DP3300 Operation Manual Basic CE 1.3
DP3300 Operation Manual Basic CE 1.3
DP3300 Operation Manual Basic CE 1.3
Operation Manual
[Basic Volume]
2005-2006 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
Product Information
Product Name: Digital Ultrasonic Diagnostic Imaging System
Model: DP-3300
Issued date of this manual: 2006-07
Version: 1.3
,
,
,
,
are the registered trademarks or
trademarks owned by Mindray in China and other countries. All other trademarks that
appear in this manual are used only for editorial purposes without the intention of improperly
using them. They are the property of their respective owners.
I-1
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national and
local requirements;
WARNING:
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
z
any Mindray product which has been subjected to misuse, negligence or accident;
any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
I-2
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to Mindray,
the following procedure should be followed:
1. Obtain return authorization: Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must appear on
the outside of the shipping container. Returned shipments will not be accepted if the
Mindray number is not clearly visible. Please provide the model number, serial number,
and a brief description of the reason for return.
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges).
3. Return address: Please send the part(s) or equipment to the address offered by
Customer Service department
Company Contact
Manufacturer:
Address:
Tel:
Fax:
EC-Representative:
Address:
Tel:
0049-40-2513175
Fax:
0049-40-255726
I-3
Important Information
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
The responsibility for maintenance and management of the product after delivery resides
with the customer who has purchased the product.
The warranty does not cover the following items, even during the warranty period:
(1) Damage or loss due to misuse or abuse.
(2) Damage or loss caused by Acts of God such as fires, earthquakes, floods, lightning,
etc.
(3) Damage or loss caused by failure to meet the specified conditions for this system,
such as inadequate power supply, improper installation, or unacceptable
environmental conditions.
(4) Damage or loss due to use outside the territory in which the system was originally
sold.
(5) Damage or loss involving system purchased from a source other than Mindray or its
authorized agents.
This system shall not be used by persons other than fully qualified and certified medical
personnel.
Do not make changes or modifications to the software or hardware of this product.
In no event shall Mindray be liable for problems, damage, or loss caused by relocation,
modification, or repair performed by personnel other than those designated by Mindray.
The purpose of this system is to provide physicians with data for clinical diagnosis.
The responsibility for diagnostic procedures lies with the physicians involved. Mindray
shall not be liable for the results of diagnostic procedures.
Important data must be backed up on external recording media such as clinical records,
notebooks etc.
Mindray shall not be liable for loss of data stored in the memory of this system caused by
operator error or accidents.
This manual contains Warnings regarding foreseeable potential dangers. Be alert at all
times to dangers other than those indicated. Mindray shall not be liable for damage or
loss that results from negligence or from ignoring the precautions and operating
instructions contained in this operation manual.
On the occasion of change of the administrator or manager for this system, be sure to
hand over this operation manual.
I-4
Introduction
This operation manual describes the operating procedures for the diagnostic ultrasonic
imaging system DP-3300. To ensure safe and correct operation of the system, carefully read
and understand the manual before operating the system.
1. Notation Conventions
In this operation manual, the following words are used in addition to the signal words related
to the safety precautions (refer to "Safety Precautions"). Please read this operation manual
before using the system.
NOTE:
2. Operation Manuals
A Mindray service person or instructor will explain the basic operating procedures for this
system at the time of delivery. However, read this operation manual carefully before using
the system in order to understand the detailed operating procedures, functions, performance,
and maintenance procedures. The organization of the documents supplied with this system
is shown below:
Main unit operation manual
NOTE:
IMPORTANT!
1.
2.
No part of this manual may be copied or reprinted, in whole or in part, without written
permission.
The contents of this manual are subject to change without prior notice and without our
legal obligation.
I-5
Safety Precautions
1.
DANGER,
WARNING,
CAUTION and NOTE are used regarding safety and other important instructions. The
signal words and their meanings are defined as follows. Please understand their meanings
clearly before reading this manual.
Signal word
DANGER
WARNING
CAUTION
NOTE
2.
Meaning
Description
Type-BF applied part
Note All ultrasound transducers can be connected to this system are
Type-BF applied parts.
"Attention" indicates the points requiring attention. Be sure to read the
operation manual concerning these points before using the equipment.
S-1
3.
Safety Precautions
Please observe the following precautions to ensure patient and operator safety when using
this system.
DANGER:
WARNING:
1 Do connect the plug of this equipment and its peripheral equipments
to the wall receptacle meeting the rating nameplate. Using adapter or
multi-functional
receptacle
may affect
system
grounding
performance and thus cause leakage current exceeding safety
requirement. Please use the supplied power cable. No other power
cables should be used.
2 Be sure to connect the potential-equalization lead wire before
inserting the equipment power plug into the receptacle. Also, be
sure to remove the equipment power plug from the receptacle before
disconnecting the wire to avoid electrical shock.
3 Connect the earth conductor only before turning ON the system.
Disconnect the grounding cable only after turning OFF the system.
Otherwise, electric shock may result.
4 For the connection of power and grounding, follow the appropriate
procedures described in this operation manual. Otherwise, there is
risk of electric shock. Do not connect the grounding cable to a gas
pipe or water pipe, otherwise functional grounding may not be
effective or there may be risk of a gas explosion.
5 Do not connect this system to outlets with the same circuit breakers
and fuses that control current to devices such as life-support
systems. If this system malfunctions and generates an overcurrent,
or when there is an instantaneous current at power ON, the circuit
breakers and fuses of the buildings supply circuit may be tripped.
6 No waterproof device is applied to this equipment. Do not use this
equipment in any place with the possibility of water ingress. There is
risk of electric shock if any water is sprayed on or into the
equipment. If carelessly spray any water onto the equipment contact
the Mindray sales office, customer service department or
representative.
7 Use the transducer carefully. In case that the body contacts the
scratched transducer surface, immediately stop using the transducer
and contact the Mindray sales office, customer service department or
representative. There is risk of electric shock if using the scratched
transducer.
8 After the sterilization or disinfection of accessories, chemicals must
be washed out must be discharged thoroughly from the accessories.
Remaining residual chemicals or gases will not only result in damage
to the accessories but also can be harmful to human bodies.
9 Do not allow this system or other equipment to come into contact
with the patient. If this system or other equipment is defective, the
patient may receive an electric shock.
S-2
10. Do not use the transducers other than those specified by Mindray. If
a transducer other than those specified by Mindray is connected, the
equipment and the transducer may be damaged, causing an accident
such as a fire in the worst case.
11. Do not subject the transducers to knocks. Use of defective
transducers may cause an electric shock.
12. Do not open the shell or front panel. If open the shell when the
machine is powered on, there may be a short circuit or electric
shock.
13. Do not use this system at the same time with other equipment such
as electric knife, high-frequency therapy equipment and defibrillator,
etc., Otherwise there is a danger of electric shock.
14. Precautions during transportation: When moving the equipment, first
turn it off and close up the keyboard, then disconnect it with other
devices(including transducer), disconnect it with power supply, and
wrap the power cable on the winding rack, finally lift it carefully by
handle and move it to a proper position.
15. Prolonged and repeated use of keyboards can result in hand or arm
nerve disorders in some individuals. Observe the local institution
work safety/health regulations on keyboard use.
16. Accessory equipment connected to the analogue and digital
interfaces must be complied with the relevant IEC standards.
Furthermore all configurations should comply with the standard
IEC60601-1-1. Anyone who connects additional equipment to the
signal input part or signal output part configures a medical system,
and is therefore responsible that the system complies with the
requirements of IEC60601-1-1. If in doubt, consult the technical
services department of your local representative.
CAUTION:
1Precautions concerning clinical examination techniques
(1) This system must be used only by medical personnel fully
trained in clinical examination techniques.
(2) This operation manual does not describe clinical examination
techniques. Selection of the proper clinical examination
technique must be based on specialized training and clinical
experience.
2Malfunctions due to radiowaves
(1) Use of radiowave-emitting devices in the proximity of this
kind of medical electronic system may interfere with its
operation. Do not bring or use devices which generate radio
waves, such as cellular telephones, transceivers, and radio
controlled toys, in the room where the system is installed.
(2) If a user brings a device which generates radio waves near
the system, they must be instructed to immediately turn OFF
the device. This is necessary to ensure the proper operation
of the system
3 Precautions concerning installation and movement of the system
(1) Do not place any objects on top of the system. They may fall,
causing injury.
S-3
(2) Confirm that the peripheral units are secured before moving
the system. Otherwise, the peripheral units may fall and
cause injury.
4Please use the supplied or recommended peripheral devices and
optional parts. Please use the supplied cables. Using other devices
or cables may degrade the system performance and even cause an
electrical shock.
5 Always keep the machine dry. Avoid transporting this machine
quickly from the cold place to the warm place, otherwise
condensation or water drops may be formed, which will cause short
circuit.
6 If the circuit breaker is tripped or the fuse is blown, it indicates that
the machine or the peripheral devices have problems. In these cases,
the user cannot repair by him but contact the Mindray sales office,
customer service department or representative.
7 There is no risk of high-temperature burns during routine ultrasound
examinations. To prevent high-temperature burns, do not apply the
transducer to the same spot on the patient for a long time. Apply the
transducer only for as long as required time for diagnosis.
8 Before cleaning the system, be sure to disconnect the power cable
from the outlet. If the system is defective, there is a risk of electric
shock.
9Before examining a new patient, press [Patient] key to delete the
patient information and data recorded in the image memory for the
previous patient. Otherwise, the new data may be confused with the
data of the previous patient.
10. Do not pull out the system and its accessories plug without turn OFF
the power. Doing so may cause these equipment damaged even
electric shock.
11. Do not turn OFF the power supply of the system during printing,
saving, or invoking. Otherwise it may result in abnormality of these
processes.
12. Some USB portable hard disks must be connected to the external
powerthe external power must be compliant with the relevant local
regulations, otherwise they can not be distinguished.
13. Press SHIFT+DEL key, all comments, bodymarks, measurement
scales and general measurement data on the screen will be cleared
up.
NOTE:
1. Do not use the machine in the vicinity of strong electromagnetic field (such as
the transformer), which may affect the performance of the monitor.
2. Do not use the machine in the vicinity of high-frequency radiation source
(such as the cellular phone), which may affect the performance of the
machine or even lead to failure.
3. To avoid damaging the machine, do not use the machine in following
environment:
(1) Locations exposed to direct sunlight;
(2) Locations subject to sudden changes in temperature
(3) Dusty locations
S-4
S-5
4.
WARNING Labels
Various WARNING labels are attached to this system in order to call the user's attention
to potential hazards.
The symbol
precautions. The WARNING labels use the same signal words as used in the
descriptions in the operation manuals.
Detailed information about the WARNING labels is given in the main body of this
operation manual. Read the operation manual carefully before using the system.
The name, appearance, and the indication of each WARNING label are shown as
follows.
No.
Label
Meaning
<1>
<2>
<3>
S-6
Contents
1
2
System Configuration...........................................................................3-1
3.1
3.2
3.3
3.4
Power ON/OFF.......................................................................................6-1
6.1 Power On........................................................................................................ 6-1
6.1.1 Check the items below before turning the power ON ........................... 6-1
6.1.2 Turning On the Power........................................................................... 6-1
6.2 Power OFF ..................................................................................................... 6-2
6.3 Power OFF/ON in the case of system failure.................................................. 6-2
10 Presets .................................................................................................10-1
10.1 Introduction ................................................................................................... 10-1
10.2 Enter/Exit Preset Mode ................................................................................. 10-1
10.2.1 Enter the preset mode: ....................................................................... 10-1
10.2.2 Exit Preset mode: ............................................................................... 10-1
10.3 Display/Modify Preset Information ................................................................ 10-1
10.3.1 Procedures to modify the preset values.............................................. 10-1
13.1
13.2
13.3
13.4
13.5
13.6
13.7
14 Body Mark............................................................................................14-1
14.1
14.2
14.3
14.4
14.5
20
21
22
23
C-4
Intended Use
The DP-3300 digital ultrasonic diagnostic imaging system can be used in whole body
including abdomen, gynecology, obstetrics, small organs (breast, thyroid, testes, etc.)
ultrasonic examinations.
Contraindication: none.
1-1
Specifications
2.1 Conditions
(1) Power
Line voltage
100240 V
Line frequency
50 / 60Hz
Power consumption
150 VA
5C to 40C
Relative humidity
Atmospheric pressure
-20C to 55C
Relative humidity
Atmospheric pressure
2-1
System Configuration
Intended use
Applicable parts
35C50EB
Body surface
65EC10EB
Gynecology ,
examinations
Transvaginal,
Transrectal
75L38EB
Body surface
65C15EA
Body surface
75L60EA
Body surface
obstetrics
and
urology
Name
Model
Mobile Trolley
UMT-100
DICOM Software
Footswitch
DICOM3.0
971-SWNOM, 68-S3
3-1
System Configuration
Name
Video printer
Model
SONY UP-895MD
SONY UP-897MD
MITSUBISHI P93W
HP DeskJet 5652/5650
Business Inkjet 1200
HP Laserjet2420d
HP DeskJet6548
3-2
System Overview
NOTE: All figures in the manuals are only for reference, what are shown in the figures may be
different from the actual system.
Monitor
Front view
4-1
System Overview
Handle
Transducer holder
Control Panel
Transducer socket
Winding rack
USB port
4-2
System Overview
No.
Name
Introduction
<1> Monitor
Display
images
and
10non-interlaced VGA
<3> Handle
rear view
4-3
parameters
dialogue,
etc.
for
System Overview
No.
Name
Introduction
<3> Control port for video Use to control video printer remotely
printer
<4> DICOM port
<6> AC input
Control panel
No.
Key name
Function
<1> Patient
<2> Info.
<3> File
<4> EXAM
4-4
System Overview
Reserved
Switch the transducerunder the precondition of the system
has been configured 2 transducer sockets
<7> Freq.
<9> TGC
<10> Cine
<11> F.position
<12> IP
<13> Print
Video print
<14> Measure
<16> Comment
<17> Menu
<18> Set
<19> Change
<20> Back
<21> Trackball
<22> Gain
<23> B
Enter B mode
<24> B/B
<25> VRev
<26> M/B
<27> M
Enter M mode
<28> HRev
<31> Freeze
4-5
System Overview
4.4 Symbols
This system uses the following symbols, whose meanings are described in the table
below. For safety symbols, refer to Safety Precautions.
Symbol
Description
AC (Alternating current)
Equipotential terminal
Main switch OFF
Main switch ON
Transducer socket A
Transducer socket B
The device is fully in conformance with the Council Directive
Concerning Medical Devices 93/42/EEC. The number
adjacent to the CE marking (0123) is the number of the
EU-notified body that certified meeting the requirements of
Annex II of the Directive.
Serial number
Date of manufacture
Manufacturer
4-6
CAUTION:
5.2.1
WARNING:
Release the lock on the transducer connector. Plug the transducer connector into the
transducer socket. See the following figure.
5-1
5.2.2
Power connection
The power system for this machine must satisfy following specifications:
100240 V
50 /60 Hz
Output power of the supply system >150 VA
5.3.2
Grounding Terminal
The power cable of the machine is three-wire cable. The grounding terminal should
be connected to the grounding protection phase of the power system. Ensure the
normal function of the grounding protection phase of the power system.
Connect the power plug to an outlet of the system. By doing this, the protective
earth line is connected.
5-2
WARNING:
5.3.3
Equipotential Terminal
is the equipotential terminal, used to balance the grounding protection potential
between this equipment and other electric devices.
WARNING:
1. Be sure to connect the potential-equalization lead wire before
inserting the equipment power plug into the receptacle. Also, be
sure to remove the equipment power plug from the receptacle before
disconnecting the wire to avoid electrical shock.
2. When there is other device connected to the equipment, the user
should use the equipotential cable to connect each equipotential
terminal, otherwise electric shock may happen.
3. Connect the earth conductor only before turning ON the system.
Disconnect the grounding cable only after turning OFF the system.
Otherwise, electric shock may result.
4. Do not connect this system to outlets with the same circuit breakers
and fuses that control current to devices such as life-support
systems. If this system malfunctions and generates an overcurrent,
or when there is an instantaneous current at power ON, the circuit
breakers and fuses of the buildings supply circuit may be tripped.
5.4.1
1.
2.
3.
4.
5.
Turn off the ultrasonic diagnostic imaging system and the video printer.
Connect the VIDEO IN port of the video printer and the Video output port of the
ultrasonic diagnostic imaging system by the data cable of the printer.
Connect the REMOTE port of the video printer and the REMOTE port of the ultrasonic
diagnostic imaging system by the remote cable of the printer.
Connect the power cable of the ultrasonic diagnostic imaging system.
After turning on the printer and the system, the printer can work normally.
5.4.2
1.
Turn off the ultrasonic diagnostic imaging system and the graph/text printer.
5-3
Connect the USB port of the printer and the USB port of the ultrasonic diagnostic imaging
system by the USB data cable of the printer.
Connect the power cable of the printer to the power supply net.
After turning on the printer and the system, the printer can work normally.
If the printers cannot be connected, set the printer type in the general preset dialog box.
For detailed operations, please refer to section 1.3 of the operation manuals Advanced
volume.
5-4
Power ON/OFF
6.1 Power On
6.1.1
6.1.2
Check all the power supplies and connecting cables for any abnormity such as
scratch or crack.
Check the control panel, display and the shell of the equipment for any abnormity
such as crack.
Check the transducer and the connecting parts for any abnormity such as scratch or
fall-off.
Check the outlet of the auxiliary power supply of this equipment and all I/O ports to
ensure that there is no abnormity such as damage or occlusion by foreign objects.
Turn on the power of the equipment (The power switch is on rear panel). The
start-up screen is firstly displayed. After 15 seconds or so, the menu and the image
are displayed. Check if the equipment is started up normally.
Check the transducer surface in the process of application for abnormal heat.
WARNING Using the transducer giving abnormal heat may burn the patient.
When turn on the system power and switch transducer, the sound of pi pa
indicates the system in normal state.
NOTE:
(1) Check the image for any abnormity such as abnormal noise or flicker.
(2) Check the control panel and ensure that the keys and rotary knob can function
normally.
6-1
Power ON/OFF
6-2
7
WARNING:
Check item
The temperature, humidity, and atmospheric pressure should meet the conditions of use.
There should be no deformation, damage, or stains to the system and peripheral units.
* If any stains are present, perform cleaning referring to subsection 19.1"Cleaning the
system".
There should be no damage to cables(such as power cable, etc.) and no looseness in the
connectors.
There should be no obstacles in the area around the system and ventilation opening.
Cleaning, disinfecting, or sterilizing the transducer. (Please refer to the operation manual
provided with the transducer.)
10
7-1
Check item
The acoustic lens surface of the transducer should not be unusually hot.
(Perform check by hand.)
7-2
<2>
<4>
<7>
<5>
<8>
<16>
<9>
<15>
<17>
<10>
<14>
<12>
<11>
<13>
In correspondence with the numbers in the figure above, each item is described as follows:
No.
Introduction
<1>
Manufacturers logo
<2>
<3>
<4>
<5>
<6>
<7>
<8>
<9>
8-1
<11>
<12>
<13>
<14>
<15>
<16>
<17>
Image area
B Mode:
B+B Mode:
8-2
M Mode:
M+B Mode:
Function
Command
Number
Switch
Character
Sub-menu
8-3
8-4
8-5
Page 1
Page 2
Page 3
Dyn Rng
Edge
Smooth
Adjust Value
Content
Frame Avg
Soften
Scan Mode
Gray Map Map1 Post Proc
Curve
Rejection
0
Ok
Cancel
Page Button
Edit Bar
Adjust Button
Command
Button
Help Bar
Part
Description
Title Bar
Page
Some dialog boxes have too much data to be put in the dialog
box. In this case, the system will divide these data into different
pages based on their content. But some other dialog boxes have
no page.
Content
[Ok] and
[Cancel]
Help Bar
The Help Bar is located in the bottom part of each dialog box, in
which the user can obtain some Prompt Information about the
operation.
8-6
Roll the trackball, there is a rectangle frame as big as dialogue box moving along with
the cursor, anchor the rectangle frame to the position where the dialogue box will be
moved,
3.
Press Set key again, the dialogue box automatically moves to where the rectangle
4.
frame is placed.
After step 2, press [Back] key, cancel the operation of dragging the dialogue box, the
dialogue box remains where it is.
8-7
Examination Beginning
Clinic No.
Inpatient No.
Diagnosis
reference
1.
Please press the Set or [Back] key to change M and F in Sex column.
2.
Enter the data in corresponding column, and then select [Ok] to affirm the data.
The characters that can be entered for ID isEnglish letters, numbers 0-9 and - can
be entered, up to 12 English characters. ID value can be empty.
3.
When [Ok] or [Cancel] is selected, the system exits the dialogue box.
9-1
10
Presets
10.1 Introduction
Preset function is used to set the system operating environment and status. The preset
values are saved in the memory inside the system, which will not be lost if power-off
occurs so as to ensure that the system operates in the user-desired status automatically
after each start-up. This chapter gives introduction about how to make system
configuration through using the preset menu in preset mode. Please refer to the
[Advanced volume] of operation manual for the detailed operation.
10-1
Presets
Procedures:
1.
Select the corresponding item, press the Set key to call up the corresponding
2.
3.
Use the Set or the [Back] key to adjust the parameter. At this time, some help
4.
5.
To cancel the modifications, just press the Set key on the [Cancel] button. This
action at the same time closes the dialog box.
6.
Move the cursor to [Return] item in the PRESET menu; press the Set key to
close the PRESET menu. The system exits the Preset mode and begins run
according to the modified preset parameters.
10-2
11
The keys on the control panel and menu items realize the image adjustment function. The
values are displayed on the right of the menu items or in the parameter area of the top of the
screen. The parameters displayed on the screen are shown in the figure below.
For the parameters adjusted by B MODE MENU or M MODE MENU, refer to the corresponding
menu items.
11-1
11.3 TGC
TGC, namely depth section gain compensation curve. Move the corresponding TGC slider
on the control panel to adjust the TGC of the corresponding scanning depth.
When adjusting TGC, the TGC curve appears automatically on the left part of the screen,
which will change as the slider moves. See figure below.
And 1.5 second after stop adjusting TGC, the TGC curve will disappear automatically.
TGC adjustment being made while the image is in frozen status is invalid temporarily, which
however will become valid after the image is unfrozen.
11-2
11.4 IP
The image processing parameterIP is adjusted by the ship-like IPkey.
Central
frequency(MHz)
35C50EB
3.5
5.0
3.5
2.5
65EC10EB
75L38EB
65C15EA
75L60EA
6.5
7.5
6.5
7.5
8.0
10
8.0
10
6.5
7.5
6.5
7.5
5.0
5.0
5.0
5.0
Frequencies (MHz)
11-3
11.7.1 Zoom
When theZoom/Depth indicator lights on, the image magnification is adjusted by the
ship-like key. The image magnification ranges from 100 to 400area.
Procedures for image zooming adjustment:
(1) Press Zoom/DepthZoom/Depth indicator lights on, a view frame for image zooming
appears in the center of image window, shown in figure below. If it is in B/B mode, there
is only zooming function of real-time response. Zoom/Depth indicator lights on,
indicating the ship-like key is in the status of zooming multiple adjustment.
(2) Roll the trackball; use the view frame to select the center of image to be zoomed.
11-4
(3) Press or of the ship-like key to change the zooming multiple, the size of view
frame will change with it. Press , the view frame will shrink, zooming multiple will
increase; Press , the view frame will expand, zooming multiple will decrease.
(4) Press Setkey, the view frame disappears, the zoomed image is displayed on the
screen.
(5) Press or of the ship-like key to change the image magnification
(6) Press Setkey again, the position of zoomed image is fixed, the cursor appears.
(7) At this time the image magnification can also be changed by pressing or of
the ship-like key.
(8) Press Zoom/Depth again, Zoom/Depth lamp lights off, image zooming status exits,
image in normal scale is resumed to be displayed.
The real-time image, frozen image and CINE review image can all be zoomed, and the
zoomed image can be measured, or inserted comment or Body Mark.
11.7.2 Depth
Confirm that the Zoom/Depth lamp lights up and then adjust the ship-like key to change
the imaging depth. Depth can be adjusted with range of 2.16 cm 24.8 cm according to
different transducers.
The depth cannot be adjusted when the image is in frozen status.
Press
key to reverse horizontally.
The status symbols in the upper left corner of the image window have following meanings:
means that the first scanning line on the left is the start scanning point of the
transducer; means that the first scanning line on the right is the start scanning point
of the transducer.
11-5
11.11 Smooth
Smooth function is provided to suppress the image noise and apply axial smooth processing
to the image in order to make the tissue look smoother. The adjusting range is 03. 0
represents minimum smooth while 3 the maximum.
The smooth of either B-mode image or M-mode image can be adjusted through Smooth
menu item in their respective menus. The current value of smooth of B or M-mode image is
also displayed in the menu item.
Smooth cannot be adjusted when the image is in frozen status.
11.13 M Speed
M Speed is provided to adjust the refresh speed of M-mode image.
Its range is 14. 1 indicates slowest scanning speed while 4 the fastest.
M Speed function is valid only on M-mode image. You can adjust it throughM Speed item
in M MODE MENU. The current value of M Speed is also displayed in the menu item.
M Speed cannot be adjusted when the image is in frozen status.
11-6
11-7
12
12.1 Introduction
CAUTION:
When an image is frozen, the images before freezing can be played back and edited
immediately. This function is called Cine Review. The stored cine images can be cleared
by turning OFF the power or unfreezing the frozen image.
In the high-density scan mode, the memory can store up to 64 B mode images. In the high
frame rate scan mode, the memory can store up to 128 B mode images. Therefore, the 64th
or the 128th image is always the latest image while the first image is always the earliest
one.
Manual Review
key to freeze the image, the B mode FREEZE MENU pops up.
Roll the trackball to call up the stored CINE images on the screen in turn. Roll the trackball
to the right and the images are displayed in the ascending order of the number of frames.
Otherwise, the images are displayed in the descending order of the number of the frames.
The indicating bar of CINE review on the screen shows the CINE review, the serial number
of the current frame, and the total number of frames. The arrow in the bar refers to the
direction that the images are played.
In Manual Review mode, press the [Cine] key, the system exits Manual Review mode.
12.2.2
Auto Review
After the system exits manual review status, click the [Review/Stop] item in the B mode
FREEZE MENU, the system displays the stored images automatically in the ascending
order of the number of frames.
Click the [Review/Stop] item again to stop Auto Review process.
Before or during the Auto Review, click the [Speed] item in the B mode FREEZE MENU to
12-1
12.2.3
Others
The magnified images can also be stored in the CINE review memory, which can be
reviewed after being frozen. The method to review the magnified images is the same as that
to review ordinary CINE images.
The images in the CINE review memory can be magnified, whose gray map can be
adjusted. Also the user can perform measurement, add annotation and Body Mark on the
CINE review images.
12-2
13
Text comments and arrows can be added to the image by typing character keys or by
pressing the Set button. Text comments can also be inputted from the comment library.
WARNING:
13-1
Press Setkey to add an arrow, there is a frame around the arrow, indicating selection
status, in which orientation can be adjusted and arrow can be deleted.
(3) Presskey or key to clockwise adjust the orientation of arrow: press once, the
angle changes 45; presskey or key to counterclockwise adjust the
orientation of arrow: press once, the angle also changes 45.
(4) Press Setkey to confirm the orientation of arrow and exit the arrow selection status.
(5) Repeat all the steps above to go on adding arrows.
13-2
Full description
Liver
LL
RL
CL
LTH
VL
Venous Ligament
PV
Portal Vein
HV
Hepatic Vein
RHV
MHV
LHV
HA
Hepatic Artery
HD
GB
Gallbladder
13-3
Sp
Spleen
SpA
Splenic Artery
SpV
Splenic Vein
Pancreas
PH
Pancreatic Head
PB
Pancreatic Body
PT
Pancreatic Tail
PD
Pancreatic Duct
Kidney
AG
Adrenal Gland
RA
Renal Artery
RV
Renal Vein
RP
Renal Pelvis
RC
Renal Calices
Pr
Pyramid
RCo
Renal Column
Ur
Ureter
Bl
Bladder
Pro
Prostate
SV
Seminal Vesicle
Sto
Stomach
Ca
Cardia
Esophagus
Bo
Bowel
Du
Duodenum
Co
Colon
Ap
Appendix
SMA
SMV
Ao
Abdominal Artery
IVC
Obstetrics
Symbol on screen
Full description
Ut
Uterus
Ov
Ovary
13-4
Cervix
Vagina
En
Endometrium
IUD
GS
Gestational Sac
Embryo
Embryo
YS
Yolk Sac
Am
Amnion
Pl
Placenta
UC
Umbilical Cord
AF
Amniotic Fluid
Fetus
FH
Fetal Head
F_Sp
Fetal Spine
F_Sto
Fetal Stomach
FK
Fetal Kidney
F_Lb
Fetal limbs
Symbol on screen
Full description
Thy
Thyroid
MG
Mammary Gland
Eye
Eye
Ts
Testicle
Ep
Epididymis
LyN
Lymph Node
CCA
IJV
ICA
ECA
VA
Vertebral Artery
IIA
IIV
EIA
EIV
FA
Femoral Artery
FV
Femoral Vein
Small Parts
13-5
Others
Symbol on screen
Full description
left
right
up
down
Anterior
anterior
Posterior
posterior
male
female
Lesion
Symbol on screen
Full description
Mass
Tumor
Sc
Scar
St
Stone
Cy
Cyst
Abs
Abscess
Hma
Hematoma
Eff
Effusion
Asc
Ascites
Nec
Necrosis
Sed
Sediment
Meta
Metastasis
Cal
Calcification
Hcc
Hepatocarcinoma
Ang
Angioma
Polyp
Polyp
As
Ascaris
FB
Foreign Body
Tb
Tuberculosis
Fe
Fecalith
Th
Thrombus
Plaque
Plaque
13-6
Myoma
HM
Hydatidiform Mole
Any
Anencephaly
Hyd
Hydrocephalus
SB
Spina Bifida
VSD
ASD
PDA
MS
Mitral Stenosis
MR
Mitral Regurgitation
MVP
MVV
LAM
Pe
Hydropericardium
AAn
Aortic Aneurysm
Asa
AS
Aortic Stenosis
PS
Pulmonic Stenosis
13-7
14
Body Mark
14.1 Introduction
Body Mark is used to point out the body part being examined and the detecting direction of
the transducer. In fact the Body Mark acts as a comment on the image.
Classification of Body Marks:
Four categories are available: abdomen, gynecology, obstetrics and small parts. Each
category has some different Body Marks. See the following figures.
14-1
Body Mark
Press Body Mark key, Body Mark lamp lights up, the dialogue box of Body Mark pops
up in the center of screen.
2.
Move the cursor to a Body Mark, there is around it to highlight the display, press
Set key, the dialogue box closes, a Body Mark is added in the left lower corner of the
window.
3.
In the step 2, if there is no Body Mark needed in the current page, move the cursor to
other page button, press Set key, open the page to look for the Body Mark needed,
operation is the same as step 2;
4.
Roll the trackball, place the transducer of Body Mark in the correct detecting position;
5.
Press
6.
PressSet key, confirm the direction and position of the transducer, add the Body Mark,
Body Mark lamp lights off, the system exits Body Mark mode.
14-2
Body Mark
7.
After the step 2 or step 3, if the direction and position of the transducer need not to
modify, directly press Body Mark key or Set key to confirm the Body Mark, Body
Mark lamp lights off, the system exits the Body Mark mode.
In B/B-mode, the user can add Body Marks respectively on the two images.
NOTE:
1.
Move the cursor onto the Body Mark. The cursor changes into a
. Press the
Roll the trackball, the frame is moving together. Move the to the target
position to which the Body Mark is to be moved.
3.
Press the Set key, the Body Mark is moved to the new position and the operation
4.
ends.
In step 2, press [Back] key, the Body Mark resumes to previous position and
operation ends.
For the Body Mark that has been entered, press the [Body Mark] key for consecutive
two times to clear the Body Mark.
When the dialog box of [Body Mark] is open, if no Body Mark is to be added, just
press the [Body Mark] key, or position the cursor to the [Close] key in the dialog box
and then press the Set key.
14-3
15
This chapter briefly describes the measurement functions of the system. For detailed
explanation of measurement and calculation, refer to Operation Manual (Advanced volume).
WARNING:
CAUTION: 1. When open a CIN or FRM file, the current patient data and the
measurements will be cleared up instead of the previous patient data
in the CIN or FRM file.
2. After open a CIN or FRM file, unfreezing the image will clear up all
comments, measurements, Body Marks and patient data.
3. During measurement, to unfreeze the image or change the exam
mode will clear the basic measured data and measurement scale.
4. All the measured data will be lost when the system is turned OFF or
Patient key is pressed.
Press
key to enter Measurement status. The menu on the right side of the
screen switches to Measurement menu.
The measurement menu is displayed on the right part of the screen. If the menu is not
displayed, press the [Menu] key.
There are 4 menus for B mode measurements and calculations.
B MEAS menu: used for general measurements and calculations of abdomen exam mode.
B-OB MEAS menu (Including B-OB MEAS and B-OB MEAS2 menu): used for calculations
of GA, fetal weight and EDD when the system is in obstetric exam mode.
B-GYN MEAS menu: used for uterine and ovary measurements & calculations in
gynecology exam mode.
B-ORTH MEAS menu: used for HIP measurements in orthopedics exam mode.
All of the measurement menus in B mode as follow:
15-1
M MEAS: used for the general measurements on the M mode image, such as distance,
heart rate, time, and slope.
M MEAS measurement menus as follows:
15-2
The keys used during measurement are shown in the figure, which are to be used in
conjunction with the trackball.
Set:
Used to start or end the measurement, or to anchor the two point of line measuring scale.
The function of the key is to be described detailed in following practice.
[Back]:
This key has two functions: to return to the previous step during measurement; to delete
the previous measurement.
[Change]:
Used for switching the fixed end and the active end in measurements.
Description
Distance
Area/circumference
Volume
Ratio
% Stenosis
15-3
ratio
between
two
OB measurement
GYN Measurement
ORTH measurement
Description
Distance
Time
Slope
Heart rate
15-4
16
File System
16.1 General
The file formats supported by the system:
z BMP
z JPG
z CIN
z FRM
z DCM (Only for DICOM optional unit)
Note: JPEG CODEC may result in the distortion of image.
The images need to be frozen before the FRM and CIN files are stored; however, other
types of files can be stored in any status.
Press the Filekey again, and the user can exit the file system.
16-1
File System
to section 1 in the Advanced volume of Operation Manual),. Or the user can select the
default type of storage media through [Driver] item of [File Menu].
If the default driver is the local disk, the default path is the root directory \Bof the local disk;
if the default driver is a removable disk, the default path is the root directory \Cof the first
logic driver.
Snapshot
The user can acquire a single-frame image and save it in the default path and in the set file
formats (JPG, BMP or DCM). The shortcut key is ShiftS.
The default file format setting: set it through the [Snapshot Type] preset item of the General
Preset dialog box, or set it through the [Snapshot Type] item in [File Menu].
The the default path setting: set it through the [Driver] preset item of the General Preset
dialog box, or set it through the [Driver] item in [File Menu].
1 Click [Snapshot] in the [File Menu] (or directly click the ShiftSkey), and the current
image will be saved in the default path and as the default format, the filename is the
default one.
2 During the file storage, a processing bar appears at the bottom of the screen. When the
storage operation finishes, the processing bar disappears automatically, and other
operations are to be performed.
Saving a FRM/CIN file:
The images need to be frozen before the files are stored.
Click the [Save FRM] or [Save CIN] in the [File Menu], and the user can save the FRM or
CIN files and other operations are similar to those of the Snapshot.
Shortcut key:
Saving FRM files: ShiftF.
Saving CIN files:ShiftC.
16-2
File System
2.
Press theFilekey to open [File Menu], and click the [Save As], and the dialog box for
file storage pops up.
3.
The default storage path is displayed in the dialog boxPlease refer to section 16.2 to
set default storage path, and the user can modify the storage path.
2
4
5
1driver list
2filename
4directory list
5file list
4.
Selecting a driver
Move the cursor onto the symbol , at the right of the Driver and press the
Setkey, and then a list box pops up as shown in the following figure. The drivers
which are available for the system appear in the list box: local disk B and
removable disk.
Move the cursor to select the desired driver, and press the Setkey, and after the
list box is closed, the selected driver becomes the current one.
16-3
File System
5.
Local disk: there is no subdirectory, so the files can only be stored under the root directory.
Removable disk: Subdirectories can be created in the removable disk, so the files can be
stored under the subdirectories.
Move the cursor onto the desired directory in the directory list box, and continuously press
the Setkey twice to enter the directory. To return to the previous level of the directory,
move the cursor onto the [..] and continuously press theSetkey twice.
6.
Inputting a filename
Anchor the cursor in the filename edit column, and press the Setkey, and input the
filename of the file to be stored.
To replace an existing file, move the cursor onto the corresponding file in the file list box, and
press theSetkey.
7.
The user can select a storage format through a drop-down list box:
8.
Move the cursor onto the symbolat the right of the filename, and press the
Set key, and a list box pops up. The available storage formats are displayed in
the list box: JPG/BMP/DCM/FRM/CIN.
Move the cursor to select a suitable file format, and press theSetkey to close the
list box.
ClickSeton theOKbutton and the dialog box closes, and then the system stores
the current screen information to the specified file
16-4
File System
Note:
When the FRM or CIN files are opened, the operation may clear all information of the
current patient, including measurement data, comments and bodymarks, etc. The patient
information in the Cine file or single-frame image file is to be set to the current patient
information.
The operation for opening a file is similar to that for storing an image file. Opening a FRM file
is taken as an example:
1.
2.
3.
4.
Move the cursor onto the [Open] item and press the Set key. A dialog box for
opening a file appears on the screen.
The files in the default format and in the default path appear in the dialog box. The user
can modify the driver and file directory, and the method is similar to that for general
saving a file. The user can select the format type in the drop-down list at the right of
the filename, and the files of this format type are displayed in the file list. When the
selected format type is ALL, all types of files under the current directory are
displayed in the file list.
Click the [Filename] or [Modify Time] button in the dialog box, to arrange the order of the
files in terms of filenames or modification time.
Double-click a file to be opened; or move the cursor onto the file to be opened and
press theSetkey, and the selected file is highlighted. Press the [OK] key again, and
the corresponding file is opened and the dialog box is closed. The system displays the
stored single-frame image on the screen.
5.
Press theFreezekey, and the opened FRM image file is closed, and the system
resumes the status prior to that of opening the file.
16.4 DICOM
If the system is configured with the DICOM unit, the [Send] item will appear in the [File
Menu].
Refer to the section 16.3. for Storing and Opening the DCM Files.
16-5
File System
Setkey, and the system start to send image.
2.
The prompt information at the bottom of the screen displays the working status of the
system.
Move the [Send File] item in the [Send] submenu, and press the Setkey. The
screen displays the corresponding dialog box.
2.
3.
The prompt information at the bottom of the screen displays the working status of the
system.
16-6
File System
Make a directory:
1
First select the drive in which a directory is to be created in the pull-down list of drive. Then
select the position at which the directory is to be created.
Move the cursor to the [Make] item in the dialog box and press the Set key. The Info
Enter the directory name into the dialog box. And press the Set key on the [Ok] button to
Rename a directory:
1
2
First select the drive in which a directory is to be created in the pull-down list of drive.
Then select the position at which the directory is to be created.
Press the Set key on the [Rename] button. The Info input dialog box pops up.
Enter the new directory name in the dialog box. Press the Set key on the [Ok] button to
close the dialog box and the previous name in the directory table is updated to the new
name. Or press the Set key on the [Cancel] button to cancel all operations.
Delete a directory:
1.
First select the drive containing the directory to be deleted in the drive table.
2.
3.
Press the Set key on the [Del] button. A dialog box pops up.
16-7
File System
4.
Press the Set key on the [Ok] button to close the dialog box.
5.
6.
If the directory is empty, the selected directory name is deleted. Or press the Set key
on the [Cancel] button to cancel all operations.
Opening a file:
1 Open the dialog box for the file manager, and select the directory of the file(the
operation is the same to the section16.3.3).)
2 The file list displays all the files under this directory. The user can click the[Name], [Type]
or [Modify Time] button, to arrange the order of the files in terms of the filename, type or
modification time.
3 To open a file, select the file and click the [Open] button; or double-click the file to be
opened, and then the file is opened.
Rename a file:
1 Select the drive and the directory under which the file to be renamed exists.
2 Select the file to be renamed in the file table. Press the Set on the [Rename] button,
the Info input dialog box pops up.
3 Enter the new file name into the dialog box and press the Set key on the [Ok] button
for confirmation or on the [Cancel] button to cancel all operations.
16-8
File System
Delete a file:
1 Select the drive and the directory under which the file to be deleted exists.
2 Select the file to be deleted in the file table and press the Set key on the [Del] button.
A dialog box pops up.
3 Press the Set key on the [Ok] button for confirmation or on the [Cancel] button to
cancel all operations.
2 If it needs to remove the USB storage device, move the trackball onto
Setkey, a dialogue box below pops up.
16-9
, press
File System
3 Select the USB storage device to be removed, press [OK] to remove it safely. If there
are two USB storage devices connected, the first connected memory is displayed on the
upper row, and the second is displayed on the lower row.
WARNING:
Do not remove the USB storage device directly from the ultrasound
system without performing the prescribed procedures. Otherwise, it
may damage the USB storage device and the ultrasound system.
16-10
17
Needle Guide
WARNING:
17-1
Needle Guide
WARNING:
1.
2.
If the positions of the needle and needle guide line are not consistent,
do not execute needle guide operation.
Calibrating method:
17-2
18
18-1
Transducer change
Field of view(sector
Angle or scan Width)
Change the sector angle or scan width may result in change to the
frame rate, and thus change the acoustic output power.
Changing the image depth changes the PRF, and thus changes the
acoustic output power.
Number
Zones
Since the number of focal zones influences the frame rate and the
actual position of the focal zones, changing the number of focal zones
changes the acoustic output power.
of
Focal
Focus position
The transmit focus location change will cause the acoustic output
power change, even though the transmitting electrical energy level and
the aperture remains the same. In most cases, the acoustic output
power will increase if the focal point is moved closer to the transducer.
Freeze
Active the freeze function stops the electrical energy transmit part of
the system, thus disabling the system from generating ultrasound
wave.
Transmit power
The transmit power level change will change the electrical output of the
system to the transducer, and thus change the acoustic output power.
Frequency change
Line density
Preset
Since the system and user presets contain all of the above imaging
parameters, changing the preset will change the acoustic output power.
(2)
18-2
19
Maintenance Check
The maintenance of the system is completed by customers and our service engineer. When the
system gets to the customer, the customer should assume all the responsibility in maintenance.
WARNING:
WARNING:
Before cleaning the system, be sure to turn off the power and
disconnect the power cable from the outlet. Cleaning the
machine when the power is On may result in electric shock.
19-1
Maintenance Check
CAUTION:
1.Be careful not to allow water or liquid to enter the system during cleaning
to avoid malfunctions or electric shock.
2.To clean connectors and TGC controls and other connectors for the
peripheral devices, contact the foreign sales distributor of Mindray. The
cleaning by the user may cause the failure or lower the performance of the
system.
Check item
Cleaning
Electric safety
Mechanical safety
Image recording
19.4 Troubleshooting
To ensure the normal operation of the machine, it is recommended to establish the
maintenance and check plan to periodically check the safety of the machine. If any
abnormity is detected, contact the foreign sales distributor of Mindray.
If some abnormal phenomena such as that after the start-up, there is no image, or there is
menu but no image, please troubleshoot first by referring to the table below. If the failure
keeps existence, please contact the distributor of Mindray.
19-2
Maintenance Check
Troubleshooting:
No.
1
Failure
The
power
switch is turned
on, but the
power indicator
does not light
on.
The power light
is on but no
image is
displayed.
Cause
Abnormal power system or
incorrect connection of the
power cable.
Measure
Check the power system and the
power cable to ensure they are in
normal status.
1.
1.
The monitor
shows the
character but
no image.
1.
2.
2.
3.
4
The image
quality is
abnormal.
1.
2.
19-3
2.
1.
2.
3.
1.
2.
20
Accuracy of Measurement
Value range
Error range
2.1611.9cm
Low Frequency transducers:
4% of full scale
4.3124.8cm
Range of M-mode image time
About 2, 4, 8,16s
Range
Distance/depth
Max. 248 mm
AreaTrace
Areaellipse, circle)
Max. 560cm2
Angle
0~180o
Error range
4%; or
<2mm if measured value is
less than 40 mm
8%; or
<130 mm2 if measured value
is less than 1600 mm2
8%; or
<130 mm2 if measured value
is less than 1600 mm2
3%
Value range
Distance
Max. 248mm
Time
Heart Rate
Slope
Max. 16s
15~999 beats/min.
Max. 999mm/s
20-1
Error range
4%; or
<2mm if measured value is
less than 40 mm
1%
5%
5%
Accuracy of Measurement
Table 20-4 Volume Measurements
Parameter
Volume
Value range
Max. 999cm
Error
12%; or
< 8000 mm3 if measured
value is less than 64000 mm3
NOTE: Measurements in any area of the selected viewing range can meet the required
precisions. Precisions given above are based on the system under worst conditions, or
on the actual tests of the system.
20-2
21
(1)
Safety Classification
(2)
(3)
(6)
21-1
22
WARNING:
NOTE: 1. DP-3300 should not be used adjacent to or stacked with other equipment. If
adjacent or tacked use is necessary, DP-3300 should be observed to verify
normal operation in the configuration in which it will be used.
2. DP-3300 needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided below.
3Preventing conducted RF immunity. Due to technological limitations, the conducted
RF immunity level are limited to 1Vrms level, conducted RF interference above
1Vrms may cause wrong diagnosis and measurements. We suggested that you
position DP-3300 further from sources of conducted RF noise.
4. Portable and mobile RF communications equipment can affects DP-3300. See
tables 1, 2, 3, and 4 below.
TABLE 1
GUIDANCE AND MINDRAY DECLARATIONELECTROMAGNETIC EMISSIONS
DP-3300 is intended for use in the electromagnetic environment specified below. The customer or
the user of DP-3300 should assure that it is used in such an environment.
EMISSIONS
COMPLIANCE
ELECTROMAGNETIC ENVIROMENTGUIDANCE
TEST
RF emissions
Group1
DP-3300 uses RF energy only for its internal function.
CISPR 11
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic
Emissions
IEC61000-3-2
Voltage
Fluctuations/
Flicker
Emissions IEC
61000-3-3
Class A
Compliance
22-1
IEC 60601
TEST LEVER
COMPLIANCE
LEVER
ELECTROMAGNETIC
ENVIRONMENT-GUIDANCE
Electrostatic
Discharge(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Electrical fast
Transient/burst
IEC 61000-4-4
2 kV for power
supply lines 1
kV for
input/output lines
(>3m).
2 kV for power
supply lines 1
kV for
input/output lines
(>3m).
Surge IEC
61000-4-5
1 kV differential
mode 2 kV
common mode
1 kV different
mode 2 kV
common mode
Voltage dips,
Short
interruptions
and voltage
variation on
power supply
input lines IEC
61000-4-11
<5% UT
(>95% dip in UT )
for 0.5 cycle
<5% UT
(>95% dip in UT )
for 0.5 cycle
40% UT
(60% dip in UT )
for 5 cycle
40% UT
(60% dip in UT )
for 5 cycle
70% UT
(30% dip in UT )
for 25 cycle
70% UT
(30% dip in UT )
for 25 cycle
<5% UT
(>95% dip in UT )
for 5 sec
<5% UT
(>95% dip in UT )
for 5 sec
3 A/m
3 A/m
Power
frequency
(50/60 HZ)
magnetic field
IEC 61000-4-8
22-2
IEC 60601-1-2
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT-GUIDANCE
Conduced RF
IEC
61000-4-6
3 Vrms 150kHz
to 80MHz
1 Vrms
Radiated RF
IEC
61000-4-3
3 V/m 80MHz
to 2.5 GHz
3V/m
Note
At 80 MHz and 800 MHz, the higher frequency range applies.
Note
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
z
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which DP-3300 is used exceeds the applicable RF compliance level above, DP-3300
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating DP-3300.
Over the frequency ranges 150kHz to 80MHz, field strengths should be less than 1V/m.
22-3
80MHz -800MHz
800MHz -2.5GHz
d = 3 .5 P
d = 1 .2 P
d = 2 .3 P
0.01
0.35
0.12
0.23
0.1
1.11
0.37
0.74
3.50
1.17
2.34
10
11.07
3.69
7.38
100
35.00
11.67
23.34
For transmitters at a maximum output power not listed above, the recommended separation distanced in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
If DP-3300 image distortion occurs, it may be necessary to position DP-3300 further from sources of
conducted RF noise or to install external power source filter to minimize RF noise to an acceptable level.
Note
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
22-4
23
Indication of Year of
Manufacture
23-1
P/N: 2302-20-34438V1.3