1. Safety Warnings
This instrument has been designed, manufactured and tested according to the following standards.
IEC 61010-1 Measurement CAT. 300V Pollution degree 2
IEC 61010-031
IEC 61326
This instruction manual contains warnings and safety rules which must be observed by the user to ensure safe operation of the instrument and retain it in safe condition. Therefore, read through these operating instructions before using the instrument.
WARNING
Read through and understand instructions contained in this manual before starting using the instrument.
Save and keep the manual handy to enable quick reference whenever necessary.
This instrument should only be used by suitably trained person and be sure to follow measurement procedures described in the manual. Kyoritsu assumes no
responsibility for damage and injury caused by misuse or not following instructions in the manual.
Be sure to understand and follow all safety instructions contained in the manual.
Be sure to observe the above instructions.
Failure to follow the above instructions may cause injury, instrument damage and/or damage to equipment under test.
The symbol
indicated on the instrument means that the user must refer to related parts in the manual for safe operation of the instrument.
Be sure to carefully read the instructions following each
symbol in the manual.
DANGER is reserved for conditions and actions that are likely to cause serious or fatal injury.
WARNING is reserved for conditions and actions that can cause serious or fatal injury.
CAUTION is reserved for conditions and actions that can cause injury or instrument damage.
DANGER
Never make measurement on the circuit in which electrical potential to ground over 300V AC/DC exists.
Do not attempt to make measurement in the presence of flammable gasses. Otherwise, the use of the instrument may cause sparking, which can lead to an
explosion.
Be sure to keep your fingers behind the Finger barrier part of test lead.
Never attempt to use the instrument if its surface or your hand is wet.
Do not open the battery cover and the instrument case when making measurement.
WARNING
Never attempt to make any measurement if any abnormal conditions are noted, such as broken case, cracked test leads and exposed metal parts.
Do not turn the Function Selector Switch with test leads connected to the instrument.
Do not install substitute parts or make any modification to the instrument.
Do not try to replace the batteries if the surface of the instrument is wet.
Make sure to disconnect test leads from the device under test when opening the Battery cover for battery replacement.
Always attach the Cap onto the tip metal parts when using the test lead in the Measurement Category III. (CAT. III) or higher environments.
When the test leads are connected to the instrument, the lower category (voltage) either of them belongs to is applied.
12
CAUTION
Always make sure to check the Function Selector Switch is setting to the appropriate range before starting measurement.
Do not expose the instrument to the direct sun, high temperature and humidity or dewfall.
When the instrument will note be in use for a long period, place it in storage after removing the batteries.
Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use abrasives or solvents.
Allowable measurement time at 10A range of current function is 15 seconds. If you carry on making measurement continuously over 15 seconds, it can cause
instrument damage.
Please refer to following explanation of the symbols marked on the instrument or in the manual.
Symbols
Ground
AC
AC and DC
Resistance
Diode
Buzzer
Capacitor
HzFrequency
F Double or reinforced insulation
DC
C/
FTemperature
Measurement category
CAT.Primary electrical circuit of equipment with power cord for connection to outlet.
CAT.Primary electrical circuit of the equipment, which is supplied power from a distribution board, and cable run from a distribution board to an outlet.
2. Features
This instrument, KEW1011, is a Digital Multimeter which designed for making measurement of the low-voltage equipment.
(1) Designed to international safety standards
(6) Auto-ranging function
IEC 61010-1 Measurement CAT. 300V Pollution degree 2 (7) Frequency measurement function
IEC 61010-031 (Requirements for the hand-held probe)
(8) DUTY measurement function
(2) REL function to check the difference of measured values
(Express Pulse width / Pulse period as a percentage)
(3) Auto power off function to save battery consumption
(9) Temperature measurement function
(4) Data hold function
(10) Current function is protected by fuse
(5) Diode and Continuity check function
(11) Instrument is protected from mechanical shock by holster
3. Specification
Measuring ranges and accuracy (235, under the 45%75%RH)
Function
Ranges
600.0mV
6.000V
60.00V
600.0V
600V
Measuring range
Accuracy
0600V
(5 Auto ranging)
0.5%rdg2dgt
0.8%rdg3dgt
13
Note
6.000V
60.00V
600.0V
0600V
(4 Auto ranging)
600V
Resistance
600.0A
6000A
60.00A
600.0A
6.000A
10.00A
600.0A
6000A
60.00mA
600.0mA
6.000A
10.00A
600.0
6.000k
60.00k
600.0k
6.000M
60.00M
Ranges
06000A
(2 Auto ranging)
0600mA
(2 Auto ranging)
010A
(2 Auto ranging)
06000A
(2 Auto ranging)
0600mA
(2 Auto ranging)
010A
(2 Auto ranging)
Capacitance
Frequency
DUTY
2.0%rdg5dgt
1.5%rdg4dgt(50/60Hz)
1.7%rdg4dgt(40400Hz)
2.2%rdg5dgt(50/60Hz)
2.5%rdg5dgt(40400Hz)
Test condition
40.00nF
400.0nF
4.000F
40.00F
400.0F
4000F
10Hz
100Hz
1000Hz
10kHz
100kHz
1000kHz
10MHz
2.0%rdg3dgt
Description
Function
1.2%rdg3dgt
1.0%rdg2dgt
060M
(6 Auto ranging)
Measuring range
Note
3.0%rdg10dgt
2.5%rdg5dgt
0.01nF4000uF
(6 Auto ranging)
5.0%rdg10dgt
1Hz10MHz
(7 Auto ranging)
0.1%rdg5dgt
0.199.9%
(Pulse width/Pulse period)
14
Ranges
C/
F
Standards
Mode of operation
Indication
Over range display
Auto-ranging
Sampling rate
Operating Environmental
conditions
Temperature & Humidity range
(guaranteed accuracy)
Operating Temperature &
Humidity range
Storage Temperature &
Humidity range
Insulation Resistance
Withstand Voltage
Overload Protection
Dimensions / Weight
Power source
Accessories
Fuse
Accuracy
500
5832
F
0150
32302
F
150700
3021292
F
Test condition
Note
1/1
F
NiCr-NiSi sensor
If can not display bar graph on temperature
range
Overload protection: 600V DC or AC rms
IEC61010-1 Measurement CAT. 300V, pollution degree 2 / Measurement CAT. 600V, pollution degree 2
IEC61010-031
IEC61326
mode
LCD maximum value 6000 (ACV/A, DCV/A, ) / maximum value 9999 (Hz) / maximum valve 4000 (F), units, symbols
"OL" symbol is displayed on the LCD.
"OL" displayed when exceeding the measuring ranges. (except for AC/DC 600V ranges and AC/DC 10A ranges and Temperature
function.)
Range shifts to upper range when indicated value is more than 6040.
Range shifts to lower range when indicated value is less than 560.
approx. 400ms
indoor use
altitude up to 2000m
235
0+40
CAUTION
The voltage shown above is the overload protection(overvoltage protection)for the instrument. Make sure not to exceed the value of voltage shown above.
In locations subject to radio frequency interference, the product may malfunction and it will recover automatically after the test.
15
4. Instrument Layout
MIN/MAX Key
SELECT Key
RANGE Key
REL Key
HOLD Key
Hz/DUTY Key
Display
Test Leads
Holster
5. Preparation
5-1
6. Measurements
6-1 Voltage Measurement (DCV, ACV)
DANGER
To avoid the danger of getting electrical shock, never make measurement on a circuit over 600V AC/DC. (electrical potential to ground 300V AC/DC)
Do not operate Function Selector Switch during measurement.
Do not make measurement when opening the battery cover and the instrument case.
6-1-1 DC Voltage Measurement (DCV)
(1) Insert the black test lead plug into the COM terminal and the red test lead plug into the VHz terminal.
(2) Set the Function Selector Switch to " " position.(Then, "DC", "Auto" and "mV" symbols are indicated on the display.)
(3) Connect the black test lead to the negative side of the circuit under the test and the red test lead to the positive side of the circuit, then the measured value
is indicated on the display. If you connect the test leads the other way, "" symbol is indicated on the display.
6-1-2 AC Voltage Measurement (ACV)
(1) Insert the black test lead plug into the COM terminal and the red test lead plug into the VHz terminal.
(2) Set the Function Selector Switch to " " position.(Then, "AC", "Auto", and "V" symbols are indicated on the display.)
(3) Connect the test leads to the circuit under test.
16
Make sure that the "OL." symbol is indicated on the display at this bout, then short the test lead tips and check "0" is indicated on the display and check if
the buzzer beeps.
(4) Connect the test leads to both ends of the resistance under test.
Measured value is indicated on the display. The buzzer beeps below about 100.
Note) Even if short the test lead tips, indicated value may not be "0". But this is because of the resistance of test leads and not a failure. In that case, by pressing
"REL" Key, "0" will be indicated.
6-3-4 Capacitance Measurement
DANGER
To avoid the danger of getting electrical shock, never make measurement of the circuit in which electric potential exists.
Do not make measurement when opening the battery cover and the instrument case.
Make sure to discharge the capacitor before making measurement.
(1) Insert the black test lead plug into the COM terminal and the red test lead plug into the VHz terminal.
(2) Set the Function Selector Switch to "" position.(Then, "Auto" and "M" symbols are indicated on the display.)
(3) Press the "SELECT" Key three times and set the instrument to Capacitance Measurement mode. (Then, "Auto" and "nF" symbols are indicated on the display.)
(4) Press the "REL" Key and "0" shall be indicated.(Then, "" symbol is indicated on the display.)
(5) Connect the test leads to both ends of the resistance under test.
Measured value is indicated on the display.
Measuring unit "nF" / "uF" is automatically chosen and indicated due to the measured value.
Note) It may take some time according to the measuring capacitance.
Measuring capacitance < 4uF ---------- Measuring time is about 2seconds
Measuring capacitance < 40uF ---------- Measuring time is about 7seconds
Measuring capacitance < 100uF ---------- Measuring time is about 15seconds
6-4 Frequency Measurement
DANGER
To avoid the danger of getting electrical shocks, never make measurement on a circuit over 600V AC/DC. (electrical potential to ground 300V AC/DC)
Do not operate Function Selector Switch during measurement.
Do not make measurement when opening the battery cover and the instrument case.
(1) Insert the black test lead plug into COM terminal and the red test lead plug into the VHz terminal.
(2) Set the Function Selector Switch to "Hz" position.(Then, "Auto" and "Hz" symbols are indicated on the display.)
(3) Connect the test leads to the circuit under test.
Measured value is indicated on the display.
Frequency can be measured at ACV and ACA functions by pressing "Hz/DUTY" Key.
Concerning with the direction for use of "Hz/ DUTY" Key, please reference to the section7-6 Hz/DUTY in this document.
Note) The minimum input can be measured is approx. 1.5V.
To measure a frequency, measure the voltage on the electrical circuit in advance. Then press the "Hz/DUTY" Key to enter into frequency measurement.
Readings of frequency may fluctuate or be influenced under noisy environment.
6-5 Temperature Measurement
(1) Connect the Banana Pin of Temperature Probe with "-" mark to COM terminal and one with "+" mark to VHz terminal respectively.
(2) Set the Function Selector Switch to desired "/
F" position.
(3) Press "SELECT" Key to choose or
F measurement.
(4) Put the sensor probe into the temperature field under measurement.
(5) Read the result from the LCD panel.
19
20
Screw
10. Maintenance