Doc316 53 01113
Doc316 53 01113
Doc316 53 01113
(Orthophosphate)
Amino Acid Method1
DOC316.53.01113
Method 8178
Reagent Solution
Adapted from Standard Methods for the Examination of Water and Wastewater.
Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows sample cell and orientation requirements for reagent addition tests, such as
powder pillow or bulk reagent tests.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information
Instrument
Sample cell
DR 6000
2495402
DR 3800
DR 2800
DR 2700
DR 1900
DR 5000
DR 3900
DR 900
2401906
Before starting
Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
For the best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water
in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results
automatically with the reagent blank adjust option.
The contents of one Amino Acid Reagent Powder Pillow can be used as an alternative to the 1 mL of Amino Acid Reagent
Solution in the test procedure.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description
Quantity
1 mL
Molybdate Reagent
1 mL
Sample cells (For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1.)
Collect samples in clean glass or plastic bottles that have been cleaned with 6 N (1:1)
hydrochloric acid and rinsed with deionized water.
Do not use a detergent that contains phosphate to clean the sample bottles. The
phosphate in the detergent will contaminate the sample.
Analyze the samples as soon as possible for best results.
If immediate analysis is not possible, immediately filter and keep the samples at or
below 6 C (43 F) for a maximum of 48 hours.
Let the sample temperature increase to room temperature before analysis.
Start
Zero
Read
Interferences
Interfering
substance
Interference level
Calcium
Chloride
Colored samples
Add 1 mL of 10 N Sulfuric Acid Standard Solution to another 25-mL sample. Use this instead of
untreated sample as the blank to zero the instrument. Use a pipet and pipet filler to measure the
sulfuric acid standard.
May cause low results. To eliminate this interference, dilute the sample until two successive dilutions
give about the same result.
Interfering
substance
Interference level
Magnesium
Nitrite (NO2)
Bleaches the blue color. Remove nitrite interference by adding 0.10 g of sulfamic acid to 50 mL
sample. Swirl to mix. Use this treated sample in the test procedure.
Phosphates, high
levels (PO43)
As the concentration of phosphate increases, the color changes from blue to green, then to yellow
and finally to brown. The brown color may suggest a concentration as high as 100,000 mg/L PO43. If
a color other than blue is formed, dilute the sample and retest.
Sulfide (S2)
Sulfide interferes. For samples with a sulfide concentration less than 5 mg/L, remove sulfide
interference as follows:
1.
2.
3.
Temperature
Turbidity
May give inconsistent results for two reasons. Some suspended particles may dissolve because of
the acid used in the test. Also, desorption of orthophosphate from particles may occur. For highly
turbid samples, add 1 mL of 10 N Sulfuric Acid Standard Solution to another 25-mL sample. Use this
instead of untreated sample as the blank to zero the instrument. Use a pipet and pipet filler to
measure the sulfuric acid standard.
Highly buffered
samples or extreme
sample pH
Can prevent the correct pH adjustment of the sample by the reagents. Sample pre-treatment may be
necessary.
Accuracy check
Standard additions method (sample spike)
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
Items to collect:
1. Use the test procedure to measure the concentration of the sample, then keep the
(unspiked) sample in the instrument.
2. Go to the Standard Additions option in the instrument menu.
3. Select the values for standard concentration, sample volume and spike volumes.
4. Open the standard solution.
5. Prepare three spiked samples: use the TenSette pipet to add 0.1 mL, 0.2 mL and
0.3 mL of the standard solution, respectively, to three 25-mL portions of fresh sample.
Mix well.
6. Use the test procedure to measure the concentration of each of the spiked samples.
Start with the smallest sample spike. Measure each of the spiked samples in the
instrument.
7. Select Graph to compare the expected results to the actual results.
Note: If the actual results are significantly different from the expected results, make sure that
the sample volumes and sample spikes are measured accurately. The sample volumes and
sample spikes that are used should agree with the selections in the standard additions menu. If
the results are not within acceptable limits, the sample may contain an interference.
1. Use the test procedure to measure the concentration of the standard solution.
2. Compare the expected result to the actual result.
Note: The factory calibration can be adjusted slightly with the standard adjust option so that the
instrument shows the expected value of the standard solution. The adjusted calibration is then
used for all test results. This adjustment can increase the test accuracy when there are slight
variations in the reagents or instruments.
Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users can get different
results under different test conditions.
Program
Standard
Sensitivity
Concentration change per 0.010 Abs change
485
Summary of method
In a highly acidic solution, ammonium molybdate reacts with orthophosphate to form
molybdophosphoric acid. This complex is then reduced by the amino acid reagent to yield
an intensely colored molybdenum blue compound. The measurement wavelength is
530 nm for spectrophotometers or 520 nm for colorimeters.
Quantity/test
Unit
Item no.
100 tests
2244100
1 mL
100 mL MDB
193432
Molybdate Reagent
1 mL
100 mL MDB
223632
Quantity/test
Unit
Item no.
each
189640
Unit
Item no.
946 mL
1420416
16/pkg
1424220
Wastewater Effluent Standard Solution, Mixed Parameter, for NH3-N, NO3-N, PO43,
COD, SO42, TOC
500 mL
2833249
Wastewater Influent Standard Solution, Mixed Parameter, for NH3-N, NO3-N, PO4,
COD, SO4, TOC
500 mL
2833149
Required apparatus
Description
Mixing cylinder, graduated, 25-mL, glass stopper
Description
Unit
Item no.
4L
27256
each
2484600
Unit
Item no.
100/pkg
80499
29 mL
221120
each
50543
500 mL
88449
29 mL
211220
each
234401
1000 mL
93153
Water, deionized
each
1970001
50/pkg
2185696
1000/pkg
2185628
100/pkg
2601300
100/pkg
69257
each
108367
each
2635700
12/pkg
2087076
Unit
Item no.
each
2196800
946 mL
2059716
100 mL
1424342
946 mL
1436716
16/pkg
17110
100 mL
1436832
16/pkg
1424210
100 mL
1424232
Optional standards
Description
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
01/2014, Edition 9