Manual Ifs
Manual Ifs
SECTION 1
CONSTRUCTION AND OPERATION
A. Components and Function
SECTION 2
COMPACT PENCIL NOZZLE SERVICING
A. Special Service Tools
B. Preparation and Cleaning
C. Hydraulic Lock Up Elimination
D. Testing
E. Needle Lift Adjustment
F. Adjusting Opening Pressure
G. Replacement of the Carbon Stop Seal, Plain
Washer and O-Ring Seal
SECTION 3
TEST BENCH REQUIREMENTS AND
PROCEDURES
A. Special Test equipment Requirements
B. Cam Box Mounting Instructions
C. IFS Test As Recieved
D. System Settings: Following System Service
E. Product Preparation for Returning to
Customer
SECTION 4
TABLES
A. Identification and Cross Reference Table
B. Compact Pencil Nozzle Shims
C. Unit Pump Timing Shims
D. Torque Values and Gap Settings
i
GENERAL Unit Pump/System Identification
A. Purpose of the Manual
Eight Digit
This manual is expressly intended to provide Stanadyne
qualified, trained diesel fuel injection techni- System S/N
cians the information necessary to prepare to
and test the Standyne Integrated Fuel Sys-
tem. In addition to this manual all special tools
and test equipment as outlined herein are
needed, as well as all service pertinent litera-
ture and individual pump specifications con- X
X
ii
SECTION 1 - CONSTRUCTION AND on the plunger position, the charging port in
the pump body.
OPERATION
During the charging cycle the unit pump tappet
follows the backside of the cam, allowing the
Compact Pencil Nozzle
spring-loaded plunger to move downward. As
the plunger moves downward the helix’s top
edge uncovers the inlet port, allowing inlet fuel
to fill the pumping chamber (Fig. 1.2).
Pumping
Chamber
Helix
1.5mm
Fig. 1.7
Fig. 2.3
Fig. 2.1
4
D. Testing 4. Seat Condition – With the nozzle tip
1. Using the 38605 CPN Calibration Line, pointed downward and the gauge closed,
connect the CPN to the nozzle tester with the operate the tester rapidly to firmly seat the
CPN tip angled downward (Figure 2.4). valve. Dry the nozzle tip thoroughly. Open the
gauge valve and raise the pressure to 250 -
350 PSI (17-24 bar) below the measured
opening pressure. While maintaining the
pressure for 10 seconds, no droplet should
separate from the nozzle tip. Slight dampness
or a drop forming is permissible with a used
nozzle.
3. Nozzle Opening Pressure – Open the Step 4 Remove the Lift Stop, Pressure
nozzle tester gauge valve and raise the Adjusting Shim, Pressure Adjusting Spring and
pressure slowly until the nozzle valve opens. Spring Seat from the Nozzle Holder Body
Note what the maximum pressure reading is Assembly. Reference Figure 2.5.
on the gauge before the pressure sharply
drops off. Repeat several times and if large NOTE: Because components of individual
variations between these readings are noted, CPN’s are matched at the factory to control
disconnect the CPN from the nozzle tester, flow and needle lift, never intermix the
drain the cap nut as described in the Hydraulic components from one CPN with another.
Lock Up Elimination section and retest.
5
Lift Stop Pressure Spring Fig. 2.5
Cap Nut Adjusting Pressure Seat
Shim Adjusting Cap Nut O-ring
Spring
6
Step 2 Place the carbon stop seal
installation tool over the nozzle tip and slide
the carbon stop seal over the 36528
installation tool (Reference Figures 2.6 and
2.7 ) until the seal drops into the groove on
the nozzle holder body assembly. Remove and
discard the installation tool.
Fig. 2.7
7
SECTION 3 - TEST BENCH
B. Cam Box Mounting
REQUIREMENTS AND
PROCEDURES These instructions are for a typical test
bench set-up. Additional pipe/tube
fittings may be required in some cases
A. Test Equipment to properly connect the cam box to the
• Test bench: test bench.
10 HP direct drive Step 1 Remove the Cam Box
Lube oil pressure capability: subassembly from the storage case and
3.5 Bar (50 PSI) inspect the contents.
Mounting plate: 107mm pilot dia.
Drive coupling: 30 mm taper Mount the cam box to an appropriate
107 mm pilot diameter mounting plate
• Laptop Computer: (Not included) (Fig. 3.2) and install the mounting plate
DAQ Card and DAQ Cable and cam box assembly onto the test
Stanadyne Proprietary DE Fuel bench pedestal.
Injection System Software
CAUTION: The cam box assembly is
• 39105 IFS Service Equipment Kit heavy - approx. 68 lbs (30.8 kg).
(Reference Fig. 3.1)
1 39106 IFS Cam Box Fig. 3.2
2 39142 Hose & Fitting Kit
3 39152 Hex Bit Socket Assy. Cam Box
Subassembl
4 39155 Cable, Pump Interface
Module
5 39156 Pump Interface Module
6 39234 Tappet Extraction Tool
Mounting Plate
7 — IFS Cam Box Storage
Case
99972 IFS Software CD-ROM Step 2 Install the drive coupling,
(Not shown)
39136 lock washer, and 39135 hex nut
(Fig. 3.3) onto the drive shaft. Tighten
Fig. 3.1 the nut to 100 ft-lbs (135 N·m). NOTE:
Drive Coupling orientation to the Drive
5 Shaft key slot is not required.
7
1
Fig. 3.3
Cam Box
6
Hex Nut
2
Drive Coupling
Lock Washer
4 (30mm Taper)
3 P/N 39136
8
Step 3 Position the mounting Install the pipe-to-tube adapter (3/8”
pedestal and connect the drive coupling NPT to 3/8” Tube) into the lube oil return
to the test bench. port (Fig. 3.6) and tighten it securely.
Step 4 Assemble the CPN Nozzle Calibrating Oil
Block onto the cam box (Fig 3.4). Index Return Port
the diamond shaped alignment pin and
install the two 5/16–18 x 1.5" cap screws. Steel Braided
Hose
Tighten the cap screws to 180-220 lbf-in
(20-25 N·m).
CPN Nozzle
Block Lube Oil
Return Port
Fig. 3.6
Alignment
Install the pipe nipple (3/8” NPT to 1/4”
NPT - Fig. 3.7a), into the Cold Start
Mounting Advance (CSA) port (Fig. 3.7b) and
Screws tighten it securely. Turn the Lube oil
(5/16-18x1.5”) Fig. 3.4 Supply 90˚ elbow until it is tight in the
position shown in Fig. 3.5 & Fig.3.7b.
Install the 39142 Hose and Fitting Kit Securely assemble the remaining
components in Figure 3.7a into the
NOTE: Use a suitable thread sealant
configuration shown in Fig. 3.7b.
on all tapered pipe thread (NPT)
fittings. Pipe Nipple
9
Before tightening the remaining fittings,
determine a safe routing for the test Nozzle Block
bench's Lube Oil and Calibrating Oil
tubing. Tighten the brass fittings SOI
securely in the cam box and connect the Pressure
Transducer
test bench hoses.
Adjust lube oil needle valve: Turn the Flow
needle in until it lightly bottoms and then Restriction
backing it out 1/4 to 1/2 turn. Fig. 3.8 Connector
NOTE: Only a small amount of oil flow is precise rack position orientation and
required for proper lubrication. Should consistency between cam boxes. The
the flow of lube oil become excessive, offset dimension is stamped on the cam
the cam shaft cavity will fill and overflow box nameplate as shown in Fig. 3.9. and
through the rack adjusting hole. Should is pre-set at the factory.
this happen, close the needle valve Micrometer Offset
slightly.
10
If the micrometer offset position is magnetic pick-up (3). Connect the DAQ
incorrect, the micrometer must be cable (5) to the Pump Interface Module
repositioned in the micrometer adapter (4) and to the DAQ card in the
as follows: computer (6).
Ensure the o-ring seal in the micrometer Fig. 3.13
2
adapter is removed. Adjust the 6
micrometer to the factory offset
dimension. Loosen the set screw. Hold 1
the adapter flush to the flat surface and
firmly push the micrometer body down in 3
the adapter until the spindle contacts the
5 4
flat surface. Verify the micrometer offset
reading and tighten the set screw
securely. Step 11 Load the IFS Service
Equipment Software 99972 by inserting
Step 8 Insert the o-ring seal (P/N the CD-ROM into the CD drive. The IFS
30343) into the groove in the Calibration Software installation wizard
micrometer adapter. Using the two 1/4”- will guide you through the installation
20 x .75” screws, mount the micrometer process which only takes a few minutes
adapter to the cam box (Fig. 3.12). to complete.
Tighten the screws to 60-70 lbf-in (6.8-
7.9 N·m). The operator will be requested to verify
a destination location for the programs
Fig. 3.12 installation. The default location is
Micrometer Mounting
C:\Program Files\IFS Calibration\.
Screws Once the installation is complete, a
(1/4”-20 x .75”)
shortcut icon is automatically placed on
the computers desktop.
Step 12 To access the IFS program,
select the IFS icon on the desktop and
open the program. The first time the
Step 9 Slowly rotate the drive while program is opened, the operator is
looking into the TDC sensor hole on the requested to input a “TDC Offset
cam box, visibly center the pin on the number” (Fig. 3.14). This number is
cam shaft in the hole. Thread the TDC
sensor all the way in until it lightly
contacts the pin then back it out
approximately 1/8 - 1/4 of a turn.
Holding the sensor stationary, tighten
the jam nut securely.
Step 10 As shown in Figure 3.13,
connect the Pump Interface Module
Cable (1). Connect the cable to the
39156 Pump Interface Module (4), the
SOI pressure Sensor (2) and the TDC Fig. 3.14
11
specific to each cam box and is Step 1 Inspect the cam follower
stamped on the cam box nameplate in roller for rotation and condition. Ensure
the location shown in Figure 3.9. Once the CSA piston moves within the tappet
this information has been entered it will body bore. Inspect the cam follower
be used as the default value and will springs on the unit pump to ensure that
not need to be entered again for future neither the inner or outer springs are
tests. broken.
Step 13 To begin a test sequence, Step 2 Check the pumping plunger
select the “New Test” button. Some for rotation without binding (Fig. 3.15).
basic identification data must be entered
Step 3 Verify that the upper spring
and the specific model number and test
seat slot location is within the width of
routine selected prior to testing.
the alignment hole diameter in the pump
Because the IFS serial number is used body (Fig. 3.15)
in the test results file name, the number
must be entered twice to verify that it is Pump Fig. 3.15
entered correctly. Once the serial Body
number entry has been verified it will Alignment
O-ring Hole
change to a Green font color.
P/N 35638
Upper
Service Dealer identification numbers Spring Seat
are matched to an internal table of Slot located
authorized service agencies. Once the O-ring
within the
number has been verified it will change width of the
P/N 35639
in font color to Green. Once this Alignment
Hole
information has been entered it will be Diameter
used as the default value and will not Plunger
need to be entered again for future should
tests. Rotate Upper
Freely Spring
The IFS model number that is to be without Seat
tested is chosen from a pull down menu, Binding Slot
accessed by selecting the arrow in the
Step 4 Measure and record the
“Pump Model Number” field.
shim thickness if supplied (Fig. 3.16).
Choose the desired test routine from the Otherwise, use a 1.65mm thick timing
pull down menu activated in the “Select shim (P/N 35998) for initial testing*.
Test Routine” field. * For this reason timing is not used
NOTE: The Manual Mode is not as a consideration for warranty.
functional at this time.
Fig. 3.16
C. System Calibration Checks: As Digital Calipers
Received for Service
When an IFS system is received for
service it should be cleaned and visually
inspected prior to mounting in the IFS
cam box. Timing Shim
12
Step 5 Replace the upper o-ring Step 4 Center the position of the
seal (Fig. 3.15) P/N 35639 (brown) and control rack metering arm pin slot within
the lower o-ring seal P/N 35638 (black). the pump mounting bore (Fig. 3.18).
Mounting the Unit Pump into the IFS Pump Mounting Bore
Cam Box:
Step 1 Remove the CPN from the
Unit Pump.
Control Rack Metering
Step 2 Look into the pump mounting Arm Pin Slot Centered in
hole and rotate the drive until the base Pump Mounting Bore
circle (low side) of the cam lobe is
visible in the hole.
Step 3 Place the tappet assembly Fig. 3.18 Control Rack
on to the magnetic end of the 39234
Tappet Extraction Tool (Fig. 3.17).
Orient the CSA feed hole position to the Step 5 Lubricate the upper and
flat side of the tool. In this position the lower o-ring seals on the pump body
tappet’s guide pin slot will align with the (Fig. 3.19).
guide pin in the tappet bore.
Step 6 Position the Metering Arm so
Fig. 3.17 Tappet that the metering arm pin lines up with
Flat Side Installation the pump body fixture hole as is shown
and Removal in Figure 3.19.
Tool
P/N 39234
Alignment
Hole
Tappet CSA Feed Orient the
Assembly Hole Lubricate Metering Arm
o-rings Pin in line
with the
Pump Alignment
Mounting Hole in the
Bore Pump Body
Metering Arm
Pin
13
Step 3 Re-check the clocking
dimension. Turn the micrometer thimble
out several turns. Turn the micrometer in
Unit Pump
Clamp until it stops. A reading of between 2.63
and 2.81mm should be attained. If the
Draw Unit Pump Down,
Tighten Screw Snugly then
micrometer reading is outside of these
Loosen and Re-Tighten to parameters, return to Step 1.
10 to 15 lbs-in (1.1 –1.7 N·m) Fig. 3.20 Mount the CPN
Step 1 Remove the Carbon Stop Seal
Step 9 After seating the pump body (P/N 16389) and Plain Washer (P/N
in the cam box, loosen the hold down 35686) as shown in Figure 3.22.
screw slightly and then tighten it to
approximately 10 to 15 lbs-in (1.1 –1.7
N•m) as shown in Figure 3.21. This will Plain Washer
provide a slight drag while rotating the (P/N 35686)
unit pump during the clocking
procedure.
Carbon Stop Seal
Fig. 3.22 (P/N 16389)
14
Step 4 Allow the Carbon Stop Seal Step 2 Select the appropriate IFS
to relax for a few minutes. Then use the system number and test routine in the
Carbon Stop Seal Re-sizing Tool in a software and follow the instructions
slight rocking motion to squeeze the displayed in the text box of the IFS
seal into its final size (Fig. 3.24). program.
Carbon Stop Seal Re-sizing Tool Step 3 Unit pump assemblies may
P/N 36474 require setting of the metering arm
maximum and minimum stop positions.
NOTE: Replacement unit pump
assemblies are not preset at the
Fig. 3.24 factory and w ill require this
adjustment. Original equipment unit
pump assemblies will have an
Step 5 Apply a thin coat of assembly
alignment line scribed on the pump
grease to the carbon stop seal.
body and upper spring seat (Fig. 3.25).
Step 6 Guide the CPN nozzle tip
into the nozzle block, align the CPN inlet Unit Pump Body
tubing with the discharge fitting of the Alignment
pump. Line
15
If the fuel delivery is acceptable, the micrometer out, then turning it in until
program will prompt the user to proceed it stops. Repeat the upper spring
to Step 3d to record it. If the fuel delivery seat setting if necessary.
is outside of the given parameters, the
Spring Seat Torque
user will be requested to proceed to Step Tool
3a where the upper spring seat is P/N 39152
adjusted as follows:
a. Adjust the bench speed to 0 RPM.
Loosen the CPN Line Nut and the
Unit Pump Clamp. Rotate the Unit
Pump in the direction shown in Fig. 3.27
Figure 3.26 to adjust the fuel
delivery. d. Set the micrometer position and the
test bench speed as specified in the
NOTE: The amount of pump rotation
test routine. Record the fuel
required for adjustment is very small.
delivery. If the fuel delivery is
Tighten the unit pump clamp and incorrect return to step a.
CPN line nut.
Step 4 The rest of the test routine
Fig. 3.26 includes check points for fuel delivery,
cold start advance operation and
injection timing at various speeds.
16
If the injection timing is out of tolerance, program will check the timing and
the program will calculate the either prompt the user to repeat the
dimensional difference of the timing timing shim replacement or
shim in use versus the shim size that is proceed to the next step.
required. A positive or negative number
in millimeters will be displayed. If Cold Start Advance Operation
positive add or, if negative subtract that
amount to or from the existing shim The test routine verifies cold start
thickness (Reference Fig. 3.16, Page advance operation by prompting the
12). Using the resulting thickness user to open the CSA activation valve
dimension determine the correct shim (Reference Figure 3.7b, Page 9).
part number from the Unit Pump Timing
Shims table at the end of this manual or Should the CSA fail to activate, ensure
on the individual IFS service that the lube oil needle valve is set to
specification. Replace the timing shim 1/2 turn out from the seated position. If
as follows: the CSA continues to fail operationally,
inspect the tappet assembly and CSA
a. Remove the CPN from the unit piston. If the piston fit is incorrect (very
pump and remove the unit pump loose, difficult to move by hand or
assembly from the Cam Box. seized in the bore) the tappet assembly
will need to be replaced.
b. Visually check the location of the
upper spring seat. If the seat Step 5 For the remainder of the
position is not within the acceptable system settings follow the latest edition
limits as shown in Figure 3.15 on of the individual IFS specification and
page 12, the IFS assembly is worn the appropriate test routine from the
beyond acceptable limits and latest version of the IFS service
should be replaced. equipment program.
If the upper spring seat position is
within the limits, replace the timing E. Preparation for Returning an IFS
shim with the correct shim. to the Customer.
The IFS seal kit (P/N 36529) contains
NOTE: If shim replacement is not the seals necessary for unit pump and
required, the inspection of the CPN engine installation. A protective
upper spring seat location will sleeve and shipping cap for the CPN are
need to be performed when the also provided to prevent contamination
pump is removed at the during handling.
conclusion of the test routine.
1. Replace the two o-ring seals on the
c. Mount the IFS assembly in the cam
unit pump.
box and clock unit pump assembly
as described earlier in part “C” of 2. Replace the carbon dam seal and
this section. the plain washer on the CPN.
d. Return the micrometer position and 3. Carefully put the protective shipping
test bench speed to the required cap (P/N 26888) over the valve end
settings, select “Verify Shim”. The of the CPN.
17
4. Hand tighten the CPN to the Unit
Pump Body.
5. Slide the Roller Tappet Assembly
into the woven protection sleeve,
then slide the Unit Pump spring into
the other end of the sleeve.
6. If possible, it is also recommended
that each assembled IFS be placed
in a poly bag.
18
SECTION 4 - TABLES
A. Identification and Cross Reference Table
37741 RE527598 4T/5T OEM, 317, 320, 325 SSL & 304J, 244J,
4WDL
37742 RE527599 5TW OEM, 328 Skid Steer Loader
38225 RE530790 4T/4TW, 5T OEM
36597 K 38226 RE530792 5TW OEM
38234 RE530795 1800 Generator Set
38236 RE530797 1500 Generator Set
38237 RE530800 4T/5T OEM, 317, 320, 325 Skid Steer Loader
38238 RE530801 5TW OEM, 328 Skid Steer Loader
19
C. Unit Pump Timing Shims
10. Lube Oil Needle Valve Turn in till contact, back out
1/2 turn
11. TDC Sensor Air Gap .005 in. (.12mm), Turn in till
contact, back out 1/4 turn
20
Stanadyne Corporation
92 Deerfield Road, Windsor, CT 06095, U.S.A
Tel: (860) 525-0821
Fax: (860) 683-4581
www.stanadyne.com
Stanadyne S.p.A.
Via Matteotti 158
25014 Castenedolo (Brescia), Italia
Tel: (39) 030 213 0070