MicrocorTools 4 4-5-100156 Rev J
MicrocorTools 4 4-5-100156 Rev J
MicrocorTools 4 4-5-100156 Rev J
Reference Manual
Table of Contents
Chapter 1 Introduction and Installation ...............................................1
Software Packages....................................................................................1
Software Installation and Activation........................................................1
Modifying the Database Location.............................................................2
Optional Data Remoting Software Add-On for Remote Data Collection(WAN/LAN)7
ii
TABLE OF CONTENTS
CHAPTER 1
Software Packages
The Microcor Tools Software offers a cost effective and simple method to collect,
save, transfer and display corrosion information from online-operated Microcor
and E9020 Transmitters, and Microcor Data Loggers. The software is offered in
two versions to accommodate different corrosion monitoring needs.
1. Microcor Tools Standard
2. Microcor Tools - Output Interfaces
The Microcor Tools Standard version is a stand-alone software package for
installation on the user computer to provide data retrieval, storage, and graphical
presentation of results from Microcor and E9020 Corrosion Transmitters and Data
Loggers.
Microcor Tools - Output Interfaces is an enhanced version of the standard
software which provides Modbus and OPC output interfaces for data transfer from
Microcor and E9020 online transmitters to Distributed Control Systems (DCS),
Process Control Systems (PCS), or Supervisory Control and Data Acquisition
(SCADA) systems. This software package is included on RCS Supplied hardware
such as the Intelligent Interface Unit (IIU).
The Microcor Tools - Output Interfaces version requires an activation key from
the manufacturer. Upon installation, run the software by navigating to Start All
Programs RCS Microcor Tools. A dialog box with the message License
error: No valid license will be displayed followed by the Software Registration
dialogue box with the Installation Code (see image below).
Choose Open With and the following dialog box will open:
Choose Select the program from a list and click OK. Choose Notepad or
WordPad from the next screen to proceed to edit the data.cfg file.
4. When the Notepad editor opens, change the path C:\aa with the
following path Z:\Userdata\Microcor\MicrocorToolsDatabase
(shown below) and Save and Exit.
Note: If the Microcor Tools Software program has already been used
and data has been collected, then this collected data can be
transferred to the new shared database location. Copy the
data file directory (not the newly renamed data.cfg file) from C:\
Program Files\MicrocorTools to the new directory.
It is also possible to share the database when a single computer is shared amongst
multiple users. Simply choose a shared location such as Shared Documents of
the All Users profile for the database and follow the above procedure.
CHAPTER 2
To launch the Microcor Tools program, click Start All Programs RCS
Microcor Tools.
Note: This reference manual will use Microcor Tools Output
Interfaces screen captures unless otherwise stated, some
of the command buttons and features may not be available
on the Microcor Tools Standard version of the software.
Onscreen help is accessible by clicking on the Help button and the Save button
will store and retain all configuration changes.
Online data collection via the RS485 network from Microcor Transmitters
Online data collection via RS485 network from Corrater E9020 Transmitters
Offline data input from Microcor Data Loggers or data-transfer units
Offline data imported from archives or files, or the previous versions of Microcor software.
Create a Site
Select the New Site option button and type a name in the text box (Existing Sites
option button is selected by default). Click the Create Site command button.
3. Locate the Address area. Use the up/down control buttons to select the unit
address or type the number in the text box (0 99).
If the data source for this location is an online transmitter, the Address must be
equal to the transmitter address. Internally, each transmitter has a DIP switch
with a selectable address (0 to 99 for Model MT 9485A and E9020 and 0 to 31 for
Model MT 9485. The preset address is marked on the outside of the transmitter as
the IN# or Instrument Number. See the transmitter manual for more details. If the
data source is a data logger or data-transfer unit, RCS recommends that the IN #
of the transmitter connecting to the data logger is used.
4. Type in the Com Port # being used on the computer to communicate with the
instrument. This may be an actual hardware port on the computer or a virtual
Com Port for operation with a TCP/IP converter via the network. This setting
is only for Online connections and is disabled for Data Logger connections.
5. Complete the Description field with the provided free-format text description
area to meaningfully identify the location (optional). The first 50 characters of
the Description field will be displayed on the History Viewer graph display
along with the Site, Address, and Tag information for easy identification.
Click the checkbox to select either Online or Data Logger data transfer method.
Data Logger checkbox is enabled only when the Start button on the Data Logger
Communication section of the Site/Communications tab is pressed and when
the instrument is set as Microcor. Data Logger Communication must be started
to be able to check or uncheck the above Data Logger box. Supported transmitters
are Microcor (MT 9485A and Microcor MT 9485) and Corrater E9020. E9020
transmitters are used with Online connections only.
Note: It is important to ensure that the Hibernate, System
Standby and Turn off hard disks on the Windows Power
Options is turned to Never so data collection will not be
interrupted on Online systems.
10
Users can choose to enable the alarm by selecting the CR/ER Alarm Enabled
button. However, it is important to note that this feature is only relevant when used
with Online systems. Alarm Level can be changed by typing in the desired value.
The measurement units are specified to the right of the alarm level value and this
can be changed in the Graphical Display (History Viewer) screen (click on the
Show Data command button; see the Retrieving and Plotting Data (History
Viewer) section for more details). Once the alarm is activated, the Parameter
3 column on the Data Display screen will turn red if the corrosion rate exceeds
the set alarm rate. A blinking red/gray indicator with text ALARM will replace
the green ALARM OK indicator on the status bar located on the bottom of the
Microcor Tools application screen (see the section on the Online Data Collection
below for more details).
The standard method for corrosion rate calculations is Linear Regression, which
can be changed by checking the 24h Filter. Checking the 24h Filter adds a
24 hour filter specifically to reduce the effects of any diurnal (24) thermal noise.
Linear Regression values are adjustable and can be set on the Graphical Display
(History Viewer) screen.
The Negative Rates Suppression enables the user to visually suppress the
negative corrosion/erosion values on the graph that are due to the noise in the
raw signal. When this option is enabled, negative values will show as zero (0) thus
depicting the minimal corrosion/erosion as zero.
7. Proceed to and complete the Probe section once the above selections are
made.
Type is an optional informational field which will be useful later when reviewing the
corrosion/erosion data. Element passes information when configuring E9020 so
it is highly recommended that this field not be empty. These pull-down menus will
provide most of the typical RCS equipment configurations. It is possible however,
Span [mils] is the useable life of the probe element when working with Microcor
Instruments. This is always defined in units of mils and is one half of the probe
element thickness (i.e. F10 or T10 probes have a 5 mil span). This required field is
used to define the physical scale of the metal loss graph.
For E9020 Instruments, Alloy Multiplier is used instead of the Span.
The appropriate Span or Multiplier is automatically set when the Probe Element
is chosen from the drop down menu; however, users may change the Span or
Multiplier by typing in a different value if necessary. If the required Probe Element
is not in the list, choose BLIND as the Element, which will set the Span or Multiplier
to 1; this can be changed as needed (set the Alloy Multiplier to the appropriate
multiplier value for the metal or alloy of the electrodes as shown in the chart of
Appendix 1).
Note: Make sure to choose the proper Element and Multiplier
values as these will be passed automatically to the E9020
Setup screen.
Note: The BLIND setting may also be used with the Microcor
Tools Output Interfaces version when used on blind
Intelligent Interface Units. This allows scaling changes for
different probe spans to be accommodated internally on
the IIU such as in DCS or SCADA systems.
8. Proceed and complete the Transmitter section (this information is necessary
for both Online and Data Logger connection types).
11
12
The two sections represent the settings stored in the computer (Database) and
in the instrument (E9020). The arrow command buttons provide settings transfer
capability from one to the other:
- from the Database to E9020
- from E9020 to the Database
To store the new configuration from the database to the E9020, choose the bottom
arrow pointing to the right. To read the current configuration of the E9020 and store
to the Database, choose the top button with the arrow pointing left.
Cycle Time is used to set the desired measurement cycle time. For water with
conductivities greater than 100 S, any cycle time of 5 minutes or more should be
satisfactory. For lower conductivity water, or when filming type inhibitors are in use,
it is recommended that the 20 minute cycle time be selected first. After consistent
readings are obtained, a shorter time cycle can be selected as long as the readings
do not increase substantially (>5%) over the readings taken with a 20 minute cycle.
The Alloy Multiplier and the Element values will be automatically updated from
the Config/DL Tranfer screen values.
The default selection on the E9020 setup for the Electrode type is Standard.
Keep the selection Standard if using a probe with protruding 5.0 square centimeter
electrodes or select Flush if using a probe with flush 0.5 square centimeter
electrodes.
The Transmitter Test command button is enabled for Online connections. It sends
requests to the transmitter and displays returned data in the Answer text field if the
communication is successful.
9. If using data loggers, first make a complete list of the IDs (Addresses), Tags
and descriptions to be used. This will help organize and manage sites.
Note: The ID (address) of the data logger is independent of the
address of the transmitter to which it is connected. The data
logger communicates with a transmitter irrespective of the
transmitter address. The ID (address) of the data logger is
allocated by this software or from the Mate or Mate II during
configuration of the data logger in order to segregate the
different sets of data.
10. The Microcor Tools software allows the direct connection from a PC to a data
logger, (this may be used ONLY in an electrically safe area). Alternatively,
the data loggers can be programmed and data retrieved with the intrinsically
safe Mate or Mate II instrument programmed in Micromate mode or with a
Checkmate DL programmed for Microcor Systems, then the data can be
transferred to the PC in the safe area.
11. When using a direct PC-to-data logger connection, the data logger screen below
is used for configuration and data retrieval after the Com Port # is enabled in
the Data Logger Communication section in the Site/Communications tab.
13
14
Read Setup command button reads and displays the data logger setup and
status information. Use this command to check the settings of the data logger
before the data transfer.
Store Setup command button reads the Tag and Sampling Interval settings
from the Transmitter section and stores this information to the Data Logger. It
will also set the data logger ID equal to the selected Address, and the data logger
real-time clock to match the computer date and time.
Note:
Read Data command button reads the collected corrosion information from the
data logger and stores it in the computer.
Clear Data command button use this command to delete the data logger memory.
When using the Micromate to configure the Microcor Data Logger, see the section
on Configuration of Data Loggers with Micromate. Refer to the Checkmate DL
User Manual for instructions on configuring Microcor Data Loggers with Checkmate
DL.
Commands
This section is used to handle history information; display, delete, move, and load
data from archives.
Show Data command button opens the Microcor Tools History Viewer (see
the Retrieving and Plotting Data (History Viewer) section for more details).
Delete Data command button deletes the corrosion data stored for this location.
Move to Archives command button moves the data file to an archives folder.
Archive files have automatically assigned names. For example, the archive file
of location Address 15, created on 13 May, 2003 is: ch15_13May03.dat. See
separate section on Archiving.
Load from Archives command button use this command to load data from
archives or import data from previous versions of Microcor. Then use the Show
Data command to display the information graphically. This command can also be
used to import data from other Microcor files (on the hard drive or floppy disk).
Apply command button only applies the changes to the current session. When
the program is restarted, the applied settings will be lost.
Note: Data files (*.DAT) that are imported do not include probe
reading interval or probe span details. This must be separately
added on the Config/DL Transfer screen. This information is
stored in the file (*.SET) of the original data folder.
12. At this point the settings for the current instrument are complete. For the next
instrument select new Address and repeat steps 1 to 11.
Note: The entries may be made for all of the instruments before
saving the settings. However, saving the settings after the
entries for each location avoids the chance of losing the data
accidentally. To save the configuration on the hard disk, select
the Site/Communications tab and click the Save command.
15
16
17
CHAPTER 3
18
3. Click the Start command button in the Data Collection section to start
collecting data from the transmitters to their respective data files. One solid
green indicator with text ALARM OK and one flashing green/yellow indicator
with text DATA will appear on the status bar to indicate that the program is
collecting data (see images below).
If the solid green indicator starts to blink red with text ALARM and gray, then it
is indicating that the corrosion rate level of at least one unit has surpassed the
set alarm level on the Rate/Alarm Settings area (only when alarm conditions are
enabled on the Rate/Alarms Settings screen).
The Parameter 2 (for E9020) and Parameter 3 (for Microcor) of the unit(s) in
alarm status (at the Data Display screen) will turn red and the Event Log will
display detailed information of the unit(s). Click on the blinking ALARM indicator
on the status bar to acknowledge the alarm. This will stop the blinking and reset
the status bar indicator to solid red and the Parameter of the unit(s) in alarm to
yellow. This means there is an ongoing alarm, and it is acknowledged. When the
alarm is turned off or when the corrosion rate resets to below the alarm rate, the
indicator light will turn to solid green with the text ALARM OK. If a unit enters an
alarm status and resets prior to user acknowledgement, red indicator will change
color to blinking green and gray and will reflect this on the Event Log; hence it is
highly recommended that the Event Log is inspected regularly.
Normal Operation
Status Bar:
First Icon Group Alarm - Solid green, indicates active data collection and
corrosion rate level below alarm value.
Second Icon - Transmitter Status - Flashing green/yellow, indicates reading of
probe transmitters is active (normal).
Active Alarm
Status Bar:
First Icon Group Alarm - Flashing red/gray, indicates corrosion rate level of at
least one unit has surpassed the set alarm value.
Second Icon - Transmitter Status - Flashing green/yellow, indicates reading of
probe transmitters is active (normal).
Event Log:
Alarm status is recorded on the Description area.
Alarm Acknowledged
Status Bar:
First Icon Group Alarm - Solid red, indicates user acknowledged the alarm
but the alarm condition still persists. This will turn to solid green when the alarm is
turned off or when the conditions return back to normalcy.
19
20
Event Log:
Description changed to show Alarm Recognized status.
Event Log:
Description will record the date, time and ID of the unit(s) which was in alarm
status.
4. When using Microcor Tools for online monitoring, it should be setup in the
Microsoft Windows Startup folder and the Autostart checkbox should
be checked. This ensures that the program will restart automatically and
commence the online communication and data collection when the computer
is restarted after maintenance or power failure.
Note: Users may require Power User or Administrative privileges
to perform the below operations and should always
consult the organizations IT Department for assistance
and permissions.
To add Microcor Tools to the Windows Startup folder, copy the Microcor Tools
icon form the desktop and paste it in the C:\Documents and Settings\All Users\
Start Menu\Programs\Startup directory.
5. The Auto Archiving feature is designed for computers or embedded
computers as in the RCS IIU with limited storage resources. If the Auto
Archiving is selected, when a site collects 2MB of data (approximately
48,620 readings or 67 days of continuous data collection (temperature
data collection enabled) at a sampling interval of 2 minutes), the data file is
automatically transferred into the archives folders replacing the previously
stored data file, and the current data file will be cleared for collection of new
data.
Note: When Auto Archiving is enabled, Microcor Tools initiates
automatic archiving when the file reaches its limit, previous
automatically stored archives will be deleted. This process
will not delete any manually archived files.
6. Microcor Tools - Output Interface software (installed on all RCS Intelligent
Interface Units) allows users to setup data communication with other
computers. Click the Interfaces Configuration button to enter the setup
screen shown below. Select the desired method of transfer and configure the
fields if necessary. Click the Enabled radio button for the selected data transfer
method and click the Save Configuration button prior to pressing Return.
Typically only one interface, either Modbus or OPC, is used at a time. However,
it is possibly to use both output interfaces at the same time if required. Microcor
Tools acts as a Modbus slave and its address is set in the Device window. The
Com Port should be set to correspond to the hardwired connection to the Modbus
Master. If Microcor Tools is supplied on an RCS computer or embedded PC, the
Modbus communications may be via RS 232, or two or four wire RS 485 according
to the unit ordered. Most commonly, Remote Terminal Unit (RTU) binary protocol
is used for economy of transmission data length. This and the baud rate, parity,
data bits, and stop bits should be set to match the Modbus master with which it
is communicating. The Modbus addresses and data format is shown in Appendix
2. Generally, floating point is not part of the Modbus Standard but there exist two
different versions for data transmission. The Swapped floating point is provided
to adapt the Microcor Tools data output to the customer format.
21
22
7. To view the collected data, switch to Config/DL Transfer screen and change
the Address to the desired ID. Click on the Show Data command button to
open the Microcor Tools History Viewer (see the Retrieving and Plotting
Data (History Viewer) section for more details). An alternate method is to
switch to Data Display screen and double click on the desired Tag to open the
Microcor Tools History Viewer.
Testing Channels
Individual transmitter channels may be checked with the Transmitter section on
the Config/DL Transfer screen.
Microcor Troubleshooting
If abnormal readings or excursions occur that cannot be explained, the system
electronics should be verified by connecting the meter prover RCS PN 20407-001
(included in Microcor Installtion Kit RCS PN 702109) in place of the probe and/or
probe adapter and checking that the reading is approximately 50% of the probe
span.
If the meter prover shows the electronics are operating correctly, then inspect the
probe-to-transmitter connections, the probe adapter and/or the probe for damage,
high resistance, corrosion or poor connections.
After initial application of power, the transmitter takes approximately one minute
to start up. During that time, it will send the error code -999,999 or -999,996 when
the data is requested.
On ICMS3 systems, these codes will show up as error messages:
-999,999 Startup Mode
-999,996 Auto Ranging
If there is a probe or probe connection failure, error code -999,998 or -999,997
will be generated.
On ICMS3 systems, this will display as follows:
Code: -999,998 Display: Probe Fault (4)
Code: -999,997 Display: Probe Fault (1)
If communication with the selected transmitter fails, a No Answer error message
will be generated in the status bar located at the bottom of the screen.
23
24
CHAPTER 4
1. For data loggers installed in non-hazardous areas, the configuration and data
retrieval may be done directly with a portable PC.
2. For data loggers installed in hazardous areas, the configuration and data retrieval
is made with the intrinsically safe Corrdata Mate or Mate II Data Transfer units
running Micromate software or with the Checkmate DL handheld instrument
programmed for Microcor Systems. Collected data is then transferred to the
PC in the safe area. In this case, the Micromate and Checkmate DL emulate
each data logger so that the software program retrieves collected data from
them as it would from a data logger.
Note:
25
26
3. Switch to the Config/DL Transfer screen and select the Address and click
the Data Logger checkbox in the Connection section and configure the
transmitter as described previously.
4. Click the Store Setup command button in the Data Logger section. This will
populate the text fields in the Data Logger area accordingly. Verify all the text
fields including Time and Date, and Battery for accuracy. Read Setup will
interrogate the data logger and display stored values and will enable verification
of these stored values.
If the data loggers are in a hazardous area, they will need to be configured with the
Corrdata Mate or Mate II programmed with the Micromate software or Checkmate
DL programmed for Microcor Systems. Make a note of the configuration information
as this will have to be entered again into the Micromate or the Checkmate DL.
The time and date shown in Time and Date box should correspond to the current
time. If there is an error or drift in the time, this can be corrected with the Store
Setup command.
The Battery text field should read OK. If it reads Replace, then the battery needs
to be replaced on the data logger and the data MUST be cleared after collection
for data collection to recommence.
Important Note:
The Sampling Interval text field should correspond to Sampling Interval in the
Transmitter section.
6. Click the Read Data button. If data from this location has already been
collected, the following message will appear:
Press Yes to add the new data onto an existing file. This is the normal
method if the data in the data logger is deleted after it is collected each
time in order to speed data transfer. It also prevents reaching the maximum
storage capacity in the data logger at which time the data collection stops.
Press No if all of the data in the Micromate/Checkmate DL is to replace
an existing file. This should be used if the data in the data logger is NOT
cleared each time it is read. In this case the existing file is automatically
archived.
Press Cancel to exit the Read Data function.
Note:
27
28
After a selection has been made, a counter in the status bar indicates the number
of samples to be transferred. Status bar will read End of Transfer when the data
is successfully transferred from the data logger to the Microcor Tools Software
program.
7. Verify that the data is transferred correctly - click the Show Data command
button to display the data with the Microcor Tools History Viewer.
8. It is recommended to clear the data in the data logger by clicking the Clear
Data command button at this point to reduce the data transfer times on later
downloads.
9. Repeat the steps 1 to 8 for each of the data loggers from which data is collected.
Make sure that the correct Address is selected for each data logger before
loading the data.
Note:
1. Verify that the Corrdata Mate or Mate II is programmed with the Micromate
software for use with Microcor Data Loggers. The Corrdata Mate programmed
as a Micromate will appear as follows when switched on:
Micromate ver x.xx
Free Readings 24500
Mate Battery: OK
More
If the Corrdata Mate is programmed for operation with Corrosometer and Corrater
probes and data loggers, the screen will appear as follows when switched on:
5. From the main menu select CommPort, and click on the Com port that will be
used for communication with the Mate. This will usually be Com 1 or Com 2.
29
30
7. Select the Quick Mode unless any of the problems described on the image
above have occurred. A pop-up screen will then ask if the Mate is already
programmed as a Mate or a Micromate. This is because the key selections
required to set up a Mate are different from setting up a Micromate.
9. Follow the instructions above and then select the button to program the Mate
as Mate or Micromate. In this case the selection would be Make Micromate.
A Scroll bar on the PC will indicate the progress of the programming and a scrolling
of addressing being programmed on the Mate will show on the Mate screen.
Note:
The Mate will beep and indicate when the programming is complete.
10. The programming of the Mate with the Micromate software also sets the Time
Zone information into the Micromate automatically from the computer. The
correct time zone is required as the underlying data format uses a GMT time
reference.
Note 1:
Note 2:
31
32
16. Enter the Tag number for this location, up to 12 characters, and press Enter
(F1) to complete the entry and move to the next screen:
Note:
Exit
19. Leave the data logger to collect data. It will take its first reading at the end of
the sample period.
20. Repeat steps 15 to 19 on the other data loggers that are to be configured.
33
34
2. Switch on the Micromate, and press F4 twice to clear the initial screens.
3. Press Read (F1) to display the following screen:
ID: 25 CP 1002
Jul 7, 2006 13:30
Read Read Set
Stat Data Clock Exit
4. If the battery on the data logger has expired, the Micromate will indicate that
there is no communication. In this case, change the data logger battery.
When the battery runs low, the data logger will first switch off power to
the transmitter, set the Battery Status to LOW, and stop taking readings.
However, it will have enough power left for some time to allow communication
with the Micromate.
WARNING:
Note:
After replacing the battery, re-read the data logger as described in steps 2 and 3.
Replacing the battery will require the data logger clock to be reset. This will be
indicated on the screen:
ID: 25 CP 1002
Clock Not Set!
Read Read Set
Stat Data Clock Exit
Press Set Time (F3) and enter the current date and time. Otherwise the default
date is Jan 1, 2000. This function is also used to correct the time in the data logger
when required.
Set D/L Clock
<
>
YYMMDDHHMMSS
Enter CLR BkSp More
WARNING:
Memory Bad indicates a problem in the data logger memory requiring repair.
Transmitter Bad indicates that the data logger is not communicating
with the transmitter. It suggests a problem with the interconnecting
cable or transmitter.
Press Exit (F4) to return to the following screen:
ID: 25 CP 1002
Jul 7, 2006 13:35
Read Read Set
Stat Data Clock Exit
6. Press Read Data (F2) to commence collecting data. If the data logger has not
had its Logger # or ID set (for example if it was configured directly from a portable
PC using the previous MS 9000 version 1.8 software), the Micromate will request
setting of an ID (Address) and Tag. If the data logger has already had the Logger
# or ID set, then it will go directly to a screen similar to the following.
Data Logger ID: 01
Tag : CDL 1003
Start
Exit
35
36
Getting Data
Please Wait
Approx 15.7 seconds (or mins)
The count-down screen will indicate how long it will take for the data transfer.
7. When data collection is complete, the screen will display the last reading, the
date and time of the last reading, and the number of readings taken (shown
below):
Last Reading +128586
July 7, 2006 13:30
Number Reads: 245
Clear
Next
The last reading confirms that the instrument is producing good numbers. If the
reading is near +262144, the probe has almost expired and needs changing. The
data logger will need to be powered down and the probe replaced. Upon power-up,
the data logger will need re-configuring to set the correct time and date.
8. Press Clear (F1) to clear the data memory in the data logger. This MUST be
done if the Battery Status indicated LOW and the battery has been changed.
If the battery has not been changed or do not wish to clear the memory, press
Next (F4).
9. Repeat steps 1 to 8 for each data logger from which data is to be collected.
Note:
5. Switch on the Micromate and press (F4) twice to go to the main screen. Then
press Xfer Data (F3). A screen such as the following will appear:
Data Logger ID: 01
Tag : CDL 1003
Start
Next Prev
Exit
The first Data Logger # or ID will be the lowest number in the list. Check that
the Data Logger # or ID on the Micromate and the Address on the PC match. If
not press Next (F2) or Prev (F3) until they do match. Press Start (F1) and the
following screen will be displayed:
Ready to Transfer
Data using Microcor
Software from RCS
Exit
At this point the Micromate is emulating the selected data logger.
6. Click the Read Setup command button in the Data Logger section on the
Microcor Tools software. This will check if the communication port is valid and
show the setup information.
The time and date shown in the Time and Date text field should correspond to the
time the data was downloaded from the data logger.
The Battery text field should read OK. If it reads Low, then the battery in the data
logger should be changed and memory cleared.
The Sampling Interval text field in the Data Logger section should correspond to
Sampling Interval text field in the Transmitter section.
Note: The Store Setup and Clear Data are not functional when
working with the Micromate even though the status bar at
the bottom of the screen indicates operation. These are
only operational when working directly with the PC and the
data logger. When using the Micromate, the configuration
and clearing of the data logger is done by the Micromate.
7. Click the Read Data command button. If data from this location has already
been collected, the following message appears:
37
38
Press Yes to add the new data onto an existing file. This is the normal
method if the data in the data logger is deleted after it is collected each time
in order to speed data transfer. It also prevents reaching the maximum
storage capacity in the data logger at which time the data collection stops.
Press No if all of the data in the Micromate is to replace an existing file. This
should be used if the data in the data logger is NOT cleared each time it is
read. In this case the existing file is automatically archived.
Press Cancel to exit the Read Data function.
Note:
After a selection has been made, a counter in the status bar indicates the number
of samples to be transferred. Status bar will read End of Transfer when the data
is successfully transferred from the data logger to the Microcor Tools program.
8. Verify that the data is transferred correctly - click the Show Data command
button to display the data with the Microcor Tools History Viewer.
9. At the end of the transfer, press Exit on the Micromate screen shown below:
Ready To Transfer
Data using Microcor
Software from RCS
Exit
This will return to the previous screen:
Data Logger ID: 01
Tag : CDL 1003
Start
Next Prev
Exit
10. Press Next (F2) to select the next Data Logger # or ID for data transfer from
the Micromate to the PC. Select the corresponding Address on the PC and
repeat steps 6 to 9 above. Repeat these steps to transfer data from other data
loggers.
Note:
11. Clear the memory in the Micromate once all of the data is properly transferred
and verified. This is done from the main screen shown below, by pressing Mate
(F4) and Clear Mate (F1):
ROHRBACK COSASCO
SYSTEMS
Xfer
Read Conf Data Mate
The Micromate is once again ready for data collection from data loggers.
Note:
The data logger checks the battery status during the maximumload time intervals. If the battery voltage is low, a battery Replace
status is set in non-volatile memory. In this state the data logger
does not power the transmitter and does not collect data. Other
functions and commands such as Read Setup, Store Setup,
Load Data, and Clear Data are not affected. The data logger
will remain in this state until Clear Data command is executed
after replacing the battery.
Clr BkSp
Exit
Enter a the desired ID number from 0 to 99 and press Enter (F1) to proceed to
the Set Data Logger Tag display. If the ID selected has already been configured,
the user will be notified; if continued with the configuration, all previously
accumulated data will be erased from the Data Logger.
Enter Data Logger Tag
>
<
Enter Clr BkSp Exit
39
40
In the Enter Data Logger Tag display, enter up to twelve (12) alpha/numeric
characters to uniquely identify the monitoring location. This can be a tag number,
location or process name. Pressing Clr (F2) clears a previously entered ID.
Pressing BkSp (F3) backs up one space for each time it is pressed so that a
change can be made. Pressing Exit (F4) returns to the Standby display.
Press Enter (F1) to proceed to the Set Sample Rate display:
Set Sample Rate
>
< minutes
Enter
Clr
BkSp Exit
Set the desired sample rate and press Enter (F1) to complete the configuration.
Data Logger is successfully configured when the Checkmate DL displays the
screen below. Date and Time settings are automatically programmed into the Data
Logger using the programmed Checkmate DL time and date. Press Exit (F4) to
return to the Standby screen from the following display:
ID: XX XXXXXXXXXXX
MMM DD, YYYY HH:MM
Samp Rate: XXXX
Exit
Note:
If the following error message appears on the Checkmate DL during the Data
Logger configuration, verify the Checkmate DL is properly connected to the Data
Logger.
ERROR
Error Reading Logger
Check Connections
Exit
ID: XX XXXXXXXXXXXX
MMM DD, YYYY HH:MM
Read Read Set
Stat
Data Clock Exit
Press Set Clock (F3) to update the Data Logger with the set time of the
Checkmate DL.
Note:
Exit
41
42
Exit
Press Start (F1) from the above screen to begin the read process. The Checkmate
DL will show the following screen during data transfer.
Getting Data
Please Wait
Approx XX.X minutes
If NO MEMORY SPACE FOR READINGS screen appears (shown below), then
the Checkmate DL memory will need to be cleared before transferring the data. It
is recommended that ALL existing data on the Checkmate DL be transferred to a
PC prior to erasing ANY data. Refer to the section on Clearing the Checkmate
DL Memory on the Checkmate DL reference manual.
NO MEMORY SPACE
FOR READINGS
Exit
Once the data is successfully read and transferred, it is displayed as shown below.
Users can at this time elect to delete the Data Logger content if desired.
Note:
Press Clear (F1) to clear Data Logger readings or Next (F4) to return back to the
Data Logger information screen without clearing the readings.
Last Reading xxxxxxx
MMM DD, YYYY HH:MM
Number Reads XXXXX
Clear
Next
WARNING! It is recommended to clear the data in the Data Logger
after transfer to the Checkmate DL. This reduces the amount
and time of data transfer for future downloads. The data is then
appended to the previous data when transferring the data from the
Checkmate DL to the PC. Refer to the Transferring Collected
Data from Checkmate DL to PC section.
Exit
Press the Start (F1) to proceed to the next screen. Use Next (F2) and Prev (F3)
to navigate the ID list (0 99) and select the desired ID to transfer data (shown
below).
Data Logger ID: XX
Tag: XXXXXXXXXXXX
Start
Next
Prev
Exit
Once the ID is selected, press Start (F1). Following screen will appear:
Ready to Transfer
Data using Microcor
Software from RCS
Exit
43
44
Note:
Make sure that the data (ID of the Data Logger) set in
the Checkmate DL to be transferred corresponds with
the Address selected in the Microcor Tools Software
before transferring the data.
Data is now ready for transfer from the Checkmate DL to the PC. On the Microcor
Tools Software program, select the Site and navigate to the Config/DL Transfer
screen, select the Address and choose Data Logger as the Connection type and
Microcor as the Instrument (see image below). Click the Read Data command
button to begin the transmission. It is also possible to click the Read Setup
command button to transfer the configuration data of the Data Logger to the Data
Logger section on the Config/DL Transfer screen (shown below).
If data already exists for the chosen address, following dialogue box will appear:
Press Yes to add the new data onto an existing file. This is the normal
method if the data in the data logger is deleted after it is collected each time
in order to speed data transfer. It also prevents reaching the maximum
storage capacity in the data logger at which time the data collection stops.
Press No if all of the data in the Checkmate DL is to replace an existing file.
This should be used if the data in the data logger is NOT cleared each time
it is read. In this case the existing file is automatically archived.
After a selection has been made, a counter in the status bar indicates the number
of samples to be transferred. Status bar will read End of Transfer when the data
is successfully transferred from the data logger to the Microcor Tools program.
During the data transfer, the following screen will be displayed on the Checkmate
DL:
Data Transfer in
Progress!
xxxxx Readings Left
Exit
The number of readings (xxxxx) counts down as the data is transferred to Microcor
Tools. Upon successful transmission, Checkmate DL will show the following screen:
Data Transfer Done!
Exit
Press Exit (F4) to return to the Data Transfer screen to choose another ID for data
transfer.
Note:
45
46
47
CHAPTER 5
Microcor Tools allows retrieving and plotting of corrosion data. Microcor Tools
History Viewer can be used to view data collected Online or by Data Logger
connection methods.
To open the Microcor Tools History Viewer (graph feature), select the site from
the Site/Communications tab, select the Address from the Config/DL Transfer
tab, and click the Show Data command button. Alternate method to enter the
History Viewer is to navigate to the Data Display tab and double-clicking the Tag
of the desired unit.
In the above screen, the top graph is the Metal Loss and the bottom graph
presents the probe Temperature (when applicable). The bottom graph can be
switched to show the Corrosion or Erosion Rate for the same set of data when
the Corrosion or Erosion Rate button is selected.
48
To show the corrosion rate graph, click the Corrosion or Erosion Rate command
button (see section below for more details). Click the Temperature command
button to display the probe temperature graph.
Note: The Temperature graph will open as the default bottom graph
if the transmitter supports temperature monitoring capabilities
and if the Probe Temperature check-box is selected; otherwise
the graph area will remain blank (to view the Corrosion or
Erosion Rate graph, Corrosion or Erosion Rate command
button must be pressed. It will not open as a default graph).
Data loggers do not support temperature measurements.
Use the upper Dimension box (top right-hand corner of the screen) to select the
Metal Loss Y-axis units in mils, m, mm or Probe Life Units (PLU). Use the
lower Dimension box (middle right-hand side of the screen) to select units for
the Temperature graph in C or F or mpy, mmpy or mpy units if Corrosion or
Erosion Rate graph is selected.
Corrosion/Erosion Rate is calculated by linear regression. This Regression Interval
may be re-adjusted and the Corrosion/Erosion Rate recalculated by clicking the
Recalculate command button. The default may suit many online systems but can
be changed to optimize noise against sensitivity of the computed corrosion or
erosion rate. The 24h Filter can be enabled at the Rate/Alarms Settings area (in
the Config/DL Transfer screen) to reduce the effects of any diurnal (24) thermal
noise. The Alarm Level units are directly related to the above selected units and
can only be changed by the above settings.
Once the settings have been optimized, the rate data may be saved with the Store
button. When closing the graphical display window (History Viewer) a dialog box
will prompt to save the settings. Select Yes to store these settings for automatic
application to all new data.
Select (or type in) values for Zero and Gain to calibrate the measurement. If
calibration parameters are unknown for probes and transmitters, select 0 for Zero
and 31 for Gain.
The Zero Offset will allow users to change the y-axis reference point to zero
(0) by changing the value. This is a helpful feature when conducting tests to
graphically view the data from a convenient starting point. It is recommended that
this be returned to original value of zero so determination of probe life can be done
accurately.
49
50
The Corrosion Rate graph is painted black and the Pitting graph is painted blue.
Use the Dimension box (center right-hand corner of the screen) to select the
Corrosion Rate units in mpy, mmpy, or mpy. Pitting tendency does not have
units of measure but is interpreted in magnitude to the corrosion rate reading.
When the pitting tendency reading is greater than the corrosion rate and erratic in
value, this is indicative of pitting type corrosion. Under general corrosion conditions,
the pitting tendency shows values much lower than the corrosion rate value and
typically be near zero. For ease of interpretation the units of pitting tendency are
set to match the corrosion rate units. The graph above shows a pitting reading
about half the value of the corrosion rate reading indicating generally little pitting.
The graph below shows just the corrosion rate with pitting de-selected.
Galvanic Probes
When using Galvanic electrodes, only the pitting (or imbalance) graph is generated.
Imbalance units can be switched between Galvanic current in microamperes (A)
or in part per billion (ppb) of oxygen based on an approximate empirical calibration
of 4 A per ppb of oxygen. Typical Galvanic graph is shown below:
Zoom Commands
To select a zoom area:
1. Navigate to the graph, press and hold the left mouse button on the starting
location.
2. Move the mouse horizontally and release the button at the desired end location.
51
52
Use the UndoZoom command button to return to the previous state of the graph.
Note:
Smoothing is a process by which data points are averaged with their neighbors
in a series to blur the sharp edges to depict smoother data points. At the default of
zero, it plots every data point. The setting is the number of points used in a moving
average to smooth the Metal Loss data. The maximum setting number is 100.
Increase and decrease the number for more or less smoothing of graphs.
Note:
This setting does not affect the data used for the
Corrosion or Erosion Rate computation, since it is
computed from the raw data.
The Regression Interval [h] control allows selecting a time interval used to
calculate the Corrosion or Erosion Rate graph (linear regression algorithm).
The Corrosion/Erosion Rate will be in units corresponding to the Metal Loss units
that have been selected in Mils per Year (mpy), Millimeters Per Year (mmpy) or
Micrometers per Year (mpy), and will be calculated by linear regression on an
initial default-running-window of 24 hours (20 hours when the collection time is
set to 2 minute intervals). This regression period is displayed in hours and may be
re-adjusted and the Corrosion/Erosion Rate re-run with the Corrosion or Erosion
Rate command button. The interval may need adjustment to optimize noise versus
sensitivity of the rate graph.
The Corrosion/Erosion Rate graph is computed from the metal loss data by
determining the slope of the best straight line through the metal loss data. The best
straight line is computed statistically by the method known as linear regression. For
example if we have 7 days of metal loss data, we may typically want to calculate
the slope or corrosion/erosion rate based on a moving 1 days worth of data. In this
case, the first corrosion/erosion rate data point is produced after the first one days
worth of data. So for 7 days worth of metal loss data there is only 6 days worth of
corrosion/erosion rate data. That is, the corrosion/erosion rate data will always be
less than the metal loss data by the amount of the regression period.
The corrosion/erosion rate essentially represents the tangent to the metal loss
curve. So if we use 1 days worth of metal loss data for computation, then the
corrosion/erosion rate value is calculated from half the regression period before and
half the regression period after that time. Consequently with 7 days of metal loss
data, the 6 days of corrosion/erosion rate data starts time-wise with day blank at
the beginning and day blank at the end of the 7 day period. In fact, the Microcor
Tools graph will show a horizontal line on the graph at each of these locations
equal to the starting and ending corrosion/erosion rate values respectively.
53
54
Data Cursor
The Data Cursor function provides information for an individual data point. The
Enable command button marks the data points on the graph with dots. Mouse
clicking on the data point will refresh the Metal Loss, Temperature/Corrosion
Rate (depending on the graph selected) and Time readings on the gray bar located
at the bottom section of the graph. To disable the Data Cursor function, click the
Disable command button.
55
56
The Corrosion/Erosion Rate graph is computed from the metal loss data by
determining the slope of the best straight line through the metal loss data. The
best straight line is computed statistically by the method known as linear regression
(this is further explained above in the Generating a Corrosion or Erosion Rate
Graph section.
With on-line systems future data to use in computations is unavailable. So latest
computed corrosion/erosion rate is really the value for half the regression period
prior to the current time. In the absence of the future data it is assumed that
this is also the best estimate of the current corrosion/erosion rate, and hence the
horizontal line on the graph over this period (at the beginning and end of the graph).
When Live Update is activated, the rate graph is purposely shifted forward by half
the regression period so that the latest computed corrosion rate value is shown
at the present time. This makes it easier to view the metal loss and the best
computed corrosion/erosion rate for the current time when zoomed-in. In the case
of 7 days worth of metal loss data, this means that the 6 days of corrosion rate data
will now be positioned with a blank period of 1 day at the start of the 7 days and no
blank period at the end of the 7 days.
Note:
Use the Notepad Print or Save As commands to print or store the report on
the hard drive.
57
58
59
CHAPTER 6
The Data Display tab is the most convenient global display for on-line systems. It
shows current readings, alarm status codes and which addresses are data loggers.
Each point on the Data Display maybe double-clicked to show the graphical display
(History Viewer). This screen allows for monitoring and troubleshooting of RS485
bus and transmitters. Communication errors and transmitter status codes are listed
in the Transmitter Status column while the column named Parameter 3 will turn
red if a unit is in alarm status. It will turn yellow when the alarm is acknowledged by
the user; yellow means there is an ongoing alarm and it is acknowledged. When
the alarm is turned off or when the corrosion rate resets to below the alarm rate,
the column will return to normal. If a unit enters an alarm status and resets prior to
user acknowledgement, red indicator will change color to blinking green and gray
(on the bottom status bar of the screen) and will log this on the Event Log.
60
Event Log
61
CHAPTER 7
The Event Log tab is intended to collect and store different types of events
in the system and to present them in a table format. The events are stored
in a 500 element buffer. The Save button permits export of the event log
information in to a text-format file. If a unit enters an alarm status and resets
prior to user acknowledgement, red indicator will change color to blinking
green and gray (on the bottom status bar of the screen) and will log this
on the Event Log; hence it is highly recommended that the Event Log is
inspected regularly.
62
EVENT LOG
63
CHAPTER 8
Collected data may be archived when the data files become large enough to
slow down the graphing function, typically over 65,000 points. Data may also be
retrieved from archives into the program, and this function may also be used to
import data files such as those from the previous versions of the Microcor software
or from other Microcor Tools software users.
Archiving Data
When moving data to the archives, this data is always stored in the Archives
folder. The path is as follows; C:\Program Files\Microcor Tools\data\Site Name\
archives.
To archive a data file, select the Address to be archived and press the Move to
Archives command button. Each time a file is archived, the date is added to the
file name to prevent overwriting of any earlier dated file. This protection function
does not operate if the data has the same date, therefore, archiving is permitted
only once per day.
64
Archive files have automatically assigned names. For example, the archive file of
location Address 15, created on 13 May, 2003 is ch15_13May03.dat. The same
address archived the following day would be ch15_14May03.dat.
Note:
Auto Archiving
The Auto Archiving feature in the main screen is designed for computers with
limited storage resources. If the Auto Archiving is selected, when a site collects
2MB of data (approximately 48,620 readings or 67 days of continuous data collection
(temperature data collection enabled) at a sampling interval of 2 minutes), the data
file is automatically transferred into the archives folders replacing previously stored
data file. Naming convention of these archive files are different compared to use
instigated archive file. For example, the archive file of location Address 15 is ch15.
dat.
Note: When Auto Archiving is enabled, Microcor Tools
initiates automatic archiving when the file reaches its
limit, previous automatically stored archives will be
deleted. This process will not delete any manually
archived files.
Appendix A
Alloy Multipliers
These factors are recommended for use with the E9020 instrument. They are
based upon use of CORRATER electrodes which have surface areas of 5cm2 for
standard probes and 0.5 cm2 for flush probes.
UNS Code
K03005
A91100
A92024
C11000
C44300
C44500
C64200
C68700
C70610
C71500
G41300
L50045
N04400
N05500
N06022
N06600
N08020
N08800
N08825
N10276
R50400
S30400
S31600
S31603
S31803
S32750
Z17001
65
Material
Pipe Grade Carbon Steel
Aluminum 1100-0
Aluminum 2024
Copper 110 ETP Comm. Pure
CDA 443 Arsenical Admiralty
CDA 445 Phosphorized Adm.
CDA 642 A1 Silicon Bronze
CDA 687 Alum. Brass Arsenical
CDA 706 90/10 Copper/Nickel
CDA 715 70/30 Copper/Nickel
AISI 4130 Alloy Steel
Lead
Monel 400 Nickel
Monel K-500 Nickel
Hastelloy C22
Inconel 600 Nickel
Carpenter 20 CB3 SST
Incolloy 800
Incolloy 825
Hastelloy C276
ASTM B-348 Grades 2-4 Titanium
AISI 304 Stainless Steel
AISI 316 Stainless Steel
AISI 316L Stainless Steel
2205 Duplex Stainless Steel
2507 Duplex Stainless Steel
Grades 1A, 1, 2, 3, or 5 Zinc
Multiplier
1.00
0.94
0.88
2.00
1.67
1.68
1.48
1.62
1.80
1.50
1.00
2.57
1.13
1.04
0.85
0.95
0.98
0.89
0.88
0.86
0.75
0.89
0.90
0.90
0.89
0.88
1.29
66
ALLOY MULTIPLIERS
67
Appendix B
Modbus Configuration
Trx
Comm
Port
Tag
Number
Range
1
Metal
Loss
(mils)
Temp
Change
(F)
Calc
Rate
(mpy
Calc
Rate w/
24h filter
(mpy)
Status
Modbus
Address
Modbus
Address
Modbus
Address
Modbus
Address
Modbus
Address
40,001
40,003
40,005
40,007
40,009
40,011
40,013
40,015
40,017
40,021
40,023
40,025
40,027
40,031
40,033
40,035
40,041
40,043
40,051
40,053
40,061
40,071
Corrosion
Rate /
Potential
(mpy)
Reserved
Reserved
Modbus
Address
Modbus
Address
Modbus
Address
Modbus
Address
Modbus
Address
40,001
40,003
40,005
40,007
40,009
40,019
40,011
40,013
40,015
40,017
40,019
40,029
40,021
40,023
40,025
40,027
40,029
40,037
40,039
40,031
40,033
40,035
40,037
40,039
40,045
40,047
40,049
40,041
40,043
40,045
40,047
40,049
40,055
40,057
40,059
40,051
40,053
40,055
40,057
40,059
40,063
40,065
40,067
40,069
40,061
40,063
40,065
40,067
40,069
40,073
40,075
40,077
40,079
40,071
40,073
40,075
40,077
40,079
40,081
40,083
40,085
40,087
40,089
40,081
40,083
40,085
40,087
40,089
40,091
40,093
40,095
40,097
40,099
40,091
40,093
40,095
40,097
40,099
10
40,101
40,103
40,105
40,107
40,109
10
40,101
40,103
40,105
40,107
40,109
11
40,111
40,113
40,115
40,117
40,119
11
40,111
40,113
40,115
40,117
40,119
12
40,121
40,123
40,125
40,127
40,129
12
40,121
40,123
40,125
40,127
40,129
13
40,131
40,133
40,135
40,137
40,139
13
40,131
40,133
40,135
40,137
40,139
14
40,141
40,143
40,145
40,147
40,149
14
40,141
40,143
40,145
40,147
40,149
15
40,151
40,153
40,155
40,157
40,159
15
40,151
40,153
40,155
40,157
40,159
16
40,161
40,163
40,165
40,167
40,169
16
40,161
40,163
40,165
40,167
40,169
17
40,171
40,173
40,175
40,177
40,179
17
40,171
40,173
40,175
40,177
40,179
18
40,181
40,183
40,185
40,187
40,189
18
40,181
40,183
40,185
40,187
40,189
19
40,191
40,193
40,195
40,197
40,199
19
40,191
40,193
40,195
40,197
40,199
20
40,201
40,203
40,205
40,207
40,209
20
40,201
40,203
40,205
40,207
40,209
21
40,211
40,213
40,215
40,217
40,219
21
40,211
40,213
40,215
40,217
40,219
22
40,221
40,223
40,225
40,227
40,229
22
40,221
40,223
40,225
40,227
40,229
23
40,231
40,233
40,235
40,237
40,239
23
40,231
40,233
40,235
40,237
40,239
24
40,241
40,243
40,245
40,247
40,249
24
40,241
40,243
40,245
40,247
40,249
25
40,251
40,253
40,255
40,257
40,259
25
40,251
40,253
40,255
40,257
40,259
26
40,261
40,263
40,265
40,267
40,269
26
40,261
40,263
40,265
40,267
40,269
27
40,271
40,273
40,275
40,277
40,279
27
40,271
40,273
40,275
40,277
40,279
Range
2
Trx
Address
(I/N)
Status
68
MODBUS CONFIGURATION
28
40,281
40,283
40,285
40,287
40,289
28
40,281
40,283
40,285
40,287
40,289
29
40,291
40,293
40,295
40,297
40,299
29
40,291
40,293
40,295
40,297
40,299
30
40,301
40,303
40,305
40,307
40,309
30
40,301
40,303
40,305
40,307
40,309
31
40,311
40,313
40,315
40,317
40,319
31
40,311
40,313
40,315
40,317
40,319
32
40,321
40,323
40,325
40,327
40,329
32
40,321
40,323
40,325
40,327
40,329
33
40,331
40,333
40,335
40,337
40,339
33
40,331
40,333
40,335
40,337
40,339
34
40,341
40,343
40,345
40,347
40,349
34
40,341
40,343
40,345
40,347
40,349
35
40,351
40,353
40,355
40,357
40,359
35
40,351
40,353
40,355
40,357
40,359
36
40,361
40,363
40,365
40,367
40,369
36
40,361
40,363
40,365
40,367
40,369
37
40,371
40,373
40,375
40,377
40,379
37
40,371
40,373
40,375
40,377
40,379
38
40,381
40,383
40,385
40,387
40,389
38
40,381
40,383
40,385
40,387
40,389
39
40,391
40,393
40,395
40,397
40,399
39
40,391
40,393
40,395
40,397
40,399
40
40,401
40,403
40,405
40,407
40,409
40
40,401
40,403
40,405
40,407
40,409
41
40,411
40,413
40,415
40,417
40,419
41
40,411
40,413
40,415
40,417
40,419
42
40,421
40,423
40,425
40,427
40,429
42
40,421
40,423
40,425
40,427
40,429
43
40,431
40,433
40,435
40,437
40,439
43
40,431
40,433
40,435
40,437
40,439
44
40,441
40,443
40,445
40,447
40,449
44
40,441
40,443
40,445
40,447
40,449
45
40,451
40,453
40,455
40,457
40,459
45
40,451
40,453
40,455
40,457
40,459
46
40,461
40,463
40,465
40,467
40,469
46
40,461
40,463
40,465
40,467
40,469
47
40,471
40,473
40,475
40,477
40,479
47
40,471
40,473
40,475
40,477
40,479
48
40,481
40,483
40,485
40,487
40,489
48
40,481
40,483
40,485
40,487
40,489
49
40,491
40,493
40,495
40,497
40,499
49
40,491
40,493
40,495
40,497
40,499
50
40,501
40,503
40,505
40,507
40,509
50
40,501
40,503
40,505
40,507
40,509
51
40,511
40,513
40,515
40,517
40,519
51
40,511
40,513
40,515
40,517
40,519
52
40,521
40,523
40,525
40,527
40,529
52
40,521
40,523
40,525
40,527
40,529
53
40,531
40,533
40,535
40,537
40,539
53
40,531
40,533
40,535
40,537
40,539
54
40,541
40,543
40,545
40,547
40,549
54
40,541
40,543
40,545
40,547
40,549
55
40,551
40,553
40,555
40,557
40,559
55
40,551
40,553
40,555
40,557
40,559
56
40,561
40,563
40,565
40,567
40,569
56
40,561
40,563
40,565
40,567
40,569
57
40,571
40,573
40,575
40,577
40,579
57
40,571
40,573
40,575
40,577
40,579
58
40,581
40,583
40,585
40,587
40,589
58
40,581
40,583
40,585
40,587
40,589
59
40,591
40,593
40,595
40,597
40,599
59
40,591
40,593
40,595
40,597
40,599
60
40,601
40,603
40,605
40,607
40,609
60
40,601
40,603
40,605
40,607
40,609
61
40,611
40,613
40,615
40,617
40,619
61
40,611
40,613
40,615
40,617
40,619
62
40,621
40,623
40,625
40,627
40,629
62
40,621
40,623
40,625
40,627
40,629
63
40,631
40,633
40,635
40,637
40,639
63
40,631
40,633
40,635
40,637
40,639
64
40,641
40,643
40,645
40,647
40,649
64
40,641
40,643
40,645
40,647
40,649
65
40,651
40,653
40,655
40,657
40,659
65
40,651
40,653
40,655
40,657
40,659
66
40,661
40,663
40,665
40,667
40,669
66
40,661
40,663
40,665
40,667
40,669
67
40,671
40,673
40,675
40,677
40,679
67
40,671
40,673
40,675
40,677
40,679
68
40,681
40,683
40,685
40,687
40,689
68
40,681
40,683
40,685
40,687
40,689
69
40,691
40,693
40,695
40,697
40,699
69
40,691
40,693
40,695
40,697
40,699
70
40,701
40,703
40,705
40,707
40,709
70
40,701
40,703
40,705
40,707
40,709
71
40,711
40,713
40,715
40,717
40,719
71
40,711
40,713
40,715
40,717
40,719
72
40,721
40,723
40,725
40,727
40,729
72
40,721
40,723
40,725
40,727
40,729
73
40,731
40,733
40,735
40,737
40,739
73
40,731
40,733
40,735
40,737
40,739
74
40,741
40,743
40,745
40,747
40,749
74
40,741
40,743
40,745
40,747
40,749
75
40,751
40,753
40,755
40,757
40,759
75
40,751
40,753
40,755
40,757
40,759
76
40,761
40,763
40,765
40,767
40,769
76
40,761
40,763
40,765
40,767
40,769
77
40,771
40,773
40,775
40,777
40,779
77
40,771
40,773
40,775
40,777
40,779
78
40,781
40,783
40,785
40,787
40,789
78
40,781
40,783
40,785
40,787
40,789
79
40,791
40,793
40,795
40,797
40,799
79
40,791
40,793
40,795
40,797
40,799
80
40,801
40,803
40,805
40,807
40,809
80
40,801
40,803
40,805
40,807
40,809
81
40,811
40,813
40,815
40,817
40,819
81
40,811
40,813
40,815
40,817
40,819
82
40,821
40,823
40,825
40,827
40,829
82
40,821
40,823
40,825
40,827
40,829
83
40,831
40,833
40,835
40,837
40,839
83
40,831
40,833
40,835
40,837
40,839
84
40,841
40,843
40,845
40,847
40,849
84
40,841
40,843
40,845
40,847
40,849
85
40,851
40,853
40,855
40,857
40,859
85
40,851
40,853
40,855
40,857
40,859
86
40,861
40,863
40,865
40,867
40,869
86
40,861
40,863
40,865
40,867
40,869
87
40,871
40,873
40,875
40,877
40,879
87
40,871
40,873
40,875
40,877
40,879
88
40,881
40,883
40,885
40,887
40,889
88
40,881
40,883
40,885
40,887
40,889
89
40,891
40,893
40,895
40,897
40,899
89
40,891
40,893
40,895
40,897
40,899
90
40,901
40,903
40,905
40,907
40,909
90
40,901
40,903
40,905
40,907
40,909
91
40,911
40,913
40,915
40,917
40,919
91
40,911
40,913
40,915
40,917
40,919
92
40,921
40,923
40,925
40,927
40,929
92
40,921
40,923
40,925
40,927
40,929
93
40,931
40,933
40,935
40,937
40,939
93
40,931
40,933
40,935
40,937
40,939
94
40,941
40,943
40,945
40,947
40,949
94
40,941
40,943
40,945
40,947
40,949
95
40,951
40,953
40,955
40,957
40,959
95
40,951
40,953
40,955
40,957
40,959
96
40,961
40,963
40,965
40,967
40,969
96
40,961
40,963
40,965
40,967
40,969
97
40,971
40,973
40,975
40,977
40,979
97
40,971
40,973
40,975
40,977
40,979
98
40,981
40,983
40,985
40,987
40,989
98
40,981
40,983
40,985
40,987
40,989
99
40,991
40,993
40,995
40,997
40,999
99
40,991
40,993
40,995
40,997
40,999
100
41,001
41,003
41,005
41,007
41,009
100
41,001
41,003
41,005
41,007
41,009
Protocol
Hardware output is RS-485 2 wire: Software output is settable ASCII or
RTU, Default setting is 19200 Baud, 8 data bits, 1 stop bits, no parity
Scaling
All outputs are scaled in engineering units as configured in the Microcor
Tools with Output Interfaces software running on the IIU
Error Codes from Microcor transmitters will produce large negative numbers in the applicable Modbus registers approx 3.8 times the probe span in
engineering units set i.e for Microcor 5 mil span probe error codes typically
around - 19.07
Error Codes from E-9020 Corrater / Galvanic Transmitter will also produce large negative numbers in the applicable Modbus registers of approx.
-762,935 on mpy units
69
70
MODBUS CONFIGURATION
Metal Loss Status (1=OK) Bit 0=1 if Bit 8=0, and Bit 9=0, and Bit 10=0
Temperature Status (1=OK)
Corrosion Rate Linear Regression (1=OK)
Corrosion Rate with 24 hr filter (1=OK)
Not Used
Not Used
Not Used
Not Used
Transmitter to IIU communication error (1=Error)
Transmitter in Startup (1=StartUp)
Probe or Probe Connection Failure (1=Failure)
Not Used
Not Used
Not Used
Not Used
Not Used
71
72
MODBUS CONFIGURATION
Appendix C
OPC Configuration
OPC refers to OLE for Process Control protocol and is based on Microsofts OLE/
COM technology which is used for data transfer between databases.
When Microcor Tools Output Interfaces is used on the RCS Intelligent Interface
Unit (IIU), the Microcor Tools acts as the OPC Server to provide data transfer
to the DCS, PCS or SCADA system running the OPC Client. While the principle
of operation is simple, configuration can be complex as OPC is integrated into
Microsoft security.
73
Network
Domain
Domain
Domain
Domain
Client OS
XP
XP
2000
2000
OPC Client
User Name
rcsuser
user2
rcsuser
user2
Network
Domain
Domain
Domain
Domain
74
OPC CONFIGURATION
2000
XP
XP
2000
user2
rcsuser
rcsuser
rcsuser
Workgroup
Workgroup
Workgroup
Workgroup
2000
XP
2000
XP
user2
rcsuser
rcsuser
rcsuser
XP
rcsuser
D LL*
XP
rcsuser
XP
rcsuser
Workgroup
2000
rcsuser
XP
rcsuser
Workgroup
XP
rcsuser
* PC is configured on the Domain but logged onto a local profile
Workgroup
Workgroup
Workgroup
Workgroup
D LL*
D LL*
D LL*
Determine whether the client computer which will connect to the IIU is set up as a
part of a domain or a workgroup network. To do this, navigate to Start Control
Panel (Classic View) System and double click the Computer Name tab of the
intended client computer (shown below):
Note:
The above image shows that this computer (with the name
028RCS) is in a Workgroup named WORKGROUP. If the
system is in a domain, the word Workgroup will be replaced
with Domain followed by the domain name.
Name tab. Click on the Change button to advance to the Computer Name
Changes screen (see image below).
On the Member of area, change from Domain/Workgroup as desired. When OK is
clicked, it will prompt for a Username and Password of an administrator. If going to
a workgroup, use a local computer administrator username and password, if going
to a domain, have the IT Administrator enter the domain Administrator username
and password.
75
76
OPC CONFIGURATION
This will open a New User dialogue box, Enter rcsuser as the User name and rcs
as the Password and confirm the password.
77
78
OPC CONFIGURATION
Caution:
The following configures the username rcsuser and password rcs to automatically
logon when the IIU is rebooted, substitute the proper username and password
where applicable. Use the Registry Editor to add the auto logon information by
following the steps below.
1. Click Start Run, type regedit and click OK.
2. Locate the following registry key:
HKEY_LOCAL_MACHINE\SOFTWARE\Microsoft\WindowsNT\CurrentVersion\Winlogon
Double-click the DefaultUserName entry,
1. Type rcsuser, and then click OK.
Double-click the DefaultPassword entry,
1. Type rcs under the value data box, and then click OK.
A. If there is no DefaultPassword value, create the value by following
the steps below (If no DefaultPassword string is specified, Windows
XP will automatically change the value of the AutoAdminLogon registry
key from 1 (true) to 0 (false) to turn off the AutoAdminLogon feature):
a. In Registry Editor, click Edit, click New, and then click String Value.
b. Type DefaultPassword as the value name of the New Value created, and then press ENTER.
c. Double-click the newly created key, and then type the password
(rcs) in the Value Data box and click OK.
Double-click the AutoAdminLogon entry, type 1 in the Value Data box,
and then click OK.
1. If there is no AutoAdminLogon entry, create the entry by performing
the following steps:
a. In Registry Editor, click Edit, click New, and then click String
Value.
b. Type AutoAdminLogon as the value name, and then press
ENTER.
c. Double-click the newly created key, and then type 1 in the
Value Data box.
3. Exit the Registry Editor.
4. Click Start Shutdown Restart, and then click OK. If the settings were
changed successfully, the computer will automatically logon to the rcsuser profile.
When using Microcor Tools for on-line monitoring, it should be setup in the
Microsoft Windows Startup folder and the Autostart checkbox should be checked.
This ensures that the program will restart automatically and commence the
online communication and data collection when the computer is restarted after
maintenance or power failure.
To add Microcor Tools to the Windows Startup folder, copy the Microcor Tools
shortcut icon form the desktop and paste it in the C:\Documents and Settings\All
Users\Start Menu\Programs\Startup directory.
192.168.3.1
255.255.255.0
192.168.3.254
This configuration enables the IIU to be plugged directly into a network and obtain
an IP Address from the DHCP Server or use the Alternate Configuration if a DHCP
Server is not found or unavailable (i.e. if a cross-over cable is used). In order for
the OPC Client computer to connect when a DHCP Server is unavailable, the
Alternate Configuration on the client computer should be setup with an incremented
IP Address such as 192.168.3.5 (IP Address range must be within 192.168.3.2 to
192.168.3.253, however, if the IIU contains more than one network interface card,
successive IP Addresses may be allocated, hence it is recommended that the
client IP Addressing begins at 192.168.3.5). Consult the IT Department for more
details.
The typical navigational path and procedure to make the TCP/IP changes is as
follows;
1.
2.
3.
4.
79
80
OPC CONFIGURATION
Note:
Note:
81
82
OPC CONFIGURATION
Setup Windows Folder Properties: Disable Use Simple Sharing IIU Default
1.
2.
3.
4.
Double click My Computer from the desktop or right-click Start and select Explorer.
Click Tools Folder Options and select View tab (shown below).
Uncheck the Use simple file sharing (Recommended).
Click Apply followed by OK.
Install Microcor Tools Software and OPC Server Application IIU Default
Install the Microcor Tools Software from the installation CD. It is important to retain
all default options when installing. Once the installation is complete, follow the
steps below to register the OPC Server.
Click Start Run, type cmd and click OK.
Type cd C:\
Type cd Program Files\MicrocorTools
Type MTXOPCServer.exe regserver
If error messages appear, the registration is unsuccessful; make sure all steps
are performed as explained above and that the user has sufficient privileges on
the computer. Otherwise Type cd C:\
6. Type cd Windows\System32
7. Type OPCEnum/regserver
8. If error messages appear, the registration is unsuccessful; make sure all steps
are performed as explained above and that the user has sufficient privileges
on the computer. Otherwise Type Exit to leave the Command Prompt window
(see image below).
1.
2.
3.
4.
5.
9. Install OPC Client Test Utility (Follow all instructions carefully when installing
this application. Failure to follow instructions may result in serious damage to
the operating system. Consult RCS for more information).
83
84
OPC CONFIGURATION
7. Select only the Run application on this computer on the Location tab and
click Apply as shown below.
9. Click the Edit button under the Launch and Activation Permissions and
ensure that rcsuser (or the appropriate username) is added.
10. Verify that the permissions on all the Groups or Users are setup as shown
below (note that the Remote Launch is disabled on all Groups or Users). Click
OK.
11. Click the Edit button under the Access Permissions and ensure that rcsuser
(or the appropriate username) is added and all Groups and Users have full
permissions allowed as shown below. Click OK.
85
86
OPC CONFIGURATION
12. Click the Edit button under the Configuration Permissions and ensure that
rcsuser (or the appropriate username) is added and all Groups and Users
have full permissions allowed as shown below. Click OK.
13. Click Apply on the Security tab before moving on to the Endpoints tab (no
changes made on the Endpoints tab, shown below).
14. Select the The Interactive User on the Identity tab and click Apply.
The OPC Server is now ready to be launched for the Workstation Network setup.
87
88
OPC CONFIGURATION
4. On the My Computer Properties, click the Com Security tab and click the
Edit Limits button under the Access Permissions area. Image shown below:
5. Make sure Everyone is added under the Group or User Names and both
Local Access and Remote Access are set to Allow for both Everyone and
ANONYMOUS LOGON.
89
90
OPC CONFIGURATION
4. When the Enabled selection is made, OPC icon will appear on the Windows
Status bar as shown below:
5. Double-click on the OPC icon show above to open the OPC Server status
screen. Users may monitor the connections from this screen. The values for
statistics will dynamically change as clients log on and off of the OPC Server.
91
92
OPC CONFIGURATION
Note:
6. The OPC Server is now set to transfer data. Make sure to Start the Online
Communication and Data Collection (shown below); the OPC Client
computer will now receive the data being transferred (make sure to use the
same credentials on the OPC Client computer as used on the OPC Server
computer for Workgroup networks).
Using the Preinstalled OPC Client Test Utility on RCS IIU to Test the OPC Server
If the IIU is configured and shipped by RCS, an OPC Client Test Utility application
will also be installed just to verify the functionality of the OPC Server. This OPC
Client Test Utility can be run by following the steps below:
1. Run the OPC Client Test Utility application from Start All Programs
VB6Client
2. Normally, in order for a client to see a server, the OPC Server PC name or the
IP address must be typed in; however in this case since both the server and the
client reside on the same machine, click on the drop-down arrow of the OPC
Servers: to automatically browse and find the OPC Server. When the Server is
found, click on it to make the selection and then click on the Connect button.
3. Click on the Add Group button of the OPC Groups area. This will browse,
find and enter the group to the text box.
4. Click on the Browse button of the Namespace area. This will browse, find
and enter the Namespace group on the text area. Expand the selection by
clicking on the plus (+) sign to show the detailed listing. Click the plus (+) sign
as required to see the individual selections. Right-click on each selection and
left-click to select and view the data.
93
94
OPC CONFIGURATION
Troubleshooting
1. OPC Server is started and the OPC Client Test Utility on the same computer
can connect, but the Remote OPC Client computer cannot connect.
Disable the Windows Firewall temporarily on both computers until the
connection is established. See Appendix 4 for details on setting the
Windows Firewall.
Ensure the IP Addresses are correct and the connection cables are secure
on the IIU and the OPC Client computer.
2. OPC Server is started and the Client is connected but Client shows 0 for
transferred values.
Ensure that the Start buttons are activated on the Online Communication
and Data Collection as shown below on the on the Site/Communications
screen.
Check the OPC Server status by clicking on the OPC icon on the computer
status bar. If the OPC Server does not show values (Unit, ML, TC, CR-LR,
CR24 and Stat) on the application status bar then check the Interfaces
Configuration of the Online Communication area to ensure the OPC
interface has been Enabled (see the following images).
If the Status bar shows no activity as shown above then the OPC Server is not
transferring data. Enable the OPC Interface and Save Configuration as shown below:
95
96
OPC CONFIGURATION
3. OPC Client computer does not receive data after the IIU reboot.
Ensure the OPC Server is started and running.
Ensure the Autostart is checked on the Online Communication area,
shown below:
Appendix D
97
98
7. Click the Add Port button under the Exceptions tab. Fill in the information as
follows on the Add a Port dialogue box;
Name:
Port number:
Protocol:
DCOM
135
TCP
99