Manual Servicio Emerald
Manual Servicio Emerald
Throughout the manual, signal words and icons appear where the nature of the
information warrants special attention.
NOTE: The note signal word appears adjacent to an important point of
information that is relevant to the current subject matter.
Operation, maintenance, and servicing of hematology systems may expose
individuals to potential safety and health hazards. All work must be performed as
described in the Operators Manual or as directed by an Abbott representative.
Information relating to potential hazards may be found in Section 8: Hazards.
WARNING: This equipment needs special precautions regarding general
requirements for safety. Please consult Section 7: Operational Precautions
and Limitations and Section 8: Hazards, before operating the
CELL-DYN Emerald.
9140840CDecember 2009
NOTES
ii
9140840CDecember 2009
Revision Status
Document Control
Number(s)
Revision Date
Section(s) Revised
New Release
9213300A
09H40-01
May/2008
9213300B
May/2008
9140859BRevision
Status
p. iii
9140846BInstallation
Procedures and Special
Requirements
9213300C
9140859DDecember 2009
February/2009
9140847BPrinciples of
Operation
9140848B
Performance
Characteristics and
Specifications
9140849B
Operating Instructions
9140853B
Service and
Maintenance
p. 9-3
9140859CRevision
Status
pp. iii, iv
9140840BForeword
9140841BTable of
Contents
ALL
9140842BList of
Figures
ALL
9140843BList of
Tables
ALL
9140845BUse or
Function
9140846CInstallation
Procedures and Special
Requirements
3-12, 3-16
iii
Document Control
Number(s)
9213300D
iv
Revision Date
December/2009
Section(s) Revised
9140847CPrinciples
of Operation
9140848C
Performance
Characteristics and
Specifications
9140849COperating
Instructions
9140850BCalibration
9140851BOperational
Precautions and
Limitations
9140852BHazards
9140853CService and
Maintenance
9140854B
Troubleshooting
9140855BQuality
Control
9140861BAppendix A
9140861BAppendix B
ALL
9140861BAppendix C
p. C-1
9140861BAppendix E
9140861BAppendix F
9140863BAppendix G
ALL
9140872BGlossary
pp. 8-11
9140862BIndex
ALL
9140859DRevision
Status
ALL
9140840CForeword
ALL
9140841CTable of
Contents
ALL
9140842CList of
Figures
ALL
9140843CList of
Tables
ALL
9140859DDecember 2009
Document Control
Number(s)
9140859DDecember 2009
Revision Date
Section(s) Revised
9140844BHow to Use
This Manual
ALL
9140845CUse or
Function
ALL
9140846DInstallation
Procedures and Special
Requirements
ALL
9140847DPrinciples
of Operation
ALL
9140848D
Performance
Characteristics and
Specifications
ALL
9140849DOperating
Instructions
ALL
9140850CCalibration
ALL
9140851COperational
Precautions and
Limitations
ALL
9140852CHazards
ALL
9140853DService and
Maintenance
ALL
9140854C
Troubleshooting
ALL
9140855CQuality
Control
ALL
9140861CAppendix A
ALL
9140861CAppendix B
ALL
9140861CAppendix C
ALL
9140861CAppendix E
ALL
9140861CAppendix F
ALL
9140863CAppendix G
ALL
9140872CGlossary
ALL
9140862CIndex
ALL
NOTES
vi
9140859DDecember 2009
Revision Log
Instructions: Use this log to provide a permanent record to verify that revised chapter(s) and/or page(s) have
been added to this manual.
1. Record the document control number of the revised section in the first column. You will find the number
in the footer. Make an entry for each chapter you receive and place in the manual.
2. Record the revision date, also found in the footer, in the second column.
3. Record the current CELL-DYN Emerald software version in the third column.
4. Write your initials or signature in the fourth column to verify that you have placed the revised page(s) in
the manual.
5. Record the date that you added the revised section to the manual in the fifth column.
Document
Control Number
Revision Date
9140859DDecember 2009
Software Version
Revision
Incorporated by
Date
Incorporated
vii
NOTES
viii
9140859DDecember 2009
Foreword
Congratulations on your purchase of the CELL-DYN Emerald System. The
CELL-DYN Emerald, which uses state-of-the-art technology, is designed to
function consistently and dependably on a daily basis.
The CELL-DYN Emerald is backed by dedicated professionals who excel in
engineering, training, and technical expertise. As a valued Abbott customer, we
will teach you how to operate, maintain, and troubleshoot your system.
Abbott Hematology is dedicated to providing the highest quality, most reliable
instrumentation available. We look forward to working with you and serving your
needs.
Customer Service
If you need information or help in diagnosing a problem, technical assistance is
available by telephone. In the USA, this service is available 24 hours a day, seven
days a week by calling Abbott Diagnostics Customer Service.
United States: 1-877-4ABBOTT (1-877-422-2688).
For customer support in Canada 1-800-387-8378.
Outside of USA and Canada: contact your Country Service and Support
representative.
For correspondence, the address in the USA is:
Abbott Diagnostics Division
Customer Service
200 Abbott Park Road
Abbott Park, IL 60064, USA
Intended Use
The CELL-DYN Emerald System is an automated hematology analyzer designed
for in-vitro diagnostic use in clinical laboratories.
9140840CDecember 2009
ix
Proprietary Statement
The entire contents of this manual are copyrighted 2008, and 2009 by Abbott
Laboratories. All rights are reserved.
The software and manual were developed solely for use with the
CELL-DYN Emerald and for in vitro diagnostic applications as specified in the
operating instructions.
The information and related graphics published herein (the Information) are the
sole property of Abbott Laboratories. Permission to use the Information is granted,
provided that:
Patent Statement
The following worldwide patents are relevant to the CELL-DYN Emerald or its
components. WO 200562015, WO 200510497. The following USA Patents are
relevant to the CELL-DYN Emerald or its components: 6,632,676. There are other
such patents and patent applications in the United States and worldwide.
Disclaimers
All samples (printouts, graphics, displays, screens, etc.) are for information and
illustration purposes only and shall not be used for clinical or maintenance
evaluations. Data shown in sample printouts and screens do not reflect actual
patient names or test results. Labels depicted in the manual may appear different
from actual product labels.
Abbott Laboratories makes no representations or warranties about the accuracy and
reliability of the information contained in the CELL-DYN Emerald Operators
Manual.
The information was developed to be used by Abbott Laboratories trained
personnel, by other persons knowledgeable or experienced with the operation and
service of the product identified, or under the direct supervision and with
cooperation from Abbott Laboratories technical sales or service representatives.
9140840CDecember 2009
In no event shall Abbott Laboratories or its affiliates be liable for any damages or
losses incurred in connection with or arising from the use of the information on this
media by persons not fully trained by Abbott Laboratories. This limitation shall not
apply to those persons knowledgeable or experienced with the operation and
service of the product identified, or under the direct supervision and with
cooperation from Abbott Laboratories technical sales or service representatives.
No confidential relationship shall be established in the event that any user of the
Information should make any oral, written or electronic response to Abbott
Laboratories (such as feedback, questions, comments, suggestions, ideas, etc.).
Such response and any information submitted therewith shall be considered nonconfidential, and Abbott shall be free to reproduce, publish, or otherwise use such
information for any purposes whatsoever including, without limitation, the
research, development, manufacture, service, use, or sale of products incorporating
such information. The sender of any information to Abbott is fully responsible for
its content, including its truthfulness and accuracy and its non-infringement of any
other persons proprietary rights.
Abbott Laboratories is not engaged in rendering medical advice or services.
Updates to the information may be provided in either paper or electronic format.
Always refer to the latest documents for the most current information.
List numbers are unique identifiers that are used when ordering products. The list
number and quantity provided in Appendix A: Accessories are intended for
guidance only and are subject to change. Contact your Abbott representative for the
most current information regarding list numbers.
All operating instructions must be followed. In no event shall Abbott be
responsible for failures, errors, or other liabilities resulting from customers noncompliance with the procedures and precautions outlined herein.
9140840CDecember 2009
xi
3. Any on-site service performed at other times and all service required to
correct defects or malfunctions not covered by this Warranty (as noted in the
paragraph below) will be billed at Abbotts labor rates then in effect.
This Warranty does not cover defects or malfunctions which:
1. Are not reported to Abbott during the Warranty Period and within one week
of occurrence.
2. Result from chemical decomposition or corrosion.
3. Are caused by customer or third party abuse, misuse, or negligence, or by
failure to comply with any requirement or instruction contained in the
applicable Abbott Operators Manual.
4. Result from maintenance, repair, or modification performed without Abbotts
authorization.
Abbotts liability for all matters arising from the supply, installation, use, repair, and
maintenance of the Instrument, whether arising under this Warranty or otherwise,
shall be limited solely to the repair or (at Abbotts sole discretion) replacement of
the Instrument or of components thereof. In no event shall Abbott be liable for
injuries sustained by third parties, incidental or consequential damages, or lost
profits. Replaced parts shall become the property of Abbott Laboratories.
THE FOREGOING IS THE SOLE WARRANTY MADE BY ABBOTT
LABORATORIES REGARDING THE INSTRUMENT, AND ABBOTT
SPECIFICALLY DISCLAIMS ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND OF FITNESS FOR A PARTICULAR PURPOSE.
The CELL-DYN Emerald Hematology System is manufactured in France for
Abbott Diagnostics Division, Abbott Laboratories, 200 Abbott Park Road, Abbott
Park, IL, 60064, USA. Please direct all inquiries concerning information in this
manual to the foregoing address.
NOTE: Direct all inquiries regarding equipment problems to Abbott Diagnostics
Customer Service. (USA customers only.)
xii
9140840CDecember 2009
98/79/EC
Legal Manufacturer
Abbott Laboratories
Abbott Park, IL 60064, USA
Authorized Representative
ABBOTT
Max-Planck-Ring 2
65205 Wiesbaden
Germany
+49 6122 580
UL
61010-1
Approved
CAN/CSA-C22.2
No. 61010-1
Approved
IEC 61010-1
Approved
UL
Listed
Trademark Statements
CELL-DYN and Emerald are trademarks of Abbott Laboratories in various
jurisdictions.
All other trademarks are the property of their respective owners.
9140840CDecember 2009
xiii
Symbols
The symbols listed below are used in labeling, including the instrument, reagents,
calibrators, controls, and this manual. Please note that Warning and Caution
symbols and statements are in this manual in Section 8: Hazards.
Table 1
Instrument/Power-related
Symbol
-5A
DILUENT
REF
REV
SN
WASTE
APPLICATION SOFTWARE
xiv
Definition/Use
Location of Label
Rear panel
Barcode/Barcode scanner
connection
Rear panel
Diluent
Rear panel
Equipotentiality
Rear panel
Ethernet/Ethernet connection
Rear panel
Catalog Number
Stand by /
Power On/Off
Front
Rear panel
Revision
Serial Number
Rear panel
Rear panel
Waste
Rear panel
Application Software
On Flash Drive
9140840CDecember 2009
Table 2
Reagent-related
Symbol
Definition/Use
CLEANER
Cleaner Reagent
DILUENT
Diluent Reagent
KEY
Key
LOT
Batch Code
LYSE
Lyse
SN
Serial Number
Use By
Table 3
Calibrator/Control-related
Symbol
Definition/Use
ASSAY VALUE
Assay Value
MEAN RANGE
Mean Range
MEAN VALUE
Mean Value
PARAMETER
Parameter
SYSTEM
TOLERANCE LIMIT
WB CAL
WB CONTROL TRI-LEVEL
WB CONTROL L/N/H or I//II/III
9140840CDecember 2009
System
Tolerance Limit
Whole Blood Calibrator
Whole Blood Control, Tri-Level
Whole Blood Control, Low, Normal or High Level; or
Whole Blood Control Level I, II or III.
xv
Table 4
Miscellaneous
Symbol
EC REP
ABBOTT
Max-Planck-Ring 2
65205 Wiesbaden
Germany
+49-6122-580
Definition/Use
Authorized Representative in the European
Community
Biohazard
IVD
i
MODEL
Model number
Temperature limitation
xvi
9140840CDecember 2009
Labeling
The following labels are affixed to the CELL-DYN Emerald System:
The CELL-DYN Emerald is CE Marked to the European In Vitro Diagnostic
Directive, which encompasses the requirements of the EMC and Safety Directives,
and have the following labels:
Shipping Container
Figure 1:
Figure 2:
Biohazard Label
Figure 3:
9140840CDecember 2009
Hazard Label
xvii
xviii
Figure 4:
Figure 5:
Figure 6:
Biohazard Label
9140840CDecember 2009
Figure 7:
Rear Panel
9140840CDecember 2009
xix
NOTES
xx
9140840CDecember 2009
1-1
1-3
1-3
1-3
1-3
1-3
1-3
9140841CDecember 2009
1-1
1-2
1-2
1-2
1-2
1-3
1-3
1-4
1-5
1-8
Table of Contents-1
AC Adapter/Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Reagent Tray. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Reagent System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
CELL-DYN Emerald Diluent Reagent. . . . . . . . . . . . . . . . . . . 1-10
CELL-DYN Emerald CN-Free Lyse Reagent . . . . . . . . . . . . . 1-11
CELL-DYN Emerald Cleaner Reagent . . . . . . . . . . . . . . . . . . 1-11
Reagent Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Reagent Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Calibrator and Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Installation Procedures and Special Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Site Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Place the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Waste Disposal Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Installation Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Inventory and Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Instrument Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
AC Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Printer installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Reagent installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Instrument Start Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Instrument Set Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Operator Log In and Log Out . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
System Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Loading or Replacing Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Replacing the Reagents Diluent, Lyse, Cleaner . . . . . . . . . . . . . . . . . . . 2-15
Using Reagent Bar Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Manual Entry of Reagent Information . . . . . . . . . . . . . . . . . . . . . . 2-16
Replacing or Emptying the Waste Container . . . . . . . . . . . . . . 2-17
Other Reagent Menu functions . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Set Up Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Advanced Set Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
Printer Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Interpretive Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Communication Button. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
Reporting Options Button. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Lab Preferences Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Table of Contents-2
9140841CDecember 2009
Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SPECIMEN CAL. ADJUSTMENTS . . . . . . . . . . . . . . . . . . . .
Thresholds. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Rename Specimen Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration Factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Other Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Version Release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup Configuration Printout . . . . . . . . . . . . . . . . . . . . . . . . . .
Instrument Quality Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Background Count . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Carryover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Quality Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-26
2-27
2-27
2-27
2-27
2-28
2-29
2-30
2-32
2-32
2-32
2-33
2-34
2-34
2-34
2-34
9140841CDecember 2009
Table of Contents-3
Table of Contents-4
9140841CDecember 2009
5-18
5-19
5-20
5-20
5-20
5-20
5-22
5-22
5-23
5-23
5-26
5-26
5-27
5-28
5-29
5-30
5-31
5-32
5-32
5-32
5-33
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
When to Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Calibration Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Calibration Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Pre-Calibration Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Automated Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Entering Calibrator Information . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Calibration Verification Procedure . . . . . . . . . . . . . . . . . . . . . . 6-10
References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Operational Precautions and Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Precautions and Requirements for System Operation . . . . . . . . . . .
Precautions Before Operation . . . . . . . . . . . . . . . . . . . . . . . . . . .
Requirements Before Operation . . . . . . . . . . . . . . . . . . . . . . . . .
Precautions During Operation . . . . . . . . . . . . . . . . . . . . . . . . . .
CELL-DYN Emerald Operators Manual
9140841CDecember 2009
7-1
7-3
7-4
7-4
7-4
7-4
Table of Contents-5
7-5
7-5
7-5
7-5
7-6
7-6
7-6
7-7
7-9
Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Operator Responsibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Safety Icons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Biological and Chemical Hazards. . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Biological Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Chemical Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
Spill Clean-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
Waste Handling and Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
Decontamination Procedure Requirements . . . . . . . . . . . . . . . . 8-8
Electrical Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
Mechanical Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Physical Hazards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Aspiration Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Heavy Objects. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Tripping Hazard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
Service and Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Saving and Printing the Event Log . . . . . . . . . . . . . . . . . . . . . . . 9-4
Maintenance Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Special Modes Button. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Fluidics Menu Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
Preventive Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
Monthly Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
Bleach Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
Materials Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
Table of Contents-6
9140841CDecember 2009
Semi-Annual Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Lubricating the Pistons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Materials Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
As Needed Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Emergency Stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Initialize System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning the cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Material Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preparing for Storage, Relocation or Shipping. . . . . . . . . . . . .
Materials Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Decontamination Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . .
Clean . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Backflush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sample Probe Removal and Replacement . . . . . . . . . . . . . . . . . . .
Materials Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sample Probe Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sample Probe Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . .
References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9-13
9-13
9-13
9-13
9-14
9-14
9-14
9-15
9-15
9-15
9-15
9-16
9-17
9-17
9-18
9-18
9-18
9-22
9-23
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Troubleshooting Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Troubleshooting Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-15
Quality Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
When to Run QC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
Quality Control Methods and Materials . . . . . . . . . . . . . . . . . . . . . 11-3
QC File Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Entering QC Assay Values and Limits . . . . . . . . . . . . . . . . . . . 11-7
QC Guidelines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-9
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-9
Control Material Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-9
Normal Specimen Guideline: The Rule of Three . . . . . . . . . . . 11-9
Running Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-10
Use of Tools in the QC Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
Printing QC Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
Sending QC Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
Deleting QC Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
CELL-DYN Emerald Operators Manual
9140841CDecember 2009
Table of Contents-7
Saving QC Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Viewing Results in the QC Menu . . . . . . . . . . . . . . . . . . . . . .
Viewing all QC Runs Within a File . . . . . . . . . . . . . . . . . . . .
Viewing all Measurands Within the QC File . . . . . . . . . . . . .
Viewing QC Statistics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Levey-Jennings Graphs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
References. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11-14
11-14
11-14
11-14
11-14
11-16
11-17
Appendix A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Appendix B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Potential Causes of Spurious Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Appendix C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
Preparation of diluted sodium hypochlorite solutions . . . . . . . . . . . . . . . . . C-1
Appendix D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
CBC Reference Intervals: Literature Sources. . . . . . . . . . . . . . . . . . . . . . . . D-1
Appendix E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
Sample Logs and Worksheets. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
Worksheet E.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2
Pre-Calibration Procedure Checklist . . . . . . . . . . . . . . . . . . . . . E-2
Worksheet E.2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-8
Calibration Verification Worksheet . . . . . . . . . . . . . . . . . . . . . . E-8
Worksheet E.3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-9
Training Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-9
Worksheet E.4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-11
Manual Calibration Factor Entry Worksheet . . . . . . . . . . . . . . . . . E-11
Appendix F. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
Manual Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Requirements for Whole Blood Specimens . . . . . . . . . . . . . . . .
Requirements for Whole Blood Calibration . . . . . . . . . . . . . . . .
Reference Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
WBC, RBC and PLT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HGB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MCV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table of Contents-8
F-1
F-1
F-2
F-2
F-2
F-2
F-2
9140841CDecember 2009
9140841CDecember 2009
Table of Contents-9
NOTES
Table of Contents-10
9140841CDecember 2009
List of Figures
List of Figures
Figure 1:
Figure 2:
Figure 3:
Figure 4:
Figure 5:
Figure 6:
Figure 7:
Use or Function
Figure 1.1
Figure 1.2
Figure 1.3
Figure 1.4
Figure 1.5
Figure 1.6
Figure 1.7
Figure 1.8
Figure 1.9
9140842CDecember 2009
List of Figures
Figure 2.19
Figure 2.20
Figure 2.21
Figure 2.22
2-28
2-28
2-29
2-31
Principles of Operation
Figure 3.1
Figure 3.2
Figure 3.3
Figure 3.4
Figure 3.5
Figure 3.6
Operating Instructions
Figure 5.1
Figure 5.2
Figure 5.3
Figure 5.4
Figure 5.5
Figure 5.6
Figure 5.7
Figure 5.8
Figure 5.9
Figure 5.10
Figure 5.11
Figure 5.12
Figure 5.13
Figure 5.14
Figure 5.15
Figure 5.16
Figure 5.17
Figure 5.18
Figure 5.19
Figure 5.20
Figure 5.21
Figure 5.22
Figure 5.23
Figure 5.24
Figure 5.25
Figure 5.26
Figure 5.27
List of Figures-2
9140842CDecember 2009
List of Figures
Figure 5.28
Figure 5.29
Figure 6.1
Figure 6.2
Figure 6.3
Figure 8.1
Figure 8.2
Figure 8.3
Calibration
Hazards
Quality Control
Figure 11.1
Figure 11.2
Figure 11.3
Figure 11.4
Figure 11.5
Figure 11.6
Figure 11.7
Figure 11.8
QC Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Lab ID Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Restoring a QC file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
Targets and Limits Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-8
QC Results Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-11
Manually Deleting QC Results . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-13
QC Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-15
Levey-Jennings Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-16
9140842CDecember 2009
List of Figures
NOTES
List of Figures-4
9140842CDecember 2009
List of Tables
List of Tables
Foreward
Table 1
Table 2
Table 3
Table 4
Instrument/Power-related . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiv
Reagent-related . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Calibrator/Control-related. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xvi
Principles of Operation
Table 3.1
Table 3.2
Table 3.3
Table 3.4
Table 3.5
Table 3.6
Table 3.7
9140843CDecember 2009
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Dimensions in Shipping Carton . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Dimensions for the AC Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Instrument Power Source Requirements. . . . . . . . . . . . . . . . . . . . . . 4-4
Reagent Consumption (in mL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Background Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Carryover Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Display Range and Analytical Measurement Range . . . . . . . . . . . . 4-9
Fresh Blood Imprecision Specifications . . . . . . . . . . . . . . . . . . . . . 4-10
Long-Term Commercial Control Imprecision: CBC . . . . . . . . . . . 4-11
Long-Term Commercial Control Imprecision: WBC Differential . 4-13
Comparability (Correlation) to CELL-DYN 1800 Internal Site. . . 4-14
Comparability (Correlation) to CELL-DYN 1800 External Site . . 4-15
Comparability (Correlation) of WBC Differential to Microscopy . 4-15
Reference Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
List of Tables-1
List of Tables
Hazards
Table 8.1
Troubleshooting
Table 10.1
Table 10.2
Table 10.3
Table A.1
Table A.2
Table A.3
Table A.4
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Consumables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
Calibrator and Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
Accessories
List of Tables-2
9140843CDecember 2009
9140844BDecember 2009
Appendices
Appendix A
This appendix lists the part numbers of components, accessories, controls, reagents
and consumables associated with the CELL-DYN Emerald System for user
convenience when placing orders.
Appendix B
This appendix contains a table on potential causes of erroneous results.
Appendix C
This appendix contains step-by-step instructions for preparing sodium
hypochlorite (bleach) solutions for use on the CELL-DYN Emerald.
Appendix D
This appendix contains a list of literature sources relating to CBC Reference
Intervals (Patient Limits).
Appendix E
This appendix contains sample logs and worksheets to copy and use for your
convenience in constructing an instrument log for your laboratory.
Appendix F
This appendix contains information about whole blood calibration and manual
calculation of calibration factors.
Index
This section contains an alphabetical listing of subject matter to help users quickly
locate specific information about the system.
How to use this Manual - 2
9140844BDecember 2009
Text Conventions
Text Conventions Used in This Manual
In this manual, procedural instructions are explained in logical groups, using
numbered steps. Illustrations and drawings appear where they are useful to the
explanation. Text conventions are as follows:
Menu Name
The menu name is printed in bold, uppercase, sans serif letters; for example, MAIN
menu.
Touch Screen Buttons
The CELL-DYN Emerald user interface includes a LCD touch screen where
buttons are selected by touching the corresponding area of the display. When
referring to these, the manual uses the term button. Touching one of these buttons
initiates an action. Touching other (non-button) areas on the display make
selections or inserts a cursor for entry. Detailed descriptions of the actions initiated
by the buttons are found in the applicable section(s) of this manual. Screen labels
are shown in bold, uppercase, sans serif letters enclosed in brackets; for example,
[QUALITY CONTROL].
Data Entry Field Names
Fields that accept data entered by the Operator have their names shown in regular,
mixed-case font enclosed within carats < >.
Keypad (Keys)
In addition to the display touch screen, the user interface for the
CELL-DYN Emerald includes keys on the front panel of the instrument. When
referring to these, the manual uses the term key. Special function keys, such as
the arrow keys, may appear as a symbol substituted for the word. Instructions for
special function keys will read, for example Press the [] arrow key.
Screen Messages
Screen messages or other screen displays will appear in bold, Courier letters, for
example, Do you wish to continue?. If the screen message requires a
response from the user, the touch screen button(s) will follow the convention as
described in Touch Screen Buttons.
9140844BDecember 2009
NOTES
9140844BDecember 2009
Section 1
Use or Function
Overview
The CELL-DYN Emerald is an automated hematology analyzer intended for in
vitro diagnostic use in the clinical laboratory. It is menu-driven and controlled by
a microprocessor.
Figure 1.1
CELL-DYN Emerald
The CELL-DYN Emerald aspirates blood from an opened collection tube held up
to the aspiration probe. The CELL-DYN Emerald can aspirate blood from several
types of collection devices which contain K2EDTA. Refer to Section 5: Operating
Instructions, Subsection: Specimen Collection and Handling.
9140845CDecember 2009
1-1
Use or Function
Section 1
Overview
Lymphocyte percent
LYM #
MID %
MID #
GRA %
Granulocyte percent
GRA #
Hematocrit
MCV
RDW
Hemoglobin measurands:
HGB
Hemoglobin concentration
MCH
MCHC
Platelet measurands:
PLT
Platelet count
MPV
1-2
9140845CDecember 2009
Use or Function
Section 1
Overview
System Components
The CELL-DYN Emerald consists of eight main components:
1.
2.
3.
4.
5.
6.
7.
8.
2
3
5
7
4
8
9140845CDecember 2009
1-3
Use or Function
Section 1
Overview
Main Menu
Menu items are discussed in detail in subsequent Sections.
1. Date/Time Displays current date
and time/
2. TOOLS ICON (Wrench) Accesses
the TOOLS menu containing
context-specific options, such as
Print and Send.
3. RETURN ICON (Curved Arrow)
Returns to the previous menu.
4. HOME ICON (House) - Returns to
the MAIN menu.
5. START UP Initiates the Start Up
Cycle.
6. EVENT LOG Accesses the Event
Log.
7. OPER. LOG IN/OUT Accesses
the Log In/Log Out screen.
8. REAGENTS Accesses the
REAGENTS menu.
9. SHUT DOWN Initiates the Shut
Down Cycle.
10.CALIBRATION Accesses the
CALIBRATION menu.
11. QUALITY CONTROL Accesses
the QUALITY CONTROL menu.
12.DATALOG Accesses stored
results.
13.SET UP Accesses the SET UP
menu.
14.MAINTENANCE Accesses the
MAINTENANCE and SERVICE
menu.
15.RUN SAMPLE Accesses the Run
Screen.
Figure 1.3
1-4
1
2
3
4
10
11
12
13
14
15
Main Menu
9140845CDecember 2009
Use or Function
Section 1
Overview
Fluidics
1. Sampling Module
2. Syringe Module
3. Counting Module
Figure 1.4
Fluidics
The fluidic components are located on the right side of the instrument and contain
the following three modules:
9140845CDecember 2009
1-5
Use or Function
Section 1
Overview
Figure 1.5
Sampling Module
1. Sampling Module consists of the rocker that performs the up/down and
forward/back movement of the aspiration probe and the aspiration probe
itself.
1-6
9140845CDecember 2009
Use or Function
Section 1
Overview
1
Figure 1.6
3 4
Syringe Module
9140845CDecember 2009
1-7
Use or Function
Section 1
Overview
WBC/HGB
RBC/PLT
Counting Chamber Counting Chamber
Figure 1.7
Counting Module
3. Counting Module This module counts the WBCs, RBCs, and PLTs and
measures the HGB. It includes:
WBC/HGB Counting Chamber with the WBC Aperture and the Hemoglobin
LED. Left bath.
RBC/PLT Counting Chamber with the RBC/PLT Aperture. Right bath.
Liquid valve manifold assembly with associated tubing. Below baths.
Main PCB Board
The Main PCB Board is located between the Fluidics and the Reagent Tray. It is
driven by a 32-bit microprocessor and manages the following components:
Aspiration probe, Rocker, and Syringe Block motors
Display and Keyboard
LIS Connection (RS232, Ethernet, etc.)
Cell counting and Hemoglobin measurement
Data Processing
Bar code reader
USB connections
1-8
9140845CDecember 2009
Use or Function
Section 1
Overview
Figure 1.8
AC Adapter/Transformer
9140845CDecember 2009
1-9
Use or Function
Section 1
Overview
Reagent Tray
The Reagent Tray holds the Cleaner and Lyse reagent bottles.
1. Cleaner Reagent
2. Cleaner Reagent line (green
collar)
3. Lyse Reagent
4. Lyse Reagent line (blue collar)
Figure 1.9
4
3
Reagent Tray
Reagent System
A reagent system, specifically formulated for the CELL-DYN Emerald instrument,
provides optimal system performance. Use of reagents other than those specified
in this manual is not recommended as instrument performance can be affected.
Each instrument is tested at the factory using the specified reagents. All
performance claims are generated using these reagents. Refer to Appendix A:
Accessories for information on ordering CELL-DYN Reagents.
CELL-DYN Emerald Diluent Reagent
CELL-DYN Emerald Diluent Reagent is formulated to do the following:
Act as the diluent for WBC, RBC, PLT, and Hemoglobin measurements
Maintain cell volume during the counting and sizing portion of the
measurement cycle
Provide a conductive medium for impedance counting and sizing of cells and
platelets
Rinse the Aspiration Probe and fluidics
1-10
9140845CDecember 2009
Use or Function
Section 1
Overview
9140845CDecember 2009
1-11
Use or Function
Overview
Section 1
1-12
9140845CDecember 2009
Section 2
Overview
An Abbott-authorized representative should perform installation of the
CELL-DYN Emerald System. This is to ensure that all system components are
functioning correctly and to verify system performance. Installation procedures
must be repeated if the instrument is moved from the original installation site.
This section provides information about installation and customization of the
CELL-DYN Emerald. The beginning of this section provides the following
requirements and guidelines for installing the system:
Site requirements
Guidelines for unpacking and inspection, connection and start up, and
relocation
Guidelines for instrument set up by the Operator
9140846DDecember 2009
2-1
Overview
NOTES
2-2
9140846DDecember 2009
Section 2
Site Requirements
Site requirements for installation include the following:
Space requirements
Power requirements
Waste disposal requirements
Installation environment
NOTE: Refer to Section 7: Operational Precautions and Limitations for general
requirements and precautions for system operation.
Space Requirements
Select an appropriate location for the CELL-DYN Emerald System. The
instrument, printer and reagents weigh approximately 88 pounds (40 kg) and
should be placed on a surface that is adequate to support the weight. Allow for
sufficient space on the counter top to place the diluent at the same level as the
instrument.
CAUTION: Diluent must be placed at the same level as the instrument for
proper operation and to prevent bubbles from entering the diluent line.
Allow at least 4 inches (10 cm) of space behind the instrument for airflow. Make
sure there is adequate space around the instrument to perform necessary procedures
or service and allow the instrument to be easily disconnected from the power
source.
NOTE: To ensure the instrument and reagents function properly, it is important to
maintain the temperature between 64o 90o F (18o 32o C).
Place the instrument:
On a stable, level surface.
On a non-porous, non-absorbing work surface and flooring that can be easily
cleaned and disinfected using recommended procedures.
In an area that will not block the ventilation openings.
Away from direct sunlight.
Do not place the instrument near a centrifuge, X-ray equipment, video
display terminal, computer, or copier.
CAUTION: Do not use mobile telephones, wireless telephones, mobile
radios, or any other radio frequency (RF) transmitting devices in the same
room as the instrument.
9140846DDecember 2009
2-3
Site Requirements
Power Requirements
CAUTION: When connections are made, check all connectors for
particles or foreign material that can impair electrical contact.
Refer to Section 4: Performance Characteristics and Specifications for specific
power requirements.
Waste Disposal Requirements
WARNING: Potential Biohazard. Observe all biosafety and chemical
hazard precautions for waste disposal. For a detailed description of the
hazards associated with the CELL-DYN Emerald, refer to Section 8:
Hazards.
Observe the following requirements for waste routing and disposal:
Users are responsible for disposing of waste in accordance with local, state,
and federal regulations.
If a waste container is used, it must be labeled as biohazardous waste.
If a drain is used, it must be suitable for waste that could present a biological
or chemical hazard.
CAUTION: If routing waste to a sink, be sure that the waste outlet tube
is placed securely in the drain hole. To prevent a possible hazard, ensure
that all system components are located away from potential waste
overflow.
Installation Environment
The following are environmental requirements:
Indoor use only
Altitude up to 6562 feet (2000 meters)
Temperature range: 64o 90o F (18o 32o C)
Maximum relative humidity 80% for temperatures up to 90o F (32o C)
2-4
9140846DDecember 2009
Section 2
Installation
Inventory and Unpacking
Inspect the packaging before unpacking the instrument. Notify the shipper if there
is any damage. An authorized Abbott representative will uncrate, inspect and place
the instrument in the specified location.
Accessories
Ensure that the following accessories have been received:
Table 2.1
Accessories Kit
Item
Table 2.2
Installation Kit
Item
9140846DDecember 2009
Quantity
Quantity
AC Adapter/Transformer
US Power Cord
2-5
Installation
Table 2.3
Service Kit
Item
Quantity
Tubing L = 500 mm 2 x 4 mm
Tubing 9
Tubing 10
Cable Ties
EMLD Grease 3 gm
Instrument Preparation
1. Open the box from the top, remove the packing material and the Accessory
kit.
2. Remove the CELL-DYN Emerald from the box and remove the plastic
covering.
3. Perform the following visual checks:
2-6
9140846DDecember 2009
Installation
Figure 2.1
Instrument Inspection
9140846DDecember 2009
2-7
Installation
CAUTION: Moving parts inside. Use care when instrument power is ON.
Electrical Connections
1.
2.
3.
4.
5.
6.
7.
8.
USB Port
USB Port
Ethernet Connection (TCP/IP)
External Bar Code Reader (RS232)
LIS Connection (RS232C)
Printer Connection
Power Cord Connection
For ground use
1
2
3
4
5
6
8
Figure 2.2
Electrical Connections
2-8
9140846DDecember 2009
Installation
9140846DDecember 2009
2-9
Installation
Reagent installation
CAUTION: Before handling reagents, refer to the information provided
in Section 1: Use or Function, Subsection: Reagent System.
1. Remove the reagent door on the left side
of the instrument.
2. Place the reagent bottles in their
designated locations and remove the
caps.
3. Place the green collar reagent tubing (2)
and cap on the CELL-DYN Emerald
Cleaner Reagent (1) bottle.
4. Place the blue collar reagent tubing (4)
and cap on the CELL-DYN Emerald
CN-free Lyse Reagent (3) bottle.
Figure 2.3
2-10
4
3
9140846DDecember 2009
Installation
7
5
Figure 2.4
9140846DDecember 2009
2-11
Installation
Instrument Start Up
CAUTION: Ensure that the reagent and waste lines have been properly
connected before starting this procedure.
1. Power ON the instrument.
Ensure the AC Adapter and power
cord are connected
Press the POWER/ON/OFF key
2. The Cycle LED turns red and
remains red until the initialization
cycle is complete. The Cycle LED
turns green and the LOGIN screen
displays when the instrument is
ready to use.
NOTE: Cycles Left indicates the
number of runs available. This is
automatically updated, using
information from the reagents bar code
label whenever a reagent is changed, or
based on available capacity in the waste
container, whichever is less. At initial
startup all reagent volumes are set to
zero. Resetting of reagent volumes will
be required as part of the installation
process.
Figure 2.5
Power ON
3. Language Selection:
a. Touch the Dropdown menu and touch the
desired language to select it. It may be
necessary to use the scroll bar at the right of
the Language Dropdown menu to view all
available languages.
b. Touch [OK] to continue.
Figure 2.6
2-12
Language Selection
9140846DDecember 2009
Installation
Instrument Set Up
Operator Log In and Log Out
Each operator should log in to the CELL-DYN Emerald when operating the
instrument. Log In names can be up to ten characters in length. Each operator
should log out when they have finished running specimens.
The CELL-DYN Emerald provides four access levels (three are passwordprotected). All passwords can be up to four characters in length.
a. User (Operator) no password is entered
b. Lab Tech (Supervisor) default password is 1-2-3. The Supervisor should
enter a new password during the initial instrument set up.
c. Service for Abbott Service Personnel Only
d. Factory for Abbott Manufacturing Personnel Only
NOTE: After the instrument is initialized,
the <LOG IN> field displays the last log in
name that was entered.
1. If the Log In name is correct, proceed to
step a. If not, proceed to step 2.
a. Touch the <PASSWORD> field and
9140846DDecember 2009
2-13
Installation
System Status
The current status of the CELL-DYN Emerald can be displayed at any time by
touching the Date/Time display in the upper left corner of the display.
The status is displayed as follows in the
Information window that opens when the Date/
Time display is touched:
1. Instrument Serial Number (ID)
2. Operator ID (USER) the current entry
3. Operators access level for the current entry
4. Setup Version
5. Software (SW) Version
6. Loader current version (used to launch the
instrument software)
Touch [EXIT] to return to the previous menu.
NOTE: The information displayed will represent
the current software version(s) installed on your
instrument and may not match the screen
shown in here.
Figure 2.8
System Status
2-14
9140846DDecember 2009
Section 2
9140846DDecember 2009
2-15
Section 2
2-16
9140846DDecember 2009
WANT TO SAVE
MODIFICATIONS? Touch [YES] to
save and return to the Reagents
screen or touch [NO] to return to the
Waste screen.
5. Touch the HOME icon to return to the
MAIN menu.
NOTE: If you are disposing of waste
directly to a drain, enter 99999 in the
capacity field, then proceed as described
in Steps 3 through 5, above. The waste
counter will need to be reset when the
system has dispensed 99999 mL of waste.
Figure 2.11 Changing the Waste Container
9140846DDecember 2009
2-17
2-18
9140846DDecember 2009
Section 2
Set Up Menu
If necessary, touch the HOME icon to return to the
MAIN menu. From the MAIN menu, touch [SET UP]
to display the screen shown at the right.
1. Date and Time: Touch <TIME> and use the
numeric keypad to enter the time in the desired
(24 hour or AM/PM) format, If using AM/PM time
format, check the box next to PM to select PM or
deselect (uncheck) for AM. Touch <DATE> and
use the numeric keypad to enter the date in the
selected format. For information about selecting
AM/PM time format and selecting date format, refer
to Subsection: Other Settings within this section.
NOTE: The date format must match the format
defined under the ADVANCED setup menu.
IMPORTANT: Changing time setting. It is possible
to change the time ahead for the start of Daylight
Savings Time without effect on data. To reset the
time back one hour at the end of Daylight Savings
Time, perform this step at the beginning of the day,
after Start Up and before any samples are run to
avoid deletion of results.
IMPORTANT: Changing the date setting. The
Emerald system will delete results (from Datalog,
QC, and Calibration files) if the date is reset to a
past date. Before results are deleted, the system
prompts the user with a warning message that must
be acknowledged. Ensure that the system date is set
correctly prior to first use of the analyzer.
2. Language: Touch the Language dropdown menu
arrow to display the available languages and
touch the desired language to select it.
NOTE: Language may be set from either this display
or at the OPER. LOG lN/OUT screen.
3. Printer: Touch the Printer dropdown menu arrow
to display the available printers and touch the
appropriate printer to select it. An option for no
printer may also be selected.
4. Paper size: Touch the Paper Size dropdown
menu to display the available paper sizes for the
selected printer and touch the desired paper size
to select it.
NOTE: Choose the correct paper size for proper
printer functionality.
Figure 2.13 Date/Time Set Up
9140846DDecember 2009
2-19
Section 2
2-20
9140846DDecember 2009
Set Up Menu
Advanced Set Up
The Advanced Set Up button is available from Lab Tech access level and higher.
CAUTION: Any changes to these options can affect the quality of results.
Do not modify these settings until you have been properly trained to do so.
1. From the MAIN menu, touch [SET UP],
and then touch [ADVANCED] to
display the screen shown at the right.
2. Each button is described in detail on
the following pages.
9140846DDecember 2009
2-21
Set Up Menu
Printer Button
The Printer Button accesses the PRINTER SET UP menu that is used to configure
the printed report.
Touch the corresponding check box to
select from the following options:
1. EXPECTED VALUES prints the
laboratorys reference range (Patient
Limits) on each report.
2. HEADER prints a specified header.
Touch the header entry fields below the
Print Box and then touch [A-Z] and
enter header information, if desired.
3. COMMENTS - provides space for
handwritten comments on the printed
report.
4. UNITS prints the selected units on
the report.
5. INTERP. RPT prints the Interpretive
Report messages based on the Patient
Limits entered in step 1. Refer to the
following sub-section for a description
of the messages.
6. FLAGS prints all flags on the report.
7. GRAPHS prints all histograms on the
report.
8. AUTOPRINT To automatically print
ANALYSIS (Patients Results), QC
(Quality Control Result) or STARTUP
(Startup Results) select the
appropriate box(es).
To deselect an option, touch the check box
again.
When all selections have been made,
touch [CONFIRM] to exit and save the
entries. Touch [ESC] to exit without saving
changes.
Figure 2.15 Printer Set Up
Interpretive Report
Interpretive messages display and are printed when the values entered for Patient
Limits are exceeded. These messages display and print only when the Interpretive
Messages (Interp. Rpt) option on the Printer Set Up screen is selected.
Refer to Section 3: Principles of Operation, Table 3.7, Interpretive Report for a
comprehensive listing of interpretive messages.
2-22
9140846DDecember 2009
Set Up Menu
Communication Button
This button is used to configure the data transmission between the CELL-DYN
Emerald and a host Laboratory Information System (LIS).
Touch the corresponding box to select the
desired option.
1. HOST SERIAL Used to select RS232
communication mode.
2. HOST NET Used to select data TCP/IP
communication mode.
3. NONE Used when no data
transmission is desired.
4. EMERALD NUMBER Used to identify
instruments when more than one
Emerald is connected to the LIS system.
5. DELAY (ms) Lag time for start of
transmission in milliseconds.
6. HANDSHAKE If this box is selected,
data will not be transmitted if there are
any errors in the acknowledgement
process. If not selected, all data will be
transmitted regardless of
acknowledgement errors.
7. SERIAL PARAM Used to access
system configuration parameters for data
transmission via the RS232 port.
8. NET. PARAM. Used to access system
configuration parameters for data
transmission via TCP/IP or UDP/IP using
the Ethernet port.
9. Touch [CONFIRM] to save entries or
touch [ESC] to return to Main Menu.
9140846DDecember 2009
2-23
Set Up Menu
2-24
9140846DDecember 2009
Set Up Menu
9140846DDecember 2009
2-25
Set Up Menu
Type
This menu allows the operator to define up to 19 different types of specimens. For
example, specimen types can be used to establish separate limits for males,
females, neonatal, pediatric and other specimen types. In the Lab Preferences
menu, one can define separate Limits & Calibration Factors for each specimen
type. After defining specimen types in the Lab Preferences menu, the user can
choose a specific specimen type from the Run Sample screen and the
pre-defined limits and calibration factors set here will be applied.
IMPORTANT: When a non-Standard specimen run is completed, be certain to
select Standard from the dropdown menu on the Next Sample set up screen.
The Standard selection re-sets the instrument to the Standard Limits and
Calibration Factors.
2-26
9140846DDecember 2009
Set Up Menu
Limits
This screen allows entry of the
laboratorys Expected (l, h) and Panic
(L, H) limits.
1. Touch each entry field and use the
numeric keypad to type in the values.
2. Touch [ ] to display the remaining
measurands as shown on the
following screen.
3. Touch each entry field and use the
numeric keypad to type in the values.
4. When all entries have been made,
touch [CONFIRM] to exit and save the
entries. Touch [ESC] to exit without
saving changes.
5. Touch [<<] or [PREVIOUS SCREEN]
to return to the previous screen.
NOTE: Limits may be defined for any
specimen type, including STANDARD.
NOTE: Stored Patient Limit Set Flagging can be affected if limits are changed.
NOTE: The CELL-DYN Emerald does not have limits Factory Set. It is the responsibility of each laboratory
to determine limit sets appropriate for its patient population. A list of references for limit sets and limit set
determination may be found in Appendix D: CBC Reference Intervals: Literature Sources.
Figure 2.18 Limits Entry
9140846DDecember 2009
2-27
Set Up Menu
WANT TO SAVE
MODIFICATIONS?
Repeat this procedure to define
additional specimen types.
Figure 2.20 Specimen Types Set Up
2-28
9140846DDecember 2009
Set Up Menu
Calibration Factors
This option allows the operator to assign manually calculated calibration factors to
the standard specimen type.
IMPORTANT: Modifying these factors without performing the calibration
procedure will adversely affect results.
To change values for a selected specimen
type:
1. Touch [CALIBRATION FACTORS].
2. Touch the entry screen next to the
desired measurand.
3. Use the numeric keypad to enter the
value.
4. When entries are complete, touch
[CONFIRM] to exit and save the
entries. Touch [ESC] to exit without
saving changes.
9140846DDecember 2009
2-29
Set Up Menu
Other Settings
The Other Settings Button displays the following screen.
The Delay Box is used to configure the following
options:
1. SHUT DOWN Sets the time in minutes
between automatic shut down cycles.
a. Touch the entry field and use the numeric
keypad to enter the time in minutes. The default
value is 180 and the range is 30-480 minutes
2. PROBE UP Sets the time in minutes to retract
the probe when the instrument is not in use.
a. Touch the entry field and use the numeric
keypad to enter the time in minutes. The
default value is 5 minutes and the range is 1
to 60 minutes.
3. SET AUTOCLEAN Configures the system to
automatically run an Auto-clean cycle when the
specified number of cycles is reached.
a. Touch the entry field next to
<SET AUTOCLEAN> to select the option.
b. Touch the entry field next to <CLEAN
INTERVAL> and use the numeric keypad to
enter the number of cycles. The default value
is 80 cycles and the range is 10 to 5000 cycles.
NOTE: The maximum number of runs
recommended between Autoclean cycles is 80.
NOTE: Automatic autoclean occurs only when the
number of samples run since Start Up on a single
day exceeds the number entered at
<CLEAN INTERVAL (CYCLES)>.
4. PRESSURE CHECKING For Abbott Service
Personnel only.
5. The LAB TECH PASSWORD field can be used
to change the current password. Touch the
entry field and use either [A-Z] or the numeric
keypad to enter a new password. Make sure to
record the password for future reference.
6. The DATE FORMAT dropdown menu is used to
select the format for the date. Available options
are:
a. DD/MM/YY
b. MM/DD/YY
c. YY/MM/DD
Touch the dropdown menu button and then
touch the desired format to select it.
2-30
9140846DDecember 2009
Set Up Menu
9140846DDecember 2009
2-31
Set Up Menu
Version Release
The VERSION RELEASE button allows the operator to perform software upgrades
from the USB storage device.
1. Insert the USB key with the software upgrade in a free USB port on the rear
of the instrument.
NOTE: If another USB key is present in the other USB port, it must be removed
before proceeding.
2. Touch the [VERSION RELEASE] button.
3. Select the software version in the table on the right side of the screen by
touching the corresponding number in the left column.
4. The selected software version will display in the text entry field on the left
side of the display, below [VERSION RELEASE]. After confirming the
correct version of software is displayed, touch the [LOAD] button.
5. A pop-up message will display asking you to confirm loading the file. Touch
[YES] to continue with the software upgrade or [NO] to escape without
loading the software.
6. Follow any instructions on the screen during the software upgrade process.
7. When the software is successfully loaded, a confirmation message will
appear on the display.
Setup Configuration Printout
This Option allows the operator to print a copy of their SETUP menu
configuration.
From the MAIN menu
touch [SET UP].
1. Touch the [TOOL]
Icon. Touch [PRINT].
2. This print-out
contains most of the
setup menu
configuration. Save
this document for
future use.
9140846DDecember 2009
Set Up Menu
Start Up
From the MAIN menu, touch [START UP] to run the Start Up Cycle. The cycle
takes approximately two minutes. The cycle primes and flushes cleaner out of the
system, and checks the mechanical and electronic systems. At the end of the cycle,
a background count is automatically run and printed (if a printer is connected and
configured). When more than one background count is performed during Start Up,
only the final count is printed. If the Start Up cycle is successful, the operator is
returned to the MAIN menu, the Status LED turns green and the instrument is ready
to run specimens. If the Start Up cycle is unsuccessful, the message START UP
FAILED displays. If START UP FAILED displays, touch [START UP] again to
repeat it a second time. If Start Up fails two times, refer to Section 10,
Troubleshooting for guidance.
9140846DDecember 2009
2-33
Section 2
Background Count
A background count is automatically run at the end of the Start Up cycle.
Background counts can be run at any time by touching the Start Switch located
behind the Aspiration Probe. Results must be within specifications before
proceeding.
Startup Specifications can be found in Section 4: Performance Characteristics
and Specifications:
If startup background counts exceed limits refer to Section 10: Troubleshooting.
NOTE: The system must be configured to print the Start Up background count.
Carryover
Check the carryover as described in Section 4: Performance Characteristics and
Specifications.
Calibration
Complete the Pre-Calibration procedures and then verify calibration as described
in Section 6: Calibration.
Quality Control
When calibration has been verified, run three levels of control material as
described in Section 11: Quality Control.
2-34
9140846DDecember 2009
Section 3
Principles of Operation
Overview
The CELL-DYN Emerald is designed to automatically perform the following
functions:
Aspirate and dilute whole blood
Count, size and classify cells present in a whole blood specimen
Measure the hemoglobin concentration of a whole blood specimen
Analyze the raw data that is collected
Output results to the display, printer and laboratory information system
Two types of measurements and several innovative techniques are used to count,
size and classify blood cells and to measure hemoglobin. The two types of
measurements are:
Electrical Impedance Counting used for WBC, RBC and PLT
measurements
Absorption Spectrophotometry used for HGB measurement
Hemoglobin measurement is carried out in the WBC Counting Chamber using
absorption spectrophotometry. The reagent used for Hemoglobin measurement is
cyanide-free. Each measurement is discussed in detail as it relates to each
measurand.
This section also discusses Instrument Alarms and Data Flags.
9140847DDecember 2009
3-1
Principles of Operation
Section 3
Overview
Measurement Principles
Impedance Counting
Electrical Impedance counting is used to count and size WBCs, RBCs and
platelets. This method is based on the measurement changes in an electrical current
produced by particles (cells), suspended in a conductive liquid, as they pass
through an aperture of known dimensions. An electrode is submerged in the liquid
on either side of the aperture to create an electrical pathway through it.
As each cell passes through the aperture, a transitory change in the resistance
between the electrodes is produced. This change produces a measurable electrical
pulse. The number of pulses indicates the number of cells that traversed the
aperture. The amplitude of each pulse is essentially proportional to the volume of
the cell that produced it.
Each pulse is amplified and compared to internal reference voltage channels. These
channels are delineated by calibrated size-discriminators to accept only pulses of
certain amplitude. Thus, pulses are sorted into various size channels according to
their amplitude.
WBC Analysis
The instrument aspirates 9.8L from the whole blood specimen. The blood is
mixed with 2mL of diluent and 0.38mL of Lyse in the WBC counting chamber. The
lyse reagent destroys the RBC and resultant stroma and perforates the WBC
cytoplasmic membrane allowing the cytoplasm to escape. The WBCs are counted
directly by impedance and the Differential measurands are obtained from the
graph:
WBC
LYM%
Lymphocyte Percent
LYM#
MID%
MID#
GRA%
Granulocyte Percent
GRA#
3-2
9140847DDecember 2009
Principles of Operation
Section 3
Overview
Figure 3.1
WBC Graph
The preceding figure shows the WBC Graph. The lymphocytes are located
between the beginning of the curve and the M1 line. Mid Cells are located between
M1 and M2. Granulocytes are located between the M2 line and the end of the
curve. When the WBC result is outside the pre-set specification, the WBC
differential measurands and the graph are suppressed.
The WBC measurands are expressed as follows:
WBC
# K/L
LYM
# K/L and %
MID
# K/L and %
GRA
# K/L and %
RBC Analysis
The instrument aspirates 20L of the dilution from the WBC Counting chamber
and adds 1.5mL of diluent to the RBC Counting Chamber for the RBC/PLT
dilution. The RBCs are counted directly by impedance and four measurands are
obtained:
9140847DDecember 2009
RBC
HCT
Hematocrit
MCV
RDW
3-3
Principles of Operation
Section 3
Overview
Figure 3.2
RBC Graph
The Hematocrit (HCT) is the ratio of red blood cells to plasma and is expressed as
a percentage of the whole blood volume. It is derived from the volume of the RBCs
that are counted during the measurement cycle.
The mean cell volume (MCV) is the average volume of individual RBCs.
MCV =
HCT x 10
RBC
# M/L
HCT
MCV
femtoliters (fL)
RDW
% CV (coefficient of variation)
Platelet Analysis
Platelets are counted directly by impedance in the RBC Counting Chamber
at the same time as RBCs.
PLT
Platelet Count
MPV
Figure 3.3
Platelet Graph
The MPV is derived from the platelet graph after the platelet count has been
determined.
3-4
9140847DDecember 2009
Principles of Operation
Section 3
Overview
# K/L
MPV
femtoliters (fL)
9140847DDecember 2009
HGB
Hemoglobin concentration
MCH
MCHC
3-5
Principles of Operation
Section 3
Overview
The Hemoglobin value is used to calculate the MCH and MCHC as follows:
The MCH is the average amount of HGB contained in the RBC. It is calculated as
follows:
MCH = HGB x 10
RBC
The MCHC is the ratio of the weight of HGB to the volume of the average RBC.
It is calculated as follows:
MCHC = HGB x 100
HCT
The HGB measurands are expressed as follows:
HGB
g/dL
MCH
picograms (pg)
MCHC
g/dL
3-6
Exponents Values
Exponents Value
Prefix
Symbol
1012
tera
109
giga
106
mega
103
kilo
10-6
micro
10-12
pico
10-15
femto
9140847DDecember 2009
Principles of Operation
Section 3
Overview
Examples of the same specimen run with each of the available unit selections are
shown in the following table.
Table 3.2
Report Units
USA
Measurand
SI
SI MOD
Value
Units
Value
Units
Value
Units
WBC*
5.32
K/L
5.32
G/L
5.32
10e9/L
RBC
5.15
M/L
5.15
T/L
5.15
10e12/L
HGB
16.2
g/dL
162
g/L
10.1
mmol/L
HCT
47.6
0.476
L/L
0.476
L/L
MCV
92.3
fL
92.3
fL
92.3
fL
MCH
31.5
pg
31.5
pg
1.96
fmol
MCHC
34.1
g/dL
341
g/L
21.2
mmol/L
RDW
12.5
12.5
%CV
12.5
%CV
PLT
323
K/L
323
G/L
323
10e9/L
MPV
8.26
fL
8.26
f/L
8.26
fL
LYM#, MID# and GRA# are reported in the same units as WBC
9140847DDecember 2009
3-7
Principles of Operation
Section 3
Overview
NOTES
3-8
9140847DDecember 2009
Section 3
Principles of Operation
Instrument Alarms
There are two instrument alarm messages:
INS_T Indicates that the ambient temperature is lower than 63 F (17 C)
or higher than 91 F (33 C). Verify that the instrument is installed in an
environment that is within the installation environment listed in Section 2:
Installation Procedures and Special Requirements.
All results are invalidated (flagged with *) when this alarm appears.
INS_H Indicates that the hemoglobin channel is saturated. All results are
invalidated. Perform a start up cycle.
Operational Alerts
There are four operational alerts:
STARTUP NOT DONE
STARTUP FAILED
QC NOT DONE
QC ALERT
STARTUP NOT DONE A Start Up cycle has not been performed on the current date.
Action: Perform a Start Up cycle. Do not perform QC or patient testing until the
issue is resolved.
STARTUP FAILED Indicates that the last Start Up cycle did not meet specifications.
Action: Perform a Start Up cycle.
WARNING: Do not perform QC or patient testing until the Start Up passes.
If the Start Up cycle has been repeated twice without resolution refer to Section 10:
Troubleshooting, for additional information.
QC NOT DONE Indicates that a QC run has not been performed on the current date.
Action: Run QC according to your laboratorys requirements.
QC ALERT Indicates QC failed for one or more measurands on one or more levels
9140847DDecember 2009
3-9
Principles of Operation
Section 3
3-10
9140847DDecember 2009
Principles of Operation
Section 3
It is suggested that one Patient Limits set or the Panic Limits set be used to enter
instrument-specific laboratory action limits. A result that falls outside a laboratory
action limit can also indicate the need for the operator to follow a laboratory
protocol, such as repeating the sample, performing a smear review or notifying the
physician. In cases where a cellular abnormality is present that alters cellular
morphology to the extent that the cells do not fit the criteria used by the instrument
to generate a flag, dispersional data alerts may be the only flag(s) that will alert the
operator to a potentially erroneous result.
Action: When a result is flagged with a Patient or Panic Limits alert, it is
recommended that you follow your laboratorys review criteria, which may include
review of a stained smear to verify the result and to check for the presence of any
additional abnormality.
If results for a measurand exceed the upper end of the Display Range, a numeric
result does not appear on the screen or printout. Instead, plus symbols (++++)
appear in the results area and the letter D appears and prints to the right of the
++++.
9140847DDecember 2009
3-11
Principles of Operation
Section 3
Table 3.3
++++
Applicable
Measurands
Trigger
Other Affected
Measurands
Affected
Measurand
result
>100K/L
LYM
GRA
MID
----------
RBC
>8.0M/L
HCT
MCV
MCH
MCHC
RDW
----------------
MCV
>150fL
HCT
RBC
MCH
MCHC
RDW
----------------
HGB
>25g/dL
MCH
MCHC
-------
HCT
>80%
MCV
RBC
MCH
MCHC
RDW
----------------
PLT
>1500K/L MPV
----
Result
Flag Color on
Color on
Display
Display
Magenta
White
XXXX
Based on
limit
definition
N/A
N/A
Magenta
Magenta
XXXX
Based on
limit
definition
N/A
N/A
Yellow
Yellow
XXXX
Based on
limit
definition
N/A
N/A
XXXX
Based on
limit
definition
N/A
N/A
3-12
9140847DDecember 2009
Principles of Operation
Section 3
WBC Flags
L1
CL1
CL1-2
L2
CL2-2
Figure 3.4
L3
CL2
CL2-3
L4
L5
CL4
CL5
L1
Platelet Aggregates
NRBCs
Giant Platelets
Cryoglobulins
Incomplete Lysis of RBC
Small Lymphocytes
Fibrin Clots
Shift in WBC cell distribution due to EDTA anticoagulant equilibration
L2
Myelocytes
Lymphoblasts
Basophils
L3
Eosinophils
Myelocytes
L4
Granulocyte volume is too low
L5
Large size cells present
The WBC Flagging Regions are shown in the previous figure. All WBC flags are
shown in white text in the Flags box at the bottom of the result screen.
WBC Suspect Measurand Flags
These flags are generated after the instrument evaluates the measured data for a
particular measurand or group of measurands. The result may be suspect due to
interfering substances or the inability of the instrument to measure a particular
measurand due to a sample abnormality. The name of each flag, how it is displayed,
the cause of the flag, and the action to be taken are given in the following
explanations.
9140847DDecember 2009
3-13
Principles of Operation
Section 3
Table 3.4
WBC Flags
Result
Flag
Measurand
Text in
Flags Box
Result
Displayed
Cause
Action
WBC and
Differential
L1 (white
text)
XXXXX
Differential
L2 (white
text)
XXXXX
Differential
L3 (white
text)
XXXXX
Differential
L4 (white
text)
XXXXX
Granulocyte volume is
too low.
Differential
L5 (white
text)
XXXXX
Large-size cells
present.
3-14
9140847DDecember 2009
Principles of Operation
Section 3
P2
CP1
Figure 3.5
CP2
CP3-2
P3
P CP3
PLT Flags
The PLT Flagging Regions are shown in the previous figure. All PLT flags are
shown in yellow text in the Flags box at the bottom of the result screen.
PLT Suspect Measurand Flags
These flags are generated after the instrument evaluates the measured data for a
particular measurand or group of measurands. The result may be suspect due to
interfering substances or the inability of the instrument to measure a particular
measurand due to a sample abnormality. The name of each flag, how it is displayed,
the cause of the flag, and the action to be taken are given in the following
explanations.
PLT Suspect Measurand Flags invalidate the PLT count and all measurands
calculated from the PLT count (MPV).
9140847DDecember 2009
3-15
Principles of Operation
Section 3
Table 3.5
Platelet Flags
Result
Flag
Measurand
Text in
Flags Box
Result
Displayed
Cause
Action
PLT, MPV
P1 (yellow
text)
XXXXX
PLT, MPV
P2 (yellow
text)
XXXXX
PLT, MPV
P3 (yellow
text)
XXXXX
3-16
9140847DDecember 2009
Principles of Operation
Section 3
Interpretation
Applicable
measurands
Flag
Color on
Display
Flag color in
Flags box
Other affected
results
R_CL
---- *
Measurement
rejected/
invalidated,
possible RBC/
PLT aperture
clog.
RBC, PLT
white
red
W_CL
---- *
Measurement
rejected/
invalidated,
possible WBC
aperture clog.
WBC
white
white
R_CLW_
CL
---- *
Measurement
rejected/
invalidated,
possible RBC/
PLT and WBC
aperture clogs.
RBC, PLT,
WBC
white
red
HGB
COUNT
INVALID
---- *
HGB blank
reading failed.
HGB
white
white
9140847DDecember 2009
3-17
Principles of Operation
Instrument Alarms, Operational Alerts, and Measurand Data Flags
Section 3
Figure 3.6
3-18
9140847DDecember 2009
Principles of Operation
Section 3
Figure 3.6
9140847DDecember 2009
3-19
Principles of Operation
Section 3
Interfering Substances
It is important to note there are commonly occurring interfering substances that can
affect the results reported by hematology analyzers. While the CELL-DYN Emerald
has been designed to detect and flag many of these substances, it may not always
be possible to do so. Appendix B: Potential Causes of Spurious Results lists some
of the substances that may interfere with specific measurands.
Interpretive Messages (Interpretive Report)
Interpretive messages display and are printed when the values entered for Expected
Values (l and h) are exceeded. These messages display and print only when the
Interpretive Messages (Interp. Rpt) option on the Printer Set Up screen is selected.
Refer to Section 2: Installation Procedures and Special Requirements,
Subsection: Advanced Set Up for instructions for entering Expected Values.
Table 3.7
Interpretive Report
Messages
3-20
Measurand
WBC
Leukopenia
Leukocytosis
LYM%
Lymphopenia
Lymphocytosis
LYM#
Lymphopenia
Lymphocytosis
MID%
no message
no message
MID#
no message
no message
GRA
Granulocytopenia
Granulocytosis
GRA#
Granulocytopenia
Granulocytosis
RBC
no message
Erythrocytosis
HGB
Anemia
no message
MCV
Microcytosis
Macrocytosis
MCH
no message
no message
MCHC
Hypochromia
Cold Agglutinin
RDW
no message
Anisocytosis
PLT
Thrombocytopenia
Thrombocytosis
MPV
no message
Giant Platelets
9140847DDecember 2009
Section 4
Overview
This section describes the specification and performance characteristics of the
CELL-DYN Emerald. The following is included in this section:
Physical Specifications
Power Specifications
Environmental Specifications
Operational Specifications
Bar Code Specifications
Performance Specifications
9140848DDecember 2009
4-1
Overview
NOTES
4-2
9140848DDecember 2009
Section 4
Specifications
Physical Specifications
Physical specifications for the CELL-DYN Emerald are listed in the following
tables.
Table 4.1
Dimensions
Dimension
Instrument
Height
Width
Depth
Weight
Table 4.2
Dimension
Instrument
Height
Width
Depth
Weight
NOTE: Refer to the Printer Manual for physical specifications for the printer.
Physical Specifications for the AC Adapter are as follows:
Table 4.3
9140848DDecember 2009
Dimension
AC Adapter
Height
Width
Depth
Weight
4-3
Specifications
Power Specifications
Power requirements are as follows:
The instrument requires a constant, non-fluctuating power source. Use of an
AC line with dimmer switches can cause electrical current fluctuations that
could affect proper instrument function and therefore, is not recommended.
A circuit dedicated to the system is recommended but not required.
Table 4.4
Operating Range
Operating Cycles
120
99 132 VAC
50/60 HZ
220 240
50/60 HZ
Environmental Specifications
Operating Temperature from 64 90 F (18 32 C)
Storage Temperature from 14 122 F (-10 50 C)
CAUTION: If the CELL-DYN Emerald has been stored at a temperature
of less than 50 F (10 C), it must remain at room temperature for 24 hours
before switching on.
Relative Humidity 80% maximum at 88 F (31 C)
4-4
9140848DDecember 2009
Specifications
Operational Specifications
Throughput
60 samples/hour (approximate)
Storage capacity
Sample volume
9.8 L (approximate)
CYCLE
DILUENT
LYSE
CLEANER
12.6
0.38
0.55
Drain All
Refill
Back flush
0.55
Initialization**
10
28.2
10
Lyse
10
Diluent
30
Cleaner
10
Clean cycle
2.2
Bleach clean
32.5
0.38
0.55
Shut Down
16
Auto Rinse
13
Run sample*
All
Reagent
Prime
Start Up
9140848DDecember 2009
Based on one background run. The amount of reagent will increase if two or three
background cycles are run during the Start Up.
4-5
Specifications
4-6
9140848DDecember 2009
Section 4
Performance Specifications
Background Counts
Background values must be within the following specifications before testing
patient samples, running QC, or performing calibration:
Table 4.6
Background Specifications
Measurand
WBC
RBC
HGB
PLT
NOTE: The Background Specification applies only to WBC, RBC, HGB, and
PLT. There are no background specifications for other measurands, so if
results are displayed they should be disregarded.
Carryover
The following table shows carryover percent for WBC, RBC, HGB and PLT.
Carryover was determined by running whole blood specimens with high target
values of WBC, RBC, HGB and PLT. Each specimen was run in triplicate followed
by three aspirations of whole blood specimens with low target values. Carryover is
calculated and expressed as a percentage using the following formula:
Percent Carryover =
9140848DDecember 2009
4-7
Performance Specifications
Table 4.7
Carryover Specifications
Target Values
Measurand*
(units)
% Carryover (95%
Confidence Limit)
Low Target
Values
High Target
Value
WBC
K/L
>90
<1%
RBC
M/L
>6.20
<1%
HGB
g/dL
>22.0
<1%
PLT
K/L
>900
<2.2%
1.
4-8
9140848DDecember 2009
Performance Specifications
Table 4.8
Measurand
Units*
Display Range
AMR
WBC
K/L
0 100
0.4 96.1
RBC
M/L
08
0.22 7.61
HGB
g/dL
0 25
3.3 24.6
HCT
0 80
5.3 75.6
MCV
fL
0 150
48.8 115
PLT
K/L
0 1500
9 1375
NOTE: Results displayed with * indicate suspect results that should not be
reported. For additional information about flagged results, refer to
Section 3: Principles of Operation, Subsection: Instrument Alarms,
Operational Alerts, and Measurand Data Flags.
9140848DDecember 2009
4-9
Performance Specifications
Imprecision (Reproducibility)
Imprecision is expressed as the standard deviation (SD) or coefficient of variation
(CV) of analytic results in a set of replicate or duplicate measurements. Fresh
whole blood specimens used to verify imprecision specifications should have mean
values that fall within the range tested in the following table.
The table below represents the results of imprecision specifications for the
hemogram parameters. The stated CV% in the table represents instrument
imprecision from N=31 runs.
Table 4.9
Measurand
(units)
Ranges Tested
Observed %CV
Range
%CV (95%
Confidence
Limit)
WBC (K/L)
4.7 10.2
1.5 3.4
3.5
RBC (M/L)
4.2 5.4
0.7 1.9
2.0
HGB (g/dL)
12.2 16.1
0.4 1.8
2.1
HCT (%)
35.7 50.7
0.9 1.6
1.7
MCV (fL)
73.4 96.0
0.3 0.8
0.8
RDW (%)
11.8 17.0
2.1 3.4
3.3
PLT (K/L)
185.2 387
2.8 5.8
6.1
MPV (fL)
7.6 9.0
1.3 2.6
2.7
LYM %
13.1 50.1
1.7 5.0
5.4
MID %
6.3 11.0
3.4 7.3
8.1
GRA %
43.1 75.8
1.1 3.0
2.9
4-10
9140848DDecember 2009
Performance Specifications
Control
Level
CD-16
WBC X 109/L
Low
5.4
Normal
3.0
High
2.4
RBC X 1012/L
Low
2.0
Normal
1.8
High
1.7
Hb g/dL
Low
3.0
Normal
1.8
High
1.6
HCT %
Low
2.2
Normal
1.9
High
1.8
* Values are the sample count-weighted averages of the individual instrument and
individual lot %CV.
9140848DDecember 2009
4-11
Performance Specifications
Table 4.10
Low
1.0
Normal
0.9
High
0.8
MCH pg
Low
2.9
Normal
1.6
High
1.4
MCHC g/dL
Low
2.9
Normal
1.6
High
1.4
RDW %
Low
3.4
Normal
3.0
High
2.8
PLT X 109/L
Low
10.4
Normal
4.9
High
4.0
MPV fL
Low
3.5
Normal
1.8
High
1.6
* Values are the sample count-weighted averages of the individual instrument and
individual lot %CV.
4-12
9140848DDecember 2009
Performance Specifications
Table 4.11
Control
Level
CD-16 Control
Lymphocytes %
Low
5.6
Normal
1.5
High
1.9
Mid %
Low
9.7
Normal
5.1
High
2.9
Granulocytes %
Low
2.9
Normal
1.5
High
2.4
Lymphocytes - absolute
Low
10.5
Normal
3.2
High
3.6
Mid - absolute
Low
19.1
Normal
6.5
High
4.2
Granulocytes - absolute
Low
5.2
Normal
3.6
High
2.7
* Values are the sample count-weighted averages of the individual instrument and
individual lot %CV.
9140848DDecember 2009
4-13
Performance Specifications
Comparability (Correlation)
Evaluation of the correlation of the CELL-DYN Emerald is shown in the tables
below. This data was computed from Passing-Bablock regression analysis of data
obtained from studies performed on whole blood analyzed against a comparative
instrument using similar technology. The results in individual laboratories may
differ from these data.
The internal Abbott site obtained the following results on 330 blood samples.
Table 4.12
Measurand*
r-value
Min
Max
Intercept
Slope
WBC (K/L)
0.994
0.4
42.3
0.578
0.905
RBC (M/L)
0.993
1.31
7.38
-0.147
1.032
HGB (g/dL)
0.997
4.8
24.4
0.222
1.004
HCT (%)
0.993
14.7
66.9
-0.158
1.036
MCV (fL)
0.921
63.6
119.6
-12.170
1.175
RDW (%)
0.758
11.8
20.9
6.320
0.558
PLT (K/L)
0.990
2.0
1039.0
1.212
1.044
MPV (fL)
0.912
6.8
11.5
2.641
0.580
LYM (%)
0.970
4.0
75.6
1.240
1.011
MID (%)
0.761
1.6
17.8
1.560
0.895
GRA (%)
0.972
21.9
94.4
-1.782
0.990
4-14
9140848DDecember 2009
Performance Specifications
The external laboratory GA-Atlanta obtained the following results on 569 blood
samples.
Table 4.13
Measurand*
r-value
Min
Max
Intercept
Slope
WBC (K/L)
0.997
0.4
82.5
0.602
0.895
RBC (M/L)
0.992
1.36
6.81
-0.270
1.074
HGB (g/dL)
0.994
4.3
19.6
-0.230
1.038
HCT (%)
0.988
12.9
57.5
-0.810
1.032
MCV (fL)
0.943
60.7
110.1
-8.067
1.097
RDW (%)
0.750
10.9
26.4
4.467
0.690
PLT (K/L)
0.982
4.0
958.0
5.554
0.995
MPV (fL)
0.916
6.0
10.9
2.582
0.561
LYM (%)
0.986
2.5
76.9
1.057
1.034
MID (%)
0.819
1.7
19.8
2.005
0.979
GRA (%)
0.982
12.6
95.7
-7.150
1.048
Table 4.14
Measurand
Range Tested*
Replicates
r-value
Slope
Y-intercept
GRA%
23.15 95.70%
180
0.932
0.943
1.302
MID%
1.800 19.25%
180
0.874
0.612
3.350
LYM%
2.500 62.10%
180
0.943
0.989
3.317
* Results are expressed in traditional US units. These values do not represent the
analytical measurement range, which is provided in another table.
Correlation coefficient, established by Passing-Bablock regression analysis.
Interfering Substances
A wide variety of interfering substances have been shown to impact the results
from automated hematology analyzers. A list of substances that can potentially
interfere with CELL-DYN Emerald results is included in Appendix B: Potential
Causes of Spurious Results.
CELL-DYN Emerald Operators Manual
9140848DDecember 2009
4-15
Performance Specifications
Reference Intervals
Table 4.15
Reference Intervals
Measurand
(units)*
Sex
Range
WBC (K/L)
M/F
270
4.70 - 10.30
RBC (M/L)
M/F
270
4.03 - 5.46
HGB (g/dL)
M/F
270
12.40 - 16.90
HCT (%)
M/F
270
36.60 - 48.30
MCV (fL)
M/F
270
81.50 - 96.80
MCH (pg)
M/F
270
27.50 - 33.10
MCHC (g/dL)
M/F
270
32.40 - 35.70
RDW (%)
M/F
270
11.80 - 14.90
PLT (K/L)
M/F
270
165 - 385
MPV (fL)
M/F
270
7.20 - 10.20
LYM %
M/F
270
12.70 - 47.80
MID %
M/F
270
6.30 - 14.00
GRA %
M/F
270
43.50 - 78.90
These ranges do not represent globally applicable reference intervals, but reflect
combined reference ranges tested in the validation study. Each laboratory should
establish/verify its own reference intervals.
4-16
9140848DDecember 2009
Section 4
References
1. Clinical and Laboratory Standards Institute. Preliminary Evaluation of
Quantitative Clinical Laboratory Methods; Approved Guideline Third
Edition. CLSI document EP10-A3 [ISBN 1-56238-622-0] Clinical and
Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne,
PA, 19087-1898 USA, 2006.
2. Clinical and Laboratory Standards Institute. Evaluation of the Linearity of
Quantitative Measurement Procedures; A Statistical Approach; Approved
Guideline. CLSI document EP6-A [ISBN 1-56238-498-8] Clinical and
Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne,
PA, 19087-1898 USA, 2003.
3. Clinical and Laboratory Standards Institute. Evaluation of Precision
Performance of Quantitative Measurement Methods; Approved Guideline
Second Edition. CLSI document EP5-A2 [ISBN 1-56238-542-9] Clinical
and Laboratory Standards Institute, 940 West Valley Road, Suite 1400,
Wayne, PA, 19087-1898 USA, 2004.
9140848DDecember 2009
4-17
References
NOTES
4-18
9140848DDecember 2009
Section 5
Operating Instructions
Overview
This section discusses the operation of the CELL-DYN Emerald System. The
following topics are included:
Software Menu
Instrument Start Up
Specimen Analysis
Routine Operation
DATALOG
USB Flash Drive
Bar Code Scanner
Additional information for operating the CELL-DYN Emerald is discussed in the
following sections:
Calibration
Section 6: Calibration
Maintenance
9140849DDecember 2009
5-1
Operating Instructions
Section 5
Overview
NOTES
5-2
9140849DDecember 2009
Section 5
Operating Instructions
Instrument Logbook
It is suggested a logbook be created for the instrument. This logbook should
contain all necessary calibration documentation and other information that is
pertinent to the instrument. Suggested sections that you may want to include are:
Installation documentation
Your laboratorys operating procedure
Quality Control
Calibration
Maintenance
Reagent Lot Number Changes
Troubleshooting and Problem resolution
Service calls and problem resolution/service performed documentation
Software upgrade information
Store the logbook near the instrument so that it is accessible to operators and
Abbott Service Personnel. Sample log sheets may be found in Appendix E:
Sample Logs and Worksheets.
9140849DDecember 2009
5-3
Operating Instructions
Section 5
Instrument Logbook
Software Menu
The software menu is shown in this section. All sub-menus are accessed from the
MAIN menu, which displays when HOME is touched. The sub-menus used for
setting up the instrument are described in Section 2: Installation Procedures and
Special Requirements, Subsection: Instrument Set Up and Advanced Set Up.
Remaining sub-menus are discussed in appropriate sections. For example, the
Calibration menu is discussed in Section 6: Calibration and Quality Control is
discussed in Section 11: Quality Control.
MAIN MENU*
START UP
OPER. LOG
IN/OUT
EVENT LOG
REAGENTS
DILUENT
LYSE
CLEANER
WASTE
PRIME
DILUENT
PRIME
LYSE
PRIME
CLEANER
RESET
SHUT DOWN
PRIME ALL
DATE
REPORTING
OPTIONS
SET UP
MAINTENANCE
RUN SAMPLE
NEXT
SAMPLE
VIEW
RUN AND
RESULTS
RUN AND
RESULTS
EDIT
COMMUNICATION
DATALOG
CYCLE
COUNTER
EDIT
PRINTER
QUALITY
CONTROL
CALIBRATION
LAB
PREFERENCES
LAB ID
INFORMATION
RESTORE
ADVANCED
PRINTER3
UPDATE
CALIBRATION
FACTORS
OTHER
SETTINGS
VERSION
RELEASE
STORAGE
TYPE
FORMAT EXT.
STORAGE DEVICE
DRAIN
ALL
REFILL
LIMITS
THRESHOLDS
INITIALIZE
SYSTEM
MOTOR INIT
CLEAN
BACKFLUSH
SPECIMEN CAL.
ADJUSTMENTS
ALARMS
FLUDICS
BLEACH
MECHANICAL
CHECKS
INIT
PRESS
RENAME
SPECIMENS
SPECIAL
MODES
SYS CLEAN
PRECISION
LINEARITY
ERROR LOG
ABBOTT
SERVICE
CHECK
VALVES
CHECK
PROBE
CHECK
ROCKER
CHECK
SYRINGE
CHECK
SENSORS
HGB LED
OFF
REMOVE
PROBE
LUBRICATION
POS.
Figure 5.1
5-4
Software Tree
9140849DDecember 2009
Operating Instructions
Section 5
Instrument Logbook
Instrument Power On
Check the reagent and waste levels before starting the days run. If reagents need
to be replaced or waste needs to be emptied, refer to instructions given in
Section 2: Installation Procedures and Special Requirements, Subsection:
Replacing the Reagents Diluent, Lyse, Cleaner. Check that there is sufficient
paper in the printer.
If the printer has been powered OFF, press the power button to turn it ON. If the
instrument has been turned off, the display is black and the LED light is off. Press
the Power button to turn the instrument on.
1. The instrument initializes and checks the motors.
2. The cycle LED turns red. The instrument cannot be cycled until initialization
is complete and the LED turns green.
3. When initialization is complete, the OPER. LOG IN/OUT screen displays.
Operator Log In
If the instrument is ON, touch [OPER. LOG IN/OUT] and log in as described in
Section 2: Installation Procedures and Special Requirements, Subsection:
Operator Log In.
Specimen Analysis
Specimen Collection and Handling
WARNING: Potential Biohazard. Consider all clinical specimens,
reagents, controls, surfaces or components that contain or have contacted
blood, serum, or other bodily fluid as potentially infectious. Wear gloves,
lab coats, and safety glasses, and follow other biosafety practices as
specified in the OSHA Bloodborne Pathogen Rule (29CFR Part 1910.1030)
or other equivalent biosafety procedures.
NOTE: For additional information on collecting venous specimens, refer to CLSI
Standards H3-A51.
9140849DDecember 2009
5-5
Operating Instructions
Section 5
Instrument Logbook
5-6
9140849DDecember 2009
Operating Instructions
Section 5
Instrument Logbook
9140849DDecember 2009
5-7
Operating Instructions
Section 5
Instrument Logbook
NOTES
5-8
9140849DDecember 2009
Section 5
Operating Instructions
Routine Operation
Daily Start Up Procedures
The Daily Start Up procedures include:
Running the Start Up Cycle and confirming that Background Counts are
within acceptable limits
Performing daily Quality Control checks
From the MAIN menu, touch [START UP] to run the Start Up Cycle. The cycle
takes approximately two minutes. It primes and cleans the system, and checks the
mechanical and electronic systems. At the end of the cycle, a background count is
automatically run and printed if a printer has been connected and configured. Up
to three (3) background counts may be automatically performed as part of the Start
Up Cycle. The final background will be automatically printed, if a printer is
configured. If the Start Up cycle is successful, the operator is returned to the MAIN
menu, the Status LED turns green, and the instrument is ready to run specimens. If
the Start Up cycle is unsuccessful, the message START UP FAILED displays. If
START UP FAILED displays, touch [START UP] again to repeat it a second time.
If Start Up fails two times, refer to Section 10: Troubleshooting for guidance.
1. If a Start Up Cycle failed the message START UP FAILED is displayed and
printed with every result when the user is logged in as an Operator or
Supervisor. The system must be configured to display these flags. See
Section 2: Installation Procedures and Special Requirements, Subsection:
Reporting Options Button.
2. Background counts must be within acceptable limits before running controls
or patient specimens.
When the Start Up Cycle is successfully completed, perform daily Quality Control
checks according to the regulations governing your laboratory before running
patient specimens. Refer to Section 11: Quality Control.
NOTE: The message QC ALERT displays and prints below the patient results
when the QC results are out of tolerance or the lot has expired. The system
must be configured to display these flags. See Section 2: Installation
Procedures and Special Requirements, Subsection: Reporting Options
Button. The message QC NOT DONE displays and prints below the
results when QC has not been run for the day.
9140849DDecember 2009
5-9
Operating Instructions
Section 5
Routine Operation
NOTES
5-10
9140849DDecember 2009
Section 5
Operating Instructions
Running Specimens
Ensure Background Counts are within acceptable limits and Quality Control results
are acceptable according to your laboratorys Quality Control program. Ensure
specimens have been properly mixed.
WARNING: Potential Biohazard. Consider all clinical specimens,
reagents, controls, surfaces or components that contain or have contacted
blood, serum, or other bodily fluid as potentially infectious. Wear gloves,
lab coats, and safety glasses, and follow other biosafety practices as specified in the OSHA Bloodborne Pathogen Rule (29CFR Part 1910.1030) or
other equivalent biosafety procedures.
From the MAIN menu, touch [RUN SAMPLE]. To enter NAME, PID, or/and SID
press the [NEXT SAMPLE] at the bottom of screen. Enter the desired information
and press confirmation. To enter all of the identification, including the Name, PID,
SID, and select the type, refer to the instructions provided in Subsection:
Specimen Identification, earlier in this section.
NOTE: Do not run linearity material in the patient mode. Run linearity material
in the LINEARITY mode. Refer to Section 9: Service and Maintenance.
9140849DDecember 2009
5-11
Operating Instructions
Running Specimens
Section 5
5-12
Results Screen
9140849DDecember 2009
Operating Instructions
Section 5
Running Specimens
Printing Results
DATALOG
The CELL-DYN Emerald Datalog stores 1500 records, including demographics,
results and graphs. Records are archived with a Sequence Number (SEQ), assigned
by the instrument software. Sequence numbers begin at 0001 for the first sample
of each day. Start Up Cycles, QC, and Precision runs are all assigned sequence
numbers, but are not stored in the Datalog. From the MAIN menu, touch
[DATALOG] to access the Datalog screen shown in the following figure. Follow the
instructions provided to access the Datalog information.
NOTE: When the Datalog is first opened from the MAIN menu, the display shows
the most current runs. If no results are stored in the internal memory and
a USB flash drive is connected, the system will display results from the
most current date stored on the drive.
9140849DDecember 2009
5-13
Operating Instructions
Running Specimens
Section 5
Datalog Screens
CELL-DYN Emerald Operators Manual
9140849DDecember 2009
Operating Instructions
Section 5
Running Specimens
Figure 5.6
DATALOG Options
9140849DDecember 2009
5-15
Operating Instructions
Section 5
Running Specimens
Figure 5.8
5-16
Autorinse
9140849DDecember 2009
Section 5
Operating Instructions
9140849DDecember 2009
5-17
Operating Instructions
Section 5
Figure 5.9
5-18
9140849DDecember 2009
Operating Instructions
Section 5
7. Select the [YES] button to proceed or select the [NO] button to exit without
formatting the USB drive.
8. If [YES] was selected, the CELL-DYN Emerald will proceed to format the
USB drive. Once the formatting is completed, an AB18 folder is created in
the USB drive. When the formatting process is complete you will be returned
to the EXT. STORAGE DEVICE screen.
9. Touch the HOME icon to return to the MAIN menu. The USB drive may now
be removed. Your USB drive is ready to be used with the
CELL-DYN Emerald.
USB File Folder Structure
Information is stored on the USB drive in the following folder structure:
Root Folder
(Emerald
analyzers only)
AB18
LOT
SETTINGS
Common Folders
(can be shared between
Emeralds)
UPDATE
PRTDRV
VERSION
XXXXXX
YYYYYY
CALI
EQC
LOG
Analyzer Serial
Number Folders
(cannot be
shared
with another
Emerald)
QC
RESULTS
2008-04
17
Result Folders
(The analyzer
results are
archived in
folders in a
YYYY-MM then
sub-folder DAY
format)
18
SETTINGS
REAGENTS
NOTE: Datalog files are saved to the USB drive in a format read by the
CELL-DYN Emerald. To easily view the contents of the files, open the
file from your CELL-DYN Emerald rather than on a PC.
From the MAIN menu, touch [DATALOG], touch [DATE], touch the dropdown
menu at the upper right, select MASS STORAGE and review the desired date.
9140849DDecember 2009
5-19
Operating Instructions
Section 5
3. Select ALL to save all data to the USB drive or select DAY and enter the
sequence numbers corresponding to the date range you want to save.
4. Touch the [SAVE] button to save the selected data to the USB drive.
5. Touch the [EXIT] button to exit the menu.
Use of the USB drive in the REAGENT menu.
Use of the USB drive in the REAGENT menu is restricted to Abbott personnel.
Installing a Printer Driver
NOTE: Follow Abbotts specific instructions provided with the printer driver to
copy a new printer driver onto the USB drive.
1. Insert the USB drive into either USB port on the rear panel of the
CELL-DYN Emerald.
2. From the MAIN menu, touch the [SET UP] button.
3. Touch [PRINTER 3 UPDATE] at the bottom of the SET UP screen.
5-20
9140849DDecember 2009
Operating Instructions
Section 5
4. The CELL-DYN Emerald will display a list of the printer drivers that are
available to be loaded.
9140849DDecember 2009
5-21
Operating Instructions
Section 5
6. Confirm the correct printer and port type appear in the upper left of the screen
and touch [LOAD].
7. A pop-up box will appear with the message: DO YOU CONFIRM TO LOAD
THE FILE selected printer name will appear here? Touch [YES] to load
the file and [NO] to exit without changes.
8. A pop-up box will appear with the message, THE PRINTER DRIVER XXX
HAS BEEN SUCCESSFULLY LOADED. Touch [OK].
Use of the USB drive in the CALIBRATION menu.
Saving current calibration data.
Data for the current calibration factors may be saved to the USB drive as a
Calibration Report. The Calibration Report contains the following information for
the current calibration stored on the instrument: Calibration date, operator ID for
the calibration, calibration factors for all calibrated parameters, lot number of
calibration material, expiration date of calibration material, assay values and limits
for all calibrated parameters.
1. From the MAIN menu, touch the [CALIBRATION] button.
The instrument will display a screen showing the current calibration factors
(FCTR column) as well as the assay values used in that calibration. Lot
number and expiration date are displayed at the top of the screen. All of this
information may be saved to the USB drive.
2. Insert the USB drive into either USB port on the rear panel of the
CELL-DYN Emerald.
3. Touch the [TOOLS] button.
5-22
9140849DDecember 2009
Operating Instructions
Section 5
5. When the pop-up box appears prompting: YOU ARE ABOUT TO SAVE
THE CALIBRATION REPORT. DO YOU CONFIRM? Touch [YES] to
save the report or [NO] to exit without saving the report.
9140849DDecember 2009
5-23
Operating Instructions
USB Flash Drive (Thumb Drive)
Section 5
NOTE: If QC information is loaded to a file that has already been used, all
existing information, including QC results, will be erased.
4. Select [EDIT].
5. Touch [LOAD].
6. Identify the desired lot number and level corresponding to the QC material
you will be using in the table on the right side of the display and touch it to
select.
The selected lot will display in the box in the upper left side of the display,
the level selected will be to the right of the lot number.
7. After verifying that this is the desired lot number and level, touch the [LOAD]
button.
5-24
9140849DDecember 2009
Operating Instructions
Section 5
8. In the pop-up box DO YOU CONFIRM TO LOAD THE FILE < >?,
select [YES] to load the file or [NO] to exit without loading the file.
9. Confirm all information on the next screen against the product labeling
provided with the Control material. Correct any information as necessary.
Once the correct information is displayed, touch [CONFIRM] or touch [ESC]
to exit without saving.
9140849DDecember 2009
5-25
Operating Instructions
Section 5
5-26
9140849DDecember 2009
Operating Instructions
Section 5
5. When all information has been correctly entered, touch [CONFIRM] to save
the information or touch [ESC] to exit without saving.
6. At the WARNING Pop-up: DO YOU WANT TO SAVE
MODIFICATIONS?, touch [YES] to save or touch [NO] to exit without
saving.
NOTE: Information must be entered into each field to successfully submit
information to the eQC program.
Saving QC data to the USB key for eQC
1. Insert a USB drive into either of the USB ports (1 or 2) on the back panel of
the CELL-DYN Emerald.
2. From the MAIN menu, touch the [QUALITY CONTROL] key.
3. Touch the TOOLS icon.
4. Select one or more lot numbers using the radio buttons and touch the [SAVE]
button.
9140849DDecember 2009
5-27
Operating Instructions
Section 5
NOTE: If a QC file containing data is selected you will see the WARNING Popup: YOU ARE GOING TO DELETE ALL ASSOCIATED RESULTS.
DO YOU WANT TO CONTINUE? Touch [YES] to continue, or touch
[NO] to exit without deleting data.
4. Select the lot and level to be restored from the list on the right side of the
display by touching the desired row.
5-28
9140849DDecember 2009
Operating Instructions
Section 5
The selected lot will appear in the box on the upper left side of the display with the
level displayed to the right.
5. After verifying that the information is correct, touch the [RESTORE] button.
6. At the WARNING Pop-up: DO YOU CONFIRM TO RESTORE THE
FILE, touch [YES] to restore the file or [NO] to exit without restoring.
Use of the USB drive in the DATALOG menu.
The information stored in the instrument DATALOG can be copied to the USB
drive.
1. From the MAIN menu, touch the [DATALOG] key.
9140849DDecember 2009
5-29
Operating Instructions
Section 5
3. Select the data to be saved by selecting the radio button for ALL or SEQ. If
using SEQ for the selection, enter the SEQ numbers in FROM and TO fields
corresponding to the data range desired.
4. Touch the [SAVE] button. At the WARNING Pop-up: YOU ARE ABOUT
TO COPY RESULTS FROM DATALOG TO THE EXTERNAL STORAGE
DEVICE. DO YOU CONFIRM? touch [YES] to save the data or [NO] to
exit without saving.
If YES is selected a pop-up box will briefly appear showing the progress of the save
operation. When the data has been saved this pop-up box disappears and it is safe
to remove the drive.
Use of the USB drive in the SET UP menu.
The information stored in instrument SET UP can be copied to the USB drive.
1. From the MAIN menu, touch the [SET UP] button.
2. Touch the TOOLS icon.
5-30
9140849DDecember 2009
Operating Instructions
Section 5
3. Touch the [SAVE] button to save the current set up to the USB drive. At the
WARNING Pop-up: YOU ARE ABOUT TO SAVE THE USER
SETTINGS. DO YOU CONFIRM?, touch [YES] to save or [NO] to exit
without saving.
NOTE: This function can be used to save SET UP information prior to service and
restore the information afterwards.
Restoring SETUP information from the USB drive.
1. From the MAIN menu, touch the [SET UP] button.
2. Touch the TOOLS icon.
9140849DDecember 2009
5-31
Operating Instructions
USB Flash Drive (Thumb Drive)
Section 5
5. Touch [YES] to load the stored settings from the USB drive. This will
overwrite all settings currently stored on the system. Touch [NO] to exit
without loading the stored settings.
Use of the USB drive in the MAINTENANCE and RUN SAMPLE menus.
There are no USB functions available to operators in the MAINTENANCE or RUN
SAMPLE menus.
Bar Code Scanner
Overview
The CELL-DYN Emerald is shipped with a bar code scanner that may be used for
information entry in the RUN SAMPLE menu. The instrument software can be
configured to input scanned information into either the PID, name, or SID fields.
Please refer to Section 2: Installation Procedures and Special Requirements
Subsection: Set Up Menu, for configuration instructions.
The barcode scanner is also used to input reagent information. Please refer to
Section 2: Installation Procedures and Special Requirements, Subsection:
Using Reagent Bar Codes for instructions on scanning reagent bar codes.
For additional information about the bar code scanner, such as instructions for use
and troubleshooting, please consult the manufacturers instructions packaged with
the bar code scanner.
5-32
9140849DDecember 2009
Section 5
Operating Instructions
References
1. Clinical and Laboratory Standards Institute/NCCLS. Procedures for the
Collection of Diagnostic Blood Specimens by Venipuncture; Approved
Standard Sixth edition. CLSI/NCCLS document H3-A6 (ISBN 1-56238650-6) CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898,
2007.
2. Clinical and Laboratory Standards Institute (CLSI). Procedures and Devices
for the Collection of Diagnostic Capillary Blood Specimens; Approved
StandardSixth Edition. CLSI document H04-A6 [ISBN 1-56238-677-8].
Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite
1400, Wayne, Pennsylvania 19087-1898 USA, 2008.
3. Thompson CB, Diaz DD, Quinn PG, Lapins M, Kurtz SR, Valeri CR. The role
of anticoagulation in the measurement of platelet volumes. Am J Clin Pathol
1983 Sep;80(3):327-32.
4. McShine RL, Sibinga S, Brozovic B. Differences between the effects of
EDTA and citrate anticoagulants on platelet count and mean platelet volume.
Clin Lab Haematol 1990;12(3):277-85.
5. International Council for Standardization in Haematology (ICSH). Protocol
for Evaluation of Automated Blood Cell Counters. Clinical and Laboratory
Hematology 1984; 6:69-84.
9140849DDecember 2009
5-33
Operating Instructions
Section 5
References
NOTES
5-34
9140849DDecember 2009
Section 6 Calibration
Section 6
Calibration
Overview
Calibration is a procedure that confirms the accuracy of the CELL-DYN Emerald.
The instrument is initially calibrated at the factory. During installation, an Abbott
representative will assist the laboratory in verifying the factory calibration.
The CELL-DYN Emerald is designed to remain stable without frequent calibration
when it is operated and maintained according to the recommendations provided in
this manual.
The following measurands can be calibrated by CELL-DYN Emerald operators:
WBC
RBC
HGB
MCV
PLT
The following measurands can be calibrated by Abbott service personnel or in the
factory:
MPV
RDW
The following information is discussed in the section:
When to Calibrate
Calibration Guidelines
Pre-Calibration Procedures
Calibration Procedures
Calibration Verification Procedure
9140850CDecember 2009
6-1
Calibration
Section 6
Overview
NOTES
6-2
9140850CDecember 2009
Section 6
Calibration
When to Calibrate
Scheduled calibration of the CELL-DYN Emerald must conform to the guidelines
established by the regulatory agencies governing the laboratory.
Calibration should be confirmed on a regular basis according to your laboratorys
protocols. The Quality Control program on the CELL-DYN Emerald provides
confirmation of instrument calibration. The decision to re-calibrate can be based on
Quality Control results. For information on Quality Control, refer to Section 11:
Quality Control.
Criteria should also be established for calibration verification. Calibration
verification criteria include:
When indicated by Quality Control data
After major maintenance and service procedures
At least every six months
As directed by the regulatory agencies governing the laboratory
A common method for calibration verification is to process a commercial
calibrator and compare the results with the assay values provided by the
manufacturer. If instrument results exceed verification criteria, the instrument
should be re-calibrated.
If needed, calibration is the last step in a troubleshooting sequence. Frequent,
unnecessary, recalibrations can mask underlying instrument problems and should
be avoided.
NOTE: If there are questions about when to re-calibrate, contact Abbott
Diagnostics Customer Service.
9140850CDecember 2009
6-3
Calibration
Section 6
When to Calibrate
NOTES
6-4
9140850CDecember 2009
Section 6
Calibration
Calibration Guidelines
Calibration Materials
The CELL-DYN Emerald can be calibrated with commercial calibrator, such as
CELL-DYN Calibrator, or with assayed whole blood specimens. Commercial
calibrator is the method of choice.
Commercial calibrator must be used before its expiration date and should be
used according to the manufacturers instructions.
Pre-Calibration Procedures
It is advisable to perform calibration when it can be completed without
interruption. Pre-Calibration Procedures ensure proper instrument performance
and a successful calibration. These steps should be completed just before starting
the CELL-DYN Emerald calibration process. If problems are detected during these
checks, do not attempt to calibrate the instrument. After the problems have been
resolved, repeat the Pre-Calibration Procedures to verify proper instrument
performance.
A Pre-Calibration Procedures Checklist is provided in Appendix E: Sample Logs
and Worksheets. The checklist outlines the procedures and is used to document the
results. It can be duplicated as needed.
Calibration Menu
From the MAIN menu, touch [CALIBRATION] to display the Calibration screen
shown in the following figure.
9140850CDecember 2009
6-5
Calibration
Calibration Guidelines
Section 6
6-6
Calibration Screen
9140850CDecember 2009
Calibration
Section 6
Calibration Guidelines
9140850CDecember 2009
6-7
Calibration
Section 6
Calibration Guidelines
9140850CDecember 2009
Calibration
Section 6
Calibration Guidelines
9140850CDecember 2009
6-9
Calibration
Calibration Guidelines
Section 6
6-10
9140850CDecember 2009
Section 6
Calibration
References
1. International Council for Standardization in Haematology (ICSH). Protocol
for Evaluation of Automated Blood Cell Counters. Clinical and Laboratory
Hematology 1984; 6:69-84.
2. Clinical and Laboratory Standards Institute/CLSI. Procedure for
Determining Packed Cell Volume by the Microhematocrit Method; Approved
Guideline Third edition. CLSI document H7-A3 (ISBN 1-56238-413-9)
CLSI, 940 West Valley Road, Suite 1400, Wayne, PA, 19087-1898 USA
2000.
9140850CDecember 2009
6-11
Calibration
Section 6
References
NOTES
6-12
9140850CDecember 2009
Section 7
Overview
This section addresses operational requirements, precautions, and limitations to
ensure operator safety and accurate test results. Failure to follow these
requirements or take these precautions may cause damage to the system, impact
system performance, or adversely affect results, or present a hazard to the operator.
Operational precautions and limitations topics include:
General requirements
Precautions and requirements for System Operation
Requirements for handling consumables
Requirements for handling specimens
Requirements for collecting, preparing and storing specimens
Interfering substances and conditions
Limitations of result interpretation
Other factors you should consider when interpreting patient test results
9140851CDecember 2009
7-1
Overview
NOTES
7-2
9140851CDecember 2009
Section 7
General Requirements
The following general CELL-DYN Emerald system requirements should be
followed to help ensure proper system performance:
Ensure the system is located out of direct sunlight, heat and drafts, and away
from any heat-generating device. Exposure to heat and drafts can interfere
with the ability of the system to maintain an operating temperature that is
within the acceptable range.
Place the instrument on a hard, level surface. Maintain the required space on
all sides of the system. For more information about space requirements, refer
to Section 2: Installation Procedures and Special Requirements. This
clearance is essential for:
Adequate ventilation and cooling of electrical components
Easy access for maintenance
Easy access for disconnecting the power cord when required
Place the instrument away from centrifuges, x-ray equipment and copiers.
CAUTION: Do not use mobile telephones, wireless telephones, mobile
radios, or any other radio frequency (RF) transmitting devices in the same
room as the instrument.
NOTE: The CELL-DYN Emerald has been evaluated to EN 55011 and EN 61000
for electromagnetic emissions and immunity, respectively.
Leave the system power on continuously unless instructed otherwise in a
maintenance or troubleshooting procedure, or unless an emergency situation
occurs.
Ensure the waste line is connected to the appropriate outlet and is routed to a
suitable waste container or drain. Dispose of all waste materials in
accordance with local, state and federal regulations.
If an external waste container is used, ensure the top of the waste container is
placed below the bottom of the analyzer.
If a drain is used for waste, ensure the waste outlet tube is secured in the drain
hole. Ensure system components are located away from potential waste
overflow.
Perform maintenance procedures as recommended in Section 9: Service and
Maintenance.
Do not attempt any maintenance or repairs that are not specified in
documentation provided by Abbott Laboratories. An Abbott-authorized
representative should perform all major service work.
9140851CDecember 2009
7-3
General Requirements
7-4
9140851CDecember 2009
General Requirements
Hg
9140851CDecember 2009
7-5
General Requirements
9140851CDecember 2009
General Requirements
9140851CDecember 2009
7-7
General Requirements
NOTES
7-8
9140851CDecember 2009
Section 7
Reference
International Council for Standardization in Haematology (ICSH). Protocol
for Evaluation of Automated Blood Cell Counters. Clin Lab Haemat. 1984;
6:69-84.
9140851CDecember 2009
7-9
Reference
NOTES
7-10
9140851CDecember 2009
Section 8 Hazards
Section 8
Hazards
Overview
9140852CDecember 2009
8-1
Hazards
Section 8
Overview
NOTES
8-2
9140852CDecember 2009
Section 8
Hazards
Operator Responsibility
Operators are responsible for using the CELL-DYN Emerald system only as
designed. Operators should be trained before being allowed to operate the system.
Failure to follow instructions for safe use of the system could cause personal injury,
damage to the system, or could adversely affect results. See also Section 7:
Operational Precautions and Limitations.
Safety Icons
Safety icons in this manual and on the CELL-DYN Emerald System identify
potentially dangerous conditions. Learn the icons and understand the type and
degree of potential hazard they represent.
The following icons may be used with text or in lieu of text. If text accompanies
the icons, it describes the nature of the hazard and is labeled with WARNING or
CAUTION.
WARNING: is defined as a physical, mechanical, or procedural condition that
could result in moderate to serious personal injury.
CAUTION: is defined as a condition that could result in minor injury or interfere
with proper functioning of the system.
Table 8.1
Safety Icons
Icon
NOTE:
Hazard
Description
CAUTION:
Important Information:
9140852CDecember 2009
8-3
Hazards
Section 8
Operator Responsibility
8-4
9140852CDecember 2009
Hazards
Section 8
Operator Responsibility
Do not eat, drink, smoke, apply cosmetics, or handle contact lenses when
handling human-sourced material or contaminated system components.
WARNING: Potential Biohazard. Identifies an activity or area where
you may be exposed to potentially infectious material.
Clean spills of potentially infectious materials and contaminated system
components with an appropriate disinfectant, such as 0.5% sodium
hypochlorite or another suitable disinfectant. Refer to Appendix C:
Preparation of diluted sodium hypochlorite solutions, Subsection:
Decontamination Procedure, for instructions for preparing the sodium
hypochlorite solution.
Decontaminate and dispose of all specimens, reagents, calibrators, controls,
and other potentially contaminated materials in accordance with local, state,
and federal regulations.
If you are exposed to biohazardous or potentially infectious materials, seek
medical attention immediately and take steps to clean the affected area.
The following label is affixed to the CELL-DYN Emerald System:
Figure 8.1
9140852CDecember 2009
Biohazard label
8-5
Hazards
Section 8
Operator Responsibility
The Biohazard label is affixed to the Start Switch below the Aspiration Probe as
shown in the following figure:
Figure 8.2
The Biohazard label is also affixed to the rear of the instrument next to the waste
outlet as shown in the following figure:
Figure 8.3
8-6
9140852CDecember 2009
Hazards
Section 8
Operator Responsibility
Chemical Hazards
Operators and others may be exposed to hazardous chemicals when handling
reagents, calibrators, and controls.
Exposure to hazardous chemicals is minimized by following instructions provided
in the manufacturers documentation, such as product labels, package inserts or
Material Safety Data Sheets (MSDS).
Precautions
In general, observe the following precautions when handling chemicals:
Consult MSDS for safe-use instructions and precautions.
Avoid contact with skin and eyes. If contact with material is anticipated, wear
impervious gloves, protective eye wear and clothing.
Maintain good housekeeping. Do not eat, drink, or store food and beverages
in areas where chemicals are used.
Seek medical attention if irritation or signs of toxicity occur after exposure.
Hazard symbols that appear on CELL-DYN Emerald System product labeling may
be accompanied by risk (R) and safety (S) numbers and represent risk and safety
phrases defined by European Community Directives. The risk and safety phrases
describe precautions to be used when working with a particular chemical or
chemical mixture.
Spill Clean-Up
Clean spills in accordance with established biosafety practices and follow
instructions provided in the MSDS. In general, use these safe work practices for
cleaning spills:
1. Wear appropriate personal protective equipment, such as a lab coat,
protective eyewear and gloves.
2. Absorb the spill with absorbent material.
3. Wipe the spill area with detergent solution.
4. Wipe the area clean with an appropriate disinfectant such as 0.5% sodium
hypochlorite or another suitable disinfectant. Refer to Section 9: Service and
Maintenance, Subsection: Decontamination Procedure, for instructions for
preparing the sodium hypochlorite solution.
5. Dispose of spilled and contaminated material in accordance with local, state,
and federal regulations.
Waste Handling and Disposal
Dispose of all waste materials in accordance with local, state, and federal
regulations.
It is the responsibility of each facility to label all waste containers and to
characterize its waste stream to ensure the waste is disposed in accordance with the
appropriate waste disposal regulations.
CELL-DYN Emerald Operators Manual
9140852CDecember 2009
8-7
Hazards
Section 8
Operator Responsibility
Electrical Hazards
The CELL-DYN Emerald System does not pose uncommon electrical hazards to
operators if it is installed and operated without alteration, and is connected to a
power source that meets required specifications. Refer to Section 4: Performance
Characteristics and Specifications for power requirements and specifications.
The electrical circuit spacing of the CELL-DYN Emerald system is based on
pollution degree 2 and altitude [up to 2000 M (6562 ft)] as per IEC 61010-15.
Pollution degree 2 is defined as an environment where normally only nonconductive pollution occurs. Occasionally, however, a temporary conductivity
caused by condensation should be expected.
Electrical Safety
WARNING: Electrical Hazard. Indicates the possibility of electrical
shock if operating or servicing instructions are not followed.
Basic electrical hazard awareness is essential to the safe operation of any system.
Only qualified personnel should perform electrical servicing. If the instrument is
used or modified in a manner not specified by the manufacturer, the protection
provided by the instrument may be impaired.
Elements of electrical safety include, but are not limited to, the following:
Inspect electrical cabling into and on the CELL-DYN Emerald system for
signs of wear and damage.
Use only approved power cords and electrical accessories, such as those
supplied with the system, to protect against electric shock.
Use a properly grounded electrical outlet of correct voltage- and currenthandling capability.
Do not disconnect any electrical connection or service any electrical or
internal components while the power is ON.
Keep liquids away from all electrical or communication component
connectors.
Do not touch with wet hands any switches or outlets.
Keep the floor under and around the CELL-DYN Emerald system dry and
clean.
Clean spilled fluids immediately. Always unplug the instrument before
cleaning up major liquid spills.
8-8
9140852CDecember 2009
Hazards
Section 8
Operator Responsibility
Mechanical Hazards
The CELL-DYN Emerald system is an automated system that operates under
computer control. As with most automated equipment, there is potential for injury
and bodily harm from moving mechanical components whenever the system is in
operation.
The CELL-DYN Emerald System minimizes mechanical hazards by providing
protective covers and locking mechanisms to protect against accidental contact
with mechanical and moving components.
During operation of the CELL-DYN Emerald system, the operator is potentially
exposed to the moving Aspiration Probe Assembly.
Basic elements of mechanical safety include, but are not limited to the following:
Never bypass or override a safety device.
Keep all protective covers in place.
Never allow any part of the body to enter the region of movement of any
mechanical component when the system is operating.
Never perform manual tasks on the surface of the system.
Open or remove covers only as directed during routine and as-needed
maintenance, component troubleshooting, or reagent removal and
replacement procedures described in Section 9: Service and Maintenance,
Section 10: Troubleshooting and Section 2: Installation Procedures and
Special Requirements, Subsection: Replacing the Reagents Diluent,
Lyse, Cleaner.
Use caution and wear a lab coat, protective eyewear and powder-free gloves
when operating the instrument, cleaning the instrument, reagent handling,
and when performing maintenance or service procedures. Always use
protective equipment when specified.
Do not wear long hair loose or articles of clothing or accessories that could
catch on the system.
Keep pockets free of items that could fall into the system.
9140852CDecember 2009
8-9
Hazards
Operator Responsibility
Section 8
Physical Hazards
Safe practices should be observed to avoid physical injury in the following
situations:
Aspiration Probes
WARNING: Aspiration probes are potentially contaminated with
infectious material. Avoid contact with the tip of the probe. Although the
CELL-DYN Emerald system presents and lowers the probe for sample
aspiration, never reach into the access area until the probe is in place.
Place the aspiration probe in an appropriately labeled, puncture-resistant, and
leakproof container before treatment and disposal.
Heavy Objects
CAUTION: Identifies an activity where you may be required to lift or
move a heavy object. Use proper lifting techniques.
The CELL-DYN Emerald Diluent and waste containers are heavy when full. Use
proper lifting techniques to reduce the risk of injury when handling the containers.
The CELL-DYN Emerald System is heavy. Ensure there is adequate help before
attempting to move the system.
Tripping Hazard
The CELL-DYN Emerald System is equipped with power cords. To avoid a
tripping hazard, ensure that cords avoid high traffic areas or are properly stowed.
8-10
9140852CDecember 2009
Section 8
Hazards
References
1. US Department of Labor, Occupational Safety and Health Administration, 29
CFR Part 1910.1030, Occupational Exposure to Bloodborne Pathogens.
2. US Department of Health and Human Services. Biosafety in Microbiological
and Biomedical Laboratories. Fourth Edition. Washington, DC: US
Government Printing Office, May 1999.
3. World Health Organization. Laboratory Biosafety Manual. Geneva: World
Health Organization, 1993.
4. Clinical and Laboratory Standards Institute. Protection of Laboratory
Workers from Occupationally Acquired Infections; Approved Guideline
Third Edition. CLSI document M29-A3 (ISBN 1-56238-567-4). Clinical and
Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne,
Pennsylvania 19087-1898, USA, 2005.
5. IEC 61010-1, International Electrotechnical Commission - World Standards
for Electrical and Electronic Engineering, 61010: - Safety Requirements for
Electrical Equipment for Measurement, Control, and Laboratory Use,
61010-1 (2001) Part 1: General Requirements.
9140852CDecember 2009
8-11
Hazards
Section 8
References
NOTES
8-12
9140852CDecember 2009
Section 9
Overview
The CELL-DYN Emerald System has been designed to require minimal routine
maintenance. To ensure optimum performance, the operator is encouraged to
routinely perform the scheduled maintenance procedures described in this section.
Failure to perform recommended maintenance can result in inaccurate or imprecise
results.
This section describes the recommended preventative maintenance procedures and
provides instructions for storing, relocating or shipping the instrument.
The maintenance schedule outlined in this section minimizes operational problems
with the CELL-DYN Emerald. The recommended intervals are based on
instruments operating in laboratories analyzing up to 50 specimens per day from a
general patient population. If more than 50 specimens per day are typically
analyzed, adjust the maintenance frequency proportionally. These intervals are
affected by several factors, including:
Number of specimens processed
Work load schedule
Operating environment
Patient population being analyzed
Each laboratory must assess its own situation and modify these recommended
intervals as necessary.
NOTE: Overdue maintenance is usually indicated by an increase in imprecision
of one or more of the directly measured measurands. This increase may
be due to carryover or dilution/sampling inconsistencies. If this occurs on
more than a random basis, the appropriate maintenance should be
performed more frequently.
WARNING: Potential Biohazard. Consider all clinical specimens,
reagents, calibrators, controls or other materials and all system surfaces or
components that contain or come in contact with human-sourced material
as potentially infectious. Wear a lab coat, gloves and safety glasses when
performing service or maintenance. Follow biosafety practices as specified
in the OSHA Bloodborne Pathogen rule (29 CFR Part 1910.1030)1 or other
equivalent biosafety practices.2, 3
Gloves should be powder-free; powder can cause instrument problems.
Maintenance procedures should not be performed without adequate training.
9140853DDecember 2009
9-1
Overview
NOTES
9-2
9140853DDecember 2009
Section 9
Event Log
The instrument provides an Event Log. This log maintains a record of certain
activities by date and holds 365 records, one for each day that the instrument is run.
Access the log from the MAIN menu by touching [EVENT LOG]. The Event Log is
shown in the following figure.
The days of operation are displayed in the
first row with day 001 being the first day of
the instrument operation - installation.
When a specific day is selected, the date,
day number, and the number of runs for
the selected day are displayed to the right
of the log. The abbreviations used in the
log are as follows:
SPR: Supervisor log in
SVC: Service log in
FCT: Factory log in
INT: Service Operation(s) performed
SUP: Start Up Cycle passed
SUF: Start Up Cycle failed
SDN: Shut Down Cycle
DIL: Diluent Reagent changed
LYS: Lyse Reagent changed
CLN: Cleaner changed
ACN: Cleaning Cycle (either automatic
or manual)
BLH: Bleach Cleaning
CAL: Calibration performed
QC: QC performed
DEL: Results deleted from Datalog
When the Event Log approaches capacity
the operator will be alerted to print the
Event Log. When the Event Log has 365
entries it will begin to delete entries on a
first-in first-out basis or the operator can
manually delete selected entries. The last
100 entries are maintained, even though
other entries are deleted.
An x will appear in the day column if the
specific event has occurred once or more
in that day.
Figure 9.1
9140853DDecember 2009
9-3
Event Log
9-4
9140853DDecember 2009
Event Log
Maintenance Menu
The MAINTENANCE Button is found on the MAIN menu. The options accessible
from the MAINTENANCE Button are described, below.
1. Initialize System. This button is used to
initialize the Emerald following an
Emergency Stop or as directed during
instrument troubleshooting.
2. Motor Init. Used for troubleshooting as
directed by an Abbott Representative.
3. Fluidics. Refer to Subsection: Fluidics
Menu Options, found later in this section.
4. Mechanical Cycles. Used for
troubleshooting as directed by an Abbott
Representative.
5. Special Modes. Used to access the
Precision and Linearity buttons,
described later in this section.
6. Sys Clean. Used as described in
Subsection: Preparing for Storage,
Relocation or Shipping, found later in
this section.
7. Error Log. An automatic log of all system
errors generated by the
CELL-DYN Emerald. For use in
troubleshooting as directed by an Abbott
Representative.
9140853DDecember 2009
9-5
Event Log
9-6
Figure 9.3
Figure 9.4
Precision Data
Figure 9.5
9140853DDecember 2009
9140853DDecember 2009
Event Log
Figure 9.6
Figure 9.7
9-7
Event Log
NOTES
9-8
9140853DDecember 2009
Section 9
Fluidics Menu
9140853DDecember 2009
9-9
NOTES
9-10
9140853DDecember 2009
Section 9
Monthly Maintenance
Bleach Cleaning
Cleaning the system with a bleach solution is performed monthly or as needed
when a measurand is repeatedly rejected.
Materials Required
Powder-free gloves, lab coat, safety glasses
Sodium hypochlorite (undiluted, unscented purchased bleach)
Plastic transfer pipette or other disposable pipette with 2 mL capacity
Flat screwdriver
Prepare a 3.6% sodium hypochlorite (bleach) solution according to the formula in
Appendix C. Transfer the prepared solution to the squirt bottle.
9140853DDecember 2009
9-11
Section 9
Procedure
Access the bleaching function as follows:
1. From the MAIN menu, touch
[MAINTENANCE] to display the
MAINTENANCE menu.
2. From the MAINTENANCE, touch
[FLUIDICS] to display the FLUIDICS menu.
(shown at the right)
3. From the FLUIDICS menu, touch [BLEACH]
to start the bleaching process and drain the
counting chambers.
9-12
Bleach Cleaning
9140853DDecember 2009
Semi-Annual Maintenance
Lubricating the Pistons
For optimal operation, the Syringe Pistons should be lubricated every six months
as described below.
Short arm T20 torx tool
NOTE: This procedure is used only for the 4 white teflon syringe pistons. The
metal sample syringe piston should not be lubricated.
Materials Required
Powder-free gloves, lab coat, safety glasses
Silicone Grease supplied in the Installation Kit
Short Arm Torx T20
Flat screwdriver / key
Procedure
Move the syringe pistons to the lubrication
position as follows:
1. From the MAIN menu, touch
[MAINTENANCE] to display the
MAINTENANCE menu.
2. From the MAINTENANCE menu, touch
[MECHANICAL CHECKS], then touch
[LUBRICATION POS.] to move the
syringe pistons to the lubrication
position.
9140853DDecember 2009
9-13
As Needed Maintenance
Emergency Stop
If there is a mechanical or fluidic problem, press the ON/OFF button to
immediately stop the instrument. Under certain conditions, the system may halt
due to an internally detected fault condition. In either situation, when the problem
is corrected, it is necessary to initialize the system. Before initializing the system,
respond to any pop-up messages displayed. This function is accessible from the
MAINTENANCE menu as shown in the following Figure.
Initialize System
To initialize the system after an Emergency
Stop:
1. From the MAIN menu touch
[MAINTENANCE].
2. Touch [INITIALIZE SYSTEM].
3. When initialization is complete, perform
the Daily Start Up Procedures as
described in Section 5: Operating
Instructions, Subsection: Daily Start
Up Procedures.
Figure 9.12 Maintenance Menu
9-14
9140853DDecember 2009
9140853DDecember 2009
9-15
Section 9
Decontamination Procedure
1. Place 1 mL of the 0.5% sodium hypochlorite solution in a clean test tube,
immerse the tip of the aspiration probe in the bleach solution, and touch [RUN
SAMPLE] to decontaminate the fluidics.
2. When the instrument is ready for the next cycle, dampen a gauze pad with the
0.5% sodium hypochlorite solution and wipe the outside of the Aspiration
Probe.
3. Initiate a CLEAN OUT cycle as follows:
NOTE: Use the 3.6% sodium hypochlorite (bleach) for these steps.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
9-16
9140853DDecember 2009
Clean
The Clean procedure may be used for troubleshooting, as described in Section 10:
Troubleshooting. This is an automated process.
1.
2.
3.
4.
Backflush
The Backflush procedure may be used for troubleshooting, as described in
Section 10: Troubleshooting. This is an automated process.
1.
2.
3.
4.
9140853DDecember 2009
9-17
9-18
9140853DDecember 2009
9140853DDecember 2009
9-19
9-20
9140853DDecember 2009
9140853DDecember 2009
9-21
Section 9
9-22
9140853DDecember 2009
Section 9
References
1. US Department of Labor, Occupational Safety and Health Administration.
29CFR Part 1910.1030. Occupational Exposure to Bloodborne Pathogens.
2. World Health Organization. Laboratory Biosafety Manual. Geneva: World
Health Organization, 1993.
3. Clinical and Laboratory Standards Institute. Protection of Laboratory
Workers from Occupationally Acquired Infections; Approved Guideline
Third Edition. CLSI document M29-A2 (ISBN 1-56238-567-4). Clinical and
Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne,
Pennsylvania 19087-1898, USA, 2005.
9140853DDecember 2009
9-23
References
NOTES
9-24
9140853DDecember 2009
Section 10 Troubleshooting
Section 10
Troubleshooting
Overview
This section provides information for use in troubleshooting instrument problems.
The CELL-DYN Emerald continuously monitors system status and provides
instrument alarms and warnings when problems are detected.
9140854CDecember 2009
10-1
Troubleshooting
Section 10
Overview
NOTES
10-2
9140854CDecember 2009
Section 10
Troubleshooting
General Information
Conditions that are instrument-or reagent-related will usually be seen on all
specimens, including controls. When patient results are suspect, it is important to
confirm instrument performance by running controls. If control results are also
suspect, troubleshoot as directed in this section.
WARNING: Potential Biohazard. Follow established biosafety practices
when performing maintenance, service, or troubleshooting procedures.
Refer to Section 8: Hazards for additional information.
Troubleshooting Guide
Understanding normal instrument operation is essential for identifying and
resolving operational problems. Effective troubleshooting requires a logical, stepby-step approach to problem solving. Logical troubleshooting can be divided into
three steps:
1. Problem Identification requires the Operator to investigate not only what
is wrong but to note what is right. The investigation should identify the
problem area and eliminate areas that are working correctly. When this step
is done, move to the next step.
2. Problem Isolation further classifies an instrument problem. Problems are
generally divided into three categories:
Measurement-related to sample analysis
Software-related
Hardware component-related
Hardware and software problems may or may not be operator-correctable with
technical assistance. Measurement problems are generally operator-correctable.
They are further divided into problems related to sample handling, maintenance or
calibration.
3. Corrective Action involves taking appropriate steps to correct the
problem. If the operator can correct the problem, with or without technical
assistance, normal operation can quickly resume.
9140854CDecember 2009
10-3
Troubleshooting
Section 10
General Information
10-4
9140854CDecember 2009
Troubleshooting
Section 10
General Information
Troubleshooting Log
An example of a Troubleshooting Log is provided in Appendix E: Sample Logs
and Worksheets. This log can be used to keep a record of troubleshooting
observations and problem resolution and can be photocopied as needed. This
record can be useful for providing troubleshooting guidance.
Table 10.1
Measurand-Related Problems
NOTE: This table can be used for troubleshooting of patient results or QC.
Measurand
Problem
No WBC and HGB results
WBC Clog Error (W-CL)
1.
2.
3.
4.
5.
WBC
9140854CDecember 2009
Corrective Action
10-5
Troubleshooting
Section 10
General Information
Table 10.1
NOTE: This table can be used for troubleshooting of patient results or QC.
Measurand
RBC
HGB
PLT
10-6
Problem
Corrective Action
1.
2.
3.
4.
No results
Imprecise results
1.
2.
3.
4.
Elevated background
1. Prime diluent.
9140854CDecember 2009
Troubleshooting
Section 10
Table 10.2
General Information
Instrument-Related Problems
Source
Problem
Corrective Action
No power
No display
CELL-DYN Emerald
Printer
9140854CDecember 2009
10-7
Troubleshooting
Section 10
General Information
Table 10.3
Alarm Messages
Message
Corrective Action
BACKUP:LAST QC RESTORED
BACKUP:SECTOR FAILED
BACKUP:SYSTEM ERROR
10-8
9140854CDecember 2009
Troubleshooting
Section 10
Table 10.3
General Information
COM:CHECKSUM ERROR
CYCLE BUSY
CYCLE:CMD FAILED
CYCLE:PRESSURE DEFAULT
CYCLE:VALVE 1 FAILED
CYCLE:VALVE 2 FAILED
CYCLE:VALVE 3 FAILED
CYCLE:VALVE 4 FAILED
CYCLE:VALVE 5 FAILED
9140854CDecember 2009
10-9
Troubleshooting
Section 10
General Information
Table 10.3
CYCLE:VALVE 6 FAILED
CYCLE:VALVE 7 FAILED
CYCLE:VALVE 8 FAILED
CYCLE:VALVE 9 FAILED
CYCLE:VALVE 10 FAILED
CYCLE:VALVE 11 FAILED
HARDWARE:SYSTEM ERROR
INIT PRINTER
10-10
9140854CDecember 2009
Troubleshooting
Section 10
Table 10.3
General Information
INTERN:MEMORY CORRUPTED
NO PRINTER RESPONSE
9140854CDecember 2009
10-11
Troubleshooting
Section 10
General Information
Table 10.3
NO PRINTER SELECTED
OUT OF RANGE
PRINTER ERROR
PRINTER IS BUSY
PRINTER:NO DRIVER ON
PRINTER:NO PAPER
PRINTER IS OFF
HOST:ACK ERROR
COM:INTERNAL ERROR
HOST:SYNCRO ERROR
SYSTEM:BUSY
SYSTEM:FATAL ERROR
10-12
9140854CDecember 2009
Troubleshooting
Section 10
Table 10.3
General Information
SYSTEM:UNKNOWN ERROR
UNKNOWN COMMAND
USB:EMPTY FILE
INVALID FORMAT
9140854CDecember 2009
10-13
Troubleshooting
Section 10
General Information
Table 10.3
10-14
9140854CDecember 2009
Section 10
Troubleshooting
References
1. US Department of Labor, Occupational Safety and Health Administration.
29CFR Part 1910.1030. Occupational Exposure to Bloodborne Pathogens.
2. World Health Organization. Laboratory Biosafety Manual. World Health
Organization, Geneva 2003.
3. Clinical and Laboratory Standards Institute. Protection of Laboratory
Workers from Occupationally Acquired Infections; Approved Guideline
Third Edition. CLSI document M29-A3 (ISBN 1-56238-567-4). CLSI, 940
West Valley Road, Suite 1400, Wayne, PA 19087-1898, 2005.
9140854CDecember 2009
10-15
Troubleshooting
Section 10
References
NOTES
10-16
9140854CDecember 2009
Section 11
Quality Control
Overview
Quality Control (QC) procedures are performed at a frequency needed to monitor
System results trueness. Quality Control procedures, both internal and external,
allow the operator to verify the performance of the analytical system. Evaluation
of results from commercial and patient controls facilitates the interpretation of
laboratory data to determine the acceptability of patient results. This section
discusses:
When to Run QC
Quality Control Methods and Materials
QC File Set Up
QC Guidelines
Running Controls
Levey-Jennings Graphs
9140855CDecember 2009
11-1
Quality Control
Section 11
Overview
NOTES
11-2
9140855CDecember 2009
Section 11
Quality Control
When to Run QC
Each laboratory should determine the frequency of performing quality control runs
with commercial controls and/or retained patient specimens. This may be specified
by the regulatory agencies governing the laboratory. Quality Control specimens
must be run and results confirmed to be within acceptable limits before reporting
patient results. Abbott recommends you, run controls:
After daily Start Up procedures are completed and background counts are
within specifications
After a reagent lot number change
After calibration (confirmatory step)
When there is an unusual shift or trend in patient results
After performing maintenance, a service call or component replacement
After a software change or upgrade
According to your laboratorys quality control protocol
According to regulatory requirements
9140855CDecember 2009
11-3
Quality Control
Section 11
When to Run QC
QC File Setup
Six QC files are available on the CELL-DYN Emerald. Each file stores 100 results
and the instrument automatically calculates the Mean, Standard Deviation and
Coefficient of Variation % for those results. Levey-Jennings graphs are
automatically plotted for each file. QC files are accessed from the MAIN menu by
touching [QUALITY CONTROL].
Overview of the QC window.
The LOTS window, shown at the right,
shows the lot number for each file.
1. The last file that was used is identified
with a dot in the circle to the left of the
lot number.
NOTE: If files have not been set up, touch
the file location (space next to the circle) to
choose a file.
NOTE: The level of control (L=low,
N=normal, H=high) material associated
with each file is displayed to the right of the
lot number. When the file is selected, a box
appears surrounding the radio button and
lot number. The selected level designation
is not within the box.
2. Touch the lot number to select a file.
3. Touch [EDIT] to enter new Assay
Values and Limits.
4. Touch [RUN AND RESULTS] to view
file data.
5. Touch the TOOLS icon to print:
Assays (and Limits), Results or
Graphics (Levey-Jennings Graphs)
6. Touch [SEND] to send results to the
LIS.
7. Touch [EXIT] to return to the previous
screen.
NOTE: If using the instrument keypad
arrows to navigate on this screen, it is
necessary to press the ENTER key to
complete the file selection.
Figure 11.1 QC Window
11-4
9140855CDecember 2009
Quality Control
Section 11
When to Run QC
WANT TO SAVE
MODIFICATIONS? Touch [YES] to
save or [NO] to return to the entry
screen. To exit without saving, touch
[ESC].
The [RESTORE] button is used to load QC
files that were backed up to an external
storage device. Files can only be restored
to the CELL-DYN Emerald on which they
were created.
Figure 11.2 Lab ID Information
9140855CDecember 2009
11-5
Quality Control
When to Run QC
Section 11
11-6
9140855CDecember 2009
Quality Control
Section 11
When to Run QC
9140855CDecember 2009
11-7
Quality Control
When to Run QC
Section 11
11-8
9140855CDecember 2009
Section 11
Quality Control
QC Guidelines
General Information
Quality Control procedures must be performed in accordance with your
laboratorys protocols, good laboratory practice, and according to regulatory
requirements according to the following general guidelines:
Before running patient specimens, run a minimum of two levels of control.
Confirm that results are within acceptable limits and monitor the data for
shifts and trends.
If results fall outside acceptable limits, try another vial from the same lot. If
results are still outside acceptable limits, refer to Section 10:
Troubleshooting.
Always run controls in the same manner as patient specimens.
Control Material Guidelines
Always mix and handle commercial controls according to the directions
provided in the package insert. Proper handling and mixing are essential for
accurate results.
Pay particular attention to the following:
Check the condition of incoming control material. Be sure that vials are at the
proper temperature and are not leaking. Check for hemolysis.
If controls are stored inside a refrigerator, place them in a central location
away from the door.
Check the expiration date and open-vial Stability dating. Do not use products
longer than recommended by the manufacturer or results can be
compromised.
Normal Specimen Guideline: The Rule of Three
Because normal erythrocytes (RBC) are biconcave discs with a uniform quantity
of hemoglobin (HGB) in each cell, there is a fairly constant numeric relationship
involving RBC, HGB, and hematocrit (HCT)5 that is informally known as the rule
of 3:
HCT = HGB X 3
HGB = RBC X 3
If there is no RBC pathology (i.e., normal cell size, normal RBC hemoglobin
concentration and no abnormal forms), a discrepancy in these numeric ratios
suggests analytic error in one or more of the measurands. For example, a lipemic
(fatty plasma) specimen may produce a falsely high HGB result due to turbidity
interference; in this case the HCT:HGB ratio will be less than 3, while the
HGB:RBC ratio will be greater than 3.
9140855CDecember 2009
11-9
Quality Control
QC Guidelines
Section 11
This rule of 3 applies only to fresh normal human specimens, and not to
commercial control materials.
Over the years, clinicians and technologists have used a quick mathematical check
to ensure that HGB and HCT values are consistent. The HCT is roughly three times
the HGB. This simple formula works only on specimens that are normochromic,
normocytic (specimens with MCV and MCHC values within typical reference
ranges). Typically, results from CELL-DYN instruments correlate within the
narrow range of 3% on normal specimens. If the calculated HCT (HGB x 3) does
not agree with the instrument value with an increasing frequency (>5%) of outliers,
the system could be miscalibrated, malfunctioning or the specimen has pathology
requiring further investigation.1
Running Controls
Before running controls, be sure the vials have been properly warmed and mixed
according to the directions given in the manufacturers package insert.
NOTE: If a Start Up cycle has not been done, an error message will display
indicating one must be run before running controls. It is not possible to
run controls until the Start Up cycle is run.
11-10
9140855CDecember 2009
Quality Control
Section 11
QC Guidelines
9140855CDecember 2009
11-11
Quality Control
Section 11
QC Guidelines
Sending QC Results
To send the contents of a QC file to the LIS, if configured, touch the TOOLS icon
followed by the [SEND] button.
NOTE: If an empty file is selected when the PRINT command is made, a blank
report will be printed.
Deleting QC Results
Each of the six control files stores up to 100 results.
The contents of a QC file are automatically deleted when certain fields of the file
(expiration date, lot number, or limits) are edited.
Changing any field within a QC file by using the [EDIT] button will cause the
system to prompt the user, DO YOU WANT TO SAVE MODIFICATIONS? If a
modification has been made to the expiration date, lot number, or limits a second
prompt will appear if the user has responded to DO YOU WANT TO SAVE
MODIFICATIONS? with [YES]. The second prompt is a warning message, YOU
ARE GOING TO DELETE ALL ASSOCIATED RESULTS. DO YOU WANT
TO CONTINUE? Responding [YES] to this prompt will delete all results within
this QC file.
11-12
9140855CDecember 2009
Quality Control
Section 11
QC Guidelines
9140855CDecember 2009
11-13
Quality Control
Section 11
QC Guidelines
Saving QC Results
QC Results can be saved to a USB drive. For additional information, refer to
Section 5: Operating Instructions, Subsection: Saving Current QC Data.
Viewing Results in the QC Menu
QC runs are added to the display in the RUN and RESULT screen at the bottom of
the table. It is only possible to view a report with histograms for a QC run
immediately after it has been tested and before any other samples have been tested.
Due to the size of the display, it is necessary to utilize scroll bars to view all
measurands and results.
NOTE: To view date/time (hour) of QC runs, use the right/left scroll bar to scroll
to the far right.
Viewing all QC Runs Within a File
The scroll bar at the right side of the display allows the operator to scroll up and
down through multiple QC runs, with the oldest runs at the top of the table and the
newest results at the bottom.
Viewing all Measurands Within the QC File
The scroll bar at the bottom of the result table is used to scroll right and left to view
results for all measurands.
Viewing QC Statistics
Statistics associated with the data in a QC file are displayed at the bottom of the
QC file screen.
11-14
9140855CDecember 2009
Quality Control
Section 11
QC Guidelines
The statistics displayed represent calculations based on the selected QC runs only.
Thus, if no QC runs are selected, all values will be 0. The mean (MEAN), standard
deviation (SD), and coefficient of variation percentage (CV%) may be rounded in
this table due to available display space.
9140855CDecember 2009
11-15
Quality Control
Section 11
QC Guidelines
Levey-Jennings Graphs
Results are automatically plotted on Levey-Jennings graphs as shown in the
following figure.
The Levey-Jennings (LJ) graphs are
displayed by touching [LJ] on the Control
Results screen.
The left column shows each
measurand with the target values and
limits shown on the right side next to
the plots.
The upper and lower dashed lines
represent the LIMITS entered during
QC FILE Setup. The solid lines
represent two times the LIMITS
entered during QC FILE Setup.
The number displayed directly under
each measurand is the result for the
run selected by the vertical line on the
graphs. To move the line to another
sample, use the [<] and [>] buttons to
the left of the time displayed at the
bottom of the screen.
The scroll bar on the bottom of the
screen moves the cursor to display
select additional measurands.
The arrow keys displayed under the
scroll bar can also be used to move the
cursor to a specific measurand.
The number of the result at the cursor
position and the date and time it was
run are shown in the box on the bottom
of the screen.
Touch [>>] or [<<] to scroll through all
results.
The date and time displayed at the
bottom of the screen correspond to the
sample indicated by the vertical line on
the graph. To move the line to another
sample, use the [<] and [>] buttons to
the left of the time displayed at the
bottom of the screen.
Only results that have been selected in
the RUN/RESULTS screen are plotted
on the L-J Graphs.
Figure 11.8 Levey-Jennings Screen
11-16
9140855CDecember 2009
Section 11
Quality Control
References
9140855CDecember 2009
11-17
Quality Control
Section 11
References
NOTES
11-18
9140855CDecember 2009
Appendix A Accessories
Appendix A
Accessories
This section provides the list numbers for components, accessories, reagents,
controls, and consumables used with the CELL-DYN Emerald System.
To place an order for the products listed or to obtain technical assistance, USA
customers contact Abbott Diagnostics Customer Service at:
1-877-4ABBOTT (1-877-422-2688)
For customer support in Canada, call:
1-800-387-8378
For customer support outside the USA and Canada, call your local Hematology
Customer Support representative.
9140861CDecember 2009
A-1
Accessories
Appendix A
NOTES
A-2
9140861CDecember 2009
Appendix A
Accessories
Accessories
List numbers are unique identifiers used when ordering products. List numbers and
quantities provided in this section are intended for guidance only and are subject to
change. USA customers, contact Abbott Diagnostics Customer Service at
1-877-4ABBOTT (1-877-422-2688) for the most current information regarding
list numbers. Customers outside the USA, contact your local Abbott Customer
Service Representative for assistance.
Table A.1
Accessories
List Number
Quantity
Name
09H40-01
CELL-DYN Emerald
Operators Manual, printed
09H52-01
25860-01
09H58-01
Emerald Grease
09H54-06
09H54-02
09H54-04
Flat Screwdriver/Key
09H50-01
09H51-01
Table A.2
Comments
Consumables
List Number
Quantity
Name
Comments
03B96-02
500/pkg
Printer paper
8.5 x 11 sheets
99650-01
1 roll
9140861CDecember 2009
A-3
Accessories
Appendix A
Accessories
Table A.3
Reagents
List Number
Quantity
Name
09H48-02
10L cubitainer
09H47-02
960 mL Bottle
09H46-02
960 mL Bottle
Table A.4
Comments
List Number
Quantity
Name
Comments
09H70-01
2 x 2.5 mL tubes
09H69-01
12 x 2.5 mL tubes
Tri-level
09H69-02
6 x 2.5 mL tubes
A-4
9140861CDecember 2009
Appendix B
Measurand
Cryoglobulin, cryofibrinogen
Heparin
Monoclonal proteins
Nucleated red cells
Platelet clumping
Unlysed red cells
Clotting
Smudge cells
Uremia plus immunosuppressants
Cryoglobulin, cryofibrinogen
Giant platelets
High white cell count (>50,000 K/L)
Autoagglutination
Clotting
Hemolysis (in vitro)
Microcytic red cells
Hemoglobin (HGB)
Carboxyhemoglobin (>10%)
Cryoglobulin, cryofibrinogen
Hemolysis (in vivo)
Heparin
High white cell count (>50,000 K/L)
Hyperbilirubinemia
Lipemia
Monoclonal proteins
Clotting
Hematocrit
(Automated) (HCT)
Cryoglobulin, cryofibrinogen
Giant platelets
High white cell count (>50,000 K/L)
Hyperglycemia (>600 mg/dL)
Autoagglutination
Clotting
Hemolysis (in vitro)
Microcytic red cells
Hematocrit (HCT)
(Microhematocrit)
Hyponatremia
Plasma trapping
Excess EDTA
Hemolysis (in vitro)
Hypernatremia
9140861CDecember 2009
B-1
Table B.1
Measurand
Autoagglutination
High white cell count (>50,000 K/L)
Hyperglycemia
Reduced red cell deformability
Swollen red cells
Cryoglobulin, cryofibrinogen
Giant platelets
Hemolysis (in vitro)
Microcytic red cells
Mean Cell
Hemoglobin (MCH)
Mean Cell
Hemoglobin
Concentration
(MCHC)
Autoagglutination
Clotting
Hemolysis (in vivo and in vitro)
Spuriously high hemoglobin
Spuriously low hematocrit
Platelets (PLT)
Cryoglobulin, cryofibrinogen
Hemolysis (in vivo and in vitro)
Microcytic red cells
Red cell inclusions
White cell fragments
Clotting
Giant platelets
Heparin
Platelet clumping
Platelet satellitosis
B-2
9140861CDecember 2009
Appendix C
X = 0.46
Add 0.46 parts deionized water to one part bleach to obtain the 3.6% sodium
hypochlorite solution. For example, add 0.46 mL of deionized water to 1.0 mL of
the 5.25% sodium hypochlorite solution (bleach) to obtain the 3.6% solution. Final
volume in this example is 1.46 mL.
NOTE: In the US, commercially available sodium hypochlorite is often
approximately 5% to 5.25%. Read the container labeling to verify the
concentration.
9140861CDecember 2009
C-1
Appendix C
NOTES
C-2
9140861CDecember 2009
Appendix D
Reference Intervals
9140861CDecember 2009
D-1
Reference Intervals
CBC Reference Intervals: Literature Sources
Appendix D
20. Kratz A, et al. Effect of marathon running on hematologic and biochemical laboratory parameters,
including cardiac markers. Am J Clin Pathol. 2002;118:856-863.
21. Pfaeffli J. Reference limits for the automated haematology analyser Sysmex XE-2100. Sysmex J Intl.
2002;12:18-23.
22. Hanson CA. Peripheral blood and bone marrow: morphology, counts and differentials, and reactive
disorders, In Clinical Laboratory Medicine, 2nd ed, KD McClatchey, ed. New York: Lippincott Williams
& Wilkins, 2002:807-815.
23. Bock BJ, et al. The data warehouse as a foundation for population-based reference intervals. Am J Clin
Pathol. 2003;120:662-670.
24. Nordin G, et al. A multicentre study of reference intervals for haemoglobin, basic blood cell counts and
erythrocyte indices in the adult population of the Nordic countries. Scand J Clin Lab Invest.
2004;64:385-398.
25. Bertholf RL. Statistical methods for establishing and validation reference intervals. Lab Med.
2006;37:306-310.
26. Mandisodza AR, et al. Haematological reference ranges for adults in Zimbabwe. Sysmex J Intl.
2006;16(suppl 1):38-41.
D-2
9140861CDecember 2009
Appendix E
9140861CDecember 2009
E-1
Worksheet E.1
Pre-Calibration Procedure Checklist
If problems are detected during these checks, DO NOT attempt to calibrate the instrument. Consult Section 10:
Troubleshooting, for assistance in resolving the issue. When problems are resolved, repeat these procedures to
verify proper instrument performance.
Date: _____________________________ Operator: _____________________________
Initial each item after it is completed.
1. _____Ensure all maintenance is complete before calibrating the instrument. Refer to
Section 9: Service and Maintenance for further information.
2. _____Confirm reagent containers are at least half full --- replace them as necessary.
3. _____Confirm the waste container is at least half empty --- replace it as necessary.
4. _____Verify all reagents have not expired. Record the lot numbers and expiration dates:
Diluent Reagent:
Lot # _____________ Exp. Date ______________
CN-Free Lyse Reagent: Lot # _____________ Exp. Date ______________
Cleaner Reagent:
Lot # _____________ Exp. Date ______________
5. _____Verify the calibrator is not expired. Record the lot number and expiration date:
Lot # ____________
Exp. Date _____________
6. _____Confirm the Background Counts are within acceptable limits. Refer to Section 4: Performance
Characteristics and Specifications for Background Count specifications. Record results below or
attach a printout to this document.
WBC
RBC
HGB
PLT
7. _____Verify instrument precision by running a fresh, normal whole blood specimen ten times into the
PRECISION file. Refer to Section 9: Service and Maintenance, Subsection: Precision for
information on using QC files. Ensure that CV% results are within the limits as provided in Section
4: Performance Characteristics and Specifications. Record the results below or attach a printout to
this document.
NOTE: You may also use the PRECISION function within the MAINTENANCE menu to run precision. N=10.
Parameter
WBC
RBC
HGB
MCV
PLT
%CV Limit
3.5%
2.0%
2.1%
0.8%
6.1%
%CV
8. _____If any problems are detected during these checks, document the problem and corrective action below:
________________________________________________________________
________________________________________________________________
________________________________________________________________
________________________________________________________________
E-2
9140861CDecember 2009
9140861CDecember 2009
Lubricate
the Pistons
SemiAnnually
As Needed
Day
Preparation
for Storage,
Relocation
or Shipping
Un-scheduled Bleach
Cleaning
Bleach
Cleaning
Day
Month:
Maintenance Log
Monthly
Table E.1
10
10
11
11
12
12
13
13
14
14
15
15
16
16
17
17
18
18
19
19
20
20
21
21
22
22
23
23
24
24
25
25
26
26
27
27
28
28
29
29
30
30
31
31
Appendix E
WARNING: Potential Biohazard. Consider all clinical specimens, reagents, calibrators, controls or other
materials and all system surfaces or components that contain or come in contact with human-sourced material
as potentially infectious. Wear a lab coat, gloves and safety glasses when performing service or maintenance.
Follow biosafety practices as specified in the OSHA Bloodborne Pathogen rule (29 CFR Part 1910.1030)1 or
other equivalent biosafety practices.2, 3 Refer to Section 8: Hazards for additional information.
Year:
________________________________________________________________
E-3
E-4
Date
Operator
Name
Control
Lot No.
Troubleshooting Log
Table E.2
Diluent
Lot No.
Lyse Lot
No.
Cleaner
Lot No.
Troubleshooting Log
Observation
Resolution
Appendix E
9140861CDecember 2009
Table E-3
Open Date
9140861CDecember 2009
Expiration Date
Operator ID
E-5
Table E-4
E-6
Open Date
Expiration Date
Operator ID
9140861CDecember 2009
Table E-5
Open Date
9140861CDecember 2009
Expiration Date
Operator ID
E-7
Worksheet E.2
Calibration Verification Worksheet
Date: _____________________________
Operator: _____________________________
Calibrator Lot # ____________________
Exp. Date: _____________________________
Calibrator
Assay
Value
Mean
Value
Difference
+/- Limit
Calibration
Needed (Y/N)
WBC
RBC
HGB
MCV
PLT
E-8
9140861CDecember 2009
Worksheet E.3
Training Checklist
TOPIC
Component and Features
Analyzer
Display and Numeric Keypads
Printer
Reagent
Barcode Reader
AC Adaptor/Transformer
Analyzers Software
Main Menu
Software Navigation
Basic Operation
Instrument Overview
Power On
Operators Login
Operators Passwords
Loading/Replacing Reagents
Replacing Waste Containers
Start-up
Running Background
Running Controls
Specimen Identification
Running Specimens
Shutdown
Daily Shutdown
9140861CDecember 2009
TOPIC
Setup
Header
Date and Time
Languages
Printer
Paper Size
Date Format
Auto Clean Interval
Auto Shutdown
Patient Limits
Reagent Log
Start-Up Alarms
QC Alarms
QC Setup
Data Interpretation
QC Results Interpretation
Accuracy and Precision
QC Statistics
Patient Result Interpretation
Patient Report Overview
Datalog
TOPIC
Calibration
When to Calibrate
Calibration Materials
Pre-calibration Checklist
Calibration
Post Calibration Procedure
Maintenance
Monthly
Semi-Annual
As-Needed
Component replacement
Quality Control
Review QC Data
QC Log Editing
Printing QC Log and
LEVEY-JENNINGS
External QC Program (eQC)
Troubleshooting
Basic
Abnormal/erratic RBC, WBC,
HGB, MCV and PLT
Background data
unacceptable
Erratic Results/Multiple
Parameters
Interfering substances
Abbott Customer Service
1-877-4ABBOTT
E-9
E-10
2008 Abbott Laboratories
NOTE: Completion of this checklist indicates the trainee has successfully completed training
on the listed topics. This document can be used for regulatory review and as
documentation of continuing education.
PLEASE PRINT
Appendix E
9140861CDecember 2009
Worksheet E.4
Reference Mean
x Current Calibration Factor = New Calibration Factor
CELL-DYN Emerald Mean
For example, if the Reference Mean Value for WBC is 6.6, the
CELL-DYN Emerald Mean is 7.1, and the current Calibration Factor is 0.98, then:
Reference
Mean
Current
Calibration
Factor
New
Calibration
Factor
Range*
WBC
0.50-2.0
RBC
0.50-2.0
HGB
0.50-2.0
MCV
0.50-2.0
PLT
0.50-2.0
9140861CDecember 2009
E-11
NOTES
E-12
9140861CDecember 2009
Appendix F
9140861CDecember 2009
F-1
Appendix F
F-2
9140861CDecember 2009
NOTE: Reference microhematocrit values can be determined using the CLSI method
for Packed Cell Volume (PCV). Use only plain, non-anticoagulated capillary
tubes. Be certain to verify proper operation of the microhematocrit centrifuge
and the timer as recommended by CLSI.
9140861CDecember 2009
F-3
Appendix F
F-4
9140861CDecember 2009
9140861CDecember 2009
F-5
Appendix F
F-6
9140861CDecember 2009
Touch [OK].
Daily Start Up
1. From the MAIN menu, touch [START UP].
At the end of the Startup cycle, a background count is automatically displayed.
2. The Background must be within specifications before proceeding to the next step. (See Table A)
NOTE: Refer to the CELL-DYN Emerald Operators Manual Section 2: Installation Procedures and Special
Requirements to set up autoprint of startup background.
3.
4.
5.
6.
Specification
WBC
RBC
HGB
PLT
9140863CDecember 2009
9140863CDecember 2009
Section 1 Glossary
Glossary
Glossary
NOTE: The references provided in this Glossary, acquired from assorted
reference works, may have been revised to reflect their meanings in
relation to functions and operations of the CELL-DYN Emerald
instrument.
absorption
accuracy
agglutination
agglutinin
aggregation
See agglutination.
algorithm
aliquot
amplitude of a pulse
analyzer
The module of the CELL-DYN Emerald through which the blood sample
passes and where measurements are made.
anticoagulant
aperture
assay
9140872CDecember 2009
Glossary-1
Glossary
background
band
basophil
blast
blood, whole
A homogenous mixture of blood that has not been separated into cellular
and liquid components.
calibration
calibration factor
calibrator
carboxyhemoglobin
carryover
CBC
Acronym for complete blood count. Includes WBC, RBC, HGB, HCT,
MCV, MCH, MCHC, and PLT. On the CELL-DYN Emerald, the CBC
also includes an automated three-part differential.
character self-checking
Glossary-2
9140872CDecember 2009
Glossary
check character/digit
An alphanumeric character that allows the decoder in the bar code reader
to mathematically determine that it has read the bar code characters
correctly. It immediately precedes the stop character in the bar code
symbol.
CLIA
CLSI
cold agglutinin
cryoglobulin
concentration
control
correlation coefficient
datalog
diff
9140872CDecember 2009
Glossary-3
Glossary
dilution ratio
drift
EDTA
eosinophil
error, random
error, systematic
erythroblast
erythrocyte
false negative
false positive
flag
gain
Gaussian
GRA
Glossary-4
9140872CDecember 2009
Glossary
G%
granulocyte
HCT
heme
hemoglobin
hemolysis
heparin
HGB
histogram
hyperglycemia
ICSH
ID
ID, operator
ID, specimen
9140872CDecember 2009
Glossary-5
Glossary
immature granulocyte
impedance method
imprecision
inaccuracy
indices, RBC
A group of calculated values for red blood cell properties: mean cell
volume (MCV), mean cell hemoglobin (MCH), and mean cell
hemoglobin concentration (MCHC).
interfering
substance/condition
in vitro
in vivo
leukocyte
Levey-Jennings graph
linearity
LIS
logarithm
LYM
Glossary-6
9140872CDecember 2009
Glossary
lymphocyte
lymphocyte, variant
L%
lysis
macrocytic RBC
MCH
The identifier for mean cell hemoglobin, the average hemoglobin mass in
the red blood cells.
MCHC
The identifier for mean cell hemoglobin concentration, the average mass
of hemoglobin per unit volume in the red blood cells.
MCV
The identifier for mean cell volume, the average volume of the red blood
cells.
metamyelocyte
A cell present in the bone marrow that gives rise to a granulocyte and is
not normally present in circulation. The maturation phase between
myelocyte and band. See also immature granulocyte.
method, reference
microcytic platelet
microcytic RBC
monocyte
mononuclear
9140872CDecember 2009
Glossary-7
Glossary
MPV
The identifier for the mean platelet volume result, the average volume of
the platelets.
myelocyte
A cell in bone marrow that gives rise to a granulocyte and is not normally
in circulation. Considered an immature granulocyte. See also immature
granulocyte.
NCCLS
neutrophil
NRBC
A nucleated cell present in the bone marrow that gives rise to a mature
red blood cell and is not normally present in circulation. Subtypes
include erythroblasts, basophilic normoblasts, polychromatophilic
normoblasts, and orthochromic normoblasts. See also NRBC.
nucleus
measurand
plasma
platelet
PLT
precision
promyelocyte
A cell present in the bone marrow that gives rise to a granulocyte and is
not normally present in circulation. See also immature granulocyte.
Glossary-8
9140872CDecember 2009
Glossary
QC file
quiet zone
The quiet zone is the area immediately adjacent to the beginning and end
of the bar code symbol. No writing or other printed material should be
present in the quiet zone.
range
RBC
An acronym for red blood cell. On the CELL-DYN Emerald, RBC is used
as the identifier for the red blood cell absolute concentration result,
calculated as the number of red blood cells per unit volume of whole blood.
RDW
reagent
A solution used to dilute, and in some cases, alter the cells in a whole
blood specimen, in preparation for measurement by the Analyzer. May
also refer to a substance used to clean the instrument fluidics.
reticulocyte
9140872CDecember 2009
Glossary-9
Glossary
RNA
Acronym for ribonucleic acid. Any of the various nucleic acids that
contain ribose and uracil as structural components. Associated with the
control of cellular chemical activities, such as protein synthesis.
sample
sample dilution
sequence number
shift
specimen
stability
Bar Code Readers use these characters to determine the order in which to
read the characters in the code and identify the bar code symbology. Start
and stop characters make it possible to scan the label in either direction
and are located at the beginning and end of each bar code symbol.
test
thrombocyte
See platelet.
throughput
trend
Glossary-10
9140872CDecember 2009
Glossary
unit of measure
verification
WBC
The identifier for the white blood cell absolute concentration result,
which is calculated as the number of white blood cells per unit volume of
whole blood.
Westgard Rules
9140872CDecember 2009
Glossary-11
Glossary
NOTES
Glossary-12
9140872CDecember 2009
Glossary
References
1.
2.
3.
4.
9140872CDecember 2009
Glossary-13
Glossary
NOTES
Glossary-14
9140872CDecember 2009
Index
Index
A
B
Background Count 2-34
Bar Code Reader 2-25
Bleach Cleaning 9-11
C
Calibration
Guidelines 6-5
Reference Methods F-2
When to Calibrate 6-3
Whole Blood F-1, G-1
Calibration Factors 2-29
Calibration Menu 6-5
Calibrator 1-12
Calibrator and Controls A-4
Cleaner 1-11
Cleaning the cover 9-15
COMMENTS 2-22
Communication 2-23
Components 1-3
Connections 2-8
Controls 1-12
Count Invalidating Flags 3-17
Counting Chambers 1-8
CURRENT SID 2-20
9140862CDecember 2009
E
Emergency Stop 9-14
Entering QC Assay Values and Limits 11-7
Ethernet Connection 2-8
Event Log 9-3
Printing 9-4
Saving 9-4
Expected Values & Patient Limits 2-22
External Bar Code Reader 2-8
F
Flags
Printing 2-22
Fluidics 1-5
Fluidics Menu 9-9
G
GRAPHS 2-22
Ground 2-8
Index-1
Index
H
Hazards 8-1
Biological 8-4
Biological & Chemical 8-4
Chemical 8-7
Electrical 8-8
Icons, Safety 8-3
Mechanical 8-9
Physical 8-10
Safety Icons 8-3
HCT 1-2
HEADER 2-22
Hemoglobin measurands 1-2
Hemoglobin Measurement 3-5
HGB 1-2
I
Icon, Safety 8-3
Impedance 3-2
Initialize System 9-14
Installation 2-5
Installation Environment 2-4
Instrument Alarms 3-9
Instrument Disclaimer x
Instrument Logbook 5-3
Interfering Substances B-1
INTERP. RPT 2-22
Interpretive Messages 3-20
Interpretive Report 2-22
K
Keypad 1-3
L
Lab Preferences 2-26
LAB TECH PASSWORD 2-30
Language 2-19
Language Selection 2-12
Levey-Jennings 11-16
Limits 2-27
Linearity 3-10
Index-2
M
Main Menu 1-4
Maintenance
As Needed 9-14
Cover Cleaning 9-15
Monthly 9-11
Piston Lubrication 9-13
Manual Calibration Factor Entry Procedure F-4
MCH 1-2
MCHC 1-2
MCV 1-2
Measurement Principles 3-2
Monthly Maintenance 9-11
MPV 1-2
O
Other Settings 2-30
P
Panic Limits 3-10
Paper size 2-19
Patient Limits 3-10
Platelet Analysis 3-4
Platelet Flags 3-15
Platelet measurands 1-2
PLT 1-2
power 2-8
Power cord connection 2-8
Power On 5-5
Power Requirements 2-4
Pre-Calibration Procedures 6-5
Precautions and Requirements for System
Operation 7-4
Precautions See Operational Precautions and
Limitations 8-1
Preparation for Shipment 8-8
CELL-DYN Emerald Operators Manual
9140862CDecember 2009
Index
Q
QC
eQC 5-26
Printing 11-12
Saving 5-26
Tools 11-12
QC ALARMS 2-24
QC not done 2-24
QC Send
Results 11-12
Quality Control
Entering Assay Values and Limits 11-7
File Setup 11-4
Levey-Jennings 11-16
Materials 11-9
Running QC 11-10
Selecting Files 11-4
Quality Control Methods and Materials 11-3
R
RBC 1-2
RBC Analysis 3-3
RDW 1-2
Reagent 8-7
installation 2-10
Reagent Bar Codes 2-15
Reagent Menu 2-18
Reagent Storage 1-11
Reagents 1-10, 2-14, 2-15, A-4
Manual Entry 2-16
9140862CDecember 2009
S
Safety i
Icons 8-3
Safety Icons 8-3
Sample Analysis Cycle 3-1
Sample Logs E-1
Sample Probe
Removal and Replacement 9-18
Sample Probe Removal and Replacement 9-18
Saving Current QC Data 5-26
Sending QC Results 11-12
SET AUTOCLEAN 2-30
Set Up Menu 2-19
Shipping 9-15
Shut Down 5-16
Shut down
Automatic 2-30
Site Requirements 7-3
Software Menu 5-4
Space Requirements 2-3
Specimen Analysis 5-5
SPECIMEN CAL. ADJUSTMENTS 2-27
Specimen Collection and Handling 5-5
Specimen Identification 5-6
Index-3
Index
USB
Backup 5-30
Restore Settings 5-31
Set Up 5-30
USB drive, Event Log
Event Log 5-20
USB Flash Drive 5-17
USB Port 2-8
Use of the USB drive in the DATALOG 5-29
technical assistance ix
Thresholds 2-27
Thumb Drive 5-17
Tools in the QC Menu 11-12
Trademarks xiii
Troubleshooting 10-3
Results 10-5
Troublshooting
QC 10-5
Type 2-26
W
Waste 2-17
Waste Disposal Requirements 2-4, 8-7
Waste Handling and Disposal 8-7
WBC 1-2
WBC Analysis 3-2
WBC Flags 3-13
White Blood Cell measurands 1-2
Worksheets E-1
UNITS 2-22
Units 2-24, 3-7
units of measurement 3-6
Index-4
9140862CDecember 2009