Deltapilot M

Download as pdf or txt
Download as pdf or txt
You are on page 1of 72

TI00437P/00/EN/19.

14
No. 71260308

Products

Solutions

Services

Technical Information
Deltapilot M
FMB50, FMB51, FMB52,
FMB53
Hydrostatic level measurement

Pressure sensor with the CONTITETM measuring cell


Condensate-resistant
Application
The device is used for the following measuring tasks:
Hydrostatic pressure measurement in liquids and paste-like media in all areas of
process engineering, process measuring technology, pharmaceuticals and the food
industry
Level, volume or mass measurements in liquids
Your benefits
Very good reproducibility and long-term stability
Maximum plant safety provided by one-of-a-kind, condensate-proofed CONTITE
measuring cell
High reference accuracy: 0.2 %
optionally 0.1 %
Turn down 100:1
Standardized platform for differential pressure, hydrostatics, and pressure
(Deltabar S Deltapilot S Cerabar S)
Simple, fast commissioning through a user interface designed for real-world
applications
Used for process pressure monitoring up to SIL2, certified to IEC 61508 Edition 2.0
and IEC 61511 by TV NORD
Usage in drinking water: KTW, NSF

Deltapilot M FMB50/51/52/53

Table of contents
Document information . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Terms and abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Total error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Warm-up period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Function and system design . . . . . . . . . . . . . . . . . . . . . .7
Device selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Level measurement in closed tanks with pressure overlay . . 10
Density measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Level measurement with automatic density correction (with
media changing in the tank) . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Electrical differential pressure measurement with gauge
pressure sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Communication and data processing . . . . . . . . . . . . . . . . . . . . 12

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Signal range 4 to 20 mA HART . . . . . . . . . . . . . . . . . . . . . . .
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Load - 4 to 20 mA HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dead time, time constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dynamic behavior: current output . . . . . . . . . . . . . . . . . . . . . . .
Dynamic behavior: HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dynamic behavior: PROFIBUS PA . . . . . . . . . . . . . . . . . . . . . . .
Dynamic behavior: FOUNDATION Fieldbus . . . . . . . . . . . . . . .
Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Firmware version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

14
14
14
14
15
15
15
16
16
16
16
17
17

Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Start-up current HART. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Residual ripple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Influence of power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overvoltage protection (optional) . . . . . . . . . . . . . . . . . . . . . . .

21
21
22
22
22
22
23
24
24
24
24
25

Performance characteristics . . . . . . . . . . . . . . . . . . . . 26
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . .
Influence of orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reference accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Thermal change in the zero output and the output span . . . .
Total performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Long-term stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

26
26
26
26
26
27
27
27

General installation instructions . . . . . . . . . . . . . . . . . . . . . . . .


FMB50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
FMB51/FMB52/FMB53 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Supplementary installation instructions . . . . . . . . . . . . . . . . . .
Wall and pipe mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
"Separate housing" version . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Oxygen applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
PWIS cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Applications with hydrogen . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Special measuring cells for acids, alkalis or sea water
(not FMB50) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

28
28
28
29
29
30
31
31
31
31

Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . . . . . . .
Storage temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . .

32
32
32
32
32
32

Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Process temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Lateral load FMB51 (static) . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Pressure specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . 34
Device height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
F31 housing, aluminum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
F15 housing, stainless steel (hygienic) . . . . . . . . . . . . . . . . . . . 35
Process connections FMB50, FMB51, FMB52 . . . . . . . . . . . . 35
Process connections FMB50, FMB51, FMB52 . . . . . . . . . . . . 36
Process connections FMB50, FMB51, FMB52 - continued . . 37
Process connections FMB50, FMB51, FMB52 - continued . . 38
Process connections FMB50 - continued . . . . . . . . . . . . . . . . . 39
Process connections FMB50 - continued . . . . . . . . . . . . . . . . 40
Process connections FMB51 (rod version) . . . . . . . . . . . . . . . . 42
Process connections FMB52 (cable version) . . . . . . . . . . . . . . 44
Dimensions of FMB53 with F31 housing, suspension clamp and
mounting bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Dimensions of FMB53 with F15 housing, suspension clamp and
mounting bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Wall and pipe mounting with "Separate housing" version . . . 48
Materials not in contact with process . . . . . . . . . . . . . . . . . . . . 51
Materials in contact with the process . . . . . . . . . . . . . . . . . . . . 54

Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Onsite operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating languages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
System integration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

57
57
59
59
60

Endress+Hauser

Deltapilot M FMB50/51/52/53

Certificates and approvals . . . . . . . . . . . . . . . . . . . . . . 61


CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
C-tick symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Ex approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Suitability for hygienic processes . . . . . . . . . . . . . . . . . . . . . . . . 61
Pharma (CoC) (FMB50 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Functional safety SIL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Overfill protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
AD2000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Pressure Equipment Directive (PED) . . . . . . . . . . . . . . . . . . . . . 62
Marine approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . 62
Drinking water approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Classification of process sealing between electrical systems and
(flammable or combustible) process fluids in accordance with
ANSI/ISA 12.27.01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Inspection certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Calibration; unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Configuration data sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Suspension clamp (FMB53 only) . . . . . . . . . . . . . . . . . . . . . . . .
Extension cable shortening kit (FMB53 only) . . . . . . . . . . . . .
M12 connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Welding necks and Weld-in tool flanges . . . . . . . . . . . . . . . . .
Adapter Uni for FMB50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

66
66
66
66
67

Documentation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Technical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Brief Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functional safety manual (SIL) . . . . . . . . . . . . . . . . . . . . . . . . .
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation/Control Drawings . . . . . . . . . . . . . . . . . . . . . . . . . .
Combination certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overfill protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

68
68
68
68
68
69
69
69

Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . . 70
HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
PROFIBUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
FOUNDATION Fieldbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

Endress+Hauser

Deltapilot M FMB50/51/52/53

Document information
Symbols used

Safety symbols
Symbol

Meaning

DANGER
A0011189-EN

WARNING
A0011190-EN

CAUTION
A0011191-EN

NOTICE
A0011192-EN

DANGER!
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTE!
This symbol contains information on procedures and other facts which do not result in
personal injury.

Electrical symbols
Symbol

Meaning
Direct current
A terminal at which DC voltage is present or through which direct current flows.
A0018335

A0018336

Alternating current
A terminal at which alternating voltage is present or through which alternating current
flows.

A0018337

Direct current and alternating current


A terminal at which alternating voltage or DC voltage is present.
A terminal through which alternating current or direct current flows.

A0018338

Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Protective ground connection
A terminal which must be connected to ground prior to establishing any other connections.

A0018339

A0011201

Equipotential connection
A connection that has to be connected to the plant grounding system: This may be a
potential equalization line or a star grounding system depending on national or company
codes of practice.

Tool symbols
Symbol

Meaning
Phillips head screwdriver
A0011219

Flat-blade screwdriver
A0011220

Torx screwdriver
A0013442

Hexagon wrench
A0011222

Allen screw
A0011221

Endress+Hauser

Deltapilot M FMB50/51/52/53

Symbols for certain types of information


Symbol

Meaning
Permitted
Indicates procedures, processes or actions that are permitted.
A0011182

Preferred
Indicates procedures, processes or actions that are preferred.
A0011183

Forbidden
Indicates procedures, processes or actions that are forbidden.
A0011184

Tip
Indicates additional information.
A0011193

Reference to documentation
Refers to the corresponding device documentation.
A0015483

Reference to page
Refers to the corresponding page number.
A0015484

Reference to graphics
Refers to the corresponding graphic number and page number.
A0015486

, ...

Series of steps
Help in the event of a problem

A0015488

Symbols in graphics
Symbol

Meaning

1, 2, 3, 4 etc.
,

Series of steps

A, B, C, D etc.

Views

A-A, B-B, etc.

Sections

Endress+Hauser

, ...

Numbering for main items

Hazardous area
Indicates the hazardous area.
A0011187

Safe area (non-hazardous area)


Indicates the non-hazardous area.
A0011188

Deltapilot M FMB50/51/52/53

Terms and abbreviations

Term/abbreviation

Explanation

MWP

The MWP (maximum working pressure) for the individual


sensors depends on the lowest-rated element, with regard to
pressure, of the selected components, i.e. the process
connection has to be taken into consideration in addition to
the measuring cell. Pay attention to the pressure-temperature
dependence also. For the relevant standards and additional
notes, see section " 33".

OPL

The OPL (over pressure limit = sensor overload limit) for the
sensor depends on the lowest-rated element, with regard to
pressure, of the selected components, i.e. the process
connection has to be taken into consideration in addition to
the measuring cell. Pay attention to the pressure-temperature
dependence also. For the relevant standards and additional
notes, see section " 33".

LRL

Lower range limit

URL

Upper range limit

LRV

Lower range value

URV

Upper range value

TD

Turn down

Case 1:
Lower range value (LRV) Upper range
value (URV)

1=2
LRL

Example:
Lower range value (LRV) = 0 mbar
Upper range value (URV) = 40 mbar (0.6 psi)
Nominal value (URL) = 400 mbar (6 psi)

LRV

URV

URL

Turn down:
TD = URL /URV:1

Set span:
URV LRV = 40 mbar (0.6 psi)
This span is based on the zero point.

5
A0019783

Example: 400 mbar (6 psi) measuring cell

Case 2:
Lower range value (LRV) Upper range
value (URV)

1=2
LRL

Example:
Lower range value (LRV) = 200 mbar (3 psi)
Upper range value (URV) = 0 bar
Nominal value (URL) = 400 mbar (6 psi)

LRV

URV

URL

Turn down:
TD = URL /(LRV) = 2:1

Set span:
URV LRV = 200 mbar (3 psi)
This span is based on the zero point.

5
A0016451

Example: 400 mbar (6 psi) measuring cell


1
2
3
4
5

Set span
Span based on zero point
Nominal value upper range limit (URL)
Nominal measuring range
Sensor measuring range

Endress+Hauser

Deltapilot M FMB50/51/52/53

Function and system design


Device selection
Deltapilot M
Product family

FMB51

FMB50

A0023537

Compact version

FMB52

A0023538

Rod version

FMB53

A0023539

Cable version

A0023540

Cable version with


Mounting bracket

Field of application

Level measurement
Pressure measurement

Industries

Food, pharmaceuticals, chemicals, power plants, automotive, etc.

Process connections

Thread
Flanges
Flush-mounted hygienic
connections

Measuring ranges

From 0.1 to +0.1 (-1.5 to +1.5 psi) bar to 1 bar to +10 bar (-15 to 150 psi)

OPL

Max. 40 bar (600 psi)

Process temperature range

10 to +100 C (+14 to
+212F) (+135 C (275F) for
30 minutes maximum)

Ambient temperature range

Without LCD display: -40 to +85 C (-40 to +185F)


With LCD display: -20 to +70 C (-4 to +158F) (extended temperature application range -40 to 85 C
(-40 to +158F) with restrictions in optical properties such as display speed and contrast)
Separate housing: -20 to +60 C (-4 to +140F)

Reference accuracy

0.2 % (option 0.1 %) depends on the measuring range 26

Supply voltage

11.5 to 45 V DC (versions with plug-in connection 35 V DC)


For intrinsically safe device versions: 11.5 to 30 V DC

Output

4 to 20 mA with superimposed HART protocol, PROFIBUS PA or FOUNDATION Fieldbus

Options

Specialties

Absolute resistance to condensate thanks to hermetically sealed CONTITETM cell


Maximum flexibility thanks to modular design
Special cleaning of the transmitter to remove paint-wetting substances, for use in paint shops

Endress+Hauser

Thread
Flanges

10 to +85 C
(+14 to +185F)

Thread
Flanges

Environment (freshwater
and wastewater)
Suspension clamp

With PE cable (unmoved): 10 to +70 C (+14 to +158F)


With FEP cable (unmoved):
10 to +80 C (+14 to +176F)

Gold/rhodium-coated process isolating diaphragm


3.1 inspection certificate
3A approval and EHEDG approval for FMB50
Specific firmware versions
Initial device settings can be ordered
Separate housing

Deltapilot M FMB50/51/52/53

FMB50, FMB51, FMB52 universal application


Modular probe program to ensure optimum process adaptation
FMB50 compact version: installation in the tank from below or from the side
FMB51, FMB52 rod and cable extension: installation from above, i.e. easy to retrofit ground tanks,
no additional opening in the vessel floor
FMB50 optimized for hygienic applications

All typical flush-mounted process connections can be supplied


Welding flanges
Stainless steel housing
All the sanitary process connections are gap-free and can be cleaned so that the unit is free of
residue, e.g. CIP cleaning
USDA/H1-approved transfer liquid as per FDA Directive
3A approval or EHEDG approvals

74 -

TYPE EL

February 2010
A0023536

A0020029

FMB53 for level measurement in water and wastewater


The housing with the electronic insert is mounted outside shafts and vessels in such a way that it is
protected from flooding. The extension cable is secured with a suspension clamp.
The measuring cell tube made of stainless steel (AISI 316L) and the Alloy process isolating
diaphragm allow use in aggressive media such as wastewater for example.
Extension cable up to 400 m (1312 ft) in length (up to 100 m (328 ft) in Ex-areas) without strain
relief.
Special measuring cell with gold/rhodium coating for applications in which severe hydrogen
formation can occur (e.g. digested sludge); 31.
Special measuring cell with gold/platinum coating for acids, alkalis or sea water; 31.

Endress+Hauser

Deltapilot M FMB50/51/52/53

Measuring principle

patm

pmess = pges - patm


pmess = (patm + phydr.) - patm

phydr.
h~p
h
h=

pges = patm + phydr.

pges

p
rg

patm

1
p = patm + phydr.

3
patm

A0023541

Deltapilot M hydrostatic level measurement and measuring principle


1
2
3

Rear isolating membrane of the CONTITE measuring cell


Measuring element
Process isolating diaphragm

g
h
p ges
p atm
p hydr.
p mess

Gravitational acceleration
Level height
Total pressure = hydrostatic pressure + atmospheric pressure
Atmospheric pressure
Hydrostatic pressure
Measured pressure in the measuring cell = hydrostatic pressure
Density of the medium

Due to its weight, a liquid column creates hydrostatic pressure. If the density is constant, the
hydrostatic pressure depends solely on the height h of the liquid column.
The CONTITE measuring cell, which works on the principle of the gauge pressure sensor, constitutes
the core of Deltapilot M. In contrast to conventional gauge pressure sensors, the precision measuring
element (2) in the CONTITE measuring cell is absolutely protected, situated between the process
isolating diaphragm (3) and the rear isolating membrane (1). Thanks to this hermetic sealing of the
measuring element, the CONTITE measuring cell is absolutely insensitive to condensate/
condensation and aggressive gases. The pressure applied is transferred from the process isolating
diaphragm to the measuring element by means of an oil without any loss in pressure.
Any measured errors due to fluctuations in temperature are compensated for in the electronics unit by
means of the sensor temperature.
A linearization function with max. 32 points, based on a table entered either manually or
semi-automatically, can be activated locally or remotely. This function facilitates measurement in
engineering units, and provides a linear output signal for spherical and horizontal cylindrical vessels,
and vessels with a conical outlet.

Endress+Hauser

Deltapilot M FMB50/51/52/53

Level measurement in closed


tanks with pressure overlay

You can determine the differential pressure in tanks with pressure overlay using two Deltapilot M. The
pressure measured values of the two probes are sent to a signal processing unit such as
Endress+Hauser RMA or a PLC. The signal processing unit or PLC determines the difference in pressure
and uses this to calculate the level and the density where necessary.

1
A0023542

Level measurement in a closed tank with pressure overlay


1
2
3

Deltapilot 1 measures the total pressure (hydrostatic pressure and top pressure)
Deltapilot 2 measures the top pressure
Signal processing unit determines the difference in pressure and uses this to calculate the level

NOTICE
Measured errors can occur.
Large measured errors can occur if the ratio of the level to the top pressure is >1:6. This does not affect
reproducibility.
When selecting the measuring cell, make sure you select measuring ranges that are sufficiently
wide ( see example).
Example:
Max. hydrostatic pressure = 600 mbar (9 psi)
Max. top pressure (Deltapilot 2) = 300 mbar (4.5 psi)
Max. total pressure, measured with Deltapilot 1 = 300 mbar (4.5 psi) + 600 mbar (9 psi) = 900 mbar
(13.5 psi) measuring cell to be selected: 0 to 1200 mbar (0 to 18 psi)
Max. pressure, measured with Deltapilot 2: 300 mbar (4.5 psi) measuring cell to be selected:
0 to 400 mbar (6 psi)
NOTICE
Possibility of probe 2 flooding during differential pressure measurement.
Measured errors can occur.
When installing, make sure that probe 2 cannot be flooded.

10

Endress+Hauser

Deltapilot M FMB50/51/52/53

Density measurement

You can measure the density in tanks with pressure overlay using two Deltapilot M and a signal
processing unit or a PLC. The signal processing unit or the PLC calculates the density from the known
distance between the two Deltapilot M devices h and the two measured values p1 and p2.

3
p2

p1
1
A0023544

Level measurement in a closed tank with pressure overlay


1
2
3

Level measurement with


automatic density correction
(with media changing in the
tank)

Deltapilot 1 determines pressure measured value p1


Deltapilot 2 determines pressure measured value p2
Signal processing unit determines the density from the two measured values p1 and p2 and the distance h

Level measurement with automatic density correction is possible in conjunction with a limit switch
such as Liquiphant and a PLC. The limit switch always switches at the same level. In the switch point,
the signal processing unit determines the corrected density from the pressure of the Deltapilot M
currently measured and the known distance between Deltapilot M and the limit switch. The signal
processing unit then calculates the level from the new density and the measured pressure of the
Deltapilot M.

1
A0023546

Level measurement with automatic density correction


1
2
3

Endress+Hauser

Deltapilot M
Liquiphant
PLC

11

Deltapilot M FMB50/51/52/53

Electrical differential
pressure measurement with
gauge pressure sensors

FXN 520

Deltapilot M

1
Fieldgate
FXA520

Multidrop-Connector
FXN520

2
A0023549

Shut-off valves

e.g. filter

In the example given, two Deltapilot M devices (each with a gauge pressure sensor) are interconnected.
The pressure difference can thus be measured using two independent Deltapilot M devices.
! WARNING
Risk of explosion!
If using intrinsically safe devices, strict compliance with the rules for interconnecting intrinsically
safe circuits as stipulated in IEC60079-14 (proof of intrinsic safety) is mandatory.

Communication and data


processing

12

4 to 20 mA with HART communication protocol


PROFIBUS PA
The Endress+Hauser devices meet the requirements of the FISCO model.
Due to the low current consumption of 11 mA 1 mA, the following number of devices can be
operated on one bus segment if installing as per FISCO:
up to 8 Deltapilot M for Ex ia, CSA IS and FM IS applications
up to 31 Deltapilot M for all other applications, e.g. in non-hazardous areas, Ex nA, etc.
Further information on PROFIBUS PA can be found in Operating Instructions BA00034S "PROFIBUS
DP/PA: Guidelines for planning and commissioning" and in the PNO Guideline.
FOUNDATION Fieldbus
The Endress+Hauser devices meet the requirements of the FISCO model.
Due to the low current consumption of 16 mA 1 mA, the following number of devices can be
operated on one bus segment if installing as per FISCO:
up to 6 Deltapilot M for Ex ia, CSA IS and FM IS applications
up to 22 Deltapilot M for all other applications, e.g. in non-hazardous areas, Ex nA, etc.
Further information on FOUNDATION Fieldbus, such as requirements for bus system components
can be found in Operating Instructions BA00013S "FOUNDATION Fieldbus Overview".

Endress+Hauser

Deltapilot M FMB50/51/52/53

Input
Measured variable

Hydrostatic pressure

Measuring range
Nominal
value

Range limit

Smallest
calibratable span
(preset at the factory) 1)

MWP

OPL

Vacuum
resistance 2)

Option 3)

lower (LRL) 4)

upper (URL)

[bar (psi)]

[bar (psi)]

[bar (psi)]

[bar (psi)]

[bar (psi)]

[bar (psi)]

[barabs (psiabs)]

0.1 (1.5)

0.1 (-1.5)

+0.1 (+1.5)

0.01 (0.15)

2.7 (40.5)

4 (60)

0.4 (6)

0.4 (-6)

+0.4 (+6)

0.02 (0.3)

5.3 (79.5)

8 (120)

0.01/0.04
(0.145/0.6)

1.2 (18)

1 (-15)

+1.2 (+18)

0.06 (1)

16 (240)

24 (360)

1H

4 (60)

1 (-15)

+4 (+60)

0.2 (3)

16 (240)

24 (360)

1M

10 (150)

1 (-15)

+10 (+150)

0.5 (7.5)

27 (405)

40 (600)

1P

Synthetic oil/
Inert oil

1C
1F

1)

Factory calibration Turn down: Max 20:1, higher on request.

2)

The vacuum resistance applies to the measuring cell at reference conditions.

3)

Product Configurator, "Sensor range" ordering feature

4)

By default, the device is set to a lower range limit of 0 bar. Please specify in the order if the lower range limit is to be set to a different default value.

Endress+Hauser

13

Deltapilot M FMB50/51/52/53

Output
Output signal

4 to 20 mA with superimposed digital communication protocol HART 6.0, 2-wire


Digital communication signal PROFIBUS PA (Profile 3.02)
Digital communication signal FOUNDATION Fieldbus
Option 1)

Output
4 to 20mA HART

PROFIBUS PA

FOUNDATION Fieldbus

1)

Product Configurator, "Output" ordering feature

Signal range
4 to 20 mA HART

3.8 to 20.5 mA

Signal on alarm

As per NAMUR NE 43
4 to 20 mA HART
Options:
Max. alarm: can be set from 21 to 23 mA (Factory setting: 22 mA)
Hold measured value: last measured value is held
Min. alarm: 3.6 mA
PROFIBUS PA: can be set in the Analog Input block,
Options: Last Valid Out Value (factory setting), Fail-safe Value, Status Bad
FOUNDATION Fieldbus: can be set in the Analog Input block,
Options: Last Good Value, Fail-safe Value (factory setting), Wrong Value

Load - 4 to 20 mA HART

R Lmax
[W]
1456
1239
804
369

11.5

20

30

40 45 U
[V]

2
3

R Lmax U 11.5 V
23 mA
A0023090

Load diagram
1
2
RLmax
U

Power supply 11.5 to 30 V DC for intrinsically safe device versions


Power supply 11.5 to 45 V DC (versions with plug-in connector 35 V DC) for other types of protection and for uncertified device
versions
Maximum load resistance
Supply voltage

When operating via a handheld terminal or via a PC with an operating program, a minimum
communication resistance of 250 must be taken into account.

14

Endress+Hauser

Deltapilot M FMB50/51/52/53

Dead time, time constant

I
100 %
90 %
63 %

t1

t2

t
t3
A0019786

Presentation of the dead time and the time constant

Dynamic behavior:
current output

Type

Dead time (t1)


[ms]

Time constant T63 (= t2) [ms]

Time constant T90 (= t3) [ms]

max.

FMB50

60

90

210

max.

FMB51
FMB52
FMB53

500

250

Dead time (t1)


[ms]

Dead time (t1) [ms] +


Time constant T63 (= t2) [ms]

Dead time (t1) [ms] +


Time constant T90 (= t3) [ms]

220

310

370

1020

1110

1170

660

910

1460

1710

Dynamic behavior: HART

Type
min.
max.
min.
max.

FMB50
FMB51
FMB52
FMB53

Reading cycle
Acyclic: max. 3/s, typical 1/s (depends on command # and number of preambles)
Cyclic (Burst): max. 3/s, typical 2/s
The Deltapilot M commands the BURST MODE function for cyclic value transmission via the HART
communication protocol.
Cycle time (update time)
Cyclic (Burst): min. 300 ms
Response time
Acyclic: min. 330 ms, typical 590 ms (depends on command # and number of preambles)
Cyclic (Burst): min. 160 ms, typical 350 ms (depends on command # and number of preambles)

Endress+Hauser

15

Deltapilot M FMB50/51/52/53

Dynamic behavior:
PROFIBUS PA

Type
min.
max.
min.
max.

FMB50
FMB51
FMB52
FMB53

Dead time (t1)


[ms]

Dead time (t1) [ms] +


Time constant T63 (= t2) [ms]

Dead time (t1) [ms] +


Time constant T90 (= t3) [ms]

95

185

245

1195

1285

1345

535

785

1635

1885

Reading cycle
Cyclic: max. 30/s (dependent on the number and type of function blocks used in a closed-control
loop)
Acyclic: typical 25/s
Cycle time (update time)
min. 100 ms
The cycle time in a bus segment in cyclic data communication depends on the number of devices, on
the segment coupler used and on the internal PLC cycle time.
Response time
Cyclic: approx. 8 to 13 ms (depends on Min. Slave Interval)
Acyclic: approx. 23 to 35 ms (depends on Min. Slave Interval)
Dynamic behavior:
FOUNDATION Fieldbus

Type
min.
max.
min.
max.

FMB50
FMB51
FMB52
FMB53

Dead time (t1)


[ms]

Dead time (t1) [ms] +


Time constant T63 (= t2) [ms]

Dead time (t1) [ms] +


Time constant T90 (= t3) [ms]

105

195

255

1105

1195

1255

545

795

1545

1795

Reading cycle
Cyclic: max. 10/s (dependent on the number and type of function blocks used in a closed-control
loop)
Acyclic: typical 5/s
Cycle time (update time)
Cyclic: min. 100 ms
Response time
Cyclic: max. 20 ms (for standard bus parameter settings)
Acyclic: typical 70 ms (for standard bus parameter settings)
Damping

A damping affects all outputs (output signal, display).


Via on-site display, handheld terminal or PC with operating program, continuous from 0...999 s
Via DIP-switch on the electronic insert, switch position
"on" (= set value) and "off" (= damping switched off)
Factory setting: 2 s

Firmware version

Designation

Option 1)

01.00.zz, FF, DevRev01

76

01.00.zz, PROFIBUS PA, DevRev01

77

01.00.zz, HART, DevRev01

78

1)

16

Product Configurator, "Firmware version" ordering feature

Endress+Hauser

Deltapilot M FMB50/51/52/53

Galvanic isolation

The following devices have a galvanic isolation between electronic and probe:
FMB51, FMB52, FMB53
FMB50 with separate housing

Protocol-specific data

HART
Manufacturer ID

17 (11 hex)

Device Type Code

35 (23 hex)

Device Revision

01 (01 hex) - SW version 01.00.zz

HART specification

DD Revision

01 (Dutch)
02 (Russian)

Device description files


(DTM, DD)

Information and files can be found:


www.endress.com
www.hartcomm.org

HART load

Min. 250

HART device variables

The measured values can be freely assigned to the device variables:


Measured values for PV (primary variable)
Pressure
Level
Tank content
Measured values for SV, TV (second and third variable)
Pressure
Level
Measured values for QV (fourth variable)
Temperature

Supported functions

Burst mode
Additional Transmitter Status
Device Locking
Alternative operating modes

PROFIBUS PA
Manufacturer ID

17 (11 hex)

Ident number

1554 hex

Profile Version

3.02
SW Version 01.00.zz

GSD Revision

DD Revision

GSD File

Information and files can be found:


www.endress.com
www.profibus.org

DD Files
Output values

Measured values for PV (via Analog Input Function Block)


Pressure
Flow
Level
Tank content
Measured values for SV
Pressure
Temperature

Endress+Hauser

17

Deltapilot M FMB50/51/52/53

Input values

Input value sent from PLC, can be shown on display

Supported functions

Identification & Maintenance


Simple device identification via control system and nameplate
Condensed status1)
Automatic ident number adaptation and switchable to following ident
numbers):
9700: Profile-specific transmitter identification number with the
"Classic" or "Condensed" status".
1503: Compatibility mode for the old Deltapilot M (DB50, DB50L,
DB51, DB52, DB53).
1555: Identification number for the new Deltapilot M (FMB50,
FMB51, FMB52, FMB53).
Device locking: The device can be locked by hardware or software.

1)

Only with Profile Version 3.02

Data of the FOUNDATION Fieldbus interface


Basic data
Device Type

0x1023

Device Revision

01 (hex)

DD Revision

0x01021

CFF Revision

0x000102

ITK Version

5.2.0

ITK Certification Driver No.

IT067500

Link-Master (LAS) capable

Yes

Link Master / Basic Device


selectable

Yes; Factory setting: Basic Device

Number of VCRs

44

Number of Link Objects in VFD

50

Number of FB-Schedule Objects

40

Virtual communication references (VCRs)


Permanent Entries

44

Client VCRs

Server VCRs

Source VCRs

Sink VCRs

Subscriber VCRs

12

Publisher VCRs

19

Link settings

18

Slot time

Min. inter PDU delay

12

Max. response delay

40

Endress+Hauser

Deltapilot M FMB50/51/52/53

Transducer Blocks
Block

Content

Output values

TRD1 Block

Contains all parameters related to the measurement

Pressure or level (channel 1)


Process temperature (channel 2)
Measured pressure value
(channel 3)
Max. pressure (channel 4)
Level before linearization
(channel 5)

Diagnostic
Block

Contains diagnostic information

Error code via DI channels


(channel 10 to 15)

Display Block

Contains parameters to configure the onsite display

No output values

Function blocks

Endress+Hauser

Block

Content

Number Execution time


of
blocks

Functionality

Resource
Block

The Resource Block contains all the data that


uniquely identify the device. It is an electronic
version of a nameplate of the device.

enhanced

Analog Input
Block 1
Analog Input
Block 2

2
The AI Block receives the measuring data from
the Sensor Block, (selectable via a channel
number) and makes the data available to other
function blocks at its output. Enhancement:
digital outputs for process alarms, fail safe mode.

25 ms

enhanced

Digital Input
Block

This block contains the discrete data of the


Diagnose Block (selectable via a channel number
10 to 15) and provides them for other blocks at
the output.

20 ms

Standard

Digital Output This block converts the discrete input and thus
Block
initiates an action (selectable via a channel
number) in the DP Flow Block or in the TRD1
Block. Channel 20 resets the counter for max.
pressure transgressions value and Channel 21
resets the Totalizer.

20 ms

Standard

PID Block

The PID Block serves as a proportional-integralderivative controller and is used almost


universally for closed-loop-control in the field
including cascade and feedforward. Input IN can
be indicated on the display. The selection is
performed in the Display Block
(DISPLAY_MAIN_LINE_CONTENT).

40 ms

Standard

Arithmetic
Block

This block is designed to permit simple use of


popular measurement math functions. The user
does not have to know how to write equations.
The math algorithm is selected by name, chosen
by the user for the function to be performed.

35 ms

Standard

Input Selector The Input Selector Block facilitates the selection 1


Block
of up to four inputs and generates an output
based on the configured action. This block
normally receives its inputs from AI Blocks. The
block performs maximum, minimum, average
and first good signal selection. Inputs IN1 to IN4
can be indicated on the display. The selection is
performed in the Display Block
(DISPLAY_MAIN_LINE_1_CONTENT).

30 ms

Standard

19

Deltapilot M FMB50/51/52/53

Block

Content

Number Execution time


of
blocks

Functionality

Signal
Characterizer
Block

The Signal Characterizer Block has two sections, 1


each with an output that is a non-linear function
of the respective input. The non-linear function is
generated by a single look-up table with
21 arbitrary x-y pairs.

40 ms

Standard

Integrator
Block

The Integrator Block integrates a variable as a


function of the time or accumulates the counts
from a Pulse Input Block. The block may be used
as a totalizer that counts up until reset or as a
batch totalizer that has a setpoint, where the
integrated or accumulated value is compared to
pre-trip and trip settings, generating a binary
signal when the setpoint is reached.

35 ms

Standard

Additional function block information:

20

Instantiate Function Block

YES

Number of instantiate blocks

20

Endress+Hauser

Deltapilot M FMB50/51/52/53

Power supply
! WARNING
Incorrect connection can limit electrical safety!
When using the measuring device in hazardous areas, installation must comply with the
corresponding national standards and regulations and the Safety Instructions or Installation or
Control Drawings 68 ff, "Safety instructions" and "Installation/Control Drawings" ordering
features.
All explosion protection data are given in separate documentation which is available upon request.
The Ex documentation is provided with all Ex-systems as standard 68 ff, "Safety instructions"
and "Installation/Control Drawings" ordering features.
According to IEC/EN61010 a suitable disconnector has to be installed for the device
HART: Overvoltage protection HAW569-DA2B for the non-hazardous area, ATEX II 2 (1) Ex ia IIC
and IEC Ex ia can be ordered as an option (see "Ordering information" ordering feature).
Protective circuits against reverse polarity, HF influences and overvoltage peaks are installed.
The digital communication signal is transmitted to the bus via a 2-wire connection. The bus also
provides the power supply.

Terminal assignment

- +
2

5
4
3
A0023505

Electrical connection
1
2
3
4
5
6
7

Supply voltage

External grounding terminal


Internal grounding terminal
Supply voltage 21
4...20 mA for HART devices
For HART and FOUNDATION Fieldbus devices: With a handheld terminal, all the parameters can be configured anywhere
along the bus line via menu operation.
Terminals
For HART devices: test terminals, see section "Taking 4 to 20 mA test signal"

4 to 20 mA HART
Type of protection

Supply voltage

Intrinsically safe

11.5 ... 30 V DC

Other types of protection


Devices without certificate

11.5 to 45 V DC (versions with plug-in connection 35 V DC)

Taking 4 to 20 mA test signal


A 4 to 20 mA test signal may be measured via the test terminals without interrupting the
measurement.

Endress+Hauser

21

Deltapilot M FMB50/51/52/53

PROFIBUS PA
Version for non-hazardous areas: 9 to 32 V DC
FOUNDATION Fieldbus
Version for non-hazardous areas: 9 to 32 V DC
Current consumption

Electrical connection

PROFIBUS PA: 11 mA 1 mA, switch-on current corresponds to IEC 61158-2, Clause 21


FOUNDATION Fieldbus: 16 mA 1 mA, switch-on current corresponds to IEC 61158-2, Clause 21
Cable entry

Degree of protection

Option 1)

M20 gland

IP66/68 NEMA 4X/6P

G " thread

IP66/68 NEMA 4X/6P

NPT " thread

IP66/68 NEMA 4X/6P

M12 plug

IP66/67 NEMA 4X/6P

7/8" plug

IP66/68 NEMA 4X/6P

IP65

HAN7D connector, 90 deg


PE cable 5m

2)

M16 valve connector

IP66/68 NEMA4X/6P + pressure compensation via cable S


IP64

1)

Product Configurator, "Electrical connection" ordering feature

2)

Only for FMB50

PROFIBUS PA
The digital communication signal is transmitted to the bus via a 2-wire connection. The bus also
provides the power supply. For further information on the network structure and grounding, and for
further bus system components such as bus cables, see the relevant documentation, e.g. Operating
Instructions BA00034S "PROFIBUS DP/PA: Guidelines for planning and commissioning" and the
PNO Guideline.
FOUNDATION Fieldbus
The digital communication signal is transmitted to the bus via a 2-wire connection. The bus also
provides the power supply. For further information on the network structure and grounding and for
further bus system components such as bus cables, see the relevant documentation, e.g. Operating
Instructions BA00013S "FOUNDATION Fieldbus Overview" and the FOUNDATION Fieldbus Guideline.
Terminals
Cable entry

For wire cross-sections of 0.5 to 2.5 mm (20 to 14 AWG)


Approval
Standard, CSA GP
ATEX II1/2G or II2G Ex ia,
IEC Ex ia Ga/Gb or Ex ia Gb,
FM/ CSA IS
ATEX II1/2D Ex t, II1/2GD Ex ia,
II3G Ex nA,
IEC Ex t Da/Db

Type

Clamping area

Plastic M20x1.5

5 to 10 mm (0.2 to 0.39 in)

Metal M20x1.5 (Ex e)

7 to 10.5 mm (0.28 to 0.41 in)

For other technical data, see the housing section 34 ff".

22

Endress+Hauser

Deltapilot M FMB50/51/52/53

Connector

Devices with valve connector

+
3

1 2 3

Han7D
BN BU GNYE

+
A0023097

BN = brown, BU = blue, GNYE = green/yellow


A
B

Electrical connection for devices with valve connector


View of the connection on the device

Material: PA 6.6
Devices with Harting plug Han7D

6
1

5
3

Han7D

+
A0019990

A
B

Electrical connection for devices with Harting plug Han7D


View of the connection on the device

Material: CuZn, gold-plated contacts of plug-in jack and connector


Devices with M12 plug
PIN assignment for M12 connector
PIN assignment for M12 connector

4
1

3
2

PIN

Meaning

Signal +

Not assigned

Signal

Earth

A0011175

Endress+Hauser offers the following accessories for devices with an M12 plug:
Plug-in jack M 12x1, straight
Material: body PA; coupling nut CuZn, nickel-plated

Endress+Hauser

23

Deltapilot M FMB50/51/52/53

Degree of protection (fully locked): IP66/67


Order number: 52006263
Plug-in jack M 12x1, elbowed
Material: body PBT/PA; coupling nut GD-Zn, nickel-plated
Degree of protection (fully locked): IP66/67
Order number: 71114212
Cable 4x0.34 mm (20 AWG) with M12 socket, elbowed, screw plug, length 5 m (16 ft)
Material: body PUR; coupling nut CuSn/Ni; cable PVC
Degree of protection (fully locked): IP66/67
Order number: 52010285
Devices with 7/8" plug
PIN assignment for 7/8" connector
PIN assignment for 7/8" connector

PIN

Meaning

Signal

Signal +

Not assigned

Shield

A0011176

External thread: 7/8 - 16 UNC


Material: 316L (1.4401)
Protection: IP66/68
Cable specification

HART
Endress+Hauser recommends using twisted, shielded two-wire cables.
The cable outer diameter depends on the cable entry used.
PROFIBUS PA
Use a twisted, shielded two-wire cable, preferably cable type A

For further information on the cable specifications, see Operating Instructions BA00034S
"PROFIBUS DP/PA: Guidelines for planning and commissioning", the PNO Guideline 2.092
"PROFIBUS PA User and Installation Guideline" and IEC 61158-2 (MBP).
FOUNDATION Fieldbus
Use a twisted, shielded two-wire cable, preferably cable type A

For further information on the cable specifications, see Operating Instructions BA00013S
"FOUNDATION Fieldbus Overview", FOUNDATION Fieldbus Guideline and IEC 61158-2 (MBP).
Start-up current HART

12 mA or 22 mA (selectable)

Residual ripple

No influence on 4 to 20 mA signal up to 5 % residual ripple within the permitted voltage range


[according to HART hardware specification HCF_SPEC-54 (DIN IEC 60381-1)]

Influence of power supply

0.001 % of URL/V

24

Endress+Hauser

Deltapilot M FMB50/51/52/53

Overvoltage protection
(optional)

The device can be fitted with overvoltage protection. The overvoltage protection is mounted at the
factory on the housing thread (M20x1.5) for the cable gland and is approx. 70 mm (2.76 in) in length
(take additional length into account when installing). The device is connected as illustrated in the
following graphic.
For details refer to TI001013KEN, XA01003KA3 and BA00304KA2.
Ordering information:
Product Configurator, "Mounted accessories" ordering feature, option NA

+
A

1
+
B

rt
sw

+
-

A0023111

A
B
1
2
3
4

Endress+Hauser

Without direct shield grounding


With direct shield grounding
Incoming connection cable
HAW569-DA2B
Unit to be protected
Connection cable

25

Deltapilot M FMB50/51/52/53

Performance characteristics
Reference operating
conditions

Influence of orientation

As per IEC 60770


Ambient temperature TA = constant, in the range of: +21 to +33 C (+70 to 91F)
Humidity = constant, in the range of: 5 to 80 % RH
Ambient pressure pA = constant, in the range of: 860 to 1060 mbar (12.47 to 15.37 psi)
Position of the measuring cell: constant, in range: FMB50: horizontally 1
FMB51/FMB52/FMB53: vertically 1
Input of LOW SENSOR TRIM and HIGH SENSOR TRIM for lower range value and upper range value
Span based on zero point
Material of the process isolating diaphragm: Alloy C276 (2.4819) and Alloy C276 (2.4819) with
coating (AuRh or AuPt)
Measuring cell material (meter body): Alloy C276 (2.4819), 316L (1.4435)
Filling oil: synthetic oil (FDA)/inert oil
Supply voltage: 24 V DC 3 V DC
Load with HART: 250

< 2.3 mbar (0.0345 psi) when using synthetic oil (FDA)
< 5 mbar (0.075 psi) when using inert oil

Position-dependent zero point shift can be corrected 28, "General installation instructions"
ordering feature.
Calibration position

A0023545

FMB50

FMB51, FMB52, FMB53

To minimize the effect of the orientation (e.g. in the case of vertical device installation), position offset
is preset at the factory.
Resolution

Current output: 1 A
Display: can be set (factory setting: presentation of the maximum accuracy of the transmitter)

Reference accuracy

The reference accuracy comprises the non-linearity according to limit point setting, hysteresis and
non-reproducibility as per IEC 60770. The data refer to the calibrated span.
Reference accuracy in % of the calibrated span
Measuring cell

TD

"Standard" option1)

"Platinum" option 1)

0.1 bar (1.5 psi)

TD 1:1 to TD 2:1
TD > 2:1 to TD 4:1

< 0.2
< 0.1 x TD

< 0.15
< 0.075 x TD

0.4 bar (6 psi)

TD 1:1 to TD 4:1
TD > 4:1 to TD 10:1

< 0.2
< 0.05 x TD

< 0.15
< 0.0375 x TD

1.2 bar (18 psi)

TD 1:1 to TD 2:1
TD > 2:1 to TD 12:1

< 0.2
< 0.1 x TD

< 0.1
< 0.05 x TD

4 bar (60 psi)

TD 1:1 to TD 4:1
TD > 4:1 to TD 20:1

< 0.2
< 0.05 x TD

< 0.1
< 0.025 x TD

10 bar (150 psi)

TD 1:1 to TD 2.5:1
TD > 2.5:1 to TD 20:1

< 0.2
< 0.08 x TD

< 0.1
< 0.04 x TD

1)

26

Product Configurator, "Reference accuracy" ordering feature

Endress+Hauser

Deltapilot M FMB50/51/52/53

Thermal change in the zero


output and the output span

Version

Measuring cell

-10 to +60 C
(+14 to +140F)

60 to 85 C
(140 to 185F)

Only FMB50:
85 to 100 C
(185 to 212F)

% of the calibrated span


FMB50
FMB51/52/53 snap-on
FMB51/52/53 welded

FMB50/51/52/53

Total performance

0.1 bar (1.5 psi)

< (0.32 + 0.30 x


TD)

< (0.34 + 0.40 x


TD)

< (0.34 + 0.55 x


TD)

0.1 bar (1.5 psi)

< (0.32 + 0.50 x


TD)

< (0.34 + 0.60 x


TD)

0.4 bar (6 psi)

< (0.31 + 0.25 x


TD)

< (0.32 + 0.30 x


TD)

1.2 bar (18 psi),


4 bar (60 psi),
10 bar (150 psi)

< (0.31 + 0.10 x


TD)

< (0.32 + 0.15 x


TD)

< (0.33 + 0.20 x


TD)

The "Total performance" specification comprises the non-linearity including hysteresis, nonreproducibility as well as the thermal change in the zero point.
Total performance in % of the URL
Version

Measuring cell

-10 to +60 C
60 to 85 C
(+14 to +140F) (140 to 185F)

85 to 100 C
(185 to 212F)

FMB50
FMB51/52/53 snap-on

0.1 bar (1.5 psi)

< 0.35

< 0.45

< 0.6

FMB51/52/53 welded

0.1 bar (1.5 psi)

< 0.8

<1

< 1.4

0.4 bar (6 psi)

< 0.35

< 0.45

< 0.6

< 0.2

< 0.25

FMB50/51/52/53

Long-term stability

Total error

1.2 bar (18 psi), 4 bar (60 psi),


< 0.15
10 bar (150 psi)

Measuring cell

Long-term stability [%]

0.1 bar (1.4 psi)

< 0.18 of the upper range limit (URL) / year


< 0.45 of the upper range limit (URL) / 5 years

0.4 bar (6 psi)


1.2 bar (18 psi)

< 0.1 of the upper range limit (URL) / year


< 0.25 of the upper range limit (URL) / 5 years

4 bar (60 psi)


10 bar (150 psi)

< 0.05 of the upper range limit (URL) / year


< 0.125 of the upper range limit (URL) / 5 years

The total error comprises the long-term stability and the total performance:
Measuring cell

Warm-up period

Endress+Hauser

% of the URL/year (in the permitted temperature range)

0.1 bar (1.5 psi)

Snap-on: 0.63
Welded: 1.0

0.4 bar (6 psi)

0.61

1.2 bar (18 psi)

0.27

4 bar (60 psi), 10 bar (150 psi)

0.25

4 to 20 mA HART:
FMB50 = 5 s
FMB51/FMB52/FMB53 = 8 s
PROFIBUS PA: 8 s
FOUNDATION Fieldbus: 20 s (after a TOTAL-reset 45 s)

27

Deltapilot M FMB50/51/52/53

Installation
General installation
instructions

The position-dependent zero point shift can be corrected:


directly at the device via an operating key
directly at the device via operating keys on the display
via digital communication if the cover is not open
WARNING
Risk of explosion!
In hazardous areas, comply strictly with the safety instructions when the housing cover is closed and
open.

The local display can be rotated in 90 stages.


FMB50

Level measurement
Always install the device below the lowest measuring point.
Do not install the device at the following positions:
in the filling curtain
in the tank outflow
or at a point in the tank that can be affected by pressure pulses from the agitator
The calibration and functional test can be carried out more easily if you mount the device
downstream of a shutoff device.
Deltapilot M must be included in the insulation for media that can harden when cold.
Pressure measurement in gases
Mount Deltapilot M with shutoff device above the tapping point so that any condensate can flow into
the process.
Pressure measurement in steams
Mount Deltapilot M with siphon above the tapping point.
Fill the siphon with liquid before commissioning.
The siphon reduces the temperature to almost the ambient temperature.
Pressure measurement in liquids
Mount Deltapilot M with the shutoff device below or at the same level as the tapping point.

28

When mounting rod and cable versions, make sure that the probe head is located at a point as free
as possible from flow. To protect the probe from impact resulting from lateral movement, mount the
probe in a guide tube (preferably made of plastic) or secure it with a clamping fixture.
In the case of devices for hazardous areas, comply strictly with the safety instructions when the
housing cover is open.
The length of the extension cable or the probe rod is based on the planned level zero point.
The height of the protective cap must be taken into consideration when designing the layout of the
measuring point. The level zero point (E) corresponds to the position of the process isolating
diaphragm.
Level zero point = E; top of the probe = L.

17 (0.67)

FMB51/FMB52/FMB53

Endress+Hauser

Deltapilot M FMB50/51/52/53

238 (9.37)

61 (2.4)

Suspension clamp (required for FMB53)


Material: 54 ff
Ordering information:
Order number: 52010869 64 ff
Product Configurator, "Accessory enclosed" ordering feature, option "PO".

A0023566

Engineering unit mm (in)

Supplementary installation
instructions

Process isolating diaphragm


Do not clean or touch process isolating diaphragms with hard or pointed objects.
The process isolating diaphragm in the rod and cable version is protected against mechanical
damage by a plastic cap.
Seal
Deltapilot M devices with a G 1 thread:
When screwing the device into the tank, the flat seal has to be positioned on the sealing surface of
the process connection. To avoid additional strain on the process isolating diaphragm, the thread
should never be sealed with hemp or similar materials.
Deltapilot M devices with NPT threads:
Wrap Teflon tape around the thread to seal it.
Tighten the device at the hexagonal bolt only. Do not turn the device at the housing.
Do not overtighten the thread when screwing in the screw. Max. torque: 20 to 30 Nm
(14.75 to 22.13 lbf ft)
Sealing the probe housing
Moisture must not penetrate the housing when mounting the device, establishing the electrical
connection and during operation.
Always firmly tighten the housing cover and the cable entries.
Lubricant is provided on the O-ring seal in the housing cover and on the thread of the aluminum
cover. To ensure that the cover seals tight, any lubricant which has been removed must be replaced.
Use silicone grease or graphite paste as the lubricant. Mineral oil-based grease can destroy the
O-ring.
PE cable length > 300 m (984 ft)
Two suspension clamps must be used for PE cables longer than 300 meters (984 ft).
Cable length tolerances
FMB52
Cable length < 5 m (16 ft): up to -35 mm (-1.38 in)
Cable length 5...10 m (16...33 ft): up to -75 mm (-2.95 in)
Cable length 10...100 m (33...328 ft): up to -100 mm (-3.94 in)
FMB53
Cable length < 5 m (16 ft): up to 17.5 mm (0.69 in)
Cable length 5...10 m (16...33 ft): up to 37.5 mm (1.48 in)
Cable length 10...100 m (33...328 ft): up to 50 mm (1.97 in)
Rod length tolerances
FMB51
Rod length < 4000 mm (157 in): up to -4 mm (-0.16 in)

Wall and pipe mounting

Endress+Hauser offers a mounting bracket for installing the device on pipes or walls. The mounting
bracket is
included in the delivery for the FMB50/51/52 with a separate housing (available for order via
feature 600) and for the FMB53
available for order as a separate accessory (Part No.: 71102216).
For the dimensions, see 48.

Endress+Hauser

29

Deltapilot M FMB50/51/52/53

"Separate housing" version

With the "separate housing" version, you are able to mount the housing with the electronics insert at a
distance from the measuring point. This allows for trouble-free measurement:
Under particularly difficult measuring conditions (at installation locations that are cramped or
difficult to access)
If rapid cleaning of the measuring point is required
If the measuring point is exposed to vibrations
You can choose between different cable versions:
PE (2 m (6.6 ft), 5 m (16 ft) and 10 m (33 ft))
FEP (5 m (16 ft)).
Ordering information:
Product Configurator, "Separate housing" ordering feature
Product Configurator, "Enclosed accessories" ordering feature, option PA
For the dimensions, see 48.

4
3

r 120 (4.72)

A0023567

Engineering unit mm (in)


In the case of the "separate housing" version, the sensor is delivered with the process connection and cable ready mounted. The housing
and a mounting bracket are enclosed as separate units. The cable is provided with a socket at both ends. These sockets are simply
connected to the housing and the sensor.
1
2
3
4

Process connection with sensor - see the following section for the degrees of protection
Cable, both ends are fitted with a socket
Mounting bracket provided, suitable for pipe and wall mounting (for pipes from 1 1/4" up to 2" diameter)
Housing with electronic insert - degrees of protection 32 ff

Degree of protection for the process connection and sensor with the use of
FEP cable:
IP 69K
IP 66 NEMA 4/6P
IP 68 (1.83 mH2O for 24 h) NEMA 4/6P
PE cable:
IP 66 NEMA 4/6P
IP 68 (1.83 mH2O for 24 h) NEMA 4/6P
Technical data of the PE and FEP cable:
Minimum bending radius: 120 mm (4.72 in)
Cable extraction force: max. 450 N (101 lbf)
Resistance to UV light
Use in hazardous area:
Intrinsically safe installations (Ex ia/IS)
FM/CSA IS: for Div.1 installation only

30

Endress+Hauser

Deltapilot M FMB50/51/52/53

Oxygen applications

Oxygen and other gases can react explosively to oils, grease and plastics. As a result, the following are
some of the precautions that must be taken:
All components of the system, such as measuring devices, must be cleaned in accordance with the
BAM (DIN 19247) requirements.
Depending on the materials used, a certain maximum temperature and maximum pressure must not
be exceeded for oxygen applications. The maximum temperature Tmax for oxygen applications is
60 C (140F).
The devices suitable for gaseous oxygen applications are listed in the following table with the
specification pmax.
Order code for devices1) cleaned for
oxygen applications

pmax for oxygen applications

FMB50 2)

Depends on the lowest-rated element, with regard to pressure, of


the selected components: over pressure limit (OPL) of the sensor or
process connection (1.5 x PN)3)
Depends on filling oil 4)

FMB51 )

Depends on the lowest-rated element, with regard to pressure, of


the selected components: over pressure limit (OPL) of the sensor or
process connection (1.5 x PN) )
Depends on filling oil )
Depends on seal material

1)

Only device, not accessory or enclosed accessory

2)

Product Configurator, "Service" ordering feature, option "HB"

3)

13, "Measuring range" ordering feature and 34 ff, "Mechanical construction" ordering feature

4)

Oxygen applications possible with FKM seal and inert oil.

PWIS cleaning

Special cleaning of the transmitter to remove paint-wetting substances, for use in paint shops, for
instance.
Ordering information:
Product Configurator, "Service" ordering feature, option HC
The stability of the materials used must be checked before using them in the medium.
The protective cap of the process isolating diaphragm must be removed if necessary (FMB51/FMB52/
FMB53).

Applications with hydrogen

With regard to materials in which hydrogen formation takes place (e.g. digested sludge), hydrogen
atoms can diffuse through the metal process isolating diaphragm. This can result in incorrect
measurement results.
Endress+Hauser offers process isolating diaphragms with a gold/rhodium coating for such instances.
Ordering information:
Product Configurator, "Process isolating diaphragm material" ordering feature, option L
NOTICE
To reduce the formation of hydrogen, you should not use galvanized assemblies.

Special measuring cells for


acids, alkalis or sea water
(not FMB50)

For acids, alkalis or sea water, Endress+Hauser offers process isolating diaphragms with a gold/
platinum coating.
NOTICE
With temperature exposure (up to 85C(185F)) there is an additional zero point deviation of 1.1
mbar (0.0165 psi).
Ordering information:
Product Configurator, "Process isolating diaphragm material" ordering feature, option N

Endress+Hauser

31

Deltapilot M FMB50/51/52/53

Environment
Ambient temperature limits

Version
Without LCD display

FMB50

FMB51

40 C to +85 C
(-40F to +185F)

With LCD display 1)


With M12 plug , elbowed

Storage temperature range

FMB53

With PE cable: -40 C to +70 C (-40F to +158F)


With FEP cable: 40 C to +80 C (-40F to +176F)
20 C to +70 C (-4F to +158F)

-25 C to +85 C
(-13F to +185F)

With separate housing


(PE and FEP cable)
1)

FMB52

With PE cable: -25 C to +70 C (-13F to +158F)


With FEP cable: -25 C to +80 C (-13F to +176F)
-20 C to +60 C (-4F to +140F)

Extended temperature application range (-40 C to +85 C (40F to +185F)) with restrictions in optical
properties such as display speed and contrast

Version

FMB50

FMB51

Without LCD display

40 C to +90 C
(-40F to +194F)

With LCD display

40 C to +85 C
(-40F to +185F)

With M12 plug , elbowed

-25 C to +90 C
(-13F to +194F)

With separate housing and


FEP cable

FMB52

FMB53

With PE cable: -40 C to +70 C (-40F to +158F)


With FEP cable: 40 C to +80 C (-40F to +176F)
With PE cable: -25 C to +70 C (-13F to +158F)
With FEP cable: -25 C to +80 C (-13F to +176F)
-20 C to +60 C (-4F to +140F)

Climate class

Class 4K4H (air temperature: 20 to 55 C (4 to +131F), relative humidity: 4 to 100%) satisfied as per
DIN EN 60721-3-4 (condensation possible)

Degree of protection

F31 housing: IP 68 (1.83 mH2O for 24 h)


Ordering information:
Product Configurator, "Electrical connection" ordering feature
Separate housing 30

Vibration resistance

Electromagnetic
compatibility

32

Device/accessory

Test standard

Vibration resistance

FMB50, FMB52, FMB53

GL VI-7-2
Part 7: Guidelines for the Performance of
Type Approvals
Chapter 2: Test Requirements for Electrical /
Electronic Equipment and Systems

Guaranteed for:
5 to 25 Hz: 1.6 mm (0.06 in);
25 to 100 Hz: 4 g
in all 3 planes

FMB50, FMB52, FMB53


with mounting bracket

IEC 61298-3

Guaranteed for:
10 to 60 Hz: 0.15 mm (0.01 in);
60 to 500 Hz: 2 g
in all 3 planes

FMB51

IEC 60068-2-6

Guaranteed for:
10 to 60 Hz: 0.075 mm (0.003 in)
60 to 150 Hz 1g in all 3 planes

Electromagnetic compatibility as per all the relevant requirements of the EN 61326 series and
NAMUR Recommendation EMC (NE21). Details can be found in the Declaration of Conformity (in
the Download area of "www.de.endress.com", "search area - Approvals and Certificates", "Manufact.
Declaration").
Max. deviation during EMC-tests < 0.5% of the span
All tests were performed with full measurement range (TD 1:1).

Endress+Hauser

Deltapilot M FMB50/51/52/53

Process
Process temperature range

FMB50

FMB51

10 C to +100 C (+14F to 212F)


135 C (275F) for 30 min. maximum

10 C to +85 C
(+14F to +185F)

FMB52

FMB53

With PE cable: -10 C to +70 C (-14F to 158F)


With FEP cable: 10 C to +80 C (-14F to 176F)

Min. process temperature when using the KALREZ seal:


-3 C (27F)

Lateral load FMB51 (static)


Pressure specifications

1)

30 Nm

! WARNING
The maximum pressure for the measuring device depends on the lowest-rated element with
regard to pressure 13 ff, "Measuring range" ordering feature and 34 ff "Mechanical
construction" ordering feature
Only operate the measuring device within the prescribed limits!
The MWP (maximum working pressure) is specified on the nameplate. This value refers to a
reference temperature of +20 C (68F), (100F (38 C) for ANSI flanges), and may be applied to the
device for an unlimited time. Pay attention to pressure-temperature dependencies.
Please refer to the following standards for pressure values permitted at higher temperatures
EN 1092-1: 2001 Tab. 18 1)
ASME B 16.5a 1998 Tab. 2-2.2 F316
ASME B 16.5a 1998 Tab. 2.3.8 N10276
JIS B 2220.
The MWP applies for the temperature ranges indicated in the "Ambient temperature limits"
( 32) and "Process temperature limits" ordering features (see above).
The test pressure corresponds to the over pressure limit of the device (OPL = 1.5 x MWP) and may
be applied for only a limited time period in order to avoid permanent damage.
The Pressure Equipment Directive (EC Directive 97/23/EC) uses the abbreviation "PS". The
abbreviation "PS" corresponds to the MWP (maximum working pressure) of the measuring device.
In the case of sensor range and process connection combinations where the OPL (over pressure
limit) of the process connection is smaller than the nominal value of the sensor, the device is set at
the factory, at the very maximum, to the OPL value of the process connection. If you want to use
the entire sensor range, select a process connection with a higher OPL value (1.5 x PN; PN = MWP).
In oxygen applications, the values for "pmax and Tmax for oxygen applications" as per 31,
"Oxygen applications" ordering feature may not be exceeded.

With regard to their stability-temperature property, the materials 1.4435 and 1.4404 are grouped together under 13EO in EN 1092-1 Tab. 18. The
chemical composition of the two materials can be identical.

Endress+Hauser

33

Deltapilot M FMB50/51/52/53

Mechanical construction
Device height

The device height is calculated from


the height of the housing and
the height of the individual process connection.
The individual heights of the components are listed in the following sections. To calculate the device
height simply add up the individual heights of the components. Where applicable also take into
consideration the installation distance (space that is used to install the device).
You can use the following table for this purpose:
Section

Page

Height

Housing height

34 ff

Process connections

35 ff

Installation distance
Device height

F31 housing, aluminum

115 (4.53)

103 (4.06)

112 (4.41)

1
2

127 (5)

94 (3.7)

A0023569

Engineering unit mm (in)


1
2

Cover with viewing window


Cover without viewing window

Front view, left-hand side view, top view

Material

Aluminum
Aluminum with glass viewing window
1)

34

Option 1)

Weight kg (lbs)
With display

Without display

1.1 (2.43)

1.0 (2.21)

I
J

Product Configurator, "Housing" ordering feature

Endress+Hauser

Deltapilot M FMB50/51/52/53

F15 housing, stainless steel


(hygienic)

95 (3.74)

1
2

114 (4.49)

100 (3.94)

76 (3)
A0023570

Engineering unit mm (in)


1
2

Cover with viewing window


Cover without viewing window

Front view, left-hand side view, top view

Material

Option 1)

Weight kg (lbs)
With display

Without display

Stainless steel

Stainless steel with glass viewing window

1.1 (2.43)

1.0 (2.21)

Stainless steel with plastic viewing window


1)

Product Configurator, "Housing" ordering feature

ISO 228 G threaded connection


B

50
1
2

59 (2.32)

25 (0.98)

50

59 (2.32)

25 (0.98)

Process connections FMB50,


FMB51, FMB52

G 1" A
55 (2.17)

G 1" A
55 (2.17)

A0023582

A0023581

Engineering unit mm (in)

Item

Designation

Material

Thread ISO 228 G 1 " A

AISI 316L (1.4435)

B
1)

Thread ISO 228 G 1 " A

1: top section AISI 316L (1.4435)


2: bottom section Alloy C276 (2.4819)

Weight kg (lbs)

Option 1)
GGJ

0.8 (1.76)

GGC

Product Configurator, Process connection section

Endress+Hauser

35

Deltapilot M FMB50/51/52/53

Threaded connection NPT

25 (0.98)

54 (2.13)

50

NPT 1"
55 (2.17)
A0023584

Engineering unit mm (in)

Designation

Material

Weight kg (lbs)

Option 1)

Thread ANSI 1 " MNPT

AISI 316L (1.4435)

0.8 (1.76)

RGJ

1)

Process connections FMB50,


FMB51, FMB52

Product Configurator, Process connection section

EN/DIN flanges, connection dimensions as per EN 1092-1/DIN 2527

41

g2

dM = 35.8
g
k
D
A0020004

Engineering unit mm

Flange 1)
Material

Boltholes
3)

Nominal Nominal
diameter pressure

Shape
4)

Diameter Thickness

Diameter Height
Quantof raised of raised ity
face
face

Weight
Diameter

Hole
circle

g2

[mm]

[mm]

[mm]

[mm]

[mm]

[mm]

[kg (lbs)]

Option
2)

AISI 316L

DN 40

PN 10/16

B1 (C)

150

18

88

18

110

3.05 (6.72)

CEJ

AISI 316L

DN 50

PN 10/16

B1 (C)

165

18

102

18

125

3.75 (8.27)

CFJ

AISI 316L

DN 80

PN 10/16

B1 (C)

200

20

138

18

160

5.55 (12.24) CGJ

AISI 316L

DN 100

PN 10/16

B1 (C)

220

20

158

18

180

6.75 (14.88) CHJ

1)

The roughness of the surface in contact with the medium, including the sealing surface of the flanges, is < Ra 0.8 m (31.5 in). Lower surface
roughness available on request.

2)

Product Configurator, Process connection section

3)

Endress+Hauser supplies DIN/EN stainless steel flanges as per AISI 316L (DIN/ EN material number 1.4404 or 14435). With regard to their
stability-temperature property, the materials 1.4404 and 1.4435 are grouped together under 13E0 in EN 1092-1: 2001 Tab.18. The chemical
composition of the two materials can be identical.

4)

Designation as per DIN 2526 in brackets

36

Endress+Hauser

Deltapilot M FMB50/51/52/53

Process connections FMB50,


FMB51, FMB52 - continued

ASME flanges, connection dimensions as per ANSI B 16.5, raised face RF

1.61

g2

dM = 1.41
g
k
D
A0022644

Engineering unit in

Flange 1)
Material 3)

Boltholes
Nominal Class
diameter

Quantity

Weight

Diameter Thickness

Diameter Height of
of raised raised face
face

g2

[in]

[in]

[kg (lbs)]

Option 2)

Diameter Hole
circle

[in]

[lb./sq in] [in]

[in]

[in]

[in]

AISI 316/316L

150

0.69

2.88

0.06

0.62

3.88

2.55 (5.62)

AEJ
(not
FMB51/52)

AISI 316/316L

150

0.75

3.62

0.06

0.75

4.75

3.45 (7.61)

AFJ

AISI 316/316L

150

7.5

0.94

0.06

0.75

6.15 (13.56) AGJ

AISI 316/316L

150

0.94

6.19

0.06

0.75

7.5

8.25 (18.19) AHJ

1)

The roughness of the surface in contact with the medium, including the sealing surface of the flanges, is < Ra 0.8 m (31.5 in). Lower surface
roughness available on request.

2)

Product Configurator, Process connection section

3)

Combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated)

Endress+Hauser

37

Deltapilot M FMB50/51/52/53

Process connections FMB50,


FMB51, FMB52 - continued

JIS flanges, connection dimensions as per JIS B 2220 BL, raised face RF

41

g2

dM = 35.8
g
k
D
A0020004

Engineering unit mm

Flange 1)
Material

AISI 316L
(1.4435)

Boltholes
Nominal
diameter

Nominal Diameter
pressure

Option 2)

Weight

Thickness

Diameter of
raised face

Height of Quantity
raised
face

Diameter Hole circle

g2

[mm]

[mm]

[mm]

[mm]

[mm]

[mm]

[kg (lbs)]

40 A

10 K

140

16

81

19

105

2.55 (5.62)

KEJ

50 A

10 K

155

16

96

19

120

2.95 (6.50)

KFJ

80 A

10 K

185

18

126

19

150

4.25 (9.37)

KGJ

100 A

10 K

210

18

151

19

175

5.35 (11.79)

KHJ

1)

The roughness of the surface in contact with the medium, including the sealing surface of the flanges (all standards), is < Ra 0.8 m (31.5 in).
Lower surface roughness available on request.

2)

Product Configurator, Process connection section

38

Endress+Hauser

Deltapilot M FMB50/51/52/53

Universal adapter
A

68.5 (2.7)

179.5 (7.07)

Process connections FMB50 continued

dM = 36 (1.42)
43.5 (1.71)

43.5 (1.71)
A0020013

A0020014

Engineering unit mm (in)

Item1)

Designation

Material

Approval

Universal process adapter incl. silicone molded seal


(spare part no.: 52023572) FDA 21CFR177.2600/USP Class VI-70C

Universal process adapter incl. EPDM molded seal


(order no.: 71100719) EPDM-70, FDA, USP Class VI; (5 pieces)
Universal process adapter, 6 inch extension incl. silicone molded seal
(spare part no.: 52023572) FDA 21CFR177.2600/USP Class VI-70C

Weight
kg (lbs)

UPJ3)

EHEDG, 3A
1: top section
AISI 316L (1.4404)
2: bottom section
AISI 316L (1.4435)

0.8 (1.76)

UNJ )

UQJ )

EHEDG, 3A
1.7 (3.75)

Universal process adapter, 6 inch extension incl. EPDM molded seal


(order no.: 71100719) EPDM-70, FDA, USP Class VI; (5 pieces)

Option 2)

UOJ )

1)

Surface roughness of the surfaces in contact with the medium Ra 0.76m (30 in) as standard. Surface roughness Ra<0.38 m electropolished
(wetted) Ordering information: Product Configurator, "Service" ordering feature, option "HK".

2)

Product Configurator, Process connection section

3)

Endress+Hauser supplies these slotted nuts in stainless steel AISI 304 (1.4301) or in AISI 304L (1.4307).

Anderson process adapter


B

78 (3.07)

188 (7.4)

2"

2"

3"

3"
A0020015

A0020016

Engineering unit mm (in)

Item1)

Designation

Material

Approval

Weight
kg (lbs)

Option 2)

Short Anderson process adapter, 2-3/16", 316L,


incl. silicone molded seal FDA 21CFR177.2600

3A

1.5 (3.31)

USJ

Long Anderson process adapter, 6-1/2", 316L,


incl. silicone molded seal FDA 21CFR177.2600

1: top section AISI 316L (1.4404)


2: bottom section AISI 316L
(1.4435)
Slotted nut AISI 316L (1.4404)

3A

2.9 (6.39)

UTJ

1)

Surface roughness of the surfaces in contact with the medium Ra 0.76m (30 in) as standard. Lower surface roughness available on request.

2)

Product Configurator, Process connection section

Endress+Hauser

39

Deltapilot M FMB50/51/52/53

56 (2.2)

68.5 (2.7)

Rd 65 x 1/6

68.5 (2.7)

Rd 78 x 1/6

Rd 78 x 1/6
A0020006

A0020007

23.5 (0.93)

20.5 (0.81)
59.5 (2.34)

4x
11.5 (0.45)

56.5 (2.22)

59.5 (2.34)

A0020005

20.5 (0.81)
59.5 (2.34)

10.2 (0.4)
59.5 (2.34)

Hygienic connections

11.2 (0.44)
59.5 (2.34)

Process connections FMB50 continued

65-1.2 (2.56-0.05)

64 (2.52)

84 (3.31)
105 (4.13)
A0020008

A0020009

Engineering unit mm (in)

Item1)

Designation

Approval

Weight
kg (lbs)

Option 2)

DIN 11851 DN 40 PN 25

EHEDG, 3A

0.7 (1.54)

MZJ3)

DIN 11851 DN 50 PN 25

EHEDG, 3A

0.9 (1.98)

MRJ )

DIN11864-1 A DN50 PN16 pipe DIN11866-A,


slotted nut, 316L

EHEDG, 3A

1 (2.21)

NDJ )

Material

AISI 316L (1.4435)

Tri-Clamp ISO 2852 DN 40 DN 51 (2"),


DIN 32676 DN 50

EHEDG, 3A

0.7 (1.54)

TDJ )

DRD DN 50 (65 mm) PN 25,


slotted nut AISI 304 (1.4301)

1.1 (1.98)

TIJ

1)

Surface roughness of the surfaces in contact with the medium Ra 0.76m (30 in) as standard. Lower surface roughness available on request.

2)

Product Configurator, Process connection section

3)

Endress+Hauser supplies these slotted nuts in stainless steel AISI 304 (DIN/EN material number 1.4301) or in AISI 304L (DIN/EN material
number 1.4307).

40

Endress+Hauser

27 (1.06)

59.5 (2.34)

20.5 (0.81)
59.5 (2.34)

9 (0.35)

68 (2.68)

59.5 (2.34)

Deltapilot M FMB50/51/52/53

49.9 (1.96)

84 (3.31)

65 (2.56)

70 (2.76)

Rd 70 x 1/6
84 (3.31)
A0020010

Engineering unit mm (in)

90 (3.54)
A0020011

Engineering unit mm (in)

Item1)

Designation

Varivent Type N for pipes 40 162, PN 40

SMS 2", PN 25

NEUMO, D50, PN 16, 316L

A0020012

Engineering unit mm (in)

Material

AISI 316L (1.4435)

Approval

Weight
kg (lbs)

Option 2)

EHEDG, 3A

1 (2.21)

TRJ

EHEDG

0.7 (1.54)

TXJ3)

3A

0.8 (1.76)

S4J

1)

Surface roughness of the surfaces in contact with the medium Ra 0.76m (30 in) as standard. Lower surface roughness available on request.

2)

Product Configurator, Process connection section

3)

Endress+Hauser supplies these slotted nuts in stainless steel AISI 304 (DIN/EN material number 1.4301) or in AISI 304L
(DIN/EN material number 1.4307).

Endress+Hauser

41

Deltapilot M FMB50/51/52/53

Threaded connection ISO 228 and NPT

Process connections FMB51


(rod version)

21,3 (0.84)

208 (8.19)

42,4 (1.67)

43,5 (1.71)
A0023592

H
L

For dimensions of process connections 34 ff


Probe length L = 0.4 m to 4 m (1.3 ft to 13 ft) - see Product Configurator, Probe connection section

Engineering unit mm (in)

Process connection incl. sensor

Weight

Housing weight

34 ff

Process connection weight

35 ff

Pipe incl. cable

0.77 kg/m (1.70 lbs/3.3 ft)

Threaded connection incl. measuring cell tube and sensor

1.65 kg (3.64 lbs)

Flange connection incl. measuring cell tube and sensor, without a flange

Total weight of device

42

Endress+Hauser

Deltapilot M FMB50/51/52/53

EN/DIN, ANSI and JIS flanges

21,3 (0.84)

208 (8.19)

42,4 (1.67)

43,5 (1.71)
A0023593

H
L

For dimensions of process connections 34 ff


Probe length L = 0.4 m to 4 m (1.3 ft to 13 ft) - see Product Configurator, Probe connection section

Engineering unit mm (in)

Process connection incl. sensor

Weight

Housing weight

34 ff

Process connection weight

35 ff

Pipe incl. cable

0.77 kg/m (1.70 lbs/3.3 ft)

Threaded connection incl. measuring cell tube and sensor

Flange connection incl. measuring cell tube and sensor, without a flange

1.3 kg (2.87 lbs)

Total weight of device

Endress+Hauser

43

Deltapilot M FMB50/51/52/53

Threaded connection ISO 228 and NPT

Process connections FMB52


(cable version)

10,4 (0.41)

236 (9.29)

42,4 (1.67)

43,5 (1.71)
A0023594

H
L

For dimensions of process connections 34 ff


Probe length L = 0.5 to 400 m (1.6 to 1312 ft) - see Product Configurator, Probe connection section

Engineering unit mm (in)

Process connection incl. sensor

Weight

Housing weight

34 ff

Process connection weight

35 ff

PE cable

0.13 kg/m (0.28 lbs/3.3 ft)

FEP cable

0.18 kg/m (0.40 lbs/3.3 ft)

Threaded connection incl. measuring cell tube and sensor

1.65 kg (3.64 lbs)

Flange connection incl. measuring cell tube and sensor, without a flange

Total weight of device

44

Endress+Hauser

Deltapilot M FMB50/51/52/53

EN/DIN, ANSI and JIS flanges

10,4 (0.41)

236 (9.29)

42,4 (1.67)

43,5 (1.71)
A0023597

H
L

For dimensions of process connections 34 ff


Probe length L = 0.5 to 400 m (1.6 to 1312 ft) - see Product Configurator, Probe connection section

Engineering unit mm (in)

Process connection incl. sensor

Weight

Housing weight

34 ff

Process connection weight

35 ff

PE cable

0.13 kg/m (0.28 lbs/3.3 ft)

FEP cable

0.18 kg/m (0.40 lbs/3.3 ft)

Flange connection incl. measuring cell tube and sensor, without a flange

1.3 kg (2.87 lbs)

Total weight of device

Endress+Hauser

45

Deltapilot M FMB50/51/52/53

Dimensions of FMB53 with


F31 housing, suspension
clamp and mounting bracket

181 (7.13)

194 (7.64)

156 (6.14)

137 (5.39)

169 (6.65)

1
2

222 (8.74)

2
0

(7
.8
7

1
2

235 (9.25)

238 (9.37)

61 (2.4)

10,4 (0.41)

236 (9.29)

42,4 (1.67)

43,5 (1.71)
A0023598

FMB53 with suspension clamp and mounting bracket (mounting bracket without marine approval)
1
2
3
4
5
L

Dimensions for housing covers with a viewing window.


Housing covers without a viewing window are approx. 15 mm (0.59 in) lower.
Mounting bracket for pipe and wall mounting (for pipes from 1 " up to 2" diameter)
Suspension clamp
Measuring cell tube
Probe length = 0.5 to 400 m (1.6 to 1312 ft) - see Product Configurator, Probe connection section

Engineering unit mm (in)

Process connection incl. sensor

Weight

Housing weight

34 ff

PE cable (cable length > 120 m (394 ft) = Delivery on cable reel)

0.13 kg/m (0.28 lbs/3.3 ft)

FEP cable (cable length > 120 m (394 ft) = Delivery on cable reel)

0.18 kg/m (0.40 lbs/3.3 ft)

Mounting bracket

0.2 kg (0.44 lbs)

Pipe bend incl. cable entry

0.65 kg (1.43 lbs)

Suspension clamp

0.4 kg (0.88 lbs)

Measuring cell tube incl. sensor

1.0 kg (2.21 lbs)

Total weight of device

46

Endress+Hauser

Deltapilot M FMB50/51/52/53

Dimensions of FMB53 with


F15 housing, suspension
clamp and mounting bracket

174 (6.85)

187 (7.36)

146 (5.75)

159 (6.26)
1
2

127 (5)

1
2

211 (8.31)

2
0

(7
.8
7

224 (8.82)

238 (9.37)

61 (2.4)

10,4 (0.41)

236 (9.29)

42,4 (1.67)

43,5 (1.71)
A0023600

FMB53 with suspension clamp and mounting bracket (mounting bracket without marine approval)
1
2
3
4
5
L

Dimensions for housing covers with a viewing window.


Housing covers without a viewing window are approx. 15 mm (0.59 in) lower.
Mounting bracket for pipe and wall mounting (for pipes from 1 " up to 2" diameter)
Suspension clamp
Measuring cell tube
Probe length = 0.5 to 400 m (1.6 to 1312 ft) - see Product Configurator, Probe connection section

Engineering unit mm (in)

Process connection incl. sensor

Weight

Housing weight

34 ff

PE cable (cable length > 120 m (394 ft) = Delivery on cable reel)

0.13 kg/m (0.28 lbs/3.3 ft)

FEP cable (cable length > 120 m (394 ft) = Delivery on cable reel)

0.18 kg/m (0.40 lbs/3.3 ft)

Mounting bracket

0.2 kg (0.44 lbs)

Pipe bend incl. cable entry

0.65 kg (1.43 lbs)

Suspension clamp

0.4 kg (0.88 lbs)

Measuring cell tube incl. sensor

1.0 kg (2.21 lbs)

Total weight of device

Endress+Hauser

47

Deltapilot M FMB50/51/52/53

42...60 (1.65...2.36)

6 (0.24)

140 (5.51)

122 (4.8)
158 (6.22)

50 (1.97)

52 (2.05)

112 (4.4)

1
2

127 (5)

69 (2.72)

70 (2.76)

86 (3.39)

Wall and pipe mounting with


"Separate housing" version

175 (6.89)

81 (3.19)

6 (0.24)

95 (3.74)

1
2

114 (4.49)

63 (2.48)

52 (2.05)

20 (0.79)
A0023601

1
2
L

Housing cover with viewing window.


Housing cover without viewing window
PE cable 2 m (6.6 ft), 5 m (16 ft) or 10 m (33 ft). FEP cable 5 m (16 ft)

Engineering unit mm (in)

Item

Designation

Dimensions with F31 housing

Dimensions with F15 housing

1)

Option 1)

Weight
Housing (F31 or F15)

Mounting
bracket

34 ff

0.5 kg (1.10)

Product Configurator, "Separate housing" ordering feature

Also available for order as a separate accessory: part number 71102216

48

Endress+Hauser

Deltapilot M FMB50/51/52/53

Reduction in installation height

1
2

36 (1.42)

F15

81 (3.19)

1
2

r 120 (4.72)

17 (0.67)

F31

44 (1.73)

r 120 (4.72)

29 (1.14)

If the separate housing is used, the mounting height of the process connection is reduced compared to
the dimensions of the standard version.

54.1 (2.13)

A0023601

Process connection adapter

Engineering unit mm (in)

Endress+Hauser

49

Deltapilot M FMB50/51/52/53

Example for a "Separate housing" version

1
20

(4
.

72

121,7 (4.79)

L1

54.1 (2.13)

A0023603

Connecting cable with process connection adapter and mounting bracket, here shown with an FMB52
1
2
3
L1
L

Mounting bracket for pipe and wall mounting (for pipes from 1 1/4" up to 2" diameter)
Process connection adapter
Measuring cell tube
PE cable = 2 m (6.6 ft), 5m (16 ft) or 10 m (33 ft)
FEP cable = 5 m (16 ft)
Probe length = 0.5 to 400 m (1.6 to 1312 ft)

Engineering unit mm (in)

Process connection incl. sensor

Weight

Separate housing for FMB50

Weight of housing ( 34 ff) + 0.5 kg (1.10 lbs)

Separate housing for FMB51 and FMB52

Weight of housing ( 34 ff) + 0.65 kg (1.43 lbs)

Process connection adapter

0.4 kg (0.88 lbs)

Mounting bracket

0.2 kg (0.44 lbs)

Pipe bend incl. cable entry

0.65 kg (1.43 lbs)

PE cable 2 m (6.6 ft)

0.16 kg (0.35 lbs)

PE cable 5 m (16 ft)

0.32 kg (0.71 lbs)

Total weight of device

Ordering information for FMB50, FMB51, FMB52: Product Configurator, "Separate housing" ordering
feature

50

Endress+Hauser

Deltapilot M FMB50/51/52/53

Materials not in contact with


process

F31 housing

14

15

16

13
6

8
9
12

11

10

2
3
4

7
A0023615

Front view, left-hand side view, top view

Item
number

Component part

Material

F31 housing, RAL 5012 (blue)

Die-cast aluminum with protective powder-coating


on polyester base

Cover, RAL 7035 (gray)

Die-cast aluminum with protective powder-coating


on polyester base

Cover seal

EPDM

Sight glass

Mineral glass

Sight glass seal

Silicone (VMQ)

External ground terminal

AISI 304 (1.4301)

Nameplates

Plastic film

Attachment for tie-on label

AISI 304 (1.4301)/ AISI 316 (1.4401)

Pressure compensation filter

AISI 316L (1.4404) and PBT-FR

10

Pressure compensation filter, O-ring

VMQ or EPDM

11

Sealing ring

EPDM

12

Snap ring

PC Plastic

13

Seal of cable gland and blind plug

EPDM/NBR

Cable gland

Polyamide (PA), for dust ignition-proof: CuZn nickelplated

Blind plug

PBT-GF30 FR

1
2

14
15

for dust ignition-proof, Ex d, FM XP and CSA XP:


AISI 316L (1.4435)
16

Endress+Hauser

Cover clamp

Clamp AISI 316L (1.4435), screw A4

51

Deltapilot M FMB50/51/52/53

F15 housing

14

7
6
2
3

11

13

12

10

4
A0023616

Front view, left-hand side view, top view

52

Item
number

Component part

Material

F15 housing

Cover

Cover seal

Silicone with PTFE coating

Sight glass for non-hazardous area, ATEX Ex


ia, NEPSI Zone 0/1 Ex ia, IECEx Zone 0/1 Ex
ia, FM NI, FM IS, CSA IS

Polycarbonate (PC)

AISI 316L (1.4404)

Sight glass for ATEX 1/2 D, ATEX 1/3 D,


ATEX 1 GD, ATEX 1/2 GD, ATEX 3 G, FM DIP, Mineral glass
CSA dust ignition-proof

Sight glass seal

Silicone (VMQ)

External ground terminal

AISI 304 (1.4301)

Attachment for tie-on label

AISI 304 (1.4301)/ AISI 316 (1.4401)

Pressure compensation filter

AISI 316L (1.4404) and PBT-FR

Pressure compensation filter, O-ring

VMQ or EPDM

10

Nameplates

lasered

11

Cable gland

Polyamide (PA), for dust ignition-proof: CuZn


nickel-plated

12

Seal of cable gland and blind plug

NBR/Silicone/EPDM

13

Sealing ring

EPDM

14

Screw

A4-50

Endress+Hauser

Deltapilot M FMB50/51/52/53

Connecting parts

4
4

4
4

1
1

5
5

2
2

1
1

3
3

6
6

10
10 11
11

7
7
8
8

9
9

5
5

6
6
A0023617

Item
number

Component part

Material

Connection between the


AISI 316L (1.4404)
housing and process connection

Mounting bracket

Bracket AISI 316L (1.4404)

Screw and nuts A4-70

Half-shells: AISI 316L (1.4404)

Seal for cable from


Separate housing

FKM, EPDM

Gland for cable from separate


housing:
Screws:

AISI 316L (1.4404)

PE cable for separate housing

Abrasion-proof cable with strain-relief Dynema members;


shielded using aluminum-coated film; insulated with
polyethylene (PE-LD), black; copper wires, twisted, UV-resistant

FEP cable for separate housing

Abrasion-proof cable; shielded using galvanized steel wire


netting; insulated with fluorinated ethylene propylene (FEP),
black; copper wires, twisted, UV-resistant

Process connection adapter for


separate housing

AISI 316L (1.4404)

10

Housing adapter

FMB50, FMB51, FMB52: AISI 316L (1.4404)


FMB53: AISI 304 (1.4301)

11

Cable gland:
Sealing insert:
O-ring:

CuZn nickel-plated
TPE-V
NBR

A2

Fill fluid
Designation

Option 1)

Inert oil

Synthetic oil polyalphaolefin FDA 21 CFR 178.3570, NSF H1

1)

Endress+Hauser

Product Configurator, Fill fluid section

53

Deltapilot M FMB50/51/52/53

Materials in contact with the


process

The wetted device components are listed in the "Mechanical construction" ( 34 ff) and "Ordering
information" ( 64 ff) sections.

FMB51

ISO228 Version GGC

FMB53

9
2

10

11

12

13

6
7
8

14
15
8
FMB52

ISO228 Version GGC

16

17

10
2
3

11

12

13

6
7
8

14
15
8

6
7
8
A0023619

54

Item
number

Component part

Material

Process connection

34 ff

Socket

AISI 316L (1.4404)

Endress+Hauser

Deltapilot M FMB50/51/52/53

Item
number

Component part

Material

Rod

AISI 316L (1.4435)

PE cable

Abrasion-proof cable; shielded using galvanized steel wire netting


and aluminum-coated film; insulated with polyethylene (PE-LD),
black/blue; copper wires, twisted, UV-resistant

PE cable
(Usage in drinking water)

Abrasion-proof cable; shielded using galvanized steel wire netting


and aluminum-coated film; insulated with polyethylene (PE-LD),
black; copper wires, twisted, UV-resistant

FEP cable

Abrasion-proof cable; shielded using galvanized steel wire netting


and aluminum-coated film; insulated with fluorinated ethylene
propylene (FEP), black; copper wires, twisted, UV-resistant

Socket

AISI 316L (1.4404)

Probe tube

AISI 316L (1.4404)

Item
number

Component part

Material

Process isolating diaphragm and meter body

64 ff

Seals

64 ff

Protection cap

PPO

Process connection

Alloy C276 (2.4819)

10

Socket

Alloy C4 (2.4610)

11

Rod

Alloy C4 (2.4610)

12

Socket

Alloy C4 (2.4610)

13

Probe tube

Alloy C22 (2.4602)

14

Process isolating diaphragm and meter body

64 ff

15

Seals

64 ff

16

Suspension clamp

AISI 316L (1.4404)

17

Clamping jaw

PA-GF

DIN/EN flanges
Endress+Hauser supplies DIN/EN flanges made of stainless steel AISI 316L as per material numbers
1.4435 or 1.4404. With regard to their stability-temperature property, the materials 1.4435 and
1.4404 are grouped together under 13EO in EN 1092-1 Tab. 18. The chemical composition of the two
materials can be identical.
Process isolating diaphragm
Material

Option 1)

Alloy C

Rhodium>Gold>AlloyC

Platinum>Gold>AlloyC (not FMB50)

1)

Endress+Hauser

Product Configurator, "Material of process isolating diaphragm" ordering


feature

55

Deltapilot M FMB50/51/52/53

Seals
Designation

Option 1)

FKM Viton

A2)

EPDM

J )

Kalrez 6375

L )

None, welded cell

1)

Product Configurator, "Seal" ordering feature

2)

Not FMB50

TSE Certificate of Suitability (Transmissible Spongiform Encephalopathy)


The following applies to all process wetted device components:
They do not contain any materials derived from animals.
No auxiliaries or operating materials derived from animals are used in production or processing.

56

Endress+Hauser

Deltapilot M FMB50/51/52/53

Operability
Operating concept

Operator-oriented menu structure for user-specific tasks

Commissioning
Operation
Diagnostics
Expert level

Quick and safe commissioning


Guided menus for applications
Reliable operation
Onsite operation possible in several languages
Standardized operation at the device and in the operating tools
Parameters relating to measured values can be locked/unlocked using the device's write protection
switch, using the device software or via remote control.
Efficient diagnostics increase measurement availability
Remedial measures are integrated in plain text
Diverse simulation options
Onsite operation

Local display (optional)


A 4-line liquid crystal display (LCD) is used for display and operation. The local display shows measured
values, dialog texts as well as fault and notice messages in plain text, thereby supporting the user at
every stage of operation. The liquid crystal display of the device can be turned in 90 stages. Depending
on the orientation of the device, this makes it easy to operate the device and read the measured values.
Functions:
8-digit measured value display including sign and decimal point, bar graph for 4 to 20 mA HART as
current display; or for PROFIBUS PA as graphic display of the standardized value of the AI Block; for
FOUNDATION Fieldbus as graphic display of the transducer output in relation to the set pressure
range.
Simple and complete menu guidance as parameters are split into several levels and groups
Each parameter is given a 3-digit ID number for easy navigation
Possibility of configuring the display to suit individual requirements and preferences, such as
language, alternating display, contrast setting, display of other measured values such as sensor
temperature etc.
Comprehensive diagnostic functions (fault and warning message etc.)

5
4
3

2
1
A0016498

1
2
3
4
5

Operating keys
Bar graph
Symbol
Header
Parameter ID number

Ordering information:
Product Configurator, "Output; operation" ordering feature

Endress+Hauser

57

Deltapilot M FMB50/51/52/53

Function

Operation via display


HART

PROFIBUS PA

FOUNDATION Fieldbus

Position adjustment (zero point correction)

Setting lower range value and upper range value - reference pressure present
at the device

Device reset

Locking and unlocking parameters relevant to the measured value

Value acceptance indicated by the green LED

Switching damping on and off

Operating keys and elements located on the electronic insert

Span

Zero

on

damp
not used
delta p on:
off:SW
only on:P2= High off:SW

F IE L D C OMMUNIC ATION P R OTO C O L

HAR T

delta p only

damping
SW / Alarm min

SW /
SW / P2=High

Display

3
on

TM

off

off

FOUNDATION

SW
HW
Address

on

off
1

damping

8 7 6 5 4

on

off

SW /
SW / P2=High

Zero

on

off

damping
SW / Alarm min
SW /
SW / P2=High

1
2
3
4
5

on

on
2

2
1

off

Display

damping
simulation
SW /
SW / P2=High
delta p only

2
1

Display

Zero

8 7 6 5 4

10 9 8 7 6
A0023125

A0023126

A0023127

HART electronic insert

PROFIBUS PA electronic insert

FOUNDATION Fieldbus electronic insert

1
2

2
3
4
5
6
7
8

Operating keys for lower range value (zero) and


upper range value (span)
Green LED to indicate successful operation
Slot for optional local display
DIP switch only for Deltabar M
DIP switch only for Deltabar M
DIP switch for alarm current SW / Alarm Min
(3.6 mA)
DIP switch for switching damping on/off
DIP switch for locking/unlocking parameters
relevant to the measured value

3
4
5
6
7
8
9
10

Green LED to indicate successful operation


Operating key for position zero adjustment or
reset (Zero)
Slot for optional local display
DIP switch for bus address SW / HW
DIP switch for hardware address
DIP switch only for Deltabar M
DIP switch only for Deltabar M
Not used
DIP switch for switching damping on/off
DIP switch for locking/unlocking parameters
relevant to the measured value

Function

off

damping
Simulation
SW /
SW / P2=High

2
3
4
5
6
7
8

Operating key for position zero adjustment or


reset (Zero)
Green LED to indicate successful operation
Slot for optional local display
DIP switch only for Deltabar M
DIP switch only for Deltabar M
DIP switch for simulation mode
DIP switch for switching damping on/off
DIP switch for locking/unlocking parameters
relevant to the measured value

Operation with operating keys and elements on the


electronic insert
HART

PROFIBUS PA

FOUNDATION Fieldbus

Position adjustment (zero point correction)

Setting lower range value and upper range value - reference pressure present
at the device

Device reset

Locking and unlocking parameters relevant to the measured value

Value acceptance indicated by the green LED

Switching damping on and off

58

Endress+Hauser

Deltapilot M FMB50/51/52/53

Ordering information:
Product Configurator, "Display, operation" ordering feature
Operating languages

You can also choose another language in addition to the standard language "English":
Option 1)

Designation
English

AA

German

AB

French

AC

Spanish

AD

Italian

AE

Dutch

AF

Chinese

AK

Japanese

AL

1)

Remote operation

Product Configurator, "Additional operating language" ordering


feature

All software parameters are accessible depending on the position of the write protection switch on the
device.
Hardware and software for remote
operation

HART

PROFIBUS PA

FOUNDATION Fieldbus

1)

2)

FieldXpert SFX100 59 ff

NI-FBUS Configurator 60 ff

FieldCare 59 ff

1)

Commubox FXA195 required 60 ff

2)

Profiboard or Proficard required 60 ff

FieldCare
FieldCare is an Endress+Hauser asset management tool based on FDT technology. With FieldCare, you
can configure all Endress+Hauser devices as well as devices from other manufacturers that support the
FDT standard.
FieldCare supports the following functions:
Configuration of transmitters in offline and online mode
Loading and saving device data (upload/download)
HistoROM/M-DAT analysis
Documentation of the measuring point
Connection options:
HART via Commubox FXA195 and the USB port on a computer
PROFIBUS PA via segment coupler and PROFIBUS interface card
Service interface with Commubox FXA291 and ToF adapter FXA291 (USB).
For further information please contact your local Endress+Hauser Sales Center.
Field Xpert SFX100
Field Xpert is an industrial PDA with integrated 3.5" touchscreen from Endress+Hauser based on
Windows Mobile. It offers wireless communication via the optional VIATOR Bluetooth modem or via
WiFi and Endress+Hauser's Fieldgate FXA520. Field Xpert also works as a stand-alone device for asset
management applications. For details refer to BA00060S/04/EN.

Endress+Hauser

59

Deltapilot M FMB50/51/52/53

Commubox FXA195
For intrinsically safe HART communication with FieldCare via the USB interface. For details refer to
TI00404F/00/EN.
Commubox FXA291
The Commubox FXA291 connects Endress+Hauser field devices with a CDI interface
(= Endress+Hauser Common Data Interface) and the USB port of a computer or laptop.
For details, see TI00405C/07/EN.

For the following Endress+Hauser devices, you also need the "ToF adapter FXA291" accessory:
Cerabar S PMC71, PMP7x
Deltabar S PMD7x, FMD7x
Deltapilot S FMB70
ToF adapter FXA291
The ToF adapter FXA291 connects the Commubox FXA291 to devices in the ToF platform, pressure
devices and Gammapilot via the USB port of a computer or laptop. For details, see KA00271F.
Profiboard
For connecting a PC to PROFIBUS.
Proficard
For connecting a laptop to PROFIBUS.
FF configuration program
FF configuration program, such as NI-FBUS Configurator, to
connect devices with "FOUNDATION Fieldbus signal" into an FF-network
set FF-specific parameters
Remote operation via:
Operation with NI-FBUS Configurator:
The NI-FBUS Configurator is an easy-to-use graphical environment for creating linkages, loops, and
a schedule based on the fieldbus concepts.
You can use the NI-FBUS Configurator to configure a fieldbus network as follows:
Set block and device tags
Set device addresses
Create and edit function block control strategies (function block applications)
Configure vendor-defined function and transducer blocks
Create and edit schedules
Read and write to function block control strategies (function block applications)
Invoke Device Description (DD) methods
Display DD menus
Download a configuration
Verify a configuration and compare it to a saved configuration
Monitor a downloaded configuration
Replace a virtual device by a real device
Save and print a configuration
System integration

The device can be given a tag name (max. 8 alphanumeric characters).


Designation

Option 1)

Measuring point (TAG), see additional specifications

Z1

Bus address, see additional specifications

Z2

1)

60

Product Configurator, "Identification" ordering feature

Endress+Hauser

Deltapilot M FMB50/51/52/53

Certificates and approvals


CE mark

The device meets the legal requirements of the relevant EC directives.


Endress+Hauser confirms that the device has been successfully tested by applying the CE mark.

C-tick symbol

The measuring system complies with the EMC requirements of the "Australian Communications and
Media Authority (ACMA)".

Ex approvals

ATEX
FM
CSA
Also combinations of different approvals

All explosion protection data are given in separate documentation which is available upon request. The
Ex documentation is supplied as standard with all devices approved for use in explosion hazardous
areas.
68 ff, "Safety Instructions" and "Installation/Control Drawings" ordering features.
Suitability for hygienic
processes

All materials in contact with foodstuffs comply with framework Regulation (EC) 1935/2004.
The Deltapilot is available with hygienic process connections (overview: see order code).
! CAUTION
Contamination in the process!
Risk of contamination if incorrect seals and parts are used!
To avoid the risk of contamination, when installing the device comply with the design principles of
EHEDG, Guideline 37 "Hygienic Design and Application of Sensors" and Guideline 16 "Hygienic Pipe
Couplings".
Suitable assemblies and seals must be used to ensure hygienic design in accordance with 3-A SSI
and EHEDG specifications.
The leak-proof connections can be cleaned with the cleaning methods typical of this industry (CIP
and SIP). Attention must be paid to the pressure and temperature specifications of the sensor and
process connections for CIP and SIP processes (clean in place/sterilize in place).

74 -

TYPE EL

February 2010
A0024005

A0020029

Gap-free connections can be cleaned without residue using the usual cleaning methods.
Pharma (CoC) (FMB50 only)

Certificate of Compliance (CoC) (according to ASME BPE-2007)


Ordering information:
Product Configurator, "Additional options 1" ordering feature, option "LW"

Functional safety SIL

Endress+Hauser

The Deltapilot M with 4 to 20 mA output signal has been developed to assessed and certified by
TV NORD CERT as per IEC 61508 Edition 2.0 and IEC 61511. These devices can be used to monitor
the process level and pressure up to SIL 2. For a detailed description of the safety functions with
Deltapilot M, settings and functional safety data, see the "Functional safety manual - Deltapilot M"
SD00347P.
Ordering information:
Product Configurator, "Additional approval:" ordering feature, option "LA".

61

Deltapilot M FMB50/51/52/53

Overfill protection

WHG (FMB50, FMB51, FMB52): see document ZE00275P


Ordering information:
Product Configurator, "Additional approval" ordering feature, option "LC".

AD2000

The pressure retaining material 316L (1.4435/1.4404) corresponds to AD2000 - W2/W10.

Pressure Equipment Directive


(PED)

The Deltapilot M device corresponds to Article 3 (3) of the EC directive 97/23/EC (Pressure Equipment
Directive) and has been designed and manufactured according to good engineering practice.

Marine approval

Designation

FMB51

FMB52

FMB53

GL

LE

ABS

LF

LR

LG

BV

LH

DNV

LI

1)

Other standards and


guidelines

Option 1)

FMB50

Product Configurator, "Additional options 1" and "Additional options 2" ordering feature

DIN EN 60770 (IEC 60770):


Transmitters for use in industrial process control systems
Part 1: Methods for inspection and routine testing
DIN 16086:
Electrical pressure measuring instruments, pressure sensors, pressure transmitters, pressure
measuring instruments, concepts, specifications on data sheets
EN 61326 series:
EMC product family standard for electrical equipment for measurement, control and laboratory use.

Drinking water approval

Designation

Option 1)

KTW

LQ

NSF61

LR

ACS (under development)

LS

1)

Classification of process
sealing between electrical
systems and (flammable or
combustible) process fluids in
accordance with ANSI/
ISA 12.27.01

Inspection certificate

62

Product Configurator, "Additional options 1" and "Additional options 2" ordering feature

Endress+Hauser instruments are designed according to ANSI/ISA 12.27.01 either as single seal or
dual seal devices with annunciation, allowing the user to waive the use and save the cost of installing
external secondary process seals in the conduit as required by the process sealing sections of ANSI/
NFPA 70 (NEC) and CSA 22.1 (CEC). These instruments comply with the North-American installation
practice and provide a very safe and cost-saving installation for pressurized applications with
hazardous fluids.
Further information can be found in the control drawings of the relevant devices.
Designation

Option 1)

FMB50

FMB51

FMB52

FMB53

3.1 Material certificate, wetted metallic parts,


EN10204-3.1 inspection certificate

JA

Conformity to NACE MR0175, wetted metallic parts

JB

Conformity to NACE MR0103, wetted metallic parts

JE

Conformity to AD2000, wetted metallic parts,


excepting process membrane

JF

Surface finish measurement ISO4287/Ra, wetted


metallic parts, Inspection certificate

KB

Endress+Hauser

Deltapilot M FMB50/51/52/53

Designation

FMB50

FMB51

FMB52

FMB53

Helium leak test, internal procedure, inspection


certificate

Pressure test, internal procedure, inspection


certificate

3.1 Material certificate+Delta-Ferrit measurement,


internal procedure, wetted metallic parts, EN102043.1 inspection certificate

3.1 Material certificate+PMI test (XRF), internal


procedure, wetted metallic parts, EN10204-3.1
inspection certificate

1)

Calibration; unit

KF

KG

Option 1)

Sensor range; %

Sensor range; mbar/bar

Sensor range; kPa/MPa

Sensor range; mm/mH2O

Sensor range; inH2O/ftH2O

Sensor range; psi

Customized pressure; see additional specification

Customized level; see additional specification

Product Configurator, "Calibration; unit" ordering feature

Option 1)

Designation

F1
2)

DKD/DAkkS calibration certificate 10-point


1)

Product Configurator, "Calibration" ordering feature

2)

Only for FMB50

F2

Option 1)

Designation
Oil and grease removed 2)
Cleaned for oxygen service

HA
)

Cleaned from PWIS (PWIS = paint wetting impairment substances)

Endress+Hauser

KE

Product Configurator, "Test, certificate" ordering feature

Factory calibration, 5-point

Service

KD

Designation

1)

Calibration

Option 1)

HB
)

HC

Configured min alarm current

IA

Configured HART Burst Mode PV

IB

1)

Product Configurator, "Service" ordering feature

2)

Only device, not accessory or enclosed accessory

63

Deltapilot M FMB50/51/52/53

Ordering information
Detailed ordering information is available from the following sources:
In the Product Configurator on the Endress+Hauser website: www.endress.com Select country
Products Select product Product page function: Configure this product
From your Endress+Hauser Sales Center: www.endress.com/worldwide
Product Configurator - the tool for individual product configuration
Up-to-the-minute configuration data
Depending on the device: direct input of information specific to measuring point, such as measuring
range or operating language
Automatic verification of exclusion criteria
Automatic creation of the order code and its breakdown in PDF or Excel output format
Ability to order directly in the Endress+Hauser Online Shop
Configuration data sheet

Pressure
The following configuration data sheet has to be filled in and included with the order if the option J Customized pressure has been selected in feature 90 "Calibration; unit" in the ordering information.
Pressure Engineering Unit

mbar
bar
psi

mmHO
mHO
ftHO
inHO

mmHg

Pa

kgf/cm2

kPa
MPa

Calibration Range / Output

Low range value (LRV):

_______________

[pressure engineering unit]

Upper range value (URV):

_______________

[pressure engineering unit]

Display Information

1st Value Display 1)


Main Value

1)

2nd Value Display 1)


none (Default)
Main Value [%]
Pressure
Current [mA] (HART only)
Temperature

Depending on sensor and communication variant

Damping

Damping:

64

__________ sec (Default 2 sec)

Endress+Hauser

Deltapilot M FMB50/51/52/53

Level
The following configuration data sheet has to be filled in and included with the order if the option K Customized level has been selected in feature 90 "Calibration; unit" in the ordering information.

Pressure Engineering Unit

mbar
bar
psi

mmHO
mHO
ftHO
inHO

Output Unit (Scaled unit)

mmHg
kgf/cm2

Pas
kPa
MPa

Mass

Length

Volume

Volume

Percent

kg
t
lb

m
dm
cm
mm

l
hl

gal
Igal

m3
ft3
in3

ft
inch

Example
Empty pressure [a]:
Low pressure value (empty) ______________
[pressure
engineering unit]

Empty calibration [a]:


Low level value (empty) ______________
[Scaled Unit]

Full pressure [b]:


High pressure value (full)

Full calibration [b]:


High level value (full)

______________
[pressure
engineering unit]

______________
[Scaled Unit]

A0023621

0 mbar / 0m

300 mbar (4.5psi) / 3 m (9.8 ft).

Display Information

1st Value Display 1)


Main Value

1)

2nd Value Display 1)


none (Default)
Main Value [%]
Pressure
Current [mA] (HART only)
Temperature

Depending on sensor and communication variant

Damping

Damping:

__________ sec (Default 2 sec)

Smallest calibratable span (preset at the factory) 13.

Endress+Hauser

65

Deltapilot M FMB50/51/52/53

Accessories
Suspension clamp
(FMB53 only)

29 ff

Extension cable shortening


kit (FMB53 only)

See Product Configurator, "Accessory enclosed" ordering feature, option "PW"


or can be ordered as a separate accessory (part no. 71125862). For details refer to SD00553P/00/A2.

M12 connector

23 ff
Ordering information:
Product Configurator, "Enclosed accessories" ordering feature, option "RL, RM, RN".

Welding necks and Weld-in


tool flanges

See technical Information TI00426F/00.

66

Endress+Hauser

Deltapilot M FMB50/51/52/53

44 (1.73)

47.5 (1.87)

Rd 52 x 1/6 "

39.5 (1.56)

47.5 (1.87)

44 (1.73)

39.5 (1.56)

Rd 52 x 1/6 "

29.5 (1.16)

26.5 (1.04)

25.5 (1)

The following adapters can be used to create a connection between the customer's process connection
and the Deltapilot S FMB50 with a universal adapter:

28.5 (1.12)

Adapter Uni for FMB50

51.6 (2.03)

51.6 (2.03)

61 (2.4)

56 (2.2)

68 (2.68)

Rd 65 x 1
78 (3.07)

Rd 78 x 1
92 (3.62)

44 (1.73)

Rd 52 x 1/6 "
44 (1.73)

39.5 (1.56)

Rd 52 x 1/6 "

A0020038

39.5 (1.56)

6 (0.24)

31.5 (1.24)

A0020037

51.6 (2.03)
65 (2.56)

51.6 (2.03)

84 (3.31)

64 (2.52)

105 (4.13)
A0020039

39.5 (1.56)

A0020040

Rd 52 x 1/6 "
44 (1.73)

68 (2.68)
84 (3.31)
A0020041

Engineering unit mm (in)

Weight
kg (lbs)

Order number

Order number with


3.1 inspection certificate

Option 1)

DIN11851 DN40

0.2 (0.44)

71114172

71114178

RA / R1

DIN11851 DN50

0.3 (0.66)

71114173

71114205

RB / R2

DRD DN50

0.6 (1.32)

71114174

71114206

RC / R3

Clamp 2"

0.2 (0.44)

71114176

71114207

RD / R4

Varivent

0.5 (1.1)

71114177

71114208

RF / R6

Item

Designation

1)

Material

AISI 316L (1.4435)

Product Configurator, Enclosed accessories section

Endress+Hauser

67

Deltapilot M FMB50/51/52/53

Documentation
Technical Information

EMC test procedures: TI00241F/00/EN


Deltabar M: TI00434P/00/EN
Cerabar M: TI00436P/00/EN

Operating Instructions

4...20 mA HART: BA00382P/00/EN


PROFIBUS PA: BA00383P/00/EN
FOUNDATION Fieldbus: BA00384P/00/EN

Brief Operating Instructions

4...20 mA HART: KA01033P/00/EN


PROFIBUS PA: KA01034P/00/EN
FOUNDATION Fieldbus: KA01035P/00/EN

Functional safety manual


(SIL)

Deltapilot M (4 to 20 mA): SD00347P/00/EN

Safety Instructions
Directive

ATEX

1)

Category

Type

Electronics

Documentation

Option 1)

Ex ia IIC

II 1/2 G

FMB50, FMB51, FMB52

4 to 20 mA HART

XA00474P/00

BA

Ex t IIC

II 1/2 D

FMB50, FMB51

4 to 20 mA HART

XA00475P/00

BB

Ex nA

II 3 G

FMB50, FMB51, FMB52, FMB53

4 to 20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus

XA00477P/00

BD

Ex ia IIC

II 2 G

FMB50, FMB51, FMB52, FMB53

4 to 20 mA HART

XA00474P/00

BE

Ex ic IIC

II 3 G

FMB50, FMB51, FMB52, FMB53

4 to 20 mA HART

XA00494P/00

BG

Ex ia IIC
Ex ia IIC

II 1/2 G
II 1/2 D

FMB50, FMB51

4 to 20 mA HART

XA00476P/00

B1

Product Configurator, "Approval" ordering feature

Directive

IECEx

1)

Approval

Approval

EPL

Type

Electronics

Documentation

Option 1)

Ex ia IIC

Ga/Gb

FMB50, FMB51, FMB52

4 to 20 mA HART

XA00478P/00

IA

Ex ia IIC

Gb

FMB52, FMB53

4 to 20 mA HART

XA00478P/00

IC

Ex t IIIC

Da/Db

FMB50, FMB51

4 to 20 mA HART

XA00479P/00

ID

Ex ic IIC

Gc

FMB50, FMB51, FMB52, FMB53

4 to 20 mA HART

XA00493P/00

IE

Ex ia IIC
Ex ia IIIC

Ga/Gb
Da/Db

FMB50, FMB51

4 to 20 mA HART

XA00480P/00

I1

Product Configurator, "Approval" ordering feature

Directive

Approval

Type

Electronics

Documentation

Option 1)

NEPSI

Ex ia IIC

FMB50, FMB51, FMB52, FMB53

4 to 20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus

XA00535P/00

NA

1)

68

Product Configurator, "Approval" ordering feature

Endress+Hauser

Deltapilot M FMB50/51/52/53

Directive

Approval

Electronics

Documentation

Option 1)

TIIS

Ex ia IIC T4

4 to 20 mA HART

TA

1)

Product Configurator, "Approval" ordering feature

Directive

Approval

Type

Electronics

Documentation

Option 1)

INMETRO

Ex ia IIC T6...T4 Ga/Gb


Ex ia IIC T6...T4 Gb

FMB50, FMB51, FMB52

4 to 20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus

XA01304P/00

MA

1)

Product Configurator, "Approval" ordering feature

Installation/Control
Drawings
Directive

FM

Approval

Type

Electronics

Documentation

FM IS Cl.I,II,III Div.1 Gr.A-G, AEx ia,


FM NI Cl.I Div.2 Gr.A-D,
FM IS: Zone 0, 1, 2, 20, 21, 22/
FM NI: Zone 2

FMB50

4 to 20 mA HART
PROFIBUS PA, FOUNDATION Fieldbus

ZD00236P/00
XA00566P/00

1)

FA

FM IS Cl.I,II,III Div.1 Gr.A-G, AEx ia,


Zone 0,1, 2, 20, 21, 22

FMB51

4 to 20 mA HART
PROFIBUS PA, FOUNDATION Fieldbus

XA00566P/00

FM DIP Cl.II,III Div.1 Gr.E-G,


Zone 21, 22

FMB50, FMB51

4 to 20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus

FC

FM NI Cl.I Div.2 Gr.A-D,


Zone 2

FMB50, FMB51

4 to 20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus

FD

FM IS Cl.I Div.1 Gr.A-D, AEx ia,


Zone 0, 1, 2

FMB52, FMB53

4 to 20 mA HART
PROFIBUS PA. FOUNDATION Fieldbus

ZD00236P/00
XA00566P/00

FE

FMB50, FMB53

4 to 20 mA HART
PROFIBUS PA, FOUNDATION Fieldbus

ZD00239P/00
XA00560P/00

C/US IS Cl.I,II,III Div.1 Gr.A-G,


C/US IS Cl.I Div.2 Gr.A-D,
Ex ia C: Zone 0, 1, 2/
US: Zone 0, 1, 2, 20, 21, 22

FMB51, FMB52

4 to 20 mA HART
PROFIBUS PA, FOUNDATION Fieldbus

XA00560P/00

CSA C/US Cl.II, III Div.1 Gr.E-G

FMB50

4 to 20 mA HART

CSA C/US Cl.II, III Div.1 Gr.E-G


US: Zone 21, 22

FMB51

4 to 20 mA HART

C/US IS Cl.I,II,III Div.1 Gr.A-G,


C/US IS Cl.I Div.2 Gr.A-D, Ex ia,
Zone 0,1, 2

CSA

Option 1)

CA

CC

Product Configurator, "Approval" ordering feature

Combination certificate

Directive

Approval

ATEX II Ex ia + FM/CSA IS
ATEX II 1/2G Ex ia IIC T6 +
KEMA/
FM / CSA FM/CSA IS CL.I Div.1 Gr.A-D,
FM/CSA: Zone 0,1,2
1)

Overfill protection

Endress+Hauser

Electronics

Documentation

4 to 20 mA HART

ZD00236P/00
ZD00239P/00

PROFIBUS PA,
FOUNDATION Fieldbus

XA00474P/00

Option 1)

8C

Product Configurator, "Approval" ordering feature

WHG (FMB50, FMB51, FMB52): ZE00275P/00/DE

69

Deltapilot M FMB50/51/52/53

Registered trademarks
HART

Registered trademark of the HART Communication Foundation, Austin, USA

PROFIBUS

Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany

FOUNDATION Fieldbus

Registered trademark of the Fieldbus Foundation, Austin, Texas, USA

70

Endress+Hauser

Deltapilot M FMB50/51/52/53

Endress+Hauser

71

71260308

www.addresses.endress.com

You might also like