BlueShadow UV/VIS Detector 40D

Variable single wavelength


Article Nr.: C55150XA.0

The BlueShadow detector 40D is a competitively priced HPLC spectrophotometer for routine HPLC applications
including fast LC methods. Besides offering excellent technical specifications, this robust detector features a
highly flexible and compact design.
The BlueShadow Detector 40D comes with an installed deuterium lamp which covers a wavelength range from
190 to 750 nm. This detector can be controlled with ClarityChrom software, as well as from the touchscreen
(standalone operation), via LAN, via RS232, or through analog input/output.
Due to a smart design the flow cell is easily accessible and can be changed very quickly. You can choose
between a wide range of flow cells for analytical or preparative LC applications with flow rates from 10 l/min
up to 10 l/min.
Key features
variable single wavelength
large choice of flow cells
wavelength range from 190 750 nm
flow rates from 10 l/min up to 10 l/min

Details
Detection
Detector type

Variable single wavelength UV/VIS detector

Detection channels

Light source

Deuterium (D2) lamp with integrated GLP chip

Wavelength range

190 750 nm

Spectral bandwidth

11 nm at H line (FWHM)

Wavelength accuracy

2.5 nm

Wavelength precision

0.3 nm (ASTM E27593)

Noise

1.5 x 105 AU at 254 nm (ASTM E165798)

Drift

3.0 x 104 AU/h at 254 nm (ASTM E165798)

Linearity

> 2.0 AU at 270 nm (ASTM E165798)

Time constants

0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s

Flow cell

Not included

Integration time

Automatic

Communication
Maximum data rate

50 Hz (LAN)
20 Hz (Analog)
10 Hz (RS232)

Inputs

Error (IN), Start (IN), Autozero, 0 10 V Analog IN

Outputs

Events 1 3, + 5 V, 24 V Valve

Analog outputs

1 x 0 5 V scalable, 20 bit, offset adjustable

Control

Digital: LANDHCP, RS232, remote connector


Analog: wavelength control
Manual: touchscreen

Programming

Timed: wavelength, events, fraction valve, links, wake up (program, link);


9 programs, 50 program lines

Technical parameters
GLP

Detailed report including lamp recognition, operating hours,


lamp operating hours, number of lamp ignitions

Display

Touchscreen TFT 2.4"

Ambient conditions

Temperature range: 4 40 C, 39.2 104 F


Humidity: below 90 % noncondensing

General
Power supply

100 240 V, 50 60 Hz, 75 W

Dimensions

242 x 169 x 399 mm (W x H x D)

Weight

5.3 kg

Downloads

Blue Shadow Detector 40D


Manual

BlueShadow Detector 40D, V7640A

English

BlueShadow Detektor 40D


Manual

BlueShadow Detector 40D, V7640

Deutsch

BlueShadow: HPLC Products for OEM customers


only

English

Certificate Lamp Replacement, V6836A

English

2014/05

English

BlueShadow Memo Card

Certificate Lamp Replacement

SN0131:UV detectors

Related products

BlueShadow UV Detector 10D

190 500 nm, small variable single wavelength UV


detector

C55120.0

BlueShadow UV Detector 10D


Fiber Optics

190 500 nm, small variable single wavelength UV


detector with fiber optic connectors

C55122.0

BlueShadow UV/VIS Detector


40D Fiber Optics

190 750 nm, variable single wavelength UV/VIS


detector with fiber optic connectors

C55152XA.0

BlueShadow UV/VIS Detector


50D

190 750 nm, variable multiple wavelength UV/VIS


detector

C55190XA.0

Technical data are subject to change without notice.


All trade names and registered trademarks are the property of their respective owners.