Service Manual: 4500 Series
Service Manual: 4500 Series
Service Manual: 4500 Series
SERVICE MANUAL
2009 Awareness Technology Inc. Unauthorized duplication is strictly prohibited. Information given
in this manual is restricted for use by authorized personnel and is to be used for the sole purpose of
providing routine instrument maintenance and repair service. No part of this manual may be copied
or redistributed without the express consent of Awareness Technology Inc. There is no transfer of
technology, copyright, trade name, patent, trade secret, or other proprietary right given or implied.
04/2009 Rev. A
Table of Contents
1. INTRODUCTION ................................................................................... 1
1.1 Warning Markings Inscriptions dAvertissement ................................................................. 1
1.1.1 Safety Symbols Le Symboles de Sret ......................................................................... 1
1.1.2 Safety Terms Terminologie de Sret............................................................................ 2
1.2 Layout ................................................................................................................................... 3
2. PRINCIPLES OF OPERATION ............................................................... 6
2.1 Power Requirements ............................................................................................................. 6
2.2 Voltage Supply...................................................................................................................... 6
2.3 Photometer ............................................................................................................................ 6
2.4 Display and Printer ............................................................................................................... 7
2.5 Temperature Control............................................................................................................. 7
2.6 System Control...................................................................................................................... 7
3. TROUBLESHOOTING ........................................................................... 9
3.1 Power Supply ........................................................................................................................ 9
3.2 Photometer ............................................................................................................................ 9
3.3 Temperature Control........................................................................................................... 10
3.4 Main PCB............................................................................................................................ 10
3.5 Internal Printer .................................................................................................................... 11
3.6 Display and Touch screen................................................................................................... 11
3.7 USB Connections................................................................................................................ 11
3.8 Error Messages.................................................................................................................... 13
4. SERVICE PROCEDURES..................................................................... 14
4.1 Opening the Instrument....................................................................................................... 14
4.2 Lamp Replacement ............................................................................................................. 15
4.3 Filter Replacement .............................................................................................................. 17
4.4 Filter Label.......................................................................................................................... 19
4.5 Printer Jam .......................................................................................................................... 20
4.6 Printer Repairs .................................................................................................................... 20
4.7 Printer Interface PCB Replacement .................................................................................... 20
4.8 Main PCB and/or Photometer Replacement ....................................................................... 20
4.8.1 Photometer Replacement ............................................................................................. 20
4.8.2 Main PCB Replacement............................................................................................... 22
5. Calibration ......................................................................................... 23
5.1 Software .............................................................................................................................. 24
5.1.1 Software Calibration .................................................................................................... 24
5.1.2 Restore Calibration Data.............................................................................................. 25
5.1.3 Reset Calibration Data ................................................................................................. 26
5.1.4 Software Calibration Temperature............................................................................ 27
5.1.5 Software Calibration Absorbance ............................................................................. 28
6. Replacement Parts ............................................................................ 30
6.1 Consumable Items and Accessories.................................................................................... 30
6.2 Non-consumable Items ....................................................................................................... 30
7. Block Diagram and Schematics ........................................................ 31
1. INTRODUCTION
This manual describes service and troubleshooting procedures for 4500 Series instruments. This
manual is written with the service technician in mind, and contains no information for clinical
chemistry analysis, or other applications. For details on instrument operation and specifications,
please refer to the operators manual specific to the instrument.
The 4500 Series instrument is a general purpose, microprocessor-controlled, bichromatic
photometer system. In the standard configuration, the instrument is equipped with six optical
filters and 37 deg C incubation.
Through the LCD touch screen display, the instrument prompts the operator to read the tubes in
the appropriate sequence. It then performs the necessary calculations and prints out test results.
It accepts standard 12mm round tubes, and includes an incubation block with 12 round tube
stations. Both the incubation block and the read well are temperature controlled to 37 deg C.
An internal thermal printer provides a printed record of all samples and test parameters. All
calibration data, test parameters, and other information is stored in battery-backed (non volatile)
RAM. An aspirating flowcell with control module is available as an accessory.
WARNING
AVERTISSEMENT
Protective Ground
La Terre Electrique
CAUTION
LATTENTION
BIOHAZARD
BIOHAZARD
Risk of Shock
(Earth) Terminal
Refer To Manual
Risk of Infection
Risque de Choc
Prise de Terre
FUSE: For continued protection against risk of fire, replace fuse only with one of the
specified type and current ratings. Disconnect equipment from supply before replacing
fuse.
FUSIBLE: Pour la protection continue contre le risque du feu, remplacez le fusible
seulement par une du type spcifique et des estimations courantes. Dmontez
lquipement de lalimentation dnergie avant de remplacer le fusible.
DANGER Le de
marque:
DANGER
WARNING
AVERTISSEMENT!
Le de marque:
WARNING
CAUTION
LATTENTION Le
de marque:
CAUTION
WARNING
AVERTISSEMENT!
Le de marque:
WARNING
CAUTION
LATTENTION Le
de marque:
CAUTION
BIOHAZARD
1.2 Layout
The instrument housing is comprised of a chassis (base and rear panel combined) and a formed
plastic cover. The cover is secured to the base by two #4 self-tapping screws.
An interactive 3.5 LCD touch screen color graphics display is mounted to the printer bracket
inside of the cover. A 20-column thermal printer and the printer PCB mount to a metal printer
bracket, which in turn mounts to the inside of the cover. The display and the printer assembly
both connect to the main PCB stack: LCD 54 conductor FFC connector; Printer 29 conductor
FFC connector; Touch Screen 4 conductor FFC connector.
USB Type A and Type B connections on rear panel to be used for mouse and PCB download
respectively. The USB Type A can also be used to upgrade firmware with a USB thumb drive.
Instrument Exterior:
The photometer assembly connects to the main PCB via a 14-conductor ribbon cable and DIP
plug as well as a 10-conductor rainbow ribbon cable and double row header. The incubation
block connects to the main PCB via a 4-conductor cable and SIP header. The 12VDC cooling
fan is mounted to the chassis base and connects to the main PCB via a 2-conductor cable and
SIP header.
An external +19v laptop PC type power supply is used to supply power via a connector on the
rear panel.
Calibration may be performed in software
EPROM device is an internal 4mb flash memory
RAM device is an internal 8mb static RAM memory; internal 512kb EEPROM
memory
PCB revisionsinstalled components, locations, and jumpers. Refer to the
appropriate PCB layout in the PCB Layouts section of this manual.
Rear panel with power supply attached:
A) Power supply
connection
B) USB Type A
connector
C) USB Type B
connector
A) Incubation
block
B) Main PCB
C) Filter Wheel
Motor
D) Lamp
Board to
board
connection
(reference
Figure 7)
C
D E
E) Photometer
Figure 3. Chassis Layout
Chassis bottom:
Fan Outlet
2. PRINCIPLES OF OPERATION
This section of the manual describes the operation and the interaction of the reader's main
systems. Most systems are controlled by the eZ80 on the main circuit board.
2.3 Photometer
The light from a tungsten-halogen lamp is passed horizontally though the tube or flowcell and
the sample it contains. The sealed box on the opposite side of the read well contains a
rotating filter wheel and photodiode. The filter wheel contains optical bandpass filters at
various wavelengths, and is speed controlled, via drivers U9A and U9B in parallel, to
approximately 4 rps under software control. As each filter passes in front of the photo
detector, an infrared optical switch generates a pulse (FCNT) at U13. C14 charges to the
peak voltage produced by the photo detector and amplifiers U16 and U14B.
The sampled voltage is buffered by U15A and then fed to comparator U15B which compares
the sampled voltage to the output of an exponential capacitor decay circuit built around Q3
and C16. The positive voltage at the comparator output enables a counter in the eZ80.
The photo detector output is proportional to the intensity of the light, whereas the width of the
positive phase of the comparator output is proportional to the absorbance. An oscilloscope
can be used to view the important signals at connector J36 while the instrument is in
operation. See Figure 6, Oscilloscope Waveforms.
Figure 5. Troubleshooting
3. TROUBLESHOOTING
The flowchart in Figure 5 describes the logical steps to take if the instrument is malfunctioning. It
is only a guide to troubleshooting, and is intended to assist in determining problems which are
routine. It is not intended to take the place of an experienced technician, nor does it attempt to
cover all possible problems. Note, for example, that the temperature control system is not
covered.
In the following section, each subassembly or component group and possible problems and
solutions are outlined.
3.2 Photometer
The photometer assembly can be examined as several component systems:
Read well assembly and cell heat PCB
Lamp and brackets
Photometer PCB
Filter wheel and motor.
The read well assembly serves to hold the tube centered on the cell block, between a fixed
plate and a spring loaded plate. A micro switch mounted to the cell heat PCB operates when
a tube is inserted.
The lamp is tungsten-halogen, rated at 6V 10W. The lamp bracket holds the lamp and lens in
alignment with the rest of the assembly, that is, the filament is centered on the optical path.
The filter wheel turns on a shoulder screw shaft and is driven by the filter wheel motor via a
neoprene belt. The filter wheel itself requires no adjustments.
The rotation of the photometer filter wheel is under micro-processor control. The processor
turns the wheel on and off and maintains it rotating at a constant average speed. If the
processor fails to detect motion of the wheel within 7 seconds of the command to start, the
Filter Wheel Err message occurs.
Possible causes:
The belt has come off the wheel.
The motor does not respond to drive circuit.
The drive circuit and or connections are missing or defective.
The wheel detection circuit connections are missing or faulty.
Analysis process:
Note: In general the wheel only rotates while the instrument is in a reading mode and the
lamp is on.
The photometer PCB contains a photodiode, a very sensitive opamp circuit (U16) and
phototransistors Q7 and Q8. Because the photodetector and related circuitry is solid-state, it
should require no service. Avoid unnecessary handling of the photometer PCB or removal of
the optical cover.
The lamp is intended to be maintenance free for the life of the instrument, since the lamp is
turned off automatically after 30 minutes of instrument inactivity. Under heavy usage the
lamp remains on for longer periods and so may require replacement. Also, a physical shock
to the instrument could cause filament breakage. The lamp must be replaced if output
becomes low or the lamp fails completely. To check for low or no lamp output, insert a
borosilicate tube filled with plain water into the read well. Do not use a soda-lime glass tube,
since it does not transmit at 340 nm. Select test #186. The instrument prints the wavelength,
position, and the detected voltage for each filter. The voltages should be between 2 and 10
volts. If all of the voltages are below 2.0 volts, the lamp is the likely suspect. As a final check,
measure the voltage at the lamp terminals. If the lamp voltage is much lower than 4.5 VDC,
the regulator circuitry on the main PCB (Q1 and U9) may be the problem. Otherwise, replace
the lamp. Refer to the section Lamp Replacement.
If only one or two filters report low detected voltages, degraded filters or a misadjusted lamp
may be the cause. To check for misadjusted lamp, refer to Lamp Replacement section. The
optical interference filters contained in the wheel are of metal-deposition type construction
and are intended to be maintenance free for the life of the instrument. Some older
instruments incorporated emulsion-type filters whose transmittance can be reduced
dramatically if subjected to high humidity over very long periods. Refer to the section Filter
Replacement.
10
11
12
Solution:
LAMP FAILURE
The following error messages indicate possible failure of the NV RAM U3. U3 can be
checked by substitution.
Problem:
Solution:
MEMORY ERROR
13
4. SERVICE PROCEDURES
4.1 Opening the Instrument
The cover must be removed to allow access to the inside of the instrument. Disconnect the
power cable and the serial cable (if connected) from the rear panel. Invert the instrument on
a soft nonabrasive padded surface such as a terrycloth towel, to prevent scratching. Refer to
Figure 4, Chassis Bottom. Locate and remove the (2) cover screws from the bottom of the
instrument. Do not remove or loosen any other screws. While holding the instrument cover in
place with both hands, return the instrument to the upright position.
Grasp the cover at the left side. Gently lift the cover upward and to the right, until the cabling
is visible. You can rest the cover on its side while you disconnect the cables. Refer to Figure
7, Cover Connections.
Remove the cover and set it aside.
To reinstall the cover, reverse the procedure. Position the cover on its side to the right of the
chassis base. Carefully fit the cover to the chassis base. Install the cover screws. Do not
over tighten the cover screws.
To main PCB
Reference
Figure 3,
Chassis layout
14
3. Loosen but do not remove the lamp terminal screws. Do not loosen or remove any other
screws. Remove the lamp by lifting it upward out of the connector.
4. Use a pair of pliers or tweezers to handle the new lamp. Do not handle with bare skin.
Insert the lamp leads into the connector until they bottom out. The lamp filament must be
centered on the lens and the lamp body must be parallel with the lamp bracket. While
holding the lamp in alignment, tighten the lamp terminal screws. You should be able to
visually confirm the lamp alignment by sighting through the lamp and lens to view the
aperture on the spring plate behind the lens. All three elements should be aligned. With
the newer lamp bracket (Figure 8a), you can adjust the height of the lamp using the
indicated screw to raise or lower the plate height.
5. Set the power switch to ON. Shield your eyes from the lamp beam and observe the
projection of the beam onto the aperture. Refer to Figure 9, Spot Alignment. The spot
should be small and centered on the aperture. The spot should be sufficiently large to
just encircle the aperture. If the spot is not centered, repeat step 4.
6. Insert a borosilicate tube filled with plain water into the read well. Do not use a soda-lime
glass tube, since it does not transmit at 340 nm. Select test #186. The instrument prints
the detected voltage for each filter position. All voltages should be between 2.00 volts
and 10.00 volts. If all the voltages report low, repeat step 5 until optimum lamp position is
obtained.
15
16
Procedure:
1. Open the instrument. Refer to Figure 3. Locate the photometer. Unplug both photometer
cables from the main PCB.
2. Use a pencil to mark the front and side edges of the photometer on the chassis so that the
photometer can be installed in its original position.
3. Locate the (3) #6 screws holding the photometer assembly to the chassis. From the bottom
of the instrument, remove these screws and washers while noting the screw lengths and
locations from which they were removed.
4. Refer to Figure 10, Filter Replacement. Remove the four 6-32 screws retaining the
photometer cover. Set the photometer cover aside.
17
5. Remove the two 4-40 screws securing the photometer PCB. Gently move the PCB aside
without disconnecting any wires.
6. Remove the belt from the filter wheel and pulley. Loosen the shaft (shoulder screw).
Remove the shaft and filter wheel. Note the number of nylon washers and the order of
installation.
Warning!
WARNING: Retaining rings will pop out from filter wheel when its
spring tension is released during removal. Shield the opening with
hand and wear safety glasses.
7. Refer to Figure 11. Locate the filter to be removed. There are currently two means of
filter retention in field use; silicone glue or a Retaining Ring. Refer to Figure 10 and
determine if a Retaining Ring is used. If a Retaining Ring is used, carefully pry two to
three of the retaining ring teeth away from the filter until the ring is dislodged and remove
the ring and filter. If silicone glue is present, remove the silicone and push the filter out of
the wheel using the eraser end of a pencil or other soft blunt object and the glue should
release. Remove any remaining filter components from that position in the wheel.
8. Locate the neutral density filters, dot screen, and transmittance filters that were included
with the replacement filter and drop them first into the cavity. Install the interference filter
with the mirror side down. Place the replacement retainer over the filter so that the tabs
are angled away from the filter and press in place with a 7/16 diameter wooden dowel or
similar object.
9. Install the filter wheel taking care not to pinch washers under the shaft. Reassemble the
shaft and washers as disassembled. Tighten the shaft securely. The filter wheel should
spin freely.
10. Install the filter wheel belt. Be sure that the belt is centered on the filter wheel and is not
twisted.
11. Install the photometer PCB taking care to center the board around the shaft.
12. Replace the photometer cover, taking care to position the cable in the slot of the cover.
Do not pinch the cable under the cover. Tighten cover screws only until snug. Do not
over tighten the photometer cover screws! Doing so will bend the photometer cover and
create light leakage.
13. Install the photometer subassembly to the chassis by replacing the (3) #6 self tap screws
through the chassis base in their original locations. Line up photometer with locating
marks made in step 2. Connect the (2) cables removed in step 1.
18
14. Attach the power cord and connect the instrument to the mains supply. Set the power
switch to ON (1). Insert a borosilicate tube filled with plain water into the read well. Do
not use a soda-lime glass tube, since it does not transmit at 340 nm. Select test #186.
The instrument prints the detected voltage for each filters. All voltages should now be
between 2 and 10 volts. If not, check lamp alignment. If the lamp alignment is properly
set, the neutral density filters may need to be changed. Add neutrals to lower the voltage.
Remove neutrals to increase the voltage.
19
CAUTION: Do not apply force to the print head! Do not scratch the
platen or pry with sharp objects!
20
A Gray ribbon
cable with red
stripe,
B Multi-color
ribbon cable
C Yellow wires
21
Remove
the two
screws
securing
the Main
PCB.
22
5. CALIBRATION
Each instrument is calibrated during manufacturing using standards that are traceable to the
National Institute for Standards and Testing (NIST), and is tested to verify its linearity to 2A. This
preset calibration is very stable. Absolute calibration can be verified with the use of NIST filters,
or by comparison to a reference instrument that is known to be calibrated to NIST filters.
Calibration may be confirmed using a commercially available calibration check set which can be
obtained from your distributor. A periodic verification of instrument accuracy and linearity is
advised. Since most lab test results are based upon standards rather than upon absolute
absorbances, the linearity of the instrument, rather than the accuracy, is the more critical
indicator of instrument performance.
Chemistry error
Operator error
Lamp failure
Degraded filters
Standards and blanks should be closely checked for accuracy and results should be compared
with a reference instrument. Use of a commercially available photometer check set to verify
photometric accuracy and linearity is strongly recommended. If field recalibration is needed, it
should only be performed with the proper reference materials and instruments.
23
5.1 Software
Calibration data is used by the instrument to compensate for variations in electronic
components. During factory calibration, any necessary hardware adjustments are made
using precision reference standards and specialized electronic equipment. The calibration
data are stored in the instrument memory and also recorded on a label located on the bottom
of the instrument. In the unlikely event the instrument must be field recalibrated to meet
specifications, new calibration data are derived using reference standards and these values
are entered via the keypad. This requires no hardware adjustments.
24
0.0
2. If the data printed in step 1 did not match the example, stop. Complete the procedure
"Reset Calibration Data", then repeat this procedure. Otherwise, locate the
calibration data label on the underside of the instrument. Copy the data from the
label for reference in the following steps.
3. If the date and time are incorrect, set the date and time as described in the Owner's
Manual.
4. Select test #210. The display shows:
BLOCK=
Enter the value from the Block Temp line on the calibration label.
5. The display shows:
CELL=
Enter the data from the Cell Temp line of the calibration label.
6. Select test #212. The display shows:
ABS FACTOR=
Enter the number from the Abs Factor line of the calibration label.
7. Select test #213 to print the calibration data. The printed data should now match the
values shown on the calibration data label. If it does not, complete the procedure
"Reset Calibration Data", then repeat this procedure.
25
0.0
2. If the data printed in step 1 matches the example shown, stop. The calibration data is
already reset. If the data printed in step 1 matches the data shown on the calibration
data label on the bottom of the instrument, and you do not wish to alter the
calibration, stop. Otherwise, make a note of the values printed in step 1.
3. Select test #210. The display shows:
BLOCK=
Enter the negative of the value printed for BLOCK TEMP. This will reset the block
temperature offset to 0.0.
4. The display shows:
CELL=
Enter the negative of the value that was printed for CELL TEMP. This will reset the
cell temperature offset to 0.0.
5. Select test #212. The display shows:
ABS FACTOR=
Enter the inverse of the value printed for ABS FACTOR in step 1. That is, calculate
1
ABS FACTOR
and enter the value.
6. Select test #213 to print the calibration data. The printed data should now match the
example in step 1. If not, repeat this procedure.
26
27
7.
8. Repeat steps 3 through 6 until the resistance is within the specified range and the
displayed temperature is 37C 0.3. Select test #213 each time to print the current
net adjustments. The net offset after the temperature is set should be 1.5C or
smaller.
9. When you are done, select test #213 to print the new calibration data. Enter these
values on the calibration data label on the bottom of the instrument.
28
8. Select the absorbance mode again. Blank as before on the diluent. The 1:2 dilution
absorbance should now agree with the reference to within .005 absorbance.
9. Read the 1:1 concentration. It should agree within 2 % of the reference. Read the
section Hardware Calibration before making any adjustments. Re-blank and
reconfirm the values.
10. When you are done, select test #213 to print the new calibration data. Enter these
values on the calibration data label on the bottom of the instrument.
29
6. REPLACEMENT PARTS
6.1 Consumable Items and Accessories
Lamp..........................................................112004
Fuses (set of 2)..........................................102006
Printer paper (5 rolls) .................................150005
Carrying case.............................................000390
Borosilicate tubes 12x75mm 250/box ........157300
Borosilicate tubes 12x75mm1000/case .....157302
Dust Cover.................................................001990
30
31