This manual provides information on installation, operation and maintenance of a radiographic table.

This manual is for the MULT-FWFTT Four-Ways Floating Top Table.

Safety symbols that may appear include symbols indicating caution, safety, type B applied part, protection against water/particulate matter, radiation hazards and dangerous voltage.

Technical Publication

SM-1103R3

Service Manual
General Purpose Rad Table

This product bears a CE marking in accordance with the provisions of the 93/42/EEC MDD dated June 14, 1993.
Este producto ostenta una marca CE de acuerdo con las disposiciones de la Directiva 93/42/CEE del 14 de Junio de 1993 sobre Productos Mdicos.
Ce produit porte la marque CE de conformit aux rglements de la Directive 93/42/CEE du 14 juin 1993 relative aux Produits mdicaux.

The information comprised in this manual applies to the following equipments


La informacin contenida en este manual se aplica a los siguientes equipos
Linformation contenue dans ce manuel est applique aux quipements suivants

MULT-FWFTT Four-Ways Floating Top Table MULTIRAD

Manufactured by:
Fabricado por:
Fabriqu par:
SEDECAL
Sociedad Espaola de Electromedicina y Calidad S.A.
Pelaya, 9 -- 13. Polgono Industrial Ro de Janeiro
28110 Algete, Madrid -- Espaa (Spain)
Phone: +34 916 280 544

Fax: +34 902 190 385

www.sedecal.com

General Purpose Rad Table


Service Manual

REVISION HISTORY
REVISION

DATE

REASON FOR CHANGE

MAY 23, 2011

First edition

AUG 09, 2013

Design changes

SEPT 09, 2013

General Update

NOV 05, 2013

Ammended distance of anchoring holes

This Document is the english original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.

SM-1103R3

General Purpose Rad Table


Service Manual

ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their
application and meaning are described below.

DANGERS ADVISE OF CONDITIONS OR SITUATIONS THAT


IF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUS
PERSONAL INJURY OR DEATH.

ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT


HEEDED OR AVOIDED COULD CAUSE SERIOUS PERSONAL
INJURY, OR CATASTROPHIC DAMAGE OF EQUIPMENT OR
DATA.

Advise of conditions or situations that if not heeded or


avoided could cause personal injury or damage to equipment
or data.

Note

Alert readers to pertinent facts and conditions. Notes represent


information that is important to know but which do not necessarily
relate to possible injury or damage to equipment.

SM-1103R3

General Purpose Rad Table


Service Manual

SAFETY SYMBOLS
The following safety symbols may appear in the equipment.
Their meaning are described below.

Caution. Consult accompanying documents.

Safety Symbol. Follow instructions for use, especially those


instructions identified with Advisory Symbols to avoid any
risk for the Patient or Operator.
(Only applies to IEC 60601--1 Standard -- Third edition)

General Mandatory action.

Type B applied part.

IPX0

Protection against harmful ingress of water or particulate matter.


IP Classification: Ordinary.

Ionizing radiation.

Non-ionizing electromagnetic radiation.

Radiation of Laser apparatus.


Do not stare into beam.
(Only applicable to equipment with Laser Pointer)

Dangerous voltage.

SM-1103R3

General Purpose Rad Table


Service Manual

General warning, caution, risk of danger.

Warning: Ionizing radiation.

Warning: Non-ionizing radiation.

Warning: Laser beam.

Warning: Dangerous voltage.

Warning: Do not place fingers between mobile and fixed parts of the equipment, it
may cause serious injuries to patient or operator.
As well, make sure the patient extremities are correctly positioned into limit areas
during operation, movement of parts may cause serious damages to patient.

Electrostatic sensitive devices.

No pushing.

No sitting.

No stepping on surface.

SM-1103R3

General Purpose Rad Table


Service Manual

Stop (of action).

Emergency stop.

ON power.

OFF power.

ON / OFF (push-push).
Each position, ON or OFF, is a stable position.

Alternating current.

Three-phase alternating current.

Three-phase alternating current with neutral conductor.

Connection point for the neutral conductor on Permanently Installed equipment.

Direct current.

SM-1103R3

General Purpose Rad Table


Service Manual

Both direct and alternating current.

Protective Earth (Ground).

Earth (Ground).

This symbol according to the European Directive indicates that the Waste of
Electrical and Electronic Equipment (WEEE) must not be disposed of as unsorted
municipal waste and must be collected separately. Please contact an authorized
representative of the manufacturer or an authorized waste management company
for information concerning the decommissioning of your equipment.

Li/Pb/Cd/Hg

This separate collection symbol is affixed to a battery or its packing, to advise that
the battery must be recycled or disposed of in accordance with local or country
laws. The letters below the symbol indicate whether certain elements (Li=Lithium,
PB=Lead, CD=Cadmium, Hg=Mercury) are contained in the battery. All batteries
removed from the equipment must be properly recycled or disposed. Please contact
an authorized representative of the manufacturer or an authorized waste
management company for information concerning the decommissioning of your
equipment.
Pollution Control. (Only applicable to Peoples Republic of China (PRC)).
This symbol indicates the product contains hazardous materials in excess of the
limits established by the Chinese Standards. It must not be disposed of as unsorted
municipal waste and must be collected separately. Please contact an authorized
representative of the manufacturer or an authorized waste management company
for information concerning the decommissioning of your equipment.

SM-1103R3

General Purpose Rad Table


Service Manual

TABLE OF CONTENTS

Section
1

Page
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1.1

Objective and Scope of this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1.2

Tools and Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

PRE-INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2.1

Pre-Installation Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2.2

Responsibilities of the Purchaser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2.3

Room Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2.3.1

Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2.3.2

Floor Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2.3.3

Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

UNPACKING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4.1

Undertable Installation of the Generator (option) . . . . . . . . . . . . . . . . . . . . . . . .

12

ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

15

5.1

Brake Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

15

5.2

Pedal Microswitch Replacing and Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . .

16

PERIODIC MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

19

6.1

Maintenance Activities Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

19

6.2

Checking of Radiographic Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

19

6.3

Check Off Maintenance Revision Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

20

RENEWAL PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

21

SYSTEM INTERCONNECTION MAPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

33

SM--1103R3

General Purpose Rad Table


Service Manual

ii

SM--1103R3

General Purpose Rad Table


Service Manual

SECTION 1

1.1

INTRODUCTION

OBJECTIVE AND SCOPE OF THIS MANUAL


This Service Manual is intended to describe the installation, adjustments,
troubleshooting and periodic maintenance of the Radiographic Table.

OPERATOR AND SERVICE MANUALS SHOULD BE


CAREFULLY READ AND UNDERSTOOD BY SERVICE
PERSONNEL BEFORE USING AND SERVICING THE
EQUIPMENT,
ESPECIALLY
THE
INSTRUCTIONS
CONCERNING SAFETY, REGULATORY, DOSAGE AND
RADIATION PROTECTION. KEEP THE MANUALS WITH THE
EQUIPMENT AT ALL TIMES AND PERIODICALLY REVIEW
THE OPERATING AND SAFETY INSTRUCTIONS.

SERVICE
PERSONNEL
MUST
HAVE
SUFFICIENT
KNOWLEDGE TO COMPETENTLY PERFORM THE SERVICE
TASKS RELATED TO X-RAY DEVICES AND PARTICULARLY
TO THE EQUIPMENT DESCRIBED IN THIS MANUAL. THIS
KNOWLEDGE IS ACQUIRED THROUGH A VARIETY OF
EDUCATIONAL
METHODS
FOR
TECHNICIANS
IN
ACCORDANCE WITH LOCAL LAWS OR REGULATIONS,
INCLUDING SPECIFIC TRAINING ON THIS EQUIPMENT.

1.2

TOOLS AND TEST EQUIPMENT


The following tools and test equipment are required for the installation:

SM-1103R3

Standard service engineers tool kit.

Standard and extended levels (1 meter for Column Base).

Electric drill motor and assorted bits.

General Purpose Rad Table


Service Manual

This page intentionally left blank.

SM-1103R3

General Purpose Rad Table


Service Manual

SECTION 2

2.1

PRE-INSTALLATION

PRE-INSTALLATION CHECK
Prior to beginning installation, it is recommended to inspect the site and verify
that the X-ray room complies with Pre-installation requirements, such as:
incoming line, main switch(es), safety devices, conduits and space
requirements.

Note

2.2

For more Pre-installation requirements,


Pre-Installation Manual of the Generator.

also

refer

to

RESPONSIBILITIES OF THE PURCHASER


The purchaser is responsible for the completion of Pre-Installation. This
includes the procurement and installation of all required materials and services
to get the room ready for the installation of the product.

2.3

ROOM REQUIREMENTS

2.3.1

ENVIRONMENTAL REQUIREMENTS

Note

STORAGE values only refer to equipment that is still in shipping


containers. If the equipment is partially or completely installed,
refer to IN USE values.

RELATIVE HUMIDITY
(Non-Condensing)
IN USE
MIN.

STORAGE

MAX.
75%

30%

Note

SM-1103R3

TEMPERATURE

MIN.
10%

IN USE

ATMOSPHERIC PRESSURE

STORAGE

IN USE

STORAGE

MAX.

MIN.

MAX.

MIN.

MAX.

MIN.

MAX.

MIN.

MAX.

100%

10o C
(50o F)

40o C
(104 oF)

--40o C
(--40o F)

70oC
(158o F)

700 hPa

1060 hPa

500 hPa

1060 hPa

These environmental conditions do not include the Digital


Detector. Refer to the Digital Detector Documentation.

General Purpose Rad Table


Service Manual

2.3.2

FLOOR REQUIREMENTS
The method of installing the Table is anchor to floor with 4 x M10 bolts.
The Drill Templates of the anchoring holes are shown in the next illustrations.

Potential for Injury and/or Equipment Damage: Floor anchors


must be a minimum of 150 mm (6) from any concrete edge
including ducts and cracks. In addition, the general condition
of the concrete in the immediate mounting area should be
inspected to ensure that anchors will be set in good quality
concrete.

The floor bearing must be concrete and the thickness to be determined by a


Structural Engineer to properly support the equipment loads. The anchors
require a minimum embedment of 57.2 mm (2 1/4) into the concrete. If the floor
thickness is less than 101.6 mm (4), it is recommended that the unit be secured
using a through-bolt method with a reinforcement plate on the back side.

Note

The maximum tolerance for the Floor levelling should be


1.5 mm per meter (0.02 per feet).

Illustration 2-1
Drill Template

20
(0.8)

12
(0.5)

DETAIL A

SM-1103R3

General Purpose Rad Table


Service Manual

2.3.3

TECHNICAL SPECIFICATIONS

POWER LINE REQUIREMENTS


Input Voltage . . . . . . . . . . . . . . . . . . . . . . . . . .

24 VDC

Input Power . . . . . . . . . . . . . . . . . . . . . . . . . . .

50 VA

DIMENSIONS AND WEIGHT

SM-1103R3

Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

770 mm (30.3)

Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2200 mm (86.6)

Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

825 mm (32.4)

Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

150 kg (330.7Ib)

Dimensions of Floating Table-Top . . . . . . . . .

2200 x 825 mm
(86.6 x 32.4)

Table-Top / Receptor distance . . . . . . . . . . . .

102 mm (4)
(4, 0.15)

Table-Top Attenuation . . . . . . . . . . . . . . . . . . .

<1.0 mm Al eq.

Longitudinal travel of Table-Top . . . . . . . . . . .

900mm (35)

Transverse travel of Table-Top . . . . . . . . . . .

230 mm (9)

Longitudinal travel of Receptor . . . . . . . . . . .

518 mm (20.3)

Grid (standard) . . . . . . . . . . . . . . . . . . . . . . . . .

40 lines/cm - 12:1

Cassettes / Detector sizes . . . . . . . . . . . . . . .

from 18 to 43 cm
(from 7 to 17)

General Purpose Rad Table


Service Manual

Illustration 2-2
Dimensions

SM-1103R3

General Purpose Rad Table


Service Manual

SECTION 3

UNPACKING
The system is shipped in different crates to facilitate transport and installation.
Upon receipt of components and parts of the equipment, inspect all shipping
crates for signs of damage. If damage is found, immediately notify the agent or
carrier.
1.

Place crates close to the location in the room where the system is
intended to be placed and remove each wood--packed part. Do not
discard any packing material (envelopes, boxes, bags) until all parts are
identified as listed in the packing list.

AT LEAST TWO PEOPLE ARE REQUIRED TO REMOVE ALL


HEAVY COMPONENTS FROM THE SHIPPING PALLET.

SM-1103R3

2.

When the equipment is unpacked, check the part and serial number of
each components identification label. Inspect all pieces for visible
damages. If any damaged part is found, repair it or order a replacement
to prevent unnecessary delay in installation.

3.

Verify that all items on the customer order are present.

4.

Leave a free working area around the equipment until installation is


complete.

General Purpose Rad Table


Service Manual

This page intentionally left blank.

SM-1103R3

General Purpose Rad Table


Service Manual

SECTION 4

Note

INSTALLATION

Follow Installation sequence process as described. At least two or


three people are required for the following tasks.

1.

Remove the transport screws that secure the Radiographic Table to the
pallet and move away the Table.

2.

Remove the front cover screws of the Table in order to access inside the
Table.

3.

Power ON the Collimator Lamp and align perfectly the Receptor axes
with the Collimator light beam. During alignment, place the
Tube-Collimator Assembly and the Receptor Carriage at different
positions to check that it is correctly aligned.

4.

Mark on the floor the four anchoring holes of the Table Base.

ANCHORING HOLES

Note

SM-1103R3

5.

Move away the Table Base and prepare its anchors.

6.

Position the Table Base at its final place and anchor the Table Base to the
floor. Check that it is properly leveled placing levels in different points of
the Upper Frame. Use leveling plates to level the Base, before attaching/
fixing/ installing definitively the Table.

Use an extended level (1 meter) or several standard levels placed


in different points for leveling the Table Base.

General Purpose Rad Table


Service Manual

7.

Note

This Table can be also power supplied by a 24 VDC External


Power Supply instead of a Generator.

8.

Note

If applicable, connect the Table Receptor Cable directly to the Terminal


Block 3TS1 in the Generator, and the Table Ion Chamber Cable to the
Generator as indicated in schematic IM-341.

Refer to the Generator Service Manual for further information.

9.

10

Connect the Power Supply cable for the Table Locks from the Table TS1
to TB7 of the Lock Board in the Generator, and connect the GND cable
from the Table Base to the GND Stud in the Generator (refer to
schematics IM-341 and 54303079).

Route the cables through the rear cables outlet of the Table Base along
the room for their later connection.

10.

Re-install the Front Cover of the Table Base.

11.

Remove the two rubber stops of the same end of the Table-Top.

SM-1103R3

General Purpose Rad Table


Service Manual

12.

Install the Table-Top from one of the Table sides (right or left) inserting the
Bearings inside their Rails. At least two people are required for this
operation.

Be sure that the Table-Top slides over the bearings and the
Longitudinal Lock Pin does not block its way.

SM-1103R3

13.

Remove the wire tie-wrap of the Longitudinal Lock Pin.

14.

Re-install the rubber stops of the Table-Top end.

11

General Purpose Rad Table


Service Manual

4.1

UNDERTABLE INSTALLATION OF THE GENERATOR (OPTION)


Follow the steps below to install the Generator underneath the Table.
1.

Remove the Front Cover after unscrewing the four (4) screws.

FRONT COVER SCREWS

2.

Disconnect the ground cable and dismount the front cover.

3.

Disconnect the three faston terminals of microswitch SW1 located inside


the Table Base.

4.

Remove the four (4) allen screws of the pedal assembly at both sides of
the Table Base (a pair on each side) and pull out the pedal assembly.

MICROSWITCH SW1
PEDAL ASSEMBLY SCREWS

12

SM-1103R3

General Purpose Rad Table


Service Manual

5.

Place the Generator Cabinet (without covers) on its carriage in front of


the Table. This component is freestanding below the Table. Seismic
areas and other conditions require the Generator Cabinet to be secured
to the floor through the mounting holes on the bottom.

GENERATOR CARRIAGE

Note

SM-1103R3

Do not slide the Generator under the Table until Configuration and
Calibration procedures are completed. Refer to the Generator
Service Manual for further information.

6.

Once configured and calibrated, push the Generator inside the Table.

7.

Reassemble all of the components, parts and connections; then replace


the Front Cover of the Table.

13

General Purpose Rad Table


Service Manual

This page intentionally left blank.

14

SM-1103R3

General Purpose Rad Table


Service Manual

SECTION 5
5.1

ADJUSTMENTS

BRAKE ADJUSTMENT
Note

As brakes come factory-adjusted, field adjustment should only be


necessary when replacing Brakes.
All brakes must have a space between the brake surface and the contact
surface of less than or equal to one (1) millimeter (1mm). If the space is too
close, the equipment may drag, or movements may be difficult. If the space is
too large, the brakes may not be strong enough to block movement.
To raise, lower or level the Brake Assembly, loosen the screws indicated in the
following illustration and move the Brake Assembly to the desired position.

Illustration 5-1
Lateral and Receptor Brakes
Screws

Lateral Brake

Screws

Receptor Brake

SM-1103R3

15

General Purpose Rad Table


Service Manual

5.2

PEDAL MICROSWITCH REPLACING AND ADJUSTMENT


1.

Remove the Table Front Cover.

2.

Disconnect Microswitch cables (check previously the cable identification


to connect the cables to the new switch in the same position).

3.

Remover nut 2, for that hold nut 1 with a 14 mm. open wrench and loose
nut 2 with another 14 mm. Remove nut 2.

4.

Remove defective Microswitch.

5.

Install the new Microswitch.

6.

Adjust the Microswitch at 1 or 2 mm. from the cam to the wheel neck with
the Table Pedal fully pressed.

Illustration 5-2
Microswitch Replacement

Nut 1

Nut 2
Adjust at 1 or 2 mm

Microswitch Wheel
Wheel neck

CAM

16

SM-1103R3

General Purpose Rad Table


Service Manual

7.

Check that the rolling wheel is perfectly aligned with the cam movement
to ensure proper activation of the wheel.

8.

Avoid wrong assembling of Microswitch as shown in illustrations below:

Illustration 5-3
Wrong Installation of Microswitch

Rolling wheel not aligned

Rolling wheel not aligned

Rolling wheel too close to cam, microswitch breaks

9.

SM-1103R3

Re-install the Table Front Cover.

17

General Purpose Rad Table


Service Manual

This page intentionally left blank.

18

SM-1103R3

General Purpose Rad Table


Service Manual

SECTION 6

PERIODIC MAINTENANCE
The maintenance service has to be performed every twelve (12) months after
the installation.

6.1

MAINTENANCE ACTIVITIES DESCRIPTION


Electrical cables and connections: Visually check for electrical cables proper
isolation and proper connection inside and outside the unit.
Functional Check: Perform a functional check of the equipment, motion of
Table-top, locks, brakes and detents. Adjust or replace as needed.
Adjustments: Perform adjustments as explained in the Adjustments Section
of this manual.
Lubrication: Lubricate mobile components such as guides, bearings, etc.
Cleaning: Clean from dust or moist inner components as grids, bearings and
rails.

6.2

CHECKING OF RADIOGRAPHIC TABLE

SM-1103R3

1.

Turn OFF the system and mains. Check that there is not voltage at Table.

2.

Disassemble Front Table cover.

3.

Check anchoring screws of the Table, they should be strongly tighten.

4.

Turn On the system and press the mushroom shaped switch. Verify that
table is disconnected and the Table travels are blocked.

5.

Release the mushroom shaped switch.

6.

Connect Table power and check Table-Top brakes and correct motion by
pressing the Table Pedal.

7.

Assemble the Front Table cover.

8.

Release the horizontal Receptor brakes and check correct traveling.

9.

Take out the Cassette film tray and check correct functioning.

19

General Purpose Rad Table


Service Manual

6.3

CHECK OFF MAINTENANCE REVISION TABLE

Hospital name:

Date:

System Name:

System Number ID:

Customer:

Field Engineer Name:

Radiographic Table
Maintenance activity

20

Comments

Electrical cables and connections

Brakes, locks and detents

Functional Check

Adjustment

Replacement

Lubrication

Cleaning

SM-1103R3

General Purpose Rad Table


Service Manual

SECTION 7

RENEWAL PARTS

A -- RADIOGRAPHIC TABLE
B -- RECEPTOR ASSEMBLY

SM-1103R3

21

General Purpose Rad Table


Service Manual

ITEM

DESCRIPTION

QTY

REFERENCE

REMARKS

RADIOGRAPHIC TABLE

A1

Tabletop Kit

SAT--A11008--01

A2

Tabletop Bumper

SAT--31617--01

A3

Eccentric Alignment Bearing Kit

SAT--A9874--01

A4

Concentric Alignment Bearing Kit

SAT--A9873--01

A5

Axial Bearing Kit

SAT--A9872--01

A6

Tabletop Lineal Bearing Kit

SAT--A9842--01

A7

Receptor Brake Kit

SAT--A9871--01

A8

Receptor Bumper

SAT--51865P04

Includes 2 units

A9

Receptor Carriage Bearing Left Kit

SAT--A9869--XX

XX= 01 Bucky standard


XX= 02 BDC and BDG

A10

Receptor Carriage Bearing Right Kit

SAT--A9870--XX

XX= 01 Bucky standard


XX= 02 BDC and BDG

A11

Tabletop Brake

SAT--S0025036

A12

Lock Pin

SAT--A11007--01

A13

Microswitch

SAT--50605015

A14

Spring

SAT--11330--01

A15

Overlay Pedal

SAT--55801403

22

Includes 4 units

SM-1103R3

General Purpose Rad Table


Service Manual

A1

A4
A12

A2

A3

A5

A6

B
A8

A7

A9

A12

A10
A11

A13
A15
A14

SM-1103R3

23

General Purpose Rad Table


Service Manual

A2

A3

A6

A5

A9

A7

A10

A12

24

A4

A8

A11

A14

A15

SM-1103R3

General Purpose Rad Table


Service Manual

ITEM

QTY

REFERENCE

Vacutec Ion Chamber

6695--21

AID Ion Chamber

6695--11

Medys Ion Chamber

6695--66

Claymount Ion Chamber

6695--56

Bucky USX--Ray

SAT--6685--11

Sensing Tray

SAT--6694--05

Manual Tray

SAT--A8894--01

1B4

Push button

SAT--50613014

1B5

Handle

SAT--54402014

1B

1B1

1B2
1B3

DESCRIPTION

REMARKS

BUCKY ASSEMBLY

1B1

1B4

1B5

1B2
1B3

SM-1103R3

25

General Purpose Rad Table


Service Manual

ITEM
2B

2B1

2B2

DESCRIPTION

QTY

REFERENCE

REMARKS

DETECTOR CABINET WITH NON ROTATING TRAY


Frame 1.5mm for Grid 1m

SAT--A10521--01

Frame 1.5mm for Grid 1.3m

SAT--A9657--02

Frame 1.5mm for Grid 1.4m

SAT--A10523--01

Frame 1.5mm for Grid 1.5m

SAT--A9656--02

Frame 1.5mm for Grid 1.8m

SAT--A10522--01

Frame 2mm for Grid 1m

SAT--A10667--01

Frame 2mm for Grid 1.3m

SAT--A10666--01

Frame 2mm for Grid 1.4m

SAT--A10665--01

Frame 2mm for Grid 1.5m

SAT--A10664--01

Frame 2mm for Grid 1.8m

SAT--A10663--01

Frame 3mm for Grid 1m

SAT--A10667--02

Frame 3mm for Grid 1.3m

SAT--A10666--02

Frame 3mm for Grid 1.4m

SAT--A10665--02

Frame 3mm for Grid 1.5m

SAT--A10664--02

Frame 3mm for Grid 1.8m

SAT--A10663--02

Vacutec Ion Chamber

6695--21

AID Ion Chamber

6695--11

Medys Ion Chamber

6695--66

6695--56

6695--57

Claymount Ion Chamber


2B3

Tray Blockage

SAT--28213--01

2B4

Grid Blockage

A525510--01

2B5

Switch Kit BDC

A520491--01

2B6

Grid Guides BDC Kit

A520492--01

Includes ITEM 2B4

2B7

Rubber Bumper

SAT--6533--01

This reference comprises 5 units

2B8

Handle Kit

SAT--A10634--02

2B9

Ball Bearing Slide

SAT--54505012

2B10

Cable Support

SAT--53530150

26

Includes 10 Units

SM-1103R3

General Purpose Rad Table


Service Manual

2B1
2B2
2B6

2B5
2B7

2B9
2B3
2B4

2B10

SM-1103R3

2B8

27

General Purpose Rad Table


Service Manual

ITEM
3B

3B1

3B2

DESCRIPTION

QTY

REFERENCE

DETECTOR CABINET WITH ROTATING TRAY


Frame 1.5mm for Grid 1m

SAT--A10521--01

Frame 1.5mm for Grid 1.3m

SAT--A9657--02

Frame 1.5mm for Grid 1.4m

SAT--A10523--01

Frame 1.5mm for Grid 1.5m

SAT--A9656--02

Frame 1.5mm for Grid 1.8m

SAT--A10522--01

Frame 2mm for Grid 1m

SAT--A10667--01

Frame 2mm for Grid 1.3m

SAT--A10666--01

Frame 2mm for Grid 1.4m

SAT--A10665--01

Frame 2mm for Grid 1.5m

SAT--A10664--01

Frame 2mm for Grid 1.8m

SAT--A10663--01

Frame 3mm for Grid 1m

SAT--A10667--02

Frame 3mm for Grid 1.3m

SAT--A10666--02

Frame 3mm for Grid 1.4m

SAT--A10665--02

Frame 3mm for Grid 1.5m

SAT--A10664--02

Frame 3mm for Grid 1.8m

SAT--A10663--02

Vacutec Ion Chamber

6695--21

AID Ion Chamber

6695--11

Medys Ion Chamber

6695--66

6695--56

6695--57

Claymount Ion Chamber


3B3

Rear bumper Kit

SAT--A10676--02

3B4

Detector Detent Kit

SAT--A10677--01

3B5

Switch Kit BCG

A520490--01

3B6

Bumper Rubber

SAT--6533--01

3B7

Grid Blockage

A520476--01

3B8

Grid Guides BDG Kit

A520493--01

3B9

Handle Kit

SAT--A10634--11

3B10

Holder for Detector

SAT--A10904--01

3B11

Ball Bearing Slide

SAT--54505012

3B12

Cable Support

SAT--53530150

3B13

Turning Kit

SAT--A10654--01

28

REMARKS

This reference comprises 5 units

Includes ITEM 3B7

Only for detector without handle

Includes 10 Units

SM-1103R3

General Purpose Rad Table


Service Manual

3B6

3B5
3B8

3B7
3B1

3B2

3B11

3B12

3B3

3B9
3B10
3B13

SM-1103R3

F4

29

General Purpose Rad Table


Service Manual

ITEM
4B

4B1

4B2

DESCRIPTION

QTY

REFERENCE

DETECTOR CABINET FIXED DETECTOR AND REMOVABLE GRID


Frame 1.5mm for Grid 1m

SAT--A10521--01

Frame 1.5mm for Grid 1.3m

SAT--A9657--02

Frame 1.5mm for Grid 1.4m

SAT--A10523--01

Frame 1.5mm for Grid 1.5m

SAT--A9656--02

Frame 1.5mm for Grid 1.8m

SAT--A10522--01

Frame 2mm for Grid 1m

SAT--A10667--01

Frame 2mm for Grid 1.3m

SAT--A10666--01

Frame 2mm for Grid 1.4m

SAT--A10665--01

Frame 2mm for Grid 1.5m

SAT--A10664--01

Frame 2mm for Grid 1.8m

SAT--A10663--01

Frame 3mm for Grid 1m

SAT--A10667--02

Frame 3mm for Grid 1.3m

SAT--A10666--02

Frame 3mm for Grid 1.4m

SAT--A10665--02

Frame 3mm for Grid 1.5m

SAT--A10664--02

Frame 3mm for Grid 1.8m

SAT--A10663--02

Vacutec Ion Chamber

6695--21

AID Ion Chamber

6695--11

Medys Ion Chamber

6695--66

6695--56

6695--57

Claymount Ion Chamber


4B3

Switch Kit BDC

A520491--01

4B4

Bumper Rubber

SAT--6533--01

4B5

Grid Blockage

A520476--01

4B6

Fans Kit

SAT--A9721--01

30

REMARKS

This reference comprises 5 units

SM-1103R3

General Purpose Rad Table


Service Manual

4B4

4B3

4B5

4B2

4B1

4B6

SM-1103R3

31

General Purpose Rad Table


Service Manual

This page intentionally left blank.

32

SM-1103R3

General Purpose Rad Table


Service Manual

SECTION 8

SYSTEM INTERCONNECTION MAPS


Refer to the following maps for details of the wire connections.
SYSTEM INTERCONNECTION

Note

SM-1103R3

General Purpose Rad System Interconnection . . . . . . . . . IM--341

Radiographic Table Interconnections . . . . . . . . . . . . . . . . . 54303079

For System Interconnection Maps I/F-001, I/F-003, I/F-045,


IM-015 refer to Generator Service Manual -- Installation Chapter.

33

General Purpose Rad Table


Service Manual

This page intentionally left blank.

34

SM-1103R3

ROT MOBILE COLUMN

Note 2.- Cable code:


A3052-XX for only Low Speed
A7014-XX for Low/High Speed

See schematic 54301085


for interconnections

TUBE STATOR CABLE (See Note 2)

X-RAY
TUBE

TS2
C

TS1

VAC SUPPLY
GENERATOR SUPPLY CABLE (See Note 1)

AC LINE

Chassis
GND

COLLIMATOR CABLE (A3111-XX)

COLLIMATOR

GND CABLE (A6383-XX)

Chassis
GND

Note 1.- Cable code:


A8410-01 for Compact Generator - single phase
A8411-01 for Compact Generator - three phase
A8412-01 for Battery Generator
A8413-01 for Capacitor Generator

HV
TANK

ANODE CABLE

GENERATOR CABINET

CATHODE CABLE

TB7

SUPPLY CABLE (A6736-XX)

LOCKS
BOARD

TB7

RAD TABLE
3

See schematic 54302065


for interconnections
TS1

SUPPLY CABLE (A7235-XX)

GND CABLE (A6383-XX)

Note 3:
AEC CABLE, AEC INTERFACE CABLE AND
AEC ADAPTATION BOARD NOT USED FOR
RAD ROOM STANDARD (without ION
CHAMBER)

Chassis
GND

Chassis
GND

GENERATOR CONSOLE

BUCKY

TABLE BUCKY CABLE (A3091-XX)

See schematic I/F-001

AEC ADAPTATION BOARD


(A3263-03)

TS1

See schematic IM-015


AEC CABLE (A3253-XX)

ION CHAMBER
(Preamplifier)

J1,J2,J3

(See Note 3)

(See Note 3)

ATP CONSOLE BOARD


AEC INTERFACE CABLE (A3251-XX)

J4

TB1

(See Note 3)
J2

AEC CABLE (A3253-XX)

ION CHAMBER
(Preamplifier)

(See Note 3)

(See Note 3)
WALL BUCKY CABLE (A6703-XX)

BUCKY
See schematic I/F-002

GND CABLE (A6383-XX)


Chassis
GND

J3
(See Note 3)

POWER I/F CABLE (A3252-03)

VERTICAL BUCKY

J5

(See Note 3)

TS1/TS3

POWER CABLE (A7021-XX)

J1

COMMUNICATION CABLE (A7022-XX)

J3

INTERFACE CABLE (A7069-XX)

J2

GND CABLE (A6383-XX)

Chassis
GND

Chassis
GND

TS1

Chassis
GND

THIS SHEET ONLY APPLIES with CONVENTIONAL CONSOLE

new interface

F. Garca

10/10/05

new interface

F. Garca

30/11/04

REV

DESCRIPTION

ISSUED BY

DATE

NAME

DATE

DRAWING

F.GARCIA

05/05/04

REVISED

A.DIAZ

02/06/04

SEDECAL
D

SHEET / OF

IM - 341

1/2

RAD ROOM STANDARD


& with AEC SYSTEM
E

REV

ROT MOBILE COLUMN

Note 2.- Cable code:


A3052-XX for only Low Speed
A7014-XX for Low/High Speed

See schematic 54301085


for interconnections

TUBE STATOR CABLE (See Note 2)

X-RAY
TUBE

TS2
C
A
Chassis
GND

TS1

GENERATOR SUPPLY CABLE (See Note 1)

AC LINE

COLLIMATOR CABLE (A3111-XX)

COLLIMATOR

VAC SUPPLY

GND CABLE (A6383-XX)

Chassis
GND

Note 1.- Cable code:


A8410-01 for Compact Generator - single phase
A8411-01 for Compact Generator - three phase
A8412-01 for Battery Generator
A8413-01 for Capacitor Generator

HV
TANK

ANODE CABLE

GENERATOR CABINET

CATHODE CABLE

LOCKS
BOARD

TB7

SUPPLY CABLE (A6736-XX)

TB7

RAD TABLE
3

See schematic 54302065


for interconnections
TS1

SUPPLY CABLE (A7235-XX)

GND CABLE (A6383-XX)

Chassis
GND

Chassis
GND
BUCKY

TABLE BUCKY CABLE (A3091-XX)

See schematic I/F-001

TS1
ATP CONSOLE BOARD

AEC CABLE (A3253-XX)

ION CHAMBER
(Preamplifier)

J5

(See Note 3)

(See Note 3)

Note 3:
AEC CABLE, AEC INTERFACE CABLE AND
AEC ADAPTATION BOARD NOT USED FOR
RAD ROOM STANDARD (without ION
CHAMBER)

AEC INTERFACE CABLE (A3251-02)


(See Note 3)
2

AEC ADAPTATION BOARD


(A3263-03)

VERTICAL BUCKY

See schematic IM-015


AEC CABLE (A3253-XX)

ION CHAMBER
(Preamplifier)

J1,J2,J3

(See Note 3)

(See Note 3)

GENERATOR CONSOLE
(TOUCH SCREEN)
J4
INTERFACE BOX CABLE (A7067-XX)

J5

TB1

(See Note 3)

TPC SUPPLY CABLE (A7059-XX)

TS1
WALL BUCKY CABLE (A6703-XX)

BUCKY
See schematic I/F-002

GND CABLE (A6383-XX)


Chassis
GND

TS1

POWER I/F CABLE (A3252-03)

AC-SUPPLY

GND CABLE (A7023-XX)

Chassis
GND

(See Note 3)

GENERATOR

Chassis
GND

Chassis
GND

THIS SHEET ONLY APPLIES with TPC/SERIAL CONSOLE

new interface

F. Garca

10/10/05

new interface

F. Garca

30/11/04

REV

DESCRIPTION

ISSUED BY

DATE

NAME

DATE

DRAWING

F.GARCIA

05/05/04

REVISED

A.DIAZ

02/06/04

SEDECAL
D

SHEET / OF

IM - 341

2/2

RAD ROOM STANDARD


& with AEC SYSTEM
E

REV

You might also like