Health Technical Memorandum 01-01
Health Technical Memorandum 01-01
March 2016
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
ii
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Preface
Introduction
This HTM supersedes the Choice Framework
for local Policy and Procedures (CFPP) series,
which was a pilot initiative by the Department of
Health.
The CFPP series of documents are reverting to
the Health Technical Memorandum title format.
This will realign them with HTM 00 Policies
and principles of healthcare engineering and
HTM 01-05: Decontamination in primary care
dental practices and the naming convention
used for other healthcare estates and facilities
related technical guidance documents within
England. It will also help to address the
recommendation to align decontamination
guidance across the four nations.
In 01-01 and 01-06 DH will be retaining the
Essential Quality Requirements and Best
Practice format, this maintains their alignment
with HTM 01-05 and the requirement of The
Health and Social Care Act 2008: Code of
Practice on the prevention and control of
infections and related guidance which requires
that decontamination policy should
demonstrate that it complies with guidance
establishing essential quality requirements and
a plan is in place for progression to best
practice. We are aware that policy within the
devolved nations differs on this particular issue
but the aim is that the technical content should
be consistent and able to be adopted by the
devolved nations so that the requirements of
the ACDP-TSE Subgroups amended guidance
can be met.
iv
Purpose
The purpose of this HTM is to help health
organisations to develop policies regarding the
management, use and decontamination of
reusable medical devices at controlled costs
using risk control, which will enable them to
comply with Regulations 12(2)(h) and 15 of the
Health and Social Care Act 2008 (Regulated
Activities) Regulations 2014 .
This HTM is designed to reflect the need to
continuously improve outcomes in terms of:
patient safety;
clinical effectiveness; and
patient experience.
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Abbreviations
vi
Executive summary
vii
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
viii
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Acknowledgements
Contents
Preface iv
Abbreviations vi
Executive summary vii
Acknowledgementsx
1 Introduction1
2 Decontamination equipment: test equipment and materials2
3 Design and pre-purchase considerations 10
4 Validation and verification20
References34
xi
1 Introduction
1 Introduction
Medical devices
Definitions
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
General considerations
2.1 This chapter reviews the key items of
portable test equipment necessary to carry out
the test procedures described in this guidance.
Specifications for instruments fitted
permanently to decontamination equipment are
given in the relevant British, European and
International Standards.
2.2 Instrumentation technology continues to
advance rapidly, making it increasingly difficult
and undesirable to provide detailed
specifications for the equipment to be used in
testing equipment. There is a clear trend
towards computer-controlled data recorders
with software, which enables the system to
verify attainment of the required conditions and
then to produce a detailed written report
accompanied by tabulated or graphed data.
Although these new systems may offer
advantages in clarity of presentation, as well as
reduced operator time, the traditional
instruments, such as chart recorders, remain
equally acceptable where they meet the
accuracy defined in this section.
Note
Retention of data for long-term use is
important. Where modern technology/
datarecording equipment is used, it should
be equipped with memory devices that
enable data to be retrieved at a later date.
Traditional chart recorders allow the retention
of the chart for long-term storage and followup.
2.3 The objectives of this section are both to
ensure that traditional measurement methods
are supported adequately and to define clearly
the essential requirements that apply to the test
equipment whether it is a traditional system or
the latest technology.
2.4 When it is proposed to use measurement
and/or recording techniques that are not
covered in this guidance, the advice of the
Authorising Engineer (Decontamination) should
be sought.
2.5 All test equipment should be calibrated by
a UKAS-accredited laboratory in accordance
with ISO/IEC 17025, with traceability to the
National Standard.
Data recorders
2.12 Test recorders are required to measure
temperature in all types of decontamination
equipment and may also be required for the
measurement of pressure, flow rates, and other
critical parameters. They should be designed
for use with the appropriate sensors,
independent of those fitted to the machine.
2.13 Sufficient connections to meet the testing
requirements of the relevant HTM and EN
standard should be provided.
2.14 Recorders should comply with the
requirements of BS EN 285 or BS EN ISO
15883.
2.15 Data from digital recorders (dataloggers)
can be presented graphically or as a listing of
numerical values, or as a combination of both.
In many cases, parts of the operating cycle can
be expanded and replotted for closer
examination.
2.16 Digital recorders should have the facility to
record data immediately that can then be
removed for secure storage. Alternatively, the
recorder may be connected to a central
computer and the data recorded to the hard
drive. Software used with digital recorders
should be developed and validated under a
recognised quality system (see BS EN ISO
13485).
2.17 The detailed specification for a test
recorder will depend upon the range of
equipment with which it is to be used. The
measurement system (recorder and sensors)
should be capable of measuring cycle variables
to the required accuracy of the instruments
fitted to the machine.
3
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Temperature measurement
Temperature sensors
2.23 Temperature sensors should be used to
sense the temperature in locations specified in
the tests described in this guidance. The
sensors should be either platinum resistance
elements and comply with class A of BS EN
60751 or thermocouples and comply with
Tolerance Class 1 of BS EN 60584-2 and have
a response time in water of t90 0.5 seconds.
2.24 Thermocouple wire should be singlestrand, not exceeding 0.7 mm diameter over
the covering of one core of a twin cable. The
width of the cable should not exceed 2 mm. If
bulkier cable is used, the tracking of steam
along the outside of the cable may invalidate
certain tests, such as those which require
temperatures to be measured in the centre of a
standard test pack (the standard test pack is
described in BS EN 285).
Instrument verification
Use of sensors
2.30 Guidance on use of sensors for sterilizers
and WDs is provided in BS EN 285 and BS EN
15883 Parts 1 and 2.
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Self-contained systems
2.33 Where it is impractical to insert sensor
leads into processing equipment, self-contained
datalogging devices may be used.
2.34 A number of different designs of small selfcontained single channel data loggers for the
measurement of temperature are commercially
available. They are independently powered,
may be programmed to take readings at the
required rate for the required duration, and are
downloaded onto a computer on completion of
the datalogging period.
2.35 Care needs to be taken in selecting units
that are capable of withstanding the high
temperatures found in sterilizers and washerdisinfectors.
2.36 The accuracy, resolution and sampling rate
requirements should be identical to those
specified for conventional recorders (see
Chapter 4, Validation and verification).
2.37 Where two or more dataloggers are used
together on the same process, the time bases
of the instruments should be synchronised.
Pressure measurement
Measurement ranges
2.38 Pressure measurement ranges for WDs
should be up to 1000 kPa (10 barA) (for
example for the water supply pressure).
Differential pressures of 0.1 to 10 kPa ( 100
mbar) may need to be measured (for example
for the determination of the pressure drop
across filters).
2.39 Pressure measurement ranges for steam
sterilizers may be from 3 to 10 kPa (in vacuum
leak testing) to typically 400 kPa at the working
pressure of a high-temperature steam sterilizer.
Sensors and gauges
2.40 Pressure sensors should be used in the
tests described in this guidance and should
conform to BS 6447. The natural frequency of
6
Mark
interval
Calibration
Application
0 to 16 kPa
0.1 kPa
Gas
Vacuum leak-testing
0 to 100 kPa
1 kPa
Liquid
Differential pressure
across water filters
0 to 500 kPa
5 kPa
Liquid
Steam sterilizers
0 to 1000 kPa
5 kPa
Liquid
Water supply
pressure.
Recirculating pump
pressure
0 to 50 kPa
1 kPa
Air
Differential pressure
across air filter
Flow measurement
Water
2.45 The volume of water used for each stage
of the operating cycle may be measured using
a water meter complying with ISO 4064-1 Class
A.
2.46 The meter should be designed to operate
at temperatures up to 90C with a supply
pressure up to 1700 kPa (16 bar).
2.47 The meter should have a minimum scale
division of 0.1 L or less and be designed to
measure flow rates over the range 1 L/min to
25 L/min.
2.48 A single jet turbine system is sufficiently
accurate for the purpose. Other systems such
as multi-jet turbine or semi-positive
displacement systems complying with ISO
4064-1 Class B or Class C may also be used.
2.49 The calibration of the flow meter should be
verified by comparing the indicated flow rate
with a measured volume collected over a
measured time period. The collected volume of
liquid may be determined by either gravimetric
or volumetric measurement. The gravimetric
method is generally more accurate as the
temperature of the liquid increases.
Chemical additives
2.50 The volume of chemical additive used for
each stage of the operating cycle may be
measured using a flow meter. A number of
commercially available flow sensors designed
to monitor flows in the range 0 to 2 L/min are
suitable for interfacing to a recorder or
datalogger.
2.51 The sensor should be suitable for use with
fluids having viscosity in the range 0.8 to 20
centiStokes and should be calibrated for the
viscosity of the fluid to be measured.
2.52 The sensor should be designed to operate
at temperatures up to 70C with a supply
pressure up to 1100 kPa (10 bar).
Note 1
When the meter is connected in the pipe
there will be a noticeable pressure drop
across the meter. Although this should be
less than 1 bar it may interfere with the
normal operation of the washer-disinfector
and therefore should not be used during
tests for other characteristics than the volume
of water used.
Note 2
A meter of the rotating vane type calibrated
using water at 20C as the flowing medium
and then subsequently used to measure the
flow of a detergent solution with a viscosity of
30 centiStokes would have an error of 15
20% if no correction was applied.
Volume measurement
2.56 The volume of chemical additives and the
volume of water used in each stage are critical
variables in the control of the washingdisinfecting process.
2.57 The volume used may be measured
directly by collection in a graduated vessel of
appropriate size.
2.58 Alternatively, for liquids of known density,
the volume may be determined by collection in
an appropriate size vessel of known mass
(empty), determination of the mass of the vessel
7
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Gas-monitoring equipment
Other instruments
Sound level meter
2.61 An integrating sound level meter is
required for the sound pressure test. It should
comply with Type 2 of BS EN 61672-1 and BS
EN 61672-2. Ten microphones are required for
the test on a single piece of equipment.
Airflow metering device
2.62 A metering device (such as a needle valve)
is required to admit air into the sterilizer
chamber for the air detector tests, and vacuum
and pressure leak tests. The device should be
capable of controlling the flow of air into an
evacuated chamber. It should be adjustable
and have a range that includes a flow of 05
mL/ min per litre volume of the sterilizer
chamber. The error in repeatability between
10% and 90% of the setting range should not
exceed 5%. The device is connected to the
chamber by a valved port provided by the
sterilizer manufacturer.
Balance
2.63 A laboratory balance is required for steam
dryness tests, load dryness tests, calibration of
flow meters (for measuring the flow of water
and/or chemical additives) and coolant quality
tests. It should be capable of measuring the
mass of loads up to 4 kg to an accuracy of 0.1
g and up to 400 g to an accuracy of 0.01 g.
2.64 An analytical balance is required for
determination of the TDS (evaporative residue)
in feed water. It should be capable of
measuring a mass of up to 100g with an
accuracy of 0.1 mg.
8
Fraction % by mass
<0.5
>20
<0.7
>50
<1.0
>75
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
post-tender analysis;
Pre-purchase considerations
Introduction
3.2 It is essential that the purchase of an item
of decontamination equipment is planned
correctly. This section aims to help the
purchaser with a step-by-step discussion of the
issues to be included. As this section is
designed to be universally applicable, it might
be necessary to vary the procedure according
to local circumstance or requirements.
Specialist advice
3.3 The efficient completion of procurement
documentation will require advice and help
from a decontamination specialist, for example
an AE(D).
3.4 This assistance should be sought in the
following areas:
10
advising manufacturer/contractor on
validation protocols;
monitoring validation performance;
auditing validation report;
Quality systems
3.5 Adherence to engineering standards and
quality systems ensures that decontamination
equipment is manufactured, installed, validated
and subject to the necessary periodic testing to
establish the initial and then on-going
satisfactory performance of the machine to
ensure optimum decontamination of surgical
instruments and safety of both operators and
patients.
Product listing
3.6 The purchaser should list all load items
proposed and projected to be re-processed by
the decontamination equipment. This should
include the following for each item:
number;
size;
temperature sensitivity;
moisture sensitivity;
materials of construction;
pressure-variation sensitivity;
Procurement of equipment
an overview
Introduction
3.12 This section gives a short overview of the
process of purchasing decontamination
equipment. It refers to more detailed information
in subsequent sections, including information
specific to each type of decontamination
equipment given in subsequent sections.
Purchasing decontamination equipment
3.13 The purchase of decontamination
equipment can be broken down into the
following sequence of steps.
What type of load will be processed?
11
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Choice of equipment
Introduction
Choice of equipment
3.33 The choice of machine should be
governed by the nature of the loads to be
reprocessed. Detailed guidance on appropriate
processes for different load items can be
obtained from the manufacturers of the medical
devices.
Sizes and numbers
3.34 Precise information on the sizes and
numbers of machines required for particular
applications is difficult to give since there are
considerable variations in patterns of use. The
number of machines required will depend on
the cycle time and the loading capacity of the
machine and in some circumstances on the
flexibility of operation that might be required, for
example whether items to be processed can
wait until there is a full load for the
decontamination equipment or need to be
processed immediately.
3.35 Consideration should be given to
contingency plans for machine usage, and
sufficient time should be included for testing,
maintenance and service. Thus reliance on a
single item of equipment is not advisable.
13
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Preparing a specification
3.37 Purchasers should seek assistance from
the AE(D) when preparing a specification for
decontamination equipment.
3.38 Standards and other specification
documents are continually being updated, and
purchasers should ensure that they consult the
latest editions of such documents, including
any amendments issued after publication, to
keep abreast of changing requirements. Advice
should be sought from the AE(D).
General design considerations
3.39 The following design considerations are
applicable to all or most types of
decontamination equipment, but are not
necessarily required by the current Standards.
Where applicable they should be included in
the specification for any decontamination
equipment to be operated in the healthcare
sector.
3.40 All decontamination equipment and
associated equipment is classed as work
equipment and should comply with the
Provision and Use of Work Equipment
Regulations 1998 (amended 2002 by the Health
and Safety (Miscellaneous Amendments)
Regulations). Purchasers are reminded that
under the Regulations it is the responsibility of
the employer, not the manufacturer, to provide
decontamination equipment that is suitable for
the purpose for which it is used or provided.
Further information is available in HTM 01-01
Part A.
3.41 All decontamination equipment made or
sold in the UK from 1 January 1996 must
conform to the emission and immunity
requirements of the current Electromagnetic
Compatibility Regulations 2005. This may be
achieved by compliance with BS EN 61000-6-3
(emission) and BS EN 61000-6-1 (immunity).
The manufacturer should be informed of any
local sources of electromagnetic disturbance
that could affect the operation of the machine.
14
Doors
3.58 The choice of design for any particular
installation will depend on the workload, space
restrictions, price and ease of maintenance.
With hinged doors there is a risk of the
Operator touching the hot inside face as the
door is opened. If hinged doors are required,
the specification should state whether they are
to be hinged on the left-hand or right-hand side
or top or bottom of the opening. Where sliding
doors are incorporated, the direction of opening
should be specified. The method of door
opening will impact on load handling equipment
design, the method of loading and unloading,
the height of the chamber above floor level, and
manual handling issues.
3.59 It should be possible to clean the contact
surfaces of the door seal without removing
parts of the machine.
Invitation to tender
3.60 Once detailed specifications have been
drawn up, manufacturers should be invited to
tender for the supply and, if required, the
installation of the decontamination equipment.
3.61 Prospective contractors should be given
the following information:
a. that each machine will be subject to a
validation process as described in the
validation and verification section (see
Chapter 4, Validation and verification);
b. that unless otherwise specified, the
installation checks and tests specified in
the validation process should be
satisfactorily completed before the
machine can be accepted;
15
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Delivery
16
Siting
3.67 A comprehensive review of the
requirements for SSDs is given in Health
Building Note 13 and for operating departments
in Health Building Note 26 Facilities for
surgical procedures.
3.68 The area in which decontamination
equipment is installed should meet the
requirements of the Workplace (Health, Safety
and Welfare) Regulations 1992 amended 2002,
which have far-reaching implications for the
design of decontamination equipment
accommodation and services to be installed.
3.69 Fire safety precautions should comply with
the current Approved Document B of the
Building Regulations and the Firecode series.
Engineering services
Introduction
3.70 Decontamination equipment installation
will require one or more external services
including steam, electricity, hot and cold water,
compressed air, ballast air, drainage, ventilation
and purified water. The manufacturer should
make clear at an early stage which services will
be needed and the detailed requirements for
each, as outlined in Table3.
Table 3 Information on services to be obtained from the manufacturer for sterilizers and washer-disinfectors
Steam
Electricity
Compressed air
Water
Drainage
Ventilation
a) the peak value during a cycle and the average value throughout a cycle of the heat in watts transmitted to the
environment when the decontamination equipment is operated in still air at an ambient temperature of 23C
2C
b) the heat in watts transmitted by a full load being unloaded from the machine into still air at an ambient
temperature of 23C 2C
c) the maximum flow of air extracted from the environment of the decontamination equipment as exhaust
ventilation
d) ventilation requirements for removal of fumes or gases from hazardous chemicals used in the process
Process chemicals
Details of all process chemicals required (e.g. detergents, rinse aids, sequestering agents, descalers,
microbicides, process gases) for the operation of the decontamination equipment, including any requirement for
regeneration of integral water treatment system, the quantity required per cycle, the nature and size of containers
in which they are supplied, the necessary storage conditions and instructions for safe handling.
17
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Electricity
Compressed air
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
General
4.1 This chapter of the guidance covers the
validation and periodic testing of the various
types of decontamination equipment used in
hospitals, laboratories and other healthcare
facilities.
Permit-to-work
4.2 The use of permit-to-work system should
be used for all maintenance and testing
procedures on decontamination equipment.
This should ensure the formal removal of
equipment from, and return to, service and will
provide certification of acceptance by the User.
A suggested permit-to-work form is shown on
the following pages.
Testing of decontamination
equipment
Introduction
4.3 Good decontamination practice is based
on four key aspects that ensure that the
required standards of performance and safety
are met and sustained:
a. all decontamination equipment is
subjected to a programme of validation;
b. all decontamination equipment is
subjected to a planned programme of
periodic tests performance;
c. all decontamination equipment is
operated by trained staff in its use in
accordance with a written procedure
20
Responsibilities
General
4.7 Decontamination equipment should be
subjected to a planned programme of testing
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Commissioning
Performance
qualification
Installation tests
Thermometric tests
Operational tests
Microbiological
tests
Residual gas/
additive tests
Works tests
4.24 Works tests before delivery of the
decontamination equipment are intended to
verify that the equipment performs in
conformity with the results obtained from type
testing in respect of various critical attributes.
(See BS EN 15883 Parts 1 and 2 for works test
details for washer-disinfectors and BS EN 285
for works test details for sterilizers.)
4.25 For one-off designs, a more extensive
programme of works tests, similar to the
programme of type tests for machines in serial
production, is required, and the purchaser
might wish to arrange for their representative
(either the AE(D) or CP(D)) to attend the factory
to witness these tests before accepting delivery
of the decontamination equipment.
Validation
4.26 Validation is the documented procedure
required for obtaining, recording and
interpreting the results needed to show that a
process will consistently yield a product
complying with a pre-determined specification.
Validation is a total process beginning with a
review of the specification against which the
equipment is purchased. This is to ensure that
24
Cleaning efficacy
tests (to extend to
protein removal
tests)
Load dryness tests
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Loading
4.43 The User should ensure that each load is
presented to the process in accordance with
documented procedures established and
tested during PQ.
4.44 Baskets or load carriers should not be
overloaded, as this will result in inefficient
cleaning and disinfection.
26
Documentation
4.48 Accurate and efficient record-keeping is
an essential part of the management of
decontamination equipment. The extent and
nature of the records that are necessary varies
with the type of machine and the use to which
it is put. Guidance is given in HTM 01-01 Parts
C and D.
Summary sheets
4.49 On completion of the validation process,
and before leaving the premises, the CP(D)
should prepare a summary report containing
the results of the commissioning and PQ tests
and essential working data.
4.50 At the request of the User the CP(D)
should also supply graphical representations of
cycle variables obtained from the thermometric
tests.
4.51 The summary report should be signed by
the CP(D) and countersigned by the AP(D) to
certify that the machine is fit for use.
4.52 Summary reports should be securely
retained by the User and be available for ready
reference.
4.53 At the same time the CP(D) should provide
the User with copies of any master process
records required for routine production.
Validation report
4.54 Within one month of the completion of the
validation process the CP(D) should prepare a
full validation report which should include:
a. all the data supplied by the contractor,
collected during the installation checks
and tests with written confirmation that
they meet the manufacturers
specification;
b. written confirmation that the calibration of
all measuring instruments fitted to the
machine have been verified;
c. all the data collected during the
commissioning tests, with written
confirmation from the CP(D) that they
meet the specified requirements;
d. data showing the correlation between the
performance of the measuring
instruments fitted to the machine and the
test instruments used during
commissioning and PQ;
e. reports containing all the data collected
during the PQ tests, with written
confirmation from the CP(D) and the User
of the loading conditions and types of
load (including, when necessary,
reference to specific individual items) that
may be satisfactorily processed in the
machine;
f. data from the instruments fitted to the
machine, independent monitoring system
data and validation instrument data,
along with comments on any changes or
adjustments made.
4.55 When data is in the form of electronic data
files, the report should include copies of disks/
DVD or CD or tapes containing the data in a
format compatible with local systems and
policies and a printout of the directory of each,
annotated to show where the data for each test
is to be found.
4.56 The CP(D) should certify that all necessary
tests have been carried out and that the results
were satisfactory.
Periodic tests
4.60 After validation and when the machine is
passed into service, it should be subject to a
schedule of periodic tests at daily, weekly,
quarterly and yearly intervals, to provide
evidence that the machine continues to operate
within the limits established during
commissioning.
4.61 The User and the CP(D) (under the
management of the AP(D)) are responsible for
the periodic tests.
4.62 The yearly test schedule is a revalidation
procedure and provides a more comprehensive
test programme than the other periodic tests; it
serves to demonstrate that data collected
during commissioning and the PQ remain valid.
Revalidation
4.63 In addition to annual revalidation,
revalidation is required under the following
circumstances:
a. when the machine is to be returned to
service after repair or component
replacement of part of the systems that
affect satisfactory attainment of the preset variables of the operating cycle;
b. when the pre-set values of the cycle
variables have been modified;
27
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Repeat validation
4.66 Revalidation and periodic tests are
designed to establish the continued
conformance of the equipment and its
performance with data established during the
original validation study.
4.67 There are occasions when it might be
necessary to repeat the full set of tests carried
out during the initial validation in order to obtain
a new set of data.
4.68 Repeat validation should be carried out if:
a. the machine is modified to such an extent
that it must be presumed that the original
data is no longer valid;
b. a machine has been moved and installed
at a new site;
c. the machine has been dismantled or
extensively overhauled;
d. a new operating cycle has been
introduced;
28
Types of test
4.70 The tests listed in the schedules fall into
the following categories:
a. Automatic control tests, which are
designed to verify the correct functioning
of the operating cycle from the readings
obtained from the instruments fitted to
the machine;
b. Thermometric tests, which are designed
to provide assurance that the
temperature requirements for the process
are met by using accurate measuring
equipment, independent of the
instruments fitted to the machine to
monitor the temperatures attained within
the chamber and reference loads;
c. Thermometric tests for a small load,
which are designed for two purposes. In
sterilizers with an active air removal
system they demonstrate that the
sterilizer is capable of removing air from a
small load in which air from a near-empty
chamber has been retained.
Thermometric tests for a full load are
designed to show that services provided
to the machine are adequate for purpose.
In certain circumstances they may also
serve as PQ tests for loading conditions
that present a lesser challenge to the
29
29
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
Principles of performance
qualification tests
Introduction
4.84 Performance qualification (PQ) is defined
as the process of obtaining and documenting
evidence that the equipment, as installed and
operated in accordance with operational
procedures, consistently performs in
accordance with predetermined criteria and
thereby yields product meeting its specification.
4.85 PQ tests should be performed as part of
the initial validation procedure, as part of any
repeat validation procedure and whenever the
User, acting on the advice of the AE(D), judges
that new loading or operating conditions require
a new PQ test.
4.86 Circumstances that might lead to new PQ
tests would include changes to the quality of
the water supply, changes to the chemical
additives used in the cleaning and disinfection
process, changes to the loading system or the
requirement to process a new type of product
or packing arrangements for decontamination
equipment.
4.87 PQ should not be undertaken on any piece
of equipment until the requirements of the
installation and operational tests specified in
HTM 01-01 Parts C and D have been met.
4.88 Soil removal efficacy tests should be
carried out on all washer-disinfectors as part of
the PQ (see HTM 01-01 Part D).
4.89 Thermometric PQ may not be required for
all loads. Consideration should be given to tests
for effective sterilization and drying of metal
loads, and of any particular loads agreed
between the User and the AE(D).
4.90 PQ tests should be carried out by the
CP(D).
30
31
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
32
Product release
33
HTM 01-01: Management and decontamination of surgical instruments: Part B Common elements
References
BS EN ISO 13485.
BS 6447.
BS EN 60751.
BS EN 60584-2.
BS EN 837-1.
ISO 4064-1.
BS 5898, ISO 384.
BS EN 61672-1.
BS EN 61672-2.
BS EN ISO 14644-1.
Health Building Note 13 Sterile services
department.
Provision and Use of Work Equipment
Regulations 1998.
ISO/IEC 17025.
BS EN 61000-6-3.
BS EN 285.
BS EN 61000-6-1.
BS EN ISO 15883.
PD 5304.
BS EN 15883-1.
BS EN 61010-1.
BS EN 15883-2.
BS EN 61010-2-040.
34
References
35