Apc 2000alw PDF
Apc 2000alw PDF
APC-2000ALW
el
Mod 8
200
133
132
Rating plate
18
91.5
18
40
38.5
Lock preventing
the casing
rotation
Process connections
G1/2 type
SW27
SW27
M type
M201.5, 4 hole
Application
25
G1/2", 12 hole
P type
M201.5, 12 hole
15
20
GP type
SW27
2 8
G1/2", 4 hole
G1/4
12
1/2NPT
25
Application
1/2NPT type
SW32
10
31
G1B
30
CG1 type
O-ring
26 2
Application
15
10.5
SW41
2 10
316Lss standard
25
35
CM302 type
Smart pressure transmitters are applicable to the measurement of the pressure, underpressure and absolute pressure of gases, vapours and
liquids. The active sensing element is a piezoresistant silicon sensor separated from the medium by a diaphragm and by specially selected type
of manometric liquid. The casing is made of aluminium alloy cast, degree of protection IP66. The design of the casing enables the use of a local
display, rotation of the display by 90, rotation of the casing by 0355 relative to the sensor, and a choice of cable direction.
Version APC-2000ALW
420 mA output signal + HART protocol
or digital Profibus PA signal (description page 7)
Possibilities of the adjusting both zero point and of the
start and end of the measuring range, characteristic
ete.with the display panel keys
Configurable display 5 digits with illumination (working
temperature range 40+85C)
ATEX Intrinsic safety EX II 1/2G Ex ia IIC T5
ATEX Explosion proof EX II 1/2G Ex d ia IIC T5
Version APC-2000ALE
020 mA, 05 mA, 420 mA, output signal + HART
protocol
Possibilities of the of the adjusting both start and end
of the measuring range according to set pressure with
the display panel keys
Configurable display (LCD) 3 digits (working
temperature range 40+85C)
No EEx or Profibus PA
Installation
Version: APC-2000ALW
Milliammeter
RS-HART converter
or communicator
RS-HART converter
or communicator
Power supply
420
mA
4 3 2 1
4 3 2 1
TEST
TEST
2 WIRE
TEST
TEST
2 WIRE
OUTPUT
3 WIRE
240 hm
OUTPUT
3 WIRE
Transmitter
Jumper
020
05
mA
R0
R0
TEST
SIGNAL
RS-HART converter
or communicator
420
mA
mA
TEST
Power supply
Power supply
Milliammeter
mA
Version: APC-2000ALE
with 05 or 020mA output signal
Measuring ranges
No.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
Accuracy
Long-term stability
(1MPa)
(300 kPa)
(160kPa)
(70 kPa)
(25 kPa)
(7 kPa)
(30 kPa)
(12 kPa)
(10 kPa)
(5 kPa)
(5 kPa)
(2.5 kPa)
(2 kPa)
(0.5 kPa)
(5 kPa abs)
(7 kPa abs)
(25 kPa abs)
(70 kPa abs)
Technical data
Metrological parameters
Electrical parameters
Power supply
12...45 V DC (EEx 13,5...28 V)
Additional voltage drop
when display illumination switched on
3V
Output signal
APC-2000ALE:
Usup[V] 12V
Materials
Casing:
Aluminium
- NACE MR 01 75 compatible
Operating conditions
0.85
250...1100
-4080C
-40...120C
Others
r0
Set range
30
100
[%]
r1
10
-40...85C
EEx version
0,02A
100:1
100:1
100:1
100:1
100:1
100:1
23:1
20:1
20:1
20:1
20:1
10:1
10:1
17:1
22:1
100:1
100:1
100:1
Thermal error
Overpressure limit
(without hysteresis)
1200 bar
(120 MPa)
450 bar
(45 MPa)
450 bar
(45 MPa)
140 bar
(14 MPa)
50 bar
(5 MPa)
14 bar
(1.4 MPa)
14 bar
(1.4 MPa)
4 bar
(400 kPa)
4 bar
(400 kPa)
2 bar
(200 kPa)
2 bar
(200 kPa)
1 bar
(100 kPa)
1 bar
(100 kPa)
0.5 bar
(50 kPa)
2 bar
(200 kPa)
14 bar
(1.4 MPa)
50 bar
(5 MPa)
140 bar
(14 MPa)
Rangeability
Communication
For pressure measurements in the areas under explosion hazard the Atex intrinsically safe transmitters,
II 1/2G EExia
IIB/T5 are available
Technical data
Metrological parameters, materials of process connection, diaphragms and casing, and operating conditions see the description page 6.
Electrical parameters
Output parameters
Electrical diagram
Configuration
Full configuration of transmitter settings, adjustment of the display mode, transmitter zeroing and calibration in relation to
pressure standards proceeds with the PDM (Process Device
Manager) software, by Siemens. The EED program library,
worked out by Aplisens for cooperation with this transmitter, is
helpful in the configuration.
Other commercial configuration software (e.g. Commuwin by
Endress and Hauser, DTM/FDT tools) make transmitter configuration possible in the range of basic commands.
Ordering procedure
Inside
earthing
clamp
Coupler
Example 1: Pressure transmitter with display, output 4..20mA + HART, version EExia, nominal measuring range 0..7bar, calibrated
range 0..6bar, process connection M20x1,5, electrical connection gland M20x1,5.
APC-2000ALW/EExia/0..7bar/0..6bar/M
Example 2: Pressure transmitter with display, output 4..20mA + HART, version EExia, version PED, nominal measuring range
0..25bar, calibrated range 0..16bar, process connection G1/2, electrical connection 1/2NPT F, mounting bracket for 2 pipe
APC-2000ALW/EExia/PED/0..25bar/0..16bar/G1/2/US/AL
Example 3: Pressure transmitter with display, Profibus PA, version EExia, nominal measuring range 0..7bar, calibrat ed range
0..6bar, process connection flange diaphragm seal DN50PN40 , electrical connection M20x1,5,
APC-2000AL/Profibus PA/W/EExia/0..7bar/0..6bar/S-P DN50.
Model
Code
Description
APC-2000
Casing,
output signal,
ALW......................
ALE.......................
AL/Profibus PA/W...............
Versions, certificates*
/EExia......................
/EExd....................
Ex II 1/2G EEx d ia IIC T5, for pressure >250mbar ( not available for ALE,
AL/Profibus PA, AL/Profibus PA/W). Packing gland available for request.
/PED..........................................
/Tlen.........................
For oxygen service (sensor filled with Fluorolube fluid), only M and G1/2 conn.
/-60+50C...........................................
/-40+80C...................
/01000bar**.....................................
/0300bar..........................................
/0160bar**...................................
/070bar.....................................
/025bar.....................................
/07bar...................................
/02bar..........................................
/01bar...................................
/00,25bar.........................................
/-0.5 +0.5bar.......................................
/-11.5bar......................................
/-16bar**.............................................
/-100100mbar.....................................
/-1570mbar......................................
/01.1bar ABS......................................
/07barABS......................................
/025barABS........................................
/070bar ABS.......................................
Process connections
/M.......................................
/G1/2........................................
/P........................................
/P (Hastelloy).............................
/GP..................................
/GP (Hastelloy)..........................
/CM30x2....................................
/CM30x2 (Hastelloy)..................
/CG1.........................................
/CG1/2......................................
/1/2NPT M................................
/1/2NPT F.................................
/code of diaphragm seal ...........
(without marking)................
/US.....................................
Electrical connection
Accessories
Other specification
The most typical specification is marked by mark.
/AL.........
/............................
01000bar ( 0100MPa)
0300bar ( 030MPa)
0160bar ( 016MPa)
070bar ( 07MPa)
0.25bar ( 02,5MPa)
07bar ( 0700KPa)
02bar ( 0200kPa)
01bar ( 0100kPa)
00,25bar ( 025kPa)
-0,50,5bar ( -5050kPa)
-11,5bar ( -100150kPa)
-16bar ( -100600kPa)
-100100mbar ( -1010kPa)
-1570mbar ( -1,570kPa)
01.1bar absolute pressure (0110kPa abs)
10bar ( 1MPa)
3bar ( 300kPa)
1,6 bar ( 160kPa)
0,7bar ( 70kPa)
0,25bar ( 25kPa)
0,07bar( 7kPa)
100mbar ( 10kPa)
50mbar ( 5kPa)
25mbar ( 2.5kPa)
50mbar ( 5kPa)
120mbar ( 12kPa)
300mbar ( 30kPa)
20mbar ( 2kPa)
5mbar ( 0.5kPa)
50mbar abs (5kPa abs)
Start and end of calibrated range in relation to 4mA and 20mA output
Thread M20x1,5 (male) with 4hole, wetted parts SS316L
Thread G1/2 (male) with 4hole , wetted parts SS316L
Thread M20x1,5 (male) with 12hole, wetted parts SS316L
Thread M20x1,5 (male) with 12hole, wetted parts Hastelloy C 276
Thread G1/2 (male) with 4hole , wetted parts SS316L
Thread G1/2 (male) with 4hole , wetted parts Hastelloy C 276
Thread M30x2 with flush diaphragm, wetted parts SS316L
Thread M30x2 with flush diaphragm, wetted parts Hastelloy C 276
Thread G1 with flush diaphragm, wetted parts SS316L
Thread G1/2 with flush diaphragm, wetted parts SS316L
Thread NPT Male, wetted parts SS316L
Thread M20x1,5 with adapter to NPT Female, wetted parts SS316L
Diaphragm seal (see chapter of diaphragm seals)
Packing gland M20x1,5
Thread 1/2NPT Female
Mounting bracket type AL. for 2 pipe, material zinced steel
Description of required parameters (e.g. non-standard process connection
G3/4, M22x1.5)