Xray Thermo
Xray Thermo
Xray Thermo
a n a l y s i s
ARL OPTIMX
X-Ray Fluorescence Spectrometer
ARL OPTIMX
Optimized Performance for Your Application
ARL OPTIMX
Optimized Performance for your Application.
The wavelength dispersive X-ray fluorescence
(WDXRF) technique is one of the most
versatile analytical methods for the chemical
analysis of solids and liquids. Elements from
beryllium to uranium can be analyzed in a
wide variety of samples with accuracy,
precision and reliability. The technique and
the analytical methods are mature enough to
establish clear qualitative and quantitative
characterization of diversified materials.
Modern technological developments have
helped to include WDXRF into reference
methods for analytical laboratories (ASTM
and ISO norms for example) by virtue of their
simplicity, flexibility, affordability and
reliability. The cost per analysis is clearly
advantageous over many traditional wet
chemical and other spectroscopic methods.
0.12
0.10
Intensity (Kcps)
0.08
0.06
0.04
0.02
0.00
0.000
0.008
0.012
Concentration
0.004
0.016
0.020
(%)
Cell 1
Cell 2
Cell 3
Cell 4
Cell 5
Cell 6
Cell 7
Average
Std. Dev.
25.4
25.7
26.4
26
25
25.9
26.7
25.9
0.58
100.6
100.7
101.8
102.3
98.4
100.5
101.5
100.8
1.27
Table 1: Excellent
reproducibility obtained for
analysis of sulfur in oils (120s)
Run nr
Al2 O3
Ca0
Fe2O3
K2O
MgO
P2O5
SO3
SiO2
1
2
3
4
5
6
7
AVG
Sd.Dev.
0.329
0.328
0.328
0.324
0.326
0.327
0.327
0.327
0.0015
50.41
50.44
50.45
50.41
50.36
50.38
50.41
50.41
0.029
0.303
0.303
0.304
0.305
0.301
0.301
0.302
0.303
0.0014
0.116
0.117
0.116
0.117
0.117
0.116
0.116
0.116
0.0005
2.76
2.77
2.74
2.76
2.75
2.74
2.74
2.75
0.011
0.014
0.015
0.014
0.014
0.014
0.014
0.014
0.014
0.0003
0.036
0.036
0.036
0.036
0.036
0.035
0.036
0.036
0.0003
2.68
2.68
2.68
2.69
2.68
2.69
2.68
2.68
0.005
Table 2: Repeatability test for a limestone sample (pressed powder, 100 s counting time)
0.025
0.02
Intensity (kcps)
ARL OPTIMX
Optimized Performance for Your Application
0.015
0.01
0.005
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
Concentration (%)
Fig. 2: Calibration curve for Na using fusion beads of oxides. Standard error of estimate is
0.011% in a range from 0.02% to 0.32%
Element
Analytical device
Typical ranges
Na
Mg
P
K
Ca
Fe
Cu
Zn
Cl
Mn
Se
Fixed channel
Fixed channel
SmartGonio
SmartGonio
SmartGonio
SmartGonio
SmartGonio
SmartGonio
SmartGonio
SmartGonio
SmartGonio
0 0.03 %
0 0.12 %
0 1.1 %
01%
0 1.6 %
0 0.33 %
0 0.012 %
0 0.2 %
0 0.48 %
0 0.0023 %
0 3.4 ppm
20 ppm*
11 ppm*
4.4 ppm
2 ppm
10 ppm
2.1 ppm
0.6 ppm
2 ppm
10 ppm
1.2 ppm
0.24 ppm
Manual introduction
of a solid sample
placed in a cassette
ARL OPTIMX
Optimized Performance for Your Application
Collimator
Exclusive SmartGonio
Sample
Optical encoder
2 theta drive
Detector
Crystal
Optical encoder
X-ray tube
1 theta drive
SmartGonio
principle of
operation
X-ray excitation
Multichromator
SmartGonio
Counting electronics
Sample loading
Sample holders
Thermo Electron Corporation reserves the right to vary these specifications without prior notice
Australia
+61 2 9898 1244 [email protected]
Austria
+43 1 333 50340 [email protected]
Belgium
+32 2 482 30 30 [email protected]
Canada
+1 800 532 4752 [email protected]
China
+86 10 5850 3588 [email protected]
France
+33 1 60 92 48 00 [email protected]
Germany
+49 6103 4080 [email protected]
Italy
+39 02 950 591 [email protected]
Japan
+81 45 453 9100 [email protected]
Netherlands
+31 76 587 98 88 [email protected]
Nordic
+46 8 556 468 00 [email protected]
South Africa
+27 11 570 1840 [email protected]
Spain
+34 91 657 4930 [email protected]
Switzerland
+41 21 694 71 11 [email protected]
UK
+44 1442 233555 [email protected]
USA
+1 800 532 4752 [email protected]
www.thermo.com
XRF-02-1062-1103-CH-e