Ut Procedure Selon Asme
Ut Procedure Selon Asme
Scope
1.1-This procedure describes the general requirements for Ultrasonic
examination of metallic welding according related approved weld maps as may
be required by the specification or under .which component is being designed
and manufactured.
1.2- This procedure provides the materials, equipment, personnel qualification,
examination process, evaluation, record and acceptance standards for XXX
project which will be fabricated in YYY Company
1.3-Ultrasonic testing / examination shall be performed in accordance with the
procedure which shall as a minimum contain the requirement of table T-421 in
ASME Section V.
2. References
2.1- ANSI, ASME Sec.VIII devision 1 & 2 Appendix 12
2.2- ANSI, ASME Sec.V article 4 & 5
2.3- ANSI recommended practices No.ASNT-TC-1A for NDT personnel
qualification and certification
3. General:
3.1-Basic Requirements and Terms
When this procedure is specified by a referencing code section, the ultrasonic
testing method described in this procedure shall be used with ASME code
section V-article 1, 4 & 5 & ASME SEC.VIII, DIV.1. When SA, SB and SE
documents are referenced, they are located in article 23.
3.2-Examination coverage (scan overlap)
Each pass of the search unit shall overlap minimum 10% of the transducer
(piezo-electric element) dimension perpendicular to the direction of the scan.
3.3- Rate of search unit movements
The rate of search unit movement for examination shall not exceed 6
inch/sec(150 mm/s).
4. Personnel Qualification:
The N.D.T. personnel whose performing ultrasonic testing / examination shall
be qualified and meets the requirements of level II, in Accordance to ASNTSNT-TC-1A.
5. Surface Preparation
The surface to be subject to ultrasonic testing shall be free from any existing
lubricants, dirt, residue, rust and sharp edges, so; to smoothly carry on the
procedure.
6. Equipments
6.1- Examination shall be conducted with a pulse echo A-Scan ultrasonic
testing instrument capable of generating frequencies over the range of at least
1 MHZ to 5 MHZ. Ultrasonic testing instrument shall be calibrated for the
examination and shall be capable of meeting the requirements of screen
height linearity and amplitude control linearity in paragraph 5.3 & 5.4 of this
procedure.(Trade name of ultrasonic testing instrument is krautkramer USN52) Ultrasonic testing instrument shall be equipped with stepped gain control in
units of 2 dB or less (according to ASME sec V-T-431).
6.2- Search Units
Search units may contain either single or dual transducer element.The search
unit size shall be selected according to the following table.
(1) Straight Beam. (either single or dual crystal)
Frequency(MHz)
Transducer size(mm)
Material thickness(mm)
4 or 5
10
25.4 or less
2 or 4
10 or 20
Over 25.4
Frequency(MHz)
Transducer size(mm)
Material thickness(mm)
70
10
12.5 or less
60 , 70
10
12.5 or 25.4
45 , 60 or 45 , 70
10
Over 25.4
(3) Search units with contoured contact wedges may be used to aid ultrasonic
coupling.
6.3-Screen Height Linearity
The ultrasonic testing instrument shall provide linear vertical presentation
within 5% (According to ASME Sec.V, Article 5 T-532) of the full screen height
for 20% to 80% of the calibrated screen height.
The procedure for evaluating screen height linearity is provided in appendix 1
of article 5, ASME code Sec.V and shall be performed at the beginning of each
period of extended use (or every 3 months, which ever is less).
6.4-Amplitude Control linearity
All recorded indications since the last valid calibration or calibration check shall be
Re-examined with the corrected calibration and their values shall be changed on the
data sheets.
7.3.2- DAC correction If a point on the distance-amplitude correction (DAC) curve
has decreased 20% or 2 db of its amplitude, all data sheets since the last calibration
or calibration check shall be marked void. A new calibration shall be made and
recorded and the area covered by the voided data shall be Re-examined. If any point
of the distance-amplitude correction (DAC) curve has increased more than 20% or 2
db of its amplitude, all recorded indications since the last valid calibration or
calibration check shall be Re-examined with the corrected calibration and their values
shall be changed on the data sheets.
8. Examination
8.1- General Requirements:
The scanning shall be performed at a gain setting at least two times the primary
reference level evaluation shall be performed with respect to the primary reference
level.
The volumes shall be scanned by straight and angle beam techniques as a described
in detail in paragraph 8 angle beams, having nominal angles as recommended in
table 6.2.1 & 6.2.2 with respect to a perpendicular to the examination surface, shall
generally be used.Other pairs of angle beams are permitted provided the measured
difference between the angle is at least 10.
8.1.1- Angle beam method scanning shall be made basically from outside of the
vessel by the direct contact method. The angle beam shall be directed
at approximately right angle to the weld axis from two direction where possible to
detect the reflectors oriented angular to the weld .In addition, the angle beam shall be
directed essentially parallel to the weld axis to detect the reflectors oriented
transverse to the weld. The search unit shall be manipulated so that the ultrasonic
energy passes through the whole volumes of weld and heat affected zone (20mm) of
base metal.The refraction angle of the angle beam transducer & exit point shall be
measured using IIW standard calibration block. The tolerance of the refraction angle
shall not be more than 2
8.1.2- Straight Beam Method
(1) The heat affected zone and weld where practicable shall be scanned with the
straight beam search unit, before scanning with angle beam.
(2) The scanning of the adjacent base metal of at least 2t tana (where t=base metal
thickness & a=1/2 groove angle) shall be performed to detect reflectors that
affect interpretation of angle beam results, and is not to be used as an acceptance
rejection examination. Locations and areas of such reflectors shall be recorded.
8.2- Surface Preparation
The base metal on each side of the weld shall be free of weld spatter, surface
irregularities, or foreign matter that might interfere with the examination.