SYSTEM FEATURES
High Image Quality
With efficient dose utilization and 17
line pair per centimeter detectors, the
CereTom delivers low contrast resolution of 3 mm at 0.3% and high contrast resolution of 0.35 mm. Sharp 3D
surface and volume renderings along
with MPR images are generated from
high definition pixel data. Using CT
Angiography of the head, a 25 cm distance can be scanned in 30 seconds
with a slice width of 1.25 mm.
The streamlined design of the CereToms internal components combined with its state of the art modular
assembly makes diagnosis and repair
fast and uncomplicated, greatly reducing down time.
Wireless Communications
DICOM images are wirelessly transferred to and from the Clarus workstation using 802.11g secure transfer
protocols.
Patient Safety
areas such as ICU rooms. It is ruggedized to easily pass over freight elevator thresholds.
Ease of Service
Low Power
Product Data
1-NL3000-065 rev 00
D-3
Product Data
1-NL3000-065 rev 00
Rotation time
Gantry speed:
Typical scan range:
Scan time:
Scan cycle time:
Slice thickness:
Number of slices per scan
Z scan widths at ISO-center:
Image reconstruction time:
Rotation time:
Gantry speed
Typical Scan Range
Typical scan time:
Scan start time delay range:
Helical scan pitch:
Image reconstruction time:
Slice Thickness
2s
30 rpm
25 cm
2s
3s
2.5 mm
4
10 mm
1s/image
1 cm
30 - 45 s
10.0 mm
3 Images
1s
60 rpm
25 cm
30 s
0 to 100s
1
1s/image
1.25 mm
Geometry
Patient Opening
Image Field of View
32 cm
253 mm
X-ray Detection
Detection system:
Main detectors:
Data acquisition:
Reference detectors:
Viewing rate:
Solid-state detectors
408 x 8 rows
408 x 8, 408 x 4, 408 x 1
1
1440 views/s
X-ray Generation
X-ray beam shape:
X-ray tube voltage:
X-ray tube current:
X-ray tube cooling time:
Focal spot size:
X-Ray Tube Type
Cone Beam
100, 120, 140 kV
7.5 and 3.75 mA
12 minutes max
1 mm x 1 mm
Fixed Anode
Data Processing
Reconstruction matrix:
Picture element (pixel) size:
Reconstruction filter functions:
Smooth
Standard
Sharp
512 x 512
.49 mm
Image Quality
Noise STD less than
Low contrast detectability
Measurement Basis: 140kv, 15 mAs, 10mm slice thickness, 8 inch CATPHAN phantom, 53mGy (5.3Rad). Dose
is measured in the center of the phantom using a pencil
probe with a 10cm chamber length (+15% tolerance)
High Contrast Spatial Resolution
Standard Kernel (Hanning Window BW=0.5)
0.65mm limiting resolution
4.8 lp/cm @ 50% MTF
7.4 lp/cm @ 10% MTF
8.2 lp/cm @ 0% MTF
High Resolution Kernel (Hanning Window BW=1.0)
0.35mm limiting resolution
6.6 lp/cm @ 50% MTF
12.2 lp/cm @ 10% MTF
16.5 lp/cm @ 0% MTF
Data processor
0.3%
3mm at 0.3%
64-bit
2 Gb
Gantry
Mass
Workstation
See Outline
< 750 lb (340kg)
7.1 lb
Product Data
1-NL3000-065 rev 00
CTDI Dose
Scan techniques for all CTDI dose specifications are at
140KV, 15 mAs, 2 second scan time, 10 mm aperture.
Dose is demonstrated with CTDI methodology in compliance with Federal Regulation 21 CFR 1020.33 . Measurements were only made with the head phantom due to
the small CereTom aperture.
CTDI expressed in mGy (Rad):
Center: 53 (5.3)
Surface: 68 (6.8)
CTDI100 expressed in mGy (Rad):
Center: 52 (5.2)
Surface: 66 (6.6)
Image Processing
ROI
Measurement of distance between two points
Mark display (scale)
Zoom
Addition/Subtraction between images
Comment and arrow insertion
Flipping
Screen save
Multi view
Connectivity
17 inch LCD
1920 x 1200
-1,000 to 3,096
Variable
12
Site Requirements
Operating temperature
Storage temperature
Operating Altitude
Operating Humidity
Floor Flatness
X-Ray Attenuation
Material
Image Display
Display monitor:
Monitor matrix:
CT number
Window:
Preset windows:
15 Deg C to 35 Deg C
-25 Deg C to 70 Deg C
0-3010m (0-10,000 ft)
20 to 80% non condensing
< 0.120 inch (3 mm)
Wireless
802.11g
USB Memory Stick 512 MB (8 studies)
DICOM 3.0
DICOM basic print
DICOM Storage SCU for CT (archive)
DICOM Media Storage Class
Power Requirements
Input
Phase
Voltage
Frequency
Battery
Standby
Scan
D-5
Single
90-264VAC/1300 watts peak
47-63Hz
2 hours
4 Scans in 1 hour
Product Data
1-NL3000-065 rev 00
60.3 (1530.7)
38.2 (970.3)
52.7 (1337.5)
Weight <800lbs
D-6
28.7 (728.7)
Product Data
1-NL3000-065 rev 00
D-7
Product Data
1-NL3000-065 rev 00
The CereTom closely resembles a portable x-ray c-arm as it relates to radiation safety. The above readings were
obtained with the scanners covers removed! The addition of covers further reduces the scatter radiation readings.
Although the CereTom incorporates a high degree of protection against x-radiation, other than the useful beam,
no practical design of equipment can provide complete protection.
It is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of
the International Commission on Radiation Protection, and take adequate steps to protect against injury.
D-8