TExas Research Data
TExas Research Data
What We Did
Measuring Pavement
Marking Thickness
Pavement marking thickness is
one of the two primary inspection
measurements made with pavement
markings, the other being
retroreflectivity. Researchers compared
the relative accuracies of thermoplastic
thickness measurement with caliper
(common field practice) versus
needlepoint micrometer (recommended
field practice). The main difference
between the two measurement devices
is that the needlepoint micrometer is
capable of measurement between the
beads, while the caliper is capable of
measuring only to the top-of-bead.
The difference between the two
devices is significant because TxDOT
specifications require contractors to
achieve minimum thermoplastic binder
thickness, not including drop-on beads.
What We Found
Measuring Pavement
Marking Thickness
Forty-seven thermoplastic
pavement marking samples of
varying thickness and beads were
taken from striping jobsites statewide
and used in the analysis. Research
report 4150-1, Analysis of TxDOT
Thickness Measurement Procedures
for Thermoplastic Pavement Markings,
summarizes the findings from the
analysis. The results show that the
caliper measured an average of
20.5 mil and 16.7 mil thicker than
the needlepoint micrometer for large
(Type III) and small (Type II) bead
samples, respectively. Based on
the research findings, the following
recommendations were made:
Thickness measurements should be
made with a needlenose micrometer
to the top of the thermoplastic.
The use of a caliper to measure
thermoplastic thickness is
discouraged.
A minimum of three measurements
should be taken.
The most accurate measurement
method is to measure diagonally
across the sample (taking at least
three measurements per sample).
Figure 2 illustrates where the
measurements should be made
on the sample.
The Researchers
Recommend
TTI.PSR0401.0304.580