Accessories
Mounting Accessories
Type 84...
85...
for Acceleration Sensors
The most important consideration when installing an accelerometer on a test structure is to have a quality mounting
surface for attachment. The primary goal is to achieve a maximum useable frequency response for performing a vibration
measurement. Since the mounting surface conditions affect
transmissibility, a variety of adhesive mounting pads, clips,
magnetic bases, studs and triaxial cubes with accurately pre-
Probe
Magnetic base
with foil
Mounting Method
Magnetic base
Accessory Type
8400_000-281e-06.14
Probe
pared mounting surfaces are available for use in a wide range
of applications. When frequencies higher than 2 3 kHz are
considered, each accelerometer instruction manual provides
specific recommendations for surface quality, sensor orientation and mounting torque. Frequency response considerations
are detailed in [2]. Note: All unit measurements are provided
in mm.
Adhesive mounting
base
Direct adhesive
mounting
Advantages
Disadvantages
Quick measurement
Lower accuracy
Isolated stud
mounting base
Stud mounting
Remarks
Magnetic base
8450A, 8452A, 8456,
KIG4662Bx, 8458,
8466K03, ...
Easy & quick installation
Adds considerably to
mass loading, lowers
resonant frequency
Limited to ferromagnetic
materials
Adhesive mounting
base
8440K0x, 8466K01,
8434, 8436, 8438, ...
Allows stud mounting,
provides electrical isolation
Lowers resonant frequency
Pads are usually exposed to test specimen
Adhesive wax
8432 Petrowax, bees wax
Ideal for lightweight units
Limited temperature
range; not suitable for
larger sensors
Fast & clean
Adhesive cement
Adhesive, Permabond
(cyanoacrylate), hot melt,
Epoxy
Good, strong coupling of
sensor to specimen; higher temperature capability
than wax
Difficult to remove sensor;
requires solvents, removal
tool and/or heat
Suited for more permanent applications & high
frequency measurements
Isolated mounting stud
8400K0x
Allows both strong coupling & ground isolation
Requires threaded hole
in specimen; base slightly
lower resonance frequency
Control mounting
torque; use Silicone
grease
Mounting stud
8402, 8404, 8406, 8410,
8411, 8416, 8418, 8421,
8451, 8453
Best coupling of sensor to
test specimen for highest
frequency response
Requires threaded hole in
specimen
Control mounting
torque; use Silicone
grease
Page 1/7
This information corresponds to the current state of knowledge. Kistler reserves
the right to make technical changes. Liability for consequential damage resulting
from the use of Kistler products is excluded.
2006 ... 2014, Kistler Group, Eulachstrasse 22, 8408 Winterthur, Switzerland
Tel. +41 52 224 11 11, Fax +41 52 224 14 14, [email protected], www.kistler.com
Kistler is a registered trademark of Kistler Holding AG.
Mounting Accessories For Acceleration Sensors
Mounting Stud & Stud Converter
Many accelerometers are specifically designed for stud mounting. The studs on many types are removable allowing both
stud and adhesive mounting. Stud mounting is the preferred
mounting of an accelerometer to test a structure. With properly prepared mounting surfaces and mounting torque correctly observed, it yields the broadest useable frequency range.
Recommended mounting torque is provided in each particular
sensor datasheet or instruction manual. In addition to each
accelerometer instruction manual, a specific technical paper [1]
outlines all the important features for mounting acelerometers.
Types 8402 ... 8453 are used to form a rigid connection between the accelerometer and the test object and are machined
from Beryllium Copper or Stainless Steel for high strength and
high modulus of elasticity coupled with high elastic limits. They
are available in different thread sizes; with English and Metric
threads at opposite ends or as a thread converting insert.
Caution: Always ensure that the stud does not contact the
bottom of the threaded hole. Always follow the recommended
mounting torque provided for each accelerometer in its corresponding instruction manual to prevent any damage to the
sensor base.
Mounting Studs
Type
THD. X
THD. Y
Material
Recommended Sensor Types
8402
7,1
2,5
2,5
10-32
10-32
BeCu
8290, 8795, 8002K, 8202, 8702, 8704,
8703, 8705, 8784, 8786, 8395, 8762, 8770
8404
7,1
2,5
2,5
10-32
10-32
17-4 PH
8044
8406
5,8
2,0
2,0
10-32
10-32
BeCu
8076K
Type
THD. X
THD. Y
Material
Recommended Sensor Types
8410
6,4
3,3
2,0
-28
10-32
BeCu
8076K, 8203, 8712A, 8784, 8786
8411
10,4
6,6
2,8
M6
10-32
BeCu
8290, 8795, 8202, 8702, 8704, 8703,
8705, 8784, 8786, 8762, 8770, 8002K
8416
6,6
3,3
2,3
10-32
5-40
316 St. Stl.
8763, 8766A250/500
8418
7,1
3,8
2,3
M6
5-40
316 St. Stl.
8763, 8766A250/500
8421
12,2
7,5
3,0
M8
-28
BeCu
8203A, 8712A
8430K03
6,9
3,6
2,3
10-32
6-32
BeCu
8766A50, 8766A050/100
8451
8,8
5,1
2,8
M5
10-32
BeCu
8688, 8290, 8795, 8202, 8702, 8704,
8703, 8705, 8762, 8784, 8786, 8770,
8002K
8453
9,8
3,8
5,1
-28
M5
BeCu
8712
Stud Converters
8400_000-281e-06.14
Type
THD. X
THD. Y
Material
Recommended Sensor Types
8414
9,5
8,1
-28
10-32
17-4 PH St. Stl.
8076K, 8484
8414E
9,5
8,1
-28
M5
17-4 PH St. Stl.
8484
5,5
3,4
10-32
5-40
17-4 PH St. Stl.
8486
5,5
3,4
10-32
M3
17-4 PH St. Stl.
Y (THD)
Type
THD. X
THD. Y
Material
Recommended Sensor Types
8412
9,5
-28
Hex
18-8 St. Stl.
8712A, 8076K
8763
8420
9,5
5-40
Hex
18-8 St. Stl.
8414M03
8,9
-28
4-40
VascoMax 300
VascoMax 300 is a registered trademark of Teledyne Vasco.
Page 2/7
This information corresponds to the current state of knowledge. Kistler reserves
the right to make technical changes. Liability for consequential damage resulting
from the use of Kistler products is excluded.
2006 ... 2014, Kistler Group, Eulachstrasse 22, 8408 Winterthur, Switzerland
Tel. +41 52 224 11 11, Fax +41 52 224 14 14, [email protected], www.kistler.com
Kistler is a registered trademark of Kistler Holding AG.
Mounting Accessories For Acceleration Sensors
Isolated Mounting Pad
Mounting pads provide ground isolation for the accelerometer.
They are attached to the test surface with adhesive or with a
screw or stud on surfaces less than ideal for direct mounting of
the accelerometer. Pads are often used in applications where
multiple locations are to be measured using only one accelerometer. Recommended mounting torque is provided in each
particular sensor data sheet or instruction manual.
Type
THD. X
Material
Recommended Sensor Types
8434
4,8
12,4
11,2
5-40
Al. Anod.
8730, 8763
8436
4,8
15,7
14,2
10-32
Al. Anod.
8202, 8203, 8274, 8702, 8703, 8704,
8705, 8774, 8784, 8786, 8766A50,
8795, 8766A250/500/050/100
8438
7,9
21,1
19,1
-28
Al. Anod.
8076K
Type
THD. X
Material
Recommended Sensor Types
800M157
2,5
6,4
11,1
10-32
Al. Anod.
8688A
800M158
2,5
5,1
9,4
5-40
Al. Anod.
8640A
8440K01
5,2
8,0
12,7 Hex
5-40
Al. Anod.
8763A, 8766A250/500
8440K02
5,7
9,0
19,1 Hex
6-32
Al. Anod.
8766A50
C
A
8440K03
5,0
8,3
14,3 Hex
10-32
Al. Anod.
8702, 8703, 8704, 8705
8440K04
5,0
8,3
15,7 Hex
6-32
Al. Anod.
8766A050/100
8466K01
6,4
10,2
22,2 Hex
10-32
Al. Anod.
8395A
Type
THD. X
THD. Y
Material
Recommended Sensor Types
8400K01
3,4
11,6
12,7 Hex
10-32
10-32
Al. Anod.
8702, 8703, 8704, 8705, 8784,
8786, 8795
8400K02
6,0
12,4
19,1 Hex
10-32
6-32
Al. Anod.
8766A50
8400K03
5,5
12,8
19,1 Hex
10-32
M6
Al. Anod.
8688, 8702, 8703, 8704, 8705,
8784, 8786, 8795
8400K04
5,2
12,3
12,7 Hex
5-40
M6
Al. Anod.
8766A250/500, 8763
8400K05
5,9
13,3
19,1 Hex
6-32
M6
Al. Anod.
8766A50
8400K06
5,3
11,4
12,7 Hex
10-32
5-40
Al. Anod.
8766A250/A500, 8763
Type
THD. X
THD. Y
Material
Recommended Sensor Types
8466K02
6,4
8,9
22,2 Hex
10-32
10-32
Al. Anod.
8395
X
A
C
Y
D
X
C
A
D
Y
B
8400_000-281e-06.14
Type
A (thickness)
THD. X
THD. Y
Material
Recommended Sensor Types
8464K01
7,6
21,6
25,4
4-40
Al. Anod.
8315A
8464K02
7,6
21,6
25,4
4-40
10-32
Al. Anod.
8315A
Type
A (thickness)
Diam. X
Material
Recommended Sensor Types
800M144
4,8
15,9
15,9
4-40
Al. Anod.
8793A, 8794A
X
C
Page 3/7
This information corresponds to the current state of knowledge. Kistler reserves
the right to make technical changes. Liability for consequential damage resulting
from the use of Kistler products is excluded.
2006 ... 2014, Kistler Group, Eulachstrasse 22, 8408 Winterthur, Switzerland
Tel. +41 52 224 11 11, Fax +41 52 224 14 14, [email protected], www.kistler.com
Kistler is a registered trademark of Kistler Holding AG.
Mounting Accessories For Acceleration Sensors
Triaxial Mounting Cubes
Allows for two or three single-axis accelerometers to be precisely mounted to perform biaxial or triaxial acceleration meas-
urements. Cube attachment to the test surface can either be
by stud or adhesive for some types. Accelerometer attachment
to the cube can be in a similar manner.
Type
THD. X
Wt (grams)
Material
Recommended Sensor Types
8502
25,4
25,4
25,4
25,4
10-32
117
303 St. Stl.
8202, 8702, 8703, 8704, 8705,
8002K
8504
14,5
14,5
14,5
14,0
10-32
20
303 St. Stl.
8044, 8742A, 8743A
8506
28,7
28,7
28,7
29,2
-28
158
303 St. Stl.
8203
8510
14,5
14,5
14,5
14,2
5-40
19
316 St. Stl.
8730
8514
17,3
17,3
17,3
18,3
10-32
35
303 St. Stl.
8202, 8702, 8704, 8774
8524
11,2
11,2
11,2
10-32
2.8
Al. Anod.
8774A, 8274
8526
11,2
11,2
11,2
2.8
Al. Anod.
8776A, 8276
Type
THD. X
8522
26,9
26,9
26,9
15,0
4-40
Weight
Material
(grams)
Mounting
Screws
28
4-40 x 7/16"
Al. Anod.
8315A
Al. Anod.
8330B
Recommended
Sensor Types
A
10-32 x 1/2"
8530K01
39,6
39,6
39,6
25,4
4-40
74
4-40 x 7/16"
-28 x 1/2"
Interface Adapter Plate
Type
A (thickness)
Diam. X
Diam. Y
Material
Recommended Sensor Types
8466K04
4,6
25,4
43,9
3,2
5,2
Al. Anod.
Adapt 8395A to old Type 8393
B
mounting hole pattern
Type
A (thickness)
THD. X
THD. Y
Diam. Z
Material
Recommended Sensor Types
8464K04
5,7
27,0
31,7
10-32
4-40
3,2
Al. Anod.
Adapt 8315A to old Types
8305/8310/8312 mounting
pattern with 10-32 stud
Magnetic Mounting Base
These offer a convenient, temporary attachment to a ferromagnetic test structure. A notched mounting unit is available
for mounting on curved surfaces. Magnetic bases find use in
8400_000-281e-06.14
Type
8450A
7,6
12,7
11,2
THD. X
5-40
Y
C
X Z
applications requiring vibration monitoring of large rotating
machinery. Note that usage of some mounting bases can add
considerable mass and reduce high frequency response.
Holding
Force
(N)
Weight
Max.
temp.
(C)
Material
Recommended Sensor Types
(grams)
26,7
1,25
170
17-4 St. Stl.
8763, 8730
8452A
11,2
17,8
15,9
10-32
55
19
260
17-4 St. Stl.
KIG4662
B-4
10,9
18,0
12,7
10-32
55
17
80
St. Stl.
8274, 8702, 8703, 8704,
8705, 8774, 8784, 8763,
8202, 8786, 8290,
8766A50/050/100/250/500
KIG4662
B-1
10,9
18,0
12,7
6-32
55
17
80
St. Stl.
8714B
KIG4662
B-5
9,9
11,9
9,9
M2.5
55
80
St. Stl.
8765, 8705
KIG4662
B-6
5,8
9,4
7,1
5-40
20
80
St. Stl.
8730, 8640
Page 4/7
This information corresponds to the current state of knowledge. Kistler reserves
the right to make technical changes. Liability for consequential damage resulting
from the use of Kistler products is excluded.
2006 ... 2014, Kistler Group, Eulachstrasse 22, 8408 Winterthur, Switzerland
Tel. +41 52 224 11 11, Fax +41 52 224 14 14, [email protected], www.kistler.com
Kistler is a registered trademark of Kistler Holding AG.
Mounting Accessories For Acceleration Sensors
Magnetic Mounting Base (continued...)
Type
THD. X
Holding
Force
(N)
Weight
Max.
temp.
(C)
Material
(grams)
Recommended Sensor Types
X
8456
11,2
24,9
-28
135
57
170
17-4 PH
St. Stl.
8203
KIG4662
B-3
14,0
18,0
10-32
50
16
80
St. Stl.
8702, 8705
KIG4662
B-2
14,0
18,0
-28
50
16
80
St. Stl.
8752
B
Type
THD. X
Holding
Force (N)
Weight
102
8458
26,9
47,2
-28
180
Type
THD. X
8466K03
6,4
8,9
24,6 Hex 10-32
Type
THD. X
Material
Recommended Sensor Types
(grams)
17-4 PH St. Stl. 8203, 8712
Holding
Force
(N)
Material
Recommended Sensor Types
100
303 St. Stl.
8395A
Holding
Force
(N)
Material
X
C
A
D
Recommended Sensor Types
800M159
2,5
6,3
11,1
10-32
40
17-4 PH
St. Stl.
8688A
800M160
2,5
5,1
9,4
5-40
30
17-4 PH
St. Stl.
8640A
C
A
D
Type
A (thickness)
THD.X
Holding
Force (N)
Material
Recommended Sensor
Types
8464K03
7,6
21,6
25,4
4-40
100
17-4 PH
8315A
St. Stl.
A
B
8400_000-281e-06.14
Page 5/7
This information corresponds to the current state of knowledge. Kistler reserves
the right to make technical changes. Liability for consequential damage resulting
from the use of Kistler products is excluded.
2006 ... 2014, Kistler Group, Eulachstrasse 22, 8408 Winterthur, Switzerland
Tel. +41 52 224 11 11, Fax +41 52 224 14 14, [email protected], www.kistler.com
Kistler is a registered trademark of Kistler Holding AG.
Mounting Accessories For Acceleration Sensors
Mounting Clips
Provide a convenient way to install single-axis, cube shaped
accelerometers on a test structure. Attached using adhesive,
they find use in large multichannel modal measurements.
Type
Material
Recommended Sensor Types
8474
19,6
17,8
18,5
Engineering
thermoplastic
8772A
Type
Material
Recommended Sensor Types
800M156
16,5
16,5
Polycarbonate
8640A
800M155
20,1
20,1
Polycarbonate
8688A
Direct Adhesive Mounting
Some Kistler accelerometers are specifically designed for adhesive mounting (Types 8278, 8776) and require no special
mounting adapters. Units furnished with stud holes can also be
used with adhesives.
Mounting Petrowax
Bee's wax has been used as a mounting agent for many years.
The recommended Petro-Wax (Type 8432) is a good replacement for bee's wax since it has been formulated to provide
improved frequency response. Wax is a good mounting agent
for lightweight sensors in temporary installations where near
room temperatures are encountered.
Cyanoacrylate, Super Glues
Liquid, low viscosity: These work best on smooth, flat and
non-porous surfaces.
Gels: These can be used on rough and porous surfaces, like
castings. The high viscosity prevents these adhesives from migrating. Operating temperatures range between 80 C and
100 C.
Two-component epoxy resin: These are used for permanent
installations. A wide variety is available for different applications and materials. Epoxies are cured or hardened with time,
light and/or elevated temperature. Kistler can provide a base,
clip or mechanical adapter that can be permanently installed
on a structure. These have a threaded hole for attaching the
accelerometer, allowing it to be unscrewed as needed. Epoxies
are available up to 260 C.
8400_000-281e-06.14
Other Adhesives Available on the Market
Hot Glue: Hot glue can be used on rough surfaces or when
higher frequencies are not of primary concern. This is ideal for
quick assembling with or without mounting adapters (plastic
clip technique). Removal and cleaning is very simple. Hot glues
are often applied in structural analysis, where many sensors
will to be applied to a structure. Hot glue guns are commercially available (line or battery powered). There are a number
of different hot glues available; some of them can be used at
elevated temperature ranges. For polymer or plastic surfaces,
low temperature hot glues are available with sufficient sticking
properties.
Page 6/7
This information corresponds to the current state of knowledge. Kistler reserves
the right to make technical changes. Liability for consequential damage resulting
from the use of Kistler products is excluded.
2006 ... 2014, Kistler Group, Eulachstrasse 22, 8408 Winterthur, Switzerland
Tel. +41 52 224 11 11, Fax +41 52 224 14 14, [email protected], www.kistler.com
Kistler is a registered trademark of Kistler Holding AG.
Mounting Accessories For Acceleration Sensors
Direct Adhesive Mounting (continued...)
Adhesives
Mounting Area
Ambient Temperature
Availability
smooth,
polished
Room Temp.
Manufacturer
rough
Temperature Range
Solvent
Kistler
Temporary/Removeable
Synthetic wax/Petro wax
Double-sided glue tape
3M (Type 465, 924)
One-sided aluminum glue tape
3M (Type 433, 431)
Not required
Thermosetting plastic glue tape
(www.nittousa.com)
up to 150 C
Nitto Denko (USA),
P-224AMB
High tensile silicone flash masking
tape (www.nittousa.com)
up to 180 C
Nitto Denko (USA),
PS-1
refer to manufacturer
instructions
Adhesive,
Henkel Teroson,
UHU GmbH
MEK (1)
refer to manufacturer
instructions
Adhesive,
Henkel Teroson,
UHU GmbH
MEK (1)
up to 65 C approx.
Adhesive,
Henkel Teroson,
UHU GmbH
up to 120 C approx.
Adhesive,
Henkel Teroson,
UHU GmbH
55 C ... 180 C
Adhesive, Henkel Teroson
Fixed/Removeable/Permanent
Super glue, liquid
Super glue, gel
Hot glue (w/hot glue gun)
Permanent
Epoxy resin, 2 components
325 Speed Bonder
(1): MEK is the abbreviation for Methyl-ethyl-ketone. In order to ensure safe usage, please refer to the "Safety" data sheet mentioned in [3].
8400_000-281e-06.14
Removal Wrench for Miniature Accelerometer:
Type
A (length)
Recommended Sensor Type
1378
63,5
8778, 8278
Literature:
[1] Special print ref 20.305, "Acceleration Mounting Considerations," www.kistler.com
[2] Special print ref 20.407, "Frequency Response Limits for Low Impedance Accelerometers," www.kistler.com
[3] Safety data sheet according to Regulation (EC) No. 1907/2006, www.merckmillipore.com
Page 7/7
This information corresponds to the current state of knowledge. Kistler reserves
the right to make technical changes. Liability for consequential damage resulting
from the use of Kistler products is excluded.
2006 ... 2014, Kistler Group, Eulachstrasse 22, 8408 Winterthur, Switzerland
Tel. +41 52 224 11 11, Fax +41 52 224 14 14, [email protected], www.kistler.com
Kistler is a registered trademark of Kistler Holding AG.