing further damage to internal parts.
Body or Shell
Brass is the most common material used in handpiece
manufacturing; it is a relatively inexpensive material and
easy to machine, although it is soft and susceptible to denting. Cosmetic protective plating is applied over the brass
but can discolor or flake off over time as a result of sterilization. Stainless steel is another material commonly used to
make handpiece bodies. Steel is lighter and stronger than
brass, but its use results in higher manufacturing costs, so
the prices are generally higher. The current state-of-the-art
in handpiece construction is titanium, which is 40 percent
lighter than stainless steel and stronger and more resistant
to the corrosive effects of autoclaving.
Figure 9. Corrosive
effects of sterilization
1.
2.
3.
4.
5.
6.
7.
www.ineedce.com
Spindle
Chuck
Rear bearing
Impeller
Front bearing
End cap
O-rings
to measure air pressure directly at the handpiece connection (the air pressure gauge on the front of the dental unit is
little more than a relative indicator and does not account for
frictional air loss through the tubing). Some manufacturers
incorporate a valve inside the handpiece that regulates incoming air to prevent excessive pressure from reaching the
turbine. The quality of the drive air can also impact turbine
life. Oil and condensation accumulate in the compressor
tank during normal operation, and without proper drying or
filtration, oil and water may travel through the air lines to the
handpiece tubing. These particles then impact the delicate
turbine at high speed, like marbles tossed against a spinning fan. You can easily check drive air quality by aiming
an empty hose at a mirror or facial tissue. Any water or oil
will show up as moisture or discoloration on the surface. If
you have doubts about your air quality or pressure, contact a
service technician to inspect dryers, filters, and connections
and drain the compressor tank.
Cutting Instruments
The cutting instrument itself can have an impact on turbine life. Overusing dull burs may cause the operator to
exert more lateral force against the tooth structure, thereby
increasing the side load on the bearings and resulting in
excessive friction while operating the handpiece. Using a
surgical-length bur, or simulating one by extending the bur
from the chuck a few millimeters, can significantly affect the
load on the turbine. The longer the cutting instrument, the
greater the negative impact on concentricity. Burs should
be seated fully into an autochuck and never extended. Conversely, short-shanked burs should never be inserted into
a chuck past the taper on the shank. Inferior-quality burs
with variable shank diameters overextend the springs in autochucks, leading to premature failure. It may seem obvious
the nature of the dental practice can also have an impact
on turbine life. A high-volume crown and bridge practice
cannot hope to achieve the same life span from a handpiece
as can a pediatric dentist or periodontist, who typically uses
a handpiece less often and with less force.
Handpiece Lubricants
The simplest lubricator design holds only one handpiece, and the operator must remain at the station to hold
the button down during the entire process. This ensures
a complete cycle of lubrication and flushing, but there
is no time savings for staff. Fully automatic designs will
hold more than one handpiece, including low-speed attachments. Similar to using an autoclave, the operator
loads the machine, closes the door, and presses a button.
This begins the maintenance cycle, and the machine will
dispense the correct amount of lubricant and then operate
each handpiece at the correct running speed. During operation the staff member is free to complete other duties while
the machine runs multiple handpieces through the maintenance cycle. While the automatic machines initially cost
more, there is absolute consistency of operation, which may
result in longer handpiece life and reduced repair costs.
Additional savings may accrue while the staff member
completes normal between-patient tasks and does not need
to use valuable time performing routine handpiece maintenance procedures. One area of concern that is not often
addressed is chuck maintenance. According to a handpiece
manufacturer product manager, Ninety-five percent of all
turbine or attachment problems are found in the head of
the handpiece. Everyone always seems to lubricate the back
end, maybe purge, but do not address the head of the handpiece or the chuck.15 When a handpiece is in operation, the
air is being expelled outward. Once the clinician takes his
or her foot off the rheostat the physics reverse, which can
bring contaminants and tooth dust back into the head of
the handpiece. If these particles are not purged, they bake
onto the internal components and will cause premature
failure. By shooting lubricant and compressed air into the
chuck, this will help dislodge the debris or at least loosen it
up. Once you run a purge cycle by attaching it to the tubing
or maintenance system, the particles are much more likely
to be expelled from the handpiece, extending longevity.
Turbine Repair/Rebuild
Rebuilding a turbine instead of replacing it saves money. As
handpiece repair costs escalated, dentists increasingly turned
to local independent repair services to extend turbine life, by
rebuilding as opposed to replacing turbines. Turbine rebuilding services have earned greater acceptance throughout the
dental industry over the years. According to a December 2003
survey of workhorse systems by Dental Products Report, 59
percent of dentists employ a dedicated handpiece repair shop
to maintain their handpieces.16 One handpiece manufacturer
has now adopted this lower-cost option when they have to
replace a turbine under warranty. At a high-quality independent repair service, a technician will evaluate the turbine to
www.ineedce.com
dentist of the substitution. This is unethical. A handpiece is a sophisticated device and can easily be compromised with improper techniques and inferior parts.
Watch out for nebulous claims such as same as manufacturers specifications or manufacturer-quality
parts. There should be no room for interpretation. If
a repair shop claims to supply original turbines, then
original packaging and instructions should accompany
these installations.
4. Relationship with the repair service: The best service results from dealing directly with the technician
who performs the repair, not with customer service or
voicemail. Dental professionals need to establish a trusting relationship with someone who can easily articulate
exactly what services are being provided.
5. Obtaining estimates: Repair options and costs should
be spelled out clearly. Communication is vital in order
for you to be sure you are receiving exactly what you are
paying for. Be wary of companies that do not provide
published pricing and of repair contracts promising to
service handpieces for a fixed fee (which result in an incentive to replace as few parts as possible instead of performing a complete overhaul of the turbine if required).
Some companies offer clean and lube services at prices
up to $45. This can be done in the office by the staff instead, at no cost and until a high-speed handpiece actually fails, daily staff maintenance is all that is required to
prolong its life.
6. Get a warranty: A quality repair operation should
provide a substantial warranty on all repairs. Another
indicator of a qualified service is the ability to facilitate
and provide original factory repairs if these are necessary
(not all handpiece repairs can be performed correctly in
the field).
7. Honest advice and commitment: An ethical service
provider should inform the office when total replacement
makes more sense. A good repair service holds a sincere
interest in serving the needs of dental professionals and
should be committed to improving the overall industry.
Look for a company that belongs to a trade association
and supports local dental society meetings and events.
Summary
Handpieces are one of the most important workhorse systems in the dental practice and require routine autoclaving,
yet they have an often undeserved reputation for excessive
breakdown and expense. If you maintain your handpieces
properly, you will get many years of optimal performance;
and, when the time comes to repair your handpiece, consideration should be given to the quality of the repair and
reliability of a qualified service provider to restore your
workhorse to optimal performance. The bottom line is that
handpieces represent a significant investment and play a
vital role in any dental practice.
www.ineedce.com
References:
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Author Profile
Glenn Williams, BS
Glenn Williams is a twenty-five year veteran of
the dental industry, specializing in handpieces
since 1987. From 1987-1994, Glenn worked for
a major handpiece manufacturer. He has owned
and operated a successful handpiece repair service in Northern California since 1994. Glenn
was one of the founding members, and is President, of the National Dental Handpiece Repair Association. The
NDHRA is an association of other dedicated independent handpiece professionals who desire to better serve the dental profession
through advancing the handpiece repair industry by establishing
and maintaining high standards of service.
Disclaimer
The author of this course owns and operates an independent repair service
and is the President of the National Dental Handpiece Repair Association.
Reader Feedback
Online Completion
Use this page to review the questions and answers. Return to www.ineedce.com and sign in. If you have not previously purchased the program select it from the Online Courses listing and complete the
online purchase. Once purchased the exam will be added to your Archives page where a Take Exam link will be provided. Click on the Take Exam link, complete all the program questions and submit your
answers. An immediate grade report will be provided and upon receiving a passing grade your Verification Form will be provided immediately for viewing and/or printing. Verification Forms can be viewed
and/or printed anytime in the future by returning to the site, sign in and return to your Archives Page.
Questions
1. Pioneering work for the air-driven
high-speed handpiece was performed by
_______.
a.
b.
c.
d.
a smoother cut
better margins
less vibration
all of the above
hybrid handpiece
air-driven handpiece
electric handpiece
all of the above
a.
b.
c.
d.
a.
b.
c.
d.
1960s
1970s
1980s
1990s
features
reliability
maintenance requirements
all of the above
improves visibility
improves access
improves seizing
a and b
a.
b.
c.
d.
a.
b.
c.
d.
resistance or drag
rotation
pressure
all of the above
www.ineedce.com
a.
b.
c.
d.
a.
b.
c.
d.
12.5 degrees
17.5 degrees
22.5 degrees
27.5 degrees
a.
b.
c.
d.
38 to 51 decibels
48 to 61 decibels
58 to 71 decibels
none of the above
1:3 step-up
1:5 step-up
1:7 step-up
any of the above
13
Questions (Continued)
16. It is imperative that you stop using an
electric handpiece immediately if you
_______.
a.
b.
c.
d.
Lubrication
Conditions of use
Maintenance practices of the individual office
All of the above
250
500
750
none of the above
a.
b.
c.
d.
a.
b.
c.
d.
no load
load
activation
a and c
a chucking mechanism
a vibratory system
a rotary system
a and c
chuck
spindle
turbine bearings
all of the above
a.
b.
c.
d.
a. is redundant
b. is necessary
c. may multiply harmful reactions when the
handpiece is subjected to heat
d. a and c
a.
b.
c.
d.
a.
b.
c.
d.
glass fiber
cellular
chip
all of the above
bacterial debris
rust and corrosion
lysed sterilization products
all of the above
impeller
turbine
chuck
none of the above
14
ethanol; 20 seconds
ethanol; 30 seconds
water; 20 seconds
water; 30 seconds
get a warranty
ensure the technician has adequate training
make sure the repair service is using the correct bearings
all of the above
ABEC 3
ABEC 5
ABEC 7
ABEC 9
a.
b.
c.
d.
49 percent
59 percent
69 percent
none of the above
a.
b.
c.
d.
www.ineedce.com
ANSWER SHEET
Title:
Address:
E-mail:
City:
State:
Telephone: Home (
Office (
Specialty:
ZIP:
Country:
Requirements for successful completion of the course and to obtain dental continuing education credits: 1) Read the entire course. 2) Complete all
information above. 3) Complete answer sheets in either pen or pencil. 4) Mark only one answer for each question. 5) A score of 70% on this test will earn
you 3 CE credits. 6) Complete the Course Evaluation below. 7) Make check payable to PennWell Corp. For Questions Call 216.398.7822
Educational Objectives
1. List and describe the major components of air-driven high-speed handpieces and electric handpieces.
2. List and describe the major differences between air-driven high-speed handpieces and electric handpieces.
3. List and describe the signs of failure for air-driven and electric handpieces.
Course Evaluation
Please evaluate this course by responding to the following statements, using a scale of Excellent = 5 to Poor = 0.
1. Were the individual course objectives met?
Yes
No
Yes
No
31.
32.
33.
34.
35.
36.
37.
38.
39.
40.
41.
42.
43.
44.
45.
46.
47.
48.
49.
50.
10. If any of the continuing education questions were unclear or ambiguous, please list them.
___________________________________________________________________
11. Was there any subject matter you found confusing? Please describe.
___________________________________________________________________
___________________________________________________________________
12. What additional continuing dental education topics would you like to see?
___________________________________________________________________
___________________________________________________________________
If not taking online, mail completed answer sheet to
www.ineedce.com
INSTRUCTIONS
All questions should have only one answer. Grading of this examination is done
manually. Participants will receive confirmation of passing by receipt of a verification
form. Verification forms will be mailed within two weeks after taking an examination.
EDUCATIONAL DISCLAIMER
The opinions of efficacy or perceived value of any products or companies mentioned
in this course and expressed herein are those of the author(s) of the course and do not
necessarily reflect those of PennWell.
Completing a single continuing education course does not provide enough information
to give the participant the feeling that s/he is an expert in the field related to the course
topic. It is a combination of many educational courses and clinical experience that
allows the participant to develop skills and expertise.
COURSE CREDITS/COST
All participants scoring at least 70% on the examination will receive a verification
form verifying 3 CE credits. The formal continuing education program of this sponsor
is accepted by the AGD for Fellowship/Mastership credit. Please contact PennWell for
current term of acceptance. Participants are urged to contact their state dental boards
for continuing education requirements. PennWell is a California Provider. The California
Provider number is 4527. The cost for courses ranges from $39.00 to $110.00.
Many PennWell self-study courses have been approved by the Dental Assisting National
Board, Inc. (DANB) and can be used by dental assistants who are DANB Certified to meet
DANBs annual continuing education requirements. To find out if this course or any other
PennWell course has been approved by DANB, please contact DANBs Recertification
Department at 1-800-FOR-DANB, ext. 445.
RECORD KEEPING
PennWell maintains records of your successful completion of any exam. Please contact our
offices for a copy of your continuing education credits report. This report, which will list
all credits earned to date, will be generated and mailed to you within five business days
of receipt.
CANCELLATION/REFUND POLICY
Any participant who is not 100% satisfied with this course can request a full refund by
contacting PennWell in writing.
2010 by the Academy of Dental Therapeutics and Stomatology, a division
of PennWell
ANAT0111C
15