PhysioQuant Manual PDF
PhysioQuant Manual PDF
PhysioQuant Manual PDF
ENVITEC
2 / 48
User-Manual
PhysioQuant Ambulatory Blood Pressure System
The greatest care has been taken in preparing this manual. However, if you still
come across any incorrect details in this manual when using the system, then
please contact us. This will allow us to correct any errors as soon as possible.
The information and images contained in this manual are subject to any
changes that may result from optical or technological developments
All trademarks stated and used in the text are the property of the present owner
and are protected by law.
All reproduction, translation and duplication in any shape or form including
extracts require the written approval of the manufacturer.
This manual is subject to updates by EnviteC-Wismar GmbH.
Doc.No.:
45-07-0500002-a
(Printed in Germany)
Tel.:
Fax:
Email:
Website:
3 / 48
4 / 48
Contents
INSTALLATION
3.1
3.2
3.3
3.4
3.5
11
Software
Settings
Attaching the connecting cable to the recorder
Network installation
GDT interface
11
14
17
18
18
PHYSIOQUANT RECORDER
4.1
4.2
4.3
4.4
4.5
4.6
4.7
21
Operation controls
Symbols on the device
LCD display with all symbols and display options
Power supply
Inserting batteries
Function control
Connecting the cuff to the recorder
21
22
23
23
24
25
26
PHYSIOQUANT CUFFS
26
27
6.1
6.2
6.3
6.4
6.5
6.6
6.7
6.8
6.9
6.10
6.11
27
27
28
29
29
30
32
33
34
35
36
5 / 48
37
37
37
38
38
38
39
40
40
40
40
40
41
41
ERROR CODES
42
10
SCOPE OF DELIVERY
43
11
43
12
44
12.1
12.2
12.3
12.4
12.5
12.6
13
TECHNICAL INFORMATION
6 / 48
44
44
45
45
46
46
47
7 / 48
This manual forms part of the device. It should always be kept near to the
device. Correctly observing the manual ensures that the device will be
used in an appropriate manner and for the purpose it is intended. It will
also ensure the health and safety of users and patients dependent on it.
Please read through the whole manual carefully, since information that is
relevant to several sections is only provided once.
The printed text of this manual is in accordance with the version of the
device, as well as the relevant safety instructions standards, at the time
that this manual was printed. All industrial property rights are reserved in
relation to any devices, circuits, processes, software programs and names
described in this manual.
In order to ensure the highest level of safety for patients and a minimum of
interference, as well as keeping in line with the relevant testing precision
level, the device should only be used in combination with original
accessories provided by EnviteC-Wismar GmbH.
8 / 48
Intended use:
PhysioQuant is a manually-operated blood pressure measuring device,
carried by the patient for long-term measurement of non-invasive blood
pressure. It can be used for adults, children and infants by applying the
corresponding cuffs. PhysioQuant must not be used for newborn babies
and is not suitable for use in Intensive Care Units. PhysioQuant can be
used to take blood pressure measurements at various intervals for up to 30
hours and is able to store measurement results.
Biocompatibility:
The product components described in this manual, including accessories,
which when used for their intended purpose come into contact with
patients, have been explained in such a way that they conform to the
biocompatibility requirements of the relevant standard when used for their
intended purpose. If you have any questions in this regard, then please
contact EnviteC-Wismar GmbH or one of its representatives.
Cleaning:
No fluid should enter the device. If any fluid has entered the device, then it
should only be used again after having been checked and approved by the
Service Department.
9 / 48
CAUTION!
The device is not suitable for use in areas where explosions may occur.
Such areas may appear as a result of using inflammable anaesthetics, skin
cleaning and skin disinfection products.
If the information provided with the device is not clear on how to make a
safe connection that will ensure the health and safety of patients, users, as
well as the environment, then please contact the manufacturer or consult a
technical expert. At all times, observe standard IEC 60601-1-1.
Before using the device on a patient, the user should check that the device
functions safely and as intended.
The device does not contain any components that need to be replaced by
users. Never open the casing of the device (please contact the Service
Department).
10 / 48
3
3.1
Installation
Software
11 / 48
Enter your name and that of your practice and confirm with [Next].
12 / 48
13 / 48
3.2
Settings
After the programme has started, the main screen will be displayed.
To configure the software, click on the 'Settings' icon to open the Settings
menu:
Settings menu:
14 / 48
SYSTEM:
CRITICAL VALUES:
Setting the critical values for day phase and night phase:
15 / 48
PRINT:
Activate print
preview.
Selecting standard
print pages.
Headings that
should be included
with every printout.
GDT:
The description of the GDT interface and the required settings can be
found in Section 3.5 GDT interface,18.
16 / 48
3.3
17 / 48
3.4
Network installation
The PhysioQuantWin software can be used in networks. This allows for
central storage of patient information and measurements (i.e. on the
server). This information can then be accessed from all workstations.
To this end, a corresponding directory is first created on the central
server. This folder must be connected as a 'drive' on all workstations.
A local programme installation is then carried out on all workstations that
will use the PhysioQuant software (as described in Section 3 Installation,
page 11).
The folder for the shared database is then created under [Settings],
whereby selection is facilitated by the integrated Explorer functionality:
3.5
GDT interface
The GDT interface is a standard produced by the German Quality
Assurance Medical Software (QMS 'Qualittsring Medizinische
Software') for system-independent data transfer between medical
devices and general practice IT systems.
The PhysioQuantWin software has an integrated GDT interface ('Device
Data Carrier interface' 'Gerte-Daten-Trger') and therefore allows for
easy data transfer with a general practice IT system.
Please contact your general practice IT dealer to find out what the
correct settings of the GDT interface are in relation to the relevant IT
system.
All standard settings in the GDT standard (Version 2.0) can be adjusted
on an individual basis.
18 / 48
Operation
This setting activates and deactivates the GDT interface.
When the GDT interface is activated, the software will check during the
start of the programme whether the defined GDT data is available and
processes it. If no data is found, then a 'normal' programme start will take
place.
Note: During GDT operation, the option for manually adding new
patients should not be active, so as to avoid any incorrect
input or non-corresponding patient master data (general
practice IT system and PhysioQuant database).
Identification
The ID is a unique identifier that consists of a minimum of 1 and a
maximum of 8 characters, which uniquely identifies the PhysioQuant
system during GDT data transfer.
19 / 48
File names
The file names that are used for the communication between general
practice IT systems and PhysioQuant software should be entered in the
'File names' field.
File names consists of a device identification code (1 - 4 characters),
e.g. PHYQ, and a EDP identification code (1 - 4 characters) for the
general practice IT systems, e.g. EDP1.
Both these identification codes are then used to create file names, which
will always have the extension *.GDT.
Data transfer
Prior selection will determine which data is transferred to general practice
IT system.
Example Short summary (one line)
LZBDM: Day 108/ 67/ 78 -
Night 0/
0/ 78 = -100/-100/-100 %
Example Table
24 h BPM
Day phase
Night phase
06:00-21:59
22:00-05:59
Difference
Day/Night phase
Mean values:
20 / 48
Ps[mmHg
108
-100%
Pd[mmHg
67
-100%
HR[P/min
78
-100%
4
4.1
PhysioQuant Recorder
Operation controls
5
2
Cuff connection
21 / 48
6
7
On/Off switch
4.2
22 / 48
4.3
4.4
Power supply
The PhysioQuant Recorder requires either two nickel metal hydride
rechargeable batteries, or two alkaline batteries.
When starting the recorder via the PhysioQuantWin software, please
ensure that the relevant power source has been selected (see 6.6
Programming the recorder, page 30)
In addition, the device also has a built-in lithium cell to indicate the time.
The capacity of two fully-charged rechargeable batteries or two new
batteries ensures at least 30 hours of operation or 200 measurements.
23 / 48
Battery capacity reduces in line with increased use. If the capacity of the
two fully-charged rechargeable batteries clearly falls below 24 operating
hours, then you need to replace them.
Always insert two fully-charged rechargeable batteries or two new
batteries, before taking a new measurement.
4.5
Inserting batteries
The battery compartment is located on the underside of the recorder.
In order to open this compartment, use your thumb to move the battery
compartment lid approx. 6 mm backwards and then remove the lid in an
upward direction:
24 / 48
4.6
Function control
25 / 48
4.7
To connect the blood pressure cuff to the recorder, plug the metal
connection of the cuff as far as possible into the connection point at the
recorder, until it clearly clicks into place.
The cuff can be removed again from the recorder by pulling back the
external metal plug cover.
PhysioQuant Cuffs
Various cuff sizes are available for the PhysioQuant (standard, large and
children size).
Please select the correct cuff size (see printed text on the cuff)
Cuffs that are too small will result in measurements that are too
high. Cuffs that are too large will result in measurements that are
too low.
Replace the cuffs at regular intervals. Damaged Velcro fasteners
may lead to incorrect measurements.
26 / 48
6
6.1
6.2
PC
>
>
>
27 / 48
6.3
Selecting a patient
The selection window of the patient database appears:
In the 'NAME' field, enter the first letter of the patient's surname. All
patients whose names start with this letter will then be displayed (e.g.
those beginning with P ):
28 / 48
6.4
6.5
29 / 48
6.6
30 / 48
every 15 min.
250 mmHg
every 30 min.
250 mmHg
NB:
Only when a new long-term measurement is started, will the recorder
delete all old patient data and measurement values.
31 / 48
6.7
CAUTION!
32 / 48
6.8
Test measurement
A test measurement must always be carried out after the cuff has been
attached and the recorder connected.
In order to avoid any incorrect measurements, please ensure that the
patient remains calm while the measurement is being taken. The patient
may stand up or sit down.
Begin the measurement by pressing:
After a brief moment, the device will inflate the cuff. Once the required
inflation pressure has been reached, the device will gradually decrease
the pressure. Cuff pressures are displayed and for every detected
oscillation the letter 'M' will be displayed. Once the measurement has
been taken, the following information is displayed sequentially:
33 / 48
6.9
34 / 48
Take off the cuff if it remains inflated after pressing the Start/Stop
button. It may be that the tube has been disconnected;
Further measurements should only be carried out after the cuff has
been correctly attached again.
35 / 48
Note:
You can only manually switch between day and night phase when 2
measurement intervals only have been programmed via the
PhysioQuantWin software when the starting sequence of the recorder
was commenced.
If more or less measurement intervals have been programmed, then the
Day/Night button has no effect on the measurement intervals.
36 / 48
7
7.1
7.2
PC
>
>
>
The measurement values are imported from the recorder and assigned
to the relevant patient.
The graphical representation of the measurement results is also
displayed (see 8.5 Single-value graph, page 40).
37 / 48
8.1
8.2
38 / 48
8.3
Representation
The saved measurement values can be represented in various ways, by
clicking on the relevant tabs.
39 / 48
8.4
Overview
The overview displays a numerical summary of the long-term blood
pressure measurements, as well as the statistical measurement results,
for the whole monitoring period, according to day and night phase.
8.5
Single-value graph
This graph displays the results of all individual measurements. The set
critical values for the day and the night phase (see 3.2 Settings, page 14)
are displayed as red lines.
8.6
Single-value table
All measurement results are listed by date/time, systole, diastole, heart
rate and mean pressure.
Additional measurements (that have been carried out manually by using
the Start/Stop button) are displayed behind the time indication and
marked with the '+' symbol.
Deleting individual measurements
In order to delete individual measurements, you need to click on the
relevant line (the line will appear in blue). Then click on [Delete]. After a
confirmation message appears, the measurement will be deleted.
8.7
8.8
Findings report
You can create a small report in relation to the findings of the long-term
blood pressure measurements. This text will be summarised and can
when activated also be forwarded to the general practice IT system.
40 / 48
8.9
Printing
You can print every page displaying measurement results by using the
[Printing] button.
The size of the printout can be determined on an individual basis. You
can define a standard format under 'Settings' (see 3.2 Settings, page
14).
The integrated printing preview allows you to view the selected pages
prior to printing.
41 / 48
Error codes
E 03
E 04
E 05
E 06
the device is not inflating beyond the set maximum pressure and is
waiting for the next measurement time.
E 07
E 08
E 10
Memory full. 200 blood pressure measurements have been carried out
and the memory of the device is full.
E 11
E 12
E 20
E 21
E 22
E 24
Difference between systole and diastole is too small (10 mmHg or less)
E 23
42 / 48
10
Scope of Delivery
PhysioQuant System
45-00-0501
Includes:
PhysioQuant, standard adult blood pressure cuff, PhysioQuantWin
software, interface cable, carrier bag, belt and manual.
PhysioQuant Recorder
45-00-0502
Includes:
PhysioQuant, standard adult blood pressure cuff, carrier bag, belt and
manual.
11
45-00-0509
45-00-0510
Cuff Child
45-00-0511
Interface cable
45-00-0515
Bag
45-00-0513
Belt
45-00-0512
43 / 48
12
CAUTION!
Use only a moist cloth and ensure that no fluid enters the device.
The most commonly used cleaning and disinfection agents in
practices and clinics can be used.
CAUTION!
44 / 48
After you have cleaned the cuff, you need to rinse it thoroughly
with water and let it dry at room temperature for approx. 15 hours.
You can also use any of the following to disinfect the device:
isopropyl alcohol 70%, ethanol 70%, Microzid; Burazon liquid,
Sporicidin or Cidex. After you have disinfected the cuff, rinse it
thoroughly with water and leave it to dry at room temperature.
Rub the tubes with a moist cloth (soapsuds). NEVER dip the tube
into any fluid.
Prior to using the device, manually check the device for any
mechanical damage.
45 / 48
Connect the pump ball over the T-piece between the supplying
tube and cuff.
46 / 48
13
Technical Information
Measurement
Measurement method
Measurement duration
Data preservation
Measurement intervals
Capacity
oscillometric
30 45 sec. (depends on patient)
unlimited
2 90 min., programmable
200 measurements or 30 hours
Measurement range
Systole
Diastole
Heart rate
60 260 mmHg
40 220 mmHg
35 240 beats/min.
Cuff
Connection
Size
Cuff pressure
Display/operation controls
Patient display
Operation controls
Interfaces
PC connection
Other
Dimensions (L x W x H)
Weight
Operating temperature
Rel. humidity
Air pressure
Power supply
47 / 48
ENVITEC
Envitec Wismar GmbH
Alter Holzhafen 18
23966 Wismar, Germany
Tel.:
+49-(0)-3841-360-1
Fax:
+49-(0)-3841-360-222
Email: [email protected]
Website: www.envitec.com
48 / 48