CK-4 Portable Corrosion Monitor Quickstart Guide
CK-4 Portable Corrosion Monitor Quickstart Guide
CK-4 Portable Corrosion Monitor Quickstart Guide
AND CORRDATA CK
SOFTWARE
QUICKSTART GUIDE
CORRDATA CK Software
for CK4 CORROSOMETER instrument,
CORROSOMETER, and CORROTEMP Probes
Introduction
The CK4 CORROSOMETER and CORROTEMP probe measuring instrument is designed to
simplify probe reading and data presentation. CORROTEMP probes incorporate temperature
measurement into the standard CORROSOMETER probes without any additional connections.
This additional independent data can be very helpful in correlation of corrosion rated with
operating temperature changes.
Setup
1.
2.
3.
Data Collection
1.
Plug the CK4 into each CORROSOMETER or CORROTEMP probe, and take the
probe reading.
2.
3.
At the P.C., start the CORRDATA CK software, read back the stored probe
readings from the CK4 lettered registers, and enter them into the program.
4.
The above steps are explained in more detail on the following pages.
Check that your P.C. is an IBM or compatible, with EGA or VGA graphics, a hard and
floppy disk drive, and running MS DOS 3.3 or higher (MS DOS 4.01 or 5.0 is preferred
for greater graphics print screen flexibility).
2.
Choose the 3" diskette or the first 5" diskette to insert into your floppy drive. At the
DOS prompt change to the floppy drive containing the diskette, and type install.
3.
Follow the on-screen instructions. The default settings install the files in the CORRDATA
directory under the C: drive root directory.
Quickstart Guide 3
NOTE: The install program also requests selection of COM1 or
COM 2 serial port. Although the serial port is not used with manual
input readings from the CK range of portable instruments, this
program is also capable of operating with the CORRDATA Remote
Data Collectors, and Mate I and II probe reading and data transfer
units. These systems allow fully automated data collection and direct
transfer to the P.C. without the need for any manual data transfer. For
further details contact Rohrback Cosasco Systems Inc.
4.
Install the batteries in the CK4 instrument by first removing the battery
compartment cover on the back of the instrument. Ensure that the correct polarity
is observed as indicated in the battery compartment.
2.
3.
Press TIME (F4). This will display the current date and time, if already set.
4.
To make changes press the MONTH (F1), DAY (F2), and YEAR (F3) keys as
required. Press once for a single change; hold down the key for scrolling; hold
down the ALPHA/NUMERIC key in addition to reverse the direction of scrolling.
5.
Use the same method to set time. When complete press Exit (F4).
2.
Press Enter to clear the logo display screen and show the main menu.
3.
From the main menu and sub-menu select SetUp, and Units. Choose the
engineering units desired from mils (0.001"), millimetres (mm), or micrometres
(m), and press Enter.
4.
From the main menu and sub-menu select Configure and New respectively. Select
Manual Co (Manual CORROSOMETER) for CORROSOMETER probe
configuration or Manual CoT (Manual CORROTEMP) for a CORROTEMP probe
configuration. In either case the entry screen is similar.
NOTE: The other selections in this sub-menu are not applicable to
use of this program with the CK4 instrument. They are part of the
CORRDATA software for the fully automated data collection that may
be achieved with Remote Data Collectors (RDC's) and either a Mate
I or Mate II data transfer unit. To see a demonstration of the capability
of this type of system, retrieve the demonstration graph DEMO.SAV
as described later in the section "Archiving and Retrieving Old Data
Files".
Quickstart Guide 5
5.
2.
3.
Use the TENS (F1) key and the ONES (F2) key to change the span. Hold down
the key to scroll up through the values; hold down the ALPHA/NUMERIC key to
scroll down the values. Enter the probe span in mils as marked on the probe.
NOTE:
A corrosion rate will only be computed on the CK4
instrument display if the probe readings are more than fourteen days
apart.
2.
Press READ (F1), and select the probe type, WIRE (F1), T/S (F2), or SPEC (F3).
Probe types are labelled on the probes as type A for WIRE , type B or C for T/S,
and type D for SPEC respectively. Alternatively the lettered keys A, B, C, or D
may also be used directly to enter the probe type.
3.
Select and press the label (A-Z) corresponding to the configuration label used on
the P.C. against which to store the data.
NOTE: Mark the probe with label A-Z to ensure correct probe
identification for data collection and transfer to the P.C.
4.
The probe reading will take approximately 2 minutes and 45 seconds. At the end
of this time the probe data will be displayed.
NOTE: If a probe is near the end of its life a warning message will
be displayed. A warning to "Replace Soon" is displayed above 80% of
probe life, and "Replace Now" above 95% of probe life.
Quickstart Guide 7
5.
If a CORROTEMP probe is being read, press TEMP (F4) to read the probe
temperature. This reading will take approximately 45 seconds to complete, and then
be added to the data display.
6.
Again select the probe label (A-Z) against which to store the reading.
7.
To exit the data display screen either turn OFF the CK4, or press any other key
except F1 to F4.
8.
The last probe readings from any probe may be reviewed on the CK4 from the main
menu by pressing DISP (F2), and then the label (A-Z) of the required probe. To
read any other probe, press its label (A-Z)
From the CORRDATA main menu select File, and from the sub-menu Man Input.
This will display the listing of configured probes.
For guidance on the data entry keys, see the menu bar at the base of the screen.
3.
Use the 8 and 9 keys to select the probe for which data is to be input.
4.
Switch ON the CK4 and press DISP (F2), and select the probe label corresponding
to probe selected on the P.C.
5.
6.
For a CORROSOMETER probe, the input fields are Date, Time, Check reading,
and Measure reading. Use the 8 and 9 keys to select the required field, and the
Enter key to enter and leave each data entry box. At initial entry the cursor is set
to the overwrite mode as indicated by the thick cursor. On a CORROTEMP probe
an additional data entry box for temperature will also appear.
Quickstart Guide 9
7.
In the Date field type the month, day, year from the CK4 screen in the form
indicated on the P.C. screen. When complete press Enter again to leave that data
entry box.
8.
The Time entry box is set to a default of 12:00 mid-day. This may be left
unchanged if the time of day that readings were taken are not significant, such as
for low corrosion rates. Otherwise type the time from the CK4 screen in the format
indicated.
9.
Type the probe Check reading from the CK4 screen. The very first probe entry is
the most critical as it is the one against which all the others will be compared, to
determine if the check reading has gone bad. This is based on 20 divisions from
the initial reading, and is indicated by a thinner graph line than the normal.
WARNING! Once the first reading has been saved with the F2
key the initial check reading cannot be reset without first deleting
the probe in the Configure menu. This applies even if the Edit
command is used to change the check reading as displayed.
10.
11.
For a CORROTEMP probe type the Temperature reading from the CK4 screen.
When satisfied with all the entries for that probe reading pres F2 to save the data.
NOTE: At least two probe readings must be entered to be able to
display a graph.
12.
From the manual probe list summary, or the individual probe data list, the graphical
display may be viewed by the shortcut key F3. Esc will then return back to the
same place. The full display features are available under Display on the main
menu.
13.
When one probe entry has been completed, press Esc return to the manual probe
summary list, and select the next probe for data entry. When all the probes have
been completed press Esc as required to return to the main menu.
14.
To edit any probe data from the manual probe summary list, use the Insert key
instead of the Enter key. To remove a line of probe data use the Delete key.
NOTE: The date of a probe reading may only be edited to within the
range of dates of the readings on either side of it, otherwise the graph
will "turn back" on itself when viewed. It is not possible to delete the
From the CORRDATA Main Menu select Display and then List to show a
complete list of probes on the system.
2.
Use the 8 and 9 arrows to move the highlight bar to the selected entry.
3.
4.
To select any other probe press Esc and List. Repeat steps 2 and 3.
Graphs initially show all the probe data up to 1131 points for CORROSOMETER
probes. The range of interest may be zoomed in by using X-Select and Y-Select
from the Ranges sub-menu.
2.
3.
Move the selection cursors with the 8 and 9 arrows. Use the Space bar to toggle
between the cursors (Note the cursors initially start on the top and bottom edges of
the graph, and the initial selection is the bottom cursor).
4.
Quickstart Guide 11
5.
To zoom in on the X-axis select X-Select. The graph will reappear with vertical
cursor lines. Move these cursors using the 6 or 7 arrow keys, and use the Space
bar to toggle between the two cursors (Note the cursors start initially at the sides
of the graph and the initial selection is the left cursor).
6.
When satisfied with the selected range press Esc twice and then View to display the
zoomed in graph.
If the cursors are positioned between data points, the CORRDATA program
calculates the point at the cursor based on the straight line connecting the data
points on either side. This point is then used in the linear regression formula.
3.
The MS DOS graphics file must be loaded BEFORE graphs can be printed from
the CORRDATA program. If the program has already been loaded without the
DOS graphics file and a graph print-out is required, quit the program, load the
graphics file as described below and restart the CORRDATA program.
2.
With MS DOS 4.01 or 5.0, loading the appropriate graphics file from the DOS
prompt before running CORRDATA software, enables Shift + Print Screen for
printing.
2.
MS DOS 5.0 provides the widest range of printer compatibility, including laserjet,
and inkjet. The common commands are as follows:
graphics/r
graphics laserjet/r
graphics laserjetii/r
graphics paintjet/r
graphics deskjet/r
NOTE: The "/r" prints the screen as it appears on the P.C. monitor
screen with a white background. If "/r" is omitted the background will
be black, thereby increasing ribbon, toner, or ink consumption.
For other options or more detail see the MS-DOS Manual.
Quickstart Guide 13
Archiving and Retrieving Old Data Files
The basic CORRDATA software provides for saving of files one at a time for archiving
purposes, and for retrieving files for viewing one at a time. Any files saved are
automatically given the extension .SAV to avoid any overwriting of data files even if they
are saved to the CORRDATA directory. For archiving purposes, set up sub-directories for
the required time periods (such as CD_MAR93) and save the archive data to this subdirectory. Each file will store up to 9,000 data points.
To save the current data to a sub-directory, first create the sub-directory in DOS at the
required location before entering the CORRDATA program. Start the CORRDATA
program and select Save, and press Enter to display the following screen.
Quickstart Guide 15