Bigdye Terminator Protocol v1.1
Bigdye Terminator Protocol v1.1
Bigdye Terminator Protocol v1.1
1
Cycle Sequencing Kit
Protocol
DRAFT
August 27, 2002 12:32 pm, 4337036A_v1.1Title.fm
Contents
Chapter 1
Introduction
Chapter Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
In This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
About the Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
New Formulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Features and Compatibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
BigDye Terminator v1.1 Cycle Sequencing Kit . . . . . . . . . . . . . . .
1-1
1-1
1-1
1-2
Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Instrument Platforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Thermal Cyclers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Required Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dye/Filter Sets and Matrix Standards for the 310
and 377 Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dye Sets and Spectral Standards for the 3700, 3100,
and 3100-Avant Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dye Set/Primer (Mobility) Files . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-4
1-6
1-6
1-6
1-7
1-4
1-4
1-5
1-6
DRAFT
iii
Chapter 2
2-1
2-1
2-1
2-1
2-3
2-3
2-3
2-3
2-4
DNA Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Poor Template Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Determining DNA Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-4
2-4
2-5
2-5
Chapter 3
iv
3-2
3-3
3-4
3-5
3-5
3-5
3-5
3-6
3-7
DRAFT
3-2
3-2
Chapter 4
4-2
4-2
4-2
4-4
4-10
4-10
4-10
4-10
4-11
DRAFT
C-8
C-8
C-9
C-9
C-9
Appendix D Troubleshooting
Appendix E Obtaining Technical Support
Services and Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
Applied Biosystems Web Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
Index
vi
DRAFT
Introduction
Chapter Summary
In This Chapter
Features and
Compatibilities
DRAFT
1-1
Chapter 1
Introduction
BigDye
Terminator v1.1
Cycle
Sequencing Kit
Buffer (5X), which has been specifically optimized for use with the
new BigDye ready reaction mixes. This buffer should be used for
any reaction optimization you undertake.
1-2
DRAFT
Instruments
Instruments
Instrument
Platforms
The BigDye Terminator v1.1 Cycle Sequencing Kit is for use with
the following instruments:
General instructions are given for using the kit reagents to generate
samples for these instruments. For more detailed instructions, refer to
the appropriate instrument users manual or chemistry guide.
Thermal Cyclers
*Includes the ABI PRISM 377, ABI PRISM 377-18, ABI PRISM 377 with XL
Upgrade, and the ABI PRISM 377 with 96-Lane Upgrade instruments.
BigDye Terminator v1.1 Cycle Sequencing Kit
DRAFT
1-3
Chapter 1
Introduction
Required Software
Dye/Filter Sets
and Matrix
Standards for the
310 and 377
Instruments
The dye/filter sets and matrix standards required for the 310 and 377
instruments are listed in the table below.
Instrument
Dye/Filter Set
Filter Set E
Filter Set E
Standards for
Instrument (Matrix)
File Generation
310/377 BigDye
Terminator v1.1 Matrix
Standards
(PN 4336805)*
*The BigDye Terminator v1.1 Cycle Sequencing Kit does not require a new
spectral calibration file if you currently have a BigDye Terminator v1.0
spectral file on your instrument.
Includes the ABI PRISM 377, ABI PRISM 377-18, ABI PRISM 377 with XL
Upgrade, and the ABI PRISM 377 with 96-Lane Upgrade instruments.
Instrument
Dye/Filter Set
Filter Set E
Filter Set E
Standards for
Instrument (Matrix)
File Generation
3700 BigDye
Terminator v1.1 Matrix
Standard
(PN 4336825)*
3100 BigDye
Terminator v1.1 Matrix
Standard
(PN 4336824)*
*The BigDye Terminator v1.1 Cycle Sequencing Kit does not require a new
spectral calibration file if you currently have a BigDye Terminator v1.0
spectral file on your instrument.
the ABI Prism 3700 DNA Analyzer Users Manual (PN 4306152) for
information on instrument (matrix) files.
1-4
DRAFT
Required Software
Instructions
For the 377 and 310 instruments, refer to the product insert (included
with matrix or sequence standards) for instructions on using the
BigDye Matrix Standards v1.1 to generate matrices.
For Performing Spectral Calibrations
For the 3700 instrument, refer to the product insert for instructions
on using the 3700/3730 BigDye Terminator v1.1 Matrix Standard
to perform spectral calibration.
For the 3100 and 3100-Avant instruments, refer to the product
insert for instructions on using the 3100 BigDye Terminator v1.1
Matrix Standard to perform spectral calibration.
DRAFT
1-5
Chapter 1
Introduction
Dye Set/Primer
(Mobility) Files
Kit
BigDye Terminator v1.1 Cycle
Sequencing Kit
Part Number
100
4337450
1000
4337451
5000
4337452
25,000
4337453
Kit Reagents
1-6
DRAFT
DRAFT
1-7
Chapter 1
Introduction
Materials for
Cycle
Sequencing
The table below lists the plates or tubes required for the
recommended Applied Biosystems thermal cyclers.
Thermal Cycler
Plate or Tube
Applied Biosystems
Part Number
4305505
N801-0560
N801-0533
N801-0534 or N8010535
4313663 or 4311971
N801-0560
N801-0533
N801-0534 N801-0535
4313663 or 4311971
N801-0533
N801-0534 N801-0535
1-8
DRAFT
Materials for
Purifying
Extension
Products
Method
Material
Supplier
Ethanol/EDTA
Precipitation
MLS
EDTA, 125 mM
Sealing taped
Ethanol/Sodium
Acetate
Precipitation
Plate Column
Purification
Spin Column
Purification
Other Equipment
MLS
Costar 6570
Thermowell Sealing
Tape
MLS
Applied Biosystems
(PN 400320)
Costar 6570
Thermowell Sealing
Tape
See Recommended
96-Well Spin Plates
on page 4-14
Costar 6570
Thermowell Sealing
Tape
Centri-Sep spin
column, 1-mL, 32
columns, 100 columns
Applied Biosystems
PN 401763,
PN 401762
Sealing tape
Costar 6570
Thermowell Sealing
Tape
You will also need a variable speed centrifuge with microtiter plate
holders capable of reach a spin speed of at least 1400 g.
Applied Biosystems recommends a Beckman Allegra 6A centrifuge
with a GH-3.8A rotor.
DRAFT
1-9
Chapter 1
Introduction
Materials for
Electrophoresis
Instrument
Material
Supplier
Applied Biosystems
(PN 4311320)
Applied Biosystems
(PN 4336825)
3700/3730 BigDye
Terminator v1.1
Sequencing Standard
Applied Biosystems
(PN 4336799)
Applied Biosystems
(PN 4311320)
Applied Biosystems
(PN 4336824)
Applied Biosystems
(PN 4336791)
Template Suppression
Reagent (TSR)
Applied Biosystems
(PN 401674)
EDTA
MLS
310/377 BigDye
Terminator v1.1 Matrix
Standards
Applied Biosystems
(PN 4336805)
Formamide
MLS
EDTA
MLS
25 mM EDTA with
50 mg/mL blue dextran
Applied Biosystems
(PN 402055)
310/377 BigDye
Terminator v1.1 Matrix
Standards
Applied Biosystems
(PN 4336805)
1-10
DRAFT
General Safety
General Safety
Documentation
User Attention
Words
Five user attention words appear in the text of all Applied Biosystems
user documentation. Each word implies a particular level of
observation or action as described below.
Note: Calls attention to useful information.
IMPORTANT! Indicates information that is necessary for proper
Site Preparation
and Safety Guide
Chemical Safety
Chemical Hazard
Warning
DRAFT
1-11
Chapter 1
Introduction
Chemical Waste
Hazard Warning
About MSDSs
1-12
DRAFT
Chemical Safety
Ordering MSDSs
2.
DRAFT
1-13
Chapter 1
1-14
Introduction
DRAFT
Chapter Summary
In This Chapter
Control DNA
Sequence
An Additional
Control Sold
Separately
DRAFT
2-1
Chapter 2
There are two forms of the v1.0 and v1.1 sequencing standard, as
shown in the table below. Please use the correct sequencing standard
for your instrument. Refer to the product inserts for instructions on
using each sequencing standard.
Instrument
Kit
PN
3700/3730 BigDye
Terminator v1.1
Sequencing Standard
4336799
BigDye Terminator
v1.1 Sequencing
Standard
4336791
2-2
DRAFT
PCR Templates
DRAFT
2-3
Chapter 2
Purifying PCR
Fragments
DNA Quality
Poor Template
Quality
2-4
DRAFT
DNA Quality
Contamination
Proteins
RNA
Chromosomal DNA
Excess PCR primers, dNTPs, enzyme, and buffer components
(from a PCR amplification used to generate the sequencing
template)
Residual salts
Residual organic chemicals such as phenol, chloroform, and
ethanol
Residual detergents
Determining DNA
Quality
Spectrophotometry
The A260/A280 ratio should be 1.7 to 1.9. Smaller ratios usually
indicate contamination by protein or organic chemicals.
Agarose gels reveal the presence of contaminating DNAs and RNAs,
but not proteins. Spectrophotometry can indicate the presence of
protein contamination, but not DNA and RNA contamination. Use
these methods together to get the most information about your DNA
before sequencing.
DRAFT
2-5
Chapter 2
DNA Quantity
Quantitating DNA
Template
Quantity
Quantity
PCR product:
100200 bp
200500 bp
5001000 bp
10002000 bp
>2000 bp
13 ng
310 ng
520 ng
1040 ng
2050 ng
Single-stranded
2550 ng
Double-stranded
150300 ng
Cosmid, BAC
0.51.0 g
23 g
The template quantities given above should work with all primers.
You may be able to use even less DNA when using capillary
instruments for detection. The amount of PCR product to use in
sequencing also depends on the length and purity of the PCR product.
2-6
DRAFT
Chapter Summary
In This Chapter
DRAFT
3-1
Chapter 3
Preparing the
Reactions for 96Well Reaction
Plates or
Microcentrifuge
Tubes
Quantity
8.0 L
Template
Primer
3.2 pmol
Deionized water
q.s.
Total Volume
20 L
3-2
DRAFT
Using BigDye
Terminator
Sequencing
Buffer
Concentration
Volume
2.5X
4 L
BigDye Sequencing
Buffer
5X
2 L
Primer
3.2 pmol
Template
See
Template
Quantity on
page 2-6.
Water
to 20 L
Final Volume
1X
20 L
*The BigDye Terminator v1.1/3.1 Sequencing Buffer is intended for use only
with BigDye Terminator v1.1/3.l Cycle Sequencing Kits.
BigDye Terminator v1.1 Cycle Sequencing Kit
DRAFT
3-3
Chapter 3
Preparing the
Reactions for
384-Well Plates
The type of tube required depends on the thermal cycler that you are
using. Refer to Materials for Cycle Sequencing on page 1-8.
Note: The wells in a 384-well reaction plate have a volume capacity
of 35 L. Therefore, Applied Biosystems recommends doing a 10 L
reaction. This allows the post-reaction cleanup step to be performed
in the same well.
To prepare the reaction mixtures:
Quantity
4.0 L
Template
Primer
3.2 pmol
Deionized water
q.s.
Total Volume
10 L
3-4
DRAFT
Cycle
Sequencing on
the System 9700,
9600, 2700, or
2400
Thermal Cyclers
Only the following thermal cyclers can be used with this protocol:
GeneAmp PCR System 9600
GeneAmp PCR System 9700 (in 9600 emulation mode)
DRAFT
3-5
Chapter 3
Preparing
Sequencing
Reactions
The type of tube required depends on the thermal cycler that you are
using. Refer to Materials Supplied by the User on page 1-7.
To prepare the sequencing reaction:
Quantity
8.0 L
DNA Template
Primer
3.2 pmol
Deionized water
q.s.
Total Volume
20 L
3-6
DRAFT
Performing Cycle
Sequencing
DRAFT
3-7
Chapter 3
3-8
DRAFT
Chapter Summary
In This Chapter
Purification
Methods
DRAFT
4-1
Chapter 4
Note: These precipitation protocols have been optimized for use with
the v1.1 formulation at the specified sequencing reaction volumes
(20 L in 96-well format and 10 L in 384-well format) and are not
recommended for other versions.
IMPORTANT! To clean up sequencing reactions at volumes less than
Ethanol/EDTA Precipitation
Recommended
Protocol
Precipitating in
96-Well Reaction
Plates
4-2
DRAFT
Ethanol/EDTA Precipitation
Then ...
a Beckman Allegra
6A centrifuge with a
GH-3.8A rotor
DRAFT
4-3
Chapter 4
Precipitating in
384-Well
Reaction Plates
4-4
DRAFT
Ethanol/EDTA Precipitation
6.
Then ...
a Beckman Allegra
6A centrifuge with a
GH-3.8A rotor
DRAFT
4-5
Chapter 4
Precipitating in
96-Well Reaction
Plates
4-6
DRAFT
7.
Then ...
a Beckman Allegra
6A centrifuge with a
GH-3.8A rotor
DRAFT
4-7
Chapter 4
Precipitating in
384-Well
Reaction Plates
4-8
DRAFT
7.
Then ...
a Beckman Allegra
6A centrifuge with a
GH-3.8A rotor
DRAFT
4-9
Chapter 4
This section describes the recommended plate and spin columns for
purifying extension products.
IMPORTANT! Extra caution is required when dispensing samples
onto the column bed. Residual dye peaks will result if samples flow
through the sides of the column.
Recommended
Spin Columns
Optimizing Spin
Column
Purification
Spin the column at 325730 g for best results. Use the following
formula to calculate the best speed for your centrifuge:
g = 11.18 r (rpm/1000)2
where:
g = relative centrifugal force
r = radius of the rotor in cm
rpm = revolutions per minute
Do not spin for more than 2 minutes.
Perform the entire procedure without interruption to ensure
optimal results. Do not allow the column to dry out.
4-10
DRAFT
Performing Spin
Column
Purification
DRAFT
4-11
Chapter 4
4-12
DRAFT
Preparing
Extension
Products
98 C for 5 min
25 C for 10 min
5. Spin down the contents of the tubes briefly.
6. Continue with 96-well plate purification.
DRAFT
4-13
Chapter 4
For large-scale procedures, you can use 96-well spin plates, such as
the Gel Filtration Kit from Edge Biosystems.
Note: Other spin plate systems may be used to successfully remove
unincorporated dye terminators. However, due to the large number of
variables associated with using spin plate systems, you will need to
optimize the performance of your system in your own laboratory.
To perform purification with the spin plate:
4-14
DRAFT
Optimizing Primer
Selection
DRAFT
A-1
A-2
DRAFT
Control Sequence
Partial Sequence
of pGEM-3Zf(+)
ACGACTCACT
ATAGGGCGAA
TTCGAGCTCG
40
GTACCCGGGG
ATCCTCTAGA
GTCGACCTGC
AGGCATGCAA
80
GCTTGAGTAT
TCTATAGTGT
CACCTAAATA
GCTTGGCGTA
120
ATCATGGTCA
TAGCTGTTTC
CTGTGTGAAA
TTGTTATCCG
160
CTCACAATTC
CACACAACAT
ACGAGCCGGA
AGCATAAAGT
200
GTAAAGCCTG
GGGTGCCTAA
TGAGTGAGCT
AACTCACATT
240
AATTGCGTTG
CGCTCACTGC
CCGCTTTCCA
GTCGGGAAAC
280
CTGTCGTGCC
AGCTGCATTA
ATGAATCGGC
CAACGCGCGG
320
GGAGAGGCGG
TTTGCGTATT
GGGCGCTCTT
CCGCTTCCTC
360
GCTCACTGAC
TCGCTGCGCT
CGGTCGTTCG
GCTGCGGCGA
400
GCGGTATCAG
CTCACTCAAA
GGCGGTAATA
CGGTTATCCA
440
CAGAATCAGG
GGATAACGCA
GGAAAGAACA
TGTGAGCAAA
480
AGGCCAGCAA
AAGGCCAGGA
ACCGTAAAAA
GGCCGCGTTG
520
CTGGCGTTTT
TCCATAGGCT
CCGCCCCCCT
GACGAGCATC
560
ACAAAAATCG
ACGCTCAAGT
CAGAGGTGGC
GAAACCCGAC
600
AGGACTATAA
AGATACCAGG
CGTTTCCCCC
TGGAAGCTCC
640
DRAFT
B-1
Appendix B
CTCGTGCGCT
CTCCTGTTCC
GACCCTGCCG
CTTACCGGAT
680
ACCTGTCCGC
CTTTCTCCCT
TCGGGAAGCG
TGGCGCTTTC
720
TCATAGCTCA
CGCTGTAGGT
ATCTCAGTTC
GGTGTAGGTC
760
GTTCGCTCCA
AGCTGGGCTG
TGTGCACGAA
CCCCCCGTTC
800
AGCCCGACCG
CTGCGCCTTA
TCCGGTAACT
ATCGTCTTGA
840
GTCCAACCCG
GTAAGACACG
ACTTATCGCC
ACTGGCAGCA
880
GCCACTGGTA
ACAGGATTAG
CAGAGCGAGG
TATGTAGGCG
920
GTGCTACAGA
GTTCTTGAAG
TGGTGGCCTA
ACTACGGCTA
960
CACTAGAAGG
ACAGTATTTG
GTATCTGCGC
TCTGCTGAAG
1000
B-2
DRAFT
Sample Electrophoresis
DRAFT
C-1
Run Modules
Configuration
Run Module
Seq1_1POP5DefaultModule
Seq1_2POP5DefaultModule
Seq1_1POP6DefaultModule
Seq1_2POP6DefaultModule
POP-5 polymer
DT3700POP5{BD}v3.mob
POP-6 polymer
DT3700POP6{BD}v5.mob
Standards
IMPORTANT! Use Dye Set E.
Dye Set
E
Standards
3700 BigDye Terminator v1.1 Matrix Standard
(PN 4336825)
C-2
DRAFT
Performing
Sample
Electrophoresis
DRAFT
C-3
Run Module
UltraSeq_POP4Default Module
LongSeq80_POP4DefaultModule
RapidSeq36_POP6DefaultModule
StdSeq50_POP6DefaultModule
POP-4 polymer
DT3100POP4LR{BD}v1.mob
POP-6 polymer
DT3100POP6{BD}v1.mob
Standards
IMPORTANT! Use Dye Set E.
Dye Set
E
Standards
3100 BigDye Terminator v1.1 Matrix Standard
(PN 4336824)
C-4
DRAFT
Performing
Sample
Electrophoresis
DRAFT
C-5
Run Module
P4StdSeq (1 mL) E
P4RapidSeq (1 mL) E
Polymer
Operating
System
POP-4 polymer
NT
POP-6 polymer
POP-4 polymer
DT310POP6{BD-LR}v3.mob
DT310POP6{BD}.mob
DT310POP4{BD}v2.mob
Macintosh
POP-6 polymer
C-6
DRAFT
DT POP6{BD Set-AnyPrimer}
Standards
IMPORTANT! The instrument (matrix) file for the BigDye
Performing
Sample
Electrophoresis
DRAFT
C-7
Run Module
*Any plate check and prerun module can be used on the ABI PRISM 377 DNA
Sequencers.
Operating
System
NT
Macintosh
*Includes the ABI PRISM 377, ABI PRISM 377-18, ABI PRISM 377 with XL Upgrade, and
the ABI PRISM 377 with 96-Lane Upgrade instruments.
C-8
DRAFT
Standards
IMPORTANT! The instrument (matrix) file for the BigDye
Using the
Lane Guide Kit
If you are using the BigDye chemistries v1.1 on the 377 instrument
in conjunction with the ABI PRISM Lane Guide Lane
Identification Kit, refer to that kits protocol (PN 4313804) for
instructions on resuspending and loading samples.
Using Long-Read
Gel and Buffer
Formulations
For longer sequencing read lengths follow the gel and buffer
formulations described in the user bulletin entitled Achieving Longer
High Accuracy Reads on the 377 Sequencer (PN 4315153).
Performing
Sample
Electrophoresis
DRAFT
C-9
C-10
DRAFT
Troubleshooting
Observation
Possible Causes
Recommended Actions
No recognizable
sequence
Insufficient template
Inhibitory contaminant in
the template
Insufficient primer
Missing reagent
Old or mishandled
reagents
Thermal cycling
conditions incorrect
DRAFT
D-1
Observation
Possible Causes
Recommended Actions
No recognizable
sequence (continued)
Electrokinetic injection
failure (capillary
instruments)
Degraded template
Old or mishandled
reagents
Thermal cycling
conditions incorrect
Electrokinetic injection
failure (capillary
instruments)
Inhibitory contaminant in
the template
Observation
Possible Causes
Recommended Actions
Incorrect instrument
(matrix) file
Primer-dimer
contamination in PCR
sequencing
Noise up to or after a
specific point in the
sequence
DRAFT
D-3
Appendix D
Troubleshooting
Observation
Possible Causes
Recommended Actions
Poor removal of
unincorporated dye
terminators
D-4
DRAFT
DRAFT
E-1
Appendix E
E-2
DRAFT
Index
Numerics
310 requirements
mobility files C-6
run modules C-6
sequencing standards C-7
3100 requirements
mobility files C-4
run modules C-4
spectral calibration standards
3700 requirements
mobility files C-2
run modules C-2
spectral calibration standards
377 requirements
matrix file standards C-9
mobility files C-8
run modules C-8
cycle sequencing
BAC DNA 3-7
protocol for GeneAmp
D
disposables
cycle sequencing 1-8
purifying 1-9
DNA
quantity characteristics 2-6
quantity to use 2-6
dye set/primer files, downloading
dye/filter set
3100 1-4
3700 1-4
377 1-4
C-4
C-2
1-6
A
attention words
3-5
ethanol/EDTA precipitation
384-well 4-4
96-well 4-3
ethanol/EDTA/sodium acetate
precipitation in 384-well 4-8
precipitation in 96-well 4-6
1-11
B
bacterial artificial chromosomes
2-3
C
centrifuge, recommended 1-9
components, sequencing kits 1-6
contaminants, of templates 2-5
control
double-stranded 2-1
electrophoresis 2-1
single-stranded 2-1
control DNA, reason for 2-1
customer support. See technical support
F
formulation, new 1-1
freeze-thaw, limits 1-7
H
hazards
chemical 1-11
waste 1-12
E-1
DRAFT
Index-1
instruments
sequencing 1-3
thermal cyclers 1-3
reaction mixtures
BAC DNA 3-7
for 384-well 3-4
for 96-well 3-2
for microcentrifuge tubes
reagents
electrophoresis 1-10
purifying 1-9
required files
matrix 1-1
spectral 1-1
L
Lane Guide C-9
long read lengths C-9
M
matrix standards
310 1-4
377 1-4
general requirements
mobility files 1-2
about C-1
downloading 1-6
MSDSs, ordering 1-13
S
1-2
P
pGEM-3Zf(+) sequence B-1
primers
importance of selection A-1
optimizing selection A-1
purification
SDS and 96-well 4-13
spin plate 4-14
purifying
methods 2-3, 4-1
methods for PCR 2-3
PCR by ultrafiltration 2-4
Q
quality problems, template 2-5
quality, examining DNA 2-5
quantitating DNA 2-6
Index-2
3-2
DRAFT
safety
chemical 1-11 to 1-13
guide 1-11 to 1-13
SDS, and purification 4-13
sequencing buffer
optimizing 1-2
sequencing kits, ordering 1-6
sequencing problems 2-4
sequencing standards
for each instrument 2-2
general requirements 1-2
shorter read lengths, and DNA quantity 2-6
signal intensities, and DNA quantity 2-6
site preparation and safety guide 1-11
spectral standards
3100 1-4
3700 1-4
spin column purification
optimizing 4-10
performing 4-11
storage temperatures 1-7
T
technical support E-1
templates
BAC DNA 2-3
PCR 2-3
poor quality characteristics 2-4
preparing 2-3
top-heavy data, and DNA quantity 2-6
U
user attention words
1-11
W
warning symbols
1-11
DRAFT
Index-3
Index-4
DRAFT
Headquarters
850 Lincoln Centre Drive
Foster City, CA 94404 USA
Phone: +1 650.638.5800
Toll Free (In North America): +1 800.345.5224
Fax: +1 650.638.5884
Worldwide Sales and Support
Applied Biosystems vast distribution and
service network, composed of highly trained
support and applications personnel, reaches
150 countries on six continents. For sales office
locations and technical support, please call our
local office or refer to our Web site at
www.appliedbiosystems.com.
www.appliedbiosystems.com