Eas - Signed (Ibu TW)

Download as pdf or txt
Download as pdf or txt
You are on page 1of 14

Esco Cleanroom Air Shower,

Model EAS-2C_.

Cleanroom Air Showers

Esco Cleanroom Air Shower,


Model EAS-2C_.

Standards
Compliance

Filtration

Electrical Safety

EN-1822 (H13), Europe


IEST-RP-CC001.3, USA
IEST-RP-CC007, USA
IEST-RP-CC034.1, USA

UL- 61010-1, USA


CAN/CSA-22.2, No.61010-1
EN-61010-1, Europe
IEC-61010-1, Worldwide

* Please refer to the specifications table on page 4,6 & 8 for the model listing.

$MFBOSPPN1SPEVDUTtCleanroom Air Showers

Main Features
High velocity shower jets in excess of 20 m/s ensure efficient scrubbing action to
remove particulate matter.
Operating modes can be programmed in the field.
Microprocessor controller supervises all functions.
Mini-pleated HEPA filtration achieves > 99.99% typical efficiency at 0.3 micron
particles.
A disposable pre-filter with 85% arrestance extends the life of the main filter.
An emergency stop button is mounted on both sides of the shower.
Indicator lights mounted on both sides of the air shower exterior regulate traffic
flow in and out of the cleanroom.
Permanently lubricated direct drive centrifugal blowers are used in conjunction
with stainless steel air nozzles.

EAS, Cleanroom Air Shower Series Comparison Chart


Model

Airflow

Door

EAS (A-Series)

Single Sided

Single Leaf

EAS (B-Series)

Double Sided

Single Leaf

EAS (C-Series)

Double Sided

Double Leaf

See the Specification Chart for more details.

Introduction
Air Showers are self contained chambers
installed at entrances to cleanrooms and
other controlled environments. They
minimize particulate matter entering
or exiting the clean space. Personnel
and materials entering or exiting the
controlled environment are scrubbed
by high velocity HEPA-filtered air jets with
velocities of 20-22m/s (4000-4300fpm).
Contaminated air is then drawn through
the base within the unit, filtered and
recirculated.
Esco is a leader in air showers for demanding
applications in the micro-electronics,
semiconductors, pharmaceutical, spray-

painting, lab animal research and food


markets. Esco filed its first Air Shower
patent in the 1980s and since then has
installed thousands of units in diverse
markets worldwide. The present Esco
Air Shower is a third-generation product
and features a field-programmable
microprocessor control that offers the
maximum application flexibility of any unit
on the market.

Cleanroom Applications: The greatest


source of particulate contamination in a
cleanroom is the operator. Air showers
are installed between change areas and
the cleanroom. The air shower enhances

cleanroom operating protocol by serving


as a reminder to all operators that they
are entering a controlled environment.
Personnel therefore develop the habit of
gowning up properly before entering the
air shower.

Pharmaceutical and Lab Animal


Research Applications: Air showers
keep pharmaceutical production and
lab animal breeding areas clean and also
minimize egress of hazardous substances
and allergens from the controlled
environment.

Air Shower Efficacy Testing


5. The operator gowns up using another
garment from the same batch.

Esco is the only company in the industry to


validate the efficacy of our air showers using
the body box test, a method pioneered
by Esco.

6. The operator proceeds into the air


shower (device under test), for a
shower cycle.

1. New cleanroom garments without


laundry processing, which have
residual particulate contamination
from manufacturing in a non cleanroom
environment, are used.

7. The operator exits the air shower and


proceeds into the body box. In the
body box the operator performs the
same series of standardized physical
movements in order to induce the
generation of particles in the body box,
for a specific duration.

2. The operator gowns up (in a jumpsuit,


booties, gloves, mask and hood), and
enters a specially sealed enclosure (the
body box). This enclosure is equipped
with vertical laminar flow.

8. Particle count levels are measured,


and compared against the baseline.
The overall efficacy of the air shower
under test and shower cycle is
calculated. Shower cleaning efficiency at various particle sizes is also
characterized.

3. The operator performs a series of


standardized physical movements in
order to induce the generation of
particles in the body box, for a specific
duration.
4. A particle counter connected to
the base of the body box measures
particle count levels. This count is the
baseline level.

9. The test is repeated multiple times, to


gather sufficient data and eliminate
any bias.

Particles Removal In EAS

Air Showers Efficiency Comparison

Exposure Time In Air Shower (SEC)

15000

Air Showers
Efficiency (%)
100

Overall Air Showers Efficiency Comparison

80

10000

60

10

15

20

25

30

35

Air Showers
Efficiency (%)

40

0.5
m

5000

ab

ov

ab

0.
1.4

ab

ab

ab

ove

20

ov

ov

ove

Particles (m)
20

Exposure Time In Air Shower (SEC)

$MFBOSPPN1SPEVDUTtCleanroom Air Showers

0.3
Esco EAS-1B1

0.5
Competitor H

1.0

3.0

5.0

10.0

Air Shower Filtration System


Blower

ULPA-filtered air

HEPA Filter

Unfiltered / potentially contaminated air

Air is forced by the blower(s) through


HEPA filter(s) which are 99.99% efficient
against particles of 0.3 microns.
Filtered air is ejected through nozzles at
high velocities into the chamber. These
turbulent air streams disperse particulate
matter on all surfaces.

The air is continuously filtered and


recirculated. The air shower is a selfcontained device and does not exchange
air with the environment it is placed in.

Dispersed particulate matter migrate


with the air stream towards the lower
areas in the air shower chamber. Air
enters the blower supply plenum through
pre-filter(s) installed at the base of the
chamber.

Air Shower Operating Sequences


Unlike conventional air showers which are delivered with a fixed operating sequence, the Esco Air Showers operating sequence may
be selected from three pre-programmed sequences:

er

Grey Room

Air Show

Air Show

Clean Room

Grey Room

er

er

Air Show

Clean Room

Grey Room

er
No Show
it
Direct Ex

Clean Room
er

Air Show

One-Way

Two-Way One-Way

Two-Way

Personnel may enter the controlled


environment but not exit through the
air shower. At the idle state, the clean
side door is locked while the grey side is
unlocked. This mode of operation is useful
for controlling traffic patterns into and out
of the controlled environment.

Personnel may enter or exit the controlled


environment through the air shower.
When entering the controlled environment
the shower is activated. When exiting the
shower is disabled to reduce throughput
time. The air shower program is able to
detect if the person is entering or exiting
the controlled environment via door
sensors and a time-sequenced control.

Personnel may enter or exit the controlled


environment through the air shower. In
both directions the air shower is activated.
This mode of operation is useful in
pharmaceutical and lab animal research
applications to prevent the egress of
hazardous substances and allergens from
the controlled environment.

Cleanroom Air Showers


High Performance Blower System
German made ebm-papst permanently lubricated,
centrifugal motor/blowers with external rotor designs.
Motors selected for energy efficiency, compact
design, and flat profile. Completely integrated
assembly optimizes motor cooling.
All rotating parts balanced for smooth, quiet,
vibration-free operation.

HEPA Filtration
System

HEPA filter(s) provide


99.99% typical
efficiency for particle
sizes of 0.1 to 0.3
microns.

Light Indicator and Emergency Switch

Indicator lamps indicate if doors are


locked or unlocked, thereby regulating
the flow of personnel in and out of the
air shower.
Emergency buttons mounted on both
external faces of the shower unlocks all
doors instantly.

Door

Heavy-duty, durable aluminium framed


door assemblies are constructed with
glass windows permitting visibility.

$MFBOSPPN1SPEVDUTtCleanroom Air Showers

Esco Cleanroom Air Shower,


Model EAS-2C_.

Key Features
Esco Air Shower filters meet the IESTRP-CC001.3 recommended practice for
HEPA performance (USA), and EN 1822
for H13 performance (EU).

Sentinel Microprocessor Control System, Programmable

The auto reset feature unlocks doors


in case personnel open the air shower
door but do not actually enter, thus
preventing accidental lock-outs.

The microprocessor control detects improper operation and


displays corresponding error messages should the integrity
of the cleanroom be violated.

In case of a power failure, all doors are


unlocked automatically, and controller
settings saved.

The LCD displays shower duration and countdown, and


reports cycle progress and operational status.
A 24 hour clock displays local time.

A disposable pre-filter with 85%


arrestance extends the life of the
main filter.
The air shower is constructed of
electro-galvanised steel sheets with an
abrasion-resistant oven-baked powder
coated finish.
Each air shower is individually factory
tested for safety and performance
in accordance with international
standards.
Robust construction qualify the air
shower for the most demanding
applications. The air shower is fully
assembled and ready to install and
operate when shipped.

Light Diffuser

Diffusers ensure even and uniform lighting


throughout the chamber.

Each unit is shipped with a


documentation outlining the tests
undertaken and the units individual
results for each unit.
All electrical components are UL listed
/ recognised.
The Air Shower is warranted for 1
year excluding consumable parts and
accessories.
FAST TRACK
FastTrack models are available for shipment within 1
week from order placement, from Esco Singapore, to
destinations around the world. The following models
are available under this program: EAS-1A1 & EAS-1B1.

Stainless Steel Nozzles

An array of stainless steel nozzles


direct high-velocity jets within
the chamber.

General Specifications, Cleanroom Air Shower, Model EAS (A-Series)


Note to customer: Insert electrical voltage number into last model number digit _ when ordering.
Model
External Dimensions (W x D x H)
Internal Work Area, Dimensions (W x D x H)

EAS-1A_

EAS-2A_

EAS-3A_

1260 x 1000 x 2050 mm


49.7" x 39.4" x 80.7"
790 x 920 x 1930 mm
31.1" x 36.2" x 76.0"

1260 x 2000 x 2050 mm


49.7" x 78.7" x 80.7"
790 x 1920 x 1930 mm
31.1" x 75.6" x 76.0"

1260 x 3000 x 2050 mm


49.7" x 118.1" x 80.7"
790 x 2920 x 1930 mm
31.1" x 115" x 76.0"

356/ Hr

351/ Hr

Air Change

371/ Hr

Initial Airflow Velocity

20-22 m/s (3,937-4,330 fpm)

Number of Nozzles

12

Air Shower Duration


Persons Per Cycle
Personnel Flow (Persons / Min. )

2-3

4-6

8-12

15-23

Above figures based on: Total Cycle Time of 16 seconds


(12 seconds of Air Shower + 4 seconds for buffer time / personnel entrance and exit)

Filtration Efficiency

Main Filter: >99.99% at 0.3 m


Pre-Filter: Arrestance 85%, efficiency 20%

Filtration Elements

Main Filter: HEPA filter


Pre-Filter: Disposable and non-washable polyester fibers

Fluorescent Lamp

17 W x 2

Air Shower Construction


Max. Power
Consumption,
Current, BTU/Hr

Electrical

245 W, 1.2 A, 500 BTU/ Hr

490 W, 2.4 A, 1000 BTU/ Hr

735 W, 3.5 A, 1499 BTU/ Hr

During Standby

113 W, 0.5 A, 231 BTU/ Hr

226 W, 1 A, 461 BTU/ Hr

339 W, 1.5 A, 692 BTU/ Hr

220-240V, AC, 50Hz, 1

EAS-1A1

EAS-2A1

EAS-3A1

110-130V, AC, 60Hz, 1

EAS-1A2

EAS-2A2

EAS-3A2

220-240V, AC, 60Hz, 1

EAS-1A3

EAS-2A3

EAS-3A3

410 kg (904 lbs)

820 kg (1808 lbs)

1230 kg (2712 lbs)

250 kg (551 lbs)

500 kg (1102 lbs)

750 kg (1653 lbs)

1450 x 1250 x 2152 mm


57.1" x 49.2" x 84.7"

1450 x 1250 x 2152 mm


57.1" x 49.2" x 84.7"

N/A

Pallet A

N/A

1450 x 1250 x 2152 mm


57.1" x 49.2" x 84.7"

1450 x 1250 x 2152 mm


57.1" x 49.2" x 84.7"

Pallet B

N/A

1450 x 1250 x 2152 mm


57.1" x 49.2" x 84.7"

1450 x 1250 x 2152 mm


57.1" x 49.2" x 84.7"

Pallet C

N/A

N/A

1450 x 1250 x 2152 mm


57.1" x 49.2" x 84.7"

Pallet A

2100 x 1300 x 778 mm


82.7" x 51.2" x 30.6"

2100 x 1300 x 924 mm


82.7" x 51.2" x 36.4"

2100 x 1300 x 1296 mm


82.7" x 51.2" x 51.0"

Pallet B

N/A

2100 x 1300 x 632 mm


82.7" x 51.2" x 24.9"

2100 x 1300 x 781 mm


82.7" x 51.2" x 30.7"

3.90 m3 (138 cu.ft.)

7.80 m3 (276 cu.ft.)

N/A

Module Form

N/A

7.80 m3 (276 cu.ft.)

11.70 m3 (414 cu.ft.)

Unassembled

2.12 m3 (75 cu.ft.)

4.24 m3 (150 cu.ft.)

5.66 m3 (200 cu.ft.)

Assembled (W x D x H)

Module Form
(W x D x H)

Unassembled
(W x D x H)
Assembled
Shipping Volume,
Maximum

17 W x 6

During Operation

Net Weight

Shipping
Dimensions,
Maximum
(W x D x H)*

17 W x 4

1.5 mm/ 0.06"/18 electro-galvanised steel / White oven-baked epoxy powder-coated finish

Gross Weight

18

Factory set at 12 seconds (adjustable)

Esco Cleanroom Air Showers, Modes of Shipment, Model EAS-2A_


Assembled

Module Form

Unassembled
Unit B
Top Frame A

Cable Connector

Top Frame B
Stainless Steel Floor

Unit A

Base Frame B
Base Frame A
Door
Pallet B
Pallet A

$MFBOSPPN1SPEVDUTtCleanroom Air Showers

Module B1
Module A1
Pallet A

Module B2
Module A2
Pallet B

Model EAS (A-Series) Cleanroom Air Shower Technical Specifications


EAS-1A_

EAS-2A_

EAS-3A_
1
2
6
7

8
4
9
5

1. Fluorescent lamp
2. Esco microprocessor system
3. Nozzles

4. Door
5. Pre-filter
6. Indicator light

7. Emergency switch
8. HEPA filter
9. Blower

Model EAS (A-Series), Door Direction


(Factory Configured. Choose One When Ordering.)
1.

2.

3.

4.

5.

6.

7.

8.

General Specifications, Cleanroom Air Shower, Model EAS (B-Series)*


Note to customer: Insert electrical voltage number into last model number digit _ when ordering.
Model
External Dimensions (W x D x H)
Internal Work Area, Dimensions (W x D x H)

EAS-1B_

EAS-2B_

EAS-3B_

1530 x 1000 x 2050 mm


60.2" x 39.4" x 80.7"

1530 x 2000 x 2050 mm


60.2" x 78.7" x 80.7"

1530 x 3000 x 2050 mm


60.2" x 118.1" x 80.7"

790 x 920 x 1930 mm


31.1" x 36.2" x 76.0"

790 x 1920 x 1930 mm


31.1" x 75.6" x 76.0"

790 x 2920 x 1930 mm


31.1" x 115" x 76.0"

712/ Hr

702/ Hr

Air Change

743/ Hr

Initial Airflow Velocity

20-22 m/s (3,937-4,330 fpm)

Number of Nozzles

12

24

Air Shower Duration


Persons Per Cycle
Personnel Flow (Persons / Min.)

2-3

4-6

8-12

15-23

Above figures based on: Total Cycle Time of 16 seconds


(12 seconds of Air Shower + 4 seconds for buffer time / personnel entrance and exit)

Filtration Efficiency

Main Filter: >99.99% at 0.3 m


Pre-Filter: Arrestance 85%, efficiency 20%

Filtration Elements

Main Filter: HEPA filter


Pre-Filter: Disposable and non-washable polyester fibers

Fluorescent Lamp

17 W x 2

Air Shower Construction

17 W x 4

17 W x 6

1.5 mm/ 0.06"/18 electro-galvanised steel / White oven-baked epoxy powder-coated finish

During Operation

500 W, 2.3 A, 1020 BTU/ Hr

1000 W, 4.6 A, 2040 BTU/ Hr

1500 W, 7 A, 3060 BTU/ Hr

During Standby

162 W, 0.7 A, 330 BTU/ Hr

200 W, 1 A, 408 BTU/ Hr

250 W, 1.1 A, 510 BTU/ Hr

220-240V, AC, 50Hz, 1

EAS-1B1

EAS-2B1

EAS-3B1

110-130V, AC, 60Hz, 1

EAS-1B2

EAS-2B2

EAS-3B2

220-240V, AC, 60Hz, 1

EAS-1B3

EAS-2B3

EAS-3B3

Gross Weight

410 kg (904 lbs)

820 kg (1808 lbs)

1230 kg (2712 lbs)

Net Weight

250 kg (551 lbs)

500 kg (1102 lbs)

750 kg (1653 lbs)

1750 x 1250 x 2152 mm


68.9" x 49.2" x 84.7"

1750 x 2500 x 2152 mm


68.9" x 98.4" x 84.7"

N/A

Pallet A

N/A

1750 x 1250 x 2152 mm


68.9" x 49.2" x 84.7"

1750 x 1250 x 2152 mm


68.9" x 49.2" x 84.7"

Pallet B

N/A

1750 x 1250 x 2152 mm


68.9" x 49.2" x 84.7"

1750 x 1250 x 2152 mm


68.9" x 49.2" x 84.7"

Pallet C

N/A

N/A

1750 x 1250 x 2152 mm


68.9" x 49.2" x 84.7"

Pallet A

2100 x 1300 x 1048 mm


82.7" x 51.2" x 41.3"

2100 x 1300 x 1296 mm


82.7" x 51.2" x 51.0"

2100 x 1300 x 1668 mm


82.7" x 51.2" x 65.7"

Pallet B

N/A

2100 x 1300 x 800 mm


82.7" x 51.2" x 31.5"

2100 x 1300 x 1219 mm


82.7" x 51.2" x 48.0"

4.70 m3 (166 cu.ft.)

9.40 m3 (333 cu.ft.)

N/A

Module Form

N/A

9.40 m (333 cu.ft.)

14.10 m (499 cu.ft.)

Unassembled

2.83 m3 (100 cu.ft.)

5.71 m3 (202 cu.ft.)

7.87 m3 (279 cu.ft.)

Max. Power
Consumption,
Current, BTU/Hr

Electrical

10

36

Factory set at 12 seconds (adjustable)

Assembled (W x D x H)

Shipping
Dimensions,
Maximum
(W x D x H)**

Module Form
(W x D x H)

Unassembled
(W x D x H)
Assembled
Shipping Volume,
Maximum

Esco Cleanroom Air Showers, Modes of Shipment, Model EAS-2B_


Assembled

Module Form

Unassembled

Unit B

Top Frame B
Stainless Steel Floor

Cable Connector

Top Frame A
Base Frame B
Base Frame A
Module B2

Unit A

Pallet B
Pallet B
Pallet A

Module B1
Module A2
Module A1
Pallet A

$MFBOSPPN1SPEVDUTtCleanroom Air Showers

Model EAS (B-Series) Cleanroom Air Shower Technical Specifications


EAS-1B_

EAS-2B_

EAS-3B_
1
2
6

11

8
4
9
5

1. Fluorescent lamp
2. Esco microprocessor system
3. Nozzles

4. Door
5. Pre-filter
6. Indicator light

7. Emergency switch
8. HEPA filter
9. Blower

Model EAS (B-Series), Door Direction


(Factory Configured. Choose One When Ordering.)
1.

2.

3.

4.

General Specifications, Cleanroom Air Shower, Model EAS (C-Series)*


Note to customer: Insert electrical voltage number into last model number digit _ when ordering.
Model

EAS-2C_

EAS-3C_

External Dimensions (W x D x H)

2330 x 2000 x 2050 mm


91.7" x 78.7" x 80.7"

2330 x 3000 x 2050 mm


91.7" x 118.1" x 80.7"

Internal Work Area, Dimensions (W x D x H)

1590 x 1920 x 1930 mm


62.6" x 75.6" x 76.0"

1590 x 2920 x 1930 mm


62.6" x 115" x 76.0"

Air Change

354/ Hr

349/ Hr

Initial Airflow Velocity

20-22 m/s (3,937-4,330 fpm)

Number of Nozzles

24

36

Air Shower Duration

Factory set at 12 seconds (adjustable)

Persons Per Cycle


Personnel Flow (Persons / Min. )

4-6

8-12

15-23

Above figures based on: Total Cycle Time of 16 seconds


(12 seconds of Air Shower + 4 seconds for buffer time / personnel entrance and exit)

Filtration Efficiency

Main Filter: >99.99% at 0.3 m


Pre-Filter: Arrestance 85%, efficiency 20%

Filtration Elements

Main Filter: HEPA filter


Pre-Filter: Disposable and non-washable polyester fibers

Fluorescent Lamp

17 W x 8

Air Shower Construction


Max. Power
Consumption,
Current, BTU/Hr

Electrical

17 W x 12

1.5 mm/ 0.06"/18 electro-galvanised steel / White oven-baked epoxy powder-coated finish

During Operation

1000 W, 4.6A, 2040 BTU/ Hr

1500 W, 7 A, 3060 BTU/ Hr

200 W, 1 A, 408 BTU/ Hr

250 W, 1.1 A, 510 BTU/ Hr

220-240V, AC, 50Hz, 1

EAS-2C1

EAS-3C1

110-130V, AC, 60Hz, 1

EAS-2C2

EAS-3C2

During Standby

220-240V, AC, 60Hz, 1

12

2-3

EAS-2C3

EAS-3C3

Gross Weight

910 kg (2006 lbs)

1660 kg (3660 lbs)

Net Weight

750 kg (1653 lbs)

1500 kg (3307 lbs)

2500 x 2500 x 2232 mm


98.4" x 98.4" x 87.9"

N/A

Pallet A

2500 x 1250 x 2232 mm


98.4" x 49.2" x 87.9"

2500 x 1250 x 2232 mm


98.4" x 49.2" x 87.9"

Pallet B

2500 x 1250 x 2232 mm


98.4" x 49.2" x 87.9"

2500 x 1250 x 2232 mm


98.4" x 49.2" x 87.9"

Pallet C

N/A

2500 x 1250 x 2232 mm


98.4" x 49.2" x 87.9"

Pallet A

2100 x 1300 x 1296 mm


82.7" x 51.2" x 51.0"

2100 x 1300 x 1668 mm


82.7" x 51.2" x 65.7"

Pallet B

2100 x 1300 x 800 mm


82.7" x 51.2" x 31.5"

2100 x 1300 x 1219 mm


82.7" x 51.2" x 48.0"

Assembled (W x D x H)

Shipping
Dimensions,
Maximum
(W x D x H)**

Module Form
(W x D x H)

Unassembled
(W x D x H)

Shipping Volume,
Maximum

Assembled

13.94 m3 (493 cu.ft.)

N/A

Module Form

13.94 m3 (493 cu.ft.)

20.91 m3 (740 cu.ft.)

Unassembled

5.71 m3 (202 cu.ft.)

7.87 m3 (279 cu.ft.)

Esco Cleanroom Air Showers, Modes of Shipment, Model EAS-2C_


Assembled

Module Form

Unassembled

Unit B

Top Frame B
Stainless Steel Floor

Cable Connector

Top Frame A
Base Frame B
Base Frame A
Module B2

Unit A

Pallet B
Pallet B
Pallet A

Module B1
Module A2
Module A1
Pallet A

$MFBOSPPN1SPEVDUTtCleanroom Air Showers

Model EAS (C-Series) Cleanroom Air Shower Technical Specifications


EAS-2C_
1
2
6
7

8
4
9
5

EAS-3C_
1

1. Fluorescent lamp
2. Esco microprocessor system
3. Nozzles
4. Door
5. Pre-filter
6. Indicator light
7. Emergency switch
8. HEPA filter
9. Blower

Model EAS (C-Series), Door Direction


(Factory Configured. Choose One When Ordering.)
1.

2.

13

ART Equipment
Biological Safety Cabinets
CO2 Incubators
Compounding Pharmacy Equipment
Containment / Pharma Products
Ductless Fume Hoods
Freeze Dryer
Lab Animal Research Products
Laboratory Fume Hoods
Laboratory Ovens and Incubators
Laminar Flow Clean Benches
PCR Cabinets
PCR Thermal Cyclers
Powder Weighing Balance Enclosures
Ultra-low Freezers
The Esco Group of Companies is a global life sciences tools provider with sales in over 100 countries. The group is active in
lab equipment, pharma equipment and medical devices. Manufacturing facilities are located in Asia and Europe. R&D is
conducted worldwide spanning the US, Europe and Asia. Sales, service and marketing subsidiaries are located in 12 major
markets including the US, UK, Singapore, Japan, China and India. Regional distribution centers are located in the US, UK,
and Singapore.

9010053_Cleanroom_EAS_Brochure_A4_vC_Apr-2013
Esco can accept no responsibility for possible errors in catalogues, brochures and other printed materials. Esco reserves the right to alter its
products and specifications without notice. All trademarks and logotypes in this material are the property of Esco and the respective companies.

-JGF 4DJFODF t $IFNJDBM 3FTFBSDI t "TTJTUFE 3FQSPEVDUJWF


5FDIOPMPHZ "35
 t 1IBSNBDFVUJDBM &RVJQNFOU t (FOFSBM &RVJQNFOU

&TDP5FDIOPMPHJFT *ODt2940 Turnpike Drive, Units 15-16tHatboro, PA 19040, USA


5PMM'SFF64"BOE$BOBEBt5FM441-9661t'BY441-9660
VTFTDPHMPCBMDPNt[email protected]
&TDP.JDSP1UF-UEt$IBOHJ4PVUI4USFFUt4JOHBQPSF
5FM t'BY tNBJM!FTDPHMPCBMDPN
www.escoglobal.com
Esco Global Offices | Beijing, China | Chengdu, China | Kuala Lumpur, Malaysia | Manama, Bahrain | (VBOH[IPV $IJOB]
Hanoi, Vietnam | Melaka, Malaysia | Mumbai, India | 1IJMBEFMQIJB 1" 64"| 4BMJTCVSZ 6,| 4IBOHIBJ $IJOB| 4FPVM ,PSFB
| %FMIJ *OEJB| 0TBLB +BQBO| Manila, Philippines |.JESBOE 4PVUI"GSJDB|+BLBSUB *OEPOFTJB|4JOHBQPSF

escoglobal com
LIVE SUPPORT

Esco Micro Pte Ltd


Cert. No: 651076

PT Esco Bintan Indonesia


Cert. No: 651333/E

You might also like